| | Class 2 Recall GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ |  |
| Date Posted |
March 11, 2013 |
| Recall Number |
Z-0934-2013 |
| Product |
GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging
systems.
For use in generating fluoroscopic images of human anatomy. |
| Code Information |
SERIAL NUMBER
592043BU4
592676BU1
593459BU1
593465BU8
592458BU4
592673BU8
592682BU9
593059BU9
593051BU6
593475BU7
593476BU5
592049BU1
593468BU2
593054BU0
592047BU5
592042BU6
592046BU7
592459BU2
593048BU2
593056BU5
593049BU0
593063BU1
593062BU3
593464BU1
594762BU7
593473BU2
592451BU9
593460BU9
597954BU7
592456BU8
592674BU6
593052BU4
593058BU1
590867BU8
592050BU9
593064BU9
592460BU0
592457BU6
592680BU3
592455BU0
593462BU5
592678BU7
594763BU5
593055BU7
592675BU3
592452BU7
593463BU3
594385BU7
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha, Wisconsin 53188-1615 |
Reason for Recall |
Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
|
| Action |
No consignee letter sent out. GE Healthcare's recall strategy was to replace the impacted cable on all affected Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ systems that were distributed. At this time, deployment of this recall is considered complete as of 9/6/2011. |
| Quantity in Commerce |
48 |
| Distribution |
Worldwide Distribution-USA including the states of FL, IL, IN, NJ, NC, OK, TX, and WI, and the countries of UNITED KINGDOM, TUNISIA, SERBIA, RUSSIAN FEDERATION, NORWAY, NIGERIA, LITHUANIA, LEBANON, KOREA, REPUBLIC OF KOREA, ITALY, INDONESIA, INDIA, FRANCE, FINLAND, COLOMBIA, CHINA, and BRAZIL. |
| |
|
|
|
|