Class 2 Device Recall HGE Healthcare Patient Monitor B30, Procare Monitor B20, Procare Monitor B40, and Monitor B40 |
|
Date Initiated by Firm |
October 11, 2013 |
Date Posted |
December 12, 2013 |
Recall Status1 |
Terminated 3 on February 03, 2016 |
Recall Number |
Z-0489-2014 |
Recall Event ID |
66970 |
510(K)Number |
K122253 K120598 K130584
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
HGE Healthcare Patient Monitor B30, Procare Monitor B20, Procare Monitor B40, and Monitor B40 K122253 (Procare Monitor B20): The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor B20 is intended for use under the direct supervision of a licensed health care practitioner. The PROCARE Monitor B20 is not intended for use during MRI. The PROCARE Monitor B20 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, EGG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esosophageal/Nasopharyngeal/Tymnpanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, and functional oxygen saturation (Sp0 2) and pulse rate via continuous monitoring , including monitoring during conditions of clinical patient motion or low perfusion. K120598 (Procare Monitor B40): The PROCARE Monitor 840 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The PROCARE Monitor 840 is intended for use under the direct supervision of a licensed health care practitioner. The PROCARE Monitor 840 is not intended for use during MRI. The PROCARE Monitor B40 monitors and displays oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esophageal/Nasopharyngeal/Tympaiic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/ Surface temperature, and functional oxygen saturation (SpO2) and pulse rate via continuous monitoring, including monitoring during conditions of clinical patient motion or low perfusion. K130584 (Monitor B40): The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B40 is not intended for use during MRI. The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network. The Monitor B40 monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring. Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/ Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy. Product Quantity Distributed (Int) 16,061 (343 US, 15,718 OUS)
|
Code Information |
Mfg Lot or Serial # SG212501463WA SG213142063WA SG212491284WA SG212491286WA SG212491320WA SG212491328WA SG213091890WA SG213142034WA SG213142065WA SG213172123WA SG213172126WA SG213172141WA SF311283756WA SF312205556WA SG212190622WA SG212190623WA SG212190624WA SG212190625WA SG212190626WA SG212190627WA SG212190628WA SG212190629WA SG212190630WA SG212190631WA SG212160603WA SG212160606WA SG212230660WA SG212260716WA SG212260735WA SG212260737WA SG212260739WA SG212270742WA SG212270747WA SG212491329WA SG212491346WA SG212421031WA SG212431128WA SG212421034WA SG213021574WA SG213031688WA SG212521557WA SG212521545WA SF312396183WA SF312396187WA SF312396189WA SF312396190WA SF312396191WA SF312396192WA SF312396193WA SG213142038WA SG212521547WA SG213021582WA SG213021585WA SG213031681WA SG213031683WA SG213031684WA SG213031685WA SG213031687WA SG213031689WA SF310200786WA SF310200788WA SG212521544WA SG212521546WA SG212521548WA SG212521549WA SG212521550WA SG212521551WA SG212521552WA SG212521553WA SG212521554WA SG212521555WA SG212521556WA SG212521558WA SG212521559WA SG212521560WA SG212521561WA SG212521563WA SG212521564WA SG212521565WA SG213021570WA SG213021571WA SG213021572WA SG213021573WA SG213021575WA SG213021576WA SG213021577WA SG213021579WA SG213021580WA SG213021581WA SG213021583WA SG213021584WA SG213021612WA SG213021614WA SG213021615WA SG213021617WA SG213021618WA SG213021619WA SG213021620WA SG213021621WA SG213021622WA SG213021626WA SG213021627WA SG213021630WA SG213031631WA SG213031632WA SG213031634WA SG213041690WA SG213041691WA SG213041692WA SG213041709WA SG213041712WA SG213041714WA SG213041715WA SG213041723WA SG213041724WA SG213041725WA SG213041726WA SG213041727WA SG213041728WA SG213041729WA SG213041730WA SG213041731WA SG213041732WA SG213041733WA SG213041734WA SG213101906WA SG213132003WA SG213132015WA SG213142039WA SG213142040WA SG213142041WA SG213142042WA SG213142044WA SG213142045WA SG213142046WA SG213142049WA SG213142051WA SG213142055WA SG213142057WA SG213142058WA SG213142060WA SG213142064WA SG213142066WA SG213142067WA SG213142069WA SG213142070WA SG213142071WA SG213142072WA SG213142073WA SG213142074WA SG213142075WA SG213142076WA SG213142077WA SG213142078WA SG213142079WA SG213142080WA SG213142081WA SG213142082WA SG213142083WA SG213142084WA SG213142085WA SG213142087WA SG213142092WA SG213142093WA SG213142096WA SGF12515079WA SGF12515082WA SGF12515093WA SGF13025268WA SGF13025274WA SGF13025278WA SGF13025282WA SGF13025283WA SGF13025284WA SGF13025289WA SGF13025297WA SGF13025298WA SGF13025300WA SGF13025301WA SGF13025303WA SGF13025304WA SGF13025305WA SGF13025306WA SGF13025307WA SGF13025308WA SGF13025309WA SGF13025310WA SGF13025311WA SGF13025312WA SGF13025313WA SGF13025314WA SGF13025315WA SGF13025317WA SGF13025318WA SGF13025319WA SGF13025320WA SGF13025321WA SGF13025322WA SGF13025323WA SGF13025324WA SGF13025325WA SGF13025326WA SGF13025327WA SGF13025328WA SGF13025329WA SGF13025330WA SGF13025331WA SGF13025332WA SGF13025333WA SGF13025334WA SGF13025335WA SGF13025336WA SGF13025337WA SGF13025338WA SGF13025339WA SGF13025340WA SGF13035372WA SGF13035373WA SGF13035374WA SGF13035375WA SGF13045480WA SGF13045481WA SGF13045482WA SGF13045483WA SGF13045484WA SGF13045486WA SGF13045487WA SGF13045488WA SGF13045489WA SGF13045490WA SGF13045491WA SGF13045492WA SGF13045493WA SGF13045494WA SG213031668WA SG213031669WA SG213031670WA SG213031671WA SG213031672WA SG213031673WA SG213031674WA SG213031675WA SG213031676WA SG213031677WA SG213031678WA SG213031680WA SG213031682WA SG213031686WA SGF13025316WA SGF13025341WA SGF13025342WA SGF13025343WA SGF13035434WA SGF13035435WA SGF13035437WA SGF13035439WA SGF13035440WA SGF13035441WA SGF13035442WA SGF13035443WA SGF13035444WA SGF13035445WA SGF13035446WA SG212521562WA SG213021578WA SG213021616WA SG213131996WA SG213131997WA SG213131999WA SG213132001WA SG213132002WA SG213132009WA SG213132010WA SG213132011WA SG213132012WA SG213132013WA SG213132014WA SG213132016WA SG213132017WA SG213132018WA SG213132020WA SG213132021WA SG213132022WA SG213132024WA SG213132025WA SG213132026WA SG213132027WA SG213142032WA SG213142033WA SG213142035WA SG213142036WA SG213142037WA SG213142043WA SG213142047WA SG213142048WA SG213142050WA SG213142052WA SG213142053WA SG213142054WA SG213142056WA SG213142059WA SG213142061WA SG213142062WA SG213142068WA SG213142086WA SG213142088WA SG213142089WA SG213142090WA SG213172122WA SG213172125WA SG213172127WA SG213172128WA SG213172129WA SG213172130WA SG213172131WA SG213172132WA SG213172133WA SG213172134WA SG213172135WA SG213172136WA SG213172137WA SG213172138WA SG213172139WA SG213172140WA SG213172142WA SG212410956WA SG212410960WA SG212410954WA SG212410955WA SG212410957WA SG212410958WA SG212410959WA SG212421051WA SG212421053WA SG212421054WA SG212481231WA SG212481235WA SG212481260WA SG212481262WA SG212481263WA SG212481264WA SG212491281WA SG212491290WA SG212491291WA SG212491292WA SG212491293WA SG212491295WA SG212491300WA SG212491302WA SG212491306WA SG212491307WA SG212491309WA SG212410968WA SG212421024WA SG212421032WA SG212421033WA SG212421059WA SG212421061WA SG212421063WA SG212421064WA SG212431103WA SG212431104WA SG212431106WA SG212431108WA SG212471130WA SG212471138WA SG212421021WA SG212421022WA SG212421023WA SG212421060WA SG212421062WA SG212431105WA SG212431107WA SG212431110WA SG212471129WA SG212471131WA SG212471132WA SG212471133WA SG212471135WA SG212471136WA SG212471137WA SG212491347WA SG212491413WA SG212491426WA SG212501437WA SG212501454WA SG212501455WA SG212511517WA SG212511518WA SG212511519WA SG212511520WA SG212511522WA SG212501477WA SG212501480WA SG212501483WA SG212501485WA SG212501490WA SG212511493WA SG212511527WA SG212511528WA SG212511530WA SG212511531WA SG212511525WA SG213031638WA SG213031640WA SG213031641WA SG213031647WA SG213031648WA SG213031649WA SG213031650WA SG213031652WA SG213031653WA SG213031654WA SG213031655WA SG213041719WA SG213041721WA SG213051743WA SG213051739WA SG213061799WA SG213091851WA SG213091852WA SG213091853WA SG213091854WA SG213091855WA SG213091856WA SG213091857WA SG213091858WA SG213091859WA SG213091860WA SG213091861WA SG213091862WA SG213091864WA SG213051745WA SG213051751WA SG213051752WA SG213051755WA SG213081846WA SG213101915WA SG213101916WA SG213101917WA SG213101921WA SG213101918WA SG213101919WA SG213101920WA SG213101922WA SG213101923WA SG213101924WA SG213101925WA SG213101926WA SG213101927WA SG213101928WA SG213101929WA SG213101953WA SG213101954WA SG213101955WA SG213101956WA SG213101905WA SG213101907WA SG213101908WA SG213091882WA SG213091883WA SG213091884WA SG213091885WA SG213091886WA SG213091887WA SG212320846WA SG212501461WA SG212501462WA SG212501467WA SG212501468WA SG212501469WA SG212501470WA SG212501471WA SG212501473WA SG212501475WA SG212501476WA SG212501478WA SG212501479WA SG212501481WA SG212501484WA SG212501486WA SG212501487WA SG212501491WA SF312466527WA SG212421058WA SG212491287WA SG212491288WA SG212491296WA SG212491297WA SG212491298WA SG212491299WA SG212491301WA SG212491303WA SG212491304WA SG212491310WA SG212491313WA SG212491314WA SG212491315WA SG212491316WA SG212491318WA SG212491322WA SG212491326WA SJF13010012WA SJF13010022WA SJF13010023WA SJF13020025WA SJF13020051WA SJF13020042WA SF313026802WA SF312526776WA SF313016780WA SF313026799WA SF310220863WA SG212431122WA SG212431127WA SG212471145WA SG212471152WA SGF13045568WA SGF13045584WA SGF13045586WA SGF13055596WA SGF13055633WA SGF13055635WA SGF13055637WA SGF13055638WA SGF13055640WA SGF13055642WA SGF13085870WA SGF13085878WA SGF13085884WA SGF13085892WA SGF13085893WA
SGF13085895WA SGF13085896WA SGF13085897WA SGF13085898WA SGF13085899WA SGF13085900WA SG212481259WA SG212481261WA SF310472212WA SF311263709WA SF312095175WA SF311143076WA SF311143078WA SF311143079WA SF311133068WA SF311143074WA SF311283761WA SF311283762WA SF311283767WA SF311283770WA SF311334256WA SF311344335WA SF311414398WA SF311414404WA SF311414430WA SF311414434WA SF311414435WA SF311414441WA SF311434505WA SF311404369WA SF311424462WA SF311424473WA SF311504768WA SF311504769WA SF311504777WA SF310472195WA SF310472210WA SF312265680WA SF312265685WA SF312265694WA SF312275701WA SF312275709WA SF312275711WA SF312275712WA SF312275713WA SF312275715WA SF312275719WA SF312275728WA SF311334241WA SF311414400WA SF311414405WA SF311414406WA SF311414407WA SF310090523WA SG212150577WA SGF12242881WA SGF12242888WA SGF12242889WA SGF12242890WA SGF12061539WA SGF12061552WA SGF12061546WA SGF12061550WA SGF12071586WA SGF12132260WA SGF12172513WA SG212020377WA SGF12112089WA SGF12112109WA SGF12112111WA SGF12313372WA SGF12313374WA SGF12333514WA SGF12081736WA SGF12152340WA SGF12152338WA SGF12051436WA SGF12051446WA SGF12051458WA SGF12051459WA SGF12051461WA SGF12051462WA SGF12051463WA SGF12071592WA SGF12071600WA SGF12071603WA SGF12071614WA SGF12071619WA SGF12071621WA SGF12071647WA SGF12081747WA SGF12081751WA SG211490206WA SG212050406WA SG212050415WA SG212050416WA SG212070419WA SG212070421WA SG212070422WA SG212070426WA SG212070432WA SG212090439WA SG212090442WA SGF12081770WA SGF12081787WA SGF12091884WA SGF12091890WA SGF12091907WA SGF12101956WA SGF12101960WA SG212280754WA SG212310812WA SG212310816WA SG212320821WA SG212320822WA SG212320823WA SG212320824WA SG212320825WA SG212310815WA SG212320820WA SG212320828WA SGF12363684WA SGF12383836WA SGF12142311WA SGF12142312WA SGF12142313WA SGF12142318WA SGF12263185WA SGF12263202WA SGF12263204WA SGF12333504WA SGF12333506WA SGF12333515WA SGF12333518WA SGF11370069WA SGF11470399WA SGF11460391WA SG212120487WA SG212120493WA SG212120497WA SG212120504WA SG212130507WA SG212130508WA SG212130509WA SG212130511WA SG212130520WA SG212150547WA SG212150549WA SG212150551WA SG212150554WA SG212150556WA SG212150560WA SG212150564WA SG212150565WA SG212150566WA SG212150568WA SG212150581WA SG212150583WA SG212150584WA SG212150586WA SG212150587WA SG212150591WA SG212150593WA SG212150594WA SG212150595WA SG212170607WA SG212170608WA SG212170609WA SG212170610WA SG212170611WA SG212170613WA SG212170614WA SG212170615WA SG212170616WA SG212190617WA SG212190619WA SG212230620WA SG212230621WA SG212230649WA SG212230650WA SG212230651WA SG212230652WA SG212230653WA SG212230654WA SGF12112127WA SGF12142271WA SGF12142275WA SGF12142279WA SGF12152341WA SGF12162449WA SGF12172486WA SGF12172494WA SGF12172563WA SGF12172568WA SGF12172571WA SGF12172596WA SGF12232672WA SGF12142330WA SGF12021263WA SG211500215WA SG211500220WA SG211500221WA SG211500224WA SG211500225WA SGF12112132WA SGF12122209WA SGF12263175WA SGF12122213WA SGF12122219WA SF311223400WA SF310200752WA SF310190695WA SF309470108WA SF310502384WA SGF12474420WA SGF12474390WA SGF12494745WA SF311324700WA SF313147104WA SF313177118WA SF312155330WA SF312155331WA SF310472224WA SF309530232WA SF310220822WA SF310230938WA SF311263715WA SF312105229WA SF312155276WA SF310271108WA SF309250001WA SF310060338WA SF310060344WA SF310060351WA SF310060352WA SF310512413WA SF311112949WA SF310411773WA SGF13176730WA SG212491417WA SG212421091WA SG212491377WA SG212491416WA SG212491420WA SG212491424WA SG212501432WA SG212501433WA SG212501438WA SG213021566WA SG213021567WA SG213021568WA SG213021623WA SG213021624WA SG213021625WA SG213021628WA SG213021629WA SF311112956WA SF311112977WA SGF12484657WA SGF12484659WA SF309530230WA SF309470111WA SF309490114WA SF309490117WA SF309490121WA SF310090506WA SF310090514WA SF310090527WA SF310230934WA SF310230935WA SF310240958WA SF310240959WA SF310240960WA SF310240961WA SF310240962WA SF310240963WA SF310240966WA SF310240971WA SF310241042WA SF310241043WA SF311072728WA SF311072724WA SF311072746WA SF311072747WA SF311072749WA SF311072754WA SF311072753WA SF311072752WA SF311072799WA SF311072803WA SF311112953WA SF311082862WA SF311263737WA SF311283740WA SF311334246WA SF311434504WA SF311434514WA SF311434517WA SF311434518WA SF311434519WA SF311434522WA SF311434525WA SF311434527WA SF312095166WA SF312095189WA SF312175413WA SF312175414WA SF312175415WA SG212410999WA SG212411019WA SG212421035WA SG212421037WA SG212421038WA SG212421039WA SG212421040WA SG212421041WA SG212421042WA SG212421043WA SG212421044WA SG212421045WA SG212421046WA SG212421047WA SG212421048WA SG212421049WA SGF12414047WA SGF12414048WA SGF12414049WA SGF12414053WA SGF12414057WA SGF12414058WA SGF12414062WA SGF12414073WA SF312446395WA SF312446396WA SF312446397WA SF312446400WA SF312446401WA SF312446402WA SF312466572WA SF312466573WA SF312466576WA SF312466577WA SF312466580WA SF312476583WA SF312476587WA SF312496603WA SGF13176644WA SGF13176665WA SF309470109WA SF310050331WA SF310060335WA SF310060336WA SF310060337WA SF310060339WA SF310060340WA SF310060341WA SF310060346WA SF310060348WA SF310060350WA SF309530223WA SJF13010011WA SGF11440228WA SGF11490535WA SGF11490536WA SGF11490537WA SGF11490538WA SGF11500539WA SGF11500562WA SGF11500563WA SGF11500564WA SGF11500566WA SGF11500568WA SGF11500569WA SGF11500570WA SGF12263186WA SGF12263192WA SGF12263208WA SGF12313370WA SGF11430195WA SGF11440248WA SGF11440252WA SGF11440253WA SGF11440254WA SGF11440255WA SG211430123WA SG211440128WA SG211440130WA SG211440131WA SG211440132WA SGF11430187WA SGF11430189WA SGF11440225WA SGF11440230WA SGF11440231WA SGF11440233WA SGF11440236WA SGF11440241WA SGF11440242WA SGF11440247WA SGF11440249WA SGF11440257WA SGF11440258WA SGF11440263WA SGF11440264WA SGF11440265WA SGF11440266WA SGF11440267WA SGF11440269WA SGF11440270WA SGF11440271WA SGF11440272WA SGF11440273WA SGF11440274WA SGF11440275WA SGF11440276WA SGF11440278WA SGF11440279WA SGF11440280WA SGF11500550WA SGF11500551WA SGF11500554WA SGF11500555WA SGF11500556WA SGF11500557WA SGF11480432WA SGF11480439WA SGF11480441WA SGF11480445WA SGF11480448WA SGF11480449WA SGF11480452WA SGF11480456WA SGF11490459WA SGF11490471WA SGF11490516WA SGF11490518WA SGF11490528WA SGF11500540WA SGF11500542WA SGF11500545WA SGF12051352WA SGF12051357WA SGF12051362WA SGF12051363WA SGF12051365WA SGF12051366WA SGF12051372WA SGF12051373WA SGF12051374WA SGF12051376WA SGF12051377WA SGF12051378WA SGF12051379WA SGF12051380WA SGF12051381WA SGF12051383WA SGF12051384WA SGF12051385WA SGF12051386WA SGF12051387WA SGF12051389WA SGF12051390WA SGF12051391WA SGF12051392WA SGF12051395WA SGF12051396WA SGF12051397WA SGF12051398WA SGF12051400WA SGF12051401WA SGF12031318WA SGF12031319WA SGF12031320WA SGF12031321WA SGF12031322WA SGF12031324WA SGF12031325WA SGF12031331WA SGF12031332WA SGF12031333WA SGF12031334WA SGF12031335WA SGF12031336WA SGF12031337WA SGF12051350WA SGF12051351WA SGF12051353WA SGF12051355WA SGF12051356WA SGF12051360WA SGF11520911WA SGF11520931WA SGF12021187WA SGF12021189WA SGF12021211WA SGF12021212WA SGF12021222WA SGF12021252WA SGF12021257WA SGF12031283WA SGF12031310WA SGF12031311WA SGF12031312WA SGF12031313WA SGF12031314WA SGF12031315WA SGF12031316WA SGF12031317WA SGF12031328WA SGF12031330WA SGF12263184WA SGF12263207WA SGF12313408WA SGF12313421WA SGF12313423WA SGF12313424WA SGF12313426WA SGF12313436WA SGF12313437WA SGF12313439WA SGF12313441WA SGF12313442WA SGF12313444WA SGF12313446WA SGF12323448WA SGF12323450WA SGF12323451WA SGF12323452WA SGF12323453WA SGF12323454WA SGF12323455WA SGF12323457WA SGF12323458WA SGF12323459WA SGF12323460WA SGF12323461WA SGF12323462WA SGF12263195WA SGF12263217WA SGF12273263WA SGF12273265WA SGF12273273WA SGF12273274WA SGF12273275WA SGF12283293WA SGF12283296WA SGF12283299WA SGF12283300WA SGF12293318WA SGF12293343WA SGF12293352WA SGF12293359WA SGF12313362WA SGF12313364WA SGF12313379WA SGF12313385WA SGF12313387WA SGF12313390WA SGF12313398WA SGF12313399WA SGF12313402WA SGF12313417WA SGF12313428WA SGF12313429WA SGF12313432WA SGF12323487WA SGF12333511WA SGF12333527WA SGF12333529WA SGF12333535WA SGF12333537WA SGF12353561WA SGF12353564WA SGF12353570WA SGF12363650WA SGF12363651WA SF310431911WA SF311504755WA SF311504761WA SF311504763WA SGF12474377WA SGF12474441WA SGF12474442WA SGF12474447WA SGF12474448WA SGF12474452WA SGF12474453WA SGF12474454WA SGF12474455WA SGF12474456WA SGF12474460WA SGF12484491WA SGF12484495WA SGF12484501WA SGF12484502WA SGF12484503WA SGF12484505WA SGF12484506WA SGF12484507WA SGF12484508WA SGF12484509WA SGF12484510WA SGF12484511WA SGF12484512WA SGF12484513WA SGF12484514WA SGF12484515WA SGF12484516WA SGF12484517WA SGF12484518WA SGF12484519WA SGF12484520WA SGF12484526WA SGF12484527WA SGF12484528WA SGF12484529WA SGF12484530WA SGF12484531WA SGF12484532WA SGF12484533WA SGF12484534WA SGF12424264WA SGF12474355WA SGF12484523WA SGF13176709WA SGF13176710WA SGF13176711WA SGF13176712WA SGF13176713WA SGF13176714WA SGF13176715WA SGF13176716WA SGF13176717WA SGF13176718WA SGF13176719WA SGF13176720WA SGF13176721WA SGF13176722WA SGF13176723WA SGF13176724WA SGF13176725WA SGF13176726WA SGF13176727WA SGF13176741WA SGF13085850WA SGF13085901WA SGF13095908WA SGF13095910WA SF310100553WA SGF12434308WA SGF12434310WA SGF12434316WA SGF12434317WA SGF12434318WA SGF12434320WA SGF12434322WA SGF12434327WA SF310361451WA SF311213353WA SF311213354WA SF311213355WA SF311213357WA SGF11440226WA SG211440133WA SG211450140WA SGF11440227WA SGF11440283WA SGF11460355WA SGF11460362WA SGF11460363WA SGF12142294WA SGF12142299WA SGF12142300WA SGF12142301WA SGF12152377WA SGF12142274WA SGF12212615WA SF310180676WA SF310180677WA SF310180679WA SF310180680WA SF310190706WA SF310190690WA SF310190696WA SF310180667WA SF310180675WA SF310180678WA SF310180681WA SF310180683WA SF310190697WA SF310220861WA SF311504736WA SF311504756WA SF311504758WA SF311504759WA SF311504754WA SF312155318WA SF312155320WA SF312195471WA SF312195472WA SF312195496WA SGF12414076WA SF312396199WA SGF12525210WA SG213031633WA SGF13025346WA SGF13025347WA SGF13025349WA SGF12273271WA SGF12273272WA SGF12313369WA SG212290766WA SG212290767WA SG212290768WA SGF12424257WA SGF12515102WA SGF12515103WA SGF12515105WA SGF12515106WA SGF12515108WA SGF12515111WA SGF12515114WA SGF12515119WA SGF12515141WA SGF12515142WA SG213081845WA SGF13085824WA SG213091847WA SG213091848WA SG213091849WA SG213091850WA SG213101938WA SGF13116163WA SGF13116165WA SGF13116166WA SGF13116167WA SGF13116168WA SGF13116169WA SGF13116170WA SGF13116171WA SGF13116172WA SGF13116175WA SGF13116176WA SGF13116177WA SGF13116178WA SGF13116179WA SGF13116180WA SGF13116181WA SGF13116182WA SGF13116183WA SGF13116184WA SGF13116185WA SGF13116186WA SGF13116187WA SGF13116188WA SGF13116189WA SGF13116190WA SGF13116191WA SGF13116192WA SGF13116193WA SGF13116195WA SGF13116196WA SGF13116197WA SGF13116198WA SGF13116199WA SGF13116200WA SGF13116201WA SGF13116202WA SGF13116203WA SGF13116220WA SGF13116222WA SGF13116223WA SF310321324WA SF310321326WA SF310321327WA SF310331330WA SF310331332WA SF310331334WA SF310331335WA SGF12323463WA SGF12323467WA SG212310810WA SG212310811WA SGF12283283WA SGF12323464WA SGF12273276WA SGF12363630WA SGF12363632WA SGF12363643WA SGF12363644WA SGF12363645WA SGF12363646WA SGF12363648WA SGF12363649WA SF312155288WA SF312155290WA SF312155291WA SF312155293WA SF312155294WA SF312155313WA
SF312155314WA SF312165357WA SF312165364WA SF312165367WA SF312165368WA SF312165369WA SF312165375WA SF312165382WA SF312165389WA SF311253660WA SF311253666WA SF311253673WA SF311253679WA SF311253681WA SF311253682WA SF311253683WA SF311253684WA SF311253685WA SF311253687WA SF311293870WA SF311293873WA SF311293874WA SF311293876WA SF311303905WA SF311303909WA SF311303911WA SF311303913WA SF311303915WA SF311303923WA SF311303927WA SF311303949WA SF311303963WA SF311303982WA SF311313998WA SF311303944WA SF311303984WA SF311313992WA SF311314012WA SF311314028WA SF311314044WA SF311314050WA SF311314061WA SF311314066WA SF311314074WA SF311314075WA SF311314083WA SF311314095WA SF311314125WA SF311334208WA SF311334216WA SF312034921WA SF312054952WA SF312054982WA SF312064997WA SF312065030WA SF312075036WA SF312075037WA SF312075041WA SF312075043WA SF312075046WA SF312085083WA SF312085084WA SF312085087WA SF312085089WA SF312085097WA SF312085098WA SF312085099WA SF312085100WA SF312085101WA SF312085102WA SF312085103WA SF312085104WA SF312085105WA SF312085106WA SF312085107WA SF312085108WA SF312085109WA SF312085110WA SF312085112WA SF312085114WA SF311303942WA SF312155275WA SF312205599WA SF312205602WA SF312205603WA SF312205611WA SF312205618WA SF311143071WA SF311143072WA SF311143106WA SF311143108WA SF311143109WA SF311143110WA SF311143111WA SF311143112WA SF311143113WA SF311143114WA SF311143115WA SF311143116WA SF311143117WA SF311143118WA SF311143119WA SF311143120WA SF311143121WA SF311143122WA SF311143123WA SF311143124WA SF311143125WA SF311143126WA SF311143127WA SF311143128WA SF311143129WA SF311143131WA SF311143133WA SF311153138WA SF311153139WA SF311153140WA SF311153141WA SF311153142WA SF312315819WA SF312315821WA SF312315822WA SF312315828WA SF312315830WA SF312315831WA SF312315808WA SF312315809WA SF312315813WA SF312315814WA SF312315815WA SF312315816WA SGF12424249WA SGF12424248WA SG212431093WA SG212431095WA SGF12424250WA SGF12424239WA SGF12424245WA SGF12414118WA SGF12424205WA SGF12424209WA SGF12424244WA SGF12424246WA SGF12484567WA SGF12484574WA SGF12484573WA SGF12484578WA SGF12484579WA SGF12484580WA SGF12484587WA SGF12484601WA SGF12484603WA SGF12484605WA SGF12484606WA SGF12484609WA SGF12484652WA SGF12484653WA SGF12484661WA SGF12484667WA SGF12484668WA SG212491355WA SGF12525196WA SGF12525217WA SGF12525218WA SGF12525220WA SGF12525224WA SGF13025265WA SGF13025266WA SGF13025267WA SGF13025269WA SGF13025276WA SGF13025270WA SGF13025271WA SGF13025275WA SGF13025277WA SGF13025281WA SGF13025285WA SGF13025287WA SGF13025288WA SGF13025272WA SGF13025279WA SGF13055641WA SGF13055697WA SGF13055698WA SGF13055699WA SGF13055700WA SGF13055701WA SGF13055702WA SGF13055706WA SGF13055707WA SGF13176731WA SF312466522WA SGF12424204WA SF312315812WA SG212521538WA SGF12525205WA SF311314089WA SF311314108WA SF311314109WA SF311314112WA SF311314116WA SF311324143WA SF311324153WA SF311324161WA SF311324175WA SF311334205WA SF311404360WA SF311404361WA SF311404371WA SF311414381WA SF311414386WA SF312466490WA SF312466498WA SF312526760WA SGF12504902WA SGF12525191WA SGF12525219WA SGF12525221WA SGF12525222WA SGF12525223WA SGF12525225WA SGF12525226WA SGF12525227WA SG212525228WA SGF12525228WA SGF12102047WA SGF12102034WA SGF12102040WA SGF12081714WA SGF12081717WA SGF12091829WA SGF12091831WA SGF12091832WA SGF12091833WA SGF12091834WA SG212240673WA SG212240682WA SG212250702WA SG212250703WA SG212360885WA SGF12273254WA SGF12273255WA SGF12273277WA SG211360037WA SG211470189WA SG211470199WA SG211490204WA SG212100403WA SG212100452WA SG212100453WA SG212100454WA SG212100455WA SG212100456WA SGF12061529WA SGF12061530WA SGF12061533WA SGF12081710WA SGF12081712WA SGF12081713WA SGF12101951WA SGF12101952WA SGF12101953WA SGF12101954WA SGF12101955WA SGF12101959WA SGF12101961WA SGF12101963WA SGF12101964WA SGF12162462WA SGF12162471WA SGF12232698WA SGF12232699WA SGF12232700WA SGF12232701WA SGF12232702WA SGF12232703WA SGF12232704WA SGF12232705WA SGF12232706WA SGF12232717WA SGF12232718WA SGF12232719WA SGF12232720WA SGF12232721WA SGF12232723WA SGF12232724WA SGF12232725WA SGF12232726WA SG212160596WA SG212160597WA SG212160598WA SG212230632WA SG212230633WA SG212230634WA SG212230635WA SG212230636WA SG212230637WA SG212230638WA SG212230639WA SG212230640WA SG212230641WA SG212230642WA SG212230643WA SG212230644WA SG212230645WA SG212230646WA SG212230647WA SG212230663WA SG212260736WA SG212310792WA SG212310793WA SG212310794WA SG212310795WA SG212310796WA SG212320842WA SG212330851WA SG212330852WA SG212330853WA SG212330854WA SGF12363591WA SGF12363592WA SGF12363593WA SGF12363610WA SGF12363611WA SGF12363612WA SGF12363613WA SGF12363614WA SGF12363615WA SGF12363616WA SGF12363617WA SGF12363618WA SGF12363620WA SGF12363622WA SGF12363623WA SGF12363624WA SGF12363625WA SGF12363626WA SGF12363627WA SGF12363631WA SGF12363639WA SGF12363640WA SGF12363606WA SGF12363607WA SGF12363609WA SGF12373691WA SG212330850WA SG212370901WA SG212370902WA SG212370905WA SG212370906WA SGF12061528WA SF311193325WA SF311213338WA SF311213339WA SF311213340WA SF311213343WA SF311213344WA SF311213345WA SF311213347WA SF311213348WA SF311213349WA SF311283750WA SF311283752WA SF311283754WA SF311283755WA SF311283757WA SF311283758WA SF311283765WA SF311283779WA SF311283789WA SF311283792WA SF311283794WA SF311283795WA SF311283796WA SF311283815WA SF311283819WA SF311283820WA SF311283821WA SF311283827WA SF311303921WA SF311313997WA SF310200751WA SF310200756WA SF310200757WA SF310200761WA SF310200770WA SF310220869WA SF310220878WA SF310220885WA SF310220886WA SF310220891WA SF310220892WA SF310230893WA SF310230894WA SF310230895WA SF310230896WA SF310230898WA SF310230899WA SF310230900WA SF310230936WA SF310230940WA SF310230944WA SF310230945WA SF310301197WA SF310301199WA SF310301200WA SF310301201WA SF310301202WA SF310301203WA SF310301204WA SF310301205WA SF310301208WA SF310311241WA SF310371591WA SF310371592WA SF310371593WA SF310371596WA SF310371598WA SF310371599WA SF310371600WA SF310371601WA SF310371602WA SF310371603WA SF310371604WA SF310371605WA SF310371595WA SF310371606WA SF310371607WA SF310371608WA SF310371609WA SF310371610WA SF310371611WA SF310371612WA SF310371613WA SF310371614WA SF311143093WA SF311143094WA SF311143095WA SF311143100WA SF311143101WA SF311143102WA SF311143104WA SF311143107WA SF311143130WA SF311143134WA SF311313996WA SF311314085WA SF311314113WA SF311314115WA SF311314119WA SF311324127WA SF311324128WA SF311324130WA SF311324140WA SF311324145WA SF311324147WA SF311324151WA SF311324152WA SF311324156WA SF311324163WA SF311324164WA SF311324167WA SF311324171WA SF311324172WA SF311324176WA SF311324178WA SF311324196WA SF311324197WA SF311324198WA SF310472232WA SF310472243WA SF310472262WA SF310472263WA SF310472264WA SF310472265WA SF310472266WA SF310472267WA SF310472268WA SF310482269WA SF312356004WA SF311474591WA SF312165393WA SF311303985WA SF311243584WA SF312245654WA SF312095171WA SF312165407WA SF311504766WA SF311253619WA SF311253637WA SF312085116WA SF312085115WA SF313106968WA SF312355998WA SF312335952WA SF312386126WA SF312195491WA SF312195493WA SF312205635WA SF312175428WA SGF11340047WA SGF11450325WA SGF11460366WA SGF11460397WA SGF11460398WA SGF11470415WA SGF11470418WA SGF11490469WA SGF11490492WA SGF11490493WA SGF11490495WA SGF11490496WA SGF11510619WA SGF11510646WA SGF11510649WA SGF11510657WA SGF11510677WA SGF11510704WA SGF11510705WA SGF11510719WA SGF11510721WA SGF11510722WA SGF11510729WA SGF11510730WA SGF11510731WA SGF11510732WA SGF11510733WA SGF11510734WA SGF11510735WA SGF11510736WA SGF11510737WA SGF11510738WA SGF11510793WA SGF11510794WA SGF11510795WA SGF11510796WA SGF11510797WA SGF11510798WA SGF11510799WA SGF11510800WA SGF11510801WA SGF11510802WA SGF11510803WA SGF11510804WA SGF11510805WA SGF11510806WA SGF11510808WA SGF11510833WA SGF11510834WA SGF11510836WA SF311344324WA SGF11490504WA SGF11500582WA SGF11510763WA SGF11510764WA SGF11510765WA SGF11510766WA SGF11510767WA SGF11510768WA SGF11510769WA SGF11510770WA SGF11510771WA SGF11510772WA SGF11510773WA SGF11510775WA SGF11510776WA SGF11510777WA SGF11510778WA SGF11510779WA SGF11510780WA SGF11510781WA SGF11510782WA SGF11510783WA SGF11510784WA SGF11510785WA SGF11510786WA SGF11510787WA SGF11510788WA SGF11510789WA SGF11510790WA SGF11510791WA SGF11510792WA SGF11510809WA SGF11510810WA SGF11510811WA SGF11510812WA SGF11510813WA SGF11510814WA SGF11510815WA SGF11510816WA SGF11510872WA SGF11510874WA SGF11510875WA SGF11510876WA SGF11510877WA SGF11510878WA SGF11510880WA SGF11510881WA SGF11510882WA SGF11510883WA SGF11510774WA SGF12011163WA SGF12011172WA SGF12011175WA SGF12011180WA SGF12011181WA SGF12011183WA SGF12021244WA SGF12021248WA SGF12051428WA SGF12051433WA SGF12061492WA SGF12061494WA SGF12061495WA SGF12061496WA SGF12061498WA SGF12061499WA SGF12061502WA SGF12081664WA SGF12081665WA SGF12081667WA SGF12081668WA SGF12081669WA SGF12081670WA SGF12081720WA SGF12091798WA SGF12091809WA SGF12091815WA SGF12091817WA SGF12091818WA SGF12091822WA SF312315866WA SGF11510827WA SF311424486WA SF311454567WA SGF12021240WA SGF12091825WA SGF11030004WA SGF11030006WA SGF11030008WA SGF11030012WA SGF11030013WA SGF11030010WA SGF11030011WA SGF11030001WA SGF11030009WA SG211430115WA SG211430116WA SG211430117WA SG211430118WA SG211430119WA SG211450141WA SG211450152WA SG211450153WA SG211450154WA SG211450155WA SG211450159WA SGF12031345WA SG211490210WA SG212020359WA SG212020361WA SGF12021224WA SGF12021226WA SG212150557WA SGF12152393WA SG212100402WA SG212150562WA SG212150569WA SG212150573WA SG212150578WA SG212160599WA SGF11080015WA SGF11080021WA SGF11080023WA SGF11090027WA SGF11090031WA SGF11090032WA SGF11090033WA SGF11090036WA SGF11080024WA SGF11090028WA SGF11090029WA SGF11090030WA SGF11090035WA SGF11090038WA SGF11090040WA SGF11090041WA SGF11080014WA SGF11080016WA SGF11080017WA SGF11080019WA SGF11080020WA SGF11080022WA SGF11090042WA SGF11080018WA SGF11090025WA SGF11090026WA SGF11090034WA SGF11090037WA SGF11090039WA SGF12273268WA SGF12383853WA SGF12383882WA SGF12393902WA SGF12393963WA SGF12393971WA SGF12393974WA SGF12383885WA SGF12393984WA SGF12393994WA SG212320844WA SG212320845WA SG212390918WA SG212310799WA SG212020365WA SG212020376WA SF312195504WA SG212160605WA SG212260738WA SG212270751WA SGF12313433WA SGF12313434WA SGF12313435WA SGF12313438WA SGF12313440WA SGF12313443WA SGF12313445WA SGF12313447WA SGF12323449WA SGF12323456WA SGF12373690WA SGF12383894WA SGF12393993WA SGF12393997WA SG212370904WA SF310120605WA SF310120609WA SG211210001WA SG211210002WA SG211210003WA SG211210004WA SG211210005WA SG211210006WA SG211210007WA SG211210008WA SG211210009WA SG211210010WA SG211210011WA SG211210012WA SG211210013WA SG211210014WA SG211210015WA SG211210016WA SG211210017WA SG211210018WA SG211210019WA SG211210020WA SG211210021WA SG211210022WA SG211210023WA SG211210024WA SG211210025WA SG211210026WA SG211210027WA SG211210028WA SG211210029WA SG211210030WA SG211210032WA SF310381677WA SF310391711WA SG211210031WA SG212020367WA SGF12051358WA SGF12011185WA SGF12021249WA SGF12061500WA SGF12091794WA SGF12091795WA SGF12091796WA SGF12091797WA SGF12091799WA SGF12091800WA SGF12091801WA SGF12091803WA SGF12091804WA SGF12091805WA SGF12091806WA SGF12091807WA SGF12091808WA SGF12091811WA SGF12091812WA SGF12091813WA SGF12091814WA SGF12091816WA SGF12091821WA SGF12091827WA SGF12091828WA SGF12091840WA SGF12091841WA SGF12091842WA
