|
Class 2 Device Recall Zenith Alpha Abdominal Endovascular Graft |
|
Date Initiated by Firm |
April 26, 2019 |
Date Posted |
June 19, 2019 |
Recall Status1 |
Terminated 3 on November 08, 2019 |
Recall Number |
Z-1844-2019 |
Recall Event ID |
82904 |
Product Classification |
System, endovascular graft, aortic aneurysm treatment - Product Code MIH
|
Product |
Zenith Alpha Abdominal Endovascular Graft, Reference Part Numbers: ZIMB-22-108, ZIMB-22-118, ZIMB-22-128, ZIMB-22-70, ZIMB-22-84, ZIMB-22-98, ZIMB-24-108, ZIMB-24-118, ZIMB-24-128, ZIMB-24-70, ZIMB-24-84, ZIMB-24-98, ZIMB-26-108, ZIMB-26-118, ZIMB-26-128, ZIMB-26-70, ZIMB-26-84, ZIMB-26-98, ZIMB-28-108, ZIMB-28-118, ZIMB-28-128, ZIMB-28-70, ZIMB-28-84, ZIMB-28-98, ZIMB-30-108, ZIMB-30-118, ZIMB-30-128, ZIMB-30-70, ZIMB-30-84, ZIMB-30-98, ZIMB-32-108, ZIMB-32-118, ZIMB-32-128, ZIMB-32-70, ZIMB-32-84, ZIMB-32-98, ZIMB-36-108, ZIMB-36-118, ZIMB-36-128, ZIMB-36-70, ZIMB-36-84, ZIMB-36-98 |
Code Information |
9221784, 9228731, 9232249, 9246529, 9272685, 9272835, 9276029, 9276400, 9303245, 9373623, 9406776, 9466908, 9506559, 9525188, 9557569, 9566160, 9566211, 9575380, 9620378, 9620379, 9620382, 9625916, 9625917, 9636371, 9640584, 9663986 9406777, 9522416, 9566161, 9575382, 9625918, 9656992 9557570, 9566162, 9575383, 9625919, 9625920, 9656978 9221629, 9232247, 9276401, 9303246, 9373624, 9394051, 9403262, 9406778, 9406779, 9406780, 9453299, 9466909, 9466910, 9482147, 9486290, 9486313, 9506616, 9522418, 9522419, 9525189, 9529047, 9534284, 9555442, 9555453, 9555475, 9566214, 9572229, 9575386, 9592482, 9592485, 9594196, 9594205, 9609105, 9617443, 9617445, 9625921, 9625922, 9625923, 9636298, 9636330, 9640585, 9642977, 9650113, 9650114, 9394051X, NS9584273, NS9584274, NS9584275, NS9584276, NS9584277, NS9584278, NS9584279, NS9584280, NS9584281, NS9584282 9214951, 9221781, 9221785, 9221786, 9221787, 9228739, 9232238, 9232239, 9232250, 9234942, 9239302, 9239303, 9246530, 9272665, 9272686, 9272687, 9272836, 9276402, 9276403, 9391303, 9403123, 9403124, 9403125, 9403126, 9403127, 9403128, 9403129, 9403130, 9403131, 9403132, 9403133, 9403134, 9403136, 9403137, 9403138, 9403139, 9404632, 9404633, 9406719, 9406720, 9406721, 9406781, 9406782, 9406783, 9464235, 9522423, 9529049, 9529050, 9529051, 9557572, 9561001, 9566215, 9575388, 9581884, 9594197, 9594206, 9598741, 9602905, 9605694, 9605695, 9625893, 9625896, 9625924, 9625925, 9625926, 9636372, 9640586, 9642978, 9656979, 9656994, 9272665X, 9403130X, 9522423X, NS9239300, NS9239301 9221788, 9289544, 9373625, 9406785, 9419801, 9482149, 9515085, 9557563, 9557573, 9561002, 9566216, 9572230, 9575390, 9594207, 9602906, 9605689, 9605696, 9611316, 9617447, 9620383, 9625897, 9625927, 9625928, 9629883, 9629884, 9640587, 9640588, 9653876, 9656980, 9671098 9228740, 9242449, 9386592, 9386660, 9386661, 9386662, 9466913, 9466914, 9466915, 9605686, 9623874, 9623879, 9625845, 9640513, 9640514, 