Class 2 Device Recall A.T.S. 2200TS Tourniquet Systems |
|
Date Initiated by Firm |
July 27, 2017 |
Create Date |
February 14, 2018 |
Recall Status1 |
Terminated 3 on August 01, 2019 |
Recall Number |
Z-0613-2018 |
Recall Event ID |
78759 |
Product Classification |
Tourniquet, pneumatic - Product Code KCY
|
Product |
A.T.S. 2200TS Tourniquet Systems, Personalized Pressure. Zimmer Surgical, Inc. Dover, OH 44622
Intended to be used by qualified professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities. |
Code Information |
Lot No. Item No. 60400040400, Lot No. 62510548 (Serial Numbers: 2213JAAA, 2213JAAB, 2213JAAC, 2213JAAD, 2213JAAE, 2213JAAF, 2213JAAG, 2213JAAH, 2213JAAJ, 2213JAAK, 2213JAAL, 2213JAAM, 2213JAAN, 2213JAAP, 2213JAAQ, 2213JAAR, 2213JAAS, 2213JAAT, 2213JAAU, 2213JAAW, 2213JABA, 2213JABB, 2213JABC, 2213JABD, 2213JABE, 2213JABF, 2213JABG, 2213JABH); Lot No. 62534445 (Serial Numbers: 2213JABJ, 2213JABK, 2213JABL, 2213JABM, 2213JABN, 2213JABP, 2213JABQ, 2213JABR, 2213JABS, 2213JABT, 2213JABU, 2213JABW, 2213JACA, 2213JACB, 2213JACC, 2213JACD, 2213JACE, 2213JACF, 2213JACG, 2213JACH); Lot No. 62631977 (Serial Numbers: 2214BAAA, 2214BAAB, 2214BAAC, 2214BAAD, 2214BAAE, 2214BAAF, 2214BAAG, 2214BAAH, 2214BAAJ, 2214BAAK, 2214BAAL, 2214BAAM, 2214BAAN, 2214BAAP, 2214BAAQ, 2214BAAR, 2214BAAS, 2214BAAT, 2214BAAU, 2214BAAW); Lot No. 62625059 (Serial Numbers: 2214BABA, 2214BABB, 2214BABC, 2214BABD, 2214BABE, 2214BABF, 2214BABG, 2214BABH, 2214BABJ, 2214BABK, 2214BABL, 2214BABM, 2214BABN, 2214BABP, 2214BABQ, 2214BABR, 2214BABS, 2214BABT, 2214BABU, 2214BABW); Lot No. 62635789 (Serial Numbers: 2214CAAA, 2214CAAB, 2214CAAC, 2214CAAD, 2214CAAE, 2214CAAF, 2214CAAG, 2214CAAH, 2214CAAJ, 2214CAAK, 2214CAAL, 2214CAAM, 2214CAAV, 2214CAAP, 2214CAAQ, 2214CAAR, 2214CAAS, 2214CAAT, 2214CAAU, 2214CAAW); Lot No. 62642067 (Serial Numbers: 2214DAAA, 2214DAAB, 2214DAAC, 2214DAAD, 2214DAAE, 2214DAAF, 2214DAAG, 2214DAAH, 2214DAAJ, 2214DAAK, 2214DAAL, 2214DAAM, 2214DAAN, 2214DAAP, 2214DAAQ, 2214DAAR, 2214DAAS, 2214DAAT, 2214DAAU, 2214DAAW); Lot No. 62642068 (Serial Numbers: 2214EAAA, 2214EAAB, 2214EAAC, 2214EAAD, 2014EAAE, 2214EAAF, 2214EAAG, 2214EAAH, 2214EAAJ, 2214EAAK, 2214EAAL, 2214EAAM, 2214EAAN, 2214EAAP, 2214EAAQ, 2214EAAR, 2214EAAS, 2214EAAT, 2214EAAU, 2214EAAW); Lot No. 62675190 (Serial Numbers: 2214DABA); Lot No. 62732714 (Serial Numbers: 2214FAAA, 2214FAAB, 2214FAAC, 2214FAAD); Lot No. 62727347 (Serial Numbers: 2214FAAE, 2214FAAF, 2214FAAG, 2214FAAH); Lot No. 62727348 (Serial Numbers: 2214FAAJ, 2214FAAK, 2214FAAL); Lot No. 62745916 (Serial Numbers: 2214FAAM, 2214FAAN); Lot No. 62723747 (Serial Numbers: 2214FAAP, 2214FAAQ, 2214FAAR, 2214FAAS, 2214FAAT, 2214FAAU, 2214FAAW, 2214FABA, 2214FABB, 2214FABC, 2214FABD, 2214FABE, 2214FABF, 2214FABG, 2214FABH, 2214FABJ, 2214FABK, 2214FABL, 2214FABM, 2214FABN); Lot No. 62723748 (Serial Numbers: 2214FABP, 2214FABQ, 2214FABR, 2214FABS, 2214FABT, 2214FABU, 2214FABW, 2214FACA, 2214FACB, 2214FACC, 2214FACD, 2214FACE, 2214FACF, 2214FACG, 2214FACH, 2214FACJ, 2214FACK, 2214FACL, 2214FACM, 2214FACN); Lot No. 62758708 (Serial Numbers: 2214GAAA, 2214GAAB, 2214GAAC, 2214GAAD); Lot No. 62758711 (Serial Numbers: 2214GAAE); Lot No. 62760907 (Serial Numbers: 2214GAAF, 2214GAAG); Lot No. 63773338 (Serial Numbers: 2214GAAH, 2214GAAJ, 2214GAAK, 2214GAAL); Lot No. 62773820 (Serial Numbers: 2214GAAM, 2214GAAN); Lot No. 62773463 (Serial Numbers: 2214GAAP); Lot No. 62773471 (Serial Numbers: 2214GAAQ, 2214GAAR, 2214GAAS); Lot No. 62773478 (Serial Numbers: 2214GAAT, 2214GAAU, 2214GAAW, 2214GABA); Lot No. 62773485 (Serial Numbers: 2214GABB, 2214GABC); Lot No. 62748403 (Serial Numbers: 2214GABD, 2214GABE, 2214GABF, 2214GABG, 2214GABH, 2214GABJ, 2214GABK, 2214GABL, 2214GABM, 2214GABN, 2214GABP, 2214GABQ, 2214GABR, 2214GABS, 2214GABT, 2214GABU, 2214GABW, 2214GACA, 2214GACB, 2214GACC); Lot No. 62779664 (Serial Numbers: 2214GACD, 2214GACE); Lot No. 62797832 (Serial Numbers: 2214HAAA, 2214HAAB, 2214HAAC); Lot No. 62797833 (Serial Numbers: 2214HAAD, 2214HAAE, 2214HAAF, 2214HAAG, 2214HAAH); Lot No. 62799449 (Serial Numbers: 2214HAAJ, 2214HAAK, 2214HAAL, 2214HAAM, 2214HAAN, 2214HAAP, 2214HAAQ, 2214HAAR, 2214HAAS, 2214HAAT, 2214HAAU, 2214HAAW, 2214HABA, 2214HABB, 2214HABC, 2214HABD, 2214HABE, 2214HABF, 