| | Class 2 Recall BD MAX (tm) (6 channel) Instruments |  |
| Date Posted |
January 22, 2013 |
| Recall Number |
Z-0701-2013 |
| Product |
Becton, Dickinson and Company, BD MAX (tm) (6 channel) Instruments, Catalog number 441916, clinical diagnostic instrument. |
| Code Information |
Serial Numbers:
MX0001 to MX0065
CM0001 to CM0105
NOTE: Serial numbers are assigned sequentially. The above ranges will include all sequential numbers in between the beginning and end points.
|
Recalling Firm/ Manufacturer |
Becton Dickinson & Co. BD Diagnostic Systems 7 Loveton Circle Sparks, Maryland 21152-0999 |
| For Additional Information Contact |
Gail Claiborne 410-316-4054
|
Reason for Recall |
Clinical diagnostic instrument may exhibit minor defects, which could lead to erroneous test results.
|
| Action |
Becton Dickinson sent an Urgent Field Corrective Action letter September 18, 2012,by fax. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised that BD Field Service Associates will contact them within 7-10 days of the notification to determine if their instrument is impacted. Customers were requested to return a response form indicating the serial number on their instrument. If customers needed further assistance they were instrycted ti contact BD Technical Services Department at 1-800-638-8663.
For questions regarding this recall call 410-316-4054. |
| Quantity in Commerce |
86 Instruments |
| Distribution |
USA including CT, FL, IL, KS, MD, MI, MO, NY, PA, RI, TX and WI. Internationally to Canada, Europe and Japan. |
| |
|
|
|
|