 Class 2 Recall
Perkin Elmer Microplate
see related information
Date Posted January 16, 2013
Recall Number Z-0690-2013
Product Microplate, truncated V-bottomed lots included in the above listed products Product Number: 1380 5465. DBS Microplate is a general purpose 96 well microplate used in the preparation and examination of specimens from the human body.
Code Information Lot Numbers: 616792 , 617141 , 617980 , 618319, 618664, 618831, 619175
Recalling Firm/
Manufacturer
Perkinelmer
940 Winter Street
Waltham, Massachusetts 02451
Reason for
Recall
Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
Action Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
Distribution Nationwide Distribution.
 
-
-