2214HABG, 2214HABH); Lot No. 62799475 (Serial Numbers: 2214HABJ, 2214HABK, 2214HABL, 2214HABM, 2214HABN, 2214HABP, 2214HABQ, 2214HABR, 2214HABS, 2214HABT, 2214HABU, 2214HABW, 2214HACA, 2214HACB, 2214HACC, 2214HACD, 2214HACE, 2214HACF, 2214HACG, 2214HACH); Lot No. 62824836 (Serial Numbers: 2214IAAA, 2214IAAB, 2214IAAC, 2214IAAD); Lot No. 62838042 (Serial Numbers: 2214IAAE, 2214IAAF, 2214IAAG, 2214IAAH); Lot No. 62812885 (Serial Numbers: 2214IAAJ, 2214IAAK, 2214IAAL, 2214IAAM, 2214IAAN, 2214IAAP, 2214IAAQ, 2214IAAR, 2214IAAS, 2214IAAT, 2214IAAU, 2214IAAW, 2214IABA, 2214IABB, 2214IABC, 2214IABD, 2214IABE, 2214IABF, 2214IABG, 2214IABH); Lot No. 62812853 (Serial Numbers: 2214IABJ, 2214IABK, 2214IABL, 2214IABM, 2214IABN, 2214IABP, 2214IABQ, 2214IABR, 2214IABS, 2214IABT, 2214IABU, 2214IABW, 2214IACA, 2214IACB, 2214IACC, 2214IACD, 2214IACE, 2214IACF, 2214IACG, 2214IACH); Lot No. 62846159 (Serial Numbers: 2214JAAA, 2214JAAB, 2214JAAC); Lot No. 62844916 (Serial Numbers: 2214JAAD); Lot No. 62869981 (Serial Numbers: 2214KAAE, 2214KAAF, 2214KAAG, 2214KAAH, 2214KAAJ); Lot No. 62871639 (Serial Numbers: 2214JAAE, 2214JAAF, 2214JAAG, 2214JAAH); Lot No. 62870928 (Serial Numbers: 2214KAAK, 2214KAAL, 2214KAAM, 2214KAAN, 2214KAAP, 2214KAAQ, 2214KAAR, 2214KAAS, 2214KAAT, 2214KAAU, 2214KAAW, 2214KABA, 2214KABB, 2214KABC, 2214KABD, 2214KABE, 2214KABF, 2214KABG, 2214KABH, 2214KABJ); Lot No. 62877932 (Serial Numbers: 221411110001, 221411110002, 221411110003, 221411110004, 221411110005, 221411110006, 221411110007, 221411110008, 221411110009, 221411110010); Lot No. 62885586 (Serial Numbers: 221411120001, 221411120002, 221411120003, 221411120004, 221411120005); Lot No. 62885587 (Serial Numbers: 221411120006, 221411120007, 221411120008, 221411120009, 221411120010, 221411120011, 221411120012, 221411120013, 221411120014, 221411120015); Lot No. 62895898 (Serial Numbers: 221411200001, 221411200002, 221411200003, 221411200004, 221411200005, 221411200006, 221411200007, 221411200008, 221411200009, 221411200010); Lot No. 62905257 (Serial Numbers: 221411240001); Lot No. 62904091 (Serial Numbers: 221411250001, 221411250002, 221411250003, 221411250004, 221411250005, 221411250006, 221411250007, 221411250008, 221411250009, 221411250010, 221411250011, 221411250012, 221411250013, 221411250014, 221411250015, 221411250016, 221411250017, 221411250018, 221411250019, 221411250020); Lot No. 62904093 (Serial Numbers: 221412010001, 221412010002, 221412010003, 221412010004, 221412010005, 221412010006, 221412010007, 221412010008, 221412010009, 221412010010, 221412010011, 221412010012, 221412010013, 221412010014, 221412010015, 221412010016, 221412010017, 221412010018, 221412010019, 221412010020); Lot No. 62879139 (Serial Numbers: 221412040001, 221412040002, 221412040003, 221412040004, 221412040005); Lot No. 62905838 (Serial Numbers: 221412080001, 221412080002, 221412080003, 221412080004, 221412080005); Lot No. 62895896 (Serial Numbers: 221412120001, 221412120002, 221412120003, 221412120004, 221412120005); Lot No. 62904123 (Serial Numbers: 221412150001, 221412150002, 221412150003, 221412150004, 221412150005, 221412150006, 221412150007, 221412150008, 221412150009, 221412150010, 221412150011, 221412150012, 221412150013, 221412150014, 221412150015, 221412150016, 221412150017, 221412150018, 221412150019, 221412150020); Lot No. 62932244 (Serial Numbers: 221412180001, 221412180002, 221412180003, 221412180004, 221412180005, 221412180006, 221412180007, 221412180008, 221412180009, 221412180010, 221412180011, 221412180012, 221412180013, 221412180014, 221412180015, 221412180016, 221412180017, 221412180018, 221412180019, 221412180020); Lot No. 62935217 (Serial Numbers: 221412190001, 221412190002, 221412190003, 221412190004, 221412190005); Lot No. 62937693 (Serial Numbers: 221501050001, 221501050002, 221501050003, 221501050004, 221501050005); Lot No. 62945082 (Serial Numbers: 221501080001, 221501080002, 221501080003, 221501080004, 221501080005); Lot No. 62945086 (Serial Numbers: 201501080006, 201501080007, 201501080008); Lot No. 62932650 (Serial Numbers: 221503090001, 221503090002, 221503090003, 221503090004, 221503090005, 221503090006, 221503090007, 221503090008, 221503090009, 221503090010, 221503090011, 221503090012, 221503090013, 221503090014, 221503090015, 