| | Class 2 Recall ABX Pentra ML |  |
| Date Posted |
January 07, 2013 |
| Recall Number |
Z-0645-2013 |
| Product |
ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2 |
| Code Information |
Lot/Serial# PML0937
|
Recalling Firm/ Manufacturer |
Horiba Instruments, Inc dba Horiba Medical 34 Bunsen Irvine, California 92618-4210 |
Reason for Recall |
Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information System. This result is from a previous run of the same patient that is stored in the ML database. This issue occurs in a very rare sequence of specific events. The affected software
|
| Action |
An "Urgent: Field Safety Notice" letter dated 11/09/2012 was sent to all customers who received the ABX PENTRA ML products. The letter informed the customers of the problem identified and the action to be taken. Customers were instructed to complete and return the enclose response form within 10 days. Customers with questions about the letter are instructed to contact their local HORIBA Medical representative. |
| Quantity in Commerce |
10 units |
| Distribution |
Nationwide in the US, including the states of CA, MI, IN, FL, and GA. |
| |
|
|
|
|