| | Class 2 Recall GE Healthcare Brivo XR385 DR-F/ X-Ray System |  |
| Date Posted |
February 26, 2013 |
| Recall Number |
Z-0761-2013 |
| Product |
GE Healthcare Brivo XR385 DR-F/ X-Ray System
X-Ray System |
| Code Information |
Model Number : 5271997
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha, Wisconsin 53188-1615 |
| Consumer Instructions |
No consumer action necessary |
Reason for Recall |
GE Healthcare has become aware of a software issue on the interface of the Brivo XR385 DR-F/ Model 5271997 radiographic product related to Visual indication as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-1-3.
|
| Action |
GE Healthcare will, without charge, remedy the issue or bring the product into
compliance with each applicable Federal and IEC standard in accordance with a planto be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to you or through a GE Healthcare field engineer site visit. A GE Healthcare Service Representative will update the software on the system to address the issue.
Contact Call Center phone number: 800-437-1171 (United States). For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification. |
| Quantity in Commerce |
9 units installed in US |
| Distribution |
Nationwide Distribution including NM, WY, FL, MO, MA, and MI. |
| |
|
|
|
|