221503090016, 221503090017, 221503090018, 221503090019, 22150309002); Lot No. 62961804 (Serial Numbers: 221503120001, 221503120002, 221503120003, 221503120004, 221503120005, 221503120006, 221503120007, 221503120008, 221503120009, 221503120010, 221503120011, 221503120012, 221503120013, 221503120014, 221503120015, 221503120016, 221503120017, 221503120018, 221503120019, 221503120020); Lot No. 62976480 (Serial Numbers: 221503190001, 221503190002, 221503190003, 221503190004, 221503190005, 221503190006, 221503190007, 221503190008, 221503190009, 221503190010, 221503190011, 221503190012, 221503190013, 221503190014, 221503190015, 221503190016, 221503190017, 221503190018, 221503190019, 221503190020); Lot No. 62976481 (Serial Numbers: 221503250001, 221503250002, 221503250003, 221503250004, 221503250005, 221503250006, 221503250007, 221503250008, 221503250009, 221503250010, 221503250011, 221503250012, 221503250013, 221503250014, 221503250015, 221503250016, 221503250017, 221503250018, 221503250019, 221503250020); Lot No. 63026460 (Serial Numbers: 221503270001, 221503270002, 221503270003, 221503270004, 221503270005); Lot No. 63007852 (Serial Numbers: 221504080001, 221504080002, 221504080003, 221504080004, 221504080005, 221504080006, 221504080007, 221504080008, 221504080009, 221504080010, 221504080011, 221504080012, 221504080013, 221504080014, 221504080015, 221504080016, 221504080017, 221504080018, 221504080019, 221504080020 Lot No. 63042181 (Serial Numbers: 221504140001, 221504140002, 221504140003, 221504140004, 221504140005); Lot No. 63042182 (Serial Numbers: 221504140006, 221504140007, 221504140008, 221504140009, 221504140010); Lot No. 63007853 (Serial Numbers: 221504140011, 221504140012, 221504140013, 221504140014, 221504140015, 221504140016, 221504140017, 221504140018, 221504140019, 221504140020, 221504140021, 221504140022, 221504140023, 221504140024, 221504140025, 221504140026, 221504140027, 221504140028, 221504140029, 221504140030); Lot No. 63042179 (Serial Numbers: 221504160001, 221504160002, 221504160003); Lot No. 63042184 (Serial Numbers: 221504220001, 221504220002, 221504220003, 221504220004, 221504220005); Lot No. 63037252 (Serial Numbers: 221504250001, 221504250002, 221504250003, 221504250004, 221504250005, 221504250006, 221504250007, 221504250008, 221504250009, 221504250010, 221504250011, 221504250012, 221504250013, 221504250014, 221504250015, 221504250016, 221504250017, 221504250018, 221504250019, 221504250020); Lot No. 63042180 (Serial Numbers: 221504280001, 221504280002, 221504280003, 221504280004, 221504280005, 221504280006, 221504280007, 221504280008, 221504280009, 221504280010, 221504280011, 221504280012, 221504280013, 221504280014, 221504280015, 221504280016, 221504280017, 221504280018, 221504280019, 221504280020); Lot No. 63042183 (Serial Numbers: 221505110001, 221505110002, 221505110003, 221505110004, 221505110005); Lot No. 63037254 (Serial Numbers: 221505120001, 221505120002, 221505120003, 221505120004, 221505120005, 221505120006, 221505120007, 221505120008, 221505120009, 221505120010, 221505120011, 221505120012, 221505120013, 221505120014, 221505120015, 221505120016, 221505120017, 221505120018, 221505120019, 221505120020); Lot No. 63092209 (Serial Numbers: 221506030001, 221506030002, 221506030003); Lot No. 63092210 (Serial Numbers: 221506050001, 221506050002, 221506050003, 221506050004, 221506050005); Lot No. 63072301 (Serial Numbers: 221506100001, 221506100002, 221506100003, 221506100004, 221506100005, 221506100006, 221506100007, 221506100008, 221506100009, 221506100010, 221506100011, 221506100012, 221506100013, 221506100014, 221506100015, 221506100016, 221506100017, 221506100018, 221506100019, 221506100020); Lot No. 63103587 (Serial Numbers: 221506130001, 221506130002, 221506130003, 221506130004, 221506130005, 221506130006, 221506130007, 221506130008, 221506130009, 221506130010, 221506130011, 221506130012, 221506130013, 221506130014, 221506130015); Lot No. 63072324 (Serial Numbers: 221506170001, 221506170002, 221506170003, 221506170004, 221506170005, 221506170006, 221506170007, 221506170008, 221506170009, 221506170010, 221506170011, 221506170012, 221506170013, 221506170014, 221506170015, 221506170016, 221506170017, 221506170018, 221506170019, 221506170020); Lot No. 63103164 (Serial