| | Class 2 Recall ABX PENTRA AST CP, |  |
| Date Posted |
January 03, 2013 |
| Recall Number |
Z-0632-2013 |
| Product |
ABX PENTRA AST CP, Part Number: A11A01629.
The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in vitro determination of Aspartate Amino Transferase (AST) in serum or plasma by colorimetry. |
| Code Information |
Lot number: 00456-4108
|
Recalling Firm/ Manufacturer |
Horiba Instruments, Inc dba Horiba Medical 34 Bunsen Irvine, California 92618-4210 |
| For Additional Information Contact |
Kim Walker 714-273-9254
|
Reason for Recall |
Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
|
| Action |
Horiba Medical sent a Field Safety Notification and Acknowledgement Letter dated November 20, 2012, to all affected customers. The letter notified the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 calendar days via mail, email or fax. Customers with questions were instructed to contact their local representative.
For questions regarding this recall call 714-273-9254. |
| Quantity in Commerce |
66886 units |
| Distribution |
Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX. |
| |
|
|
|
|