| Date Initiated by Firm | December 18, 2015 |
| Date Posted | January 15, 2016 |
| Recall Status1 |
Terminated 3 on January 04, 2017 |
| Recall Number | Z-0630-2016 |
| Recall Event ID |
72860 |
| 510(K)Number | K041427 |
| Product Classification |
Glucose oxidase, glucose - Product Code CGA
|
| Product | arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424.
For the in vitro quantitative determination of glucose in human plasma from whole blood samples. |
| Code Information |
Lot Number EA4M78 |
| FEI Number |
1832816
|
Recalling Firm/ Manufacturer |
Arkray Factory USA, Inc. 5182 W 76th St Minneapolis MN 55439-2900
|
| For Additional Information Contact | Customer Service 877-538-8872 Ext. 3243 |
Manufacturer Reason for Recall | One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL. |
FDA Determined Cause 2 | Component design/selection |
| Action | The firm sent an Urgent Medical Device Recall letter to customers beginning 12/18/2015. The letter identified affected product, stated the reason for recall and asked for unused product to be returned. For instructions on returning unused product contact Lesty Jaranilla 1-877-538-8872 x3243. |
| Quantity in Commerce | 88 boxes (25 foil packaged test strips per box) |
| Distribution | Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CGA
|