9671040, 9671042, 9671043, 9671044, 9242450 9228741, 9228742, 9403263, 9419802, 9470240, 9506618, 9515086, 9557574, 9561003, 9566217, 9575392, 9609106, 9623930, 9625929, 9629885, 9640589, 9640590, 9656981, 9660204, 9663989 9373626, 9557564, 9557575, 9569073, 9575332, 9578369, 9629871, 9629886, 9640591, 9656995, 9373626X 8405214, 9221798, 9221799, 9226415, 9226416, 9226417, 9228743, 9232252, 9239304, 9242451, 9248832, 9248833, 9248834, 9276031, 9276032, 9276406, 9276407, 9276408, 9276409, 9303242, 9373627, 9373628, 9409576, 9419022, 9442622, 9464237, 9470243, 9470244, 9470245, 9482150, 9486314, 9486315, 9508711, 9508712, 9508713, 9515058, 9515059, 9515060, 9515088, 9515089, 9515090, 9515091, 9515092, 9522430, 9522431, 9522432, 9522433, 9529053, 9529054, 9557576, 9557577, 9566163, 9569089, 9578373, 9581886, 9584296, 9602907, 9602908, 9602909, 9602910, 9602911, 9617436, 9617449, 9617451, 9617453, 9617455, 9617458, 9620384, 9625898, 9629838, 9629839, 9629872, 9629873, 9629874, 9629875, 9629876, 9629887, 9629888, 9636331, 9636332, 9636333, NS9584286, NS9584287, NS9584288, NS9584289, NS9584290, NS9584291, NS9584292, NS9584293, NS9584294, NS9584295 9226360, 9226418, 9226420, 9226421, 9226422, 9226424, 9226425, 9226426, 9226427, 9228744, 9228745, 9228827, 9232253, 9239284, 9242433, 9242452, 9242453, 9242454, 9242455, 9242456, 9246532, 9246533, 9246534, 9246535, 9246536, 9386666, 9391304, 9391324, 9391325, 9391326, 9391327, 9391454, 9391455, 9391456, 9394052, 9394053, 9403264, 9403265, 9403266, 9419025, 9419026, 9419027, 9456698, 9464238, 9470209, 9470211, 9470246, 9470247, 9470248, 9474588, 9474589, 9474590, 9478694, 9478695, 9482116, 9486316, 9486324, 9506565, 9506566, 9506567, 9506568, 9508715, 9508716, 9508717, 9508718, 9508719, 9508720, 9508721, 9508722, 9512198, 9517953, 9517954, 9525193, 9525194, 9529058, 9529060, 9561004, 9569074, 9569095, 9575333, 9578381, 9598744, 9598745, 9598747, 9602912, 9602913, 9605697, 9623931, 9625899, 9629877, 9629889, 9640556, 9640557, 9640592, 9640593, 9640594, 9640595, 9640596, 9640597, 9640598, 9640599, 9642979, 9642980, 9647086, 9647087, 9650115, 9656963, 9656982, 9228745X, 9391324X, 9470248X, NS9584297, NS9584298, NS9584299, NS9584300, NS9584301, NS9584302, NS9584303, NS9584304, NS9584305, NS9584306 9228746, 9228748, 9232388, 9246537, 9246538, 9246539, 9246540, 9300906, 9303247, 9303248, 9303249, 9303281, 9370215, 9373476, 9373636, 9373638, 9391459, 9391460, 9391461, 9391462, 9403267, 9419029, 9419030, 9419031, 9419032, 9419033, 9419804, 9474595, 9486326, 9486327, 9486328, 9515098, 9515099, 9515100, 9515101, 9522445, 9522446, 9522447, 9522448, 9522449, 9529061, 9529062, 9529064, 9561005, 9566164, 9578386, 9581891, 9581892, 9587377, 9587406, 9647037, 9650044, 9650045, 9373637X 9217948, 9217949, 9217950, 9226365, 9228749, 9228750, 9246541, 9246542, 9246543, 9248835, 9248836, 9373644, 9391463, 9391464, 9394054, 9394055, 9403268, 9403269, 