SGF12091843WA SGF12091845WA SGF12091847WA SGF12091849WA SGF12091850WA SGF12091851WA SGF12091852WA SGF12091853WA SGF12101966WA SGF12101967WA SGF12101968WA SGF12101975WA SGF12101976WA SGF12101980WA SGF12101981WA SGF12101982WA SGF12101984WA SGF12101990WA SGF12102006WA SGF12102010WA SGF12102013WA SGF11510817WA SGF11510818WA SGF11510819WA SGF11510820WA SGF11510821WA SGF11510822WA SGF11510823WA SGF11510824WA SGF11510825WA SGF11510826WA SGF11510828WA SGF11510829WA SGF11510830WA SGF11510831WA SGF11510873WA SGF11510879WA SGF11510884WA SGF11510885WA SGF11510886WA SGF11510887WA SGF11510888WA SGF11510889WA SGF11510890WA SGF11520893WA SGF11520894WA SGF11520981WA SGF11520982WA SGF11520983WA SGF11520984WA SGF11520986WA SGF11520987WA SGF11520988WA SGF11520989WA SGF11520990WA SGF11520991WA SGF11520992WA SGF11520993WA SGF11521061WA SGF11521062WA SGF11521064WA SGF11521065WA SGF11521066WA SGF11521067WA SGF11521068WA SGF11521069WA SGF11521070WA SGF11521071WA SGF11521072WA SGF11521073WA SGF11520895WA SGF11520896WA SGF11520897WA SGF11520898WA SGF11520899WA SGF11520985WA SGF11521063WA SGF11521074WA SGF11521075WA SGF11521076WA SGF11521077WA SGF11521078WA SGF11521079WA SGF11521080WA SGF11521081WA SGF11521082WA SGF11521084WA SGF11521085WA SGF11521086WA SGF11521087WA SGF12031323WA SGF12051354WA SGF12051359WA SGF12051368WA SGF12051369WA SGF12051370WA SGF12051375WA SGF12363653WA SGF12363654WA SGF12363655WA SGF12363656WA SGF12363657WA SGF12363686WA SGF12363687WA SGF12363688WA SGF12383907WA SGF12383908WA SGF12383909WA SGF12383910WA SGF12383911WA SGF12383912WA SGF12383913WA SGF12383914WA SGF12383915WA SGF12383916WA SGF12383919WA SGF12383921WA SGF12383924WA SGF12383926WA SGF12383927WA SGF12383929WA SGF12383931WA SGF12393938WA SGF12393939WA SGF12393940WA SGF12393941WA SGF12393942WA SGF12393943WA SGF12393944WA SGF12393945WA SGF12393946WA SGF12393947WA SGF12393948WA SGF12393949WA SGF12393950WA SGF12393951WA SGF12393952WA SGF12393953WA SGF12393954WA SGF12393955WA SGF12393956WA SGF12393957WA SGF12393958WA SGF12393959WA SGF12393960WA SGF12393961WA SGF12393962WA SF312105224WA SGF11480443WA SGF11510622WA SGF11510626WA SGF11510659WA SGF11510660WA SGF11510664WA SGF11510668WA SGF11510669WA SGF11510670WA SGF11510673WA SGF11510674WA SGF11510676WA SGF11510683WA SGF11510684WA SGF11510686WA SGF11510687WA SGF11510695WA SGF11510697WA SGF11510698WA SGF11510699WA SGF11510700WA SGF11510701WA SGF11510702WA SGF11510703WA SGF11510706WA SGF11510707WA SGF11510708WA SGF11510709WA SGF11510710WA SGF11510711WA SGF11510712WA SGF11510713WA SGF11510714WA SGF11510715WA SGF11510717WA SGF11510718WA SGF11510720WA SGF11510724WA SGF11510725WA SGF11510726WA SGF11510727WA SGF11510728WA SGF11510739WA SGF11510740WA SGF11510741WA SGF11510742WA SGF12253022WA SGF12253024WA SGF12253025WA SGF12253026WA SGF12253071WA SGF12253074WA SGF12253076WA SGF12253077WA SGF12253078WA SGF12253080WA SGF12253083WA SGF12253085WA SGF12253089WA SGF12253091WA SGF12253092WA SGF12253093WA SGF12253102WA SGF12253104WA SGF12253105WA SGF12253106WA SGF12253107WA SGF12253108WA SGF12253109WA SGF12253111WA SGF12253112WA SGF12253114WA SGF12253115WA SGF12253117WA SGF12253120WA SGF12253121WA SGF12253125WA SGF12253126WA SGF12253127WA SGF12263133WA SGF12263134WA SGF12263135WA SGF12263136WA SGF12263137WA SGF12263141WA SGF12263142WA SGF12263166WA SGF12263167WA SGF12263168WA SGF12263218WA SGF12263219WA SGF12263220WA SGF12263221WA SGF12263222WA SGF12263223WA SGF12263224WA SF312065028WA SGF11510672WA SGF11510675WA SGF11510685WA SGF11510696WA SGF11450343WA SGF11450344WA SGF11450346WA SGF11490497WA SGF11490498WA SGF11490499WA SGF11490500WA SGF11490501WA SGF11490503WA SGF11490505WA SGF11490506WA SGF11500547WA SGF11500548WA SGF11500558WA SGF11500559WA SGF11500560WA SGF11500561WA SGF11500571WA SGF11500572WA SGF11500573WA SGF11500574WA SGF11500578WA SGF11500579WA SGF11500583WA SGF11500584WA SGF11500585WA SGF11500586WA SGF11500587WA SGF11500588WA SGF11500589WA SGF11500590WA SGF11500591WA SGF11500592WA SGF11500593WA SGF11500594WA SGF11500595WA SGF11510744WA SGF11510745WA SGF11510746WA SGF11510748WA SGF11510751WA SGF11510752WA SGF11510753WA SGF11510754WA SGF11510755WA SGF11510759WA SGF11510760WA SGF11510761WA SGF12373754WA SGF12373758WA SGF12373759WA SGF12373760WA SGF12373761WA SGF12373762WA SGF12373763WA SGF12373765WA SGF12373766WA SGF12373768WA SGF12373769WA SGF12373770WA SGF12373772WA SGF12373774WA SGF12373775WA SGF12373777WA SGF12373779WA SGF12373780WA SGF12373781WA SGF12373782WA SGF12373784WA SGF12373785WA SGF12373786WA SGF12373787WA SGF12373788WA SGF12373790WA SGF12373791WA SGF12373792WA SGF12373794WA SGF12373795WA SGF12373796WA SGF12373798WA SGF12373799WA SGF12373800WA SGF12373803WA SGF12373809WA SGF12373813WA SGF12373814WA SGF12373818WA SGF12373819WA SGF11500581WA SGF11520926WA SGF11510853WA SGF12253012WA SGF12253013WA SGF12253014WA SGF12253016WA SGF12253018WA SGF12253019WA SGF12253020WA SGF12253021WA SGF12253023WA SGF12253028WA SGF12253033WA SGF12253034WA SGF12253035WA SGF12253037WA SGF12253044WA SGF12253045WA SGF12253046WA SGF12253053WA SGF12253055WA SGF12253057WA SGF12253060WA SGF12253062WA SGF12253063WA SGF12253064WA SGF12263138WA SGF12263139WA SGF12263140WA SGF12263143WA SGF12263144WA SGF12263145WA SGF12263146WA SGF12263147WA SGF12263148WA SGF12263149WA SGF12263150WA SGF12263151WA SGF12263152WA SGF12263153WA SGF12263154WA SGF12263155WA SGF12263156WA SGF12263157WA SGF12263158WA SGF12263159WA SGF12263160WA SGF12263161WA SGF12263162WA SGF12263163WA SGF12263164WA SGF12263165WA SGF12212618WA SGF12212620WA SGF12232758WA SGF12242852WA SGF12242853WA SGF12242854WA SGF12242855WA SGF12242856WA SGF12242857WA SGF12242865WA SGF12242866WA SGF12242867WA SGF12242869WA SGF12242870WA SGF12242871WA SGF12242872WA SGF12242873WA SGF12242874WA SGF12242875WA SGF12242876WA SGF12242877WA SGF12242878WA SGF12242892WA SGF12242893WA SGF12242894WA SGF12242895WA SGF12242896WA SGF12242897WA SGF12242898WA SGF12242899WA SGF12242900WA SGF12242901WA SGF12242902WA SGF12242903WA SGF12242908WA SGF12242909WA SGF12242910WA SGF12242921WA SGF12252987WA SGF12252990WA SGF12252993WA SGF12252995WA SGF12252996WA SGF12252997WA SGF12252998WA SGF12252999WA SGF12253000WA SGF12253001WA SGF12242905WA SGF12242906WA SGF12242907WA SGF12242911WA SGF12242912WA SGF12242913WA SGF12242914WA SGF12242915WA SGF12242916WA SGF12242917WA SGF12242918WA SGF12242919WA SGF12242920WA SGF12242922WA SGF12242923WA SGF12242924WA SGF12242925WA SGF12242926WA SGF12242927WA SGF12242928WA SGF12242929WA SGF12242930WA SGF12242931WA SGF12252964WA SGF12252965WA SGF12252966WA SGF12252967WA SGF12252968WA SGF12252969WA SGF12252970WA SGF12252971WA SGF12252972WA SGF12252973WA SGF12252974WA SGF12252975WA SGF12252976WA SGF12252977WA SGF12252978WA SGF12252979WA SGF12252980WA SGF12252981WA SGF12252982WA SGF12252983WA SGF12252984WA SGF12252985WA SGF12252986WA SGF12253009WA SGF12253010WA SGF12253011WA SGF12253017WA SGF11510839WA SGF11510840WA SGF11510841WA SGF11510842WA SGF11510843WA SGF11510844WA SGF11510846WA SGF11510847WA SGF11510848WA SGF11510850WA SGF11510851WA SGF11510852WA SGF11510854WA SGF11510855WA SGF11510856WA SGF11510857WA SGF11510858WA SGF11510859WA SGF11510860WA SGF11510861WA SGF11510862WA SGF11510866WA SGF11510868WA SGF11510869WA SGF11510870WA SGF11510871WA SGF11520904WA SGF11520905WA SGF11520906WA SGF11520907WA SGF11520908WA SGF11520909WA SGF11520910WA SGF11520912WA SGF11520913WA SGF11520914WA SGF11520915WA SGF11520917WA SGF11520918WA SGF11520919WA SGF11520920WA SGF11520921WA SGF11520923WA SGF11520924WA SGF11520925WA SGF11520928WA SGF11520929WA SGF11520930WA SGF11500567WA SGF11510845WA SGF11520916WA SGF11520922WA SGF11520932WA SGF11520933WA SGF11520934WA SGF11520936WA SGF11520937WA SGF11520938WA SGF11520939WA SGF11520940WA SGF11520941WA SGF11520942WA SGF11520943WA SGF11520944WA SGF11520945WA SGF11520946WA SGF11520947WA SGF11520953WA SGF11520954WA SGF11520955WA SGF11520957WA SGF11520959WA SGF11520961WA SGF11520964WA SGF11520966WA SGF11520967WA SGF11520968WA SGF11520969WA SGF11520970WA SGF11520971WA SGF11520972WA SGF11520974WA SGF11520976WA SGF11520994WA SGF11520995WA SGF11520996WA SGF11520997WA SGF11520998WA SGF11520999WA SGF11521000WA SGF11521001WA SGF11521002WA SGF11521003WA SGF11521004WA SGF11521028WA SGF11521055WA SGF11521057WA SGF11521058WA SF312245666WA SGF12313400WA SGF12313416WA SG212270749WA SGF11360064WA SGF11370068WA SGF11360065WA SGF11360067WA SGF12081672WA SGF12091819WA SGF12091820WA SGF12091823WA SGF12091824WA SGF12101986WA SGF12101993WA SGF12102016WA SGF12102019WA SGF12122180WA SGF12122181WA SGF12122182WA SGF12122183WA SGF12122184WA SGF12122185WA SGF12122186WA SGF12122187WA SGF12122188WA SGF12122189WA SGF12122190WA SGF12122191WA SGF12122192WA SGF12122193WA SGF12122194WA SGF12122195WA SGF12122196WA SGF12122197WA SGF12122198WA SGF12122199WA SGF12122200WA SGF12122201WA SGF12122202WA SGF12122203WA SGF12122204WA SGF12122222WA SGF12122223WA SGF12122224WA SGF12122225WA SGF12122226WA SGF12122227WA SGF12081671WA SGF12081673WA SGF12101987WA SGF12101996WA SGF12102007WA SGF12102008WA SGF12102009WA SGF12102011WA SGF12102012WA SGF12102015WA SF310492366WA SGF12383925WA SGF12393985WA SG212310801WA SG212370907WA SG212370908WA SG212370909WA SG212370910WA SGF12293336WA SG212370911WA SGF12393988WA SGF12393989WA SGF12393990WA SGF12393991WA SGF12393992WA SGF12393995WA SGF12393996WA SG212410926WA SG212410928WA SG212410929WA SG212410935WA SG212410937WA SG212410938WA SG212410939WA SG212410940WA SG212410942WA SG212410943WA SG212370903WA SGF12242904WA SGF12152356WA SGF12162447WA SGF12172480WA SGF12172495WA SGF12172510WA SGF12172535WA SGF12172557WA SGF12172561WA SGF12172577WA SGF12172583WA SGF12172606WA SGF12172607WA SGF12242763WA SGF12242766WA SGF12242768WA SGF12242770WA SGF12242774WA SGF12242775WA SGF12242777WA SGF12242778WA SGF12242779WA SGF12242780WA SGF12242782WA SGF12242783WA SGF12242785WA SGF12242787WA SGF12242788WA SGF12242789WA SGF12242790WA SGF12242794WA SGF12242796WA SGF12242805WA SGF12242807WA SGF12242809WA SGF12242811WA SGF12242817WA SGF12242819WA SGF12242820WA SGF12242821WA SGF12242826WA SGF12242882WA SGF12242884WA SGF12242885WA SGF12242887WA SGF12242891WA SGF12252937WA SGF12252938WA SGF12252939WA SGF12252940WA SGF12252941WA SGF12142284WA SGF12142323WA SGF12152395WA SGF12172501WA SGF12172543WA SGF12172566WA SGF12162475WA SGF12253002WA SGF12253003WA SGF12253004WA SGF12253005WA SGF12253006WA SGF12253007WA SGF12253008WA SGF12253015WA SGF12253027WA SGF12253030WA SGF12253031WA SGF12253032WA SGF12253036WA SGF12253039WA SGF12253041WA SGF12253042WA SGF12253050WA SGF12253052WA SGF12253059WA SGF12253061WA SGF12253067WA SGF12253069WA SGF12253070WA SGF12253072WA SGF12253073WA SGF12253075WA SGF12253079WA SGF12253081WA SGF12253082WA SGF12253084WA SGF12253086WA SGF12253087WA SGF12253088WA SGF12253090WA SGF12253094WA SGF12253095WA SGF12253096WA SGF12253097WA SGF12253098WA SGF12253099WA SGF12253100WA SGF12253101WA SGF12253103WA SGF12253110WA SGF12253116WA SGF12253118WA SGF12253119WA SGF12253122WA SGF12253123WA SGF12253124WA SGF12373704WA SGF12373722WA SGF12373730WA SGF12373789WA SGF12373804WA SGF12383826WA SGF12383827WA SGF12383828WA SGF12383829WA SGF12383834WA SGF12383835WA SGF12383837WA SGF12383838WA SGF12383841WA SGF12383842WA SGF12383843WA SGF12383844WA SGF12383845WA SGF12383848WA SGF12383850WA
SGF12383851WA SGF12383852WA SGF12383854WA SGF12383855WA SGF12383856WA SGF12383857WA SGF12383858WA SGF12383859WA SGF12383861WA SGF12383862WA SGF12383863WA SGF12383864WA SGF12383865WA SGF12383866WA SGF12383867WA SGF12383868WA SGF12383869WA SGF12383870WA SGF12383871WA SGF12383872WA SGF12383873WA SGF12383874WA SGF12383875WA SGF12383877WA SGF12383878WA SGF12383879WA SGF12383880WA SGF12383881WA SGF12383883WA SGF12383884WA SGF12061508WA SGF12061509WA SGF12061510WA SGF12071606WA SGF12071607WA SGF12071612WA SGF12071632WA SGF12071634WA SGF12071635WA SGF12071636WA SGF12071637WA SGF12071640WA SGF12071642WA SGF12071644WA SGF12071646WA SGF12071648WA SGF12071657WA SGF12071658WA SGF12081662WA SGF12081674WA SGF12081675WA SGF12081676WA SGF12081678WA SGF12081679WA SGF12081684WA SGF12081685WA SGF12081704WA SGF12071643WA SGF12081729WA SGF12081774WA SGF12091861WA SGF12091894WA SGF12091895WA SGF12091897WA SGF12091898WA SGF12091904WA SGF12091920WA SGF12091921WA SGF12091925WA SGF12091929WA SGF12091931WA SGF12091934WA SGF12091935WA SGF12091941WA SGF12091945WA SGF12091947WA SGF12091948WA SGF12061479WA SGF12061521WA SGF12071602WA SF311253662WA SF313137064WA SF311253671WA SF311253665WA SGF12383830WA SGF12383832WA SGF12383833WA SGF12383839WA SGF12383840WA SGF12383846WA SGF12383847WA SGF12383849WA SGF12383860WA SGF12383876WA SGF12383886WA SGF12383887WA SGF12383888WA SGF12383889WA SGF12383890WA SGF12383891WA SGF12383892WA SGF12383893WA SGF12393895WA SGF12393896WA SGF12393897WA SGF12393898WA SGF12393899WA SGF12393900WA SGF12393901WA SGF12393903WA SGF12393932WA SGF12393933WA SGF12393934WA SGF12393935WA SGF12393936WA SGF12393937WA SGF12393964WA SGF12393965WA SGF12393966WA SGF12393967WA SGF12393968WA SGF12393969WA SGF12393970WA SGF12393972WA SGF12393973WA SGF12393975WA SGF12393977WA SGF12393978WA SGF12393979WA SGF12393980WA SGF12393981WA SGF12393982WA SGF12393983WA SGF12393987WA SF312085130WA SF311283801WA SGF11500580WA SF311303979WA SF311293867WA SF312295773WA SGF12393976WA SGF12393986WA SF311243611WA SF311253686WA SF311042658WA SF311514801WA SGF12484650WA SG212491386WA SGF12505020WA SGF12505028WA SG212501434WA SG212501441WA SG212501450WA SF310361530WA SF310361534WA SF310341435WA SF311143096WA SF312245645WA SF309380046WA SF309410073WA SF309410074WA SF309410081WA SF309380043WA SF310060349WA SF310060353WA SF310060355WA SF310060356WA SF310060357WA SF310060361WA SF310060362WA SF310060370WA SF310060359WA SF310060365WA SF310060368WA SF310060374WA SF310060358WA SF310060376WA SF310060363WA SF310110582WA SF310110583WA SF310110585WA SF310110587WA SF310180632WA SF310180633WA SF310180636WA SF310180639WA SF310180640WA SF310180644WA SF310200758WA SF310190692WA SF310190698WA SF310190699WA SF310190703WA SF310190704WA SF310190705WA SF310190708WA SF310190713WA SF310190716WA SF310200750WA SF310200760WA SF310200766WA SF310200769WA SF310200772WA SF310200773WA SF310200753WA SF310190747WA SF310200759WA SF310200765WA SF310200767WA SF310220810WA SF310220813WA SF310220820WA SF310220824WA SF310220826WA SF310230928WA SF310230933WA SF310230937WA SF310230939WA SF310230941WA SF310230943WA SF310241014WA SF310241015WA SF310241018WA SF310241019WA SF310241020WA SF310241022WA SF310241023WA SF310241024WA SF310241025WA SF310241009WA SF310241010WA SF310241011WA SF310241013WA SF310271122WA SF310271124WA SF310311226WA SF310311237WA SF310311254WA SF310311265WA SF310311266WA SF310311269WA SF310311273WA SF310311274WA SF310311275WA SF310311277WA SF310311278WA SF310311286WA SF310311295WA SF310311297WA SF310271133WA SF310311258WA SF310301193WA SF310331346WA SF310331351WA SF310371594WA SF310371597WA SF310361438WA SF310361470WA SF310361482WA SF310361472WA SF310361476WA SF310371623WA SF310371629WA SF310371631WA SF310371632WA SF310371633WA SF310371644WA SF310371648WA SF310381649WA SF310381650WA SF310381652WA SF310381701WA SF310411763WA SF310411757WA SF310411765WA SF310411766WA SF310411776WA SF310411777WA SF310411778WA SF310411741WA SF310411747WA SF310411742WA SF310411774WA SF310431887WA SF310431892WA SF310431897WA SF310411770WA SF310421789WA SF310421790WA SF310421793WA SF310421795WA SF310421796WA SF310421797WA SF310421798WA SF310421799WA SF310421800WA SF310421814WA SF310421815WA SF310421816WA SF310421818WA SF310421819WA SF310421820WA SF310431942WA SF310431969WA SF310441978WA SF310431932WA SF310441988WA SF310441989WA SF310441993WA SF310431966WA SF310442015WA SF310452040WA SF310452045WA SF310452051WA SF310452052WA SF310452055WA SF310452024WA SF310452019WA SF310452023WA SF310452025WA SF310462177WA SF310482312WA SF310482317WA SF310482318WA SF310482326WA SF310482329WA SF310492345WA SF310482335WA SF310492340WA SF310512442WA SF310512433WA SF310502367WA SF310502372WA SF310502396WA SF310512414WA SF310512415WA SF310512416WA SF310512417WA SF310512423WA SF310512425WA SF310512426WA SF310512429WA SF310512430WA SF310502394WA SF311133051WA SF311193323WA SF311193329WA SF311223402WA SF311223403WA SF311223405WA SF311223407WA SF311223408WA SF311223409WA SF311223410WA SF311223412WA SF311223413WA SF311223414WA SF311223404WA SF311223417WA SF311223422WA SF311223423WA SF311223428WA SF311223429WA SF311223432WA SF311223433WA SF311223434WA SF311223435WA SF311233445WA SF311233437WA SF311233438WA SF311233439WA SF311233440WA SF311233441WA SF311233442WA SF311233444WA SF311233451WA SF311243598WA SF311243596WA SF311253674WA SF311263714WA SF311263720WA SF311263723WA SF311283785WA SF311283799WA SF311283769WA SF311283778WA SF311283808WA SF311283812WA SF311283818WA SF311293846WA SF311293847WA SF311293856WA SF311293888WA SF311293889WA SF311303903WA SF311293865WA SF311314035WA SF311314063WA SF311314067WA SF311314076WA SF311314092WA SF311314093WA SF311314098WA SF311314101WA SF311324138WA SF311324158WA SF311324159WA SF311324162WA SF311324174WA SF311324177WA SF311324188WA SF311324189WA SF311324199WA SF311334278WA SF311334290WA SF311334296WA SF311334302WA SF311324168WA SF311314107WA SF311324134WA SF311324148WA SF311324183WA SF311324190WA SF311334206WA SF311334212WA SF311334224WA SF311334261WA SF312065023WA SF312075038WA SF312075044WA SF312075045WA SF311494619WA SF312095176WA SF312165355WA SF312165358WA SF312165359WA SF312165363WA SF312165365WA SF312165366WA SF312165374WA SF312165376WA SF312165380WA SF312165390WA SF312165397WA SF312115245WA SF312115247WA SF312125272WA SF312125264WA SF312125266WA SF312125268WA SF312155347WA SF312165353WA SF312165354WA SF312165377WA SF312165378WA SF312165379WA SF312165383WA SF312165384WA SF312165387WA SF312165391WA SF312165392WA SF312165396WA SF312165398WA SF312165400WA SF312165401WA SF312165402WA SF312165403WA SF312165404WA SF312165405WA SF312165406WA SF312165408WA SF312165409WA SF312165410WA SF312165411WA SF312165412WA SF312165386WA SF312165394WA SF312175418WA SF312175429WA SF312175430WA SF312175420WA SF312195500WA SF312195506WA SF312245641WA SF312245646WA SF312265684WA SF312275703WA SF312275704WA SF312275705WA SF312275707WA SF312275699WA SF312275700WA SF312275702WA SF312275706WA SF312275708WA SF312396195WA SF312416201WA SGF12424210WA SGF12424214WA SF312416247WA SF312416248WA SF312416250WA SF312426255WA SGF12434299WA SGF12434300WA SGF12434301WA SGF12434302WA SF312436354WA SF312436355WA SGF12474419WA SGF12474427WA SF312466524WA SF312466533WA SF312466537WA SF312466569WA SGF12494808WA SGF12494750WA SGF12504834WA SGF12504845WA SGF12504848WA SF312496641WA SGF12504992WA SF312516733WA SF312516741WA SF312516742WA SGF12515155WA SF313026798WA SGF13035392WA SGF13035395WA SGF13035399WA SGF13035401WA SGF13055620WA SGF13085815WA SGF13085817WA SGF13085894WA SF313106973WA SF313106974WA SGF13106024WA SGF13106026WA SGF13106029WA SGF13106030WA SGF13106031WA SGF13106032WA SGF13106033WA SGF13106034WA SGF13106035WA SGF13106036WA SGF13106039WA SGF13176612WA SGF13176613WA SJF13180143WA SF310321320WA SJF13010006WA SJF13010014WA SJF13010017WA SJF13010018WA SJF13010020WA SJF13020024WA SJF13020032WA SJF13020033WA SJF13020037WA SJF13020038WA SJF13020039WA SJF13020040WA SJF13020041WA SJF13020043WA SJF13020044WA SJF13030045WA SJF13030046WA SJF13030051WA SJF13010015WA SJF13010016WA SJF13010019WA SJF13010021WA SJF13020027WA SJF13020030WA SJF13020031WA SJF13020034WA SJF13020035WA SJF13020036WA SF313147068WA SF313147069WA SF313177115WA SJF13030048WA SJF13030049WA SGF13035376WA SGF13035426WA SF311474575WA SF311474578WA SF311474594WA SGF11030003WA SF313117020WA SF310381699WA SF312446357WA SF312335943WA SF312335945WA SF312335965WA SF312335971WA SGF13136469WA SGF13176670WA SGF13176674WA SGF13176676WA SF310371582WA SF310371586WA SF310371588WA SF310371589WA SF310371615WA SF310371590WA SF310381662WA SF310381664WA SF310381665WA SF310381667WA SF310381669WA SF310381695WA SF310381696WA SF310381697WA SF310381698WA SF310431916WA SF310431935WA SF310431949WA SF310431951WA SF310431952WA SF310431953WA SF310431955WA SF310431956WA SF310431957WA SF310431958WA SF310431959WA SF310431961WA SF310431962WA SF310431963WA SF310431968WA SF310462149WA SF310462189WA SF310502377WA SF310502380WA SF310502381WA SF310502390WA SF310502393WA SF310502382WA SF310502385WA SF310502376WA SF310502378WA SF310502379WA SF310502383WA SF310502391WA SF310502374WA SF310502375WA SF310431954WA SF310462192WA SF310512434WA SF310512445WA SF310512458WA SF310512471WA SF310512469WA SF311082859WA SF311153162WA SF311153163WA SF311153169WA SF311213341WA SF311213342WA SF311213346WA SF311213350WA SF311213351WA SF311213358WA SF311213359WA SF311213363WA SF311213364WA SF311213366WA SF311213367WA SF311253608WA SF311253610WA SF311253611WA SF311253613WA SF311253614WA SF311243537WA SF311243543WA SF311243546WA SF311243547WA SF311243603WA SF311243601WA SF311253636WA SF311253620WA SF311253621WA SF311253616WA SF311253639WA SF311253640WA SF311253642WA SF311253643WA SF311253644WA SF311253646WA SF311253648WA SF311253650WA SF311253651WA SF311253652WA SF311253645WA SF311253647WA SF311253649WA SF311253653WA SF311253654WA SF311253655WA SF311253656WA SF311253657WA SF311253658WA SF311253659WA SF311253661WA SF311253664WA SF311253667WA SF311253668WA SF311253669WA SF311253670WA SF311253672WA SF311253675WA SF311253677WA SF311253678WA SF311253689WA SF311283781WA SF311263727WA SF311283773WA SGF12414077WA SF311263733WA SF311314102WA SF311314106WA SF311314111WA SF311314123WA SF311324154WA SF311334204WA SF311334248WA SF311344338WA SF311263712WA SF311263736WA SF311263739WA SF311283744WA SF311283747WA SF311283749WA SF311303947WA SF311303961WA SF311314065WA SF311314086WA SF311303904WA SF311303926WA SF311303966WA SF311303977WA SF311313991WA SF311313994WA SF311313999WA SF311314013WA SF311263732WA SF311283741WA SF311283745WA SF311283751WA SF311283753WA SF311293878WA SF311283805WA SF311303967WA SF311283777WA SF311263738WA SF311283810WA SF311283811WA SF311283823WA SF311283829WA SF311293868WA SF311303899WA SF311303914WA SF311303916WA SF311303918WA SF311303952WA SF311263729WA SF311263730WA SF311263731WA SF311314014WA SF311293850WA SF311293853WA SF311293857WA SF311293859WA SF311293883WA SF311293885WA SF311303988WA SF311314020WA SF311314038WA SF311314048WA SF311334250WA SF311303935WA SF311313995WA SF311303938WA SF311293855WA SF311314079WA SF311314062WA SF311314096WA SF311314071WA SF311314031WA SF311314072WA SF311314026WA SF312416217WA SF312416242WA SF312426254WA SF311314034WA SF311303989WA SF311314011WA SF311314029WA
