 Class 2 Recall
MEDRAD(R) Hand Controller Sheath
see related information
Date Posted February 21, 2013
Recall Number Z-0870-2013
Product MEDRAD Hand Controller Sheath, Catalog number AVA 500 HCS/3010903, for cardiac catheterization procedures. The Sterile Sheath is specifically intended for use in the x-ray angiography environment to cover the Medrad Avanta Sterile Hand Controller to prevent cross contamination between patients.
Code Information Catalog number AVA 500 HCS/3010903 - Lot numbers D72211, DA82711, DA93131, D101091, D101681, D110591, D111961, D113341A, D120731, D101641, D82741, D93131, DA101091, DA101691, D110601, D111991, D113351, D120831A, D72251, D83011, D93151, D101101, D101701, D110611, D112061, D113351A, D120741, D72261, DA83021, DA93151, DA101101, DA101721, D110621, D112131, D113461, D120761, D72281, DA83031, DA93221, D101111, D101721, D110631, D112131A, D113461A, D120771, D72291, D83041, D93231, DA101111, D101731, D110661, D112341, D113471, D120831, D72651, DA83121, D93411, D101121, D101741, D110731, D112341A, D113471A, D120861, D72671, D83151, DA93411, DA101121, D101861, D110741, D112351, D113491, DA72681, D83651, D93421, D101131, DA101861, D110751, D112351A, D113501, DA72691, D90021, DA93421, DA101131, D101881, D110871, D112361, D120161, D72751, D90541, D93431, D101171, DA101881, D110881, D112361A, D120161A, D120891, D90561, DA93431, D101301, DA101901, D110901, D112411, D120171, D80221, D90611, D93561, D101311, D101911, D111201, D112411A, D120171A, DA80231, DA90631, D93631, DA101441, DA101931, D111221, D112421, D120181, DA80241, D90821, DA93631, D101441, D101931, D111231, D112421A, D120311, D80281, DA90831, DA100141, D101461, D101951, D111241, D112431, D120311A, D80391, D90981, D100151, DA101461, DA101951, D111261, D112431A, D120321, DA80561, D91031, D100181, D101471, D101971, D111261A, D112441, D120321A, D80631, DA91041, DA100191, DA101481, D102111, D111291, D112441A, D120331, DA80641, DA91061, DA100321, D101481, D102141, D111291A, D112481, D120341, D80871, D91071, D100331, DA101511, DA102141, D111301, D112481A, D120371, D80981, D91181, DA100341, D101511, D102151, D111301A, D112491, D120371A, DA81281, D91191, D100351, D101531, DA102151, D111311, D112501, D120381, DA81421, D91281, DA100461, DA101531, D102351, D111311A, D112621, D120381A, D81471, D91331, D100561, D101541, DA102351, D111321, D112621A, D120391, D81541, D91351, D100671, D101551, D102811, D111321A, D112631, D120451, DA81821, D91381, DA100671, DA101551, D102821, D111361, D112631A, D120511, D81971, D91561, DA100691, D101581, D102921, D120861A, D112641, D120521, D81991, D91901, D100691, DA101581, D103361, D111351, D112721, D120521A, DA82031, D92061, D100711, D101591, D103401, D111371, D112791, D120531, D82131, D92091, DA100711, DA101601, D103481, D111371A, D113071, D120541, D82141, D92111, DA100741, D101601, D103631, D111381, D113081, D120541A, D82171, D120881, D100751, D101611, D103611, D111711, D113111, D120551, D82181, D92501, D101021, D101621, D103571, D111721, D113111A, D120591, D82191, D92521, DA101021, DA101621, D103621, D111731, D113291, D120601, DA82201, DA92781, D101031, D101631, D110181, D111751, D113291A, D120671, D120881A, D92791, D101041, D101651, D110191, D111861, D113321, D120671A, D82241, D93081, DA101041, DA101651, D110321, D111871, D113321A, D120681, DA82251, D93091, D101061, D101661, D110331, D111891, D113331, D120681A, D82671, D93101, DA101061, D101671, D110351, D111921, D113331A, D120691, D82701, D93111, D101071, DA101671, D110541, D111921A, D113341, and D120721.
Recalling Firm/
Manufacturer
Medrad Inc
1 Medrad Dr
Indianola, Pennsylvania 15051-9759
Reason for
Recall
The seal between the white polyethylene and the clear plastic may be missing on the Hand Controller Sheath package, resulting in a potential breach of the sterility of this accessory.
Action Medrad sent a Recall letter dated November 5, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer were instructed to review their inventory and quarantine any packages with an expiration date prior to April, 2017. Customers were also given the option of conducting a visual inspection on site to determine if the product should be returned. Alternatively, the customer has the option of returning the affected product. If the customer has affected product, they are instructed to contact Medrad Customer Support to make arrangements for the return and replacement of the product. For further questions please call our Customer Support Team at 1-877-229-3767.
Quantity in Commerce 832,175
Distribution Worldwide Distribution including USA nationwide.
 
-
-