|
Class 2 Device Recall Euphora Rapid Exchange Balloon Dilatation Catheter |
|
Date Initiated by Firm |
July 13, 2017 |
Date Posted |
September 20, 2017 |
Recall Status1 |
Terminated 3 on August 19, 2020 |
Recall Number |
Z-3136-2017 |
Recall Event ID |
77976 |
510(K)Number |
K143480
|
Product Classification |
Catheters, transluminal coronary angioplasty, percutaneous - Product Code LOX
|
Product |
Euphora Rapid Exchange Balloon Dilatation Catheter, a Percutaneous Transluminal Coronary Angioplasty (PTCA) Rapid Exchange System. |
Code Information |
Product Number/CFN GTIN Lot EUP22512X 00643169559950 212561150 EUP3015X 00643169560154 212561157 EUP1512X 00643169559837 212561158 EUP1515X 00643169559844 212561152 EUP3015X 00643169618916 212572601 EUP2012X 00643169618527 212572603 EUP2520X 00643169560031 212572479 EUP1510X 00643169618633 212572600 EUP2512X 00643169560017 212572471 EUP22515X 00643169559967 212572472 EUP1515X 00643169618688 212577123 EUP2020X 00643169618442 212577128 EUP27520X 00643169560109 212572474 EUP3020X 00643169560161 212572476 EUP1520X 00643169618923 212577129 EUP2512X 00643169618817 212577127 EUP27515X 00643169618855 212553130 EUP2506X 00643169559998 212572475 EUP2510X 00643169560000 212577156 EUP2510X 00643169618732 212577125 EUP2012X 00643169618527 212577130 EUP2015X 00643169618480 212577124 EUP2010X 00643169618862 212577131 EUP2020X 00643169559905 212610574 EUP2512X 00643169560017 212577157 EUP3012X 00643169560147 212577155 EUP2006X 00643169559868 212577159 EUP27515X 00643169560093 212572477 EUP27512X 00643169560086 212577158 EUP2010X 00643169559875 212610579 EUP3515X 00643169560284 212572478 EUP2006X 00643169559868 212610578 EUP22520X 00643169559974 212610565 EUP22512X 00643169559950 212610575 EUP22510X 00643169559943 212610572 EUP3010X 00643169560130 212610518 EUP3006X 00643169560123 212610566 EUP3520X 00643169560291 212610564 EUP2512X 00643169560017 212613013 EUP2520X 00643169560031 212613011 EUP1506X 00643169559813 212577780 EUP1515X 00643169559844 212610576 EUP2030X 00643169559929 212610567 EUP2515X 00643169618435 212577126 EUP2015X 00643169559899 212610577 EUP2512X 00643169560017 212610569 EUP2515X 00643169560024 212610563 EUP2530X 00643169560055 212613004 EUP3512X 00643169560277 212610560 EUP1506X 00643169559813 212613006 EUP2510X 00643169560000 212613009 EUP3515X 00643169560284 212610573 EUP3020X 00643169560161 212613010 EUP3510X 00643169560260 212613005 EUP27512X 00643169560086 212613012 EUP4015X 00643169560413 212610517 EUP3015X 00643169560154 212613003 EUP4012X 00643169560406 212610570 EUP4015X 00643169618381 212613017 EUP2012X 00643169559882 212613000 EUP1520X 00643169559851 212610561 EUP2515X 00643169560024 212613002 EUP22515X 00643169559967 212613008 EUP27515X 00643169560093 212613014 EUP2010X 00643169618862 212613022 EUP2015X 00643169618480 212613024 EUP2520X 00643169618596 212613020 EUP3012X 00643169560147 212613007 EUP3520X 00643169618701 212613018 EUP3030X 00643169560185 212610562 EUP2020X 00643169618442 212621104 EUP2015X 00643169618480 212621109 EUP3515X 00643169560284 212613016 EUP3515X 00643169618497 212613021 EUP1515X 00643169618688 212621111 EUP2520X 00643169618596 212621110 EUP3020X 00643169618510 212621100 EUP1510X 00643169559820 212572473 EUP22515X 00643169618466 212621112 EUP2515X 00643169618435 212621108 EUP27515X 00643169618855 212620619 EUP4030X 00643169560444 212610568 EUP4020X 00643169560420 212610519 EUP4020X 00643169618565 212621103 EUP4010X 00643169618718 212613019 EUP3012X 00643169618985 212621106 EUP2012X 00643169559882 212621114 EUP22512X 00643169559950 212621120 EUP22515X 00643169559967 212621126 EUP2512X 00643169560017 212621118 EUP2510X 00643169560000 212621125 EUP3512X 00643169618664 212620618 EUP3515X 00643169618497 212621102 EUP2530X 00643169618947 212621107 EUP2015X 00643169559899 212621116 EUP2010X 00643169559875 212621117 EUP2510X 00643169618732 212633107 EUP2515X 00643169560024 212621115 EUP2525X 00643169560048 212621113 EUP3020X 00643169560161 212621124 EUP2010X 00643169559875 212633087 EUP3015X 00643169560154 212621122 EUP2520X 00643169560031 212621121 EUP3010X 00643169560130 212621119 EUP2012X 00643169559882 212633095 EUP2512X 00643169560017 212633085 EUP2515X 00643169560024 212633088 EUP1510X 00643169559820 212613015 EUP1515X 00643169559844 212621123 EUP2510X 00643169560000 212633090 EUP2520X 00643169560031 212633097 EUP2506X 00643169559998 212633084 EUP2015X 00643169559899 212633094 EUP2020X 00643169559905 212633092 EUP2012X 00643169618527 212633101 EUP22510X 00643169559943 212633091 EUP3520X 00643169560291 212633086 EUP3012X 00643169560147 212633089 EUP2015X 00643169618480 212633102 EUP2020X 00643169618442 212633108 EUP2010X 00643169618862 212653764 EUP2515X 00643169618435 212633106 EUP2530X 00643169560055 212633096 EUP3006X 00643169560123 212633093 EUP22512X 00643169618893 212633103 EUP2512X 00643169618817 212633104 EUP2515X 00643169618435 212653770 EUP2012X 00643169559882 212663029 EUP3015X 00643169618916 212621101 EUP1510X 00643169618633 212653765 EUP3010X 00643169618831 212633105 EUP1515X 00643169618688 212653766 EUP2015X 00643169618480 212653767 EUP2010X 00643169559875 212663035 EUP3010X 00643169618831 212653768 EUP27520X 00643169618671 212653769 EUP1506X 00643169559813 212663039 EUP2512X 00643169560017 212663026 EUP2015X 00643169559899 212663031 EUP2020X 00643169559905 212663033 EUP2515X 00643169560024 212663030 EUP2510X 00643169560000 212663037 EUP2520X 00643169560031 212663036 EUP3015X 00643169560154 212663032 EUP22512X 00643169559950 212663027 EUP3012X 00643169560147 212663034 EUP2030X 00643169559929 212663240 EUP22520X 00643169559974 212663025 EUP2510X 00643169618732 212653772 EUP2520X 00643169618596 212653775 EUP3020X 00643169618510 212663248 EUP2012X 00643169618527 212653771 EUP3020X 00643169560161 212663038 EUP3520X 00643169618701 212663241 EUP3510X 00643169560260 212663028 EUP2512X 00643169618817 212653773 EUP27520X 00643169560109 212663024 EUP2025X 00643169618756 212663245 EUP2525X 00643169618572 212663246 EUP3515X 00643169560284 212671881 EUP3012X 00643169618985 212663243 EUP2525X 00643169560048 212671882 EUP2030X 00643169619012 212663242 EUP2020X 00643169618442 212653774 