SF311314051WA SF311314053WA SF311314055WA SF311314077WA SF311314078WA SF311314081WA SF311314082WA SF311314090WA SF311314094WA SF311314100WA SF311334245WA SF312446383WA SF312446384WA SF312446387WA SF312446388WA SF312446390WA SF312446391WA SF311324150WA SF311324129WA SF311324200WA SF311324135WA SF311324137WA SF311303901WA SF311324144WA SF311324149WA SF311324173WA SF311324180WA SF311324181WA SF311324192WA SF311334213WA SF311334217WA SF311334222WA SF311334236WA SF311344311WA SF311324179WA SF311324191WA SF312446381WA SF312446389WA SF312446398WA SF312446399WA SF311324142WA SF311324146WA SF311334207WA SF311314018WA SF312526775WA SF313026795WA SF313026801WA SF313026812WA SF313026813WA SF313036816WA SF311303981WA SF311324155WA SF311324157WA SF311334228WA SF311334257WA SF311414408WA SF311414409WA SF311414412WA SF311414413WA SF311424476WA SF311424478WA SF311424483WA SF311424487WA SF311414436WA SF311414411WA SF311424472WA SF311414427WA SF311414455WA SF311434546WA SF311404380WA SF311414402WA SF311414415WA SF311414422WA SF311414426WA SF311424489WA SF311494694WA SF311494696WA SF311504752WA SF311514795WA SF311514805WA SF311514807WA SF311514809WA SF311514812WA SF311514813WA SF311514814WA SF311514816WA SF311514817WA SF311514819WA SF311514820WA SF311514821WA SF311514822WA SF311514824WA SF311514825WA SF311514826WA SF311514827WA SF311514828WA SF311514829WA SF311514830WA SF311514831WA SF311514832WA SF311514833WA SF311514834WA SF311504774WA SF311504780WA SF311504775WA SF311504757WA SF311504776WA SF311504782WA SF311504783WA SF311524854WA SF311504771WA SF311514786WA SF311524835WA SF311524849WA SF311524851WA SF311524853WA SF312024859WA SF312024861WA SF312024875WA SF312024876WA SF312024887WA SF311524837WA SF311524852WA SF312024894WA SF312024896WA SF312034935WA SGF12504979WA SGF12504980WA SGF12504988WA SGF12504994WA SGF12504995WA SGF12504999WA SGF12505000WA SF312496618WA SF312496621WA SF312496623WA SF312496624WA SF312496625WA SF312496626WA SF312496627WA SF312496646WA SF312496649WA SF312496651WA SF312496655WA SF312496656WA SF312496658WA SF312496659WA SF312496661WA SF312496662WA SF312496664WA SF312496665WA SF312065008WA SF312065011WA SF312054942WA SF312054975WA SF312054988WA SF312054990WA SF312065031WA SF312075033WA SF312075034WA SF312075035WA SF312075047WA SF312075048WA SF312075051WA SF312085056WA SF312085121WA SF312526744WA SF312526747WA SF312526748WA SF312526750WA SF312526751WA SF312526753WA SF312526754WA SF312085120WA SF312095133WA SF312095137WA SF312095140WA SF312095141WA SF312095149WA SF312095178WA SF312095200WA SF312105208WA SGF12504903WA SF312085067WA SF312085071WA SF312085127WA SF312085131WA SF312095151WA SF312085122WA SF312085118WA SF312105209WA SF312105226WA SF312125256WA SF312125260WA SF312125261WA SF312125263WA SF312125265WA SF312125257WA SF312125262WA SF312105218WA SF312105222WA SF312125255WA SF312115246WA SF312155277WA SF312526767WA SF312526769WA SF313016782WA SF312155333WA SF312155348WA SF312165360WA SF312165361WA SF312155325WA SF312175427WA SF312195475WA SF312195477WA SF312175426WA SF312175434WA SF312175438WA SF312205543WA SF312205565WA SF312205555WA SF312155323WA SF312195516WA SF312205531WA SF312205538WA SF312205541WA SF312205553WA SF312205559WA SF312205571WA SF312205573WA SF312205574WA SF312205575WA SF312205596WA SF312205621WA SF312205637WA SF312205594WA SF312205612WA SF312195521WA SF312195522WA SF312195517WA SF312205582WA SF312195524WA SF312205554WA SF312195474WA SF312195476WA SF312195483WA SF312195484WA SF312195485WA SF312195499WA SF312195503WA SF312195511WA SF312195512WA SF312195518WA SF312205569WA SF312205617WA SF312205623WA SF312245657WA SF312245660WA SF312195497WA SF312195502WA SF312195514WA SF312195525WA SF312195527WA SF312205568WA SF312205570WA SF312205600WA SF312205629WA SF312195494WA SF312195490WA SF312195492WA SF312195513WA SF312195520WA SF312205528WA SF312205529WA SF312205601WA SF312205614WA SF312205625WA SF312205626WA SF312205630WA SF312205633WA SF312205636WA SF312245649WA SF312205581WA SF312205610WA SF312205615WA SF312205619WA SF312245650WA SF312245658WA SF312245663WA SF312205579WA SF312245642WA SF312245662WA SF312245647WA SF312245648WA SF312195507WA SF312205632WA SF312205634WA SF312245651WA SF312245656WA SF312245659WA SF312245661WA SF312245665WA SF312245667WA SF312245668WA SF312295774WA SF312315842WA SF312315852WA SF312315856WA SF312325888WA SF312335944WA SF312335950WA SF312335972WA SF312335979WA SF312335939WA SF312335974WA SF312335959WA SF312335961WA SF312335964WA SF312356010WA SF312356012WA SF312356015WA SF312356017WA SF312356007WA SF312356009WA SF312356013WA SF312356020WA SF312356021WA SF312356022WA SF312356024WA SF312356025WA SF312366034WA SF312366050WA SF312366030WA SF312366037WA SF312366038WA SF312366041WA SF312366044WA SF312366052WA SF312366053WA SF312366054WA SF312366055WA SF312376066WA SF312376069WA SF312376071WA SF312376072WA SF312366029WA SF312376108WA SF312376118WA SF312376062WA SGF13116240WA SGF13116244WA SGF13116236WA SGF13116239WA SGF13116242WA SF312376115WA SF312376119WA SF312356023WA SF312376076WA SF312376080WA SF312376085WA SF312376086WA SF312376087WA SF312376093WA SF312376123WA SF312376124WA SF312386127WA SF312386128WA SF312386129WA SF312386131WA SF312376090WA SF312396141WA SF312356014WA SF312386132WA SF312396135WA SF312396138WA SF312436294WA SGF12474478WA SGF12474479WA SGF12484687WA SGF12484688WA SGF12484689WA SGF12484690WA SGF12484691WA SGF12494694WA SGF12494695WA SGF12494696WA SGF12494697WA SGF12494698WA SGF12494737WA SGF12494738WA SF312446403WA SF312446405WA SF312446406WA SF312446407WA SF312446408WA SF312446409WA SF312446410WA SF312456422WA SF312456424WA SF312456425WA SF312456426WA SF312456427WA SF312456441WA SF312456484WA SGF13176662WA SF312466526WA SF312476596WA SF312496605WA SF312496640WA SF312456453WA SF312466544WA SF312506712WA SF312506722WA SF312506706WA SF312506709WA SF313036818WA SF313036834WA SF313046835WA SF313046839WA SF313046843WA SF313046847WA SF313036822WA SF313036823WA SF313036831WA SF313046844WA SF313046845WA SF313046848WA SF313046850WA SF313046851WA SF313046852WA SF313046853WA SF313056888WA SF313056907WA SF313066929WA SF313066935WA SF313066938WA SF313066945WA SF313106960WA SF313106962WA SF313106969WA SF313117006WA SF313117012WA SF313117008WA SF313117037WA SF313117039WA SF313117040WA SGF13126298WA SGF13126330WA SGF13126331WA SGF13126332WA SGF13126333WA SGF13126334WA SGF13126335WA SGF13126336WA SGF13126337WA SGF13126338WA SGF13126339WA SGF13126342WA SGF13126343WA SGF13126344WA SGF13126345WA SGF13126346WA SGF13126348WA SGF13126349WA SGF13126351WA SGF13126371WA SGF13126372WA SGF13126383WA SGF13126392WA SGF13126401WA SGF13126402WA SGF13126403WA SGF13126405WA SGF13126406WA SGF13126407WA SGF13126408WA SGF13126410WA SGF13126411WA SGF13126412WA SGF13126413WA SGF13126414WA SGF13126415WA SGF13126416WA SGF13126417WA SGF13126418WA SGF13126419WA SGF13126420WA SGF13126421WA SGF13126422WA SGF13126427WA SGF13126428WA SGF13126429WA SGF13126430WA SGF13126431WA SGF13126434WA SGF13126440WA SGF13116292WA SGF13126340WA SGF13126341WA SGF13126350WA SGF13126355WA SGF13126370WA SGF13126388WA SGF13126391WA SGF13126399WA SGF13126400WA SGF13126432WA SGF13126433WA SGF13126435WA SGF13126436WA SGF13126437WA SGF13126438WA SGF13126439WA SGF13136441WA SGF13136443WA SGF13136447WA SGF13136449WA SGF13136450WA SGF13136452WA SGF13136455WA SGF13136457WA SGF13136459WA SGF13136460WA SGF13136461WA SGF13136463WA SGF13136464WA SGF13136465WA SGF13136466WA SGF13136467WA SGF13136468WA SGF13136470WA SGF13136471WA SGF13136472WA SGF13136473WA SGF13136474WA SGF13136475WA SGF13136476WA SGF13136477WA SGF13136478WA SGF13136479WA SGF13136480WA SGF13136481WA SGF13136482WA SGF13136484WA SGF13136491WA SGF13136492WA SF313096950WA SF313096953WA SF313096954WA SF313096956WA SF313096958WA SF313096959WA SF313117030WA SF313117033WA SF313117034WA SF313117035WA SF313117036WA SF313117038WA SF313096955WA SF313096957WA SF313117032WA SF313117044WA SF313177113WA SGF13176697WA SF313147079WA SF313147072WA SGF13126409WA SGF13176559WA SGF13176560WA SGF13176562WA SGF13176563WA SGF13176568WA SGF13176570WA SGF13176571WA SGF13176576WA SF313177116WA SF313177120WA SF310361449WA SF310371584WA SF310371585WA SF312195463WA SF310331391WA SF310331392WA SF312095147WA SF312095148WA SGF12525184WA SGF12525190WA SF312315859WA SF312315865WA SF312325886WA SF312325891WA SF312325894WA SF312325896WA SF312325898WA SF310020240WA SF311494670WA SF311504760WA SF311504764WA SF311504765WA SF312526761WA SF312526762WA SF312526763WA SF312526764WA SF312526765WA SF311072815WA SF311082816WA SGF12504977WA SGF12504978WA SGF12504982WA SGF12504993WA SGF12504996WA SGF12504997WA SGF12504998WA SGF12505001WA SGF12505002WA SGF12505003WA SGF12505004WA SGF12505005WA SGF12505006WA SGF12505007WA SGF12505008WA SGF12505009WA SGF12505011WA SGF12505012WA SGF12505013WA SGF12505016WA SGF12505022WA SGF12505035WA SGF12505036WA SGF12505038WA SGF12515054WA SGF12515060WA SGF12515061WA SGF12515063WA SGF12515068WA SGF12515071WA SGF12515072WA SGF12515073WA SGF12515074WA SGF12515075WA SGF12515104WA SGF12515109WA SGF12515110WA SGF12515127WA SGF12515143WA SGF12515144WA SGF12515145WA SGF12515146WA SGF12515147WA SGF12515148WA SGF12515149WA SGF12515150WA SGF12515151WA SGF12515152WA SGF12515156WA SGF12525197WA SF311072763WA SF311082860WA SF311112887WA SF310502401WA SF310502405WA SF310502397WA SF310502398WA SF310502399WA SF310502400WA SF310502402WA SF310502403WA SF310502404WA SF310502406WA SF310512407WA SF310512409WA SF310512410WA SF310512412WA SF310512474WA SF310512475WA SF311072790WA SF310512411WA SF311112895WA SF311112897WA SF311112930WA SF311112936WA SF312335968WA SF312335969WA SF312335970WA SF312335975WA SF312335978WA SF312024871WA SF312024895WA SF312024867WA SF312024883WA SF312024882WA SF312024888WA SF312024889WA SF312024892WA SF312526779WA SJF13030050WA SF313117026WA SF311072814WA SF311072806WA SF311072809WA SF311072811WA SF311072812WA SF311082817WA SF311082818WA SF311434531WA SF313016788WA SF311072813WA SF311072776WA SF311072783WA SF311072784WA SF311072787WA SF312386133WA SF312396134WA SF312396136WA SGF12273270WA SGF12263212WA SGF12283284WA SGF12293327WA SGF12293341WA SGF12293342WA SGF12293344WA SGF12293349WA SGF12293355WA SGF12293360WA SGF12313368WA SGF12313371WA SGF12313373WA SGF12313375WA SGF12313376WA SGF12313382WA SGF12313386WA SGF12313388WA SGF12323498WA SGF12333499WA SGF12333500WA SGF12333501WA SGF12333503WA SGF12333505WA SGF12333507WA SGF12333509WA SGF12333516WA SGF12333517WA SGF12333520WA SGF12333521WA SGF12333523WA SGF12333526WA SGF12333528WA SGF12333533WA SGF12333534WA SGF12333539WA SGF12263183WA SGF12263193WA SGF12263200WA SGF12273242WA SGF12273243WA SGF12273244WA SGF12273245WA SGF12273246WA SGF12273247WA SGF12273248WA SGF12273259WA SGF12273262WA SGF12273266WA SGF12283291WA SGF12283294WA SGF12283295WA SGF12283301WA SGF12283309WA SGF12293329WA SGF12293354WA SGF12323465WA SGF12323466WA SGF12323475WA SGF12323476WA SGF12323477WA SGF12323478WA SGF12323479WA SGF12323480WA SGF12323481WA SGF12323482WA SGF12323483WA SGF12323484WA SGF12323485WA SGF12323489WA SGF12323491WA SF311143075WA SF311474579WA SF311494615WA SF311504762WA SF311524855WA SF312024890WA SF312085124WA SF312095174WA SF312095179WA SF312095192WA SF312095193WA SF312095203WA SF312095205WA SF312095206WA SF312095207WA
SF312105223WA SF312105225WA SF312105234WA SF312115235WA SF312115237WA SF312115238WA SF312115248WA SF312115253WA SF312125267WA SF312125269WA SF312155273WA SF311263716WA SF311263718WA SF311263734WA SF311263735WA SF312095180WA SF312095182WA SF312095187WA SF312095191WA SF312095194WA SF312095196WA SF312095204WA SF312105213WA SF312125270WA SF312125271WA SF312155317WA SF312155319WA SF312155321WA SF312155322WA SF312155324WA SF312155327WA SF312155334WA SF312155335WA SF312155336WA SF312155337WA SF312155338WA SF312155339WA SF312155340WA SF312155341WA SF312155342WA SF312155343WA SF312155344WA SF312155345WA SF312155346WA SF312155349WA SF312155350WA SF312155351WA SF312155352WA SF312165356WA SF312195509WA SGF12484617WA SGF12484620WA SGF12484622WA SGF12484624WA SF310100533WA SF310100544WA SF310100551WA SF311052684WA SF311052691WA SF311052693WA SF311243605WA SF311283742WA SF311283743WA SF311283746WA SF311283771WA SF311313993WA SF311303956WA SF311293877WA SF311314110WA SGF12132264WA SGF12242795WA SGF12242798WA SGF12242810WA SGF12242815WA SGF12373748WA SGF12373751WA SGF12373753WA SGF12373755WA SF310492342WA SF310492351WA SF310492352WA SF310492354WA SF309410078WA SF310431919WA SGF13085885WA SG211500214WA SG211500217WA SGF11420161WA SGF11480434WA SGF11490507WA SGF11490508WA SGF11490509WA SGF11510807WA SGF11510832WA SGF11510835WA SGF11510838WA SG212290769WA SGF11420138WA SGF12061547WA SF311233452WA SF311233453WA SF311233464WA SF311233466WA SF310090498WA SF310090500WA SF310090501WA SF310090505WA SF310090507WA SG212431098WA SF309410089WA SF310421836WA SF310431894WA SF310431890WA SF310452072WA SF310462157WA SF310462161WA SF310462163WA SF310462168WA SF310462169WA SF311263725WA SF311283822WA SGF12494707WA SG212501439WA SG213031658WA SG213031661WA SG213031663WA SF311414446WA SF310030287WA SF310050327WA SF310050328WA SF310050332WA SF310050333WA SJF13170132WA SJF13170139WA SJF13170136WA SG212010283WA SG211520268WA SG211520273WA SGF12011177WA SG212050390WA SG212010290WA SG212010296WA SGF12373793WA SGF12373805WA SG212050398WA SG212260709WA SG212260710WA SG212260712WA SG213031635WA SG213031642WA SF310431886WA SF310431880WA SF310431885WA SF310482321WA SF310482322WA SF310482323WA SF310482319WA SF310482320WA SF310482332WA SF311253633WA SF311253634WA SF311313990WA SF311314040WA SF311514797WA SF311514798WA SF311514803WA SF311514804WA SF311514806WA SF311514808WA SF311514810WA SF311514815WA SF311504779WA SF311514790WA SF311514791WA SF311514793WA SF311514796WA SF311514799WA SF311514800WA SF311514802WA SF311514811WA SF311514823WA SF313066932WA SGF12504989WA SG212481224WA SG212481226WA SG212511533WA SF311082839WA SJF13010005WA SF311283798WA SF311283787WA SGF12383824WA SGF12383825WA SG212290773WA SG212290774WA SG212290775WA SG212320831WA SG212320833WA SG212320837WA SG212320838WA SG212320839WA SG212360866WA SG212360870WA SGF12152366WA SGF12152369WA SGF12152378WA SGF12424152WA SF311233457WA SGF13176572WA SGF13176573WA SG213172124WA SG213172145WA SG213172146WA SGF13095951WA SGF13095955WA SGF13095956WA SGF13095957WA SGF13095958WA SGF13095959WA SGF13095960WA SGF13095961WA SGF13095966WA SGF13095976WA SGF13095979WA SGF13095981WA SGF13095982WA SGF13095998WA SGF13095999WA SGF13096000WA SGF13096001WA SGF13096002WA SGF13096003WA SGF13096005WA SGF13096006WA SGF13096007WA SGF13096008WA SGF13096009WA SG213051753WA SG213051754WA SG213051762WA SG213051763WA SG213051764WA SG213051765WA SG213051766WA SG213051767WA SG213051768WA SG213051769WA SG213051770WA SG213051772WA SG213051773WA SG213051774WA SG213051775WA SGF13055672WA SGF13055674WA SGF13055675WA SGF13055676WA SGF13055677WA SGF13055678WA SGF13055679WA SGF13055680WA SGF13055681WA SGF13055682WA SGF13055683WA SGF13055684WA SGF13055685WA SGF13055686WA SGF13055687WA SGF12242860WA SGF12252947WA SGF12252948WA SGF12252949WA SGF12252950WA SGF12484535WA SGF12484536WA SGF12484537WA SGF12484538WA SGF12484539WA SGF12484540WA SGF12484541WA SGF12484542WA SGF12484543WA SGF12484548WA SF310462133WA SF310462134WA SF310462138WA SF310462155WA SF310462172WA SF310462173WA SF311283793WA SF311283797WA SF311283800WA SF311283802WA SF311283803WA SF311283804WA SF311283806WA SF311303948WA SF311303954WA SF311303964WA SF311494625WA SF311494626WA SF311494631WA SF311494632WA SF311494640WA SF311494642WA SF311494643WA SF312285753WA SF312285758WA SF312285759WA SF312285750WA SF312285751WA SG211490209WA SG211490212WA SG212020366WA SG212020369WA SG212020370WA SG212020379WA SGF11510647WA SGF11510661WA SGF11510663WA SG211400077WA SG212050412WA SGF12061501WA SGF12061532WA SGF12142298WA SG211490203WA SGF12071616WA SGF12071623WA SGF12071627WA SGF12101985WA SGF12101998WA SGF11510645WA SGF12142229WA SGF12142295WA SGF12242868WA SGF12263236WA SGF12414074WA SGF12414075WA SF310240969WA SF310240972WA SF310240973WA SF310240984WA SF310240985WA SF310240986WA SF310240987WA SF310240988WA SF310240993WA SF310240994WA SF310240995WA SF310240996WA SF310240997WA SF310240998WA SF310240999WA SF310241000WA SF310241001WA SF310241002WA SF310241007WA SF310241008WA SF311334282WA SF311344317WA SF311484596WA SF311494656WA SGF12414088WA SGF12424240WA SGF12424247WA SGF12424251WA SGF12424252WA SGF12424253WA SGF12434270WA SGF12434271WA SGF12434272WA SGF12434273WA SGF12434274WA SGF12434275WA SGF12434276WA SGF12434277WA SGF12434278WA SGF12434279WA SGF12434280WA SGF12434281WA SGF12434282WA SGF12434283WA SGF12434284WA SGF12434285WA SGF12434286WA SGF12434287WA SGF12434288WA SGF12434289WA SGF12434290WA SGF12434291WA SGF12434292WA SGF12434293WA SF312275723WA SF312275732WA SF312275736WA SF312275738WA SF312275739WA SF312275740WA SF312275741WA SF312275742WA SF312275743WA SF312275744WA SF312275747WA SF312285752WA SF312285757WA SF312285761WA SF312285762WA SF312285763WA SF312285764WA SF312285765WA SF312285766WA SF312285767WA SF312376070WA SF312376079WA SF312376081WA SF312376082WA SF312376111WA SF312376122WA SF312396137WA SF312396140WA SF312396142WA SF312396143WA SF312396144WA SF312396145WA SF312396146WA SF312396147WA SF312396148WA SF312396149WA SF312396150WA SF312396151WA SF312396152WA SF312396153WA SF312396169WA SF312396172WA SF312396196WA SGF12474410WA SGF12474415WA SGF12474416WA SGF12474417WA SG212491294WA SG212491305WA SG212491308WA SG212491311WA SG212491312WA SGF12474418WA SGF12484647WA SGF12484655WA SGF12484660WA SGF12484662WA SGF12484663WA SG212481234WA SG212481236WA SG212481237WA SG212481238WA SG212481245WA SG212491376WA SG212491380WA SGF12494799WA SGF12494800WA SGF12494801WA SGF12494802WA SGF12494803WA SGF12494804WA SGF12494806WA SGF12494807WA SGF12494809WA SGF12494810WA SGF12494811WA SGF12494816WA SGF12494760WA SGF12494761WA SGF12494762WA SGF12494766WA SGF12504874WA SGF12504875WA SGF12504876WA SGF12504877WA SGF12504878WA SGF12504879WA SGF12504880WA SGF12504881WA SGF12504882WA SGF12504883WA SGF12515080WA SGF12515081WA SGF12515083WA SGF12515084WA SGF12515085WA SGF12515086WA SGF12515087WA SGF12515088WA SGF12515090WA SGF12515091WA SGF12515092WA SGF12515094WA SGF12515095WA SGF12515096WA SGF12515097WA SGF12515098WA SG212421050WA SG212421056WA SG212491392WA SG212491393WA SG212491394WA SG212491395WA SG212491396WA SG212491397WA SG212491398WA SG212491399WA SG212491410WA SGF12505010WA SGF12505017WA SGF12505018WA SGF12505021WA SGF12515133WA SGF12515130WA SGF12515132WA SGF13045591WA SGF13045593WA SGF13095950WA SGF13095952WA SGF13095953WA SGF13095954WA SGF13095963WA SGF13095977WA SGF13095978WA SGF13095980WA SG213041713WA SG213041716WA SG213041717WA SG213041720WA SGF13035433WA SGF13035436WA SGF13035438WA SGF13045559WA SGF13045560WA SGF13045562WA SGF13045563WA SGF13045564WA SG213031662WA SG213031664WA SG213031665WA SG213031666WA SG213031667WA SG213051746WA SG213051747WA SG213051748WA SG213051749WA SG213051750WA SGF13116206WA SGF13116207WA SGF13116208WA SGF13116211WA SGF13116212WA SGF13116213WA SGF13116214WA SGF13116215WA SGF13116216WA SGF13116217WA SGF13116218WA SGF13116219WA SGF13126313WA SGF13126314WA SGF13126315WA SGF13126316WA SGF13126317WA SGF13126318WA SGF13126319WA SGF13126320WA SGF13126321WA SGF13126322WA SGF13085864WA SGF13085865WA SGF13085866WA SGF13085867WA SGF13085868WA SGF13085869WA SGF13085871WA SGF13085872WA SGF13085873WA SGF13085875WA SGF13085792WA SGF13085793WA SGF13085794WA SGF13085795WA SGF13085796WA SGF13085797WA SGF13085798WA SGF13085801WA SGF13085802WA SGF13085804WA SGF13106108WA SGF13106118WA SGF13106120WA SGF13106123WA SGF13106129WA SGF13085814WA SGF13085819WA SGF13085820WA SGF13085821WA SGF13085822WA SGF13085823WA SGF13085827WA SGF13085828WA SGF13085833WA SGF13085834WA SGF13085835WA SGF13085836WA SGF13085837WA SGF13085838WA SGF13085839WA SGF13085840WA SGF13085841WA SGF13085842WA SGF13085843WA SGF13085844WA SGF13085845WA SGF13085846WA SGF13085847WA SGF13085848WA SGF13085849WA SGF13085851WA SGF13085852WA SGF13085853WA SGF13085854WA SGF13085855WA SGF13085856WA SGF13085857WA SGF13085858WA SGF13085859WA SGF13085860WA SGF13085861WA SF313106989WA SF313106990WA SF313106991WA SF313106992WA SF313106993WA SF313106994WA SF313106995WA SF313106996WA SF313106997WA SF313106998WA SF313106999WA SF313117000WA SF313117001WA SF313117002WA SF313117003WA SF313117004WA SF313117005WA SF313117007WA SF313117009WA SF313117010WA SGF13126374WA SGF13126376WA SGF13126377WA SGF13126379WA SGF13126381WA SGF13126386WA SGF13126393WA SGF13126394WA SGF13126395WA SGF13126396WA SGF13126397WA SGF13126398WA SGF13126404WA SGF13126357WA SGF13126368WA SGF13126369WA SGF13126373WA SJF13120094WA SJF13120095WA SJF13120096WA SJF13120097WA SJF13120098WA SJF13120099WA SJF13120100WA SJF13120101WA SJF13120102WA SJF13120103WA SJF13120104WA SJF13120105WA SJF13120106WA SJF13120107WA SJF13120108WA SJF13120109WA SJF13120110WA SJF13120111WA SJF13110081WA SJF13110083WA SJF13110087WA SJF13110091WA SGF13126367WA SG213091863WA SG213131992WA SG213131993WA SG213131994WA SG213131995WA SG213131998WA SG213132000WA SG213132004WA SG213132005WA SG213132006WA SG213132007WA SG213132008WA SG213132019WA SG213132023WA SG213132028WA SF313096951WA SF313117014WA SF313137048WA SF313137052WA SF313137054WA SF313137055WA SF313137057WA SF313147081WA SF313147082WA SF313147083WA SF313147084WA SF313147085WA SF313147088WA SF313147090WA SF313147103WA SG213172143WA SG213172144WA SGF13176677WA SGF13176679WA SGF13176668WA SGF13176671WA SGF13176673WA SGF13176685WA SGF13176686WA SGF13176690WA SGF13176691WA SGF13176695WA SGF13176735WA SGF13176736WA SGF13176737WA SGF13176738WA SGF13176739WA SGF13176740WA SG213142091WA SG213142095WA SGF12494814WA SF311072791WA SF311072794WA SF311072802WA SF311112954WA SF311112959WA SF311112960WA SF311112961WA SGF12525238WA SGF13015263WA SGF13015264WA SGF12525206WA SGF13035424WA SJF13010007WA SJF13020026WA SJF13020028WA SJF13020029WA SJF13010010WA SJF13010013WA SG212421027WA SG212421020WA SF312496608WA SGF12212627WA SGF12222651WA SGF12222652WA SGF12222653WA SGF12222654WA SGF12222656WA SGF12222657WA SGF12222658WA SGF12222659WA SGF12222660WA SGF12222663WA SGF12212626WA SGF12222642WA SGF12222643WA SGF12222644WA SG212150574WA SG211390049WA SG211390052WA SG211390057WA SG211430093WA SG211430094WA SG211430095WA SG211430096WA SG211430097WA SG211430098WA SG211430099WA SG211430100WA SG211430101WA SG211430102WA SG211430122WA SG211430124WA SG211430125WA SG211430126WA SGF12051382WA SGF12051440WA SGF12051452WA SGF12051453WA SGF12051455WA
SF310361539WA SF310361540WA SF310361541WA SF310361542WA SF310361543WA SF310361544WA SF310361545WA SF310361546WA SF310361547WA SF310361548WA SF310361550WA SF310361551WA SF310361552WA SF310361555WA SF310361556WA SF310361557WA SF310361558WA SF310361559WA SF310361560WA SF310381659WA SF310381661WA SF310381666WA SF310381668WA SF310381670WA SF310381671WA SF310381672WA SF310381673WA SG211430104WA SG211430105WA SG211430106WA SG211430107WA SG211430111WA SG211430112WA SG211430113WA SG211430114WA SG211440137WA SG211440138WA SG211470188WA SG211470190WA SG211470191WA SG211470192WA SG211470193WA SG211470194WA SG211470195WA SG211470196WA SG211470197WA SG211470198WA SG211450160WA SG211450170WA SG211450172WA SG211460173WA SG211460174WA SG211460175WA SG211460176WA SG211460177WA SG211460178WA SG211460181WA SG211460183WA SG211460184WA SG211460186WA SGF11360062WA SGF11430188WA SGF11460384WA SGF11460385WA SGF11460386WA SGF11460387WA SGF11460388WA SGF11470400WA SGF11470401WA SGF11470402WA SF311474581WA SG211440139WA SG212020368WA SG212020374WA SGF12051417WA SGF12051439WA SGF12051475WA SGF12021242WA SGF12021245WA SGF12021250WA SGF12051427WA SGF12061490WA SGF12061491WA SGF12061493WA SG212090440WA SGF12222655WA SGF12222664WA SGF12172525WA SG212270745WA SG212270746WA SGF12273279WA SG212230659WA SG212230661WA SG212230662WA SGF12273278WA SG211450165WA SGF12011094WA SGF12011096WA SGF12011110WA SGF12011135WA SGF12011173WA SGF12051430WA SGF12061497WA SGF12011168WA SGF12212621WA SG212160602WA SGF12273256WA SGF12011182WA SG213101930WA SF310080473WA SF310080484WA SF310080488WA SF310080492WA SF310080493WA SF310080494WA SF310120606WA SF310120607WA SF310120608WA SF310130610WA SF310180682WA SF310190741WA SF310200777WA SF310200778WA SF310200790WA SF310241012WA SF310241045WA SF310241046WA SF310241047WA SF310241048WA SF310241049WA SF310241050WA SF310241052WA SF310241053WA SF310241054WA SF310241056WA SF310241057WA SF310301152WA SF310301172WA SF310301173WA SF310301174WA SF310301175WA SF310301176WA SF310301177WA SF310301178WA SF310301180WA SF310301182WA SF310301184WA SF310301185WA SF310301186WA SF310301187WA SF310301188WA SF310301189WA SF310301190WA SF310301191WA SF310301192WA SF310301194WA SF310301195WA SF310301206WA SF310301207WA SF310311209WA SF310311210WA SF310311211WA SF310311212WA SF310311213WA SF310311214WA SF310311215WA SF310311216WA SF310311217WA SF310311218WA SF310311219WA SF310311221WA SF310311222WA SF310311223WA SF310311225WA SF310311240WA SF310311243WA SF310311244WA SF310311245WA SF310311246WA SF310311247WA SF310311248WA SF310311249WA SF310311229WA SF310482294WA SF310431888WA SF310431898WA SF310431905WA SF310431899WA SF310431901WA SF310431908WA SF310431910WA SF310431917WA SF310431921WA SF310431928WA SF310431972WA SF310431977WA SF310431931WA SF310441985WA SF310441991WA SF310441996WA SF310431895WA SF310452058WA SF310452069WA SF310452070WA SF310452111WA SF310462121WA SF310462132WA SF310462135WA SF310462136WA SF310462137WA SF310462187WA SF310462188WA SF310482324WA SF310482330WA SF310492344WA SF310492355WA SF310492358WA SF310492359WA SF310482336WA SF310492341WA SF311153155WA SF311193317WA SF311193335WA SF311193331WA SF311283748WA SF311283760WA SF311283766WA SF311283772WA SF311283776WA SF311263700WA SF311283825WA SF311293863WA SF311293886WA SF311293887WA SF311303896WA SF311324166WA SF311324186WA SF311324139WA SF311324141WA SF311324169WA SF311344314WA SF311344325WA SF311344329WA SF311344331WA SF311324170WA SF311334215WA SF311324184WA SF311404367WA SF311414383WA SF311404342WA SF311414399WA SF311414401WA SF311414403WA SF311414416WA SF311434516WA SF311434520WA SF311434521WA SF311434523WA SF311434526WA SF311454565WA SF311454566WA SF311454571WA SF311454572WA SF311474576WA SF311474582WA SF311474583WA SF311474589WA SF311474590WA SF311454560WA SF311454561WA SF311454563WA SF311454555WA SF311454556WA SF311454557WA SF311454558WA SF311454559WA SF311454564WA SF311454568WA SF311454570WA SF311454573WA SF311454574WA SF311484602WA SF311484606WA SF311474577WA SF311474585WA SF311474586WA SF311474587WA SF311474588WA SF311474592WA SF311484597WA SF311484601WA SF311484605WA SF311484607WA SF311484611WA SF311484614WA SF311494617WA SF311504701WA SF311504702WA SF311504703WA SF311504705WA SF311504706WA SF311504707WA SF311494664WA SF311494665WA SF311494666WA SF311494667WA SF311494669WA SF311494671WA SF311494673WA SF311494675WA SF311494680WA SF311494683WA SF311494688WA SF311494689WA SF311494690WA SF311504781WA SF311524844WA SF311524846WA SF311524847WA SF312065003WA SF312085085WA SF312085113WA SF312085117WA SF312095132WA SF312095134WA SF312095135WA SF312155279WA SF312155332WA SF312195508WA SF312195510WA SF312265669WA SF312275710WA SF312275717WA SF312275721WA SF312275722WA SF312275725WA SF312275726WA SF312275729WA SF312275731WA SF312275714WA SF312275720WA SF312275724WA SF312275733WA SF312275734WA SF312275748WA SF312285754WA SF312285755WA SF312285756WA SF312295768WA SF312295771WA SF312295772WA SF312315817WA SF312315818WA SF312315826WA SF312315827WA SF312315829WA SF312315833WA SF312315834WA SF312315838WA SF312315840WA SF312315841WA SF312366049WA SF312376064WA SF312376065WA SF312376068WA SF312376121WA SG212410966WA SG212421025WA SG212421026WA SG212421028WA SG212421029WA SG212421030WA SF312416215WA SGF12424172WA SGF12424175WA SGF12424259WA SGF12424260WA SGF12424261WA SGF12424262WA SGF12424263WA SGF12424265WA SGF12424267WA SGF12424269WA SGF12424254WA SGF12424255WA SGF12424266WA SGF12474413WA SGF12484648WA SGF12484611WA SGF12484612WA SF312476599WA SF312496629WA SF312496642WA SF312506699WA SF312506705WA SF312506714WA SF312506715WA SF312506716WA SF312506717WA SF312506718WA SF312506719WA SF312506723WA SF312506724WA SF312506725WA SF312506732WA SGF13035448WA SGF13035449WA SGF13035450WA SGF13035451WA SGF13035391WA SF312526772WA SF312526777WA SGF13035370WA SGF13035371WA SGF13035357WA SGF13035358WA SGF13035360WA SGF13035361WA SGF13035362WA SGF13035363WA SGF13035364WA SGF13035369WA SF313066912WA SF313066913WA SF313066914WA SF313066915WA SF313066916WA SF313066917WA SF313066919WA SF313066920WA SF313066922WA SF313066924WA SF313056899WA SF313066928WA SGF13146517WA SF313106975WA SG213121961WA SG213121962WA SG213121963WA SG213121964WA SGF13106044WA SF313147091WA SF313147092WA SF313147093WA SF313147094WA SF313147099WA SF311344330WA SF312295793WA SGF12424206WA SF309490113WA SF310060342WA SF310060347WA SF310060366WA SF310060367WA SF310060371WA SF310060378WA SF310301169WA SF310301179WA SF310301181WA SF310301183WA SF311072764WA SF311072765WA SF311072800WA SF311072808WA SF311072807WA SF311324136WA SF311504700WA SF311504704WA SF312446393WA SF312446394WA SGF13025273WA SGF13025280WA SGF13025286WA SF312456436WA SGF12122228WA SGF12152407WA SGF12152408WA SGF12152409WA SGF12152410WA SGF12152411WA SGF12152413WA SGF12152418WA SF311253663WA SF311253692WA SF311484608WA SF311494662WA SF311494672WA SF311494678WA SF311494685WA SF311494686WA SF311494693WA SF310431920WA SF310431907WA SF310482280WA SF310482284WA SF310502368WA SF310502369WA SF310502370WA SF310502371WA SF310502373WA SF310190701WA SF311253676WA SF310431914WA SF310431915WA SF311243549WA SF311243550WA SF310311228WA SF310301198WA SF310311253WA SF310311255WA SF310311256WA SF310311257WA SF311133054WA SF311133055WA SF310311251WA SF310311252WA SGF12283289WA SGF12263226WA SGF12313425WA SG212150558WA SGF12313419WA SGF12071596WA SGF11380081WA SGF11430190WA SGF11460382WA SGF11460383WA SGF11460389WA SG211360033WA SG211360034WA SG211360040WA SG211360045WA SG211380046WA SG211380048WA SG211390050WA SGF11350055WA SGF11350056WA SGF11360057WA SG211360041WA SG212020372WA SG212020373WA SG212050387WA SG212050399WA SG212050400WA SGF12031343WA SGF12051394WA SGF12051415WA SGF12051411WA SGF12051412WA SG212020355WA SGF12021221WA SGF12051423WA SG212050401WA SF312054950WA SF312054962WA SF312054971WA SGF12061503WA SGF12142286WA SGF12242842WA SGF12293333WA SF310361523WA SGF11460390WA SGF12051414WA SGF12051416WA SGF12051422WA SGF12051424WA SGF12051425WA SGF12051426WA SGF12061483WA SGF12061484WA SGF12061485WA SGF12061486WA SGF12061487WA SGF12061488WA SGF12061489WA SGF12051486WA SGF12152419WA SGF12152420WA SGF12162476WA SGF12162477WA SGF12212616WA SGF12212617WA SGF12212622WA SGF12212634WA SGF12212637WA SGF12222639WA SGF12222640WA SGF12222641WA SGF11490464WA SGF11490466WA SGF11490479WA SGF11490513WA SGF11490514WA SGF11490524WA SGF11490526WA SGF11490529WA SGF11490530WA SGF11500541WA SGF12011092WA SGF12011101WA SGF12011103WA SGF12011114WA SGF12011116WA SGF12011153WA SGF12011167WA SGF12011169WA SGF12011171WA SGF12011179WA SGF12263229WA SGF12263230WA SGF12263233WA SGF12263234WA SGF12263235WA SGF12263225WA SG212240684WA SG212240687WA SG212240690WA SG212240693WA SF310361529WA SGF12293332WA SG212150588WA SGF12051413WA SGF12011147WA SGF12011148WA SGF12011149WA SGF12011146WA SF311283759WA SF311283764WA SF311283809WA SF311293836WA SF311293843WA SF311293844WA SF311414442WA SF311424463WA SF311424496WA SF311434547WA SF311434554WA SF311133050WA SF311133052WA SF311133053WA SF311143091WA SF311143098WA SF311143099WA SF310482286WA SF310482289WA SF310482291WA SF310482292WA SF310482309WA SF310482313WA SF312195501WA SF312205530WA SF312165381WA SF312195462WA SF312195467WA SF312085066WA SF312085068WA SF312085069WA SF312085070WA SGF13146539WA SF310452041WA SF310452042WA SF310452043WA SF310452046WA SF310452047WA SF310452048WA SF310220855WA SF310220858WA SF310220860WA SF310220862WA SF312075032WA SF312075050WA SF311042630WA SF311042640WA SF311042641WA SF311042642WA SF311042643WA SF311042644WA SF311042645WA SF311042646WA SF311042647WA SF311042648WA SF311042649WA SF311042652WA SF311042655WA SF311042669WA SF311042671WA SF311042673WA SF311042674WA SF311042676WA SF311042678WA SF311042679WA SF311042680WA SF311042681WA SF311062702WA SF311062704WA SF311062706WA SF311072719WA SF311303946WA SF311314027WA SF311303957WA SF311303962WA SF311303969WA SF311303970WA SF311303974WA SF311303986WA SF311314039WA SF311314047WA SF311314049WA SF311314056WA SF311143080WA SF311143081WA SF311143083WA SF311143084WA SF311143086WA SF311143087WA SF311143092WA SF311524848WA SF310442000WA SF310110572WA SF310110584WA SF310110588WA SF310120590WA SF310120592WA SF310120594WA SF310120597WA SF310120600WA SF310120601WA SF310120604WA SF310190719WA SF310190720WA SF310220854WA SF310271106WA SF310271107WA SF310271118WA SF310431893WA SF310431924WA SF310431925WA SF310431926WA SF310431927WA SF310462150WA SF310482334WA SF310492337WA SF310492339WA SF311123008WA SF311123014WA SF311143082WA SF311143077WA SF311153144WA SF311193310WA SF311303941WA SF311303958WA SF311414454WA SF311424495WA SF311434524WA SF311434528WA SF311434532WA SF311434542WA SF311434544WA SF311494652WA SF311494658WA SF311494663WA SF311494668WA SF311494695WA SF312105211WA SF312105231WA SF312205564WA SF312205584WA SF312205587WA SF312295806WA SF312315823WA SF312315824WA SF312315825WA SG212431097WA SF312396194WA SF312416219WA SF312416234WA SF312416235WA SF312416236WA SF312416237WA SF312416239WA SF312416241WA SF312416243WA SF312416251WA SGF12474364WA SGF12474409WA SGF12474411WA SGF12474412WA SF312446367WA SF312446385WA SF312446386WA SGF12474414WA SGF12484634WA SF312466493WA SF312466531WA SF312466540WA SF312466549WA SF312466559WA SG212491352WA SG212491353WA SG212491354WA SG212491356WA