9403270, 9474596, 9525196, 9529066, 9529067, 9529068, 9534296, 9534297, 9557578, 9602871, 9650039, 9650040 9232389, 9246544, 9525197, 9525198, 9566166, 9578388, 9587407, 9625900, 9629891, 9640558, 9642981 9228753, 9232390, 9248837, 9396561, 9525171, 9525200, 9525202, 9557579, 9566167, 9578390, 9594198, 9594209, 9594210, 9629892, 9629893, 9640600, 9642982, 9642983, 9660205, 9664012 9226366, 9228755, 9228756, 9246545, 9246546, 9246547, 9246548, 9272837, 9276416, 9276417, 9300916, 9370209, 9370220, 9370221, 9370222, 9391465, 9391466, 9391467, 9464239, 9474597, 9474598, 9486330, 9506569, 9508723, 9508724, 9508725, 9529069, 9529070, 9566168, 9569103, 9569104, 9569105, 9578391, 9578392, 9581895, 9581896, 9581897, 9584308, 9584309, 9587408, 9598750, 9598751, 9602914, 9611317, 9617464, 9617466, 9617468, 9623932, 9625894, 9625901, 9629894, 9629895, 9629896, 9629897, 9629898, 9640559, 9640560, 9642967, 9642984, 9642985, 9642986, 9642987, 9642988, 9650116, 9653828, 9653858, 9653877, 9653878, 9653879, 9653880, 9656983 9217952, 9228734, 9228828, 9228829, 9228830, 9228831, 9228832, 9228833, 9228834, 9228835, 9228836, 9228837, 9232391, 9232392, 9234943, 9246549, 9246550, 9248838, 9279007, 9300908, 9300913, 9300917, 9303251, 9303283, 9303284, 9370225, 9380032, 9391468, 9391469, 9394056, 9394185, 9394186, 9394187, 9403271, 9403272, 9403273, 9403274, 9403275, 9409622, 9409623, 9419805, 9419807, 9456708, 9456713, 9470212, 9474599, 9474748, 9474749, 9474750, 9474751, 9474752, 9474753, 9506570, 9506571, 9506572, 9508726, 9508727, 9508728, 9508729, 9508730, 9508731, 9508732, 9508733, 9515108, 9515109, 9515110, 9515111, 9515112, 9515113, 9515114, 9525207, 9525208, 9525210, 9557580, 9620332, 9640511, 9642954, 9642955, 9650041, 9671028, 9394185X 9217983, 9217984, 9228735, 9228838, 9228839, 9228840, 9228841, 9228842, 9228843, 9232393, 9232394, 9248839, 9248840, 9272838, 9383423, 9394035, 9394190, 9394191, 9394192, 9394193, 9394194, 9394195, 9396562, 9404637, 9404638, 9404639, 9404640, 9404641, 9447170, 9464240, 9470213, 9470214, 9470215, 9470216, 9474754, 9474755, 9474756, 9474757, 9474758, 9474759, 9482118, 9508734, 9508735, 9515115, 9515116, 9515117, 9515118, 9515119, 9525214, 9525216, 9525218, 9525220, 9525221, 9529071, 9529072, 9529073, 9529074, 9529075, 9557581, 9566170, 9569110, 9572234, 9578357, 9578394, 9578395, 9578396, 9578397, 9578398, 9578399, 9578400, 9581898, 9581899, 9581900, 9587409, 9594199, 9594211, 9594212, 9594213, 9594214, 9594215, 9594216, 9598726, 9598727, 9598752, 9602915, 9602916, 9602917, 9602918, 9602919, 9602920, 9602921, 9609107, 9609108, 9611318, 9617479, 9617481, 9617483, 9617485, 9617488, 9625902, 9629899, 9629900, 9629901, 9629902, 9629903, 9629904, 9629905, 9629906, 9633300, 9633301, 9636334, 9636335, 9640561, 9642989, 9642990, 9642991, 9642992, 9642993, 9647088, 9647089, 9647090, 9647091, 9647092, 9650117, 9650118, 9650119, 9653859, 9653874, 9653881, 