EUP4025X 00643169560437 212663023 EUP3515X 00643169560284 212691101 EUP22512X 00643169618893 212691109 EUP22515X 00643169618466 212691108 EUP22515X 00643169559967 212690995 EUP4015X 00643169560413 212691102 EUP22510X 00643169618398 212691107 EUP3520X 00643169618701 212691106 EUP4020X 00643169560420 212690996 EUP3512X 00643169560277 212690998 EUP3512X 00643169618664 212691103 EUP3515X 00643169618497 212691112 EUP1512X 00643169618558 212691113 EUP1512X 00643169559837 212690997 EUP37512X 00643169560345 212690999 EUP3510X 00643169560260 212691100 EUP27515X 00643169618855 212691110 EUP27512X 00643169618404 212691111 EUP2506X 00643169618770 212691115 EUP27520X 00643169618671 212691116 EUP4006X 00643169618954 212663247 EUP3515X 00643169618497 212663244 EUP22506X 00643169618657 212701877 EUP2506X 00643169559998 212701870 EUP22520X 00643169618824 212702003 EUP4015X 00643169618381 212691104 EUP22510X 00643169618398 212702009 EUP2012X 00643169559882 212710806 EUP3030X 00643169560185 212701871 EUP22512X 00643169618893 212702007 EUP3030X 00643169618534 212702006 EUP2530X 00643169618947 212702010 EUP32512X 00643169618909 212701876 EUP3530X 00643169618619 212701874 EUP4020X 00643169618565 212691114 EUP32506X 00643169618602 212702000 EUP27520X 00643169618671 212702011 EUP27506X 00643169618800 212701878 EUP27512X 00643169618404 212702012 EUP3515X 00643169618497 212702013 EUP27515X 00643169618855 212702014 EUP2512X 00643169560017 212710699 EUP2515X 00643169560024 212710697 EUP2015X 00643169559899 212710805 EUP32525X 00643169618411 212709022 EUP32506X 00643169618602 212709023 EUP32510X 00643169618879 212702004 EUP32520X 00643169618978 212702005 EUP3525X 00643169618886 212702002 EUP32525X 00643169618411 212701879 EUP37515X 00643169618992 212701873 EUP37506X 00643169618848 212702001 EUP1510X 00643169559820 212701872 EUP37525X 00643169560376 212724523 EUP3012X 00643169560147 212710808 EUP3006X 00643169560123 212710810 EUP27512X 00643169560086 212710812 EUP27520X 00643169560109 212710803 EUP3525X 00643169618886 212709024 EUP2530X 00643169560055 212710804 EUP3010X 00643169560130 212710800 EUP4020X 00643169618565 212702015 EUP4015X 00643169618381 212702008 EUP37512X 00643169618640 212701875 EUP3020X 00643169560161 212710802 EUP3512X 00643169560277 212710801 EUP1510X 00643169618633 212710899 EUP27515X 00643169560093 212710807 EUP1520X 00643169618923 212691105 EUP1512X 00643169618558 212710894 EUP2012X 00643169618527 212710904 EUP32515X 00643169560222 212710696 EUP2515X 00643169618435 212710903 EUP1515X 00643169618688 212710896 EUP2015X 00643169618480 212710901 EUP3015X 00643169618916 212710902 EUP2025X 00643169618756 212710895 EUP2030X 00643169619012 212710898 EUP27515X 00643169618855 212710900 EUP2525X 00643169618572 212710893 EUP3520X 00643169618701 212621105 EUP2530X 00643169618947 212710897 EUP32525X 00643169618411 212710891 EUP2006X 00643169559868 212724527 EUP22515X 00643169618466 212724810 EUP4010X 00643169618718 212724806 EUP1506X 00643169618695 212724801 EUP4030X 00643169618541 212724803 EUP4030X 00643169560444 212724525 EUP3510X 00643169618961 212724802 EUP1515X 00643169559844 212724529 EUP1520X 00643169559851 212724528 EUP2515X 00643169560024 212751952 EUP32506X 00643169618602 212710892 EUP2020X 00643169559905 212710809 EUP2520X 00643169618596 212760548 EUP2015X 00643169618480 212760547 EUP2512X 00643169618817 212773508 EUP2012X 00643169618527 212773507 EUP2512X 00643169560017 212760549 EUP2010X 00643169618862 212773512 EUP2006X 00643169618794 212724807 EUP27510X 00643169560079 212724524 EUP27510X 00643169618589 212724800 EUP27525X 00643169618763 212724804 EUP37525X 00643169618787 212724805 EUP3530X 00643169560314 212751951 EUP2512X 00643169618817 212751956 EUP22515X 00643169559967 212751946 EUP3520X 00643169560291 212751947 EUP3010X 00643169618831 212751955 EUP3515X 00643169560284 212751950 EUP2515X 00643169560024 212773509 EUP2020X 00643169618442 212773501 EUP2012X 00643169559882 212773510 EUP2515X 00643169618435 212773500 EUP2010X 00643169559875 212773502 EUP2510X 00643169560000 212773505 EUP4020X 00643169560420 212751948 EUP2520X 00643169560031 212773503 EUP2015X 00643169559899 212773506 EUP2515X 00643169560024 212779763 EUP3020X 00643169560161 212779639 EUP1520X 00643169618923 212751953 EUP3510X 00643169560260 212751945 EUP2512X 00643169560017 212779762 EUP3015X 00643169560154 212773504 EUP22512X 00643169559950 212779631 EUP1515X 00643169618688 212779633 EUP3012X 00643169560147 212773511 EUP1512X 00643169559837 212724526 EUP3012X 00643169618985 212779761 EUP2015X 00643169618480 212779638 EUP2020X 00643169618442 212779634 EUP2515X 00643169618435 212779635 EUP2530X 00643169618947 212779637 EUP3020X 00643169618510 212779636 EUP2515X 00643169618435 212786709 EUP2020X 00643169618442 212786713 EUP2015X 00643169618480 212786714 EUP3512X 00643169560277 212786704 EUP2515X 00643169560024 212803427 EUP2015X 00643169559899 212803275 EUP1510X 00643169559820 212751949 EUP22520X 00643169559974 212786706 EUP2525X 00643169618572 212786708 EUP22510X 00643169559943 212786703 EUP4012X 00643169618503 212786700 EUP1520X 00643169618923 212779760 EUP4012X 00643169560406 212786701 EUP2510X 00643169618732 212779632 EUP4015X 00643169560413 212786705 EUP3030X 00643169618534 212786711 EUP2015X 00643169559899 212786710 EUP2512X 00643169618817 212786712 EUP2512X 00643169560017 212803426 EUP2515X 00643169618435 212803430 EUP2015X 00643169618480 212803429 EUP2012X 00643169559882 212803428 EUP2020X 00643169618442 212803431 EUP2520X 00643169560031 212803421 EUP3015X 00643169560154 212803276 EUP1512X 00643169618558 212786702 EUP2520X 00643169618596 212803425 EUP3015X 00643169618916 212803424 EUP2012X 00643169618527 212814779 EUP2512X 00643169618817 212803423 EUP2020X 00643169559905 212803277 EUP2010X 00643169559875 212803420 EUP3012X 00643169560147 212803422 EUP3020X 00643169560161 212814943 EUP3515X 00643169618497 212814776 EUP22512X 00643169559950 212814942 EUP2530X 00643169560055 212814770 EUP3515X 00643169560284 212814773 EUP3512X 00643169560277 212814771 EUP2006X 00643169559868 212814945 EUP3010X 00643169560130 212814768 EUP2506X 00643169559998 212814767 EUP1506X 00643169559813 212786707 EUP2030X 00643169559929 