|
Class 2 Device Recall PrisMax System |
|
Date Initiated by Firm |
January 07, 2021 |
Create Date |
February 26, 2021 |
Recall Status1 |
Terminated 3 on April 30, 2024 |
Recall Number |
Z-1151-2021 |
Recall Event ID |
87147 |
510(K)Number |
K190910
|
Product Classification |
Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
|
Product |
The PrisMax System Version II Hemodialysis Delivery System: Automatic Repositioning System Pump Assembly, specifically tubing, Model Numbers: a) 955626 b) 955558 c) 955701 PrisMax V2 Preventive Maintenance Kit, Product Code SC616 ARPS PUMP, Product Code SC6035 ARPS Pump Assembly, Product Code SC6053
|
Code Information |
Model Numbers: serial number, UDI; a) 955626: 101227, (01) 00085412639499 (21) 101227; 101280, (01) 00085412639499 (21) 101280; 101292, (01) 00085412639499 (21) 101292; 101293, (01) 00085412639499 (21) 101293; 101327, (01) 00085412639499 (21) 101327; 101348, (01) 00085412639499 (21) 101348; 101352, (01) 00085412639499 (21) 101352; 101353, (01) 00085412639499 (21) 101353; 101389, (01) 00085412639499 (21) 101389; 101391, (01) 00085412639499 (21) 101391; 101393, (01) 00085412639499 (21) 101393; 101394, (01) 00085412639499 (21) 101394; 101398, (01) 00085412639499 (21) 101398; 101422, (01) 00085412639499 (21) 101422; 101423, (01) 00085412639499 (21) 101423; 101450, (01) 00085412639499 (21) 101450; 101452, (01) 00085412639499 (21) 101452; 101456, (01) 00085412639499 (21) 101456; 101791, (01) 00085412639499 (21) 101791; 102142, (01) 00085412639499 (21) 102142; 102151, (01) 00085412639499 (21) 102151; 102152, (01) 00085412639499 (21) 102152; 102153, (01) 00085412639499 (21) 102153; 102154, (01) 00085412639499 (21) 102154; 102155, (01) 00085412639499 (21) 102155; 102159, (01) 00085412639499 (21) 102159; 102312, (01) 00085412639499 (21) 102312; 102318, (01) 00085412639499 (21) 102318; 102326, (01) 00085412639499 (21) 102326; 102328, (01) 00085412639499 (21) 102328; 102329, (01) 00085412639499 (21) 102329; 102330, (01) 00085412639499 (21) 102330; 102353, (01) 00085412639499 (21) 102353; 102356, (01) 00085412639499 (21) 102356; 102364, (01) 00085412639499 (21) 102364; 102365, (01) 00085412639499 (21) 102365; 102397, (01) 00085412639499 (21) 102397; 102399, (01) 00085412639499 (21) 102399; 102400, (01) 00085412639499 (21) 102400; 102402, (01) 00085412639499 (21) 102402; 102403, (01) 00085412639499 (21) 102403; 102404, (01) 00085412639499 (21) 102404; 102406, (01) 00085412639499 (21) 102406; 102407, (01) 00085412639499 (21) 102407; 102408, (01) 00085412639499 (21) 102408; 102409, (01) 00085412639499 (21) 102409; 102410, (01) 00085412639499 (21) 102410; 102411, (01) 00085412639499 (21) 102411; 102413, (01) 00085412639499 (21) 102413; 102414, (01) 00085412639499 (21) 102414; 102415, (01) 00085412639499 (21) 102415; 102416, (01) 00085412639499 (21) 102416; 102417, (01) 00085412639499 (21) 102417; 102418, (01) 00085412639499 (21) 102418; 102419, (01) 00085412639499 (21) 102419; 102420, (01) 00085412639499 (21) 102420; 102422, (01) 00085412639499 (21) 102422; 102423, (01) 00085412639499 (21) 102423; 102424, (01) 00085412639499 (21) 102424; 102425, (01) 00085412639499 (21) 102425; 102426, (01) 00085412639499 (21) 102426; 102457, (01) 00085412639499 (21) 102457; 102458, (01) 00085412639499 (21) 102458; 102459, (01) 00085412639499 (21) 102459; 102460, (01) 00085412639499 (21) 102460; 102461, (01) 00085412639499 (21) 102461; 102462, (01) 00085412639499 (21) 102462; 102463, (01) 00085412639499 (21) 102463; 102464, (01) 00085412639499 (21) 102464; 102465, (01) 00085412639499 (21) 102465; 102466, (01) 00085412639499 (21) 102466; 102467, (01) 00085412639499 (21) 102467; 102468, (01) 00085412639499 (21) 102468; 102469, (01) 00085412639499 (21) 102469; 102470, (01) 00085412639499 (21) 102470; 102471, (01) 00085412639499 (21) 102471; 102487, (01) 00085412639499 (21) 102487; 102488, (01) 00085412639499 (21) 102488; 102489, (01) 00085412639499 (21) 102489; 102490, (01) 00085412639499 (21) 102490; 102491, (01) 00085412639499 (21) 102491; 102492, (01) 00085412639499 (21) 102492; 102493, (01) 00085412639499 (21) 102493; 102494, (01) 00085412639499 (21) 102494; 102495, (01) 00085412639499 (21) 102495; 102496, (01) 00085412639499 (21) 102496; 102497, (01) 00085412639499 (21) 102497; 102498, (01) 00085412639499 (21) 102498; 102499, (01) 00085412639499 (21) 102499; 102500, (01) 00085412639499 (21) 102500; 102501, (01) 00085412639499 (21) 102501; 102876, (01) 00085412639499 (21) 102876; 102877, (01) 00085412639499 (21) 102877; 102878, (01) 00085412639499 (21) 102878; 102879, (01) 00085412639499 (21) 102879; 102880, (01) 00085412639499 (21) 102880; 102881, (01) 00085412639499 (21) 102881; 102882, (01) 00085412639499 (21) 102882; 102883, (01) 00085412639499 (21) 102883; 102884, (01) 00085412639499 (21) 102884; 102885, (01) 00085412639499 (21) 102885; 102886, (01) 00085412639499 (21) 102886; 102887, (01) 00085412639499 (21) 102887; 102888, (01) 00085412639499 (21) 102888; 102889, (01) 00085412639499 (21) 102889; 102890, (01) 00085412639499 (21) 102890; 102891, (01) 00085412639499 (21) 102891; 102892, (01) 00085412639499 (21) 102892; 102893, (01) 00085412639499 (21) 102893; 102894, (01) 00085412639499 (21) 102894; 102895, (01) 00085412639499 (21) 102895; 102896, (01) 00085412639499 (21) 102896; 102897, (01) 00085412639499 (21) 102897; 102898, (01) 00085412639499 (21) 102898; 102899, (01) 00085412639499 (21) 102899; 102900, (01) 00085412639499 (21) 102900; 102921, (01) 00085412639499 (21) 102921; 102922, (01) 00085412639499 (21) 102922; 102923, (01) 00085412639499 (21) 102923; 102924, (01) 00085412639499 (21) 102924; 102925, (01) 00085412639499 (21) 102925; 102926, (01) 00085412639499 (21) 102926; 102927, (01) 00085412639499 (21) 102927; 102928, (01) 00085412639499 (21) 102928; 102929, (01) 00085412639499 (21) 102929; 102930, (01) 00085412639499 (21) 102930; 103186, (01) 00085412639499 (21) 103186; 103187, (01) 00085412639499 (21) 103187; 103188, (01) 00085412639499 (21) 103188; 103189, (01) 00085412639499 (21) 103189; 103190, (01) 00085412639499 (21) 103190; 103191, (01) 00085412639499 (21) 103191; 103192, (01) 00085412639499 (21) 103192; 103193, (01) 00085412639499 (21) 103193; 103194, (01) 00085412639499 (21) 103194; 103195, (01) 00085412639499 (21) 103195; 103196, (01) 00085412639499 (21) 103196; 103197, (01) 00085412639499 (21) 103197; 103198, (01) 00085412639499 (21) 103198; 103199, (01) 00085412639499 (21) 103199; 103200, (01) 00085412639499 (21) 103200; 103201, (01) 00085412639499 (21) 103201; 103202, (01) 00085412639499 (21) 103202; 103203, (01) 00085412639499 (21) 103203; 103204, (01) 00085412639499 (21) 103204; 103205, (01) 00085412639499 (21) 103205; 103206, (01) 00085412639499 (21) 103206; 103207, (01) 00085412639499 (21) 103207; 103208, (01) 00085412639499 (21) 103208; 103209, (01) 00085412639499 (21) 103209; 103210, (01) 00085412639499 (21) 103210; 103211, (01) 00085412639499 (21) 103211; 103212, (01) 00085412639499 (21) 103212; 103213, (01) 00085412639499 (21) 103213; 103214, (01) 00085412639499 (21) 103214; 103215, (01) 00085412639499 (21) 103215; 103216, (01) 00085412639499 (21) 103216; 103217, (01) 00085412639499 (21) 103217; 103218, (01) 00085412639499 (21) 103218; 103219, (01) 00085412639499 (21) 103219; 103220, (01) 00085412639499 (21) 103220; 103221, (01) 00085412639499 (21) 103221; 103222, (01) 00085412639499 (21) 103222; 103223, (01) 00085412639499 (21) 103223; 103224, (01) 00085412639499 (21) 103224; 103225, (01) 00085412639499 (21) 103225; 103226, (01) 00085412639499 (21) 103226; 103227, (01) 00085412639499 (21) 103227; 103228, (01) 00085412639499 (21) 103228; 103229, (01) 00085412639499 (21) 103229; 103230, (01) 00085412639499 (21) 103230; 103231, (01) 00085412639499 (21) 103231; 103232, (01) 00085412639499 (21) 103232; 103233, (01) 00085412639499 (21) 103233; 103234, (01) 00085412639499 (21) 103234; 103235, (01) 00085412639499 (21) 103235; 103326, (01) 00085412639499 (21) 103326; 103327, (01) 00085412639499 (21) 103327; 103328, (01) 00085412639499 (21) 103328; 103329, (01) 00085412639499 (21) 103329; 103330, (01) 00085412639499 (21) 103330; 103331, (01) 00085412639499 (21) 103331; 103332, (01) 00085412639499 (21) 103332; 103333, (01) 00085412639499 (21) 103333; 103334, (01) 00085412639499 (21) 103334; 103335, (01) 00085412639499 (21) 103335; 103336, (01) 00085412639499 (21) 103336; 103337, (01) 00085412639499 (21) 103337; 103338, (01) 00085412639499 (21) 103338; 103339, (01) 00085412639499 (21) 103339; 103340, (01) 00085412639499 (21) 103340; 103341, (01) 00085412639499 (21) 103341; 103342, (01) 00085412639499 (21) 103342; 103343, (01) 00085412639499 (21) 103343; 103344, (01) 00085412639499 (21) 103344; 103345, (01) 00085412639499 (21) 103345; 103366, (01) 00085412639499 (21) 103366; 103367, (01) 00085412639499 (21) 103367; 103368, (01) 00085412639499 (21) 103368; 103369, (01) 00085412639499 (21) 103369; 103370, (01) 00085412639499 (21) 103370; 103371, (01) 00085412639499 (21) 103371; 103372, (01) 00085412639499 (21) 103372; 103373, (01) 00085412639499 (21) 103373; 103374, (01) 00085412639499 (21) 103374; 103375, (01) 00085412639499 (21) 103375; 103426, (01) 00085412639499 (21) 103426; 103427, (01) 00085412639499 (21) 103427; 103428, (01) 00085412639499 (21) 103428; 103429, (01) 00085412639499 (21) 103429; 103430, (01) 00085412639499 (21) 103430; 103431, (01) 00085412639499 (21) 103431; 103432, (01) 00085412639499 (21) 103432; 103433, (01) 00085412639499 (21) 103433; 103434, (01) 00085412639499 (21) 103434; 103435, (01) 00085412639499 (21) 103435; 103456, (01) 00085412639499 (21) 103456; 103457, (01) 00085412639499 (21) 103457; 103458, (01) 00085412639499 (21) 103458; 103459, (01) 00085412639499 (21) 103459; 103460, (01) 00085412639499 (21) 103460; 103461, (01) 00085412639499 (21) 103461; 103462, (01) 00085412639499 (21) 103462; 103463, (01) 00085412639499 (21) 103463; 103464, (01) 00085412639499 (21) 103464; 103465, (01) 00085412639499 (21) 103465; 103536, (01) 00085412639499 (21) 103536; 103537, (01) 00085412639499 (21) 103537; 103538, (01) 00085412639499 (21) 103538; 103539, (01) 00085412639499 (21) 103539; 103540, (01) 00085412639499 (21) 103540; 103541, (01) 00085412639499 (21) 103541; 103542, (01) 00085412639499 (21) 103542; 103543, (01) 00085412639499 (21) 103543; 103544, (01) 00085412639499 (21) 103544; 103545, (01) 00085412639499 (21) 103545; 103546, (01) 00085412639499 (21) 103546; 103547, (01) 00085412639499 (21) 103547; 103548, (01) 00085412639499 (21) 103548; 103549, (01) 00085412639499 (21) 103549; 103550, (01) 00085412639499 (21) 103550; 103551, (01) 00085412639499 (21) 103551; 103552, (01) 00085412639499 (21) 103552; 103553, (01) 00085412639499 (21) 103553; 103554, (01) 00085412639499 (21) 103554; 103555, (01) 00085412639499 (21) 103555; 103696, (01) 00085412639499 (21) 103696; 103697, (01) 00085412639499 (21) 103697; 103698, (01) 00085412639499 (21) 103698; 103699, (01) 00085412639499 (21) 103699; 103700, (01) 00085412639499 (21) 103700; 103701, (01) 00085412639499 (21) 103701; 103702, (01) 00085412639499 (21) 103702; 103703, (01) 00085412639499 (21) 103703; 103704, (01) 00085412639499 (21) 103704; 103705, (01) 00085412639499 (21) 103705; 103716, (01) 00085412639499 (21) 103716; 103717, (01) 00085412639499 (21) 103717; 103718, (01) 00085412639499 (21) 103718; 103719, (01) 00085412639499 (21) 103719; 103720, (01) 00085412639499 (21) 103720; 103721, (01) 00085412639499 (21) 103721; 103722, (01) 00085412639499 (21) 103722; 103723, (01) 00085412639499 (21) 103723; 103724, (01) 00085412639499 (21) 103724; 103725, (01) 00085412639499 (21) 103725; 103726, (01) 00085412639499 (21) 103726; 103727, (01) 00085412639499 (21) 103727; 103728, (01) 00085412639499 (21) 103728; 103729, (01) 00085412639499 (21) 103729; 103730, (01) 00085412639499 (21) 103730; 103731, (01) 00085412639499 (21) 103731; 103732, (01) 00085412639499 (21) 103732; 103733, (01) 00085412639499 (21) 103733; 103734, (01) 00085412639499 (21) 103734; 103735, (01) 00085412639499 (21) 103735; 103746, (01) 00085412639499 (21) 103746; 103747, (01) 00085412639499 (21) 103747; 103748, (01) 00085412639499 (21) 103748; 103749, (01) 00085412639499 (21) 103749; 103750, (01) 00085412639499 (21) 103750; 103751, (01) 00085412639499 (21) 103751; 103752, (01) 00085412639499 (21) 103752; 103753, (01) 00085412639499 (21) 103753; 103754, (01) 00085412639499 (21) 103754; 103755, (01) 00085412639499 (21) 103755; 103866, (01) 00085412639499 (21) 103866; 103867, (01) 00085412639499 (21) 103867; 103868, (01) 00085412639499 (21) 103868; 103869, (01) 00085412639499 (21) 103869; 103870, (01) 00085412639499 (21) 103870; 103871, (01) 00085412639499 (21) 103871; 103872, (01) 00085412639499 (21) 103872; 103873, (01) 00085412639499 (21) 103873; 103874, (01) 00085412639499 (21) 103874; 103875, (01) 00085412639499 (21) 103875; 103876, (01) 00085412639499 (21) 103876; 103877, (01) 00085412639499 (21) 103877; 103878, (01) 00085412639499 (21) 103878; 103879, (01) 00085412639499 (21) 103879; 103880, (01) 00085412639499 (21) 103880; 103881, (01) 00085412639499 (21) 103881; 103882, (01) 00085412639499 (21) 103882; 103883, (01) 00085412639499 (21) 103883; 103884, (01) 00085412639499 (21) 103884; 103885, (01) 00085412639499 (21) 103885; 103886, (01) 00085412639499 (21) 103886; 103887, (01) 00085412639499 (21) 103887; 103888, (01) 00085412639499 (21) 103888; 103889, (01) 00085412639499 (21) 103889; 103890, (01) 00085412639499 (21) 103890; 103891, (01) 00085412639499 (21) 103891; 103892, (01) 00085412639499 (21) 103892; 103893, (01) 00085412639499 (21) 103893; 103894, (01) 00085412639499 (21) 103894; 103895, (01) 00085412639499 (21) 103895; 103896, (01) 00085412639499 (21) 103896; 103897, (01) 00085412639499 (21) 103897; 103898, (01) 00085412639499 (21) 103898; 103899, (01) 00085412639499 (21) 103899; 103900, (01) 00085412639499 (21) 103900; 103901, (01) 00085412639499 (21) 103901; 103902, (01) 00085412639499 (21) 103902; 103903, (01) 00085412639499 (21) 103903; 103904, (01) 00085412639499 (21) 103904; 103905, (01) 00085412639499 (21) 103905; 103916, (01) 00085412639499 (21) 103916; 103917, (01) 00085412639499 (21) 103917; 103918, (01) 00085412639499 (21) 103918; 103919, (01) 00085412639499 (21) 103919; 103920, (01) 00085412639499 (21) 103920; 103921, (01) 00085412639499 (21) 103921; 103922, (01) 00085412639499 (21) 103922; 103923, (01) 00085412639499 (21) 103923; 103924, (01) 00085412639499 (21) 103924; 103925, (01) 00085412639499 (21) 103925; 103936, (01) 00085412639499 (21) 103936; 103937, (01) 00085412639499 (21) 103937; 103938, (01) 00085412639499 (21) 103938; 103939, (01) 00085412639499 (21) 103939; 103940, (01) 00085412639499 (21) 103940; 103941, (01) 00085412639499 (21) 103941; 103942, (01) 00085412639499 (21) 103942; 103943, (01) 00085412639499 (21) 103943; 103944, (01) 00085412639499 (21) 103944; 103945, (01) 00085412639499 (21) 103945; 104076, (01) 00085412639499 (21) 104076; 104077, (01) 00085412639499 (21) 104077; 104078, (01) 00085412639499 (21) 104078; 104079, (01) 00085412639499 (21) 104079; 104080, (01) 00085412639499 (21) 104080; 104081, (01) 00085412639499 (21) 104081; 104082, (01) 00085412639499 (21) 104082; 104083, (01) 00085412639499 (21) 104083; 104084, (01) 00085412639499 (21) 104084; 104085, (01) 00085412639499 (21) 104085; 104106, (01) 00085412639499 (21) 104106; 104107, (01) 00085412639499 (21) 104107; 104108, (01) 00085412639499 (21) 104108; 104109, (01) 00085412639499 (21) 104109; 104110, (01) 00085412639499 (21) 104110; 104111, (01) 00085412639499 (21) 104111; 104112, (01) 00085412639499 (21) 104112; 104113, (01) 00085412639499 (21) 104113; 104114, (01) 00085412639499 (21) 104114; 104115, (01) 00085412639499 (21) 104115; 104126, (01) 00085412639499 (21) 104126; 104127, (01) 00085412639499 (21) 104127; 104128, (01) 00085412639499 (21) 104128; 104129, (01) 00085412639499 (21) 104129; 104130, (01) 00085412639499 (21) 104130; 104131, (01) 00085412639499 (21) 104131; 104132, (01) 00085412639499 (21) 104132; 104133, (01) 00085412639499 (21) 104133; 104134, (01) 00085412639499 (21) 104134; 104135, (01) 00085412639499 (21) 104135; 104146, (01) 00085412639499 (21) 104146; 104147, (01) 00085412639499 (21) 104147; 104148, (01) 00085412639499 (21) 104148; 104149, (01) 00085412639499 (21) 104149; 104150, (01) 00085412639499 (21) 104150; 104151, (01) 00085412639499 (21) 104151; 104152, (01) 00085412639499 (21) 104152; 104153, (01) 00085412639499 (21) 104153; 104154, (01) 00085412639499 (21) 104154; 104155, (01) 00085412639499 (21) 104155; 104206, (01) 00085412639499 (21) 104206; 104207, (01) 00085412639499 (21) 104207; 104208, (01) 00085412639499 (21) 104208; 104209, (01) 00085412639499 (21) 104209; 104210, (01) 00085412639499 (21) 104210; 104211, (01) 00085412639499 (21) 104211; 104212, (01) 00085412639499 (21) 104212; 104213, (01) 00085412639499 (21) 104213; 104214, (01) 00085412639499 (21) 104214; 104215, (01) 00085412639499 (21) 104215; 104226, (01) 00085412639499 (21) 104226; 104227, (01) 00085412639499 (21) 104227; 104228, (01) 00085412639499 (21) 104228; 104229, (01) 00085412639499 (21) 104229; 104230, (01) 00085412639499 (21) 104230; 104231, (01) 00085412639499 (21) 104231; 104232, (01) 00085412639499 (21) 104232; 104233, (01) 00085412639499 (21) 104233; 104234, (01) 00085412639499 (21) 104234; 104235, (01) 00085412639499 (21) 104235; 104326, (01) 00085412639499 (21) 104326; 104327, (01) 00085412639499 (21) 104327; 104328, (01) 00085412639499 (21) 104328; 104329, (01) 00085412639499 (21) 104329; 104330, (01) 00085412639499 (21) 104330; 104331, (01) 00085412639499 (21) 104331; 104332, (01) 00085412639499 (21) 104332; 104333, (01) 00085412639499 (21) 104333; 104334, (01) 00085412639499 (21) 104334; 104335, (01) 00085412639499 (21) 104335; 104346, (01) 00085412639499 (21) 104346; 104347, (01) 00085412639499 (21) 104347; 104348, (01) 00085412639499 (21) 104348; 104349, (01) 00085412639499 (21) 104349; 104350, (01) 00085412639499 (21) 104350; 104351, (01) 00085412639499 (21) 104351; 104352, (01) 00085412639499 (21) 104352; 104353, (01) 00085412639499 (21) 104353; 104354, (01) 00085412639499 (21) 104354; 104355, (01) 00085412639499 (21) 104355; 104376, (01) 00085412639499 (21) 104376; 104377, (01) 00085412639499 (21) 104377; 104378, (01) 00085412639499 (21) 104378; 104379, (01) 00085412639499 (21) 104379; 104380, (01) 00085412639499 (21) 104380; 104381, (01) 00085412639499 (21) 104381; 104382, (01) 00085412639499 (21) 104382; 104383, (01) 00085412639499 (21) 104383; 104384, (01) 00085412639499 (21) 104384; 104385, (01) 00085412639499 (21) 104385; 104656, (01) 00085412639499 (21) 104656; 104657, (01) 00085412639499 (21) 104657; 104658, (01) 00085412639499 (21) 104658; 104659, (01) 00085412639499 (21) 104659; 104660, (01) 00085412639499 (21) 104660; 104661, (01) 00085412639499 (21) 104661; 104662, (01) 00085412639499 (21) 104662; 104663, (01) 00085412639499 (21) 104663; 104664, (01) 00085412639499 (21) 104664; 104665, (01) 00085412639499 (21) 104665; 104696, (01) 00085412639499 (21) 104696; 104697, (01) 00085412639499 (21) 104697; 104698, (01) 00085412639499 (21) 104698; 104699, (01) 00085412639499 (21) 104699; 104700, (01) 00085412639499 (21) 104700; 104701, (01) 00085412639499 (21) 104701; 104702, (01) 00085412639499 (21) 104702; 104703, (01) 00085412639499 (21) 104703; 104704, (01) 00085412639499 (21) 104704; 104705, (01) 00085412639499 (21) 104705; 104716, (01) 00085412639499 (21) 104716; 104717, (01) 00085412639499 (21) 104717; 104718, (01) 00085412639499 (21) 104718; 104719, (01) 00085412639499 (21) 104719; 104720, (01) 00085412639499 (21) 104720; 104721, (01) 00085412639499 (21) 104721; 104722, (01) 00085412639499 (21) 104722; 104723, (01) 00085412639499 (21) 104723; 104724, (01) 00085412639499 (21) 104724; 104725, (01) 00085412639499 (21) 104725; 104746, (01) 00085412639499 (21) 104746; 104747, (01) 00085412639499 (21) 104747; 104748, (01) 00085412639499 (21) 104748; 104749, (01) 00085412639499 (21) 104749; 104750, (01) 00085412639499 (21) 104750; 104751, (01) 00085412639499 (21) 104751; 104752, (01) 00085412639499 (21) 104752; 104753, (01) 00085412639499 (21) 104753; 104754, (01) 00085412639499 (21) 104754; 104755, (01) 00085412639499 (21) 104755; 104856, (01) 00085412639499 (21) 104856; 104857, (01) 00085412639499 (21) 104857; 104858, (01) 00085412639499 (21) 104858; 104859, (01) 00085412639499 (21) 104859; 104860, (01) 00085412639499 (21) 104860; 104861, (01) 00085412639499 (21) 104861; 104862, (01) 00085412639499 (21) 104862; 104863, (01) 00085412639499 (21) 104863; 104864, (01) 00085412639499 (21) 104864; 104865, (01) 00085412639499 (21) 104865; 104886, (01) 00085412639499 (21) 104886; 104887, (01) 00085412639499 (21) 104887; 104888, (01) 00085412639499 (21) 104888; 104889, (01) 00085412639499 (21) 104889; 104890, (01) 00085412639499 (21) 104890; 104891, (01) 00085412639499 (21) 104891; 104892, (01) 00085412639499 (21) 104892; 104893, (01) 00085412639499 (21) 104893; 104894, (01) 00085412639499 (21) 104894; 104895, (01) 00085412639499 (21) 104895; 105006, (01) 00085412639499 (21) 105006; 105007, (01) 00085412639499 (21) 105007; 105008, (01) 00085412639499 (21) 105008; 105009, (01) 00085412639499 (21) 105009; 105010, (01) 00085412639499 (21) 105010; 105011, (01) 00085412639499 (21) 105011; 105012, (01) 00085412639499 (21) 105012; 105013, (01) 00085412639499 (21) 105013; 105014, (01) 00085412639499 (21) 105014; 105015, (01) 00085412639499 (21) 105015; 105016, (01) 00085412639499 (21) 105016; 105017, (01) 00085412639499 (21) 105017; 105018, (01) 00085412639499 (21) 105018; 105019, (01) 00085412639499 (21) 105019; 105020, (01) 00085412639499 (21) 105020; 105021, (01) 00085412639499 (21) 105021; 105022, (01) 00085412639499 (21) 105022; 105023, (01) 00085412639499 (21) 105023; 105024, (01) 00085412639499 (21) 105024; 105025, (01) 00085412639499 (21) 105025; 105036, (01) 00085412639499 (21) 105036; 105037, (01) 00085412639499 (21) 105037; 105038, (01) 00085412639499 (21) 105038; 105039, (01) 00085412639499 (21) 105039; 105040, (01) 00085412639499 (21) 105040; 105041, (01) 00085412639499 (21) 105041; 105042, (01) 00085412639499 (21) 105042; 105043, (01) 00085412639499 (21) 105043; 105044, (01) 00085412639499 (21) 105044; 105045, (01) 00085412639499 (21) 105045; 105246, (01) 00085412639499 (21) 105246; 105247, (01) 00085412639499 (21) 105247; 105248, (01) 00085412639499 (21) 105248; 105249, (01) 00085412639499 (21) 105249; 105250, (01) 00085412639499 (21) 105250; 105251, (01) 00085412639499 (21) 105251; 105252, (01) 00085412639499 (21) 105252; 105253, (01) 00085412639499 (21) 105253; 105254, (01) 00085412639499 (21) 105254; 105255, (01) 00085412639499 (21) 105255; 105256, (01) 00085412639499 (21) 105256; 105257, (01) 00085412639499 (21) 105257; 105258, (01) 00085412639499 (21) 105258; 105259, (01) 00085412639499 (21) 105259; 105260, (01) 00085412639499 (21) 105260; 105261, (01) 00085412639499 (21) 105261; 105262, (01) 00085412639499 (21) 105262; 105263, (01) 00085412639499 (21) 105263; 105264, (01) 00085412639499 (21) 105264; 105265, (01) 00085412639499 (21) 105265; 105276, (01) 00085412639499 (21) 105276; 105277, (01) 00085412639499 (21) 105277; 105278, (01) 00085412639499 (21) 105278; 105279, (01) 00085412639499 (21) 105279; 105280, (01) 00085412639499 (21) 105280; 105281, (01) 00085412639499 (21) 105281; 105282, (01) 00085412639499 (21) 105282; 105283, (01) 00085412639499 (21) 105283; 105284, (01) 00085412639499 (21) 105284; 105285, (01) 00085412639499 (21) 105285; 105496, (01) 00085412639499 (21) 105496; 105497, (01) 00085412639499 (21) 105497; 105498, (01) 00085412639499 (21) 105498; 105499, (01) 00085412639499 (21) 105499; 105500, (01) 00085412639499 (21) 105500; 105501, (01) 00085412639499 (21) 105501; 105502, (01) 00085412639499 (21) 105502; 105503, (01) 00085412639499 (21) 105503; 105504, (01) 00085412639499 (21) 105504; 105505, (01) 00085412639499 (21) 105505; 105526, (01) 00085412639499 (21) 105526; 105527, (01) 00085412639499 (21) 105527; 105528, (01) 00085412639499 (21) 105528; 105529, (01) 00085412639499 (21) 105529; 105530, (01) 00085412639499 (21) 105530; 105531, (01) 00085412639499 (21) 105531; 105532, (01) 00085412639499 (21) 105532; 105533, (01) 00085412639499 (21) 105533; 105534, (01) 00085412639499 (21) 105534; 105535, (01) 00085412639499 (21) 105535; 105626, (01) 00085412639499 (21) 105626; 105627, (01) 00085412639499 (21) 105627; 105628, (01) 00085412639499 (21) 105628; 105629, (01) 00085412639499 (21) 105629; 105630, (01) 00085412639499 (21) 105630; 105631, (01) 00085412639499 (21) 105631; 105632, (01) 00085412639499 (21) 105632; 105633, (01) 00085412639499 (21) 105633; 105634, (01) 00085412639499 (21) 105634; 105635, (01) 00085412639499 (21) 105635; 105822, (01) 00085412639499 (21) 105822; 105823, (01) 00085412639499 (21) 105823; 105824, (01) 00085412639499 (21) 105824; 105825, (01) 00085412639499 (21) 105825; 105826, (01) 00085412639499 (21) 105826; 105827, (01) 00085412639499 (21) 105827; 105828, (01) 00085412639499 (21) 105828; 105829, (01) 00085412639499 (21) 105829; 105830, (01) 00085412639499 (21) 105830; 105831, (01) 00085412639499 (21) 105831; 106062, (01) 00085412639499 (21) 106062; 106063, (01) 00085412639499 (21) 106063; 106064, (01) 00085412639499 (21) 106064; 106067, (01) 00085412639499 (21) 106067; 106068, (01) 00085412639499 (21) 106068; 106069, (01) 00085412639499 (21) 106069; 106070, (01) 00085412639499 (21) 106070 (603 units). b) 955558: 100266, (01) 07332414124359 (21) 100266; 100267, (01) 07332414124359 (21) 100267; 100269, (01) 07332414124359 (21) 100269; 100271, (01) 07332414124359 (21) 100271; 100272, (01) 07332414124359 (21) 100272; 100273, (01) 07332414124359 (21) 100273; 100274, (01) 07332414124359 (21) 100274; 100275, (01) 07332414124359 (21) 100275; 100279, (01) 07332414124359 (21) 100279; 100300, (01) 07332414124359 (21) 100300; 100301, (01) 07332414124359 (21) 100301; 100303, (01) 07332414124359 (21) 100303; 100304, (01) 07332414124359 (21) 100304; 100316, (01) 07332414124359 (21) 100316; 100317, (01) 07332414124359 (21) 100317; 100318, (01) 07332414124359 (21) 100318; 100320, (01) 07332414124359 (21) 100320; 100331, (01) 07332414124359 (21) 100331; 100332, (01) 07332414124359 (21) 100332; 100333, (01) 07332414124359 (21) 100333; 100335, (01) 07332414124359 (21) 100335; 100336, (01) 07332414124359 (21) 100336; 100337, (01) 07332414124359 (21) 100337; 100340, (01) 07332414124359 (21) 100340; 100351, (01) 07332414124359 (21) 100351; 100352, (01) 07332414124359 (21) 100352; 100353, (01) 07332414124359 (21) 100353; 100371, (01) 07332414124359 (21) 100371; 100373, (01) 07332414124359 (21) 100373; 100375, (01) 07332414124359 (21) 100375; 100376, (01) 07332414124359 (21) 100376; 