SG212491358WA SG212491359WA SG212491360WA SG212491362WA SG212491363WA SF312466536WA SF312466517WA SF312466546WA SF312466563WA SF312466579WA SGF12494756WA SGF12494757WA SGF12494758WA SGF12494759WA SGF12494783WA SGF12494787WA SF312496650WA SF312506692WA SGF12515160WA SGF12515166WA SGF12525193WA SGF12525194WA SGF12525195WA SGF13065747WA SGF13065748WA SGF13065749WA SGF13065752WA SG213061810WA SGF13055724WA SGF13065762WA SGF13095903WA SGF13095904WA SGF13095907WA SGF13096012WA SGF13096011WA SF313117045WA SGF13136505WA SGF13126327WA SGF13126358WA SGF13126362WA SGF13126363WA SGF13146519WA SF313137049WA SF313137051WA SF313177119WA SJF13130122WA SJF13130123WA SJF13170133WA SJF13170140WA SJF13170141WA SF310120598WA SF311344328WA SF312245652WA SF310441995WA SF311052686WA SF311052689WA SF311052690WA SF312085076WA SF312085078WA SF312085079WA SF312085082WA SF312085086WA SF312085088WA SF312085090WA SF312085091WA SF312085092WA SF312085093WA SF312085094WA SF312085095WA SF312085096WA SF310060381WA SF310060400WA SF310060415WA SF310060428WA SF310060430WA SF309490116WA SF310060369WA SF310060372WA SF310060380WA SF310060387WA SF310060394WA SF310060402WA SF310060403WA SF310060412WA SF310060419WA SF310060429WA SF310060401WA SF310060405WA SF310060406WA SF310060408WA SF310060426WA SF310060427WA SF310090510WA SF310090517WA SF310090519WA SF310090521WA SF310090525WA SF310060407WA SF310090515WA SF310090518WA SF310060421WA SF310060424WA SF310060425WA SF310060422WA SF310241064WA SF310241067WA SF310241068WA SF310241069WA SF310241070WA SF310241071WA SF310462140WA SF310462143WA SF310462148WA SF310462151WA SF310462153WA SF310462154WA SF310452103WA SF310482303WA SF310482305WA SF311072789WA SF311112884WA SF311112886WA SF311112889WA SF311112894WA SF311112903WA SF311112917WA SF311112940WA SF311112964WA SF311112965WA SF311112967WA SF311112968WA SF311112969WA SF311112976WA SF311193295WA SF311183267WA SF311193292WA SF311193293WA SF311233456WA SF311243591WA SF311243606WA SF311253695WA SF311303924WA SF311303983WA SF311314060WA SF311314021WA SF311314022WA SF311314033WA SF311314041WA SF311314064WA SF311314080WA SF311314124WA SF311334300WA SF311334274WA SF311334237WA SF311334252WA SF311334227WA SF311334233WA SF311334239WA SF311334247WA SF311344307WA SF311414428WA SF311414432WA SF311414438WA SF311414439WA SF311434501WA SF311334255WA SF311404350WA SF311404373WA SF311404377WA SF311414382WA SF311414384WA SF311414388WA SF311414390WA SF311404370WA SF311414387WA SF311334232WA SF311334258WA SF311344333WA SF311344334WA SF311344337WA SF311404363WA SF311404364WA SF311404376WA SF311414389WA SF311414395WA SF311414425WA SF311414431WA SF311414437WA SF311394341WA SF311404343WA SF311404344WA SF311404345WA SF311404346WA SF311404347WA SF311404348WA SF311404351WA SF311404352WA SF311404353WA SF311404354WA SF311404355WA SF311404356WA SF311404358WA SF311404359WA SF311404365WA SF311404368WA SF311404372WA SF311404374WA SF311404375WA SF311404378WA SF311414391WA SF311414392WA SF311414396WA SF311414397WA SF311414414WA SF311414417WA SF311414418WA SF311414419WA SF311414420WA SF311414444WA SF311414445WA SF311414450WA SF311414451WA SF311414452WA SF311424460WA SF311424464WA SF311424468WA SF311424469WA SF311424470WA SF311424474WA SF311424475WA SF311424477WA SF311424479WA SF311424481WA SF311424484WA SF311424485WA SF311424488WA SF311424490WA SF311424491WA SF311424492WA SF311424493WA SF311424494WA SF311424497WA SF311434498WA SF311434499WA SF311434502WA SF311434503WA SF311434513WA SF311434530WA SF311434533WA SF311434541WA SF311434545WA SF311434548WA SF311434549WA SF311434550WA SF311434551WA SF311434552WA SF311434553WA SF311414421WA SF311414449WA SF311424458WA SF311414448WA SF311434529WA SF311434535WA SF311494644WA SF311494645WA SF311494649WA SGF12414011WA SGF12414025WA SGF12414056WA SGF12414078WA SF312396200WA SGF12414085WA SGF12414087WA SG212410993WA SG212411002WA SGF12424189WA SGF12424195WA SGF12424197WA SGF12424198WA SGF12424199WA SF312426256WA SF312426257WA SF312426258WA SF312426259WA SF312426260WA SF312426261WA SF312436267WA SF312436269WA SF312436271WA SF312456428WA SF312456431WA SF312456432WA SF312456434WA SF312456435WA SF312456442WA SF312456445WA SF312456446WA SF312456447WA SF312456449WA SF312456450WA SF312456451WA SF312436356WA SF312456429WA SF312456430WA SF313026800WA SF313036815WA SF312476581WA SF312476582WA SF313016781WA SF313026797WA SF313026811WA SF313036817WA SF313036819WA SGF12494749WA SG212491419WA SG212491409WA SG212491414WA SF313016789WA SF313016791WA SF313016792WA SF313026796WA SF312526773WA SF312526778WA SF313016786WA SF313016783WA SF313016784WA SG212511532WA SG213041705WA SG213041706WA SG213041707WA SG213041708WA SG213041710WA SG213041711WA SF313046864WA SF313046867WA SF313046870WA SF313046879WA SF313056880WA SF313056881WA SF313056882WA SF313056883WA SF313056884WA SF313056885WA SF313056887WA SF313056889WA SF313056890WA SF313056891WA SF313056892WA SF313056893WA SF313056894WA SF313056896WA SF313056897WA SF313056898WA SF313056900WA SF313056901WA SF313056902WA SF313056903WA SF313056904WA SF313056905WA SF313056906WA SF313056908WA SF313056909WA SF313066911WA SGF13096004WA SGF13116205WA SGF13116209WA SF313117042WA SJF13130125WA SF313096952WA SF313137053WA SJF13170131WA SJF13170135WA SF312075042WA SF312095146WA SF312095162WA SF312095164WA SF312095167WA SF312165362WA SF312205545WA SF312275718WA SF312275735WA SF312275737WA SF312275746WA SF312325893WA SF312325884WA SF312325911WA SF312325919WA SF312325921WA SF312356016WA SF311052683WA SF311052685WA SF311052696WA SF311052697WA SF311303951WA SF311303955WA SF312195468WA SF312195469WA SF312195470WA SG213101947WA SF313147097WA SF313147100WA SF313147101WA SF313147102WA SF313147105WA SF313147107WA SF310190742WA SF310200755WA SF310200771WA SF310200774WA SF310200775WA SF310200776WA SF310200779WA SF310200780WA SF310200783WA SF310200784WA SF310210792WA SF310210793WA SF310210794WA SF310210795WA SF310210796WA SF310210797WA SF310210798WA SF310190715WA SF311223419WA SF311223424WA SG212100449WA SG212100461WA SG212240676WA SG212240678WA SG212240681WA SG212240683WA SG212240685WA SG212240686WA SG212240689WA SG212250700WA SG212250701WA SG212250704WA SF311494618WA SF311494620WA SF311494623WA SF311494634WA SG212521539WA SG212521540WA SG212521541WA SG212521542WA SG212521543WA SG213021586WA SG213021587WA SG213021588WA SG213021589WA SG213021590WA SF310482300WA SGF12363642WA SG211500222WA SGF12142314WA SGF12162434WA SGF12373756WA SF311314046WA SF312095170WA SJF13010008WA SJF13010009WA SF309520181WA SF311102877WA SF311112891WA SF310472196WA SF310472260WA SF310361483WA SF311223398WA SF311223399WA SF311243524WA SF309520191WA SF309520192WA SF309520195WA SF309520196WA SF309520200WA SF309520201WA SF309520204WA SF310361478WA SF310361481WA SGF12021199WA SF310020249WA SF311424480WA SF310060360WA SF310060390WA SF310060409WA SF310060411WA SF310070432WA SF310070434WA SF310070438WA SF310070440WA SF310070442WA SF310070443WA SF310070446WA SF310070448WA SF309520174WA SF309520184WA SF310462191WA SF310472194WA SF310472198WA SF310472200WA SF310472204WA SF310472205WA SF310472206WA SF310472207WA SF310472211WA SF310472214WA SF310472215WA SF310472216WA SF310472219WA SF310482276WA SF310482270WA SF310482273WA SF310482279WA SF310492356WA SF310482327WA SF310492343WA SF310492357WA SF310492360WA SF310492361WA SF310492362WA SF310492364WA SF310482328WA SF309390069WA sf309520172wa SF310030288wa SG212240679WA SGF11450298WA SGF11450302WA SGF11450303WA SGF11450304WA SGF11450305WA SGF11450307WA SGF11450308WA SGF11450309WA SGF11450310WA SGF11450314WA SGF11450319WA SGF11450320WA SGF12091886WA SGF12232682WA SGF11460349WA SGF11470404WA SGF11470405WA SGF11470406WA SGF11470411WA SGF11470413WA SGF11470419WA SGF11470421WA SGF11470422WA SGF11470424WA SGF11470425WA SGF11470427WA SGF11470428WA SGF11470429WA SGF11470431WA SGF12011130WA SGF12021190WA SGF12021254WA SGF12031269WA SGF12031274WA SGF12031276WA SGF12031290WA SGF12031298WA SGF12031303WA SGF12031305WA SGF12031307WA SGF12031309WA SGF12031289WA SGF12031297WA SF310401736WA SF310401737WA SF310411738WA SF310411743WA SF310411745WA SF310030303WA SF310030304WA SF31003034WA SF310060410WA SF310070441WA SF310472209WA SF310472213WA SF310381676WA SGF12252959WA SGF120919423 SF310361440WA SF309520198WA SF311082831WA SF311243359WA SGF12152404WA SGF1211205WA SGF12242829WA SGF12242837WA SGF12242838WA SGF12242841WA SGF12242850WA SGF12081716WA SF310220833WA SF310220864WA SG211520266WA SG211520267WA SG211520271WA SG211520272WA SG211520274WA SGF11460356WA SGF11460357WA SGF11460358WA SGF11460359WA SGF11460364WA SGF11460365WA SGF11520950WA SGF11521033WA SGF11521035WA SGF11521053WA SGF11380079WA SGF11440277WA SGF11450312WA SGF11450313WA SGF11450316WA SGF11450317WA SGF11450324WA SGF11450326WA SGF11450328WA SGF11450332WA SGF11450333WA SGF11450334WA SGF11450335WA SGF11450336WA SGF12112055WA SGF12112121WA SGF12112136WA SGF12112138WA SGF12112140WA SGF12122146WA SGF12122148WA SGF12122149WA SGF12242800WA SGF12242786WA SF311233484WA SF311233458WA SF310341407WA SF310341425WA SF311233499WA SF311243572WA SF310090511WA SF310090516WA SF310180620WA SF310240949WA SF309530215WA SF310100531WA SGF11520965WA SGF12031295WA SGF12081730WA SGF12091942WA SF310100537WA SGF11360066WA SGF13176606WA SF309420098WA SGF11380083WA SGF11360058WA SGF11370075WA SGF11370076WA SGF11370077WA SGF11380082WA SGF11430171WA SGF11430172WA SGF11430173WA SGF11430174WA SGF11430175WA SGF11430194WA SGF11430196WA SGF11430198WA SGF11430199WA SGF11430200WA SGF11430201WA SGF11430202WA SGF11430203WA SGF11430204WA SGF11430205WA SGF11430206WA SGF11440207WA SGF11440208WA SGF11440209WA SGF11440210WA SGF11440211WA SGF11440212WA SGF11440213WA SGF11440214WA SGF11440215WA SGF11440216WA SGF11440217WA SGF11440218WA SGF11440219WA SGF11440220WA SGF11440221WA SGF11440222WA SGF11440223WA SGF11440224WA SF310180635WA SF310180637WA SF310180641WA SF310180642WA SGF12363601WA SGF12363608WA SGF12363619WA SGF12363621WA SGF12363629WA SGF12363635WA SGF12363662WA SGF12363668WA SGF12363669WA SGF12363670WA SGF12363672WA SGF12363673WA SGF12363674WA SGF12363675WA SGF12363676WA SGF12363677WA SGF12363678WA SGF12363679WA SGF12363680WA SGF12363681WA SGF12363682WA SGF12363685WA SF310110567WA SF309520178WA SF309520179WA SF309520187WA SF309520189WA SF311303898WA SF311303907WA SF311303930WA SF311303931WA SF311303932WA SF311303933WA SF311303936WA SF311303937WA SF311303940WA SF311303945WA SF311303950WA SF311303972WA SF311303980WA SF311303987WA SF310230912WA SGF12363586WA SGF12363598WA SF309520206WA SF309530213WA SF309520199WA SF309520203WA SF309520205WA SF309520207WA SF309520208WA SF309530214WA SF309530216WA SGF12031301WA SGF12031308WA SGF12031326WA SGF11521014WA SGF11521019WA SGF11520960WA SGF12021191WA SGF12031275WA SGF12031280WA SGF12232732WA SGF13035455WA SGF13055671WA SF311223418WA SF311223420WA SF311223421WA SGF12232729WA SGF12232734WA SGF12232739WA SGF12232740WA SGF12232741WA SGF12232744WA SGF12232752WA SF310110554WA SF310110565WA SF310110566WA SF310110569WA SF311233468WA SF311082834WA SF310030295WA SF310472240WA SF310472246WA sf310331396wa SF310190743WA SF310220847WA SF310472251WA SF311243556WA SF311243573WA SF310361450WA SF310110580WA SF310020252WA SF310371646WA SF31052413WA SF311062687WA SF310180626WA SF310090513WA SF311233476WA SF311233483WA SF310331379WA SF310331398WA SF310341401WA SF310341414WA SF310341421WA SF310341428WA SF310070444WA SF311213389WA SF311223431WA SF311193307WA SF311213356WA SF309520197WA SF309520210WA SF310251086WA SF310271098WA SF310080477WA SF310220880WA SF310220849WA SF310070447WA SF311243570WA SF309520194WA SF310431941WA SF310331386WA SF310180625WA SF310180628WA SF310180629WA SF310230942WA SF310100549WA SF310240967WA SF310240978WA SGF12484666WA SF310401713WA SF310411739WA SGF12515069WA SGF12515070WA SGF12515138WA SGF12515124WA SF310361489WA SF310361495WA SF310361497WA SF310361498WA SF310361501WA SF310361504WA SF310361511WA SF310361514WA SF309420105WA SF310361455WA SF310050326WA SF310070449WA SF310070459WA SF310070467WA SF310361436WA SF310341433WA SF310090497WA SGF13015251WA SGF12515140WA SF311223393WA SF310020245WA SF310020266WA SF310030325WA SGF12474378WA SGF12434324WA SF311293880WA SF310462175WA SF311072717WA SF309390055WA SF310030281WA SF310030286WA SF310030315WA SF310030317WA SG212491405WA SF311334265WA SF310020244WA SF310020246WA SF309410086WA SF311213360WA SF310220832WA SF309520188WA SF310331373WA SF311223396WA SF310190731WA SF310190700WA SF310411750WA SF310411752WA SF311334214WA SF310230902WA SF311344313WA SF310331382WA SF310240965WA SF310331389WA SF310030289WA SF310030299WA SG213031679WA SF310492363WA SF310261091WA SF311082833WA SF311233487WA SF311243508WA SF311334263WA SF311233489WA SF311233500WA SF309500138WA SF310472201WA SG212491364WA SF310411744WA SF310331395WA SF310341406WA SF310361493WA SF310361507WA SF310361515WA SF310361516WA SF310361521WA SF310472234WA SGF12242816WA SGF12102020WA SGF12102053WA SGF11520949WA SF311243522WA SGF12263214WA SGF12363589WA SGF12363595WA SGF12363599WA SF310431947WA SF311072734WA SF309510156WA SF310361446WA SF309500146WA SF310472225WA SGF11520927WA SGF12011124WA SGF12021214WA SF310230917WA SF311334288WA SF310462141WA SGF12253029WA SGF12253038WA SGF12253040WA SGF12253043WA SGF12253047WA SGF12253048WA SGF12253049WA SGF12253051WA SGF12253054WA SGF12253056WA SGF12253058WA SGF12102005WA SGF12172503WA SGF12172534WA SGF12112112WA SGF12112095WA SGF12091875WA SGF12363577WA SGF12252957WA SGF12252960WA SGF12021205WA SG212320840WA SGF12293353WA SGF12293356WA SGF12293357WA SGF12313361WA SGF12313363WA SGF12313366WA SGF12313367WA SGF12313377WA SGF12313391WA SGF12313394WA SGF12313396WA SGF12313397WA SGF12313401WA SGF12313414WA SGF12313415WA SGF12313430WA SGF12313431WA SF311243577WA SGF12102021WA SGF12091946WA SGF12112077WA SGF12132255WA SGF12132259WA SGF12132261WA SGF12142282WA SGF12142283WA SGF12142288WA SGF12142302WA SGF12142307WA SGF12142320WA SGF12142321WA SGF12142331WA SGF12142333WA SGF12142335WA SGF12152365WA SGF12152373WA SGF12152375WA SGF12152376WA SGF12152381WA SGF12152387WA SGF12152424WA SGF12152425WA SGF12152431WA SGF12162432WA SGF12162443WA SGF12162448WA SGF12162450WA SGF12162454WA SGF12162455WA SGF12172498WA SGF12172507WA SGF12172508WA SGF12172512WA SGF12172531WA SGF12172549WA SGF12172560WA SGF12172567WA SGF12172589WA SGF12172595WA SGF12172597WA SGF12172598WA SGF12172601WA SGF12172608WA SGF12232666WA SGF12333525WA SF310452107WA SF311243529WA SF311243541WA SGF12071589WA SGF12071595WA SF310341416WA SF310311282WA SF310311283WA SF310311284WA SF310411754WA SF310411755WA SF310411756WA SF310411758WA SF310411760WA SF310411761WA SF310411762WA SGF11460396WA SGF11480457WA SGF11490480WA SGF11490487WA SGF11500576WA SGF11500577WA SGF11500598WA SGF11510610WA SGF11510615WA SGF11510617WA SGF11510620WA SGF11510629WA SGF11510633WA SGF11510634WA SGF11510636WA SGF11510638WA SGF11510639WA SGF1150577WA SGF11521031WA SGF12263216WA SGF12333519WA SGF12363578WA SGF12373745WA SGF12373747WA SF310331387WA SF310341400WA SF310341405WA SF310341408WA SF310341426WA
SGF12263213WA SF311082824WA SF311072804WA SGF11520962WA SGF11521029WA SF311193321WA SGF13116251WA SGF13176589WA SGF13176601WA SGF13176602WA SGF12232707WA SGF12232710WA SGF12162438WA SF310090508WA SF311344315WA SGF12474426WA SF311082842WA SGF11460348WA SGF11460352WA SGF11460360WA SGF12102042WA SGF12102052WA SGF12102054WA SF311112915WA SF311112926WA SGF12363596WA SGF12363683WA SGF13025290WA SGF13106025WA SGF13106021WA SGF12373726WA SG211510246WA SGF11510688WA SGF12021193WA SGF12021203WA SGF12021208WA SGF12021220WA SGF12263205WA SGF12273251WA SGF12273257WA SGF12283282WA SGF12323474WA SGF11510650WA SG213091888WA SGF1302562WA SGF12031278WA SF310110579WA SF311082830WA SF312065006WA SGF12373724WA SGF12373728WA SF310020257WA SGF12363575WA SF310341431WA SF310341418WA SGF12061554WA SGF12061555WA SGF12061556WA SGF12061557WA SGF12081715WA SGF12112115WA SGF12132266WA SGF12142324WA SGF12152337WA SGF12152344WA SGF12152354WA SGF12152361WA SGF12152363WA SGF12152367WA SGF12152385WA SGF12172485WA SGF12172505WA SGF12172515WA SGF12172528WA SGF12172545WA SGF12172547WA SGF12172552WA SGF12172554WA SGF12172558WA SGF12172576WA SGF12172584WA SGF12172590WA SGF12172591WA SGF12172593WA SGF12172599WA SGF12172600WA SGF12172605WA SGF12172609WA SGF12232681WA SGF12232686WA SGF12232746WA SGF12232747WA SGF12232750WA SGF12232753WA SGF12232754WA SGF12242759WA SGF12242760WA SGF12242761WA SGF12242762WA SGF12242764WA SGF12242765WA SGF12242767WA SGF12242769WA SGF12242771WA SGF12242773WA SGF12242776WA SGF12242781WA SGF12242791WA SGF12242797WA SGF12242801WA SGF12242803WA SGF12242814WA SGF12242818WA SGF12091926WA SGF12091916WA SGF12112064WA SGF12112113WA SGF11520958WA SGF12091862WA SGF12091930WA SG211410079WA SG211410080WA SG211410087WA SG211410088WA SG211410089WA SGF11380084WA SGF11380085WA SGF11390091WA SGF11390119WA SGF11390120WA SGF11390123WA SGF11410127WA SGF11410128WA SGF11410129WA SGF12102031WA SGF12031284WA SGF12031285WA SGF12242862WA SGF12252951WA SGF12273253WA SGF12273280WA SGF12273281WA SGF12031260WA SGF11510616WA SGF11510642WA SGF11510648WA SGF11460375WA SGF11470414WA SGF11470420WA SGF11470430WA SGF11490521WA SGF11490522WA SGF11490533WA SGF11500549WA SGF11500552WA SGF11500565WA SGF11490481WA SGF11490520WA SGF11490523WA SGF11490534WA SGF11500543WA SGF11500544WA SGF11460354WA SGF11460367WA SGF11480444WA SGF11490465WA SGF11490472WA SGF11490473WA SGF11490482WA SGF11490483WA SGF11490484WA SGF11490485WA SGF11490486WA SGF11490490WA SGF12071631WA SGF12081761WA SGF12081767WA SGF12081776WA SGF12081702WA SGF12081758WA SGF12081777WA SGF12081782WA SGF12091906WA SGF12091909WA SGF12091917WA SGF12091923WA SGF12091932WA SGF12102000WA SGF12102022WA SGF12102049WA SGF12112058WA SGF12112067WA SGF12112086WA SGF12112096WA SGF12112101WA SGF12112106WA SGF12112114WA SGF12112116WA SGF12071609WA SGF12091866WA SGF12091868WA SGF12091871WA SGF12091879WA SGF12091883WA SGF12091891WA SGF12091892WA SGF12091903WA SGF12091918WA SGF12091919WA SGF12091944WA SGF12102028WA SGF12102030WA SGF12102033WA SGF12102035WA SGF12102003WA SGF12102043WA SGF12102046WA SGF12172502WA SGF12132251WA SGF12142310WA SGF12142317WA SGF12152357WA SGF12152386WA SGF12152398WA SGF12162440WA SGF12162445WA SGF12162446WA SGF12172527WA SGF12172538WA SGF12172551WA SGF12142285WA SGF12172497WA SGF12172553WA SGF12172559WA SGF12172572WA SGF12172575WA SGF12172585WA SGF12172588WA SGF12172592WA SGF12172594WA SGF12242879WA SGF12242880WA SGF12242883WA SGF12242886WA SGF12252932WA SGF12252933WA SGF12252935WA SGF12252942WA SGF12252943WA SGF12252944WA SGF12252945WA SGF12252988WA SGF12252989WA SGF12252991WA SGF12252992WA SGF12252994WA SGF12112103WA SGF11460376WA SGF11470403WA SGF11470407WA SGF11470410WA SGF11470417WA SGF11470423WA SGF11470426WA SGF11480433WA SGF11480435WA SGF11480436WA SGF11390092WA SGF11390094WA SGF11390096WA SGF11390099WA SGF11390100WA SGF11390101WA SGF11390102WA SGF11390103WA SGF11390104WA SGF11390105WA SGF11390108WA SGF11390110WA SGF11390111WA SGF11390118WA SGF11390121WA SGF12091854WA SGF12091855WA SGF12091881WA SGF12091910WA SGF12091911WA SGF12091913WA SGF12091915WA SGF12091949WA SGF12102025WA SGF12102026WA SGF11410135WA SGF11460380WA SGF11510611WA SGF11510612WA SGF11510625WA SGF11510627WA SGF11510635WA SGF11510637WA SGF12091887WA SGF12091924WA SGF12102001WA SGF12102023WA SGF12102041WA SGF12102051WA SGF12102050WA SGF12112066WA SGF12112068WA SGF12112072WA SF310331393WA SF310341402WA SGF12172573WA SGF12232668WA SGF12232669WA SGF12232670WA SGF12232675WA SGF12232676WA SGF12232677WA SGF12232685WA SGF12232687WA SGF12232691WA SGF12232695WA SGF12232697WA SGF11520975WA SGF12222645WA SGF12222646WA SGF12222647WA SGF12222648WA SGF12222649WA SGF12222650WA SGF11430176WA SGF11430177WA SGF11430178WA SGF11430179WA SGF11430180WA SGF11430181WA SGF11430182WA SGF11430183WA SGF11440237WA SGF11440239WA SGF11440240WA SGF11440243WA SGF11440244WA SGF11440245WA SGF11440246WA SGF11520948WA SGF11521030WA SGF11521048WA SGF12051443WA SGF12051465WA SGF12061505WA SGF12061506WA SGF12071591WA SGF12071593WA SGF12071611WA SGF12071617WA SGF11521032WA SGF12091899WA SGF12091902WA SGF11510864WA SGF12081725WA SGF12071641WA SGF12081692WA SGF12081701WA SGF12081703WA SGF12081706WA SGF12081707WA SGF12081708WA SGF12081724WA SGF12081726WA SGF12081727WA SGF12081731WA SGF12081744WA SGF12081746WA SGF12081749WA SGF12081750WA SGF12081752WA SGF12081754WA SGF12081757WA SGF12081762WA SGF12081763WA SGF12081766WA SGF12081769WA SGF12081778WA SGF12081779WA SGF12081780WA SGF12081781WA SGF12081783WA SGF12081784WA SGF12081785WA SGF12081786WA SGF12081789WA SGF12081791WA SGF12081792WA SGF12081793WA SGF12091856WA SGF12091857WA SGF12091860WA SGF12091867WA SGF12091869WA SGF12081759WA SGF12091912WA SGF12091914WA SGF12091933WA SGF12091927WA SGF12102039WA SGF11520952WA SGF12091896WA SGF12273249WA SF309520209WA SGF11390095WA SGF11450331WA SGF11460381WA SGF11460393WA SGF11460394WA SGF11460395WA SGF11470412WA SGF11470416WA SGF11480437WA SGF11480438WA SGF11480442WA SGF11480446WA SGF11480451WA SGF11480454WA SGF11480455WA SGF11480458WA SGF11490460WA SGF11490462WA SGF11490468WA SGF11490470WA SGF11490474WA SGF11490476WA SGF11490477WA SGF11490488WA SGF11490489WA SGF11490491WA SGF11490512WA SGF11490515WA SGF11490517WA SGF11490525WA SGF11490527WA SGF11490531WA SGF11490532WA SGF11500546WA SGF11500553WA SGF11500596WA SGF11500597WA SGF11500599WA SGF11500600WA SGF11500601WA SGF11500602WA SGF11500603WA SGF11500605WA SGF11500606WA SGF11500607WA SGF11500608WA SGF11500609WA SGF11510613WA SGF11510614WA SGF11510618WA SGF11510624WA SGF11510630WA SGF11460347WA SGF11460350WA SGF11460351WA SGF11460353WA SGF11460370WA SG212130528WA SG212150548WA SG212150582WA SG212150585WA SG212190618WA SG212240667WA SG212240668WA SG212240671WA SG212240674WA SG212240688WA SG212240692WA SG212240694WA SG212240695WA SG212240696WA SG212240697WA SG212240698WA SF310472239WA SF310341432WA SGF12071608WA SGF12081788WA SGF12102027WA SGF12112056WA SGF12112060WA SGF12112069WA SGF12112073WA SGF12112082WA SGF12112087WA SGF12112092WA SF310331371WA SF310452082WA SGF11521006WA SGF11521007WA SGF11521009WA SGF11521013WA SGF11521017WA SGF11521018WA SGF11521020WA SGF11521021WA SGF11521022WA SGF11521023WA SGF11521024WA SGF11521025WA SGF11521026WA SGF11521027WA SGF11521040WA SGF11521041WA SGF11521050WA SGF11521052WA SGF11521059WA SGF11521060WA SGF11521089WA SGF11521054WA SGF12081773WA SGF12081790WA SGF12091877WA SGF12091882WA SGF12091936WA SGF11390093WA SGF12112099WA SGF12142281WA SGF12142289WA SGF12142334WA SGF12152401WA SGF12152426WA SGF12162433WA SGF12172517WA SGF12172540WA SGF12172542WA SGF12172544WA SGF12172546WA SGF12172550WA SGF12172555WA SGF12172562WA SGF12172564WA SGF12172569WA SGF12172570WA SGF12172579WA SGF12172580WA SGF12172602WA SGF12172603WA SGF12172611WA SGF12232665WA SGF12232667WA SGF12172565WA SGF12172574WA SGF12172604WA SGF12232673WA SGF12232679WA SGF12232680WA SGF12232683WA SGF12232688WA SGF12232689WA SGF12232690WA SGF12232692WA SGF12232693WA SGF12232694WA SGF12232696WA SGF12232708WA SGF12232709WA SGF12232711WA SGF12232712WA SGF12232713WA SGF12232714WA SGF12232715WA SGF12232716WA SGF12232730WA SGF12232731WA SGF12232743WA SGF12132249WA SGF11521005WA SGF11521008WA SGF11521010WA SGF11521011WA SGF11521015WA SGF11521016WA SGF11521034WA SGF11521036WA SGF11521037WA SGF11521038WA SGF11521039WA SGF11521042WA SGF11521045WA SGF11521046WA SGF11521047WA SGF11521049WA SGF11521051WA SGF11521056WA SGF11350052WA SGF11360063WA SGF11410131WA SGF11410132WA SGF11410133WA SGF11410134WA SGF11420136WA SGF11420137WA SGF11420139WA SGF11420140WA SGF11420141WA SGF11420142WA SGF11420143WA SGF11420144WA SGF11420145WA SGF11420146WA SGF11420147WA SGF11420148WA SGF11420149WA SGF11420150WA SGF11420151WA SGF11420152WA SGF11420153WA SGF11420154WA SGF11420155WA SGF11420156WA SGF11420157WA SGF11420158WA SGF11420159WA SGF11420160WA SGF11420162WA SGF11420163WA SGF11420164WA SGF11420165WA SGF11420166WA SGF11420167WA SGF11420168WA SGF11420169WA SGF11420170WA SGF12091859WA SF310331347WA SGF13116229WA SGF12071615WA SGF12091900WA SGF12112091WA SGF12112097WA SGF12112098WA SGF12112105WA SF310341434WA SGF12031288WA SGF12102044WA SGF12112057WA SGF11520963WA SGF11390114WA SGF11410130WA SGF11380086WA SGF12091876WA SG212360884WA SGF12363641WA SGF12363665WA SGF12071613WA SGF12071633WA SGF12081760WA SGF12081765WA SGF12081768WA SGF12081771WA SGF1209188SWA SGF12011108WA SGF12112100WA SF311223436WA SGF12091865WA SGF12081734WA SGF12091872WA SG212110471WA SGF12112104WA SGF12112117WA SGF12112123WA SGF12112125WA SGF12112128WA SGF12112135WA SGF12112137WA SGF12122145WA SGF12122150WA SGF12122151WA SGF12122152WA SGF12122153WA SGF12122154WA SGF12122155WA SGF12122160WA SGF12122161WA SGF12122162WA SGF12122163WA SGF12122166WA SGF12122206WA SGF12102045WA SG212240691WA SG212280753WA SG212280755WA SG212280758WA SG212280759WA SG212280760WA SG212280761WA SG212310783WA SG212310784WA SG212310785WA SG212310786WA SG212310787WA SG212310788WA SG212310790WA SG212310791WA SG212310802WA SG212310805WA SG212310806WA SG211520246WA SG212240666WA SF310462142WA SF310030302WA SF310371640WA SF310371638WA SGF11510621WA SGF12112108WA SGF12102038WA SG212360876WA SG212360886WA SG212360888WA SG212360889WA SGF11390090WA SGF12142326WA SGF12162436WA SGF12172610WA SGF12232671WA SGF12232674WA SGF12232678WA SGF11521012WA SGF12011091WA SGF12011095WA SGF12011097WA SGF12011105WA SGF12011106WA SGF12011118WA SGF12011119WA SGF12011120WA SGF12011122WA SGF12011126WA SGF12011127WA SGF12011128WA SGF12011129WA SGF12011133WA SGF12011138WA SGF12021194WA SGF12021206WA SGF12021207WA SGF12021209WA SGF12021210WA SGF12021216WA SGF12021218WA SGF12011125WA SGF12021195WA SGF12021198WA SGF12021200WA SGF12021215WA SGF12031293WA SGF12112074WA SGF12091858WA SGF12091874WA SGF12112094WA SGF12091885WA SGF12112065WA SG212120477WA SG212120478WA SG212120480WA SG212120481WA SG212120482WA SG212120484WA SGF12253068WA SGF12273241WA SGF12283297WA SGF12283303WA SGF12323490WA SGF12333531WA SGF12333532WA SGF12363585WA SGF12363594WA SGF12313378WA SGF12363588WA SGF12363597WA SGF12363671WA SGF12373718WA SGF12373721WA SGF12373725WA SG212360887WA SGF11520973WA SG212360882WA SG212360890WA SG212360891WA SG212360895WA SGF11390124WA SGF12091943WA SGF12373717WA SGF12363652WA SGF12363658WA SGF12363659WA SGF12363689WA SGF12373744WA SGF12373746WA SGF12373749WA SGF12373750WA SGF12373752WA SGF12373720WA SF310110571WA SF311213388WA SF310371618WA SF309520193WA SF310361484WA SF311344303WA SF311243581WA SGF13015252WA SGF12504846WA