212820599 EUP27515X 00643169560093 212820840 EUP2525X 00643169560048 212820595 EUP27512X 00643169560086 212820845 EUP27520X 00643169560109 212820843 EUP22510X 00643169559943 212820844 EUP4020X 00643169560420 212820598 EUP27512X 00643169618404 212820596 EUP3530X 00643169560314 212820597 EUP27510X 00643169618589 212820842 EUP3510X 00643169560260 212820592 EUP4030X 00643169560444 212820590 EUP4006X 00643169618954 212820589 EUP22510X 00643169618398 212820846 EUP3006X 00643169560123 212820591 EUP32515X 00643169560222 212820594 EUP2510X 00643169560000 212803278 EUP1515X 00643169618688 212803279 EUP27520X 00643169618671 212814777 EUP3520X 00643169560291 212814772 EUP1520X 00643169618923 212814944 EUP22520X 00643169559974 212814946 EUP1515X 00643169559844 212814774 EUP1512X 00643169559837 212814778 EUP1510X 00643169559820 212814940 EUP37520X 00643169560369 212820841 EUP37512X 00643169560345 212820593 EUP1510X 00643169618633 212814947 EUP1506X 00643169559813 212814941 EUP2512X 00643169560017 212835751 EUP1520X 00643169559851 212814769 EUP2525X 00643169560048 212835748 EUP2520X 00643169618596 212835807 EUP1510X 00643169559820 212835744 EUP1506X 00643169559813 212835749 EUP1506X 00643169618695 212814775 EUP2030X 00643169559929 212835740 EUP2506X 00643169559998 212835742 EUP2530X 00643169560055 212835746 EUP27510X 00643169560079 212835735 EUP3006X 00643169560123 212835736 EUP27515X 00643169560093 212835747 EUP3010X 00643169560130 212835752 EUP32525X 00643169560246 212835734 EUP3512X 00643169560277 212835738 EUP2030X 00643169619012 212835756 EUP3515X 00643169560284 212835750 EUP3510X 00643169560260 212835732 EUP2530X 00643169618947 212835803 EUP2525X 00643169618572 212835802 EUP22512X 00643169618893 212835759 EUP4015X 00643169618381 212835800 EUP3520X 00643169560291 212835743 EUP37515X 00643169560352 212835737 EUP4025X 00643169618749 212835754 EUP4015X 00643169560413 212835739 EUP4025X 00643169560437 212835731 EUP4010X 00643169618718 212835755 EUP4030X 00643169560444 212835741 EUP37510X 00643169560338 212835733 EUP4006X 006 3169560383 212846533 EUP3010X 00643169618831 212835801 EUP27515X 00643169618855 212835806 EUP2012X 00643169559882 212846536 EUP3012X 00643169618985 212835805 EUP2515X 00643169560024 212846534 EUP2012X 00643169618527 212846665 EUP2006X 00643169618794 212835758 EUP2520X 00643169560031 212846660 EUP3520X 00643169618701 212835804 EUP22520X 00643169559974 212846535 EUP2020X 00643169559905 212846538 EUP2020X 00643169618442 212846667 EUP3020X 00643169560161 212846661 EUP2006X 00643169559868 212846532 EUP22506X 00643169559936 212846663 EUP27520X 00643169560109 212846539 EUP1510X 00643169618633 212846670 EUP1515X 00643169559844 212846537 EUP2512X 00643169618817 212846669 EUP27515X 00643169618855 212846675 EUP2515X 00643169618435 212846668 EUP2506X 00643169618770 212835757 EUP2015X 00643169618480 212853378 EUP27520X 00643169618671 212846666 EUP3010X 00643169618831 212846674 EUP2010X 00643169618862 212853376 EUP27510X 00643169618589 212846664 EUP2012X 00643169618527 212855142 EUP2520X 00643169618596 212846671 EUP2520X 00643169618596 212855143 EUP2015X 00643169618480 212855340 EUP2020X 00643169618442 212855141 EUP3015X 00643169618916 212853374 EUP22520X 00643169618824 212853377 EUP3012X 00643169618985 212853379 EUP3025X 00643169618428 212853372 EUP22515X 00643169618466 212855158 EUP2510X 00643169618732 212855157 EUP3030X 00643169560185 212883773 EUP2015X 00643169559899 212853365 EUP22512X 00643169559950 212853366 EUP2510X 00643169560000 212855345 EUP22515X 00643169559967 212855342 EUP2515X 00643169560024 212853369 EUP2012X 00643169559882 212853368 EUP2520X 00643169560031 212853370 EUP3030X 00643169618534 212855156 EUP2512X 00643169560017 212853367 EUP3012X 00643169560147 212855144 EUP3015X 00643169560154 212855344 EUP2515X 00643169618435 212855140 EUP2515X 00643169560024 212866525 EUP2006X 00643169559868 212866527 EUP22506X 00643169559936 212866522 EUP2512X 00643169560017 212866528 EUP2020X 00643169559905 212866526 EUP2012X 00643169559882 212866529 EUP1520X 00643169559851 212855343 EUP1510X 00643169559820 212866524 EUP22506X 00643169559936 212883775 EUP3015X 00643169560154 212866523 EUP27520X 00643169618671 212883768 EUP3012X 00643169560147 212866530 EUP22520X 00643169618824 212883767 EUP27515X 00643169618855 212866532 EUP2520X 00643169560031 212866531 EUP3025X 00643169618428 212883765 EUP22512X 00643169618893 212883764 EUP2525X 00643169560048 212883774 EUP2515X 00643169618435 212884678 EUP22510X 00643169559943 212904277 EUP32525X 00643169560246 212904272 EUP3512X 00643169560277 212904275 EUP2030X 00643169559929 212904342 EUP3525X 00643169560307 212904266 EUP3510X 00643169560260 212904268 EUP4012X 00643169560406 212904267 EUP27512X 00643169618404 212904348 EUP37520X 00643169560369 212904270 EUP4020X 00643169560420 212904276 EUP3020X 00643169618510 212904355 EUP3515X 00643169560284 212904340 EUP4020X 00643169618565 212904347 EUP3520X 00643169560291 212904278 EUP3015X 00643169618916 212904357 EUP37510X 00643169560338 212904273 EUP4030X 00643169560444 212904274 EUP3520X 00643169618701 212904353 EUP2025X 00643169559912 212904279 EUP32515X 00643169560222 212917017 EUP32525X 00643169560246 212917087 EUP3010X 00643169560130 212917082 EUP22515X 00643169559967 212917019 EUP2506X 00643169618770 212917092 EUP2010X 00643169618862 212917089 EUP1506X 00643169618695 212883766 EUP4025X 00643169560437 212904271 EUP3515X 00643169618497 212904351 EUP27506X 00643169618800 212904352 EUP3512X 00643169618664 212904346 EUP22525X 00643169618930 212917090 EUP2020X 00643169618442 212917143 EUP2006X 00643169618794 212917096 EUP22510X 00643169618398 212917095 EUP27512X 00643169618404 212917142 EUP2506X 00643169559998 212917080 EUP32515X 00643169618626 212917094 EUP22515X 00643169618466 212917097 EUP2025X 00643169559912 212917084 EUP2030X 00643169619012 212917098 EUP32520X 00643169618978 212904345 EUP22506X 00643169618657 212904356 EUP2030X 00643169559929 212931082 EUP2006X 00643169559868 212917086 EUP37515X 00643169618992 212917093 EUP37515X 00643169560352 212917018 EUP37525X 00643169560376 212917014 EUP1510X 00643169618633 212883772 EUP1520X 00643169618923 212883771 EUP1515X 00643169618688 212883770 