100377, (01) 07332414124359 (21) 100377; 100381, (01) 07332414124359 (21) 100381; 100384, (01) 07332414124359 (21) 100384; 100385, (01) 07332414124359 (21) 100385; 100387, (01) 07332414124359 (21) 100387; 100388, (01) 07332414124359 (21) 100388; 100389, (01) 07332414124359 (21) 100389; 100390, (01) 07332414124359 (21) 100390; 100406, (01) 07332414124359 (21) 100406; 100408, (01) 07332414124359 (21) 100408; 100410, (01) 07332414124359 (21) 100410; 100411, (01) 07332414124359 (21) 100411; 100412, (01) 07332414124359 (21) 100412; 100413, (01) 07332414124359 (21) 100413; 100414, (01) 07332414124359 (21) 100414; 100432, (01) 07332414124359 (21) 100432; 100433, (01) 07332414124359 (21) 100433; 100434, (01) 07332414124359 (21) 100434; 100435, (01) 07332414124359 (21) 100435; 100461, (01) 07332414124359 (21) 100461; 100465, (01) 07332414124359 (21) 100465; 100467, (01) 07332414124359 (21) 100467; 100469, (01) 07332414124359 (21) 100469; 100471, (01) 07332414124359 (21) 100471; 100480, (01) 07332414124359 (21) 100480; 100481, (01) 07332414124359 (21) 100481; 100483, (01) 07332414124359 (21) 100483; 100484, (01) 07332414124359 (21) 100484; 100485, (01) 07332414124359 (21) 100485; 100508, (01) 07332414124359 (21) 100508; 100510, (01) 07332414124359 (21) 100510; 100511, (01) 07332414124359 (21) 100511; 100512, (01) 07332414124359 (21) 100512; 100513, (01) 07332414124359 (21) 100513; 100514, (01) 07332414124359 (21) 100514; 100515, (01) 07332414124359 (21) 100515; 100517, (01) 07332414124359 (21) 100517; 100518, (01) 07332414124359 (21) 100518; 100519, (01) 07332414124359 (21) 100519; 100520, (01) 07332414124359 (21) 100520; 100521, (01) 07332414124359 (21) 100521; 100523, (01) 07332414124359 (21) 100523; 100524, (01) 07332414124359 (21) 100524; 100525, (01) 07332414124359 (21) 100525; 100526, (01) 07332414124359 (21) 100526; 100528, (01) 07332414124359 (21) 100528; 100556, (01) 07332414124359 (21) 100556; 100557, (01) 07332414124359 (21) 100557; 100558, (01) 07332414124359 (21) 100558; 100559, (01) 07332414124359 (21) 100559; 100560, (01) 07332414124359 (21) 100560; 100561, (01) 07332414124359 (21) 100561; 100563, (01) 07332414124359 (21) 100563; 100565, (01) 07332414124359 (21) 100565; 100568, (01) 07332414124359 (21) 100568; 100570, (01) 07332414124359 (21) 100570; 100571, (01) 07332414124359 (21) 100571; 100573, (01) 07332414124359 (21) 100573; 100574, (01) 07332414124359 (21) 100574; 100576, (01) 07332414124359 (21) 100576; 100577, (01) 07332414124359 (21) 100577; 100578, (01) 07332414124359 (21) 100578; 100609, (01) 07332414124359 (21) 100609; 100611, (01) 07332414124359 (21) 100611; 100613, (01) 07332414124359 (21) 100613; 100615, (01) 07332414124359 (21) 100615; 100616, (01) 07332414124359 (21) 100616; 100617, (01) 07332414124359 (21) 100617; 100618, (01) 07332414124359 (21) 100618; 100620, (01) 07332414124359 (21) 100620; 100621, (01) 07332414124359 (21) 100621; 100622, (01) 07332414124359 (21) 100622; 100623, (01) 07332414124359 (21) 100623; 100624, (01) 07332414124359 (21) 100624; 100625, (01) 07332414124359 (21) 100625; 100629, (01) 07332414124359 (21) 100629; 100656, (01) 07332414124359 (21) 100656; 100657, (01) 07332414124359 (21) 100657; 100659, (01) 07332414124359 (21) 100659; 100660, (01) 07332414124359 (21) 100660; 100661, (01) 07332414124359 (21) 100661; 100662, (01) 07332414124359 (21) 100662; 100664, (01) 07332414124359 (21) 100664; 100667, (01) 07332414124359 (21) 100667; 100668, (01) 07332414124359 (21) 100668; 100670, (01) 07332414124359 (21) 100670; 100671, (01) 07332414124359 (21) 100671; 100673, (01) 07332414124359 (21) 100673; 100674, (01) 07332414124359 (21) 100674; 100676, (01) 07332414124359 (21) 100676; 100678, (01) 07332414124359 (21) 100678; 100679, (01) 07332414124359 (21) 100679; 100681, (01) 07332414124359 (21) 100681; 100682, (01) 07332414124359 (21) 100682; 100683, (01) 07332414124359 (21) 100683; 100685, (01) 07332414124359 (21) 100685; 100687, (01) 07332414124359 (21) 100687; 100688, (01) 07332414124359 (21) 100688; 100689, (01) 07332414124359 (21) 100689; 100692, (01) 07332414124359 (21) 100692; 100693, (01) 07332414124359 (21) 100693; 100696, (01) 07332414124359 (21) 100696; 100697, (01) 07332414124359 (21) 100697; 100698, (01) 07332414124359 (21) 100698; 100699, (01) 07332414124359 (21) 100699; 100700, (01) 07332414124359 (21) 100700; 100701, (01) 07332414124359 (21) 100701; 100702, (01) 07332414124359 (21) 100702; 100703, (01) 07332414124359 (21) 100703; 100705, (01) 07332414124359 (21) 100705; 100707, (01) 07332414124359 (21) 100707; 100712, (01) 07332414124359 (21) 100712; 100713, (01) 07332414124359 (21) 100713; 100715, (01) 07332414124359 (21) 100715; 100716, (01) 07332414124359 (21) 100716; 100721, (01) 07332414124359 (21) 100721; 100722, (01) 07332414124359 (21) 100722; 100728, (01) 07332414124359 (21) 100728; 100730, (01) 07332414124359 (21) 100730; 100731, (01) 07332414124359 (21) 100731; 100733, (01) 07332414124359 (21) 100733; 100736, (01) 07332414124359 (21) 100736; 100745, (01) 07332414124359 (21) 100745; 100746, (01) 07332414124359 (21) 100746; 100750, (01) 07332414124359 (21) 100750; 100751, (01) 07332414124359 (21) 100751; 100754.(01) 07332414124359 (21) 100754; 100834, (01) 07332414124359 (21) 100834; 100835, (01) 07332414124359 (21) 100835; 100836, (01) 07332414124359 (21) 100836; 100838, (01) 07332414124359 (21) 100838; 100842, (01) 07332414124359 (21) 100842; 100843, (01) 07332414124359 (21) 100843; 100846, (01) 07332414124359 (21) 100846; 100847, (01) 07332414124359 (21) 100847; 100907, (01) 07332414124359 (21) 100907; 100908, (01) 07332414124359 (21) 100908; 100909, (01) 07332414124359 (21) 100909; 100912, (01) 07332414124359 (21) 100912; 100913, (01) 07332414124359 (21) 100913; 100929, (01) 07332414124359 (21) 100929; 100930, (01) 07332414124359 (21) 100930; 100956, (01) 07332414124359 (21) 100956; 100958, (01) 07332414124359 (21) 100958; 100959, (01) 07332414124359 (21) 100959; 100962, (01) 07332414124359 (21) 100962; 100963, (01) 07332414124359 (21) 100963; 100964, (01) 07332414124359 (21) 100964; 100966, (01) 07332414124359 (21) 100966; 100967, (01) 07332414124359 (21) 100967; 100968, (01) 07332414124359 (21) 100968; 100970, (01) 07332414124359 (21) 100970; 100971, (01) 07332414124359 (21) 100971; 100972, (01) 07332414124359 (21) 100972; 100974, (01) 07332414124359 (21) 100974; 100975, (01) 07332414124359 (21) 100975; 101036, (01) 07332414124359 (21) 101036; 101037, (01) 07332414124359 (21) 101037; 101038, (01) 07332414124359 (21) 101038; 101039, (01) 07332414124359 (21) 101039; 101040, (01) 07332414124359 (21) 101040; 101041, (01) 07332414124359 (21) 101041; 101042, (01) 07332414124359 (21) 101042; 