SGF12515055WA SGF12515053WA SF311334266WA SF311334209WA SF311334299WA SGF13025261WA SF310240964WA SF310020255WA SF310020274WA SGF12232733WA SGF12373723WA SGF12373729WA SG212310782WA SG212310789WA SG212310814WA SG212320817WA SG212320818WA SG212320819WA SGF12283290WA SGF12293346WA SGF12293347WA SGF12293348WA SGF12293350WA SGF12293351WA SGF12293358WA SGF12313365WA SGF12313380WA SGF12313383WA SGF12313384WA SGF12313389WA SGF12313392WA SGF12313395WA SGF12313413WA SGF12333502WA SG212290770WA SG212290771WA SG212290778WA SG212310803WA SGF12263228WA SGF12313409WA SGF12313410WA SGF12313411WA SGF12313412WA SGF12313393WA SGF12313418WA SGF12313420WA SGF12353558WA SGF12353560WA SGF12353562WA SGF12353569WA SGF11440229WA SGF11440232WA SGF11440234WA SGF11440235WA SGF11440256WA SGF11440259WA SGF11440260WA SGF11440261WA SGF11440262WA SGF11430197WA SGF12505025WA SGF12515040WA SGF12515117WA SF311193298WA SGF11521043WA SGF11521044WA SGF12061553WA SGF12071585WA SF310030312WA SF312085058WA SGF12011123WA SGF12021192WA SGF12021196WA SGF12021197WA SGF12021201WA SGF12021202WA SGF12021204WA SGF12263206WA SGF12363628WA SGF12363634WA SGF12363636WA SGF12363637WA SGF12363638WA SGF12363663WA SGF12363666WA SGF12031329WA SGF12051418WA SGF12051437WA SGF12051441WA SGF12051448WA SGF12051456WA SGF12051471WA SGF12061534WA SGF12061538WA SGF12061543WA SGF12061544WA SGF12061548WA SGF12061549WA SGF12061559WA SGF12061561WA SGF12061562WA SGF12061563WA SGF12061564WA SGF12061565WA SGF12061566WA SGF12061567WA SGF12061568WA SGF12061570WA SGF12061571WA SGF12061572WA SGF12061573WA SGF12061574WA SGF12061575WA SGF12061576WA SGF12061577WA SGF12061581WA SGF12071584WA SGF12071587WA SGF12071588WA SGF12071590WA SGF12071594WA SGF12071597WA SGF12071598WA SGF12071599WA SGF12071604WA SGF12071610WA SGF12071618WA SGF12071622WA SGF12071624WA SGF12091893WA SF311233446WA SF310341429WA SF310361439WA SF310472227WA SF310472228WA SGF12061578WA SGF12061579WA SGF12061580WA SGF12061582WA SGF12061583WA SGF12313422WA SGF12313427WA SGF12353550WA SGF12353551WA SGF12353552WA SGF12353553WA SGF12353554WA SGF12353555WA SGF12353556WA SGF12353557WA SGF12353559WA SGF12353563WA SGF12353565WA SGF12353566WA SGF12353567WA SGF12353568WA SGF12353571WA SGF12353572WA SGF12353573WA SGF12353574WA SF310391708WA SF310391709WA SF310391710WA SF310391712WA SF310401716WA SF310401717WA SF310401718WA SF310401719WA SF310401721WA SF310401724WA SF310401726WA SF310401727WA SF310220845WA SF311082854WA SF311072785WA SGF12293339WA SGF12293340WA SGF12232735WA SF310331337WA SF310331357WA SGF12293334WA SGF12293335WA SF310130611WA SGF12373694WA SGF12373696WA SGF12373700WA SGF12373701WA SGF12373702WA SGF12373703WA SGF12373705WA SGF12373706WA SGF12373707WA SGF12373708WA SGF12373709WA SGF12373710WA SGF12373711WA SGF12373712WA SGF12373713WA SGF12373714WA SGF12373715WA SGF12373716WA SGF12373719WA SG212260717WA SG212260718WA SG212260719WA SG212260720WA SG212260721WA SG212260722WA SG212260723WA SG212260724WA SG212260725WA SG212260726WA SG212260727WA SGF12232684WA SGF12112083WA SGF12112085WA SGF12061537WA SGF12071605WA SGF12071625WA SGF12071626WA SGF12071628WA SGF12071629WA SGF12071630WA SGF12112062WA SGF12112093WA SF309520185WA SF310512418WA SF310512419WA SF310512420WA SF310512421WA SF310512422WA SF310512424WA SF310512427WA SF310512428WA SF310512436WA SF310512439WA SF310512449WA SF310512450WA SF310512451WA SF310512453WA SF310512460WA SGF11520951WA SGF12021188WA SGF12031277WA SGF12031279WA SGF12031281WA SGF12031286WA SGF12031287WA SGF12031291WA SGF12031296WA SGF12031299WA SGF12031300WA SGF12031304WA SGF12051361WA SGF12051364WA SGF12051367WA SGF12263128WA SF310401720WA SF310411748WA SF310361492WA SF310462180WA SF310472247WA SF309520177WA SF310512447WA SF310512452WA SF310381653WA SGF12242812WA SF311213390WA SF310361508WA SF310030279WA SF310030280WA SF310030282WA SF310030283WA SF310030290WA SF310030297WA SF310030309WA SF310030313WA SF310030316WA SF310030324WA SF310020277WA SF310030314WA SF310030318WA SF310030319WA SF309500141WA SF310220888WA SGF12071601WA SGF11480447WA SGF11480450WA SGF11500604WA SG212240665WA SG212240669WA SG212240670WA SG212240672WA SG212240675WA SG212240677WA SG212240680WA SG212240699WA SGF12232756WA SGF12242784WA SGF12242792WA SGF12242793WA SGF12242799WA SGF12242802WA SGF12242804WA SGF12242806WA SGF12242808WA SGF12242813WA SGF12242822WA SGF12242823WA SGF12242824WA SGF12242825WA SGF12242830WA SGF12242831WA SGF12242832WA SGF12242833WA SGF12242834WA SGF12242835WA SGF12242836WA SGF12242843WA SGF12242845WA SGF12242849WA SGF12242851WA SGF12373727WA SGF12373731WA SGF12373733WA SGF12373734WA SGF12373735WA SGF12373736WA SGF12373737WA SGF12373738WA SGF12373739WA SGF12373740WA SGF12373743WA SGF12373732WA SGF12373741WA SGF12373742WA SGF12051472WA SGF12051474WA SGF12051476WA SGF12051477WA SGF12051478WA SF310100535WA SF311243528WA SF311243553WA SF311243525WA SF311243545WA SF311243552WA SF311243554WA SF311243555WA SF311243557WA SF311243558WA SF311243559WA SF311243560WA SF311243561WA SF311243562WA SF311243563WA SF311243564WA SF311243565WA SF311243566WA SF311243567WA SF311243568WA SF311243569WA SF311243571WA SF311243574WA SF311243575WA SF311243576WA SF311243578WA SF311243579WA SF311243580WA SF311243582WA SF311243583WA SF310110555WA SF310110556WA SF310401729WA SF310401731WA SF310401732WA SF310401733WA SF310472226WA SF311293872WA SF311303925WA SF311283774WA SF311263726WA SGF12333536WA SG211390055WA SG211410083WA SG211410084WA SG211410085WA SG211410086WA SGF11390122WA SGF12333538WA SGF12363576WA SGF12363580WA SGF12363582WA SGF12363590WA SGF12021260WA SGF12031282WA SGF12031292WA SGF11440250WA SGF11460361WA SGF11460368WA SGF11460369WA SGF11460371WA SGF11460372WA SGF11460373WA SGF11460374WA SGF11460377WA SGF11460378WA SGF11460379WA SGF11460392WA SGF12021237WA SGF12031302WA SF311112942WA SF309410085WA SF310190727WA SGF12142273WA SGF12142276WA SGF12142290WA SGF12142328WA SGF12152360WA SGF12152370WA SGF12152379WA SGF12152380WA SGF12152384WA SGF12152391WA SGF12152427WA SGF12162437WA SGF12172491WA SGF12172496WA SGF12172504WA SGF12172514WA SF309520175WA SGF13045578WA SGF13055594WA SGF13055597WA SGF13055599WA SGF13055600WA SGF13176728WA SGF13176729WA SGF13176732WA SGF13176733WA SGF12494747WA SGF12494748WA SGF12494753WA SGF12494754WA SGF12494755WA SGF12494798WA SF310361471WA SF310361473WA SF310361477WA SF310361480WA SF310361485WA SF310361486WA SF310361487WA SF310361488WA SF310361490WA SF311112925WA SF310070463WA SF309390057WA SGF13025262WA SF310080481WA SF310080486WA SF309410083WA SF309410075WA SF309510157WA SGF12494740WA SGF12494833WA SGF12504842WA SGF12504844WA SGF12504847WA SGF12504849WA SGF12504850WA SGF12504852WA SGF12504853WA SGF12504854WA SGF12504855WA SGF12504856WA SGF12504857WA SGF12504858WA SGF12504859WA SGF12504861WA SGF12504862WA SGF12504863WA SGF12504866WA SGF12504868WA SG213031657WA SG213031659WA SG213031660WA SGF13045457WA SGF13045458WA SGF13045459WA SGF13045460WA SGF13045463WA SGF13045465WA SGF13045466WA SGF13045496WA SGF13045499WA SGF13045500WA SGF13045501WA SGF13045503WA SGF13045504WA SGF13045505WA SGF13045506WA SGF13045507WA SGF13045508WA SGF13045509WA SGF13045510WA SGF13045511WA SGF13045512WA SGF13045513WA SGF13045514WA SGF13045515WA SGF13045516WA SGF13045517WA SGF13045518WA SGF13045519WA SGF13045520WA SGF13045523WA SGF13045525WA SGF13045528WA SGF13045530WA SGF13126359WA SGF13126360WA SGF13126361WA SGF13126364WA SGF13126365WA SGF13126366WA SGF13136444WA SGF13136445WA SGF13136446WA SGF13136448WA SGF13136451WA SGF13136453WA SGF13136454WA SGF13136456WA SGF13136458WA SGF13136506WA SJF13110060WA SJF13110076WA SGF13095984WA SF310220889WA SF310230897WA SF310230901WA SF310230904WA SF310230908WA SF310230911WA SF310230916WA SF310230918WA SF310230919WA SF310230920WA SF310230921WA SF310230922WA SF310230923WA SF310230924WA SF310230925WA SF310230926WA SF310230927WA SF310401714WA SF310401728WA SF310401730WA SF310401734WA SF310401735WA SF310411759WA SF310080491WA SF310090509WA SF311334238WA SF311334292WA SF311344312WA SF310361437WA SF310361441WA SF310361442WA SF310361443WA SF310361444WA SF310361445WA SF310361447WA SF309520166WA SF309520186WA SF309520190WA SJF13110092WA SJF13120116WA SF310050329WA SF310050334WA SF310060389WA SF310060398WA SF310060404WA SF310060413WA SF310060420WA SF310060423WA SF310070431WA SF310070433WA SF310070436WA SF310070437WA SF310070439WA SF310070445WA SF310070450WA SF310070452WA SF310070453WA SF310070455WA SF310070456WA SF310070458WA SF310070460WA SF310070461WA SF310070468WA SF309500145WA SF309500147WA SF309500148WA SF309500149WA SF309500150WA SF309500151WA SF309500152WA SF309500153WA SF309500154WA SF309510155WA SF309510158WA SF309510159WA SF309510160WA SF309510161WA SF309510162WA SF309510163WA SF309510165WA SGF13035367WA SGF13035368WA SGF13035378WA SGF13035379WA SGF13035380WA SGF13035381WA SGF13035382WA SGF13035385WA SGF13035386WA SGF13035387WA SGF13085862WA SGF13085902WA SGF13095909WA SGF13095911WA SGF13095912WA SGF13095913WA SGF13095914WA SGF13095915WA SGF13095916WA SGF13095917WA SGF13095918WA SGF13095919WA SGF13095920WA SGF13095921WA SGF13095922WA SGF13095923WA SGF13095924WA SGF13095925WA SGF13095926WA SGF13095927WA SGF13095928WA SGF13095929WA SGF13095930WA SGF13095931WA SGF13095932WA SGF13095933WA SGF13095934WA SGF13095935WA SGF13095936WA SGF13095937WA SGF13095938WA SGF13095939WA SGF13095940WA SGF13095941WA SGF13095942WA SGF13095944WA SGF13095945WA SF310361448WA SF310361454WA SF310361459WA SF310361460WA SF310361461WA SF310361462WA SF310361463WA SF310361464WA SF310361465WA SF310361466WA SF310361467WA SF310361468WA SF310361469WA SF310361479WA SGF13045545WA SGF13045546WA SGF13045547WA SGF13045580WA SGF13045583WA SGF13045585WA SGF13045587WA SGF13045588WA SGF13045589WA SGF13045590WA SF310200781WA SF310200789WA SGF13085803WA SGF13085805WA SGF13085807WA SGF13085809WA SGF13085810WA SGF13085811WA SGF13085812WA SGF13085813WA SGF13085816WA SGF13085818WA SGF13085787WA SGF13085789WA SGF13085790WA SGF13085791WA SJF13110053WA SJF13110054WA SJF13110055WA SJF13110056WA SJF13110058WA SJF13110061WA SJF13110062WA SJF13110063WA SJF13110068WA SJF13110069WA SJF13110070WA SJF13110071WA SJF13110072WA SJF13110073WA SJF13110075WA SJF13110077WA SJF13110079WA SJF13110089WA SJF13110067WA SJF13110074WA SF311082867WA SF311102881WA SF311233506WA SF311243507WA SF311243509WA SF311243512WA SF311233494WA SF311233497WA SF311233505WA SF311233498WA SF311233501WA SF311233502WA SF311233503WA SF311243511WA SF311243533WA SF311243534WA SF311243535WA SF311243536WA SF311243538WA SF311243539WA SF311233478WA SF311233496WA SF311243513WA SF311243515WA SF311243518WA SF311243519WA SF311243523WA SF311243527WA SF311243530WA SF311243585WA SF311243586WA SF311243587WA SF311243589WA SF311243592WA SF311243593WA SF311243594WA SF311243595WA SF311334226WA SF311334231WA SF311334264WA SF311334273WA SF311334275WA SF311334280WA SF311334284WA SF311334293WA SF311334297WA SF311344304WA SF311344305WA SF311344306WA SF311344308WA SF311344309WA SF311344310WA SF311344318WA SF311344319WA SF311344320WA SF311344321WA SF311344322WA SF311344332WA SF312355995WA SGF12464332WA SGF12474388WA SGF12474389WA SGF12474395WA SGF12474398WA SGF12474421WA SGF12474422WA SGF12474423WA SGF12474424WA SGF12474425WA SGF12474428WA SGF12474429WA SGF12474430WA SGF12474431WA SGF12474432WA SGF12474433WA SGF12474434WA SGF12474435WA SGF12474436WA SGF12474438WA SGF12474439WA SGF12474440WA SGF12474443WA SGF12474444WA SGF12474446WA SG212421055WA SG212421057WA SG212491348WA SG212491349WA SG212491351WA SG212491367WA SG212491368WA SG212491369WA SG212491415WA SG212491428WA SG212491400WA SG212491402WA SG212491403WA SG212491404WA SG212491406WA SG212491407WA SG212491408WA SG212491411WA SG212491412WA SG212481268WA SG212491423WA SG212491427WA SG212501440WA SG212501447WA SG212501448WA SG212501449WA SG212501451WA SG212501452WA SG212501453WA SG212501456WA SGF13035430WA SGF13035452WA SGF13035453WA SGF13045456WA SGF13045461WA SGF13045462WA SGF13045464WA SGF13045467WA SGF13045468WA SGF13045469WA SGF13045470WA SGF13045471WA SGF13045472WA SGF13045473WA SGF13045474WA SGF13045475WA SGF13045476WA SGF13045477WA SGF13045478WA SGF13045479WA SGF13045495WA SGF13045497WA SGF13045498WA SGF13045502WA SGF12434314WA SGF12434326WA SGF12474333WA SGF12474334WA SGF12474338WA SGF12474340WA SGF12474342WA SGF12474344WA SGF12474347WA SGF12474348WA SGF12474349WA SGF12474350WA SGF12424256WA SGF12434323WA SGF12434325WA SGF12474363WA SGF12474366WA SGF12474367WA SGF12474370WA SGF12474371WA SGF12474372WA SGF12474373WA SGF12474392WA SGF12474394WA SGF12474397WA SGF12434311WA SGF12434321WA SGF12474360WA SGF12474361WA SGF12474362WA SGF12474365WA SGF12474368WA SGF12474369WA SGF12484610WA SGF12484616WA SGF12484618WA SGF12484621WA SGF12474383WA SGF12474396WA SGF12484570WA SGF12484577WA SGF12474351WA SGF12474382WA SGF12474386WA SGF12474387WA SGF12474445WA SGF12474391WA SGF12474393WA SGF12494710WA SGF12494711WA SGF12494715WA SGF12494717WA SGF12494719WA SGF12494721WA SGF12494727WA SGF12494729WA SGF12494732WA SGF12494733WA SGF12434305WA SGF12434312WA SGF12434315WA SGF12434304WA SGF12494702WA SGF12484561WA SGF12484682WA SGF12494713WA SGF12494718WA SGF12494722WA SGF12494734WA SGF12494735WA SGF12494724WA SGF12494743WA SGF12494769WA SGF12494770WA SGF12494771WA SGF12494772WA SGF12494773WA SGF12494774WA SGF12494775WA SGF12494776WA SGF12494777WA SGF12494778WA SGF12494779WA SGF12494780WA SGF12494781WA SGF12494789WA SGF12505015WA SGF12505019WA SGF12505023WA SGF12505027WA SGF12515101WA SGF12515107WA SGF12515112WA SGF12515113WA SGF12515115WA SGF12515118WA SGF12515120WA SGF12515122WA SGF12494813WA SGF12494823WA SGF12504860WA SGF12504864WA SGF12504865WA SGF12504867WA SGF12504869WA SGF12504870WA SGF12504871WA SGF12504872WA SGF12504873WA SGF12504884WA SGF12504885WA SGF12504886WA SGF12504888WA SGF12504889WA SGF12504896WA SGF12504898WA SGF12494736WA SGF12504835WA SGF12504837WA SGF12504839WA SGF12504843WA SGF12504851WA SGF12504904WA SGF12504905WA SGF12504906WA SGF12504907WA SGF12504908WA SGF12504909WA SGF12504910WA SGF12504911WA SGF12504912WA SGF12504913WA SGF12504914WA SGF12504915WA SGF12504916WA SGF12504917WA SGF12504918WA SGF12504919WA SGF12504920WA SGF12504922WA SGF12504923WA SGF12504924WA SGF12504925WA SGF12504926WA SGF12504927WA SGF12504928WA SGF12504890WA SGF12504901WA SGF12504930WA SGF12504931WA SGF12504932WA SGF12504933WA SGF12504934WA SGF12504940WA SGF12504943WA SGF12504945WA SGF12504961WA SGF12504965WA SGF12515121WA SGF12504892WA SGF12504894WA SGF12504895WA SGF12504897WA SGF12504899WA SGF12504887WA SGF12504891WA SGF12504893WA SGF12504900WA SGF12504929WA SGF12504935WA SGF12504936WA SGF12504937WA SGF12504938WA SGF12504939WA SGF12504941WA SGF12504944WA SGF12504947WA SGF12504948WA SGF12504951WA SGF12504952WA SGF12504953WA SGF12504956WA SGF12504957WA SGF12504958WA SGF12504968WA SGF12504969WA SGF12504972WA SGF12504973WA SGF12504974WA SGF12504976WA SGF12504983WA SGF12504984WA SGF12504986WA SGF12504942WA SGF12504946WA SGF12504949WA SGF12504950WA SGF12504954WA SGF12504955WA SGF12504959WA SGF12504960WA SGF12504962WA SGF12504963WA SGF12504964WA SGF12504966WA SGF12504967WA SGF12504970WA SGF12504971WA SGF12504975WA SGF12504981WA SGF13106060WA SGF13106066WA SGF13106065WA SGF13106062WA SGF13106071WA SGF13106069WA SGF13106061WA SGF13106067WA SGF13106068WA SGF13106072WA SGF13106070WA SGF13106073WA SGF13106075WA SGF13106076WA SGF13106077WA SGF13106079WA SGF13106081WA SGF13106082WA SGF13106083WA SGF13106084WA SGF13106085WA SGF13106086WA SGF13106087WA SGF13106088WA SGF13106089WA SGF13106090WA SGF13106091WA SGF13106092WA SGF13106093WA SGF13106094WA SGF13106095WA SGF13106096WA SGF13106097WA SGF13106100WA SGF13106128WA SGF13106130WA SGF13106131WA SGF13106132WA SGF13106133WA SGF13106134WA SGF13106138WA SGF13106139WA SGF13106064WA SGF13106154WA SGF13106155WA SGF13106156WA SGF13106157WA SGF13106158WA SGF13106159WA SGF13116160WA SGF13116161WA SGF13116162WA SGF13116164WA SGF13116231WA SGF13116250WA SGF13106135WA SGF13106136WA SGF13106137WA SGF13106140WA SGF13106141WA SGF13106142WA SGF13106143WA SG213091897WA SG213101931WA SG213101933WA SG213101939WA SG213101940WA SG213101941WA SG213101942WA SG213101946WA SG213101948WA SG213101957WA SGF13116227WA SGF13116232WA SGF13116233WA SGF13116234WA SGF13116235WA SGF13116237WA SGF13116238WA SGF13116243WA SGF13116252WA SGF13116253WA SGF13116254WA SGF13116255WA SGF13116257WA SGF13116259WA SGF13116261WA SGF13116262WA SGF13116263WA SGF13116265WA SF310271102WA SF310271103WA SF310271104WA SF310080490WA SF310080495WA SG212491425WA SG212491430WA SG212491431WA SG212501435WA SG212501436WA SF310180616WA SF310180618WA SF310180619WA SF310180622WA SF310180624WA SF310180630WA SF310331394WA SF310341399WA SF310341415WA SF310080483WA SGF13015255WA SGF13025257WA SGF13025258WA SGF13025259WA SGF13025260WA SGF13025291WA SGF13025292WA SGF13025293WA SGF13025294WA SGF13025295WA SGF13025296WA SF310020248WA SF310020253WA SF310020269WA SF310020276WA SF310030278WA SF310030285WA SF310030291WA SF310030292WA SF310030296WA SF310030298WA SF310030300WA SF310030301WA SF310030307WA SF310030308WA SF310030310WA SF310030311WA SF310030321WA SF310030322WA SF310030323WA SF312085119WA SF312085125WA SF310472244WA SF310190736WA SF311213362WA SF310220890WA SF310361453WA SF310361456WA SGF12434294WA SGF12434295WA SGF12434296WA SG213121966WA SG213121967WA SG213121968WA SG213121969WA SG213121971WA SG213121973WA SGF13106057WA SGF13106056WA SGF13035431WA SGF13045521WA SGF13045522WA SGF13045524WA SGF13045526WA SGF13045527WA SGF13045529WA SGF13045531WA SGF13045532WA SGF13045533WA SGF13045534WA SGF13045535WA SGF13045536WA SGF13045537WA SGF13045538WA SGF13045539WA SGF13045540WA SGF13045542WA SGF13055673WA SF310200785WA SF310472261WA SF309380040WA SF309390059WA SF309390061WA SF309390062WA SF309390063WA SF309390064WA SF309390065WA SF309390066WA SF309390067WA SF309390068WA SF309390070WA SF310361474WA SF310361491WA SF310361494WA SF310361496WA SF310361499WA SF310361500WA SF310361502WA SF310361503WA SF310361505WA SF310361506WA SF310361509WA SF310361510WA SF310361513WA SF310361517WA SF310361518WA SF310361519WA SF310361520WA SF310361452WA SF310452108WA SF309520176WA SF309520183WA SF310251087WA SF310251088WA SF310261090WA SF310261092WA SF310261093WA SF310261094WA SF310261095WA SF310271096WA SF310271097WA SF310271099WA SF310271100WA SF310271101WA SF310190732WA SGF13106037WA SGF13106038WA SGF13106040WA SGF13106041WA SGF13106042WA SF312175423WA SF312175431WA SF312175435WA SF312175439WA SF312175441WA SF312195519WA SF312205533WA SF312205535WA SF312205542WA SF312205547WA SF312205548WA SF312205552WA SF312205562WA SF312205563WA SF312205567WA SF312205583WA SF312205597WA SF312205605WA SF312205627WA SF312215638WA SF310230929WA SF310230930WA SF310230931WA SF310230932WA SF309520182WA SF310241004WA SF310241005WA SF310331388WA SF310331348WA SF310130612WA SF310130613WA SF310130614WA SF310180615WA SF310180617WA SGF12474437WA SGF12474449WA SGF12474450WA SGF12474451WA SGF12474457WA SGF12474458WA SGF12474459WA SGF12474461WA SGF12474462WA SGF12474463WA SGF12474464WA SGF12474465WA SGF12474466WA SGF12474467WA SGF12474468WA SGF12474469WA SGF12474470WA SGF12474471WA SGF12474472WA SGF12474473WA SGF12474474WA SGF12474475WA SGF12484481WA SGF12484482WA SGF12484483WA SGF12484484WA SGF12484485WA SGF12484486WA SGF12484488WA SGF12484489WA SGF12484490WA SGF12484492WA SGF12484493WA SGF12484494WA SGF12484496WA SGF12484497WA SGF12484498WA SGF12484499WA SGF12484500WA SGF12484504WA SF310090520WA SF310431936WA SF310452060WA SF310411740WA SF310411746WA SF310411749WA SF310411751WA SF310331378WA SF310331380WA SF310331381WA SF310331385WA SF310331383WA SF310331384WA SF312366026WA SF312366028WA SF312366031WA SF312366032WA SF312366033WA SF312366035WA SF312366036WA SF312366042WA SF312366043WA SF312366045WA SF312376075WA SF312376078WA SF312376084WA SF312376088WA SF312376089WA SF312376091WA SF312376092WA SF312376094WA SF312376096WA SF312376098WA SF312376099WA SF312376101WA SF312376102WA SF312376105WA SF312376112WA SF312376116WA SF312376077WA SF312366046WA SF312366047WA SF312366051WA SF312376056WA SF312376057WA SF312376058WA SF312376060WA SF312376061WA SF312376063WA SF310331390WA SF310080485WA SF310090526WA SF310080480WA SF310030284WA SF310431933WA SF310431937WA SF310431938WA SF310431939WA SF310431940WA SF310431946WA SF310431948WA SF310441994WA SF310441997WA SF310220805WA SF310220839WA SF310220840WA SF310220848WA SF310220850WA SF310220851WA SF310220852WA SF310220853WA SF310220856WA SF310220857WA SF310220859WA SF310220865WA SF310220866WA SF310220867WA SF310220868WA SF311082841WA SF311082827WA SG212491381WA SF311082836WA SGF13055595WA SGF13055601WA SF309520170WA SF310020256WA SF310452094WA SF310452091WA SF310452098WA SF310452100WA SF310452101WA SF310452080WA SF310431930WA SF310431970WA SF310441986WA SF310441987WA SF310452068WA SF310452073WA SF310452074WA SF310452078WA SF310452083WA SF310452087WA SF310452089WA SF310452093WA SF310462165WA SF310361512WA SF310411753WA SF309390060WA SF310200768WA SF311233459WA SG213061802WA SG213061803WA SG213061804WA SG213061805WA SG213061806WA SG213061807WA SF310452110WA SF309500137WA SF310442001WA SF310441999WA SF310442011WA SF310442014WA SF310070465WA SF310070469WA SF310180634WA SGF13055602WA SGF13055603WA SGF13055604WA SGF13055605WA SGF13055606WA SGF13055607WA SGF13055608WA SGF13055609WA SGF13055610WA SGF13055611WA SGF13055612WA SGF13055613WA SGF13055614WA SGF13055615WA SGF13055629WA SGF13055630WA SGF13055631WA SGF13055632WA SGF13055634WA SGF13055636WA SGF13095988WA SGF13095989WA SGF13095990WA SGF13095991WA SGF13095992WA SGF13095993WA SGF13095994WA SGF13095995WA SGF13095996WA SGF13095997WA SGF13106013WA SGF13106014WA SGF13106015WA SGF13106016WA SGF13106017WA SGF13106018WA SGF13106019WA SGF13106020WA SGF13106022WA SGF13106028WA SF311293871WA SF311293875WA SF311303920WA SF311303928WA SF311303975WA SF311314030WA SG213061808WA SF311082838WA SF311082840WA SF311082844WA SF310371634WA SF311233463WA SF309520202WA SF309530211WA SF309530212WA SF310381702WA SF310381704WA SF310391705WA SF310391706WA SF310391707WA SG213091889WA SG212491390WA SG212491391WA SG212501442WA SG212501445WA SG212501457WA SG212501458WA SG212501459WA SG212501460WA SG212501464WA SG212501465WA SG212501466WA SG212511492WA SG212511494WA SG212511495WA SG212511496WA SG212511497WA SG212511498WA SG212511499WA SG212511500WA SG212511501WA SG212511502WA SG212511503WA SG212511504WA SG212511505WA SG212511506WA SG212511507WA SG212511508WA SG212511509WA SG212511510WA SG212511511WA SG212511512WA SG212511514WA SG212511515WA SG212511516WA SG213101935WA SG213101937WA SF310251089WA SF310331397WA SF310341403WA SF310341404WA SF310341409WA SF310341410WA SF310341411WA SF310341412WA SF310341413WA SF310341417WA SF310341419WA SF310341420WA SF310341422WA SF310341423WA SF310341424WA SF310341427WA SF310341430WA SGF12434307WA SGF12474339WA SF310472233WA SF310361458WA SGF13095943WA SGF13095946WA SGF13055696WA SGF13055708WA SGF13055710WA SGF13055712WA SGF13055713WA SGF13055714WA SGF13055715WA SGF13055716WA SGF13055717WA SGF13065743WA SGF13065745WA SGF13065746WA SGF13065750WA SGF13065751WA SG213091892WA SG213091893WA SG213091894WA SG213091895WA SG213091896WA SG213091898WA SG213091899WA SG213091900WA SG213091901WA SG213091902WA SG213091903WA SGF13106048WA SGF13106049WA SGF13106050WA SGF13106051WA SGF13106052WA SGF13106053WA SGF13106054WA SGF13106055WA SGF13106058WA SGF13106059WA SGF13106063WA SF310070464WA SF310070466WA SGF13116260WA SGF13116267WA SGF13116269WA SGF13035383WA SGF13035384WA SGF13035389WA SGF13035390WA SGF13035393WA SGF13035394WA SGF13035396WA SGF13035397WA SGF13106080WA SGF13106078WA SGF11030002WA SGF11030005WA SGF11030007WA SGF12474343WA SGF12474345WA SF310220843WA SF311092872WA SF311102875WA SF311102876WA SF311102879WA SF311102880WA SF311102882WA SF311102883WA SF311112892WA SGF13176620WA SGF13176734WA SF309520167WA SF309520168WA SF309520171WA SF309520180WA SF309380037WA SF309380044WA SF309410077WA SF309410082WA SF309420093WA SF309420103WA SF309420104WA SF309420106WA SF309510164WA SF309520169WA SF309520173WA SF310230915WA SF311072810WA SGF13055718WA SGF13055720WA SGF13055723WA SGF13055725WA SGF13055726WA SGF13055729WA SGF13055730WA SGF13055731WA SGF13055732WA SGF13055733WA SGF13055737WA SGF13065753WA SGF13065754WA SGF13065755WA SGF13065757WA SF312466505WA SF312466512WA SF312466521WA SF312466525WA SF312466528WA SF312466529WA SF312466535WA SG213051771WA SG213051776WA SG213051777WA SG213051778WA SG213051779WA SG213051780WA SG213051781WA SG213061792WA SG213061794WA SG213061795WA SG213061796WA SG213061797WA SG213061798WA SG213061800WA SG213061801WA SGF13055727WA SGF13055728WA SGF13055734WA SGF13055735WA SGF13055736WA SGF13055738WA SGF13055739WA SGF13055740WA SGF13055741WA SGF13055742WA SGF13065756WA SGF13065758WA SGF13065759WA SGF13065760WA SGF13065761WA SGF13065763WA SGF13065764WA SGF13065765WA SGF13065766WA SGF13065767WA SF310431943WA SF310431944WA SF310090512WA SF310090522WA SF310090524WA SF311233480WA SF311233481WA SF311233485WA SF311233490WA SF311233493WA SF311233495WA SF311233504WA SF311243510WA SF311072766WA SF311082828WA SF311082832WA SF311082837WA SF311082843WA SF312376083WA SF311233491WA SF311233492WA SGF12494822WA SGF12494825WA SF310240968WA SF310240970WA SF310240974WA SF310240990WA SF309500136WA SF309500140WA SF309500144WA SF309470107WA SF309470110WA SF309500135WA SF309500139WA SF309500142WA SF309500143WA SF311112904WA SF311112906WA SF311112908WA SF311112914WA SF311112918WA SF311112920WA SF311112921WA SF311112922WA SF311112923WA SF311112924WA SF311112927WA SF311112928WA SF311112929WA SF311112932WA SF311112933WA SF311112934WA SF311112935WA SF311112937WA SF311112938WA SF311112939WA SF311112948WA SF310110563WA SF311052687WA SF311052688WA SF310452076WA SF310452077WA SF310452084WA SF310452085WA SF310452086WA SF310452092WA SF310452097WA SGF12505030WA SGF12515039WA SGF12515041WA SGF12515045WA SGF12515047WA SGF12515048WA SGF12515049WA SGF12515050WA SGF12515051WA SGF12515057WA SGF12515058WA SGF12515059WA SGF12515062WA SGF12515064WA SGF12515065WA SGF12515066WA SGF12515067WA SGF12515076WA SGF12515134WA SGF12515135WA SGF12515136WA SGF12515137WA SGF12515139WA SGF12515042WA SGF12515043WA SGF12515044WA SGF12515046WA SGF12515052WA SGF12515056WA SGF12515077WA SGF12515078WA SGF12515099WA SGF12515100WA SGF12515116WA SGF12515123WA SGF12515125WA SGF12515126WA SGF12515128WA SGF12515129WA SGF13116294WA SGF13126297WA SF310070451WA SF310431974WA SF311494679WA SF311153175WA SGF12434303WA SGF12434319WA SGF13116247WA SF310452081WA SGF13116246WA SGF13116248WA SGF13116256WA SGF13116258WA SGF13116264WA SF310100528WA SF310100534WA SF310100536WA SF310100538WA SF310100542WA SF310100546WA SGF13015245WA SGF13015248WA SGF13015253WA SGF13015254WA SGF13015256WA SF311082835WA SF311112941WA SF311112945WA SF311112946WA SF311112947WA SF311112950WA SF311112951WA SF310030320WA SGF13116249WA SGF13126296WA SGF13126299WA SGF13126301WA SGF13126302WA SGF13126303WA SGF13126304WA SGF13126305WA SGF13126307WA SGF13126308WA SGF13126309WA SF310331369WA SJF13120112WA SJF13120113WA SJF13120114WA SJF13120115WA SJF13120117WA SF311072798WA SGF13126375WA SGF13126378WA SGF13126380WA SGF13126382WA SGF13126384WA SGF13126385WA SGF13126387WA SGF13126389WA SGF13126390WA SGF12434306WA SGF12434309WA SGF12474335WA SF310020250WA SF309530231WA SF310020233WA SF310020234WA SF310020235WA SF310020237WA SF310020238WA SF310020239WA SF310020241WA SF310020242WA SF310020243WA SF310020247WA SF310020251WA SF310020258WA SF310020259WA SF310020261WA SF310020263WA SF310020264WA SF310020265WA SF310020267WA SF310020268WA SF310020270WA SF310020271WA SF310020272WA SF310020273WA SF310020275WA SF310331361WA SF310331364WA SF310331365WA SF310331366WA SF310331367WA SF310331368WA SF310331370WA SF310331372WA SF310331374WA SF310331375WA SF310331376WA SF310331377WA SF311193309WA SGF12132249WA SF311052700WA SF311052701WA SF311072711WA SF311072712WA SF311072713WA SF311072714WA SF311072716WA SF311072718WA SF311072720WA SF311072723WA SF311072725WA SF311072726WA SF311072730WA SF311072741WA SF311072742WA SF311334277WA SF311112944WA SGF13116266WA SF311233447WA SF311233449WA SF310030293WA SF310472203WA SGF13035427WA SF309410072WA SF311082819WA SF311082820WA SF311082821WA SF311082825WA SF311082829WA SF311243532WA SGF13055648WA SGF13055649WA SGF13055650WA SGF13055651WA SGF13055652WA SGF13055654WA SGF13055655WA SGF13055656WA SGF13055657WA SGF13055659WA SF310100543WA SF310110562WA SF310110564WA SF310110570WA SGF12424211WA SGF12424216WA SGF12424217WA SGF12424218WA SGF12424219WA SF310371617WA SF310371622WA SF310371624WA SGF13055703WA SGF13055704WA SGF13055705WA SGF13055709WA SGF13055711WA SGF13065768WA SGF13065769WA SGF13065770WA SGF13065771WA SGF13065772WA SJF13110078WA SJF13110082WA SJF13110084WA SJF13110085WA SJF13110088WA SJF13110090WA SJF13120118WA SJF13120119WA