| | Class 2 Recall Mobile X - Ray Series |  |
| Date Posted |
November 16, 2012 |
| Recall Number |
Z-0197-2013 |
| Product |
GE Healthcare Optima XR 200amx with digital upgrade
Mobile general purpose radiographic imaging of the human head and body. |
| Code Information |
part number: 5421698
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha, Wisconsin 53188-1615 |
Reason for Recall |
It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
|
| Action |
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue.
Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard.
Further questions please call (262) 513-4122. |
| Quantity in Commerce |
364 total units installed in US |
| Distribution |
Nationwide Distribution |
| |
|
|
|
|