101045, (01) 07332414124359 (21) 101045; 101046, (01) 07332414124359 (21) 101046; 101047, (01) 07332414124359 (21) 101047; 101048, (01) 07332414124359 (21) 101048; 101049, (01) 07332414124359 (21) 101049; 101050, (01) 07332414124359 (21) 101050; 101051, (01) 07332414124359 (21) 101051; 101052, (01) 07332414124359 (21) 101052; 101053, (01) 07332414124359 (21) 101053; 101054, (01) 07332414124359 (21) 101054; 101057, (01) 07332414124359 (21) 101057; 101058, (01) 07332414124359 (21) 101058; 101059, (01) 07332414124359 (21) 101059; 101063, (01) 07332414124359 (21) 101063; 101064, (01) 07332414124359 (21) 101064; 101066, (01) 07332414124359 (21) 101066; 101067, (01) 07332414124359 (21) 101067; 101071, (01) 07332414124359 (21) 101071; 101072, (01) 07332414124359 (21) 101072; 101073, (01) 07332414124359 (21) 101073; 101074, (01) 07332414124359 (21) 101074; 101075, (01) 07332414124359 (21) 101075; 101076, (01) 07332414124359 (21) 101076; 101078, (01) 07332414124359 (21) 101078; 101080, (01) 07332414124359 (21) 101080; 101163, (01) 07332414124359 (21) 101163; 101164, (01) 07332414124359 (21) 101164; 101165, (01) 07332414124359 (21) 101165; 101166, (01) 07332414124359 (21) 101166; 101167, (01) 07332414124359 (21) 101167; 101193, (01) 07332414124359 (21) 101193; 101194, (01) 07332414124359 (21) 101194; 101195, (01) 07332414124359 (21) 101195; 101196, (01) 07332414124359 (21) 101196; 101197, (01) 07332414124359 (21) 101197; 101201, (01) 07332414124359 (21) 101201; 101204, (01) 07332414124359 (21) 101204; 101205, (01) 07332414124359 (21) 101205; 101207, (01) 07332414124359 (21) 101207; 101210, (01) 07332414124359 (21) 101210; 101213, (01) 07332414124359 (21) 101213; 101215, (01) 07332414124359 (21) 101215; 101216, (01) 07332414124359 (21) 101216; 101217, (01) 07332414124359 (21) 101217; 101297, (01) 07332414124359 (21) 101297; 101298, (01) 07332414124359 (21) 101298; 101299, (01) 07332414124359 (21) 101299; 101300, (01) 07332414124359 (21) 101300; 101301, (01) 07332414124359 (21) 101301; 101302, (01) 07332414124359 (21) 101302; 101305, (01) 07332414124359 (21) 101305; 101306, (01) 07332414124359 (21) 101306; 101307, (01) 07332414124359 (21) 101307; 101308, (01) 07332414124359 (21) 101308; 101309, (01) 07332414124359 (21) 101309; 101310, (01) 07332414124359 (21) 101310; 101312, (01) 07332414124359 (21) 101312; 101315, (01) 07332414124359 (21) 101315; 101318, (01) 07332414124359 (21) 101318; 101493, (01) 07332414124359 (21) 101493; 101497, (01) 07332414124359 (21) 101497; 101498, (01) 07332414124359 (21) 101498; 101500, (01) 07332414124359 (21) 101500; 101501, (01) 07332414124359 (21) 101501; 101502, (01) 07332414124359 (21) 101502; 101503, (01) 07332414124359 (21) 101503; 101505, (01) 07332414124359 (21) 101505; 101506, (01) 07332414124359 (21) 101506; 101507, (01) 07332414124359 (21) 101507; 101525, (01) 07332414124359 (21) 101525; 101526, (01) 07332414124359 (21) 101526; 101527, (01) 07332414124359 (21) 101527; 101528, (01) 07332414124359 (21) 101528; 101530, (01) 07332414124359 (21) 101530; 101535, (01) 07332414124359 (21) 101535; 101536, (01) 07332414124359 (21) 101536; 101563, (01) 07332414124359 (21) 101563; 101565, (01) 07332414124359 (21) 101565; 101566, (01) 07332414124359 (21) 101566; 101569, (01) 07332414124359 (21) 101569; 101572, (01) 07332414124359 (21) 101572; 101576, (01) 07332414124359 (21) 101576; 101578, (01) 07332414124359 (21) 101578; 101579, (01) 07332414124359 (21) 101579; 101587, (01) 07332414124359 (21) 101587; 101591, (01) 07332414124359 (21) 101591; 101592, (01) 07332414124359 (21) 101592; 101618, (01) 07332414124359 (21) 101618; 101619, (01) 07332414124359 (21) 101619; 101621, (01) 07332414124359 (21) 101621; 101624, (01) 07332414124359 (21) 101624; 101625, (01) 07332414124359 (21) 101625; 101626, (01) 07332414124359 (21) 101626; 101627, (01) 07332414124359 (21) 101627; 101628, (01) 07332414124359 (21) 101628; 101629, (01) 07332414124359 (21) 101629; 101631, (01) 07332414124359 (21) 101631; 101650, (01) 07332414124359 (21) 101650; 101654, (01) 07332414124359 (21) 101654; 101655, (01) 07332414124359 (21) 101655; 101661, (01) 07332414124359 (21) 101661; 101662, (01) 07332414124359 (21) 101662; 101663, (01) 07332414124359 (21) 101663; 101665, (01) 07332414124359 (21) 101665; 101670, (01) 07332414124359 (21) 101670; 101674, (01) 07332414124359 (21) 101674; 101677, (01) 07332414124359 (21) 101677; 101687, (01) 07332414124359 (21) 101687; 101703, (01) 07332414124359 (21) 101703; 101712, (01) 07332414124359 (21) 101712; 101713, (01) 07332414124359 (21) 101713; 101717, (01) 07332414124359 (21) 101717; 101718, (01) 07332414124359 (21) 101718; 101719, (01) 07332414124359 (21) 101719; 101722, (01) 07332414124359 (21) 101722; 101725, (01) 07332414124359 (21) 101725; 101726, (01) 07332414124359 (21) 101726; 101727, (01) 07332414124359 (21) 101727; 101764, (01) 07332414124359 (21) 101764; 101772, (01) 07332414124359 (21) 101772; 101773, (01) 07332414124359 (21) 101773; 101775, (01) 07332414124359 (21) 101775; 101776, (01) 07332414124359 (21) 101776; 101778, (01) 07332414124359 (21) 101778; 101787, (01) 07332414124359 (21) 101787; 101788, (01) 07332414124359 (21) 101788; 101799, (01) 07332414124359 (21) 101799; 101802, (01) 07332414124359 (21) 101802; 101804, (01) 07332414124359 (21) 101804; 101805, (01) 07332414124359 (21) 101805; 101809, (01) 07332414124359 (21) 101809; 101811, (01) 07332414124359 (21) 101811; 101812, (01) 07332414124359 (21) 101812; 101815, (01) 07332414124359 (21) 101815; 101822, (01) 07332414124359 (21) 101822; 101823, (01) 07332414124359 (21) 101823; 101824, (01) 07332414124359 (21) 101824; 101866, (01) 07332414124359 (21) 101866; 101872, (01) 07332414124359 (21) 101872; 101892, (01) 07332414124359 (21) 101892; 101903, (01) 07332414124359 (21) 101903; 101926, (01) 07332414124359 (21) 101926; 101928, (01) 07332414124359 (21) 101928; 101931, (01) 07332414124359 (21) 101931; 101937, (01) 07332414124359 (21) 101937; 101939, (01) 07332414124359 (21) 101939; 101941, (01) 07332414124359 (21) 101941; 101956, (01) 07332414124359 (21) 101956; 101958, (01) 07332414124359 (21) 101958; 101959, (01) 07332414124359 (21) 101959; 101965, (01) 07332414124359 (21) 101965; 101966, (01) 07332414124359 (21) 101966; 101970, (01) 07332414124359 (21) 101970; 101997, (01) 07332414124359 (21) 101997; 102004, (01) 07332414124359 (21) 102004; 102006, (01) 07332414124359 (21) 102006; 102011, (01) 07332414124359 (21) 102011; 102012, (01) 07332414124359 (21) 102012; 102014, (01) 07332414124359 (21) 102014; 102042, (01) 07332414124359 (21) 102042; 102044, (01) 07332414124359 (21) 102044; 102045, (01) 07332414124359 (21) 102045; 102170, (01) 07332414124359 (21) 102170; 