9656984, 9671059, 9671101, 9671102, 9671103, 9671104 9370226, 9376518, 9386593, 9386594, 9394036, 9394037, 9394197, 9394198, 9394199, 9394200, 9394201, 9394202, 9470217, 9470218, 9478696, 9478697, 9486297, 9508741, 9512199, 9515122, 9515123, 9515124, 9522377, 9525227, 9525229, 9525231, 9525233, 9529077, 9529078, 9529079, 9561008, 9566171, 9569113, 9578405, 9587415, 9594217, 9602922, 9602923, 9602924, 9605698, 9605699, 9605701, 9605702, 9609109, 9611319, 9617490, 9620385, 9620386, 9623933, 9629878, 9629907, 9636336, 9636337, 9636338, 9640562, 9640563, 9642968, 9642994, 9642995, 9647093, 9650120, 9660206, 9660207, 9660208, 9660209, 9660210 9279021, 9279022, 9279023, 9279024, 9394203, 9522378, 9529009, 9560984, 9561009, 9566172, 9569116, 9575338, 9578408, 9587417, 9594218, 9605703, 9609099, 9609110, 9611320, 9620387, 9620388, 9629908, 9636339, 9640564, 9642969, 9642996, 9642997, 9650121, 9650122, 9653882, 9656985, 9671105, 9671106, 9370227, 9370227X 9279025, 9370228, 9370229, 9400627, 9508744, 9515126, 9515127, 9534247, 9557565, 9560985, 9566173, 9575339, 9578409, 9578411, 9581849, 9587419, 9620389, 9633333, 9642970, 9647072, 9647095, 9650123, 9650124, 9656964, 9330485, 9330485X 9279026, 9279027, 9386653, 9478700, 9482119, 9486299, 9508745, 9508746, 9508747, 9515128, 9515129, 9515130, 9515131, 9522379, 9525248, 9534303, 9534304, 9566174, 9566175, 9569119, 9569122, 9569123, 9578412, 9578413, 9578414, 9578415, 9578416, 9578417, 9578418, 9587421, 9587422, 9594219, 9598729, 9602925, 9602926, 9602927, 9602928, 9602930, 9605704, 9623925, 9623926, 9623934, 9623935, 9623936, 9623937, 9633334, 9636299, 9636340, 9636341, 9636342, 9636343, 9640565, 9640566, 9647073, 9647074, 9647097, 9647098, 9647099, 9647100, 9656986, 9515130X 9276033, 9276034, 9279028, 9279110, 9279118, 9279119, 9279121, 9279122, 9279123, 9279124, 9279125, 9279126, 9279127, 9365243, 9365244, 9365245, 9370230, 9370231, 9370232, 9370238, 9370239, 9370240, 9370241, 9370242, 9386654, 9461549, 9461553, 9464246, 9466882, 9470221, 9470222, 9474577, 9474578, 9478701, 9478702, 9478703, 9478704, 9478705, 9478706, 9478707, 9478708, 9478709, 9482120, 9482121, 9482122, 9482124, 9482125, 9482126, 9482127, 9486300, 9486301, 9486302, 9506575, 9506576, 9506577, 9506578, 9506579, 9506581, 9508748, 9508749, 9508750, 9512178, 9515062, 9515063, 9515132, 9517941, 9517965, 9517966, 9517968, 9517969, 9517970, 9517974, 9561011, 9566155, 9566176, 9566177, 9569135, 9569136, 9572239, 9572240, 9575340, 9581850, 9581851, 9581852, 9584258, 9584260, 9587392, 9587393, 9587394, 9587395, 9587424, 9587425, 9587426, 9587427, 9587428, 9587429, 9587430, 9587431, 9592483, 9594200, 9594220, 9594221, 9598730, 9602931, 9602932, 9605706, 9605707, 9605708, 9605709, 9605710, 9605711, 9609111, 9609112, 9620381, 9620391, 9620392, 9620393, 9620394, 9625904, 9629879, 9633299, 9633302, 9633303, 9633304, 9633305, 9633306, 9633307, 9633308, 9633309, 9633335, 