Date Initiated by Firm |
August 28, 2018 |
Create Date |
October 29, 2018 |
Recall Status1 |
Terminated 3 on February 04, 2021 |
Recall Number |
Z-0291-2019 |
Recall Event ID |
81087 |
510(K)Number |
K163656
|
Product Classification |
Appliance, fixation, nail/blade/plate combination, multiple component - Product Code KTT
|
Product |
Hoffman LRF Hexapod Strut, Extra Short 183mm-295mm, Model Number 4935-0-040. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity Correction of segmental bony or soft tissue defects Joint arthrodesis Management of communicated intra-articular fractures of the distal radius Bone transport The Hoffmann LRF System is indicated in adults for: Osteotomy Revision procedure where other treatments or devices have been unsuccessful Bone reconstruction procedures Fusions and replantations of the foot Charcot foot reconstruction Lisfranc dislocations The typical LRF construct features two or more rings bridged by three to four telescopic struts with multiplanar, ball jointed fixation bolts. Once applied, the LRF can be unlocked for gross frame adjustment and fracture reduction. The LRF can be finely adjusted to aid in limb alignment and definitively locked to hold compression and reduction. If deemed appropriate, the LRF Telescopic Struts can provide up to 5mm of controlled dynamization. Original Statement OP-Tech H-ST-1_Rev2_HoffmannLRF_OpTech_2016-10690] |
Code Information |
Lots: L18239 L13690 L22250 L26334 L31272 L35847 L35848 L46136 L46137 L46495 L46497 L50179 L51363 L53107 L53490 L59090 Y02173 Y07632 Y08556 Y08558 Y08557 Y14369 Y14370 Y16719 Y14646 Y14642 |
Recalling Firm/ Manufacturer |
Stryker GmbH Bohnackerweg 1 Selzach Switzerland
|
For Additional Information Contact |
Eric Petschler 201-831-6693
|
Manufacturer Reason for Recall |
The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Urgent Product Recall notification letters dated 8/28/18 were sent to customers. |
Quantity in Commerce |
867 |
Distribution |
US Nationwide in the states of CO, FL, MD, MI, NY, OK, and PA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KTT and Original Applicant = Stryker GmbH
|