| | Class 2 Recall Restep DVT System |  |
| Date Posted |
February 20, 2013 |
| Recall Number |
Z-0840-2013 |
| Product |
STRYKER Sustainability Solutions, AC Power Adaptor for Restep Deep Vein Thrombosis (DVT) System, an external compression system. PUMPKIT (Pump Kit), and RSP159-01 (AC Power Adaptor). |
| Code Information |
108851407, 108851408, 108870192, 108874144, 108883271, 108886340, and 108888023.
|
Recalling Firm/ Manufacturer |
Stryker Sustainability Solutions 1810 W Drake Dr Tempe, Arizona 85283-4327 |
| Consumer Instructions |
No consumer action necessary |
| For Additional Information Contact |
Moira Barton-Varty 888-888-3433 Ext. 5519
|
Reason for Recall |
Reports that the AC Power Adaptor may not operate due to wires coming loose in the unit.
|
| Action |
Stryker Sustainability Solutions sent a Urgent Medical Device Removal Notification letter dated September 5, 2012, to all affected customers. The letter identified the product, problem and the action to be taken by the customer.
The firm notified that affected AC Power Adaptors would be replaced with improved Adaptors. Any affected lots would be removed and shipped back to Stryker Sustainability Solutions for final disposition.
Customers were instructed to fill out the attached Recall Effectiveness Form even if no affected product was found and fax to Stryker Sustainability Solutions at (480) 809-4957. |
| Quantity in Commerce |
1,311 |
| Distribution |
Distributed in Arizona and New Jersey. |
| |
|
|
|
|