Date Initiated by Firm |
January 09, 2017 |
Create Date |
March 24, 2017 |
Recall Status1 |
Terminated 3 on September 28, 2018 |
Recall Number |
Z-1560-2017 |
Recall Event ID |
76255 |
Product Classification |
General surgery tray - Product Code LRO
|
Product |
Regard, Item Number: 880323002, Sterile, LD0754B - C Section - OSU |
Code Information |
037321B Exp Date 2018-08-12 |
Recalling Firm/ Manufacturer |
Resource Optimization & Innovation Llc 2909 N Neergard Ave Springfield MO 65803-6317
|
For Additional Information Contact |
Debbie Seilling-Ferguson 417-820-2793
|
Manufacturer Reason for Recall |
Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
ROi sent an Urgent Recall Notice dated January 9, 2017, to all affected consignees. The recalling firm instructed consignees to quarantine affected product and complete the response form. The recalling firm stated that they will issue over lableing for the affected packs. Sterile replacement light covers will be sent directly to the end user if available. Consignees were instructed to inform their subaccounts of the recall. For questions regarding this recall call 417-820-2793. |
Quantity in Commerce |
30 kits |
Distribution |
US Distribution to LA and TX. |