EUP4025X 00643169618749 212904344 EUP37510X 00643169618459 212904343 EUP3010X 00643169560130 212931080 EUP2010X 00643169618862 212937999 EUP3510X 00643169560260 212931037 EUP3006X 00643169560123 212931035 EUP1515X 00643169618688 212917140 EUP1515X 00643169559844 212917085 EUP37525X 00643169618787 212917091 EUP1512X 00643169618558 212904354 EUP37512X 00643169560345 212931033 EUP2020X 00643169618442 212931097 EUP3510X 00643169618961 212931088 EUP3020X 00643169618510 212931091 EUP3015X 00643169618916 212931094 EUP1510X 00643169618633 212917141 EUP4010X 00643169560390 212917016 EUP1520X 00643169618923 212904358 EUP2020X 00643169618442 212948302 EUP2020X 00643169618442 212938422 EUP2030X 00643169619012 212938416 EUP2510X 00643169618732 212938429 EUP2520X 00643169618596 212948301 EUP2520X 00643169618596 212938418 EUP3015X 00643169618916 212948300 EUP2506X 00643169618770 212947379 EUP2515X 00643169618435 212938428 EUP27510X 00643169618589 212938409 EUP2515X 00643169618435 212947378 EUP3015X 00643169618916 212938426 EUP3020X 00643169618510 212938420 EUP3006X 00643169618473 212938414 EUP22515X 00643169559967 212954717 EUP2010X 00643169618862 212955144 EUP22510X 00643169559943 212954716 EUP2510X 00643169618732 212931092 EUP3506X 00643169560253 212931034 EUP27525X 00643169618763 212931085 EUP3010X 00643169618831 212931093 EUP27510X 00643169560079 212938682 EUP22520X 00643169618824 212931090 EUP2506X 00643169559998 212931039 EUP32512X 00643169618909 212931087 EUP2520X 00643169618596 212931095 EUP3006X 00643169618473 212931084 EUP1506X 00643169618695 212938412 EUP1520X 00643169559851 212931038 EUP1520X 00643169618923 212931096 EUP1515X 00643169618688 212931098 EUP22512X 00643169559950 212938689 EUP37512X 00643169618640 212931089 EUP1515X 00643169618688 212938424 EUP4006X 00643169618954 212931086 EUP4006X 00643169560383 212931036 EUP22512X 00643169559950 212954719 EUP3025X 00643169618428 212938415 EUP22520X 00643169618824 212938430 EUP2020X 00643169618442 212955151 EUP2525X 00643169560048 212954718 EUP22515X 00643169618466 212955146 EUP3030X 00643169560185 212954715 EUP27515X 00643169560093 212954714 EUP27510X 00643169560079 212955141 EUP22510X 00643169618398 212955145 EUP3020X 00643169618510 212955148 EUP3030X 00643169618534 212955143 EUP2515X 00643169618435 212955149 EUP1515X 00643169559844 212938694 EUP1506X 00643169559813 212931081 EUP1512X 00643169618558 212917099 EUP2520X 00643169618596 212955155 EUP22525X 00643169618930 212938411 EUP1520X 00643169618923 212955153 EUP1515X 00643169618688 212955154 EUP1520X 00643169559851 212955140 EUP3015X 00643169618916 212955152 EUP2010X 00643169618862 212967568 EUP2530X 00643169560055 212938687 EUP22512X 00643169618893 212955150 EUP2515X 00643169618435 212967571 EUP27520X 00643169560109 212967272 EUP22512X 00643169559950 212967277 EUP27512X 00643169560086 212967274 EUP2020X 00643169618442 212967576 EUP2512X 00643169618817 212967572 EUP2520X 00643169618596 212967574 EUP22510X 00643169559943 212967276 EUP2006X 00643169618794 212982866 EUP22512X 00643169618893 212938413 EUP3506X 00643169619005 212931083 EUP4012X 00643169618503 212938404 EUP4010X 00643169618718 212938408 EUP37510X 00643169618459 212938400 EUP1515X 00643169559844 212967275 EUP32510X 00643169618879 212938403 EUP3020X 00643169618510 212967569 EUP2030X 00643169559929 212982863 EUP27520X 00643169618671 212967570 EUP27515X 00643169618855 212955147 EUP3030X 00643169560185 212982862 EUP3015X 00643169618916 212967573 EUP1520X 00643169618923 212967575 EUP2020X 00643169618442 212982878 EUP3530X 00643169560314 212938681 EUP32512X 00643169560215 212938683 EUP37525X 00643169618787 212938402 EUP37520X 00643169618725 212938405 EUP2515X 00643169618435 212982876 EUP2010X 00643169618862 212982872 EUP2520X 00643169618596 212982901 EUP22515X 00643169618466 212982871 EUP27510X 00643169618589 212982870 EUP2512X 00643169618817 212982874 EUP2530X 00643169560055 212982861 EUP3510X 00643169560260 212967273 EUP1512X 00643169618558 212982867 EUP1510X 00643169618633 212982877 EUP1515X 00643169618688 212982902 EUP3510X 00643169560260 212938691 EUP37520X 00643169560369 212938688 EUP37510X 00643169560338 212938685 EUP4010X 00643169560390 212938690 EUP4012X 00643169560406 212938692 EUP4015X 00643169560413 212938686 EUP2515X 00643169560024 213012238 EUP2512X 00643169560017 213012237 EUP2012X 00643169559882 213012283 EUP2015X 00643169559899 213012281 EUP27520X 00643169618671 213012285 EUP2530X 00643169560055 213012280 EUP3020X 00643169560161 213012239 EUP4025X 00643169560437 212938684 EUP4025X 00643169618749 212938401 EUP3012X 00643169618985 212982900 EUP3025X 00643169618428 212982868 EUP27525X 00643169618763 212982865 EUP3020X 00643169618510 212982873 EUP2530X 00643169618947 212982864 EUP2006X 00643169559868 212990777 EUP1512X 00643169559837 212990773 EUP27515X 00643169560093 212990780 EUP3010X 00643169560130 212990775 EUP37525X 00643169560376 212938693 EUP22512X 00643169559950 213012282 EUP1510X 00643169618633 212990792 EUP3015X 00643169618916 212990793 EUP3525X 00643169618886 212947377 EUP27515X 00643169618855 212982875 EUP22515X 00643169559967 212990784 EUP2010X 00643169618862 212990796 EUP22510X 00643169618398 212990794 EUP1510X 00643169559820 212990776 EUP27520X 00643169618671 212990795 EUP1520X 00643169559851 212990783 EUP3506X 00643169560253 212990762 EUP3515X 00643169560284 212990769 EUP3512X 00643169560277 212990772 EUP3520X 00643169560291 212990767 EUP32520X 00643169560239 212990779 EUP3525X 00643169560307 212990763 EUP32515X 00643169560222 212990768 EUP32512X 00643169560215 212990781 EUP3530X 00643169560314 212990774 EUP37512X 00643169560345 212990770 EUP3512X 00643169618664 212990789 EUP4020X 00643169618565 212990788 EUP3515X 00643169618497 212990790 EUP3530X 00643169618619 212990787 EUP3520X 00643169618701 212990791 EUP1515X 00643169559844 213018314 EUP4020X 00643169560420 212990766 EUP1506X 00643169559813 213018308 EUP4006X 00643169560383 212990771 EUP4030X 00643169560444 212990764 EUP4015X 00643169560413 212990778 EUP2012X 00643169559882 213018320 EUP2515X 00643169560024 213018311 EUP2512X 00643169560017 213018305 EUP32520X 00643169618978 212990786 EUP2520X 00643169560031 213018323 EUP3012X 