• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CoaguChek XS System (PST)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CoaguChek XS System (PST) see related information
Date Posted June 24, 2013
Recall Status1 Terminated on April 16, 2014
Recall Number Z-1581-2013
Recall Event ID 65246
510(K)Number K062925 
Product Classification Test, time, prothrombin - Product Code GJS
Product CoaguChek XS System (PST)

The CoaguChek XS PT test strips are part of the CoaguChek XS System. The CoaguChek XS System measures blood-clotting time for people who are taking anticoagulation medications such as Coumadin¿ or warfarin.
Code Information Model/Catalog/Part Number: 04837738001
Recalling Firm/
Manufacturer
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis IN 46256-1025
317-576-2000 Ext. 249
For Additional Information Contact Todd Siesky
317-521-3911
Manufacturer Reason
for Recall
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
FDA Determined
Cause 2
Software design
Action Rochet sent an Urgent Medical Device Correction letter dated May 16, 2013, via UPS Ground to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were issued the following instructions: If you are using Warfarin therapy in combination with antibiotics and/or chemotherapy drugs, which could potentially lead to an extremely high test result (> 10 INR), you may in rare cases receive repeated ERROR 6 messages instead of a test result. If the ERROR 6 message is displayed repeatedly on your CoaguChek XS meter, immediately contact your treating physician. The CoaguChek XS PT Test Strip package insert; the online version of the CoaguChek XS PST system users manual; and the CoaguChek XS systems are being updated to provide additional ERROR 6 message information and actions. Updated versions will be available on the www.poc.roche.com website by the end of June 2013. Health Care Professionals are instructed to use an alternative method to determine INR results if an ERROR 6 message is displayed repeatedly on the CoaguChek XS, CoaguChek XS Plus, or CoaguChek XS Pro meters for the same patient. Self testing patients are instructed to contact their Health Care provider immendiately if the ERROR 6 message is displayed repeatedly on the CoaguChek XS. Please contact your Roche Account Manager or Roche Diagnostics Point of Care Technical Service, 24 hours a day, seven days a week at 1-800-428-4674 if you have questions about the information contained in this UMDC. For questions regarding this recall call 317-521-3711.
Quantity in Commerce 204,205 meters total
Distribution Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = GJS and Original Applicant = ROCHE DIAGNOSTICS CORP.
-
-