9636344, 9636345, 9636346, 9636347, 9640567, 9640568, 9640569, 9640570, 9640571, 9647075, 9647101, 9647102, 9647103, 9647104, 9647105, 9647106, 9647107, 9647108, 9650089, 9650090, 9653860, 9653884, 9653885, 9653886, 9653887, 9653888, 9653889, 9653890, 9653891, 9276034X, 9370230X 9272844, 9272846, 9272847, 9272848, 9272849, 9272850, 9272851, 9272852, 9276035, 9279129, 9279130, 9283230, 9370248, 9370249, 9370250, 9386595, 9386655, 9394038, 9396612, 9396613, 9396614, 9396615, 9396616, 9426564, 9426565, 9426567, 9426568, 9426571, 9466883, 9466884, 9470223, 9470237, 9478710, 9478711, 9478712, 9482111, 9482112, 9482115, 9482129, 9482130, 9482131, 9486303, 9506585, 9512179, 9512180, 9512181, 9512182, 9512183, 9512184, 9512185, 9512186, 9515064, 9517977, 9517978, 9517979, 9517981, 9517984, 9517985, 9534268, 9534269, 9534270, 9534316, 9534317, 9534318, 9534319, 9534320, 9534321, 9534322, 9534323, 9534324, 9534325, 9534326, 9569137, 9572241, 9572242, 9572243, 9572244, 9572245, 9572246, 9572247, 9572248, 9575341, 9575342, 9581853, 9605687, 9605688, 9633257, 9633258, 9642956, 9647036, 9660191, 9506585X 9276036, 9276037, 9276038, 9276039, 9283232, 9283233, 9283234, 9283235, 9391306, 9396541, 9396620, 9426572, 9464251, 9466885, 9466886, 9466887, 9474579, 9478713, 9478714, 9482132, 9512187, 9515065, 9517988, 9517989, 9517990, 9517991, 9529012, 9529014, 9534329, 9561012, 9566178, 9572252, 9572253, 9572254, 9575343, 9575344, 9584261, 9594223, 9594224, 9594225, 9594226, 9605691, 9605712, 9605713, 9605714, 9605715, 9609113, 9620395, 9620396, 9623938, 9625895, 9625905, 9636348, 9636349, 9636350, 9640572, 9640573, 9640574, 9640575, 9647076, 9647109, 9647110, 9647111, 9647112, 9647113, 9650091, 9650092, 9650093, 9653861, 9653892, 9660211, 9664015 9400629, 9400630, 9486304, 9515066, 9517943, 9517994, 9517995, 9517996, 9534334, 9534335, 9534336, 9561013, 9572255, 9584262, 9584263, 9594201, 9594227, 9602901, 9605716, 9636300, 9636351, 9636352, 9647043, 9647077, 9647078, 9647114, 9647115, 9660201, 9660212, 9660213 9276040, 9400632, 9426580, 9517944, 9517998, 9557566, 9560987, 9566179, 9594229, 9605717, 9609114, 9611322, 9617438, 9620397, 9642971, 9653893, 9656987, 9656988, 9656993, 9656996 9276041, 9283222, 9283237, 9283238, 9365246, 9386600, 9386601, 9426581, 9426582, 9466889, 9478715, 9478716, 9529018, 9557567, 9560988, 9561014, 9572261, 9575346, 9594230, 9594231, 9609115, 9611323, 9625906, 9625907, 9633310, 9636353, 9647116, 9647117, 9653862, 9653863, 9653894 9276021, 9279005, 9279006, 9365247, 9365248, 9365249, 9365250, 9365251, 9365252, 9386602, 9386657, 9391309, 9391310, 9396543, 9396544, 9400672, 9426584, 9426585, 9464257, 9470225, 9478717, 9482133, 9482134, 9515067, 9518002, 9518003, 9518004, 9522380, 9529019, 9529020, 9529021, 9534274, 9566180, 9566181, 9569080, 9572262, 9572263, 9572264, 9572265, 9572266, 9572267, 9572268, 9581857, 9581858, 9581859, 9581860, 9581861, 9581862, 