| | Class 2 Recall Philips Digital Diagnost R2.0.2 |  |
| Date Posted |
December 05, 2012 |
| Recall Number |
Z-0475-2013 |
| Product |
Philips Digital Diagnost Software Version R2.0.2
Product Usage:
This system is used for making X-ray exposures for diagnostics |
| Code Information |
All Digital Diagnost systems with release 2.0.2 and 2.0.2 SP1
Models 712020, 712022
|
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover, Massachusetts 01810 |
| For Additional Information Contact |
Beth St. Germain 978-659-4519
|
Reason for Recall |
With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.
|
| Action |
Philips Healthcare initiated a recall on October 25, 2012 via Urgent - Field Safety Notice for the Digital Diagnost X-ray System. The letter identified the affected product, problem and actions to be taken. The letter informs customers that Philips will supply a software update kit to version 2.1.4. For questions please contact your local Philips representative 1-800-722-9377. |
| Quantity in Commerce |
623 units |
| Distribution |
Worldwide Distribution - USA Nationwide and the countries of Argentian. Australia, Austria, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Eastonia, Finland, France, Germany, India, Indonesia, Iraq, Italy, Japan, Latvia, Lithuania, Lybia, Malta, Mexico, Netherlands, New Zealnd,Norway, Oman, Polnad, Protugual, Quatar, Romania, Russia, Saudi Arabia,Singapore, Slovakia, Soth Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE and UK. |
| |
|
|
|
|