| Class 2 Device Recall Mindray | |
Date Initiated by Firm | September 24, 2012 |
Date Posted | December 17, 2012 |
Recall Status1 |
Terminated 3 on August 15, 2013 |
Recall Number | Z-0528-2013 |
Recall Event ID |
63652 |
510(K)Number | K102004 |
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product | Mindray V12 Size: 320 mm X 320 mm X 450 mm N.W.: 3 kg G.W.: 6 kg Qty:1 Manufactured in China
Mindray V21 Size: 515 mm X 335 mm X 685 mm N.W.: 8 kg G.W.: 11 kg Qty:1 Manufactured in China
One product, V Series monitor, available in two sizes: The V12 has a 12 inch screen, the V21 has a 21 inch screen.
The V Series Monitor is a Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices. It also monitors of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements, 2) Heart Rate, 3) Pulse Oximetry (Sp O2), 4) ST Segment Analysis, 5) Arrhythmia Detection, 6) Non Invasive Blood Pressure (NIBP), 7) Invasive Blood Pressure (IBP), 8) Cardiac Output (CO), 9) Respiratory Gasses, 10) Respiration Rate, 11) Temperature, It is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. |
Code Information |
V Series Monitor P/N's 0998-00-1800-101 and 0998-00-1800- 201. |
Recalling Firm/ Manufacturer |
Mindray DS USA, Inc. d.b.a. Mindray North America 800 MacArthur Blvd. Mahwah NJ 07430
|
For Additional Information Contact | Diana Arpino 201-995-8407 |
Manufacturer Reason for Recall | Mindray has identified an issue with the V Series Monitor where the monitors touch screen may stop responding to touch. |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | Mindray issued an "Urgent: V Series Monitoring System Corrective Action" letter dated 9/24/2012 via certified mail with return receipt verification to all hospitals.
The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their Mindray Service Representative at 1-800-288-2121 to arrange for relplacement of the affected product. For questions regarding this recall call 201-995-8407. |
Quantity in Commerce | 206 units (180 units domestic, 26 units foreign) |
Distribution | Worldwide Distribution - USA (nationwide) and internationally to Great Britian, Colombia, and Australia. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MHX
|
|
|
|