9594233, 9594234, 9594235, 9594236, 9594237, 9594238, 9594239, 9594240, 9609116, 9609117, 9609118, 9609119, 9609120, 9609121, 9609122, 9611324, 9611325, 9611326, 9620398, 9620399, 9620400, 9620401, 9620402, 9620403, 9620404, 9623939, 9633311, 9636354, 9636355, 9636356, 9636357, 9640576, 9640577, 9647079, 9647118, 9647119, 9647120, 9647121, 9650094, 9650095, 9650096, 9656989, 9660214, 9660215, 9664020 9283224, 9283239, 9283240, 9365253, 9391314, 9396546, 9464258, 9474583, 9478691, 9478718, 9478719, 9478720, 9478721, 9486305, 9486306, 9512191, 9512192, 9515068, 9515069, 9515070, 9515071, 9517945, 9518013, 9518015, 9522381, 9522382, 9522383, 9522384, 9529022, 9529023, 9529024, 9529025, 9529026, 9529027, 9555457, 9555458, 9555459, 9555460, 9555461, 9566184, 9572271, 9572272, 9572273, 9572274, 9572275, 9572276, 9575347, 9578358, 9581846, 9581863, 9581864, 9581865, 9581866, 9581867, 9581868, 9594242, 9594243, 9594244, 9594245, 9594246, 9609100, 9609123, 9609124, 9620405, 9620406, 9620407, 9620408, 9620409, 9620410, 9629880, 9633312, 9633313, 9633314, 9633315, 9633316, 9633317, 9636284, 9636358, 9640522, 9640578, 9640579, 9640580, 9647080, 9647081, 9650097, 9650098, 9650099, 9650100, 9650101, 9650108, 9653864, 9653899, 9653900, 9653901, 9663993, 9663994, 9663995, 9664021, 9664022, 9664023, 9664024, 9664025 9386604, 9386605, 9478722, 9478723, 9482135, 9486308, 9512193, 9515073, 9522385, 9522386, 9525174, 9529028, 9561015, 9561016, 9566186, 9572281, 9575348, 9578359, 9581869, 9581870, 9584267, 9594202, 9594249, 9594250, 9609101, 9609125, 9609126, 9609127, 9611327, 9620411, 9620412, 9386604X, 9478723X 9478724, 9478725, 9522388, 9557568, 9560989, 9561017, 9566187, 9569086, 9572283, 9572284, 9575349, 9578360, 9598725, 9602900, 9602902, 9602903, 9611328, 9611329, 9611330, 9620413, 9647082, 9647083, 9650056, 9650102, 9650103, 9656968, 9656969, 9671109, 9671112 9412948, 9466894, 9522390, 9560991, 9566188, 9569087, 9611331, 9620414, 9633318, 9650104, 9650105, 9650109, 9656970, 9656971, 9671110, 9671111 9276043, 9276044, 9515075, 9522391, 9560992, 9561020, 9575350, 9611315, 9611332, 9611333, 9620415, 9625908, 9625909, 9636359, 9636360, 9650110, 9656972 9276023, 9276045, 9276298, 9276299, 9276300, 9276301, 9361560, 9386610, 9386611, 9400693, 9464262, 9464263, 9592449, 9592451, 9592452, 9592453, 9592454, 9592455, 9592456, 9592457, 9592458, 9592459, 9592460, 9592461, 9598718, 9598719, 9598720, 9598721, 9598722, 9605683, 9605684, 9605685, 9609072, 9609073, 9609074, 9609075, 9623875, 9623876, 9623877, 9623878, 9625846, 9633256, 9640512, 9671029, 9671030, 9671032, 9671033, 9386610X, 9386613 9276302, 9276303, 9276304, 9276305, 9276306, 9276307, 9365254, 9370212, 9386614, 9386615, 9386616, 9386642, 9386643, 9386645, 9415644, 9461576, 9478729, 9478730, 9486310, 9486311, 9515076, 9515078, 9515079, 9515080, 9522394, 9522395, 9522396, 9522397, 9522398, 9525176, 9525177, 9529029, 9529031, 