00643169560147 213018326 EUP2015X 00643169559899 213037277 EUP2512X 00643169560017 213037279 EUP22520X 00643169559974 213037278 EUP2515X 00643169560024 213037300 EUP4015X 00643169618381 212938406 EUP37515X 00643169560352 212990765 EUP4012X 00643169560406 212990782 EUP4030X 00643169618541 212990785 EUP1510X 00643169559820 213037276 EUP2010X 00643169618862 213037303 EUP3012X 00643169618985 213037302 EUP3015X 00643169618916 213037304 EUP2512X 00643169560017 213042932 EUP2012X 00643169559882 213043301 EUP2515X 00643169560024 213043302 EUP2020X 00643169559905 213042935 EUP3010X 00643169560130 213042936 EUP1520X 00643169559851 213042933 EUP27512X 00643169560086 213042938 EUP3006X 00643169560123 213037301 EUP22515X 00643169559967 213042937 EUP3015X 00643169560154 213042939 EUP2010X 00643169618862 213043305 EUP3020X 00643169560161 213043303 EUP2512X 00643169560017 213055809 EUP27520X 00643169618671 213043304 EUP2012X 00643169559882 213055902 EUP2515X 00643169560024 213055818 EUP2506X 00643169559998 213043300 EUP22510X 00643169559943 213055813 EUP22512X 00643169559950 213055819 EUP3015X 00643169618916 213043306 EUP2010X 00643169559875 213069086 EUP2015X 00643169618480 213069127 EUP27520X 00643169618671 213055903 EUP2010X 00643169618862 213055904 EUP3030X 00643169560185 213055814 EUP1506X 00643169559813 213055816 EUP27515X 00643169560093 213055815 EUP2012X 00643169559882 213069087 EUP1512X 00643169559837 213055811 EUP2030X 00643169619012 213069120 EUP2512X 00643169560017 213069095 EUP3515X 00643169560284 213055812 EUP32510X 00643169560208 213069105 EUP4015X 00643169560413 213042934 EUP22506X 00643169559936 213069080 EUP3020X 00643169560161 213069103 EUP2515X 00643169560024 213069096 EUP2025X 00643169559912 213055817 EUP2006X 00643169559868 213069085 EUP3512X 00643169560277 213055810 EUP1515X 00643169559844 213055900 EUP1510X 00643169618633 213069124 EUP2030X 00643169559929 213069090 EUP2525X 00643169560048 213055901 EUP22512X 00643169559950 213069091 EUP2012X 00643169618527 213069116 EUP27520X 00643169560109 213069100 EUP3010X 00643169560130 213069101 EUP3015X 00643169560154 213069102 EUP2520X 00643169560031 213069097 EUP1512X 00643169618558 213069128 EUP2020X 00643169618442 213069123 EUP1520X 00643169618923 213069132 EUP2530X 00643169560055 213069098 EUP22520X 00643169559974 213069093 EUP2506X 00643169559998 213069094 EUP22515X 00643169559967 213069092 EUP27506X 00643169560062 213069099 EUP2020X 00643169559905 213069089 EUP3515X 00643169560284 213069107 EUP3512X 00643169560277 213069106 EUP3520X 00643169560291 213069108 EUP1512X 00643169559837 213085241 EUP2015X 00643169559899 213085242 EUP4025X 00643169560437 213084959 EUP4015X 00643169618381 213069122 EUP4006X 00643169560383 213069110 EUP27515X 00643169618855 213085243 EUP3012X 00643169560147 213085240 EUP3025X 00643169560178 213069104 EUP2015X 00643169559899 213069088 EUP22512X 00643169618893 213069131 EUP2515X 00643169618435 213069118 EUP2520X 00643169618596 213069119 EUP2512X 00643169618817 213069114 EUP4012X 00643169560406 213069111 EUP2530X 00643169618947 213069126 EUP3010X 00643169618831 213069125 EUP3030X 00643169618534 213069115 EUP3020X 00643169618510 213069117 EUP4020X 00643169560420 213069113 EUP4015X 00643169560413 213069112 EUP2525X 00643169618572 213069129 EUP37510X 00643169560338 213069109 EUP3515X 00643169618497 213069121 EUP2512X 00643169560017 213079177 EUP3512X 00643169618664 213069130 EUP1515X 00643169618688 213079346 EUP2515X 00643169618435 213079349 EUP1506X 00643169618695 213079350 EUP4010X 00643169618718 213079347 EUP4006X 00643169618954 213079354 EUP1510X 00643169559820 213079176 EUP2515X 00643169560024 213079179 EUP3510X 00643169618961 213069133 EUP3006X 00643169560123 213079175 EUP22520X 00643169559974 213079178 EUP3520X 00643169618701 213069134 EUP2520X 00643169618596 213079351 EUP27520X 00643169618671 213079352 EUP2520X 00643169618596 213085244 EUP4010X 00643169560390 213079342 EUP2015X 00643169559899 213079341 EUP3015X 00643169560154 213079340 EUP32506X 00643169618602 213079344 EUP2510X 00643169560000 213118315 EUP2012X 00643169559882 213092749 EUP32512X 00643169618909 213079355 EUP3506X 00643169619005 213079345 EUP4012X 00643169618503 213079348 EUP2020X 00643169559905 213092753 EUP2010X 00643169559875 213092757 EUP2512X 00643169560017 213092750 EUP2515X 00643169560024 213092747 EUP2520X 00643169560031 213092758 EUP3015X 00643169560154 213092748 EUP2530X 00643169560055 213092751 EUP2006X 00643169559868 213092746 EUP2012X 00643169618527 213079353 EUP22512X 00643169559950 213092752 EUP4012X 00643169618503 213093647 EUP3020X 00643169560161 213092756 EUP4010X 00643169618718 213093645 EUP4006X 00643169618954 213102876 EUP2506X 00643169559998 213092754 EUP3506X 00643169619005 213093641 EUP3510X 00643169560260 213092755 EUP3510X 00643169618961 213093650 EUP2006X 00643169618794 213093648 EUP2012X 00643169559882 213102868 EUP37520X 00643169618725 213093643 EUP37525X 00643169618787 213093644 EUP4025X 00643169618749 213102924 EUP2512X 00643169560017 213102863 EUP32512X 00643169618909 213093646 EUP32515X 00643169618626 213093642 EUP2515X 00643169560024 213102865 EUP3012X 00643169560147 213102872 EUP1512X 00643169559837 213102873 EUP2020X 00643169618442 213102920 EUP2510X 00643169618732 213118206 EUP2020X 00643169559905 213102874 EUP32510X 00643169618879 213093651 EUP2025X 00643169559912 213102675 EUP2015X 00643169559899 213102871 EUP32506X 00643169618602 213093640 EUP4030X 00643169560444 213125623 EUP2512X 00643169618817 213102923 EUP27520X 00643169560109 213102679 EUP22525X 00643169559981 213102676 EUP22515X 00643169559967 213102860 EUP27515X 00643169560093 213102674 EUP2030X 00643169559929 213102861 EUP3512X 00643169560277 213102862 EUP3020X 00643169618510 213102879 EUP2515X 00643169618435 213102878 EUP1510X 00643169559820 213092759 EUP22520X 00643169559974 213102870 EUP2015X 00643169559899 213118203 EUP1506X 00643169618695 213093649 EUP2520X 00643169560031 213118205 EUP27512X 00643169560086 213117558 EUP1515X 00643169559844 213102864 EUP3006X 00643169560123 213117557 EUP1515X 00643169618688 213102921 EUP2520X 00643169618596 213102925 EUP1520X 00643169618923 213118215 EUP2510X 00643169560000 213145118 