Numbers: 221506170021, 221506170022, 221506170023); Lot No. 63072352 (Serial Numbers: 221506200001, 221506200002, 221506200003, 221506200004, 221506200005, 221506200006, 221506200007, 221506200008, 221506200009, 221506200010, 221506200011, 221506200012, 221506200013, 221506200014, 221506200015, 221506200016, 221506200017, 221506200018, 221506200019, 221506200020); Lot No. 63096468 (Serial Numbers: 221507150001, 221507150002, 221507150003, 221507150004, 221507150005, 221507150006, 221507150007, 221507150008, 221507150009, 221507150010, 221507150011, 221507150012, 221507150013, 221507150014, 221507150015, 221507150016, 221507150017, 221507150018, 221507150019, 221507150020); Lot No. 63109655 (Serial Numbers: 221507160001, 221507160002, 221507160003); Lot No. 63103590 (Serial Numbers: 221507210001, 221507210002, 221507210003, 221507210004, 221507210005, 221507210006, 221507210007, 221507210008, 221507210009, 221507210010, 221507210011, 221507210012, 221507210013, 221507210014, 221507210015); Lot No. 63109651 (Serial Numbers: 221507280001, 221507280002, 221507280003, 221507280004, 221507280005, 221507280006, 221507280007, 221507280008, 221507280009, 221507280010); Lot No. 63116357 (Serial Numbers: 221507290001, 221507290002, 221507290003, 221507290004, 221507290005, 221507290006, 221507290007, 221507290008, 221507290009, 221507290010, 221507290011, 221507290012, 221507290013, 221507290014, 221507290015, 221507290016, 221507290017, 221507290018, 221507290019, 221507290020); Lot No. 63123305 (Serial Numbers: 221508040001, 221508040002, 221508040003, 221508040004, 221508040005, 221508040006, 221508040007, 221508040008, 221508040009, 221508040010, 221508040011, 221508040012, 221508040013, 221508040014, 221508040015, 221508040016, 221508040017, 221508040018, 221508040019, 221508040020); Lot No. 63138553 (Serial Numbers: 221508260001, 221508260002, 221508260003, 221508260004, 221508260005, 221508260006, 221508260007, 221508260008, 221508260009, 221508260010, 221508260011, 221508260012, 221508260013, 221508260014, 221508260015, 221508260016, 221508260017, 221508260018, 221508260019, 221508260020); Lot No. 63138631 (Serial Numbers: 221509090001, 221509090002, 221509090003, 221509090004, 221509090005, 221509090006, 221509090007, 221509090008, 221509090009, 221509090010, 221509090011, 221509090012, 221509090013, 221509090014, 221509090015, 221509090016, 221509090017, 221509090018, 221509090019, 221509090020); Lot No. 63171089 (Serial Numbers: 221509240001, 221509240002, 221509240003, 221509240004, 221509240005, 221509240006, 221509240007, 221509240008, 221509240009, 221509240010, 221509240011, 221509240012, 221509240013, 221509240014, 221509240015, 221509240016, 221509240017, 221509240018, 221509240019, 221509240020); Lot No. 63202480 (Serial Numbers: 221510120001, 221510120002, 221510120003, 221510120004, 221510120005); Lot No. 63202478 (Serial Numbers: 221510220001, 221510220002, 221510220003, 221510220004, 221510220005, 221510220006, 221510220007, 221510220008, 221510220009, 221510220010, 221510220011, 221510220012, 221510220013, 221510220014, 221510220015, 221510220016, 221510220017, 221510220018, 221510220019, 221510220020); Lot No. 63234563 (Serial Numbers: 221511230001, 221511230002, 221511230003, 221511230004, 221511230005);
Lot No. 63234565 (Serial Numbers: 221512020001, 221512020002, 221512020003, 221512020004, 221512020005); Lot No. 63212407 (Serial Numbers: 221512070001, 221512070002, 221512070003, 221512070004, 221512070005, 221512070006, 221512070007, 221512070008, 221512070009, 221512070010, 221512070011, 221512070012, 221512070013, 221512070014, 221512070015, 221512070016, 221512070017, 221512070018, 221512070019, 221512070020); Lot No. 63245752 (Serial Numbers: 221512160001, 221512160002, 221512160003, 221512160004, 221512160005, 221512160006, 221512160007); Lot No. 63253256 (Serial Numbers: 221512180001, 221512180002, 221512180003, 221512180004, 221512180005); Lot No. 63247425 (Serial Numbers: 221601060001, 221601060002, 221601060003, 221601060004, 221601060005, 221601060006, 221601060007, 221601060008, 221601060009, 221601060010, 221601060011, 221601060012, 221601060013, 221601060014, 221601060015, 221601060016, 221601060017, 221601060018, 221601060019, 221601060020); Lot No. 63526287 (Serial