|
Class 2 Device Recall BECKMAN COULTER iChemVELOCITY |
|
Date Initiated by Firm |
March 02, 2018 |
Create Date |
May 04, 2018 |
Recall Status1 |
Terminated 3 on July 27, 2023 |
Recall Number |
Z-1643-2018 |
Recall Event ID |
79575 |
510(K)Number |
K101852
|
Product Classification |
Automated urinalysis system - Product Code KQO
|
Product |
BECKMAN COULTER iChemVELOCITY, Urine Chemistry System, Catalog Numbers: 800-3530, 800-7167, 800-7719, 800-7720, 800- 7106, 800-3049, 800-3050, 800-3079, 800-3080, 800-7190, 800-7713, 800-7714, 800-7715, 700- 7176-001, 700-7177-001
Product Usage: The iChemVELOCITY automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi- quantitative results for glucose, blood, leukocyte esterade, bilirubin, urobilinogen, pH, protein, ketones, and ascorbic acid; and qualitative results for nitrites, color, and clarity. iChemVELOCITY strips are intended for use only with the iChemVELOCITY analyzer. In particular they are not intended for visual reading. The iChemVELOCITY is not intended to be used as a Point of Care (POC) analyzer. The measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver finction. Tests performed using the iChemVELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.
|
Code Information |
All Fielded Serial Numbers with North American Configuration settings |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
For Additional Information Contact |
469-961-4941
|
Manufacturer Reason for Recall |
Beckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers.
1. The installation of International settings can result in the generation of erroneous, false low results for some of the analytes at some concentrations.
2. The installation of colors different from those stated in the Instructions for Use (Colorless, Straw, Yellow, Amber, Red, Blue) for output settings will result in incorrect reporting of colors. For example, if Green is the color choice for the output setting instead of Blue, Green will be reported.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Beckman Coulter sent an Urgent Field Safety Notice letter dated February 22, 2018 to the customers. The letter identified the affected product, problem and actions to be taken. Beckman Coulter will schedule a site visit to verify the reporting units of measurement for the laboratory within 90 days of this recall initiation. For questions call 800-854-3633. |
Quantity in Commerce |
2225 units |
Distribution |
Worldwide Distribution - US Nationwide and the countries of Brazil, Canada, Malaysia, Singapore, Taiwan, Thailand, Turkey |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KQO and Original Applicant = IRIS INTERNATIONAL INC., A DIVISION OF IRIS IRIS I
|
|
|
|