Class 2 Device Recall Plum A Single Channel Infusion Pumps with MedNet Software

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards | Inspections
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Recall
Plum A Single Channel Infusion Pumps with MedNet Software
see related information
Date Posted December 07, 2012
Recall Status1 Open
Recall Number Z-0483-2013
Recall Event ID 63518
Premarket Notification
510(K) Number
K042081 
Product Classification Pump, Infusion - Product Code FRN
Product Plum A+ Single Channel Infusion Pumps with Hospira Mednet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local area networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list #20679 and 20792, module list number 20677. The Plum A+ Infusion System with Hospira MedNet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local area networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature.
Code Information pump list numbers 20679 and 20792, and module list number 20677; all serial numbers
Recalling Firm/
Manufacturer
Hospira Inc.
275 N Field Dr
Lake Forest, Illinois 60045-2579
Consumer Instructions Contact the recalling firm for information
For Additional Information Contact Mr. Joseph Sener
224-212-2000
Manufacturer Reason
for Recall
The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise.
FDA Determined
Cause 2
DESIGN: Device Design
Action The firm, Hospira Inc. sent "URGENT DEVICE FIELD CORRECTION" letters dated October 26, 2012 to all affected customers. The letters identified the products, problem and action(s) to be taken. The customers were instructed to read the information provided carefully, complete the attached Reply Form and return it to the firm via fax at 1-877-650-8667. The letter states that the System Operating Manual will be updated to clarify the instructions for setting the volume, and that the updated manuals will be sent as soon as they are available. Customers were instructed to notify their consignees if they have further distributed the affected product. The customers were requested to contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 4, for additional information and technical assistance.
Distribution Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Lavenia, South Africa, Spain, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates and Vietnam.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = FRN and Original Applicant = HOSPIRA, INC.
-
-