102172, (01) 07332414124359 (21) 102172; 102197, (01) 07332414124359 (21) 102197; 102198, (01) 07332414124359 (21) 102198; 102200, (01) 07332414124359 (21) 102200; 102201, (01) 07332414124359 (21) 102201; 102204, (01) 07332414124359 (21) 102204; 102206, (01) 07332414124359 (21) 102206; 102210, (01) 07332414124359 (21) 102210; 102211, (01) 07332414124359 (21) 102211; 102212, (01) 07332414124359 (21) 102212; 102213, (01) 07332414124359 (21) 102213; 102214, (01) 07332414124359 (21) 102214; 102255, (01) 07332414124359 (21) 102255; 102256, (01) 07332414124359 (21) 102256; 102257, (01) 07332414124359 (21) 102257; 102258, (01) 07332414124359 (21) 102258; 102259, (01) 07332414124359 (21) 102259; 102260, (01) 07332414124359 (21) 102260; 102261, (01) 07332414124359 (21) 102261; 102262, (01) 07332414124359 (21) 102262; 102263, (01) 07332414124359 (21) 102263; 102264, (01) 07332414124359 (21) 102264; 102265, (01) 07332414124359 (21) 102265; 102266, (01) 07332414124359 (21) 102266; 102267, (01) 07332414124359 (21) 102267; 102268, (01) 07332414124359 (21) 102268; 102270, (01) 07332414124359 (21) 102270; 102271, (01) 07332414124359 (21) 102271; 102272, (01) 07332414124359 (21) 102272; 102273, (01) 07332414124359 (21) 102273; 102274, (01) 07332414124359 (21) 102274; 102275, (01) 07332414124359 (21) 102275; 102276, (01) 07332414124359 (21) 102276; 102277, (01) 07332414124359 (21) 102277; 102278, (01) 07332414124359 (21) 102278; 102279, (01) 07332414124359 (21) 102279; 102280, (01) 07332414124359 (21) 102280; 102281, (01) 07332414124359 (21) 102281; 102282, (01) 07332414124359 (21) 102282; 102283, (01) 07332414124359 (21) 102283; 102284, (01) 07332414124359 (21) 102284; 102285, (01) 07332414124359 (21) 102285; 102286, (01) 07332414124359 (21) 102286; 102287, (01) 07332414124359 (21) 102287; 102288, (01) 07332414124359 (21) 102288; 102289, (01) 07332414124359 (21) 102289; 102337, (01) 07332414124359 (21) 102337; 102338, (01) 07332414124359 (21) 102338; 102339, (01) 07332414124359 (21) 102339; 102340, (01) 07332414124359 (21) 102340; 102341, (01) 07332414124359 (21) 102341; 102342, (01) 07332414124359 (21) 102342; 102343, (01) 07332414124359 (21) 102343; 102344, (01) 07332414124359 (21) 102344; 102345, (01) 07332414124359 (21) 102345; 102346, (01) 07332414124359 (21) 102346; 102347, (01) 07332414124359 (21) 102347; 102348, (01) 07332414124359 (21) 102348; 102349, (01) 07332414124359 (21) 102349; 102350, (01) 07332414124359 (21) 102350; 102351, (01) 07332414124359 (21) 102351; 102532, (01) 07332414124359 (21) 102532; 102533, (01) 07332414124359 (21) 102533; 102534, (01) 07332414124359 (21) 102534; 102535, (01) 07332414124359 (21) 102535; 102536, (01) 07332414124359 (21) 102536; 102537, (01) 07332414124359 (21) 102537; 102538, (01) 07332414124359 (21) 102538; 102539, (01) 07332414124359 (21) 102539; 102540, (01) 07332414124359 (21) 102540; 102541, (01) 07332414124359 (21) 102541; 102542, (01) 07332414124359 (21) 102542; 102543, (01) 07332414124359 (21) 102543; 102544, (01) 07332414124359 (21) 102544; 102545, (01) 07332414124359 (21) 102545; 102546, (01) 07332414124359 (21) 102546; 102562, (01) 07332414124359 (21) 102562; 102563, (01) 07332414124359 (21) 102563; 102564, (01) 07332414124359 (21) 102564; 102565, (01) 07332414124359 (21) 102565; 102566, (01) 07332414124359 (21) 102566; 102567, (01) 07332414124359 (21) 102567; 102568, (01) 07332414124359 (21) 102568; 102569, (01) 07332414124359 (21) 102569; 102570, (01) 07332414124359 (21) 102570; 102571, (01) 07332414124359 (21) 102571; 102572, (01) 07332414124359 (21) 102572; 102573, (01) 07332414124359 (21) 102573; 102574, (01) 07332414124359 (21) 102574; 102575, (01) 07332414124359 (21) 102575; 102576, (01) 07332414124359 (21) 102576; 102592, (01) 07332414124359 (21) 102592; 102593, (01) 07332414124359 (21) 102593; 102594, (01) 07332414124359 (21) 102594; 102595, (01) 07332414124359 (21) 102595; 102596, (01) 07332414124359 (21) 102596; 102597, (01) 07332414124359 (21) 102597; 102598, (01) 07332414124359 (21) 102598; 102599, (01) 07332414124359 (21) 102599; 102600, (01) 07332414124359 (21) 102600; 102601, (01) 07332414124359 (21) 102601; 102602, (01) 07332414124359 (21) 102602; 102603, (01) 07332414124359 (21) 102603; 102604, (01) 07332414124359 (21) 102604; 102605, (01) 07332414124359 (21) 102605; 102606, (01) 07332414124359 (21) 102606; 102607, (01) 07332414124359 (21) 102607; 102608, (01) 07332414124359 (21) 102608; 102609, (01) 07332414124359 (21) 102609; 102610, (01) 07332414124359 (21) 102610; 102611, (01) 07332414124359 (21) 102611; 102612, (01) 07332414124359 (21) 102612; 102613, (01) 07332414124359 (21) 102613; 102614, (01) 07332414124359 (21) 102614; 102615, (01) 07332414124359 (21) 102615; 102616, (01) 07332414124359 (21) 102616; 102617, (01) 07332414124359 (21) 102617; 102618, (01) 07332414124359 (21) 102618; 102619, (01) 07332414124359 (21) 102619; 102620, (01) 07332414124359 (21) 102620; 102621, (01) 07332414124359 (21) 102621; 102637, (01) 07332414124359 (21) 102637; 102638, (01) 07332414124359 (21) 102638; 102639, (01) 07332414124359 (21) 102639; 102640, (01) 07332414124359 (21) 102640; 102641, (01) 07332414124359 (21) 102641; 102642, (01) 07332414124359 (21) 102642; 102643, (01) 07332414124359 (21) 102643; 102644, (01) 07332414124359 (21) 102644; 102645, (01) 07332414124359 (21) 102645; 102646, (01) 07332414124359 (21) 102646; 102647, (01) 07332414124359 (21) 102647; 102648, (01) 07332414124359 (21) 102648; 102649, (01) 07332414124359 (21) 102649; 102650, (01) 07332414124359 (21) 102650; 102651, (01) 07332414124359 (21) 102651; 102652, (01) 07332414124359 (21) 102652; 102653, (01) 07332414124359 (21) 102653; 102654, (01) 07332414124359 (21) 102654; 102655, (01) 07332414124359 (21) 102655; 102656, (01) 07332414124359 (21) 102656; 102657, (01) 07332414124359 (21) 102657; 102658, (01) 07332414124359 (21) 102658; 102659, (01) 07332414124359 (21) 102659; 102660, (01) 07332414124359 (21) 102660; 102661, (01) 07332414124359 (21) 102661; 102662, (01) 07332414124359 (21) 102662; 102663, (01) 07332414124359 (21) 102663; 102664, (01) 07332414124359 (21) 102664; 102665, (01) 07332414124359 (21) 102665; 102666, (01) 07332414124359 (21) 102666; 102706, (01) 07332414124359 (21) 102706; 102707, (01) 07332414124359 (21) 102707; 102708, (01) 07332414124359 (21) 102708; 102709, (01) 07332414124359 (21) 102709; 102710, (01) 07332414124359 (21) 102710; 102711, (01) 07332414124359 (21) 102711; 102712, (01) 07332414124359 (21) 102712; 102713, (01) 07332414124359 (21) 102713; 102714, (01) 07332414124359 (21) 102714; 102715, (01) 07332414124359 (21) 102715; 102716, (01) 07332414124359 (21) 102716; 102717, (01) 07332414124359 (21) 102717; 