SJF13120120WA SGF13176575WA SGF13176581WA SGF13176583WA SGF13176585WA SGF13176586WA SGF13176588WA SGF13176590WA SGF13176591WA SGF13176592WA SGF13176593WA SGF13176594WA SGF13176595WA SGF13176596WA SGF13176597WA SGF13176598WA SGF13176599WA SGF13176600WA SGF13176603WA SGF13176605WA SGF13176607WA SF310090470WA SF310090502WA SF310090503WA SF310090504WA SF310060343WA SF310060345WA SF310060354WA SF310100547WA SF310100550WA SF310110558WA SF310110560WA SF310110573WA SF310110574WA SF310110575WA SF310110576WA SF310110577WA SF310110578WA SF310110581WA SF310110568WA SF310381663WA SF310442003WA SF310442005WA SF310442009WA SF310442012WA SF310442013WA SF310442016WA SF310452030WA SF310462152WA SF310472220WA SF310472221WA SF310472223WA SF310472252WA SF310472255WA SF310472256WA SF310472257WA SF310472258WA SF310482272WA SF311052695WA SF311062703WA SF311062707WA SF311062705WA SF311072729WA SF311072751WA SF311082826WA SF311082822WA SF311082823WA SF311072756WA SF311072795WA SF311112899WA SF311193294WA SF311193297WA SF311193296WA SF311193299WA SF311193300WA SF311193301WA SF311193302WA SF311193303WA SF311193304WA SF311193305WA SF311193306WA SF311193313WA SF311213361WA SF311213369WA SF311213370WA SF311213371WA SF311213373WA SF311213380WA SF311213381WA SF311213382WA SF311213383WA SF311213384WA SF311213385WA SF311213386WA SF311213387WA SF311213391WA SF311213392WA SF311223394WA SF311223395WA SF311223397WA SF311223401WA SF311233454WA SF311233467WA SF311233469WA SF311233471WA SF311233473WA SF311233477WA SF311233450WA SF311233475WA SF311233474WA SF311233479WA SF311233482WA SF311233486WA SF311233488WA SF311243516WA SF311243521WA SF311243531WA SGF12484664WA SG213101944WA SG213101945WA SG213101949WA SG213101950WA SG213101951WA SG213101952WA SGF13116274WA SGF13116275WA SGF13116276WA SGF13116277WA SGF13116278WA SGF13116279WA SGF13116280WA SGF13116281WA SGF13116282WA SGF13116283WA SGF13116284WA SGF13116285WA SGF13116286WA SGF13116287WA SGF13116288WA SGF13116289WA SGF13116290WA SGF13116291WA SGF13116293WA SGF13116295WA SF310200763WA SF310190721WA SF310190724WA SF310190726WA SF310190728WA SF310190729WA SF310190730WA SF310190733WA SF310190734WA SF310190735WA SF310190737WA SF310190738WA SF310190739WA SF310190740WA SF310190744WA SF310190745WA SF310190746WA SF310190748WA SF310190749WA SF310200764WA SF310230905WA SF310230906WA SF310230907WA SF310230909WA SF310230910WA SF310230913WA SF310230914WA SF310371579WA SF310371625WA SF310371626WA SF310371628WA SF310371630WA SF310371635WA SF310371636WA SF310371637WA SF310371639WA SF310371641WA SF310371642WA SF310371643WA SF310371645WA SF310371647WA SF310381700WA SF310381703WA SF310462139WA SF310472197WA SF310472199WA SF310472202WA SF310462176WA SF310462178WA SF310472237WA SF310472238WA SF310472245WA SF310472248WA SF310472249WA SF310371616WA SF310371619WA SF310371620WA SF310371621WA SF310371627WA SF309380050WA SF309390054WA SF309530220WA SF310020236WA SF310020254WA SF310020260WA SF310020262WA SF309410076WA SF309410087WA SF309410088WA SF309410091WA SF309410092WA SGF12071653WA SGF12091878WA SGF12112081WA SGF12112088WA SGF12112090WA SG212260728WA SG212260729WA SG212260730WA SG212260733WA SG212280752WA SGF13116268WA SGF13116270WA SGF13116271WA SGF13116272WA SGF13116273WA SGF13126306WA SGF13126310WA SGF13126311WA SGF13126312WA SGF13126328WA SF312315835WA SF312315839WA SF312315843WA SF312315844WA SF312315845WA SF312315846WA SF312315847WA SF312315849WA SF312315850WA SF312315851WA SF312315853WA SF312315854WA SF312315855WA SF312315860WA SF312315861WA SF312315862WA SF312315863WA SF312315867WA SF312315868WA SF312315870WA SF312315871WA SF312315872WA SF312315873WA SF312315874WA SF312315875WA SF312325876WA SF312325877WA SF312325878WA SF312325879WA SF312325880WA SF312325881WA SF312325882WA SF312325885WA SF312325887WA SF312325889WA SF312325890WA SF312325892WA SF312315832WA SF311112931WA SF311334242WA SF31133426WA SF311314087WA SF311324126WA SGF12071620WA SGF12263227WA SG212270748WA SG212270750WA SF310100540WA SF310120591WA SF310120593WA SF310120595WA SF310120596WA SF310120599WA SF310120602WA SF310120603WA SF311334272WA SF311334276WA SG212410961WA SG212410962WA SG212410963WA SG212410964WA SG212410965WA SG212410967WA SG212410969WA SG212410970WA SGF11490519WA SGF12021247WA SGF12091802WA SGF12091844WA SGF12091848WA SGF12101965WA SGF12101977WA SGF12101979WA SGF12101983WA SGF12101988WA SGF12101989WA SGF12101991WA SGF12101994WA SG211500216WA SGF11490511WA SGF12081700WA SGF12091901WA SGF12102037WA SG212310797WA SG212310798WA SG212310800WA SG212320847WA SG212320848WA SG212320849WA SG212340855WA SG212340856WA SG212340857WA SG212340858WA SG212340859WA SG212340860WA SGF12424161WA SF311143085WA SF311143103WA SF311143105WA SF311143132WA SF311283782WA SF311283783WA SF311283784WA SF311303897WA SF311303910WA SF311334251WA SF311404362WA SF311414443WA SF312034907WA SF312054967WA SF312064999WA SF312065002WA SF312065007WA SF312205606WA SGF13085863WA SF312205546WA SF312205566WA SGF13106074WA SF313137047WA SF313137050WA SF312335977WA SF312335981WA SGF13126323WA SGF13126324WA SGF13126325WA SGF12212619WA SGF12232757WA SGF12021246WA SG211430103WA SG211450161WA SG211450164WA SGF11430191WA SGF11440285WA SGF11440295WA SGF11440296WA SGF11450345WA SGF11521083WA SG212150555WA SG212110473WA SG212150559WA SG212150552WA SG212150553WA SG211450162WA SGF12142297WA SGF12021217WA SGF12021258WA SGF11440294WA SGF12263201WA SGF12373821WA SG212070430WA SGF12031339WA SGF12031340WA SGF12031341WA SGF12031342WA SGF12031344WA SGF12031346WA SGF12031347WA SGF12031348WA SG211450163WA SGF11450311WA SGF11450341WA SGF11450342WA SGF12081682WA SGF12081693WA SGF12323488WA SG211430108WA SF310361549WA SF310361553WA SF310371562WA SF310371587WA SGF12222662WA SGF12263231WA SGF12263232WA SGF12273252WA SGF12313404WA SGF12363647WA SGF12363660WA SG212020382WA SG212110474WA SG212150563WA SG212120490WA SG212310808WA SG212310809WA SGF12152364WA SGF11440292WA SGF12152412WA SGF12152416WA SGF12152417WA SGF12152421WA SGF12011134WA SGF12031265WA SGF12031306WA SGF12152414WA SGF12152415WA SG212290762WA SG212290763WA SG212290764WA SG212290765WA SGF12011158WA SGF12011170WA SGF12011178WA SGF12021243WA SGF12051429WA SG211450157WA SG211450158WA SGF12101972WA SGF12101997WA SGF11450338WA SGF12101970WA SGF12101974WA SGF12101992WA SGF12102014WA SGF12142278WA SGF12101969WA SF310482304WA SF310482307WA SF310482308WA SF310482310WA SF310482311WA SF310472208WA SF310502386WA SF310502387WA SF310502388WA SF310502389WA SF310502392WA SF310512473WA SF311042660WA SF311042663WA SF311223425WA SF311223426WA SF311223427WA SF312085077WA SF312085081WA SF312085129WA SF312105221WA SF312155274WA SF312155316WA SF312205540WA SF312205560WA SF312325909WA SF312325918WA SF312325920WA SF312366027WA SF312366048WA SF312376113WA SF312416226WA SG212410994WA SG212411001WA SGF12424268WA SG212491345WA SG212491383WA SG212491418WA SG212491421WA SGF12484594WA SGF12484595WA SGF12484600WA SGF12484602WA SGF12484644WA SGF12484646WA SGF12484656WA SGF12484607WA SGF12484608WA SGF12484613WA SGF12484614WA SGF12494701WA SGF12494795WA SGF12494821WA SGF12494827WA SGF12494828WA SGF12494829WA SGF12494830WA SGF12494831WA SGF12504836WA SGF12504840WA SGF12504841WA SGF12494818WA SGF12494820WA SGF12494826WA SGF12494832WA SGF12515154WA SGF12515157WA SGF12515158WA SGF12504921WA SGF12515131WA SGF12515177WA SGF12515179WA SGF12515181WA SGF12525209WA SGF12525211WA SGF12525212WA SGF12525213WA SGF12525214WA SGF12525215WA SGF12525216WA SGF12515164WA SGF12515167WA SGF12515169WA SGF12515170WA SGF12515173WA SGF12515175WA SGF12515176WA SGF12525182WA SGF12525183WA SGF12525185WA SGF12525186WA SGF12525188WA SGF12515174WA SGF13015243WA SGF13015244WA SGF13015246WA SGF13015247WA SGF13015249WA SGF13015250WA SGF12515089WA SGF13055668WA SGF13055670WA SGF13055692WA SGF13065744WA SGF13085781WA SGF13085782WA SGF13085799WA SG213091870WA SG213091871WA SG213091874WA SG213091875WA SG213091876WA SG213091877WA SG213091878WA SG213091879WA SG213091880WA SG213091881WA SF313066937WA SGF13085831WA SGF13085832WA SGF13116241WA SGF13116245WA SGF13126329WA SGF13126300WA SGF13126326WA SGF13136511WA SGF13136512WA SGF13136513WA SGF13136514WA SJF13110052WA SJF13110057WA SJF13110066WA SJF13110086WA SJF13120093WA SJF13130128WA SJF13130129WA SG213131986WA SG213131987WA SF313147071WA SF313147070WA SF309490118WA SF309490119WA SF310301196WA SF310311259WA SF310311260WA SF310311261WA SF310462131WA SF310452095WA SF310462128WA SF310512454WA SF310512455WA SF310512456WA SF310512457WA SF310512459WA SF311153143WA SF311153181WA SF311153188WA SF311153191WA SF311173236WA SF311183239WA SF311183245WA SF311183247WA SF311183248WA SF311183250WA SF311233448WA SF311293852WA SF311293854WA SF311303894WA SF311303900WA SF311303929WA SF311314017WA SF311293866WA SF311303971WA SF311303973WA SF311303976WA SF311303978WA SF311314019WA SF310060375WA SF310060386WA SF310060395WA SF310060396WA SF310060399WA SF310060379WA SF310060388WA SF310060391WA SF310060392WA SF310462171WA SF310462174WA SF310462144WA SF310462166WA SF311072792WA SF311494653WA SF311494657WA SF311494674WA SF312295800WA SF312295792WA SF312295807WA SF312356006WA SGF12414082WA SGF12414083WA SGF12414086WA SGF12424190WA SGF12424191WA SGF12424192WA SGF12424193WA SGF12424194WA SGF12424196WA SGF12525187WA SG212521536WA SGF13176678WA SGF13176689WA SGF13176692WA SGF13176693WA SGF13176675WA SGF13176680WA SGF13176681WA SGF13176682WA SGF13176683WA SGF13176684WA SGF12242846WA SF311414423WA SF311414424WA SG212431121WA SF313117046WA SF312095145WA SF312245655WA SF310100548WA SF310452022WA SF310452027WA SF310452028WA SF310452031WA SF310452057WA SF310452059WA SF310452061WA SF310452064WA SF310452067WA SF310462167WA SF310512431WA SF310512432WA SF310512435WA SF310512437WA SF310512438WA SF310512441WA SF311133047WA SF311112962WA SF311123022WA SF311123024WA SF311133026WA SF311133027WA SF311133029WA SF311133030WA SF311133031WA SF311133032WA SF311133033WA SF311133034WA SF311133035WA SF311133036WA SF311133037WA SF311133038WA SF311133041WA SF311133042WA SF311133043WA SF311133044WA SF311133045WA SF311133048WA SF311133039WA SF311133040WA SF311133046WA SF311082861WA SF311122980WA SF311153149WA SF311153150WA SF311153165WA SF311153166WA SF311153167WA SF311153176WA SF311153153WA SF311153156WA SF311153159WA SF311153160WA SF311153161WA SF311153168WA SF311153172WA SF311153170WA SF311434515WA SF311474584WA SF311474593WA SF311524850WA SF311524841WA SF312024866WA SF312024870WA SF312024872WA SF312024873WA SF312024874WA SF312024877WA SF312024878WA SF312024879WA SF312024880WA SF312024881WA SF312095152WA SF312095154WA SF312095155WA SF312095156WA SF312095158WA SF312095177WA SF312095195WA SF312105230WA SF312105232WA SF312165370WA SF312165373WA SF312165385WA SF312165388WA SF312165395WA SF312205578WA SF312205580WA SF312205593WA SF312205598WA SF312205616WA SF312205620WA SF312205537WA SF312205622WA SF312205536WA SF312205544WA SF312275730WA SF312275745WA SF312295785WA SF312295786WA SF312295788WA SF312295789WA SF312295790WA SF312295791WA SF312295795WA SF312295797WA SF312295798WA SF312335940WA SF312335946WA SF312335947WA SF312335982WA SF312335983WA SF312335928WA SF312335930WA SF312335942WA SF312335958WA SF312335962WA SF312396181WA SF312396182WA SF312396184WA SF312396186WA SF312396188WA SF312416233WA SF312436289WA SF312466543WA SF312476585WA SF312476592WA SF312496602WA SF312496648WA SF312496653WA SF312496654WA SF312496657WA SF312496660WA SF313046836WA SF313046837WA SF313046838WA SF313046840WA SF313046841WA SF313046846WA SF313117024WA SF313117027WA SF313117029WA SF313117031WA SF313117043WA SF312205631WA SF311112888WA SF311112890WA SF311334289WA SGF13085800WA SF310180665WA SF310452104WA SF310452106WA SF310452114WA SF310452115WA SF310452117WA SF310180621WA SF310180623WA SF310180627WA SF310180631WA SF310180638WA SF310180643WA SF310180645WA SF310180646WA SF310180647WA SF310180648WA SF310180649WA SF310180650WA SF310180651WA SF310180652WA SF310180653WA SF310180654WA SF310180655WA SF310180656WA SF310180657WA SF310180658WA SF310180659WA SF310180660WA SF310180661WA SF310180662WA SF310180663WA SF310180664WA SF310180666WA SF310180668WA SF310180669WA SF310180670WA SF310180671WA SF310180672WA SF310180673WA SF310180674WA SGF12142272WA SGF12142277WA SGF12152429WA SGF12162435WA SGF12162452WA SGF12242772WA SGF12212633WA SGF12212635WA SGF12212636WA SGF12212613WA SGF12212629WA SGF12212630WA SGF12212631WA SGF12212632WA SGF12162441WA SGF12172506WA SG212481197WA SG212481202WA SG212481225WA SG212481227WA SG212481228WA SGF12424164WA SGF12474406WA SGF12474476WA SGF12474477WA SGF13095905WA SGF13106045WA SGF13106046WA SGF13106047WA SJF13110064WA SJF13110065WA SJF13110080WA SJF13130124WA SJF13130126WA SGF12142280WA SGF12142303WA SGF12152351WA SGF12323486WA SGF12323494WA SGF12323495WA SGF12323497WA SF309410080WA SGF12424184WA SGF12424186WA SGF12424187WA SGF12424188WA SGF12424200WA SGF12424201WA SGF12424202WA SGF12424203WA SGF12424207WA SGF12424208WA SGF13055639WA SGF13085780WA SGF13085783WA SGF13085784WA SGF13085785WA SGF13085786WA SGF13085788WA SGF11340045WA SGF11350048WA SG211490213WA SG211510231WA SG211510232WA SG211510239WA SG211510241WA SGF11470432WA SGF11510628WA SGF11510641WA SGF11510652WA SGF11510656WA SGF11510690WA SGF11510691WA SGF11510692WA SGF11510693WA SGF11510694WA SG212090444WA SG212150571WA SG212150572WA SG212150575WA SG212150579WA SGF12122241WA SGF12122242WA SGF12152402WA SGF12152405WA SGF12152406WA SGF12333510WA SGF12333512WA SGF12333522WA SGF12333524WA SGF12333530WA SG212290772WA SG212290776WA SG212320827WA SG212320829WA SG212320830WA SG212320832WA SG212320834WA SG212320835WA SG212320836WA SG212320841WA SF309530219WA SF309530221WA SF309530224WA SF309530225WA SF309420094WA SF309420095WA SF309420096WA SF309420097WA SF309380047WA SF309380048WA SF309380049WA SF309390052WA SF309390053WA SF309390056WA SF309390058WA SF311072777WA SG212380912WA SG212380913WA SG212380914WA SGF12383904WA SGF12383905WA SGF12383906WA SGF12263194WA SGF12263203WA SGF12021255WA SGF12021262WA SGF12031266WA SGF12031271WA SGF12031272WA SGF12031273WA SG212320826WA SGF12484651WA SGF12484670WA SGF12484671WA SGF12484615WA SGF12494725WA SGF12494790WA SGF12494791WA SGF12494792WA SGF12494793WA SGF12494794WA SGF12494796WA SGF12494797WA SGF13055598WA SF311494654WA SF311494659WA SF311494676WA SF310100529WA SF311022577WA SF311022579WA SF311022581WA SF311022583WA SF311022584WA SF311022586WA SF311022587WA SF311022590WA SF311022591WA SF311022592WA SF311022594WA SF311022595WA SF311022596WA SF311022597WA SF311022598WA SF311022599WA SF311022600WA SF311022601WA SF311022602WA SF311022603WA SF311022604WA SF311022605WA SF311022606WA SF311022607WA SF311022608WA SF311022609WA SF311022610WA SF311022611WA SF311022612WA SF311022613WA SF311022614WA SF311022615WA SF311022616WA SF311022617WA SF311022618WA SF311022619WA SF311022620WA SF311022621WA SF311022622WA SF311022623WA SF311022624WA SF311022625WA SF311022626WA SF311042629WA SF311042633WA SF311042634WA SF311042635WA SF311042636WA SF311042637WA SF310522560WA SF311022589WA SF310100530WA SGF12142309WA SGF12142315WA SGF12263199WA SGF12273240WA SF311324182WA SGF12132250WA SGF12132252WA SGF12132256WA SGF12132258WA SGF12132263WA SGF12132267WA SGF12273250WA SG212360892WA SG212360893WA SGF12313381WA SF310240946WA SF310240947WA SF310240948WA SF310240951WA SF310240952WA SF311303902WA SF310452109WA SF310452112WA SF310452113WA SF310452116WA SF310452118WA SGF13095985WA SGF13095986WA SGF12091922WA SGF12112059WA SGF12112075WA SGF12112078WA SGF12112079WA SGF12112102WA SGF12112107WA SGF12051432WA SGF12091863WA SGF12102024WA SGF12031327WA SGF12051431WA SGF12091826WA SGF12091940WA SGF12101978WA SGF12102017WA SGF12102018WA SGF12112061WA SGF12112084WA SGF12091838WA SGF11521088WA SGF12011141WA SGF12011142WA SGF12011143WA SG212020353WA SGF12051403WA SGF12051454WA SF312054943WA SF310030305WA SF310100539WA SF310100541WA SF310100552WA SF310220807WA SF310431918WA SF310431912WA SF310472229WA SF310472230WA SF310472231WA SF310472235WA SF310472236WA SF310472241WA SF310472242WA SF310482271WA SF310482274WA SF310482315WA SF310482316WA SF311233460WA SF311233461WA SF311233465WA SF311283775WA SF311293862WA SF311314036WA SF312024886WA SF312024897WA SF312024901WA SF312024905WA SF312034920WA SF312085063WA SF312085072WA SF312085073WA SF312085074WA SF312085075WA SF312085080WA SF312085052WA SF312085053WA SF312085054WA SF312085055WA SF312085057WA SF312085059WA SF312085060WA SF312085061WA SF312085062WA SF312085064WA SF312085065WA SF312105233WA SF312155329WA SGF12504990WA SGF12504991WA SJF13180144WA SF311112901WA SG213121960WA SG213121978WA SG211360038WA SGF12091835WA SG211400073WA SG211400074WA SG211400075WA SG211400076WA SG211450142WA SG211450143WA SG211450144WA SG211450145WA SG211450146WA SG211450148WA SG211450149WA SG211450150WA SG211450151WA SG211510240WA SF312205585WA SF312195523WA SF311123004WA SF311123023WA SF312205604WA SF312195479WA SF312195488WA SF312245644WA SF312245653WA SF312205592WA SF312205624WA SF312205628WA SF312245640WA SF312245643WA SF312315848WA SF312396139WA SG212431125WA SG212431126WA SF312466515WA SG213021591WA SG213021592WA SG213021593WA SG213021594WA SG213021595WA SG213021596WA SG213021597WA SG213021598WA SG213021599WA SG213021600WA SG213021601WA SG213021603WA SG213021604WA SG213021606WA SG213021569WA SG213091891WA SG213101959WA SG213121970WA SG213121974WA SG213121975WA SG213121977WA SG213131980WA SG213131981WA SG213131982WA SG213131983WA SG213131984WA SF311253680WA SF310090499WA SF311143097WA SF311153151WA SF311253691WA SF311263705WA SF311263711WA SF311253690WA SF311263710WA SF311263713WA SF311263701WA SF311253698WA SF311263702WA SF311494651WA SF311123012WA SF311123015WA SG213121965WA SG213121972WA SF311324193WA SF311454562WA SF311484604WA SF311484609WA SF311494616WA SF311494621WA SF311494630WA SF311494633WA SG212481217WA SG212491371WA SGF12484562WA SGF12484563WA SGF13025348WA SG212150570WA SG212050417WA SG212110472WA SG212260713WA SG212260732WA SG212260734WA SG212070418WA SG212110464WA SG212110465WA SG212110466WA SG212110468WA SGF12253113WA SF312396175WA SF312396176WA SF312396178WA SF312396179WA SF310241016WA SF310431923WA SF310452033WA SF310452065WA SF310452075WA SF310452079WA SF310462159WA SF310462160WA SF310462162WA SF310462164WA SF310472250WA SF310472253WA SF310472254WA SF310472259WA SF311143135WA SF311153136WA SF311293832WA SF311334249WA SF311334253WA SF312034919WA SF312034927WA SF312034928WA SF312075039WA SF312064998WA SF312095190WA SF312115236WA SF312155328WA SF312195473WA SF312335966WA SF312506731WA SGF13035402WA SG213091904WA SG213101911WA SG213101912WA SG213101913WA SG213101910WA SG213101914WA SGF13136462WA SG212481207WA SG212481209WA SG212481187WA SG212481201WA SF312466497WA SF312466500WA SF310462158WA SF311334298WA SF311414447WA SF311434534WA SF311434537WA SF311434538WA SF311474580WA SF312265695WA SGF13035454WA SGF12253065WA SGF12253066WA SGF12263129WA SGF12263131WA SGF12263132WA SGF12273238WA SGF12273258WA SGF12273261WA SGF12333541WA SGF12252961WA SGF12252963WA SGF12263130WA SGF12263169WA SGF12263170WA SGF12263176WA SGF12273260WA SF310220827WA SF310220870WA SF310220871WA SF310220872WA SF310220873WA SF310220874WA SF310220875WA SF310220876WA SF310220877WA SF310220879WA SF310220881WA SF310220882WA SF310220883WA SF310220884WA SF310220887WA SF311283813WA SF311283814WA SF311283816WA SF311283824WA SF311283826WA SF311283828WA SF311293831WA SF311293833WA SF311293834WA SF311293835WA SF311293837WA SF311293838WA SF311293841WA SF311293842WA SF311293845WA SF311293848WA SF311293849WA SF311293851WA SF311293858WA SF311303895WA SF311303912WA SF311303917WA SF311303919WA SF311303922WA SF311303959WA SF312175450WA SF312175432WA SF312175433WA SF312175436WA SF312175437WA SF312175443WA SF312175445WA SF312175448WA SF312175449WA SF312175451WA SF312175453WA SGF12515162WA SGF12515163WA SGF12515165WA SGF12515168WA SGF12515172WA SGF12515178WA SGF12515180WA SGF12525229WA SGF12525230WA SGF12525231WA SGF12525232WA SGF12525233WA SGF12525234WA SGF12525235WA SGF12525236WA SGF12525237WA SGF12525239WA SGF12525240WA SGF12525241WA SGF12525242WA SG212491429WA SGF12504838WA SGF13146518WA SGF13146520WA SGF13146521WA SGF13146522WA SGF13146523WA SGF13146524WA SGF13146525WA SGF13146526WA SGF13146527WA SGF13146528WA SGF13146529WA SGF13146530WA SGF13146531WA SGF13146532WA SGF13146533WA SGF13146534WA SGF13146535WA SGF13146536WA SGF13146537WA SGF13146538WA SGF13146540WA SGF13146541WA SGF13146542WA SGF13176543WA SGF13176574WA SGF13176577WA SGF13176578WA SGF13176579WA SGF13176580WA SGF13176582WA SGF12142270WA SGF12142304WA SG212150589WA SG212150592WA SF310361457WA SGF12263198WA SG212260711WA SF310452032WA SF310452096WA SF310452099WA SF310452102WA SF311414453WA SF311434543WA SF312265676WA SGF13176556WA SGF13176687WA SGF13176688WA SG211470200WA SG211510233WA SGF11490510WA SGF11510623WA SGF11510689WA SG211450147WA SG211490200WA SG211490211WA SG212020371WA SG212090441WA SG212090443WA SG212090445WA SG212090446WA SG212090447WA SG212090448WA SGF12071638WA SGF12071639WA SGF12071645WA SGF12071649WA SGF12071650WA SGF12071651WA SGF12071652WA SGF12071654WA SGF12071655WA SGF12071656WA SGF12081659WA SGF12081660WA SGF12081661WA SGF12081663WA SGF12081772WA SGF12152394WA SGF12152403WA SGF12162456WA SGF12162463WA SGF12162466WA SGF12162467WA SGF12162468WA SGF12162469WA SGF12172509WA SGF12172520WA SGF12172521WA SGF12172522WA SGF12172526WA SGF12172529WA SGF12172530WA SGF12172533WA SGF12172539WA SGF12172548WA SGF12172556WA SGF12172582WA SGF12172587WA SGF12212623WA SGF12212624WA SGF12212625WA SGF12212628WA SF309380033WA SF309380034WA SF309380035WA SF309380036WA SF309380038WA SF309380039WA SF309380041WA SF309380042WA SF309380045WA SF309380051WA SF309530222WA SF309530226WA SF309530227WA SF309530228WA SF309530229WA SF310311227WA SF310311250WA SF310311270WA SF310311276WA SF310311279WA SF310311280WA SF310311281WA SF310311285WA SF310311287WA SF310311288WA SF310311289WA SF310311290WA SF310311291WA SF310311292WA SF310311293WA SF310311296WA SF310311298WA SF310311300WA SF310311302WA SF310321310WA SGF11380087WA SG212360877WA SF310361528WA SG211510238WA SGF11340046WA SGF11380088WA SGF11390098WA SGF11390117WA SG211360043WA SG211390054WA SG211390056WA SG211390064WA SG211390065WA SG211390066WA SG211360042WA SG211360044WA SG211400067WA SG211430109WA SG211430110WA SG211440127WA SG211440129WA SG211440134WA SG211440135WA SG211440136WA SGF11350050WA SGF11350051WA SGF11360059WA SGF11360060WA SGF11360061WA SGF11370078WA SGF11410126WA SGF11440286WA SGF11440287WA SGF11440288WA SGF11440289WA SGF11440290WA SGF11440291WA SGF11440293WA SGF11450337WA SGF11450339WA SGF11450340WA SG211400068WA SG211400069WA SG211400070WA SG211400071WA SG211400072WA SGF11350054WA SGF11410125WA SGF11430184WA SGF11430185WA SGF11430186WA SGF11430192WA SGF11430193WA SGF11440268WA SGF11440281WA SGF11440284WA SGF11450297WA SGF11450299WA SG211450166WA SG211450167WA SG211450168WA SG211450169WA SG211450171WA SG211460179WA SG211460180WA SG211460182WA SG211460185WA SG211460187WA SGF12313405WA SGF12413998WA SGF12413999WA SGF12414000WA SGF12414003WA SGF12383917WA SGF12383918WA SGF12383920WA SGF12383922WA SGF12383923WA SGF12383928WA SGF12383930WA SGF12414013WA SG212511521WA SG212511523WA SG212511524WA SG212511526WA SG212521534WA SG212521535WA SG212521537WA SGF11370072WA SGF11370073WA SGF11370070WA SGF11370071WA SGF11370074WA SGF11350053WA SF309390071WA SF309410090WA SF310070435WA SF310070454WA SF310070457WA SF310080471WA SF310080472WA SF310080474WA SF310080475WA SF310080476WA SF310080478WA SF310080479WA SF310080482WA SF310080487WA SF310080489WA SF310080496WA SF310070462WA SF310421851WA SF310421854WA SF310421862WA SF310431863WA SF310431864WA SF310431865WA SF310431866WA SF310431867WA SF310321325WA SF310331329WA SF310331331WA SF310331342WA SF310321328WA SF310331333WA SF310331336WA SF310331338WA SF310331339WA SF310331340WA SF310331341WA SF310331344WA SF310442017WA SF310452050WA SF310452054WA SF310452088WA SF311414456WA SF311424461WA SF311424465WA SF311424482WA SF311434536WA SGF13116210WA SJF13030047WA SF311072735WA SF311072736WA SF311072737WA SF311072738WA SF311072739WA SF311072744WA SF311072768WA SF311072769WA SF311072770WA SF311072773WA SF311072774WA SF311072781WA SGF12011137WA SGF12011139WA SGF12011140WA SG211520260WA SG211520261WA SG211520263WA SG211520264WA SG211520265WA SG211520269WA SG212010276WA SG212010277WA SG212010278WA SG212010279WA SG212010280WA SG212010281WA SG212010282WA SG212010284WA SG212010285WA SG212010286WA SG212010287WA SG212010288WA SG212010291WA SG212010292WA SG212010293WA SG212010294WA SG212010297WA SG212010298WA SG212010300WA SG212010301WA SG212010302WA SG212010303WA SG212010304WA SG212010305WA SG212010306WA SG212010307WA SG212010308WA SG212010309WA SG212010310WA SG212010311WA SG212010312WA SG212010313WA SG212010314WA SG212010315WA SG212010316WA SG212010317WA SG212010318WA SG212010319WA SG212010320WA SG212010321WA SG212010322WA SG212010323WA SG212010324WA SG212010325WA SG212010326WA SG212010327WA SG212010328WA SG212010329WA SG212010330WA SG212010331WA SG212010332WA SG212020333WA SG212020334WA SG212020335WA SG212020336WA SG212020337WA SG212020338WA SG212020339WA SG212020340WA SG212020341WA SG212020342WA SG212020343WA SG212020344WA SG212020345WA SG212020346WA SG212020347WA SG212020348WA SG212020349WA SG212020350WA SG212020352WA SG212020354WA SG212020356WA SGF12011144WA SGF12011145WA SGF12011150WA SGF12011151WA SGF12011152WA SGF12011154WA SGF12011155WA SGF12011156WA SGF12011157WA SGF12011159WA SGF12011160WA SGF12011161WA SGF12011162WA SGF12011164WA SGF12011165WA SGF12011166WA SGF12011099WA SGF12011112WA SGF12021223WA SGF12021227WA SGF12021229WA SGF12021230WA SGF12021231WA SGF12021232WA SGF12021233WA SGF12021234WA SGF12021235WA SGF12021236WA SG211520259WA SG211520275WA SG212020357WA SG212020360WA SG212020362WA SG212020364WA SGF12383831WA SF310311220WA SF310311230WA SF310311232WA SF310311233WA SF310311234WA SF310311235WA SF310311236WA SF310311238WA SF310311239WA SF310311242WA SF310311262WA SF310311263WA SF310311264WA SF310311267WA SF310311268WA SF310311271WA SF310311272WA SF310311294WA SF311243520WA SF311243526WA SF311243540WA SF311243542WA SGF13045550WA SGF13045551WA SGF13045552WA SGF13045553WA SGF13045554WA SGF13045555WA SGF13045556WA SGF13045557WA SGF13045558WA SGF13045561WA SGF13045566WA SF310060416WA SGF12414084WA SF311072771WA SF311072772WA SF311072775WA SF311072778WA SF311072779WA SF311072780WA SF311072782WA SF311072755WA SF311072761WA SF311072762WA SGF12101957WA SGF12122236WA SGF12011176WA SGF12142293WA SGF12212614WA SGF12212638WA SGF12142305WA SGF12152422WA SGF12152423WA SGF12172578WA SGF12021261WA SG212100457WA SG212100460WA SGF12091839WA SGF12091870WA SGF12091880WA SGF12101971WA SGF12101973WA SGF12101995WA SGF12101999WA SGF12102002WA SGF12263197WA SGF12263215WA SGF12273237WA SGF12273239WA SG211490202WA SG211510234WA SG211510236WA SG211510258WA SGF11480440WA SGF11500575WA SGF11510631WA SGF11510640WA SGF11510643WA SGF11510644WA SGF11510651WA SGF11510653WA SGF11510654WA SGF11510655WA SGF11520891WA SGF11520892WA SGF11520900WA SGF11520901WA SGF11520902WA SGF11520903WA SGF11520977WA SGF11520978WA SGF11520979WA SGF11520980WA SGF12102048WA SGF12112071WA SGF12112080WA SGF12263196WA SF309500126WA SF311042659WA SF311042661WA SF311042665WA SG212491370WA SG212491372WA SG212491373WA SG212491374WA SF309410084WA SF309490115WA SF309500122WA SF309500123WA SF309500125WA SF309500127WA SF309500128WA SF309500132WA SF309500134WA SF310462156WA SF310462170WA SF311233443WA SF311233455WA SF311233462WA SF311233470WA SF311233472WA SF311253638WA SF311454569WA SF311494628WA SF311494655WA SF311494660WA SF311494677WA SF311494681WA SF311494682WA SF311494684WA SF311494687WA SF311494691WA SF311494692WA SF312205586WA SF312205588WA SF312205589WA SF312205590WA SF312205607WA SG212410971WA SG212410972WA SG212410973WA SG212410974WA SG212410975WA SG212410976WA SG212410977WA SG212410978WA SG212410979WA SG212410980WA SG212410981WA SG212410982WA SG212410983WA SG212410984WA SG212410985WA SG212410986WA SG212410987WA SG212410988WA SG212410989WA SG212410990WA SG212410991WA SG212410992WA SG212410995WA SG212410997WA SG212410998WA SG212411004WA SG212411005WA SG212411009WA SG212411010WA SG212411011WA SG212491366WA SGF12515159WA SGF12525192WA SF312095143WA SF312456444WA SF312456452WA SF312456465WA SF312456474WA SF312466501WA SF312466502WA SF312466504WA SF312466506WA SF312466508WA SF312466510WA SF312466514WA SF312466551WA SF312466561WA SF312466571WA SF312496610WA SF312496613WA SF312496614WA SF312496615WA SF312496617WA SF312496630WA SF312496671WA SF312496673WA SF312496676WA SF312496677WA SF312496678WA SF312496679WA SF312496680WA SF312496681WA SF312496682WA SF312496684WA SF312496685WA SF312506686WA SF312506687WA SF312506688WA SF312506689WA SF312506697WA SF312506701WA