Numbers: 221612140001, 221612140002, 221612140003, 221612140004, 221612140005, 221612140006, 221612140007, 221612140008, 221612140009, 221612140010, 221612140011, 221612140012, 221612140013, 221612140014, 221612140015, 221612140016, 221612140017, 221612140018, 221612140019, 221612140020); Lot No. 63531691 (Serial Numbers: 221612150001, 221612150002, 221612150003, 221612150004, 221612150005, 221612150006, 221612150007); Lot No. 63526321 (Serial Numbers: 221612170001, 221612170002, 221612170003, 221612170004, 221612170005, 221612170006, 221612170007, 221612170008, 221612170009, 221612170010, 221612170011, 221612170012, 221612170013, 221612170014, 221612170015, 221612170016, 221612170017, 221612170018, 221612170019, 221612170020); Lot No. 63261584 (Serial Numbers: 221601070001, 221601070002, 221601070003, 221601070004, 221601070005); Lot No. 63264374 (Serial Numbers: 221601110001, 221601110002, 221601110003, 221601110004, 221601110005, 221601110006, 221601110007, 221601110008, 221601110009, 221601110010, 221601110011, 221601110012, 221601110013, 221601110014, 221601110015, 221601110016, 221601110017, 221601110018, 221601110019, 221601110020); Lot No. 63265280 (Serial Numbers: 221601140001, 221601140002, 221601140003, 221601140004, 221601140005, 221601140006, 221601140007); Lot No. 63257987 (Serial Numbers: 221601180001, 221601180002, 221601180003, 221601180004, 221601180005, 221601180006, 221601180007, 221601180008, 221601180009, 221601180010, 221601180011, 221601180012, 221601180013, 221601180014, 221601180015, 221601180016, 221601180017, 221601180018, 221601180019, 221601180020); Lot No. 63271202 (Serial Numbers: 221601220001, 221601220002, 221601220003, 221601220004, 221601220005); Lot No. 63258039 (Serial Numbers: 221602060001, 221602060002, 221602060003, 221602060004, 221602060005, 221602060006, 221602060007, 221602060008, 221602060009, 221602060010, 221602060011, 221602060012, 221602060013, 221602060014, 221602060015, 221602060016, 221602060017, 221602060018, 221602060019, 221602060020); Lot No. 63278388 (Serial Numbers: 221602150001, 221602150002, 221602150003, 221602150004, 221602150005, 221602150006, 221602150007, 221602150008, 221602150009, 221602150010, 221602150011, 221602150012, 221602150013, 221602150014, 221602150015, 221602150016, 221602150017, 221602150018, 221602150019, 221602150020); Lot No. 63278436 (Serial Numbers: 221602240001, 221602240002, 221602240003, 221602240004, 221602240005, 221602240006, 221602240007, 221602240008, 221602240009, 221602240010, 221602240011, 221602240012, 221602240013, 221602240014, 221602240015, 221602240016, 221602240017, 221602240018, 221602240019, 221602240020); Lot No. 63314121 (Serial Numbers: 221603090001, 221603090002, 221603090003); Lot No. 63314118 (Serial Numbers: 221603110001, 221603110002, 221603110003); Lot No. 63324625 (Serial Numbers: 221603220001, 221603220002, 221603220003); Lot No. 63336959 (Serial Numbers: 221603300001, 221603300002, 221603300003, 221603300004, 221603300005, 221603300006, 221603300007, 221603300008, 221603300009, 221603300010); Lot No. 63314042 (Serial Numbers: 221603300011, 221603300012, 221603300013, 221603300014, 221603300015, 221603300016, 221603300017, 221603300018, 221603300019, 221603300020, 221603300021, 221603300022, 221603300023, 221603300024, 221603300025, 221603300026, 221603300027, 221603300028, 221603300029, 221603300030); Lot No. 63341344 (Serial Numbers: 221604040001, 221604040002, 221604040003, 221604040004, 221604040005); Lot No. 63341346 (Serial Numbers: 221604050001, 221604050002, 221604050003, 221604050004, 221604050005, 221604050006, 221604050007, 221604050008, 221604050009, 221604050010); Lot No. 63314117 (Serial Numbers: 221604220001, 221604220002, 221604220003, 221604220004, 221604220005, 221604220006, 221604220007, 221604220008, 221604220009, 221604220010, 221604220011, 221604220012, 221604220013, 221604220014, 221604220015, 221604220016, 221604220017, 221604220018, 221604220019, 221604220020); Lot No. 63352349 (Serial Numbers: 221604250001, 221604250002, 221604250003); Lot No. 63339440 (Serial Numbers: 221605020001, 221605020002, 221605020003, 221605020004, 221605020005, 221605020006, 221605020007, 221605020008, 221605020009, 221605020010, 221605020011, 221605020012, 221605020013, 