 Class 2 Device Recall Osteoraptor Suture Anchor see related information
Date Posted September 17, 2013
Recall Status1 Open
Recall Number Z-2223-2013
Recall Event ID 66037
510(K)Number K082215 
Product Classification Staple, fixation, bone - Product Code JDR
Product OSTEORAPTOR 2.3 mm Suture Anchor with 1 ULTRABRAID COBRAID Suture, Black; Part Number: 72201992

Biodegradable suture anchor
Code Information 50272332
50272462
50272750
50274144
50274350
50277105
50278907
50281510
50282171
50282827
50283849
50284637
50285830
50286658
50287767
50289371
50290043
50291934
50292091
50295383
50296070
50297228
50297782
50298793
50300135
50301932
50306856
50307462
50308037
50309210
50310436
50310780
50312090
50313439
50313818
50314915
50315807
50316601
50316986
50317483
50318001
50318687
50319907
50320679
50321916
50322244
50323257
50323630
50323780
50324608
50325611
50326067
50327821
50329242
50331546
50332253
50333092
50334400
50335620
50336405
50338241
50339832
50341062
50341605
50343271
50344262
50344618
50346215
50347305
50349093
50350513
50350978
50352032
50352701
50352997
50353606
50355123
50355672
50356987
50357104
50358206
50359714
50360454
50360914
50361413
50362647
50363190
50364588
50364909
50364915
50370316
50370385
50372675
50372728
50372892
50374025
50374503
50375408
50376044
50376238
50377316
50378706
50380530
50381197
50382647
50384163
50384950
50385362
50387220
50387886
50387896
50388742
50390996
50391608
50392266
50394701
50395538
50397273
50398231
50399415
50400574
50401262
50401918
50402459
50402823
50403367
50403718
50404238
50404920
50405868
50405869
50410769
50412026
50412957
50414254
50416317
50416526
50417204
50418001
50418331
50418992
50421031
50422577
50423435
50423591
50426460
50426933
50427032
50428116
50428624
50429271
50429999
50430000
50430400
50430503
50430504
50430505
50430657
50431379
50433038
50433039
50433040
50433041
50433127
50434653
50435732
50435733
50435734
50435809
50436890
50438659
50440452
50441439
50450277
50450752
50451781
50452210
50452585
50453453
50454295
50454418
50455195
50455483
50455813
50456909
50457385
50458416
50458475
50459739
50460097
50460240
50461060
50462482
50463458
50464108
Recalling Firm/
Manufacturer
Smith & Nephew, Inc. Endoscopy Division
150 Minuteman Drive
Andover MA 01810-1031
978-749-1073
For Additional Information Contact Albert A. Pytka
978-749-1073
Manufacturer Reason
for Recall
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
FDA Determined
Cause 2
Packaging
Action Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
Quantity in Commerce 18919
Distribution Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.