EUP2030X 00643169619012 213102922 EUP1506X 00643169559813 213102678 EUP22520X 00643169618824 213102877 EUP2012X 00643169618527 213118209 EUP22506X 00643169618657 213102875 EUP2506X 00643169618770 213118208 EUP2025X 00643169618756 213118213 EUP2020X 00643169559905 213145117 EUP2010X 00643169559875 213145116 EUP27512X 00643169618404 213118207 EUP27515X 00643169618855 213118211 EUP2515X 00643169618435 213145982 EUP2512X 00643169560017 213145109 EUP3520X 00643169560291 213102677 EUP4012X 00643169560406 213118201 EUP3015X 00643169560154 213118204 EUP4020X 00643169560420 213118202 EUP22512X 00643169559950 213145103 EUP22515X 00643169559967 213145104 EUP27520X 00643169618671 213145986 EUP3020X 00643169560161 213145115 EUP2506X 00643169559998 213145105 EUP2530X 00643169560055 213145112 EUP37512X 00643169560345 213117559 EUP1515X 00643169559844 213145108 EUP1520X 00643169559851 213102866 EUP1512X 00643169618558 213145988 EUP1510X 00643169559820 213145107 EUP3530X 00643169560314 213102867 EUP3012X 00643169560147 213145114 EUP27515X 00643169560093 213145111 EUP27510X 00643169560079 213145106 EUP3512X 00643169560277 213145113 EUP2015X 00643169618480 213145987 EUP4010X 00643169560390 213145102 EUP4015X 00643169560413 213145110 EUP2020X 00643169618442 213145985 EUP3012X 00643169618985 213145984 EUP22525X 00643169618930 213145989 EUP3015X 00643169618916 213145983 EUP3006X 00643169618473 213145119 EUP2012X 00643169559882 213151442 EUP3515X 00643169560284 213102869 EUP22515X 00643169618466 213151449 EUP2020X 00643169618442 213151457 EUP22512X 00643169618893 213151450 EUP2025X 00643169559912 213151434 EUP2515X 00643169560024 213151441 EUP2015X 00643169559899 213151447 EUP3510X 00643169560260 213163995 EUP1515X 00643169618688 213151451 EUP3515X 00643169560284 213151435 EUP2006X 00643169559868 213163997 EUP32512X 00643169560215 213163996 EUP3015X 00643169560154 213151445 EUP2520X 00643169560031 213151444 EUP3520X 00643169560291 213151446 EUP3520X 00643169618701 213145981 EUP2010X 00643169618862 213151455 EUP3030X 00643169618534 213145980 EUP22510X 00643169559943 213151439 EUP1506X 00643169559813 213151438 EUP1520X 00643169559851 213151443 EUP3525X 00643169560307 213151437 EUP3506X 00643169560253 213118200 EUP27512X 00643169560086 213151436 EUP2515X 00643169618435 213151452 EUP2510X 00643169618732 213151448 EUP3010X 00643169618831 213151454 EUP2530X 00643169618947 213151459 EUP3020X 00643169618510 213151461 EUP2520X 00643169560031 213158142 EUP2520X 00643169618596 213151453 EUP2012X 00643169559882 213158145 EUP2010X 00643169559875 213158140 EUP2512X 00643169560017 213157534 EUP2015X 00643169559899 213157539 EUP2020X 00643169559905 213157535 EUP3530X 00643169618619 213151460 EUP2525X 00643169618572 213151456 EUP2015X 00643169618480 213158159 EUP2030X 00643169559929 213157532 EUP2012X 00643169618527 213158156 EUP4006X 00643169560383 213151433 EUP4025X 00643169560437 213151440 EUP3010X 00643169560130 213157531 EUP2515X 00643169560024 213157537 EUP2010X 00643169618862 213158185 EUP3020X 00643169560161 213157538 EUP2020X 00643169618442 213158183 EUP1510X 00643169559820 213158144 EUP3512X 00643169618664 213151458 EUP1512X 00643169559837 213157530 EUP22520X 00643169618824 213158153 EUP2530X 00643169618947 213158187 EUP3015X 00643169560154 213158143 EUP1512X 00643169618558 213158186 EUP2530X 00643169560055 213158146 EUP3006X 00643169560123 213171523 EUP2030X 00643169619012 213158151 EUP2520X 00643169618596 213158181 EUP32515X 00643169560222 213186770 EUP32520X 00643169560239 213163998 EUP3515X 00643169618497 213158150 EUP27510X 00643169618589 213158148 EUP1515X 00643169559844 213157536 EUP1520X 00643169618923 213158182 EUP1510X 00643169618633 213158154 EUP3512X 00643169560277 213158141 EUP2525X 00643169560048 213157533 EUP37515X 00643169560352 213157529 EUP2515X 00643169618435 213158158 EUP3025X 00643169618428 213158149 EUP27515X 00643169618855 213158180 EUP3530X 00643169618619 213158152 EUP3525X 00643169618886 213158147 EUP2510X 00643169560000 213163999 EUP27512X 00643169618404 213158155 EUP2012X 00643169559882 213164685
EUP22512X 00643169559950 213164684 EUP3012X 00643169560147 213164681 EUP2512X 00643169560017 213164680 EUP2510X 00643169618732 213164694 EUP3015X 00643169560154 213164689 EUP2020X 00643169618442 213164945 EUP2015X 00643169618480 213164944 EUP1515X 00643169618688 213164948 EUP2515X 00643169560024 213164687 EUP3520X 00643169618701 213158184 EUP1510X 00643169618633 213164946 EUP1520X 00643169618923 213164947 EUP22515X 00643169559967 213164682 EUP22515X 00643169618466 213164695 EUP2012X 00643169618527 213164697 EUP22510X 00643169618398 213164698 EUP2520X 00643169560031 213164691 EUP2015X 00643169559899 213164688 EUP2512X 00643169618817 213164696 EUP2520X 00643169618596 213164941 EUP2515X 00643169560024 213171529 EUP2012X 00643169559882 213171528 EUP2015X 00643169559899 213171534 EUP2010X 00643169559875 213164692 EUP2020X 00643169559905 213164686 EUP1520X 00643169559851 213164690 EUP3515X 00643169560284 213164683 EUP4015X 00643169618381 213158157 EUP2010X 00643169559875 213171536 EUP2510X 00643169560000 213171532 EUP3015X 00643169560154 213171533 EUP2020X 00643169559905 213171535 EUP27525X 00643169618763 213164693 EUP3012X 00643169560147 213171537 EUP3012X 00643169618985 213164942 EUP2515X 00643169618435 213164943 EUP22506X 00643169559936 213171531 EUP22520X 00643169559974 213171527 EUP3020X 00643169560161 213171526 EUP2510X 00643169618732 213171606 EUP2010X 00643169618862 213171622 EUP2020X 00643169618442 213171624 EUP3020X 00643169618510 213164699 EUP2012X 00643169618527 213171614 EUP2030X 00643169619012 213171607 EUP2015X 00643169618480 213171620 EUP2506X 00643169618770 213171605 EUP2006X 00643169618794 213171612 EUP2515X 00643169618435 213171616 EUP27515X 00643169618855 213171619 EUP1515X 00643169618688 213171611 EUP3010X 00643169618831 213171602 EUP2520X 00643169618596 213171621 EUP3015X 00643169618916 213171610 EUP1512X 00643169559837 213171530 EUP1512X 00643169618558 213171613 EUP3020X 00643169618510 213171618 EUP3510X 00643169618961 213171603 EUP3015X 00643169618916 213164940 EUP2530X 00643169618947 213171623 EUP3520X 00643169618701 213171617 EUP3512X 00643169618664 213171604 