221605020014, 221605020015, 221605020016, 221605020017, 221605020018, 221605020019, 221605020020); Lot No. 63343366 (Serial Numbers: 221605060001, 221605060002, 221605060003, 221605060004, 221605060005, 221605060006, 221605060007, 221605060008, 221605060009, 221605060010, 221605060011, 221605060012, 221605060013, 221605060014, 221605060015, 221605060016, 221605060017, 221605060018, 221605060019, 221605060020, 221605060021, 221605060022); Lot No. 63352347 (Serial Numbers: 221605110001, 221605110002, 221605110003, 221605110004, 221605110005); Lot No. 63339497 (Serial Numbers: 221605120001, 221605120002, 221605120003, 221605120004, 221605120005, 221605120006, 221605120007, 221605120008, 221605120009, 221605120010, 221605120011, 221605120012, 221605120013, 221605120014, 221605120015, 221605120016, 221605120017, 221605120018, 221605120019, 221605120020); Lot No. 63378855 (Serial Numbers: 221606020001, 221606020002, 221606020003); Lot No. 63369605 (Serial Numbers: 221606110001, 221606110002, 221606110003, 221606110004, 221606110005, 221606110006, 221606110007, 221606110008, 221606110009, 221606110010, 221606110011, 221606110012, 221606110013, 221606110014, 221606110015, 221606110016, 221606110017, 221606110018, 221606110019, 221606110020); Lot No. 63403010 (Serial Numbers: 221606110021, 221606110022, 221606110023); Lot No. 63403012 (Serial Numbers: 221606150001, 221606150002, 221606150003, 221606150004, 221606150005, 221606150006, 221606150007, 221606150008); Lot No. 63403026 (Serial Numbers: 221606210001, 221606210002, 221606210003, 221606210004, 221606210005, 221606210006, 221606210007, 221606210008, 221606210009, 221606210010, 221606210011, 221606210012, 221606210013, 221606210014, 221606210015, 221606210016, 221606210017, 221606210018, 221606210019, 221606210020); Lot No. 63403009 (Serial Numbers: 221606220001, 221606220002, 221606220003); Lot No. 63403024 (Serial Numbers: 221606230001, 221606230002, 221606230003, 221606230004, 221606230005, 221606230006, 221606230007, 221606230008, 221606230009, 221606230010, 221606230011, 221606230012, 221606230013, 221606230014, 221606230015, 221606230016, 221606230017, 221606230018, 221606230019, 221606230020); Lot No. 63403025 (Serial Numbers: 221606240001, 221606240002, 221606240003, 221606240004, 221606240005, 221606240006, 221606240007, 221606240008, 221606240009, 221606240010); Lot No. 63403017 (Serial Numbers: 221606240011, 221606240012, 221606240013, 221606240014, 221606240015); Lot No. 63394148 (Serial Numbers: 221606290001, 221606290002, 221606290003, 221606290004, 221606290005, 221606290006, 221606290007, 221606290008, 221606290009, 221606290010, 221606290011, 221606290012, 221606290013, 221606290014, 221606290015, 221606290016, 221606290017, 221606290018, 221606290019, 221606290020); Lot No. 63410582 (Serial Numbers: 221607050001, 221607050002, 221607050003, 221607050004, 221607050005, 221607050006, 221607050007, 221607050008, 221607050009, 221607050010, 221607050011, 221607050012, 221607050013, 221607050014, 221607050015); Lot No. 63422767 (Serial Numbers: 221607080001, 221607080002, 221607080003, 221607080004, 221607080005, 221607080006, 221607080007, 221607080008, 221607080009, 221607080010, 221607080011, 221607080012, 221607080013, 221607080014, 221607080015, 221607080016, 221607080017, 221607080018, 221607080019, 221607080020); Lot No. 63410583 (Serial Numbers: 221607210001, 221607210002, 221607210003, 221607210004, 221607210005, 221607210006, 221607210007, 221607210008, 221607210009, 221607210010); Lot No. 63410585 (Serial Numbers: 221607220001, 221607220002, 221607220003, 221607220004, 221607220005, 221607220006, 221607220007, 221607220008, 221607220009, 221607220010); Lot No. 63394176 (Serial Numbers: 221607260001, 221607260002, 221607260003, 221607260004, 221607260005, 221607260006, 221607260007, 221607260008, 221607260009, 221607260010, 221607260011, 221607260012, 221607260013, 221607260014, 221607260015, 221607260016, 221607260017, 221607260018, 221607260019, 221607260020); Lot No. 63430499 (Serial Numbers: 221608080001, 221608080002, 221608080003, 221608080004, 221608080005, 221608080006, 221608080007, 221608080008, 221608080009, 221608080010, 221608080011, 221608080012, 221608080013, 221608080014, 221608080015, 221608080016, 221608080017, 