9529032, 9529033, 9529044, 9560994, 9560995, 9561022, 9566189, 9566191, 9566192, 9566194, 9575352, 9575353, 9575354, 9575355, 9575356, 9575357, 9584268, 9598734, 9598735, 9611334, 9611335, 9611336, 9611337, 9611338, 9620416, 9623927, 9623941, 9623942, 9623943, 9623944, 9623945, 9623946, 9625910, 9629881, 9633319, 9633320, 9633321, 9633322, 9633323, 9633324, 9633325, 9636361, 9636362, 9642972, 9642973, 9642974, 9653865, 9656973, 9656974, 9656975, 9656976, 9656977, 9656991, 9664000 9376505, 9376506, 9376507, 9376580, 9376581, 9383408, 9383409, 9461581, 9522401, 9560996, 9566198, 9575358, 9575359, 9575360, 9575361, 9581871, 9581872, 9592484, 9594203, 9611340, 9611342, 9611344, 9611346, 9633326, 9633327 9376508, 9376509, 9376582, 9376583, 9560997, 9566199, 9609087, 9609102, 9609103, 9611347, 9611348, 9620380, 9623928, 9623947, 9623948, 9636363, 9636364, 9647084, 9650057, 9650058, 9653829, 9653866, 9653867, 9660202, 9664002, 9664003, 9664029, 9664030 9560998, 9566200, 9609104, 9611349, 9611350, 9623949, 9636365, 9650059, 9653868, 9664004, 9664031, 9664782 9555469, 9566201, 9569088, 9575365, 9575366, 9581873, 9623950, 9629882, 9633328, 9636366, 9642975, 9647085, 9650042, 9650046, 9650047, 9650051, 9650060, 9653869, 9653870, 9663988, 9664005, 9664032, 9664033 9365288, 9365289, 9376585, 9376586, 9376587, 9466898, 9560999, 9566206, 9575371, 9578361, 9581874, 9581875, 9598738, 9611351, 9611352, 9611353, 9611354, 9611355, 9623929, 9623951, 9623952, 9623953, 9623954, 9636367, 9640581, 9640582, 9642976, 9650106, 9650107, 9650111, 9660203, 9664006, 9664007, 9664008, 9664009, 9664034, 9664035 9365290, 9365291, 9376589, 9376590, 9380004, 9383412, 9386651, 9566153, 9566158, 9566210, 9575374, 9575376, 9575377, 9581877, 9584270, 9605692, 9605693, 9611357, 9611358, 9611359, 9611360, 9623955, 9623956, 9623957, 9623958, 9625911, 9625912, 9633329, 9633330, 9633331, 9633332, 9636368, 9636369, 9640583, 9650112, 9653871, 9653872, 9653873, 9664010, 9664011, 9664780, 9664781 |
Recalling Firm/ Manufacturer |
Cook Inc. 750 N Daniels Way Bloomington IN 47404-9120
|
For Additional Information Contact |
Cook Medical Customer Relations Department 812-339-2235
|
Manufacturer Reason for Recall |
may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may occur if an affected product is used include a prolonged procedure and open surgical intervention.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Urgent Medical Device Recall notification letters dated 4/26/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions or concerns, contact
Cook Medical Customer Relations at 800-457-4500 or 812-339-2235. |
Quantity in Commerce |
1027 |
Distribution |
No US distribution.
Product distributed to the following foreign countries: Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile Columbia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, UK |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|