• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Medical & Radiation Emitting Device Recalls

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Recall
SOL SYS 8IN CALC 1.5/15.0
see related information
Date Posted December 20, 2012
Recall Number Z-0575-2013
Product SOL SYS 8IN CALC 1.5/15.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Code Information Product Code: 157215150 and Lot codes: 111620, DD3BR1000, 126364, DD3BRA000, 134668A, DG1N91000, A37A71000, DN5H11000, A37A7A000, DN5H1B000, A37A7C, DT6ED1000, A3RBC1000, E1RHJ1000, A3RBCA000, E4HSE1000, A3YAV1000, E4LLG1000, A4VAK1000, E5EB41000, A6ACH1000, E5EB4A000, A88AA1000, E5WJ81000, AP2BB1000, E73B51, AV3DP1000, EC3DN1000, B2LAB1000, ED5KV1000, B3FAE1000, EJ4EG1000, B46AH1000, ER8EY1000, B54FC1000, EW6G31000, BS2GG1000, EY2ET1000, BY3BD1000, FB3E21, C1PC91000, FB8P81, C34B71000, FD4SX1, C3KA31000, FF8DT1, C4JCP1000, FG7DP1, C4VCP1000, FJ9L91, C5FH71000, X32EY1000, C5FH7A, X32EYA000, CC5JP1000, X32EYB000, CE6FL1000, X49FD1000, CH7EF1000, X49FDA000, CL5PE1000, X4ABK1000, CL5PE1A, X4ABKA000, CV5BS1000, X4ABKB000, CV5BSA000, X4GD91000, CX8CV1000, YA6GR1000, CX8CVA000, YF5DJ1000, CY7B11000, YH2C21000, D1TLF1000, YH2C2A000, D32J51000, Z41AS1000, D32J5A000, Z41ASA000, D37GC1000, Z5TH71000, D46LF1000, Z6BB81000, D52LX1000, Z6BB8A000, D81NN1000, Z83A11000, DA2EM1000, ZN7HS1000, DA2EMA000, ZN7HSA000, DA2EMB000, ZN7HSB000, and DA8MD1000.
Recalling Firm/
Manufacturer
DePuy Orthopaedics, Inc.
700 Orthopaedic Dr
Warsaw, Indiana 46582-3994
Reason for
Recall
An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a
Action Deputh Orthopaedics sent an Urgent Information - Medical Device Recall Notice dated December , 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The notification letter instructed consignees to please cease further distribution or use of products immediately. The purpose of the communication was to inform them of this recall and request acknowledgement of receipt of this letter by their facility. Reconciliation forms should be returned to your DePuy Sales Representative or faxed to the following Fax Number: 574-371-4939. For questions regarding this recall call 574-267-8143.
Quantity in Commerce 1732 total all product codes
Distribution Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR
 
-
-