• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Dimension Vista(R) System Vista Drug 4 Calibrator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Dimension Vista(R) System Vista Drug 4 Calibratorsee related information
Date Initiated by FirmFebruary 21, 2013
Date PostedApril 18, 2013
Recall Status1 Terminated 3 on January 07, 2014
Recall NumberZ-1151-2013
Recall Event ID 64596
510(K)NumberK093441 K112142 
Product Classification Enzyme immunoassay, digoxin - Product Code KXT
ProductSiemens Dimension Vista(R) System Vista Drug 4 Calibrator. The device is used as an in vitro diagnostic product.
Code Information Catalog Number KC460, SMN 10472284, lot 2KD052, exp 5/1/14 and Catalog Number KC460A, SMN 10476471, lot 2KD053, exp 5/1/14.
FEI Number 2517506
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
500 Gbc Dr Ms 514
PO BOX 6101
Newark DE 19702-2466
For Additional Information ContactAydee Crawford
302-631-6312
Manufacturer Reason
for Recall
Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(R) Digoxin(DIGXN) Flex(R) reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL.
FDA Determined
Cause 2
Process control
ActionThe firm initiated their recall of this product by sending an Urgent Medical Device Recall letter dated February 2013 to all Dimension Vista Drug 4 Calibrator customers who received the affected lots. The letter notified them of the issue, the potential risk to health, and actions to be taken by customers.
Quantity in Commerce1,638
DistributionDistributed within the US to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = KXT
-
-