EUP2525X 00643169618572 213171608 EUP2515X 00643169560024 213180027 EUP2015X 00643169559899 213180033 EUP2520X 00643169560031 213180028 EUP3015X 00643169560154 213180024 EUP2012X 00643169559882 213180031 EUP27515X 00643169560093 213180025 EUP27512X 00643169560086 213180021 EUP1515X 00643169559844 213180034 EUP4030X 00643169560444 213171524 EUP1506X 00643169559813 213180026 EUP2520X 00643169618596 213180036 EUP27520X 00643169560109 213180029 EUP4015X 00643169618381 213171615 EUP3520X 00643169560291 213180032 EUP37520X 00643169560369 213179559 EUP2515X 00643169618435 213180035 EUP2512X 00643169560017 213186772 EUP37512X 00643169560345 213180020 EUP22510X 00643169559943 213180022 EUP2015X 00643169618480 213180038 EUP2506X 00643169559998 213180023 EUP3015X 00643169618916 213180120 EUP2012X 00643169559882 213186778 EUP2020X 00643169618442 213180037 EUP22525X 00643169559981 213186771 EUP22512X 00643169559950 213180030 EUP2515X 00643169560024 213186774 EUP2512X 00643169618817 213186984 EUP2015X 00643169559899 213186981 EUP2520X 00643169560031 213186982 EUP22515X 00643169559967 213186775 EUP3015X 00643169560154 213186779 EUP2020X 00643169559905 213186776 EUP2010X 00643169559875 213191126 EUP3510X 00643169560260 213186777 EUP2015X 00643169559899 213191130 EUP2012X 00643169618527 213180039 EUP2012X 00643169559882 213191124 EUP3020X 00643169560161 213186980 EUP2012X 00643169559882 213191144 EUP1510X 00643169618633 213186986 EUP2010X 00643169559875 213191148 EUP1506X 00643169618695 213186985 EUP2010X 00643169559875 213205549 EUP2015X 00643169559899 213191147 EUP2512X 00643169560017 213191142 EUP2020X 00643169559905 213191141 EUP27510X 00643169560079 213205548 EUP2515X 00643169560024 213191125 EUP4020X 00643169618565 213186983 EUP2510X 00643169560000 213191131 EUP2512X 00643169560017 213191127 EUP2515X 00643169560024 213191143 EUP2520X 00643169560031 213191146 EUP2530X 00643169560055 213191128 EUP3015X 00643169560154 213191145 EUP2015X 00643169618480 213191161 EUP3512X 00643169560277 213191129 EUP2012X 00643169618527 213191154 EUP2010X 00643169618862 213191153 EUP2015X 00643169618480 213189257 EUP2020X 00643169618442 213191158 EUP2020X 00643169618442 213191120 EUP4012X 00643169560406 213186773 EUP2515X 00643169618435 213191159 EUP2010X 00643169618862 213189255 EUP3010X 00643169560130 213191123 EUP1520X 00643169618923 213191149 EUP1515X 00643169618688 213191121 EUP1512X 00643169618558 213191151 EUP3012X 00643169618985 213171609 EUP2520X 00643169618596 213191152 EUP2012X 00643169559882 213196855 EUP2512X 00643169618817 213191160 EUP2015X 00643169559899 213196858 EUP3015X 00643169618916 213191155 EUP27520X 00643169618671 213189256 EUP2010X 00643169559875 213196859 EUP2520X 00643169618596 213191122 EUP2510X 00643169560000 213196847 EUP2512X 00643169560017 213196856 EUP2020X 00643169559905 213196880 EUP3520X 00643169618701 213191156 EUP2520X 00643169560031 213196857 EUP2515X 00643169560024 213196853 EUP3015X 00643169618916 213189258 EUP22512X 00643169559950 213196848 EUP22515X 00643169559967 213196854 EUP3020X 00643169560161 213196845 EUP2510X 00643169618732 213189259 EUP2006X 00643169559868 213196850 EUP3020X 00643169618510 213191157 EUP1512X 00643169559837 213196844 EUP2030X 00643169559929 213196843 EUP1510X 00643169559820 213196846 EUP3030X 00643169560185 213196842 EUP1515X 00643169618688 213196888 EUP4012X 00643169618503 213196911 EUP4030X 00643169618541 213196887 EUP4010X 00643169560390 213205551 EUP2015X 00643169618480 213196905 EUP4015X 00643169560413 213196841 EUP1520X 00643169618923 213196892 EUP2510X 00643169618732 213196896 EUP2020X 00643169618442 213196904 EUP2012X 00643169618527 213196899 EUP22520X 00643169618824 213196891 EUP3515X 00643169618497 213191150 EUP1515X 00643169559844 213196852 EUP1520X 00643169559851 213196851 EUP2025X 00643169618756 213196889 EUP2520X 00643169618596 213196906 EUP2515X 00643169618435 213196902 EUP3015X 00643169618916 213196907 EUP2512X 00643169618817 213196901 EUP2530X 00643169618947 213196890 EUP3020X 00643169618510 213196908 EUP1510X 00643169618633 213196895 EUP2025X 00643169559912 213205553 EUP22515X 00643169618466 213196893 EUP1506X 00643169618695 213196910 EUP27520X 00643169618671 213196897 EUP3515X 00643169560284 213196849 EUP3515X 00643169618497 213196900 EUP27515X 00643169618855 213196909 EUP3012X 00643169560147 213240546 EUP1512X 00643169618558 213196903 EUP3512X 00643169618664 213196894 EUP3506X 00643169560253 213205550 EUP3010X 00643169618831 213196898 EUP3010X 00643169560130 213205552 EUP3010X 00643169560130 213240549 EUP3006X 00643169560123 213205555 EUP3015X 00643169560154 213240555 EUP32515X 00643169560222 213240548 EUP27515X 00643169560093 213240553 EUP27512X 00643169560086 213240547 EUP2506X 00643169559998 213240557 EUP22520X 00643169559974 213240556 EUP2512X 00643169560017 213240663 EUP2012X 00643169559882 213240664 EUP2006X 00643169559868 213240551 EUP2530X 00643169560055 213240660 EUP3012X 00643169618985 213240916 EUP3530X 00643169560314 213205554 EUP2010X 00643169559875 213240559 EUP3520X 00643169560291 213240550 EUP27520X 00643169560109 213240665 EUP2512X 00643169560017 213247279 EUP2515X 00643169560024 213248761 EUP2012X 00643169559882 213247278 EUP2015X 00643169559899 213247277 EUP2520X 00643169560031 213248765 EUP2020X 00643169559905 213248762 EUP2010X 00643169618862 213248776 EUP2015X 00643169559899 213240666 EUP32512X 00643169560215 213240545 EUP3512X 00643169560277 213205556 EUP3512X 00643169560277 213240554 EUP22510X 00643169559943 213240552 EUP3515X 00643169560284 213240662 EUP2030X 00643169619012 213240917 EUP2010X 00643169559875 213248764 EUP3020X 00643169560161 213248763 EUP2510X 00643169560000 213247276 EUP3012X 00643169560147 213248766 EUP3015X 00643169560154 213248760 EUP1520X 00643169618923 213248768 EUP2512X 00643169618817 213248767 EUP2012X 00643169618527 213248774 EUP2020X 00643169618442 213248775 EUP2510X 00643169618732 213248773 EUP3015X 00643169618916 213248771 EUP3020X 00643169618510 213248772 EUP2012X 00643169559882 213258488 EUP2520X 00643169618596 213248770 EUP1512X 00643169559837 213258413 EUP1512X 00643169618558 213258496 EUP2010X 00643169559875 213258490 EUP2006X 00643169559868 213258418 