221608080018, 221608080019, 221608080020); Lot No. 63437710 (Serial Numbers: 221608120001, 221608120002, 221608120003, 221608120004, 221608120005); Lot No. 63436338 (Serial Numbers: 221608160001, 221608160002, 221608160003, 221608160004, 221608160005); Lot No. 63445193 (Serial Numbers: 221608170001, 221608170002, 221608170003, 221608170004, 221608170005, 221608170006, 221608170007, 221608170008, 221608170009, 221608170010, 221608170011, 221608170012, 221608170013, 221608170014, 221608170015, 221608170016, 221608170017, 221608170018, 221608170019, 221608170020); Lot No. 63430540 (Serial Numbers: 221608190001, 221608190002, 221608190003, 221608190004, 221608190005, 221608190006, 221608190007, 221608190008, 221608190009, 221608190010, 221608190011, 221608190012, 221608190013, 221608190014, 221608190015, 221608190016, 221608190017, 221608190018, 221608190019, 221608190020); Lot No. 63445194 (Serial Numbers: 221608230001, 221608230002, 221608230003, 221608230004, 221608230005, 221608230006, 221608230007, 221608230008, 221608230009, 221608230010, 221608230011, 221608230012, 221608230013, 221608230014, 221608230015, 221608230016, 221608230017, 221608230018, 221608230019, 221608230020); Lot No. 63455066 (Serial Numbers: 221609010001, 221609010002, 221609010003, 221609010004, 221609010005, 221609010006, 221609010007, 221609010008, 221609010009, 221609010010, 221609010011, 221609010012, 221609010013, 221609010014, 221609010015, 221609010016, 221609010017, 221609010018, 221609010019, 221609010020); Lot No. 63474281 (Serial Numbers: 221609080001, 221609080002, 221609080003, 221609080004, 221609080005, 221609080006, 221609080007, 221609080008, 221609080009, 221609080010, 221609080011, 221609080012, 221609080013, 221609080014, 221609080015, 221609080016, 221609080017, 221609080018, 221609080019, 221609080020); Lot No. 63474283 (Serial Numbers: 221609080021, 221609080022, 221609080023, 221609080024, 221609080025, 221609080026); Lot No. 63474284 (Serial Numbers: 221609150001, 221609150002, 221609150003, 221609150004, 221609150005, 221609150006, 221609150007, 221609150008); Lot No. 63474287 (Serial Numbers: 221609210001, 221609210002, 221609210003, 221609210004, 221609210005, 221609210006, 221609210007, 221609210008, 221609210009, 221609210010); Lot No. 63474285 (Serial Numbers: 221610010001, 221610010002, 221610010003, 221610010004, 221610010005, 221610010006, 221610010007, 221610010008, 221610010009, 221610010010, 221610010011, 221610010012, 221610010013, 221610010014, 221610010015); Lot No. 63455098 (Serial Numbers: 221610050001, 221610050002, 221610050003, 221610050004, 221610050005, 221610050006, 221610050007, 221610050008, 221610050009, 221610050010, 221610050011, 221610050012, 221610050013, 221610050014, 221610050015, 221610050016, 221610050017, 221610050018, 221610050019, 221610050020); Lot No. 63474282 (Serial Numbers: 221610110001, 221610110002, 221610110003, 221610110004, 221610110005, 221610110006, 221610110007, 221610110008, 221610110009, 221610110010, 221610110011, 221610110012, 221610110013, 221610110014, 221610110015, 221610110016, 221610110017, 221610110018, 221610110019, 221610110020); Lot No. 63484747 (Serial Numbers: 221610140001, 221610140002, 221610140003, 221610140004, 221610140005); Lot No. 63482404 (Serial Numbers: 221610180001, 221610180002, 221610180003, 221610180004, 221610180005, 221610180006, 221610180007, 221610180008, 221610180009, 221610180010, 221610180011, 221610180012, 221610180013, 221610180014, 221610180015, 221610180016, 221610180017, 221610180018, 221610180019, 221610180020); Lot No. 63482405 (Serial Numbers: 221610200001, 221610200002, 221610200003, 221610200004, 221610200005, 221610200006, 221610200007, 221610200008, 221610200009, 221610200010, 221610200011, 221610200012, 221610200013, 221610200014, 221610200015, 221610200016, 221610200017, 221610200018, 221610200019, 221610200020); Lot No. 63482408 (Serial Numbers: 221611210001, 221611210002, 221611210003, 221611210004, 221611210005, 221611210006, 221611210007, 221611210008, 221611210009, 221611210010, 221611210011,221611210012, 221611210013, 221611210014, 221611210015, 221611210016, 221611210017, 221611210018, 221611210019, 221611210020); Lot No. 63501470 (Serial Numbers: 221611210021, 221611210022, 221611210023, 