|
Class 2 Device Recall UniCel DxH 600 Coulter Cellular Analysis System |
|
Date Initiated by Firm |
November 03, 2017 |
Create Date |
December 13, 2017 |
Recall Status1 |
Terminated 3 on June 19, 2018 |
Recall Number |
Z-0237-2018 |
Recall Event ID |
78611 |
510(K)Number |
K140911
|
Product Classification |
Counter, differential cell - Product Code GKZ
|
Product |
UniCel DxH600, Part Number B23858, automated hematology analyzer
(Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities) |
Code Information |
Serial Number AZ06682 |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
For Additional Information Contact |
469-961-4941
|
Manufacturer Reason for Recall |
The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions:
¿ "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed.
¿ You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance." |
Quantity in Commerce |
1 unit |
Distribution |
CA, AR, HI, CA, OH, WA, MD, Canada, Japan |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = GKZ and Original Applicant = BECKMAN COULTER, INC.
|
|
|
|