Class 2 Device Recall Trilogy F/M Acetabular Shell without holes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Trilogy F/M Acetabular Shell without holes see related information
Date Posted February 04, 2013
Recall Status1 Terminated on June 20, 2014
Recall Number Z-0769-2013
Recall Event ID 63531
510(K)Number K934765 
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
Product Trilogy F/M Acetabular Shell without holes. Several ODs: 40MM, 42MM, 44MM, 46MM, 48MM, 50MM, 52MM, 54MM, 56MM, 58MM, 60MM, 62MM, 64MM, 66MM, 68MM, 70MM.

The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
Code Information 61216907
61157229
00369603
61452214
00370029
00369698
61176450
00369643
00370031
61454835
00369732
00369699
61216908
61673516
61642322
62065259
61906448
61263152
61454836
61476896
00369630
61775736
61943339
61403525
61642325
61732773
61263539
61642324
61906449
62022702
62076908
61885843
61309789
61928501
61237709
61292681
61407876
61825843
61452216
61817457
61237707
61237708
61271325
61327754
61452215
61452217
61489313
61519975
61536413
61564182
61598285
61642323
00370315
00369632
61744558
61885845
62079957
61732772
61928502
61783457
61825844
61869694
61906453
62010604
61846487
61993146
61642326
61779030
61783460
61906450
62052790
61536692
61487482
61487481
61598295
61632192
61309790
61511677
61455839
61263540
61367254
61321073
61237710
61281183
61455840
00369649
61976102
61788120
61885847
61928503
61801103
61840614
61947562
61996712
61813950
61899040
61911390
62001304
61864069
61536693
61473713
61417315
62076907
61487483
61489314
61281184
61237711
61237712
61642328
61260584
61306613
61329094
61403527
61455843
61598296
61642327
00369605
61642329
61813945
61980779
61947563
62076909
61928504
61955589
61292682
62006523
61801104
61813948
61864070
61885848
61906451
61476897
61612715
61642330
61470887
61417316
61473724
61567576
61256570
61455841
61215903
61237713
61281185
61309791
61397913
61536694
00369645
61728663
61775735
61801102
61864071
61996713
61292684
61642332
61891249
61526509
61611815
61976103
61760310
61768196
61951207
62043439
61235829
61237714
61271326
61429593
61500680
61526508
61611819
61817458
62079956
61260587
61452212
61321074
61403528
61417317
61469130
61500681
61801101
61825845
61911391
62065258
61350519
00369738
00369635
61548179
61928505
61532294
61548180
61642333
61803766
61983681
62036846
61237716
61309792
61403529
61412049
61889258
61256571
61424318
61455846
61476898
62006522
00369803
61921682
61873604
61500682
61318326
61987182
61350520
00369741
00369702
00369639
61452213
00369640
61724206
00369705
Recalling Firm/
Manufacturer
Zimmer, Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information Contact Jaime L. Weeks
574-372-4807
Manufacturer Reason
for Recall
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
FDA Determined
Cause 2
Process control
Action Zimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons.
Quantity in Commerce 275,952 units
Distribution Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = LPH and Original Applicant = ZIMMER, INC.
-
-