221611210024, 221611210025); Lot No. 63501471 (Serial Numbers: 221611210021, 221611210022, 221611210023, 221611210024, 221611210025); Lot No. 63501459 (Serial Numbers: 221611280001, 221611280002, 221611280003, 221611280004, 221611280005, 221611280006, 221611280007, 221611280008, 221611280009, 221611280010, 221611280011, 221611280012, 221611280013, 221611280014, 221611280015, 221611280016, 221611280017, 221611280018, 221611280019, 221611280020); Lot No. 63501486 (Serial Numbers: 221612020001, 221612020002, 221612020003, 221612020004, 221612020005, 221612020006, 221612020007, 221612020008, 221612020009, 221612020010, 221612020011, 221612020012, 221612020013, 221612020014, 221612020015, 221612020016, 221612020017, 221612020018, 221612020019, 221612020020); Lot No. 63501469 (Serial Numbers: 221612030001, 221612030002, 221612030003, 221612030004, 221612030005, 221612030006, 221612030007, 221612030008, 221612030009, 221612030010, 221612030011, 221612030012, 221612030013, 221612030014, 221612030015, 221612030016, 221612030017, 221612030018); Lot No. 63531693 (Serial Numbers: 221612050001, 221612050002, 221612050003, 221612050004, 221612050005); Lot No. 63501509 (Serial Numbers: 221612070001, 221612070002, 221612070003, 221612070004, 221612070005, 221612070006, 221612070007, 221612070008, 221612070009, 221612070010, 221612070011, 221612070012, 221612070013, 221612070014, 221612070015, 221612070016, 221612070017, 221612070018, 221612070019, 221612070020); Lot No. 63501547 (Serial Numbers: 221612100001, 221612100002, 221612100003, 221612100004, 221612100005, 221612100006, 221612100007, 221612100008, 221612100009, 221612100010, 221612100011, 221612100012, 221612100013, 221612100014, 221612100015, 221612100016, 221612100017, 221612100018, 221612100019, 221612100020); Lot No. 63501493 (Serial Numbers: 221612130001, 221612130002, 221612130003, 221612130004, 221612130005, 221612130006, 221612130007, 221612130008) |
Recalling Firm/ Manufacturer |
Zimmer Surgical Inc 200 W Ohio Ave Dover OH 44622-9642
|
For Additional Information Contact |
330-343-8801
|
Manufacturer Reason for Recall |
Updated on-device label and 3 pages of an updated Operator/Service Manual. Also, decrease from 3.8 meters (m) to 1.0 separation distance between the device and RF communication equipment.
|
FDA Determined Cause 2 |
Use error |
Action |
The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE CORRECTIONS" letter dated 7/27/2017 to its customers via email or mail. The letter describes the product, problem and actions to be taken. Instructions for distributors include ensure all affected personnel are aware of the recall, confirm receipt of new device label and updated Operator/Service Manual, place the label on the device in the location specified in the notice, replace the 3 pages of the Operator/Service Manual with the updated pages, review the hospital list of customers that accompanied the notice and return contact information for any additional hospitals product was distributed to.
Instructions for Risk Managers at hospitals include ensure all affected personnel are aware of the recall, confirm receipt of new device label and updated Operator/Service Manual, place the label on the device in the location specified in the notice, replace the 3 pages of the Operator/Service Manual with the updated pages, and complete and return the Attachment 1 form via email to corporatequality.postmarket@zimmerbiomet.com or ZimmerBiomet8287@stericycle.com or fax to 866-338-2611.
If after reviewing this notification you have further questions or concerns please call +1(330) 364-0989 between 8:00 am and 5:00 pm EST, Monday through Friday.Calls received outside of the call center operating hours will receive a prompt to record a voicemail or be transferred to an on-call representative in the case of an emergency. |
Quantity in Commerce |
8,231 devices total |
Distribution |
Worldwide Distribution: US (nationwide) to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Guam; and countries to Australia, Brazil, Canada, Chile, China, Denmark, Dominican, El Salvador, England, Finland, France, Hong Kong, India, Italy, Japan, Malaysia, Mexico, Netherlands, Panama, Singapore, Sweden, Taiwan, and Thailand.. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|