Class 2 Device Recall Perkin Elmer NEOBASE NONDERIVATIZED MSMS KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Perkin Elmer NEOBASE NONDERIVATIZED MSMS KIT
see related information
Date Posted January 16, 2013
Recall Status1 Terminated on March 07, 2014
Recall Number Z-0687-2013
Recall Event ID 63771
Premarket Notification
510(K) Number
K083130 
Product Classification System, Test, Amino Acids, Free Carnitines And Acylcarnitines Tandem Mass Spectrometry - Product Code NQL
Product NEOBASE NON-DERIVATIZED MSMS KIT Product Number: 3040-001U. This tandem mass spectrometry (MSMS) reagent kit is intended for the measurement and evaluation of amino acids, succinylacetone, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. Quantitative analysis of these analytes ) and their relationship with each other is intended to provide analyte concentration profiles that may aid in screening newborns for metabolic disorders.
Code Information Lot Numbers: 617253 , 617346, 618494 , 618653, 618764 , 619453
Recalling Firm/
Manufacturer
Perkinelmer
940 Winter Street
Waltham, Massachusetts 02451
Manufacturer Reason
for Recall
Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
FDA Determined
Cause 2
PRODUCTION CONTROLS: Process Control
Action Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
Quantity in Commerce 819 kits (10 plates/kit)
Distribution Nationwide Distribution.