SF312506702WA SF312506703WA SF312376074WA SF312396158WA SF312396164WA SF312396165WA SF312396167WA SF312396168WA SF312335953WA SF312335954WA SF312335955WA SF312335956WA SF312335960WA SF312335963WA SF312355986WA SF312355987WA SF312355988WA SF312355990WA SF312355991WA SF312355992WA SF312355993WA SF312355994WA SF312355996WA SF312115254WA SF312396173WA SF312396174WA SF311253688WA SF311253694WA SF311334243WA SF312466530WA SF312466532WA SF312466534WA SF312396197WA SF312396198WA SF312416202WA SF312416203WA SF312416204WA SF312416205WA SF312416206WA SF312416207WA SF312416208WA SF312416209WA SF312416210WA SF312416211WA SF312416212WA SF311324202WA SF312396166WA SF311494629WA SF311494639WA SF311494627WA SF311494635WA SF311314045WA SF311334225WA SF311334270WA SF311334287WA SF311494624WA SF311494638WA SF311494646WA SF311494647WA SF311494648WA SF311494650WA SF312436282WA SF312436283WA SF312436284WA SF312436285WA SF312436287WA SF312436288WA SF312436291WA SF312436292WA SF312436293WA SF312436300WA SF312436306WA SF312436307WA SF312436333WA SF312446404WA SF312446411WA SF312456416WA SF312456417WA SF312456418WA SF312456419WA SF312456420WA SF312456421WA SF312456423WA SF312456433WA SF312456437WA SF312456438WA SF312456439WA SF312456440WA SF312456443WA SF312456448WA SF312456454WA SF312456455WA SF312456456WA SF312456457WA SF312456458WA SF312456459WA SF312456460WA SF312456461WA SF312456462WA SF312456464WA SF312456466WA SF312456469WA SF312456470WA SF312456471WA SF312456472WA SF312456473WA SF312456476WA SF312456478WA SF312456480WA SF312456481WA SF312456485WA SF312466487WA SF312466488WA SF312466491WA SF312466492WA SF312466496WA SF312466499WA SF312466503WA SF312466507WA SF312466509WA SF312466511WA SF312466516WA SF312466518WA SF312466519WA SF312466520WA SF312466523WA SF312446379WA SF312446380WA SF312446382WA SF312436315WA SF312436323WA SF312436326WA SF312436327WA SF312436328WA SF312436329WA SF312436332WA SF312436334WA SF312436335WA SF312436336WA SF312436337WA SF312436338WA SF312436339WA SF312436340WA SF312436341WA SF312436342WA SF312436343WA SF312436344WA SF312436345WA SF312436346WA SF312436347WA SF312436348WA SF312436350WA SF312436351WA SF312436352WA SF312436353WA SF312446358WA SF312446359WA SF312446360WA SF312446361WA SF312446362WA SF312446363WA SF312446364WA SF312446365WA SF312446366WA SF312446368WA SF312446369WA SF312446370WA SF312446371WA SF312446372WA SF312446373WA SF312446374WA SF312446375WA SF312446376WA SF312446377WA SF312446378WA SF312436274WA SF312436295WA SF312436296WA SF312436297WA SF312436298WA SF312436299WA SF312436301WA SF312436302WA SF312436303WA SF312436304WA SF312436305WA SF312436308WA SF312436309WA SF312436310WA SF312436311WA SF312436312WA SF312436313WA SF312436314WA SF312436316WA SF312436317WA SF312436318WA SF312436319WA SF312436321WA SF312436322WA SF312436324WA SF312436325WA SF312436330WA SF312436349WA SF312456477WA SF312466541WA SF312466548WA SF312466553WA SF312466556WA SF312466557WA SF312466558WA SF312466562WA SF312466564WA SF312466565WA SF312466566WA SF312466567WA SF312466568WA SF312466570WA SF312466574WA SF312466575WA SF312476586WA SF312436278WA SF312436280WA SF312436281WA SF312436286WA SF312436290WA SF312466494WA SF312496628WA SF312496631WA SF312496632WA SF312496633WA SF312496635WA SF312496639WA SF312496644WA SF312496670WA SF312456412WA SF312456413WA SF312456414WA SF312456415WA SF312476593WA SF312476594WA SF312476597WA SF312476598WA SF312496601WA SF312496606WA SF312496611WA SF312496636WA SF312496637WA SF312496638WA SF312496645WA SF312496668WA SF312496669WA SF312466486WA SGF12484544WA SGF12484545WA SGF12484546WA SGF12484547WA SGF12484549WA SGF12484550WA SGF12484551WA SGF12484552WA SGF12484553WA SGF12484554WA SGF12484555WA SGF12484564WA SF311283786WA SF311283807WA SF312295780WA SF312295782WA SF312295783WA SF312295784WA SF312295794WA SF312295796WA SF312396185WA SF312426262WA SF312426263WA SF312376067WA SF312376107WA SF312376125WA SF312396154WA SF312396155WA SF312396156WA SF312396157WA SF312396159WA SF312396160WA SF312396161WA SF312396162WA SF312396163WA SF312295787WA SF312295799WA SF312295801WA SF312295802WA SF312396170WA SF312396171WA SF312396177WA SF312416214WA SF312416220WA SF312416224WA SF312416227WA SF312416228WA SF312416249WA SF312416225WA SF312416240WA SF312416216WA SF312416221WA SF312416223WA SF312416231WA SF311344323WA SF311344339WA SF311404349WA SF311404357WA SF311404366WA SF311404379WA SF311414393WA SF311414429WA SF311414433WA SF313147076WA SF313147077WA SF313147073WA SF313147074WA SF313137058WA SF311324187WA SF311334244WA SF312155278WA SF312034933WA SF312034934WA SF312054956WA SF312054963WA SF312054993WA SF312054966WA SF312466545WA SF312466547WA SF312456463WA SF312456468WA SF312456479WA SF312466489WA SF312466550WA SF312466552WA SF312466555WA SF312466560WA SF312466578WA SF312476589WA SF312506720WA SF312506726WA SF312506727WA SF312506728WA SF312506730WA SF312526745WA SF312526749WA SF312526752WA SF312526756WA SF312526757WA SF313046875WA SF313046878WA SF313046863WA SF313046865WA SF313046874WA SF313046873WA SF313046877WA SF313046842WA SF313046858WA SF313046872WA SF313036827WA SF313046855WA SF313046860WA SF313046862WA SF313046868WA SF313046869WA SF313046871WA SF313046876WA SF313056895WA SF313066927WA SF313066930WA SF313066944WA SF313066947WA SGF13085883WA SGF13085888WA SGF13085891WA SGF13085886WA SGF13085887WA SGF13085889WA SGF13085890WA SF313106966WA SF312416213WA SF312416218WA SF312416229WA SF312436268WA SF312436270WA SF312436272WA SF312436273WA SF312436275WA SF312436276WA SF312335948WA SF312335949WA SF312335976WA SF312335980WA SF312335984WA SF312335985WA SF312315857WA SF311283768WA SF311283817WA SF312065000WA SF312095188WA SF312285760WA SF312335941WA SF312436277WA SF312436320WA SF312436331WA SF311494661WA SF312024857WA SF312024858WA SF312024869WA SF311324194WA SGF12363664WA SG212320843WA SGF12333542WA SG211380047WA SGF11390089WA SF311042650WA SF311042651WA SF311042653WA SF311042654WA SF311042656WA SF311042657WA SF311042662WA SF311042666WA SF311042667WA SF311042668WA SF311042670WA SF311042672WA SF311042675WA SF311042677WA SF311143088WA SF311193288WA SF311193322WA SF311193324WA SF311193327WA SF311193328WA SF311193330WA SF311193332WA SF311193333WA SF311193334WA SF311213352WA SF311213365WA SF311213368WA SF311213372WA SF311213374WA SF311213375WA SF311213376WA SF311213377WA SF311213378WA SF311213379WA SF311293861WA SF311303891WA SF311303892WA SF311324165WA SF311334291WA SF311494697WA SF311504708WA SF311504709WA SF311504710WA SF311504711WA SF311504712WA SF311504713WA SF311504734WA SF311504735WA SF311504737WA SF311504738WA SF311504739WA SF311504740WA SF311504741WA SF311504742WA SF311504743WA SF311504744WA SF311504745WA SF311504746WA SF311504747WA SF311504748WA SF311504749WA SF311504750WA SF311504751WA SF311504753WA SF312315864WA SF312315869WA SF312325883WA SF312325895WA SF312325897WA SF312325899WA SF312325900WA SF312325901WA SF312325902WA SF312325903WA SF312325904WA SF312325905WA SF312325906WA SF312325908WA SF312325910WA SF312325912WA SF312325913WA SF312325914WA SF312325915WA SF312325916WA SF312325917WA SF312325922WA SF312325923WA SF312325924WA SF312325925WA SF312335926WA SF312054984WA SF312065024WA SF312065027WA SF312085111WA SF312085128WA SF312125258WA SF312195505WA SF312245639WA SF312265670WA SF312265671WA SF312265672WA SF312265679WA SF312265681WA SF312265682WA SF312265683WA SF312265687WA SF312335957WA SF312335973WA SF312355997WA SF312355999WA SF312356000WA SF312356001WA SF312356002WA SF312356003WA SF312356005WA SF312356008WA SF312356011WA SGF12424162WA SGF12424163WA SGF12424165WA SGF12424166WA SGF12424167WA SGF12424169WA SGF12424171WA SGF12424173WA SGF12424174WA SG212431109WA SG212481254WA SG212481255WA SGF12464330WA SGF12474337WA SGF12474375WA SGF12474376WA SGF12474379WA SGF12424168WA SGF12474399WA SGF12474400WA SGF12474401WA SGF12474402WA SGF12474403WA SGF12474404WA SGF12474405WA SGF12474407WA SGF12474408WA SG212481184WA SG212481185WA SG212481186WA SG212481188WA SG212481189WA SG212481194WA SG212481195WA SG212481198WA SG212481199WA SG212481203WA SGF12484524WA SGF12484525WA SGF12484556WA SGF12484558WA SGF12484559WA SGF12484677WA SGF12484678WA SGF12484679WA SGF12484680WA SGF12484681WA SGF12484683WA SGF12484684WA SGF12484685WA SGF12484686WA SGF12494692WA SGF12494693WA SF312456467WA SF312456475WA SF312466495WA SF312466513WA SG212421052WA SG212481257WA SG212481271WA SG212481272WA SG212481249WA SG212481250WA SG212481252WA SG212481253WA SG212481265WA SG212481270WA SG212501443WA SG212501444WA SG212501446WA SG212501472WA SG212501474WA SG212501482WA SG212501488WA SG212501489WA SG212511513WA SG212511529WA SGF12494763WA SGF12494764WA SGF12494765WA SGF12494767WA SGF12494768WA SGF12494782WA SGF12494784WA SGF12494785WA SGF12494786WA SGF12494788WA SGF12504985WA SGF12504987WA SGF12505024WA SGF12505026WA SGF12505029WA SGF12505031WA SGF12505032WA SGF12505033WA SGF12505034WA SGF12505037WA SF312506698WA SF312506729WA SF312516734WA SG213021602WA SG213021605WA SG213021607WA SG213021608WA SG213021609WA SG213021610WA SG213021611WA SG213021613WA SGF13045485WA SGF13055619WA SGF13055623WA SGF13055625WA SGF13055626WA SGF13055627WA SGF13055628WA SGF13055643WA SGF13055645WA SGF13055646WA SGF13055647WA SGF13045565WA SGF13055616WA SGF13055617WA SGF13055618WA SGF13055621WA SGF13055644WA SGF13055691WA SGF13055694WA SGF13095948WA SGF13095962WA SGF13095964WA SGF13095965WA SGF13095967WA SGF13095970WA SGF13095971WA SGF13045567WA SGF13055622WA SGF13055624WA SGF13095947WA SGF13095949WA SGF13095968WA SGF13095969WA SGF13095972WA SGF13095973WA SGF13095974WA SGF13095975WA SGF13055693WA SF313066946WA SF313066948WA SF313066949WA SF313106963WA SF313106967WA SGF13106115WA SGF13106117WA SGF13106114WA SF313106961WA SGF13106098WA SGF13106099WA SGF13106101WA SGF13106102WA SGF13106103WA SGF13106104WA SGF13106105WA SF313106970WA SF313106977WA SF313106978WA SG213121979WA SG213131985WA SG213132029WA SG213132030WA SG213132031WA SGF13176544WA SGF13176545WA SGF13176546WA SGF13176547WA SGF13176548WA SGF13146516WA SGF13176549WA SGF13176550WA SGF13176551WA SGF13176552WA SGF13176553WA SGF13176554WA SGF13176555WA SGF13176557WA SGF13176558WA SF313147086WA SF312175457WA SF312175442WA SF312195465WA SF312195464WA SF312195458WA SF312175454WA SF312175444WA SF312175456WA SF312195460WA SF312195459WA SF312195461WA SF312195466WA SGF12484521WA SGF12484522WA SGF12484565WA SGF12484566WA SGF12484568WA SGF12484569WA SGF12484571WA SGF12484575WA SGF13045541WA SGF13045543WA SGF13045544WA SGF13045548WA SGF13045570WA SGF13045571WA SGF13045572WA SGF13045574WA SGF13045576WA SGF13045579WA SF312095150WA SF312095153WA SF312095157WA SF312095159WA SF312095160WA SF312095161WA SF312095163WA SF312095165WA SF312095169WA SF312095173WA SF312295803WA SF312295804WA SF312295805WA SF312315810WA SF312315811WA SG213172097WA SG213172105WA SG213172109WA SG213172110WA SG213172111WA SG213172112WA SG213172113WA SG213172114WA SG213172115WA SG213172116WA SG213172117WA SG213172118WA SG213172119WA SG213172120WA SG213172121WA SF309530217WA SF310200787WA SF310190691WA SF310190693WA SF310190694WA SF310190723WA SF310190725WA SF310200791WA SF310241060WA SF310241061WA SF310241063WA SF310241065WA SF310241066WA SF310301168WA SF310311299WA SF310311301WA SF310311303WA SF310321304WA SF310321305WA SF310321306WA SF310321307WA SF310321308WA SF310321309WA SF310321311WA SF310321312WA SF310321313WA SF310321314WA SF310321315WA SF310321316WA SF310321317WA SF310321318WA SF310321319WA SF310321321WA SF310321322WA SF310361524WA SF310361554WA SF310371561WA SF310371563WA SF310371564WA SF310371566WA SF310371567WA SF310371568WA SF310371569WA SF310371571WA SF310371572WA SF310371573WA SF310371574WA SF310371575WA SF310371576WA SF310371577WA SF310371578WA SF310371580WA SF310371581WA SF310371583WA SF310361475WA SF310381651WA SF310381658WA SF310381681WA SF310411764WA SF310411767WA SF310411768WA SF310411769WA SF310411771WA SF310401725WA SF310411772WA SF310411781WA SF310421788WA SF310421791WA SF310421792WA SF310421794WA SF310421801WA SF310421802WA SF310421803WA SF310421804WA SF310421805WA SF310421806WA SF310421807WA SF310421808WA SF310421809WA SF310421810WA SF310421811WA SF310421812WA SF310421813WA SF310401722WA SF310421860WA SF310431875WA SF310431877WA SF310431879WA SF310431882WA SF310431883WA SF310431891WA SF310431900WA SF310431902WA SF310441981WA SF310442002WA SF310442004WA SF310442006WA SF310442007WA SF310442010WA SF310431971WA SF310452020WA SF310452035WA SF310452036WA SF310452037WA SF310452038WA SF310452039WA SF310452044WA SF310452053WA SF310462122WA SF310462193WA SF310462124WA SF310462125WA SF310462127WA SF310462129WA SF310462130WA SF310462123WA SF310482275WA SF310482277WA SF310482287WA SF310482288WA SF310482290WA SF310482293WA SF310482296WA SF310482299WA SF310482301WA SF310482302WA SF310482314WA SF311112963WA SF311122997WA SF311123001WA SF311123011WA SF311123021WA SF313036824WA SG213101936WA SG213101943WA SG213101958WA SGF11490502WA SGF11510658WA SGF11510743WA SGF11510747WA SGF11510749WA SGF11510750WA SGF11510756WA SGF11510757WA SGF11510758WA SGF11510762WA SGF12011174WA SGF12011184WA SGF12011186WA SGF12021239WA SGF12021241WA SGF12021251WA SGF12081666WA SGF12081719WA SGF12081721WA SGF12081722WA SGF12081755WA SGF12081756WA SGF12081764WA SGF12081775WA SGF12091810WA SF309470112WA SF309490120WA SF309500124WA SF309500129WA SF309500130WA SF309500131WA SF309500133WA SF310190702WA SF310190707WA SF310190710WA SF310190711WA SF310190714WA SF311153179WA SF312024885WA SF312205561WA SF312205572WA SF312205613WA SGF12142296WA SGF12172532WA SG212020378WA SGF12021225WA SG212100450WA SG212100458WA SG212100459WA SGF12091905WA SGF12112063WA SGF12112070WA SGF12373776WA SGF12091873WA SGF12091908WA SGF12122170WA SGF12091888WA SGF12091889WA SGF12091928WA SF311334259WA SF310190718WA SF311243544WA SF311243548WA SF311243551WA SF310361533WA SF310361535WA SF310190709WA SF310190712WA SF310190717WA SF310361536WA SF310361538WA SF310100532WA SF310100545WA SF311303960WA SF311324185WA SF311334229WA SF312125259WA SGF12434313WA SG212481256WA SG212481258WA SGF12474480WA SG212481242WA SGF12484619WA SGF12484623WA SF313046849WA SF313046854WA SGF13035377WA SGF13045549WA SGF13045569WA SGF13045573WA SGF13045575WA SGF13045577WA SGF13045581WA SGF13045582WA SG213051735WA SG213051736WA SG213051737WA SG213051738WA SGF13045592WA SGF13055719WA SGF13055721WA SGF13055722WA SGF13055690WA SGF13116194WA SGF13116204WA SGF13116221WA SGF13116224WA SGF13116225WA SGF13116226WA SGF13116228WA SGF13116230WA SF311112885WA SF311112893WA SF311112900WA SF311112905WA SF311112907WA SF311112909WA SF311112910WA SF311112911WA SF311112912WA SF311112913WA SF311112916WA SF311112919WA SGF12484658WA SGF12494700WA SGF12011100WA SGF12162472WA SGF12162473WA SGF12162474WA SGF12373767WA SGF12293337WA SGF12293338WA SGF12333540WA SGF12333543WA SGF12373822WA SGF12373823WA SGF13096010WA SF310220808WA SGF11340044WA SGF12484557WA SG213101909WA SGF12051457WA SGF12242844WA SGF12242840WA SGF12051442WA SGF12373764WA SGF11510716WA SGF11510837WA SGF11510849WA SGF11510863WA SGF11510865WA SGF11510867WA SGF12142316WA SGF12142319WA SGF12142327WA SGF12152347WA SGF12152348WA SGF12152349WA SF310361527WA SGF12051393WA SGF12051407WA SGF12051420WA SGF12051435WA SGF12051450WA SGF12051451WA SGF12172511WA SGF12172519WA SF310361522WA SF310361525WA SGF12112129WA SGF12112141WA SGF12112143WA SGF12081728WA SGF12081735WA SGF12081737WA SGF12051399WA SGF12051402WA SGF11390106WA SGF11390107WA SGF11390109WA SGF11390115WA SGF11390116WA SGF11380080WA SGF11390112WA SG212020381WA SG212050383WA SG212050384WA SG212050386WA SG212050392WA SG212050395WA SG212020380WA SGF12263210WA SGF12263188WA SGF12293331WA SGF12283298WA SGF12283307WA SGF12051444WA SGF12051468WA SGF12242828WA SGF12242847WA SF310381655WA SF310381656WA SF310381660WA SG212050396WA SG212050404WA SG212050405WA SG212050407WA SG212050408WA SG212050409WA SG212050410WA SG212050414WA SGF12373757WA SGF12373810WA SGF12373811WA SGF12363579WA SGF12363581WA SGF12252934WA SGF12252936WA SGF12252946WA SGF12252952WA SGF12252953WA SGF12252954WA SGF12252955WA SGF12252956WA SGF12252958WA SGF12252962WA SGF12363583WA SGF12363584WA SGF12363587WA SGF12373771WA SGF12373773WA SGF12373783WA SG212360894WA SG212360883WA SGF12373797WA SGF12373808WA SGF12273267WA SGF12273269WA SGF12283305WA SGF11520935WA SGF11520956WA SGF11390113WA SGF12081723WA SG212491361WA SF311524836WA SF311524838WA SF311524839WA SF311524840WA SF311524842WA SGF13055695WA SGF13095987WA SGF13095983WA SGF13106043WA SGF13106023WA SG213101932WA SG213101934WA SGF13106109WA SGF13106112WA SGF13106119WA SGF13106121WA SGF13106124WA SGF13106126WA SGF13106106WA SGF13106107WA SGF13106110WA SGF13106111WA SGF13106122WA SGF13106125WA SGF13106127WA SF313117016WA SGF13176604WA SF313147095WA SF313147106WA SF313177112WA SF313177114WA SF313177117WA SJF13170142WA SF309420099WA SF309420101WA SF310220809WA SF310220811WA SF310220812WA SF310220815WA SF310220817WA SF310220802WA SF310220804WA SF310200782WA SF310210799WA SF310210800WA SF310220806WA SF310220816WA SF310220818WA SF310220819WA SF310220821WA SF310220814WA SF310220823WA SF310220828WA SF310220801WA SF310220803WA SF310241003WA SF310241006WA SF310361526WA SF310361531WA SF310381654WA SF310431868WA SF310431869WA SF310431870WA SF310431871WA SF310431872WA SF310431873WA SF310431874WA SF310431876WA SF310431878WA SF310431881WA SF310431973WA SF310452021WA SF310452056WA SF310452062WA SF310452063WA SF310452066WA SF310452071WA SF310431965WA SF310452105WA SF310462119WA SF310472218WA SF310482306WA SF310512440WA SF310512443WA SF310512444WA SF310512446WA SF310512448WA SF311112970WA SF311112971WA SF311112973WA SF311122979WA SF311112972WA SF311112974WA SF311112975WA SF311153158WA SF311153164WA SF311153174WA SF311153186WA SF311153177WA SF311153173WA SF311153171WA SF311153178WA SF311153182WA SF311153183WA SF311153187WA SF311153189WA SF311193291WA SF311243590WA SF311253607WA SF311253609WA SF311253612WA SF311253615WA SF311253618WA SF311253623WA SF311253624WA SF311253625WA SF311253626WA SF311253627WA SF311253628WA SF311253629WA SF311253635WA SF311253641WA SF311253693WA SF311263706WA SF311263707WA SF311263708WA SF311263719WA SF311263717WA SF311283780WA SF311314054WA SF311314105WA SF311314121WA SF311303939WA SF311314015WA SF311314037WA SF311314059WA SF311314068WA SF311314070WA SF311314073WA SF311314069WA SF311303943WA SF311314043WA SF311314103WA SF311334260WA SF311344336WA SF311344340WA SF311334218WA SF311334235WA SF311494622WA SF311494637WA SF311494636WA SF311484612WA SF312034908WA SF312034914WA SF312034924WA SF312034909WA SF312034923WA SF312054979WA SF312054995WA SF312065005WA SF312054958WA SF312366039WA SF312366040WA SF312376059WA SGF12414079WA SGF12414081WA SGF12414080WA SGF12424170WA SGF12474336WA SGF12474353WA SG212481246WA SG212481243WA SG212481244WA SGF12484604WA SG212491337WA SGF12484649WA SGF12484665WA SGF12484669WA SGF12484672WA SGF12484673WA SGF12484674WA SGF12484675WA SGF12484676WA SG212491378WA SG212491384WA SG212491387WA SG212491388WA SG212491389WA SGF12494805WA SGF12494720WA SGF12494726WA SGF12494728WA SGF12494731WA SGF12494741WA SGF12494723WA SGF12494730WA SGF12494751WA SGF12494752WA SGF12494739WA SGF12494742WA SGF12494746WA SGF12494744WA SG212491375WA SG212491382WA SG212491385WA SG212491401WA SF311524856WA SG212491422WA SGF12505014WA SGF13035388WA SGF13055653WA SGF13055658WA SGF13055660WA SGF13055661WA SGF13055662WA SGF13055663WA SGF13055664WA SGF13055665WA SGF13055666WA SGF13055667WA SGF13055669WA SGF13055688WA SGF13055689WA SGF13085806WA SGF13085808WA SF310240950WA SF310240953WA SF310240954WA SF310240955WA SF310240956WA SF310240957WA SF310241027WA SF310241028WA SF310241029WA SF310241030WA SF310241031WA SF310241034WA SF311072715WA SF311072788WA SF311112896WA SF311112898WA SF311112958WA SF311253617WA SF311243597WA SF311243599WA SF311334271WA SF311414385WA SF311414440WA SF311434540WA SF311484595WA SF311484600WA SF311484610WA SF311484613WA SF311484598WA SF311484599WA SF311494641WA SGF12494812WA SGF12494815WA SGF12494817WA SGF12494819WA SGF12494824WA SGF13035428WA SGF13035429WA SF312065009WA SF310462179WA SF310462181WA SF310462182WA SF310462183WA SF310462184WA SF310462185WA SF310462186WA SF310462190WA SF312315820WA SF312315836WA SF312315837WA SG212150561WA SG212010289WA SG211520262WA SGF11450300WA SGF11450301WA SG212050394WA SG211430121WA SGF12162451WA SG212050413WA SG212070423WA SG212070427WA SG212070428WA SG212070429WA SG212070431WA SG211510230WA SG211510235WA SG212010295WA SGF12112118WA SGF12132257WA SGF12132262WA SGF12142329WA SGF12152362WA SGF12152368WA SGF12152382WA SGF12172524WA SGF12172586WA SGF11521090WA SGF12011098WA SGF12011109WA SGF12011115WA SG212260705WA SG212260707WA SG212270740WA SG212270743WA SG212270744WA SG212280756WA SGF12011104WA SGF12102032WA SGF12242839WA SGF12242848WA SG212150550WA SG212150567WA SGF11450315WA SGF12011102WA SGF12172518WA SG212130512WA SG212290781WA SG212290777WA SGF12011107WA SGF12011111WA SGF12011093WA SGF12011121WA SGF12011132WA SGF12112124WA SGF12122207WA SF310371570WA SG212050388WA SG212050389WA SG212050391WA SG212050393WA SG212150590WA SG211450156WA SGF11470408WA SGF11470409WA SF310190722WA SF310431896WA SF310431884WA SF310431889WA SF310431922WA SF310431934WA SF310431929WA SF310441992WA SF310441998WA SF310442008WA SF310431945WA SF310431964WA SF310452018WA SF310452026WA SF310452029WA SF310462120WA SF310452090WA SF310462126WA SF310462146WA SF310482295WA SF310482297WA SF310482298WA SF310482278WA SF310482281WA SF310482282WA SF310482283WA SF310512408WA SF310482333WA SF310492338WA SF310492347WA SF310492348WA SF310502395WA SF311042664WA SF311133049WA SF311133057WA SF311133059WA SF311133060WA SF311133061WA SF311153152WA SF311153154WA SF311153157WA SF311253631WA SF311253622WA SF311253630WA SF311253632WA SF311253699WA SF311263703WA SF311263704WA SF311263722WA SF311263724WA SF311263728WA SF311253696WA SF311253697WA SF311293882WA SF311303890WA SF311303953WA SF311293830WA SF311293864WA SF311293869WA SF311303893WA SF311303965WA SF311314025WA SF311334203WA SF311334268WA SF311334269WA SF311334285WA SF311334286WA SF311344326WA SF311334281WA SF311334294WA SF311344316WA SF311344327WA SF312034912WA SF312034931WA SF312054944WA SF312054946WA SF312155287WA SF312155315WA SF312165372WA SF312165399WA SF312195498WA SF312205532WA SF312205539WA SF312205591WA SF312205608WA SF312265673WA SF312265674WA SF312265675WA SF312265677WA SF312265678WA SF312265686WA SF312265688WA SF312265689WA SF312265690WA SF312265691WA SF312265692WA SF312265693WA SF312265696WA SF312265698WA SF312335967WA SF312355989WA SF312496634WA SF312506707WA SF312506708WA SF312506693WA SF312506695WA SF312506696WA SF312506704WA SF312506711WA SF312506710WA SF312506690WA SF312506691WA SF312506694WA SF312496675WA SF312506700WA SF312506713WA SF312476584WA SJF13130121WA SF313016794WA SF313026803WA SF313026805WA SF313026806WA SF313026809WA SF313026810WA SF313056886WA SF313066934WA SF313066940WA SF313066941WA SF313106984WA SF313106985WA SF313106986WA SF313106987WA SF313106988WA SF313106979WA SF313106980WA SF313106981WA SF313106982WA SF313106983WA SF313106971WA SF313106972WA SF313117013WA SF313117015WA SF313117017WA SF313117018WA SF313117019WA SF313117022WA SF313117041WA SGF13136504WA SGF13136507WA SGF13136508WA SGF13136509WA SGF13136510WA SF313147096WA SF313147098WA SGF13176706WA SGF13176707WA SGF13176708WA SF313147108WA SG212491280WA SG212491334WA SF311062708WA SF311072710WA SF311072750WA SF311072759WA SF311072801WA SF311324201WA SF311263721WA SGF12051473WA SG212070420WA SG212070424WA SG212070425WA SG211400078WA SG211490201WA SGF12172536WA SGF12373812WA SGF12373816WA SGF12373815WA SGF12373817WA SGF12373820WA SF310361532WA SF310110557WA SF310110561WA SF310110586WA SF310110589WA SF310200762WA SF310241040WA SF310241044WA SF310241041WA SF310241037WA SF310241039WA SF310271115WA SF310271117WA SF310311224WA SF310311231WA SF310431960WA SF310441990WA SF311042639WA SF311042631WA SF311052682WA SGF13116173WA SGF13116174WA SG213031651WA SG212471134WA SGF12051421WA SGF12061527WA SGF12061531WA SGF12101958WA SGF12081709WA SGF12081711WA SGF12081718WA SGF12091830WA SGF12091836WA SGF12101962WA SGF12232722WA SGF12242858WA SGF12242859WA SGF12242861WA SGF12242864WA SGF12283288WA SGF12313403WA SGF12313407WA SGF12323468WA SGF12323469WA SGF12323470WA SGF12323471WA SGF12323472WA SGF12323473WA SGF12333544WA SGF12333545WA SGF12333546WA SF309530218WA SG213142094WA SGF13176664WA SGF13176667WA SF311123016WA SF311123017WA SF311123018WA SF311123019WA SF311123020WA SF311123025WA SGF12484560WA SGF12283306WA SGF12122238WA SGF12122239WA SGF12162457WA SGF12162459WA SGF12162461WA SGF12162464WA SGF12162458WA SGF12162460WA SGF12122235WA SGF12122237WA SGF12162465WA SGF12162470WA SGF12313406WA SGF12333547WA SGF12333548WA SGF12333549WA SGF12363600WA SGF12363602WA SGF12363603WA SGF12363604WA SGF12363605WA SF311283791WA SG212120485WA SG212120486WA SG212120488WA SG212120489WA SG212120491WA SG212120496WA SG212120498WA SG212120502WA SG212120505WA SG212120506WA SG212130523WA SG212130524WA SG212130525WA SG212130526WA SG212130527WA SG212130530WA SG212130532WA SG212130533WA SG212130534WA SGF12122157WA SGF12122164WA SGF12122165WA SGF12122167WA SGF12122168WA SGF12122169WA SGF12122171WA SGF12122176WA SGF12122178WA SGF12122179WA SGF12122208WA SGF12122210WA SGF12122211WA SGF12122212WA SGF12122214WA SGF12122215WA SGF12122216WA SGF12122220WA SGF12122221WA SG212120494WA SGF12283292WA SGF12122232WA SGF12122233WA SG212310804WA SG212310813WA SGF12152359WA SGF12152389WA SGF12172581WA SGF12172612WA SGF12232727WA SGF12232728WA SGF12232736WA SGF12232737WA SGF12232738WA SGF12232742WA SGF12232749WA SGF12232751WA SGF12232755WA SG211520270WA SGF12051408WA SGF12011131WA SGF12021228WA SGF12021253WA SGF12021256WA SGF12021259WA SGF12021264WA SGF12031267WA SGF12031268WA SGF12031270WA SG211490205WA SG212020375WA SGF12021238WA SGF12051466WA SGF12051447WA SG212120501WA SG212130535WA SG212130539WA SG212130542WA SGF12112134WA SGF12142287WA SGF12142292WA SGF12172483WA SGF12172479WA SGF12172488WA SGF12172537WA SGF12132244WA SGF12172516WA SGF12263173WA SG212260731WA SG212270741WA SG212280757WA SGF12283302WA SGF12283304WA SGF12283310WA SG212290780WA SGF12333508WA SGF12293326WA SGF12293328WA SGF12242827WA SGF12142306WA SGF12142308WA SG212150580WA SG212230658WA SG212120499WA SG212130529WA SGF12293316WA SGF12293323WA SGF12293324WA SGF12293325WA SGF12293330WA SGF12293313WA SGF12333513WA SGF12142268WA SGF12142269WA SGF11490475WA SGF11510632WA SGF11510662WA SGF11510665WA SGF11510666WA SGF11510667WA SGF11510671WA SGF11510678WA SGF11510679WA SGF11510680WA SGF11510681WA SGF11510682WA SGF12031294WA SGF12031338WA SGF12051349WA SGF12051371WA SGF12051405WA SGF12051409WA SGF12051410WA SGF12051419WA SGF12081732WA SGF12081733WA SGF12081739WA SGF12081740WA SGF12081741WA SGF12081742WA SGF12081743WA SGF12081745WA SGF12081748WA SGF12091938WA SGF12232745WA SG212150576WA SG212160601WA SG212230655WA SG212230656WA SGF12112119WA SGF12132265WA SGF12142291WA SGF12142325WA SGF12142336WA SGF12152383WA SGF12152428WA SGF12152430WA SGF12172499WA SGF12172523WA SG212290779WA SG212360861WA SG212360862WA SG212360863WA SG212360864WA SG212360865WA SG212360867WA SG212360868WA SG212360869WA SG212360871WA SG212360872WA SG212360873WA SG212360874WA SG212360875WA SG212360879WA SG212360880WA SG212360881WA SGF12142322WA SGF12152339WA SGF12152342WA SGF12152343WA SGF12152345WA SGF12152346WA SGF12152352WA SGF12152353WA SGF12152358WA SGF12152372WA SGF12363633WA SGF12363661WA SG212160600WA SG212230657WA SGF12232748WA SGF12273264WA SG211510237WA SG211510242WA SG211510243WA SG211510244WA SG211510245WA SG211510247WA SG211510248WA SG211510249WA SG211510250WA SG211510251WA SG211510252WA SG211510253WA SG211510254WA SG211510255WA SG211510256WA SG211510257WA SG212100438WA SG212100451WA SG212110462WA SG212110463WA SG212110467WA SG212110469WA SG212110470WA SG212110475WA SG212110476WA SG212120479WA SG212120483WA SG212120500WA SG212130510WA SG212130513WA SG212130514WA SG212130515WA SG212130516WA SG212130517WA SG212130518WA SG212130519WA SG212130521WA SG212130522WA SG212130531WA SG212130536WA SG212130537WA SG212130538WA SG212130540WA SG212130541WA SG212130543WA SG212130544WA SG212130545WA SG212130546WA SGF12102004WA SGF12102029WA SGF12102036WA SGF12112076WA SGF12112120WA SGF12112122WA SGF12112126WA SGF12112130WA SGF12112131WA SGF12112133WA SGF12112139WA SGF12112142WA SGF12112144WA SGF12122147WA SGF12122156WA SGF12122158WA SGF12122159WA SGF12122172WA SGF12122173WA SGF12122174WA SGF12122175WA SGF12122177WA SGF12122205WA SGF12122217WA SGF12122218WA SGF12122230WA SGF12122231WA SGF12122234WA SGF12132246WA SGF12132247WA SGF12132248WA SGF12132253WA SGF12132254WA SGF12081705WA SGF12081738WA SGF12132243WA SGF12142332WA SGF12152350WA SG212260708WA SG212260714WA SGF12172541WA