EUP1510X 00643169559820 213258417 EUP2015X 00643169559899 213258491 EUP2512X 00643169560017 213258483 EUP2515X 00643169560024 213258489 EUP2020X 00643169559905 213258485 EUP22512X 00643169559950 213258419 EUP22506X 00643169559936 213258412 EUP2012X 00643169618527 213258494 EUP2510X 00643169560000 213258486 EUP2520X 00643169560031 213258484 EUP22515X 00643169559967 213258415 EUP3015X 00643169560154 213258492 EUP3012X 00643169560147 213258487 EUP2530X 00643169618947 213248769 EUP2015X 00643169618480 213258495 EUP3020X 00643169560161 213258482 EUP1515X 00643169559844 213240661 EUP37510X 00643169560338 213240543 EUP4020X 00643169560420 213240544 EUP4012X 00643169560406 213247275 EUP2010X 00643169618862 213258523 EUP1520X 00643169559851 213258411 EUP27512X 00643169560086 213258481 EUP1520X 00643169618923 213258520 EUP1506X 00643169559813 213240558 EUP2512X 00643169618817 213258498 EUP2015X 00643169618480 213266352 EUP2015X 00643169559899 213274735 EUP2520X 00643169618596 213258521 EUP3012X 00643169618985 213258497 EUP3015X 00643169618916 213258499 EUP2515X 00643169618435 213258493 EUP3512X 00643169560277 213258480 EUP2012X 00643169618527 213266351 EUP2020X 00643169618442 213266354 EUP2530X 00643169618947 213266357 EUP3010X 00643169560130 213258410 EUP2012X 00643169559882 213274736 EUP2506X 00643169559998 213274726 EUP2010X 00643169559875 213274738 EUP2512X 00643169560017 213274731 EUP2510X 00643169560000 213274733 EUP2030X 00643169559929 213274722 EUP2515X 00643169560024 213274730 EUP3015X 00643169618916 213266349 EUP3030X 00643169560185 213258416 EUP22510X 00643169559943 213266358 EUP3012X 00643169618985 213266350 EUP2020X 00643169618442 213258522 EUP3510X 00643169560260 213258414 EUP3520X 00643169618701 213266355 EUP3020X 00643169618510 213266356 EUP2525X 00643169560048 213274721 EUP2010X 00643169618862 213274745 EUP3015X 00643169560154 213274737 EUP2530X 00643169560055 213274724 EUP2012X 00643169618527 213274748 EUP2520X 00643169618596 213266348 EUP3012X 00643169618985 213274750 EUP3010X 00643169618831 213274739 EUP2525X 00643169618572 213274744 EUP2015X 00643169618480 213274752 EUP2520X 00643169560031 213274728 EUP2515X 00643169618435 213274741 EUP2512X 00643169618817 213274743 EUP3015X 00643169618916 213274742 EUP1515X 00643169618688 213274740 EUP1515X 00643169559844 213274725 EUP1510X 00643169559820 213283434 EUP1512X 00643169618558 213274749 EUP2012X 00643169559882 213283522 EUP3515X 00643169560284 213274732 EUP1510X 00643169618633 213266353 EUP1506X 00643169559813 213274734 EUP1512X 00643169618558 213266347 EUP2520X 00643169618596 213274746 EUP2015X 00643169559899 213283438 EUP1520X 00643169618923 213274751 EUP1512X 00643169559837 213283428 EUP2020X 00643169559905 213283437 EUP1520X 00643169559851 213283432 EUP2512X 00643169560017 213283435 EUP22512X 00643169559950 213283429 EUP2515X 00643169560024 213283521 EUP2510X 00643169560000 213283436 EUP2520X 00643169560031 213283520 EUP22515X 00643169559967 213283431 EUP27515X 00643169560093 213283427 EUP2012X 00643169618527 213283535 EUP2015X 00643169618480 213283542 EUP3020X 00643169560161 213283430 EUP22520X 00643169559974 213283426 EUP3010X 00643169560130 213283433 EUP32520X 00643169560239 213283422 EUP2020X 00643169618442 213283538 EUP2510X 00643169618732 213283532 EUP2515X 00643169618435 213283531 EUP2520X 00643169618596 213283540 EUP22515X 00643169618466 213283539 EUP3020X 00643169618510 213283533 EUP2030X 00643169619012 213283534 EUP32506X 00643169560192 213283423 |
Recalling Firm/ Manufacturer |
Medtronic Vascular Galway DBA Medtronic Ireland Parkmore Business Park West Galway Ireland
|
For Additional Information Contact |
Joey Lomicky 763-526-2494
|
Manufacturer Reason for Recall |
Stylette removal difficulties on the Euphora and Solarice products.
|
FDA Determined Cause 2 |
Process design |
Action |
Medtronic sent a communication informing their consignees about the recall and for the customers to take the following actions:
Medtronics records indicate that your facility has received product potentially affected by this issue. As a result, Medtronic requests that you immediately take the following actions:
1. Identify and quarantine all unused affected product as listed in your inventory.
2. Return all unused affected listed product in your inventory to Medtronic. Contact Medtronic Customer Service at 888-283-7868 to initiate a product return and credit. Your local Medtronic Representative can assist you in the return and replacement of this product as necessary.
3. Complete the enclosed Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality or scan and email to RS.CFQFCA@medtronic.com.
Medtronic is communicating this information to the U.S. FDA and other appropriate regulatory agencies as required.
Please share this notification with others who need to be aware in your organization or to any organization where affected product may have potentially been transferred. We are committed to acting responsibly in the best interest of patient safety. We sincerely appreciate your patience as it relates to this matter. If you have any questions, please contact your Medtronic Field Representative or Technical Services at 1-877-526-7890 Ext. 27220. Adverse reactions or quality problems experienced with the use of this product may be reported to your Medtronic Representative and the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. |
Quantity in Commerce |
69,081 (US) |
Distribution |
Worldwide distribution. US Nationwide, including Puerto Rico; ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, BOSNIA AND HERZEGOWINA, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REPUBLIC, ESTONIA, FRANCE, GERMANY, GREAT BRITAIN, GREECE, GUADELOUPE, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, REPUBLIC OF KOREA, LATVIA, LEBANON, LIECHTENSTEIN, LITHUANIA, MALAYSIA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PAPUA NEW GUINEA, PHILIPPINES, POLAND, PORTUGAL, REUNION, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA (SLOVAK REPUBLIC), SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM and VIET NAM |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LOX and Original Applicant = MEDTRONIC INC.
|
|
|
|