| Date Posted |
February 28, 2013 |
| Recall Number |
Z-0897-2013 |
| Product |
Ingenuity TF PET/CT. Diagnostic imaging system |
| Code Information |
Model # 882456, Serial # 2004. Units currently running Ingenuity TF PET/CT system software version 4.0.0.26635.
|
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland, Ohio 44143-2131 |
| For Additional Information Contact |
Michael McAndrew 800-722-9377 Ext. 5
|
Reason for Recall |
The system may not default to the same gating trigger with which the data were acquired. This could cause a risk for patients.
|
| Action |
Philips Medical Systems (Cleveland), Inc. sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated January 21, 2013 to the affected customer. The letter identified the product, problem, and actions to be taken by the customer. Contact your local Philips representative or Philips Healthcare office for further information concerning this issue. |
| Quantity in Commerce |
1 Unit |
| Distribution |
Distributed only in Ohio. |
| |
|