SGF12363667WA SGF12152355WA SGF12152374WA SGF12152388WA SGF12152390WA SGF12152392WA SGF12152396WA SGF12152400WA SGF12172482WA SGF12172484WA SGF12172487WA SGF12172489WA SGF12172490WA SGF12172492WA SGF12172500WA SGF12132245WA SGF12152371WA SGF12152397WA SGF12152399WA SGF12162439WA SGF12162442WA SGF12162444WA SGF12162453WA SGF12172478WA SGF12172481WA SGF12172493WA SGF12051445WA SGF12051449WA SGF12051467WA SGF12051388WA SGF12061481WA SGF12061482WA SGF12061507WA SGF12061511WA SGF12061513WA SGF12061514WA SGF12061516WA SGF12061517WA SGF12061519WA SGF12061520WA SGF12061522WA SGF12061523WA SGF12061524WA SGF12061525WA SGF12061526WA SGF12061535WA SGF12061536WA SGF12061540WA SGF12061541WA SGF12061542WA SGF12061504WA SGF12061512WA SGF12061515WA SGF12061518WA SGF12373778WA SGF12373801WA SGF12373802WA SGF12373806WA SGF12373807WA SGF12263171WA SGF12263172WA SGF12263174WA SGF12263177WA SGF12263178WA SGF12263179WA SGF12263180WA SGF12263181WA SGF12263182WA SGF12263187WA SGF12263189WA SGF12263190WA SGF12263191WA SF310381679WA SF310361537WA SF310381657WA SF310381674WA SF310381675WA SF310381678WA SF310381680WA SF310381682WA SF310381683WA SF310381684WA SF310381685WA SF310381686WA SF310381687WA SF310381688WA SF310381689WA SF310381690WA SF310381691WA SF310381692WA SF310381693WA SF310371565WA SF310381694WA SGF12283308WA SGF12293311WA SGF12293312WA SGF12293314WA SGF12293315WA SGF12293317WA SGF12293319WA SGF12293320WA SGF12293321WA SGF12293322WA SGF12283285WA SGF12283286WA SGF12283287WA SGF11340043WA SGF12021213WA SGF12021219WA sf310060393wa sf310060364wa sf310060385wa sf310060414wa sf310060417wa sf310060418wa SGF13126352WA SGF13126353WA SGF13126354WA SGF13126356WA SGF13126347WA SGF13136488WA SGF13136490WA SGF13136493WA SGF13136494WA SGF13136495WA SGF13136498WA SGF13136499WA SGF13136500WA SGF13136501WA SGF13136502WA SGF13136503WA SGF13136485WA SGF13136486WA SGF13136487WA SGF13126423WA SGF13126424WA SGF13126425WA SGF13126426WA SGF13176694WA SGF13176696WA SGF13176698WA SGF13136489WA SGF13176561WA SGF13176564WA SGF13176565WA SGF13176566WA SG213131988WA SG213131989WA SG213131990WA SG213131991WA SG213172098WA SG213172099WA SG213172100WA SG213172101WA SG213172102WA SG213172103WA SG213172104WA SG213172106WA SG213172107WA SG213172108WA SGF13176587WA SGF13176608WA SGF13176609WA SGF13176610WA SGF13176611WA SGF13176614WA SGF13176615WA SGF13176616WA SGF13176617WA SGF13176618WA SGF13176619WA SGF13176621WA SGF13176622WA SGF13176623WA SGF13176624WA SGF13176625WA SGF13176626WA SGF13176627WA SGF13176628WA SGF13176629WA SGF13176630WA SGF13176631WA SGF13176632WA SGF13176633WA SGF13176634WA SGF13176635WA SGF13176636WA SGF13176637WA SGF13176638WA SGF13176639WA SGF13176640WA SGF13176641WA SGF13176642WA SGF13176643WA SGF13176645WA SGF13176646WA SGF13176647WA SGF13176648WA SGF13176649WA SGF13176650WA SGF13176651WA SGF13176652WA SGF13176653WA SGF13176654WA SGF13176655WA SGF13176656WA SGF13176657WA SGF13176658WA SGF13176659WA SGF13176660WA SGF13176661WA SGF13176663WA SGF13176567WA SGF13176569WA SGF13136442WA SGF13136483WA SGF13136496WA SGF13136497WA SGF13136515WA SF309420100WA SF309420102WA SF310060397WA SF310180684WA SF310180685WA SF310180686WA SF310180687WA SF310180688WA SF310180689WA SF310200754WA SF310220825WA SF310220829WA SF310220830WA SF310220831WA SF310220834WA SF310220835WA SF310220836WA SF310220837WA SF310220838WA SF310220841WA SF310220842WA SF310220844WA SF310220846WA SF310240975WA SF310240976WA SF310240977WA SF310240979WA SF310240980WA SF310240981WA SF310240982WA SF310240983WA SF310240989WA SF310240991WA SF310240992WA SF310241017WA SF310241038WA SF310241051WA SF310241055WA SF310241058WA SF310241059WA SF310241062WA SF310271110WA SF310271111WA SF310271113WA SF310271125WA SF310271126WA SF310271129WA SF310271130WA SF310271132WA SF310271121WA SF310271127WA SF310271112WA SF310271120WA SF310271123WA SF310421843WA SF310421838WA SF310421839WA SF310421840WA SF310421841WA SF310421842WA SF310421844WA SF310421845WA SF310421847WA SF310421848WA SF310421849WA SF310421850WA SF310421852WA SF310421853WA SF310421855WA SF310421856WA SF310421857WA SF310421858WA SF310421859WA SF310421861WA SF310401715WA SF310411775WA SF310411779WA SF310411780WA SF310411782WA SF310411783WA SF310411784WA SF310411785WA SF310411786WA SF310411787WA SF310421817WA SF310421821WA SF310421822WA SF310421823WA SF310421824WA SF310421825WA SF310421826WA SF310421827WA SF310421828WA SF310421829WA SF310421830WA SF310421831WA SF310421832WA SF310421833WA SF310421834WA SF310421835WA SF310421837WA SF310421846WA SF310431903WA SF310431904WA SF310431906WA SF310431913WA SF310492346WA SF310512461WA SF310512462WA SF310512463WA SF310512464WA SF310512465WA SF310512466WA SF310512467WA SF310512468WA SF310512470WA SF310512472WA SF310512476WA SF310512477WA SF310512478WA SF310512479WA SF310512480WA SF310512481WA SF310512482WA SF310512483WA SF310512484WA SF310512485WA SF310512486WA SF310512487WA SF310512488WA SF310512489WA SF310512490WA SF310512491WA SF310512492WA SF310512493WA SF310512494WA SF310512495WA SF310512496WA SF310512497WA SF310512498WA SF310512499WA SF310512500WA SF310512501WA SF310512502WA SF310512503WA SF310512504WA SF310512505WA SF310512506WA SF310512507WA SF310512508WA SF310512509WA SF310512510WA SF310512511WA SF310512512WA SF310512513WA SF310512514WA SF310512515WA SF310512516WA SF310512517WA SF310512518WA SF310512519WA SF310512520WA SF310512521WA SF310512522WA SF310512523WA SF310512524WA SF310512525WA SF310512526WA SF310522527WA SF310522528WA SF310522529WA SF310522530WA SF310522531WA SF310522532WA SF310522533WA SF310522534WA SF310522535WA SF310522536WA SF310522537WA SF310522538WA SF310522539WA SF310522540WA SF310522541WA SF310522542WA SF310522543WA SF310522544WA SF310522545WA SF310522546WA SF310522547WA SF310522548WA SF310522549WA SF310522557WA SF310522558WA SF310522559WA SF310522561WA SF310522562WA SF310522565WA SF310522566WA SF310522567WA SF310522568WA SF310522569WA SF310522570WA SF310522572WA SF310522573WA SF310522574WA SF310522575WA SF310522576WA SF310522550WA SF310522551WA SF310522552WA SF310522553WA SF310522554WA SF310522555WA SF310522556WA SF310522563WA SF310522564WA SF310522571WA SF311022578WA SF311022580WA SF311022582WA SF311022585WA SF311022588WA SF311022593WA SF311112978WA SF311122981WA SF311122982WA SF311122983WA SF311122984WA SF311122985WA SF311122986WA SF311122987WA SF311122988WA SF311122989WA SF311122990WA SF311122991WA SF311122992WA SF311122993WA SF311122994WA SF311122995WA SF311122996WA SF311122998WA SF311122999WA SF311123000WA SF311123002WA SF311123003WA SF311123005WA SF311123006WA SF311123007WA SF311123009WA SF311123010WA SF311123013WA SF311133056WA SF311133058WA SF311133062WA SF311133063WA SF311133064WA SF311133065WA SF311133066WA SF311133067WA SF311133069WA SF311133070WA SF311143073WA SF311143089WA SF311143090WA SF311133028WA SF311153180WA SF311153184WA SF311153185WA SF311153190WA SF311153192WA SF311153193WA SF311153194WA SF311153195WA SF311173196WA SF311173197WA SF311173198WA SF311173199WA SF311173200WA SF311173201WA SF311173202WA SF311173203WA SF311173204WA SF311173205WA SF311173206WA SF311173207WA SF311173208WA SF311173209WA SF311173211WA SF311173212WA SF311173213WA SF311173214WA SF311173215WA SF311173216WA SF311173217WA SF311173218WA SF311173219WA SF311173220WA SF311173221WA SF311173222WA SF311173223WA SF311173224WA SF311173225WA SF311173226WA SF311173227WA SF311173228WA SF311173229WA SF311173230WA SF311173231WA SF311173232WA SF311173233WA SF311173234WA SF311173235WA SF311183237WA SF311183238WA SF311183240WA SF311183241WA SF311183242WA SF311183243WA SF311183244WA SF311183246WA SF311183249WA SF311183251WA SF311183252WA SF311183253WA SF311183254WA SF311183255WA SF311183256WA SF311183257WA SF311183258WA SF311183259WA SF311183260WA SF311183261WA SF311183262WA SF311183263WA SF311183264WA SF311183265WA SF311183266WA SF311183268WA SF311183269WA SF311183270WA SF311183271WA SF311183272WA SF311183273WA SF311183274WA SF311183275WA SF311183276WA SF311183277WA SF311183278WA SF311183279WA SF311183280WA SF311183281WA SF311183282WA SF311193283WA SF311193284WA SF311193285WA SF311193286WA SF311193287WA SF311193289WA SF311193308WA SF311193316WA SF311193318WA SF311193319WA SF311193320WA SF311193336WA SF311193337WA SF311283763WA SF311293840WA SF311283788WA SF311283790WA SF311293839WA SF311303968WA SF311314016WA SF311314024WA SF311314042WA SF311293881WA SF311293860WA SF311303906WA SF311303908WA SF311314023WA SF311314052WA SF311314057WA SF311314058WA SF311314088WA SF311314091WA SF311314114WA SF311324195WA SF311334210WA SF311334211WA SF311334219WA SF311334220WA SF311334234WA SF311334267WA SF311334279WA SF311334283WA SF311334295WA SF311314084WA SF311314097WA SF311314104WA SF311314120WA SF311314122WA SF311334301WA SF311414457WA SF311504767WA SF311504772WA SF311504773WA SF311504784WA SF311514785WA SF311514787WA SF311514788WA SF311514789WA SF311514792WA SF311514794WA SF311524843WA SF312024868WA SF312034910WA SF312034911WA SF312034915WA SF312034922WA SF312034925WA SF312034926WA SF312034929WA SF312034930WA SF312034932WA SF312034936WA SF312054937WA SF312054938WA SF312054939WA SF312054940WA SF312054941WA SF312054945WA SF312054949WA SF312054951WA SF312054953WA SF312054954WA SF312054955WA SF312054957WA SF312054959WA SF312054960WA SF312054961WA SF312054964WA SF312054965WA SF312054969WA SF312054970WA SF312054974WA SF312054976WA SF312054977WA SF312054978WA SF312054980WA SF312054983WA SF312054985WA SF312054986WA SF312054987WA SF312054989WA SF312054991WA SF312054994WA SF312065004WA SF312065010WA SF312054947WA SF312054996WA SF312075040WA SF312075049WA SF312205534WA SF312205558WA SF312205609WA SF312376097WA SF312376100WA SF312376110WA SF312376114WA SF312376117WA SF312376120WA SG212410996WA SG212411000WA SG212411003WA SG212411006WA SG212411007WA SG212411008WA SG212411012WA SG212411013WA SG212411014WA SG212411015WA SG212411016WA SG212411017WA SG212411018WA SGF12414089WA SGF12414090WA SGF12414091WA SGF12414092WA SGF12414093WA SGF12414094WA SGF12414095WA SGF12414096WA SGF12414097WA SGF12414098WA SGF12414099WA SGF12414100WA SGF12414101WA SGF12414102WA SGF12414103WA SGF12414104WA SGF12414105WA SGF12414106WA SGF12414107WA SGF12414108WA SGF12414109WA SGF12414110WA SGF12414111WA SGF12414112WA SGF12414113WA SGF12414114WA SGF12414115WA SGF12414116WA SGF12414117WA SGF12414119WA SGF12414120WA SGF12414121WA SGF12414122WA SGF12414123WA SGF12414124WA SGF12414125WA SGF12414126WA SGF12414127WA SGF12414128WA SGF12414129WA SGF12414130WA SGF12414131WA SGF12414132WA SGF12414133WA SGF12414134WA SGF12414135WA SGF12414136WA SGF12424139WA SGF12424140WA SGF12424141WA SGF12424148WA SGF12424149WA SGF12424153WA SGF12424154WA SGF12424155WA SGF12424156WA SGF12424157WA SGF12424158WA SGF12424159WA SGF12414052WA SGF12414059WA SGF12414065WA SGF12414069WA SGF12414001WA SGF12414002WA SGF12414005WA SGF12414006WA SGF12414060WA SGF12414070WA SGF12414071WA SGF12414072WA SG212410927WA SG212410944WA SG212410945WA SG212410947WA SG212410948WA SG212410949WA SG212410950WA SG212410951WA SG212410952WA SG212410953WA SGF12414027WA SGF12414030WA SGF12414035WA SGF12414038WA SGF12414040WA SGF12414042WA SGF12414043WA SGF12414044WA SGF12414045WA SGF12414046WA SGF12414050WA SGF12414051WA SGF12414054WA SGF12414055WA SGF12414061WA SGF12414063WA SGF12414064WA SGF12414066WA SGF12414067WA SGF12414068WA SG212410925WA SG212410930WA SG212410931WA SG212410932WA SG212410933WA SG212410934WA SG212410936WA SG212410941WA SGF12414004WA SGF12414007WA SGF12414008WA SGF12414009WA SGF12414010WA SGF12414012WA SGF12414014WA SGF12414015WA SGF12414016WA SGF12414017WA SGF12414018WA SGF12414019WA SGF12414020WA SGF12414021WA SGF12414022WA SGF12414023WA SGF12414024WA SGF12414026WA SGF12414031WA SGF12414032WA SGF12414034WA SGF12414036WA SGF12414028WA SGF12414029WA SGF12414033WA SGF12414037WA SGF12414039WA SGF12414041WA SGF12424137WA SGF12424138WA SGF12424142WA SGF12424143WA SGF12424144WA SGF12424145WA SGF12424146WA SGF12424151WA SGF12424160WA SGF12424147WA SGF12424150WA SGF12424176WA SGF12424177WA SGF12424178WA SGF12424179WA SGF12424180WA SGF12424181WA SGF12424182WA SGF12424183WA SGF12424185WA SF312416222WA SF312416232WA SG212421065WA SG212421066WA SG212421067WA SG212421068WA SG212421070WA SG212421071WA SG212421072WA SG212421073WA SG212421074WA SG212421075WA SG212421076WA SG212421077WA SG212421078WA SG212421079WA SG212421080WA SG212421081WA SG212421082WA SG212421083WA SG212421084WA SG212421085WA SG212421086WA SG212421087WA SG212421088WA SG212421089WA SG212421090WA SG212421036WA SG212421069WA SG212421092WA SGF12424212WA SGF12424213WA SGF12424215WA SGF12424220WA SGF12424221WA SGF12424222WA SGF12424223WA SGF12424224WA SGF12424225WA SGF12424226WA SGF12424227WA SGF12424228WA SGF12424229WA SGF12424230WA SGF12424231WA SGF12424232WA SGF12424233WA SGF12424234WA SGF12424235WA SGF12424236WA SGF12424237WA SGF12424238WA SGF12424241WA SGF12424242WA SGF12424243WA SG212431094WA SG212431096WA SG212431099WA SG212431100WA SG212431102WA SG212431111WA SG212431112WA SG212431113WA SG212431114WA SG212431115WA SG212431116WA SG212431117WA SG212431118WA SG212431119WA SG212431120WA SG212431123WA SF312436279WA SGF12424258WA SGF12474354WA SGF12474356WA SGF12474357WA SGF12474358WA SGF12474359WA SGF12464331WA SGF12474346WA SG212471140WA SG212471142WA SG212471143WA SG212471144WA SG212471146WA SG212471147WA SG212471149WA SG212471150WA SG212471151WA SG212471153WA SG212471154WA SG212471158WA SG212471159WA SG212471160WA SG212471163WA SG212471164WA SG212471165WA SG212471166WA SG212471167WA SG212471168WA SG212471169WA SG212471170WA SG212471171WA SG212471172WA SG212471173WA SG212471174WA SG212471175WA SG212471176WA SG212471177WA SG212471179WA SG212471180WA SG212471181WA SG212471182WA SG212471183WA SG212481190WA SG212471139WA SG212481191WA SG212481192WA SG212481193WA SG212481196WA SG212481200WA SG212481204WA SG212481205WA SG212481206WA SG212481208WA SG212481210WA SG212481211WA SG212481212WA SG212481213WA SG212481214WA SG212481215WA SG212481216WA SG212481218WA SG212481219WA SG212481220WA SG212481221WA SG212481222WA SG212481223WA SG212481229WA SG212481230WA SG212481232WA SG212481233WA SG212481239WA SG212481240WA SG212481241WA SG212481247WA SG212481248WA SGF12484487WA SGF12484572WA SGF12484576WA SGF12484581WA SGF12484583WA SGF12484584WA SGF12484585WA SGF12484586WA SGF12484588WA SGF12484589WA SGF12484590WA SGF12484591WA SGF12484592WA SGF12484593WA SGF12484596WA SGF12484597WA SGF12484598WA SGF12484599WA SG212471141WA SG212471148WA SG212471155WA SG212471156WA SG212471157WA SG212471161WA SG212471162WA SG212471178WA SG212481251WA SGF12484629WA SGF12484630WA SGF12484632WA SGF12484633WA SGF12484636WA SGF12484639WA SGF12484640WA SGF12484641WA SGF12484642WA SGF12484643WA SGF12484645WA SGF12484654WA SGF12474352WA SGF12474374WA SGF12474380WA SGF12474381WA SGF12474384WA SGF12474385WA SGF12484582WA SGF12484625WA SGF12484626WA SGF12484627WA SGF12484628WA SGF12484631WA SGF12484635WA SGF12484637WA SGF12484638WA SG212491350WA SG212491357WA SG213041718WA SG213051722WA SG213051740WA SG213051741WA SG213051742WA SG213051744WA SG213051756WA SG213051757WA SG213051758WA SG213051759WA SG213051760WA SG213051761WA SG213051782WA SG213051783WA SG213051784WA SG213051785WA SG213051786WA SG213051787WA SG213051788WA SG213051789WA SG213051790WA SG213051791WA SG213061814WA SG213061815WA SG213061816WA SG213061817WA SG213061818WA SG213061819WA SG213061820WA SG213061821WA SG213061822WA SG213061823WA SG213061824WA SG213061825WA SG213061826WA SG213061827WA SG213061828WA SG213061829WA SG213061830WA SG213061831WA SG213061832WA SG213061833WA SG213061834WA SG213061835WA SG213061836WA SG213061837WA SG213061838WA SG213061839WA SG213061840WA SG213061841WA SG213061842WA SG213061843WA SG212491379WA SG213031636WA SG213031637WA SG213031639WA SG213031643WA SG213031644WA SG213031645WA SG213031646WA SGF13025299WA SGF13025302WA SGF13035350WA SGF13035351WA SGF13035352WA SGF13035353WA SGF13035354WA SGF13035355WA SGF13035356WA SGF13035359WA SGF13035365WA SGF13035366WA SGF13035398WA SGF13035400WA SGF13035404WA SGF13035406WA SGF13035407WA SGF13035408WA SGF13035409WA SGF13035410WA SGF13035411WA SGF13035412WA SGF13035413WA SGF13035432WA SGF13035403WA SGF13035405WA SGF13035414WA SGF13035415WA SGF13035416WA SGF13035417WA SGF13035418WA SGF13035419WA SGF13035420WA SGF13035421WA SGF13035422WA SGF13035423WA SGF13035425WA SG213031656WA SF310030294WA SF310030306WA SF312024891WA SF312024893WA SF312024898WA SF312024899WA SF312024900WA SF312024902WA SF312024903WA SF312024904WA SF312024906WA SF312034917WA SF310060373WA SF310060377WA SF310060382WA SF310060383WA SF310060384WA SF311484603WA SF312024884WA SF311082845WA SF311082846WA SF311082847WA SF311082848WA SF311082849WA SF311082850WA SF311082851WA SF311082852WA SF311082853WA SF311082855WA SF311082856WA SF311082857WA SF311082858WA SF311082863WA SF311082864WA SF311082865WA SF311082866WA SF311082871WA SF311092873WA SF311102874WA SF311112952WA SF311112957WA SF311072733WA SF311072740WA SG212410946WA SG212481266WA SG212481267WA SG212481269WA SG212491274WA SG212491275WA SG212491276WA SG212491277WA SG212491278WA SG212491279WA SG212491282WA SG212491283WA SG212491317WA SG212491319WA SG212491321WA SG212491323WA SG212491324WA SG212491325WA SG212491327WA SG212491330WA SG212491331WA SG212491332WA SG212491333WA SG212491335WA SG212491336WA SG212491338WA SG212491339WA SG212491340WA SG212491341WA SG212491342WA SG212491343WA SG212491344WA SG212491365WA SGF12494699WA SGF12494703WA SGF12494708WA SG213041693WA SG213041694WA SG213041695WA SG213041696WA SG213041697WA SG213041698WA SG213041699WA SG213041700WA SG213041701WA SG213041702WA SG213041703WA SG213041704WA SF310241021WA SF310241026WA SF310241032WA SF310241033WA SF310241035WA SF310241036WA SF310241072WA SF310241073WA SF310241074WA SF310251075WA SF310251076WA SF310251077WA SF310251078WA SF310251079WA SF310251080WA SF310251081WA SF310251082WA SF310251083WA SF310251084WA SF310251085WA SF311062709WA SF311052694WA SF311072727WA SF311072748WA SF311072757WA SF311072758WA SF311072760WA SF311072767WA SF311112902WA SF311193312WA SF311193314WA SF311193315WA SGF12515153WA SGF12515161WA SGF12515171WA SGF12525189WA SGF12525198WA SGF12525199WA SGF12525200WA SGF12525201WA SGF12525202WA SGF12525203WA SGF12525204WA SGF13176666WA SGF13176669WA SGF13176672WA SG211360035WA SG211360039WA SGF11350049WA SGF11390097WA SF310271105WA SF310271109WA SF310271114WA SF310271116WA SF310271119WA SF312155295WA SF312155296WA SF312496609WA SF312496672WA SF312496674WA SGF12494704WA SGF12494705WA SF313066923WA SF313106964WA SF312496647WA SF312506721WA SGF13106027WA SF311414410WA SF310331359WA SF311524845WA SF310452049WA SGF12323496WA SG212310807WA SG212050411WA SG212080433WA SG212080434WA SG212080435WA SG212080436WA SG212080437WA SGF12081677WA SGF12081680WA SGF12081681WA SGF12081683WA SGF12081686WA SGF12081687WA SGF12081688WA SGF12081689WA SGF12081690WA SGF12081691WA SGF12081694WA SGF12081695WA SGF12081696WA SGF12081697WA SGF12081698WA SGF12081699WA SGF12323492WA SGF12323493WA SGF12091846WA SG212160604WA SGF12061558WA SGF12091837WA SGF12122240WA SGF11440251WA SGF11440282WA SGF11450306WA SGF11450318WA SGF11450321WA SGF11450322WA SGF11450323WA SGF11450327WA SGF11450329WA SGF11450330WA SG212050385WA SG212050397WA SGF12051404WA SG212260706WA SG212260715WA SG212370896WA SG212370897WA SG212370898WA SG212370899WA SG212370900WA SG212390915WA SG212390916WA SG212390917WA SG212390919WA SG212390920WA SG212390921WA SG212390922WA SG212390923WA SG212390924WA SG211390058WA SG211390059WA SG211390060WA SG211390061WA SG211390062WA SG211410081WA SG211410082WA SG211410090WA SG211410091WA SG211410092WA SG211500223WA SG211500228WA SG211510229WA SG212360878WA SGF12222661WA SGF12242863WA SGF12373692WA SGF12373693WA SGF12081753WA SGF12091864WA SGF12091937WA SGF12091939WA SGF12091950WA SGF12112110WA SGF12373695WA SGF12373698WA SGF12373699WA SF312496667WA SF313026804WA SF313036820WA SGF12051406WA SGF12051434WA SGF12051438WA SGF12061545WA SGF12061551WA SGF12061560WA SGF12061569WA SGF12373697WA SGF13106145WA SF312095136WA SF311303934WA SF311072786WA SF310321323WA SF312496604WA SF312526755WA SF311504770WA SF310331343WA SF310331345WA SF310331349WA SF310331350WA SF310331352WA SF310331353WA SF310331354WA SF310331355WA SF310331356WA SF310331358WA SF310331360WA SF312386130WA SF312195515WA SF312195526WA SF312205549WA SF312205550WA SF312205551WA SF312205557WA SF312205576WA SF312205577WA SF312496683WA SF312516735WA SF312516736WA SF312516737WA SF312516738WA SF312516739WA SF312516740WA SF312526746WA SF310110559WA SF310271128WA SF310271131WA SF310301134WA SF310301135WA SF310301136WA SF310301137WA SF310301138WA SF310301139WA SF310301140WA SF310301141WA SF310301142WA SF310301143WA SF310301144WA SF310301145WA SF310301146WA SF310301147WA SF310301148WA SF310301149WA SF310301150WA SF310301151WA SF310301153WA SF310301154WA SF310301155WA SF310301156WA SF310301157WA SF310301158WA SF310301159WA SF310301160WA SF310301161WA SF310301162WA SF310301163WA SF310301164WA SF310301165WA SF310301166WA SF310301167WA SF310301170WA SF310301171WA SF310331363WA SF310331362WA SF310452034WA SF310472217WA SF310472222WA SF310482331WA SF310492350WA SF310492349WA SF310492353WA SF310492365WA SF310482285WA SF310482325WA SF311042628WA SF311042627WA SF311042638WA SF311042632WA SF311153137WA SF311153145WA SF311153146WA SF311153147WA SF311153148WA SF311173210WA SF311193290WA SF311223406WA SF311223411WA SF311223415WA SF311223416WA SF311223430WA SF311243514WA SF311243517WA SF311243588WA SF311243600WA SF311243602WA SF311243604WA SF311314032WA SF311314099WA SF311334221WA SF311334223WA SF311293879WA SF311293884WA SF311314010WA SF311324160WA SF311334230WA SF311334254WA SF311334262WA SF311434500WA SF311434539WA SF311424459WA SF311424471WA SF311504699WA SF311504698WA SF311514818WA SF312024860WA SF312024862WA SF312024864WA SF312024863WA SF312024865WA SF312065001WA SF312085123WA SF312085126WA SF312095138WA SF312095139WA SF312095144WA SF312105212WA SF312105214WA SF312105215WA SF312105216WA SF312105217WA SF312105219WA SF312105220WA SF312105227WA SF312105228WA SF312095168WA SF312095172WA SF312095181WA SF312095183WA SF312095185WA SF312095186WA SF312095197WA SF312095198WA SF312095201WA SF312095202WA SF312095184WA SF312095199WA SF312105210WA SF312155280WA SF312155281WA SF312155282WA SF312155283WA SF312155284WA SF312155285WA SF312155286WA SF312155289WA SF312155292WA SF312155297WA SF312155298WA SF312155299WA SF312155300WA SF312155301WA SF312155302WA SF312155303WA SF312155304WA SF312155305WA SF312155306WA SF312155307WA SF312155308WA SF312155309WA SF312155310WA SF312155311WA SF312155312WA SF312155326WA SF312165371WA SF312175416WA SF312175417WA SF312175419WA SF312175421WA SF312175422WA SF312175424WA SF312175425WA SF312175440WA SF312175446WA SF312175447WA SF312175452WA SF312175455WA SF312195478WA SF312195480WA SF312195481WA SF312195482WA SF312195486WA SF312195487WA SF312195489WA SF312195495WA SF312205595WA SF312265697WA SF312325907WA SF312335927WA SF312335929WA SF312335931WA SF312335932WA SF312335933WA SF312335934WA SF312335935WA SF312335936WA SF312335937WA SF312335938WA SF312275716WA SF312275727WA SF312285749WA SF312295769WA SF312295770WA SF312295775WA SF312295776WA SF312295777WA SF312295778WA SF312295779WA SF312295781WA SF312335951WA SF312356018WA SF312356019WA SF312376095WA SF312376103WA SF312376104WA SF312376106WA SF312376109WA SF312376073WA SF312396180WA SGF12434297WA SGF12434298WA SF312416238WA SGF12434328WA SGF12434329WA SG212431101WA SG212431124WA SF312456482WA SGF12474341WA SG212491285WA SG212491289WA SGF12494706WA SGF12494709WA SGF12494712WA SGF12494714WA SGF12494716WA SF312416253WA SF312466538WA SF312466539WA SF312466542WA SF312476590WA SF312476591WA SF312476595WA SF312496607WA SF312476600WA SF312496612WA SF312496616WA SF312496643WA SF312496666WA SF312496663WA SF312526743WA SF312526758WA SF312526766WA SF312526768WA SF312526770WA SF312526771WA SF313016793WA SF313026808WA SF313036821WA SF313036825WA SF313036826WA SF313036829WA SF313036830WA SF313036832WA SF313036833WA SGF13025344WA SGF13025345WA SGF13035447WA SF312526759WA SF312526774WA SF313026807WA SF313026814WA SF313066910WA SF313066918WA SF313066921WA SF313066931WA SF313066933WA SF313066936WA SF313066942WA SF313036828WA SF313046856WA SF313046857WA SF313046859WA SF313046861WA SF313046866WA SF313066925WA SF313066926WA SF313066939WA SG213061793WA SG213061812WA SG213061813WA SG213081844WA SG213061809WA SG213061811WA SG213091865WA SG213091866WA SG213091867WA SG213091868WA SG213091869WA SG213091872WA SG213091873WA SGF13085825WA SGF13085826WA SGF13085829WA SGF13085830WA SF313066943WA SGF13085874WA SGF13085876WA SGF13085877WA SGF13085879WA SGF13085880WA SGF13085881WA SGF13085882WA SGF13095906WA SF313106976WA SGF13106153WA SGF13106144WA SGF13106146WA SGF13106147WA SGF13106148WA SGF13106149WA SGF13106150WA SGF13106151WA SGF13106152WA SF313117021WA SF313117023WA SF313117025WA SF313117028WA SF313117011WA SF313137056WA SF313137059WA SF313137060WA SF313137061WA SF313137062WA SF313137063WA SF313137065WA SF313137066WA SF313147067WA SF313147075WA SF313147078WA SF313147080WA SGF13106113WA SGF13106116WA SG213121976WA SJF13110059WA SJF13130127WA SJF13130130WA SF313147089WA SJF13170134WA SJF13170137WA SJF13170138WA SF311504778WA SG212230648WA SG212230664WA SF312456483WA SF312466554WA SF312476588WA SF311072721WA SF311072722WA SF311072731WA SF311072732WA SF311072743WA SF311072745WA SF313106965WA SF311334240WA SF312416230WA SF312446392WA SF311072793WA SF311072796WA SF311072797WA SF311072805WA SF311082868WA SF311082869WA SF311082870WA SF311102878WA SF311112943WA SF311112955WA SF311112966WA SF311424466WA SF311414394WA SF311424467WA SF312034916WA SF312034918WA SF312034913WA SF312095142WA SF312315858WA SF310462145WA SF310462147WA SF311052698WA SF311052699WA SF311052692WA SF312416244WA SF312416245WA SF312416246WA SF312416252WA SF313016785WA SF313016787WA SF313016790WA SF312496619WA SF312496620WA SF312496622WA SF312496652WA SF311324131WA SF311324132WA SF311324133WA SGF12263209WA SGF12263211WA SF311193311WA SF312054948WA SF312054968WA SF312054972WA SF312054973WA SF312054981WA SF312054992WA SF312065012WA SF312065013WA SF312065014WA SF312065015WA SF312065016WA SF312065017WA SF312065018WA SF312065019WA SF312065020WA SF312065021WA SF312065022WA SF312065025WA SF312065026WA SF312065029WA SF312115239WA SF312115240WA SF312115241WA SF312115242WA SF312115243WA SF312115244WA SF312115249WA SF312115250WA SF312115251WA SF312115252WA SF310431909WA SF310431950WA SF310431967WA SF310431975WA SF310431976WA SF310441979WA SF310441980WA SF310441982WA SF310441983WA SF310441984WA SF311193326WA SG211500218WA SG211500219WA SG211500226WA SG211500227WA SGF11490461WA SGF11490463WA SGF11490467WA SGF11490478WA SGF12051460WA SGF12051464WA SGF12051469WA SGF12051470WA SGF12061480WA SG212120492WA SG212120495WA SG212120503WA SG212170612WA |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
GE Healthcare has recently become aware of a potential safety issue due to ECG filter settings associated with the B20/B40 V1, B30 and B40 V2 Patient Monitors.
The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly if the monitor starts with impedance respiration set in the ON position. Instead of the user selected filter, an additional 0.5 Hz high pass filter is used. The
|
FDA Determined Cause 2 |
Software design |
Action |
GE Healthcare sent an Urgent Medical Device Correction letter dated October 10, 2013, to all affected consignees.
The letter described the safety Issue, Safety Instructions, Affected Product, Details, Product Correction and Contact Information (Technical Support at 1-800-558-7044).Consignees were asked to retain a copy of the instructions with the user manual.
For questions regarding this recall call 262-513-4122. |
Quantity in Commerce |
16,061 (343 US, 15,718 OUS) |
Distribution |
Worldwide Distribution - USA (nationwide) to CA, CT, FL, GA, MO, NC, NJ, PA, TX, WI and Internationally to:
FRANCE, PORTUGAL, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELGIUM, BOLIVIA, BOSNIA 7 HERZEGOVINA, BRAZIL, BRUNEI DARUSSALAM, BULGARIA, NEW ZEALAND, CHILE, CHINA, COLOMBIA, CONGO, COSTA RICA, COTE DLVOIRE, CROATIA, CYPRUS,. CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, GABON, GERMANY, GREECE, GUATEMALA, HONDURAS, HUNGARIA, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, MOLDOVA, MOROCCO, MYAMMARA, NAMBIA, NAPAL, NETHERLANDS, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL ROMANIA, RUSSIA, SENEGAL SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SIR LANKA, SWEDEN, SWITZERLAND, THAILAND, THE NETHERLANDS, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES , UNITED KINGDOM, URUGUAY, UZEBEKISTAN, VENEZUELA, and VIETNAM. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS CHINA CO., LTD.
|
|
|
|