102718, (01) 07332414124359 (21) 102718; 102719, (01) 07332414124359 (21) 102719; 102720, (01) 07332414124359 (21) 102720; 102721, (01) 07332414124359 (21) 102721; 102722, (01) 07332414124359 (21) 102722; 102723, (01) 07332414124359 (21) 102723; 102724, (01) 07332414124359 (21) 102724; 102725, (01) 07332414124359 (21) 102725; 102726, (01) 07332414124359 (21) 102726; 102727, (01) 07332414124359 (21) 102727; 102728, (01) 07332414124359 (21) 102728; 102729, (01) 07332414124359 (21) 102729; 102730, (01) 07332414124359 (21) 102730; 102731, (01) 07332414124359 (21) 102731; 102732, (01) 07332414124359 (21) 102732; 102733, (01) 07332414124359 (21) 102733; 102734, (01) 07332414124359 (21) 102734; 102735, (01) 07332414124359 (21) 102735; 102736, (01) 07332414124359 (21) 102736; 102737, (01) 07332414124359 (21) 102737; 102738, (01) 07332414124359 (21) 102738; 102739, (01) 07332414124359 (21) 102739; 102740, (01) 07332414124359 (21) 102740; 102761, (01) 07332414124359 (21) 102761; 102762, (01) 07332414124359 (21) 102762; 102763, (01) 07332414124359 (21) 102763; 102764, (01) 07332414124359 (21) 102764; 102765, (01) 07332414124359 (21) 102765; 102766, (01) 07332414124359 (21) 102766; 102767, (01) 07332414124359 (21) 102767; 102768, (01) 07332414124359 (21) 102768; 102769, (01) 07332414124359 (21) 102769; 102770, (01) 07332414124359 (21) 102770; 102771, (01) 07332414124359 (21) 102771; 102772, (01) 07332414124359 (21) 102772; 102773, (01) 07332414124359 (21) 102773; 102774, (01) 07332414124359 (21) 102774; 102775, (01) 07332414124359 (21) 102775; 102776, (01) 07332414124359 (21) 102776; 102777, (01) 07332414124359 (21) 102777; 102778, (01) 07332414124359 (21) 102778; 102779, (01) 07332414124359 (21) 102779; 102780, (01) 07332414124359 (21) 102780; 102821, (01) 07332414124359 (21) 102821; 102822, (01) 07332414124359 (21) 102822; 102823, (01) 07332414124359 (21) 102823; 102824, (01) 07332414124359 (21) 102824; 102825, (01) 07332414124359 (21) 102825; 102826, (01) 07332414124359 (21) 102826; 102827, (01) 07332414124359 (21) 102827; 102828, (01) 07332414124359 (21) 102828; 102829, (01) 07332414124359 (21) 102829; 102830, (01) 07332414124359 (21) 102830; 102846, (01) 07332414124359 (21) 102846; 102847, (01) 07332414124359 (21) 102847; 102848, (01) 07332414124359 (21) 102848; 102849, (01) 07332414124359 (21) 102849; 102850, (01) 07332414124359 (21) 102850; 102851, (01) 07332414124359 (21) 102851; 102852, (01) 07332414124359 (21) 102852; 102853, (01) 07332414124359 (21) 102853; 102854, (01) 07332414124359 (21) 102854; 102855, (01) 07332414124359 (21) 102855; 102856, (01) 07332414124359 (21) 102856; 102857, (01) 07332414124359 (21) 102857; 102858, (01) 07332414124359 (21) 102858; 102859, (01) 07332414124359 (21) 102859; 102860, (01) 07332414124359 (21) 102860; 102861, (01) 07332414124359 (21) 102861; 102862, (01) 07332414124359 (21) 102862; 102863, (01) 07332414124359 (21) 102863; 102864, (01) 07332414124359 (21) 102864; 102865, (01) 07332414124359 (21) 102865; 102866, (01) 07332414124359 (21) 102866; 102867, (01) 07332414124359 (21) 102867; 102868, (01) 07332414124359 (21) 102868; 102869, (01) 07332414124359 (21) 102869; 102870, (01) 07332414124359 (21) 102870; 102871, (01) 07332414124359 (21) 102871; 102872, (01) 07332414124359 (21) 102872; 102873, (01) 07332414124359 (21) 102873; 102874, (01) 07332414124359 (21) 102874; 102875, (01) 07332414124359 (21) 102875; 102971, (01) 07332414124359 (21) 102971; 102972, (01) 07332414124359 (21) 102972; 102973, (01) 07332414124359 (21) 102973; 102974, (01) 07332414124359 (21) 102974; 102975, (01) 07332414124359 (21) 102975; 102976, (01) 07332414124359 (21) 102976; 102977, (01) 07332414124359 (21) 102977; 102978, (01) 07332414124359 (21) 102978; 102979, (01) 07332414124359 (21) 102979; 102980, (01) 07332414124359 (21) 102980; 102981, (01) 07332414124359 (21) 102981; 102982, (01) 07332414124359 (21) 102982; 102983, (01) 07332414124359 (21) 102983; 102984, (01) 07332414124359 (21) 102984; 102985, (01) 07332414124359 (21) 102985; 102986, (01) 07332414124359 (21) 102986; 102987, (01) 07332414124359 (21) 102987; 102988, (01) 07332414124359 (21) 102988; 102989, (01) 07332414124359 (21) 102989; 102990, (01) 07332414124359 (21) 102990; 102991, (01) 07332414124359 (21) 102991; 102992, (01) 07332414124359 (21) 102992; 102993, (01) 07332414124359 (21) 102993; 102994, (01) 07332414124359 (21) 102994; 102995, (01) 07332414124359 (21) 102995; 102996, (01) 07332414124359 (21) 102996; 102997, (01) 07332414124359 (21) 102997; 102998, (01) 07332414124359 (21) 102998; 102999, (01) 07332414124359 (21) 102999; 103000, (01) 07332414124359 (21) 103000; 103001, (01) 07332414124359 (21) 103001; 103002, (01) 07332414124359 (21) 103002; 103003, (01) 07332414124359 (21) 103003; 103004, (01) 07332414124359 (21) 103004; 103005, (01) 07332414124359 (21) 103005; 103006, (01) 07332414124359 (21) 103006; 103007, (01) 07332414124359 (21) 103007; 103008, (01) 07332414124359 (21) 103008; 103009, (01) 07332414124359 (21) 103009; 103010, (01) 07332414124359 (21) 103010; 103011, (01) 07332414124359 (21) 103011; 103012, (01) 07332414124359 (21) 103012; 103013, (01) 07332414124359 (21) 103013; 103014, (01) 07332414124359 (21) 103014; 103015, (01) 07332414124359 (21) 103015; 103036, (01) 07332414124359 (21) 103036; 103037, (01) 07332414124359 (21) 103037; 103038, (01) 07332414124359 (21) 103038; 103039, (01) 07332414124359 (21) 103039; 103040, (01) 07332414124359 (21) 103040; 103041, (01) 07332414124359 (21) 103041; 103042, (01) 07332414124359 (21) 103042; 103043, (01) 07332414124359 (21) 103043; 103044, (01) 07332414124359 (21) 103044; 103045, (01) 07332414124359 (21) 103045; 103046, (01) 07332414124359 (21) 103046; 103047, (01) 07332414124359 (21) 103047; 103048, (01) 07332414124359 (21) 103048; 103049, (01) 07332414124359 (21) 103049; 103050, (01) 07332414124359 (21) 103050; 103051, (01) 07332414124359 (21) 103051; 103052, (01) 07332414124359 (21) 103052; 103053, (01) 07332414124359 (21) 103053; 103054, (01) 07332414124359 (21) 103054; 103055, (01) 07332414124359 (21) 103055; 103056, (01) 07332414124359 (21) 103056; 103057, (01) 07332414124359 (21) 103057; 103058, (01) 07332414124359 (21) 103058; 103059, (01) 07332414124359 (21) 103059; 103060, (01) 07332414124359 (21) 103060; 103061, (01) 07332414124359 (21) 103061; 103062, (01) 07332414124359 (21) 103062; 103063, (01) 07332414124359 (21) 103063; 103064, (01) 07332414124359 (21) 103064; 103065, (01) 07332414124359 (21) 103065; 103066, (01) 07332414124359 (21) 103066; 103067, (01) 07332414124359 (21) 103067; 103068, (01) 07332414124359 (21) 103068; 103069, (01) 07332414124359 (21) 103069; 103070, (01) 07332414124359 (21) 103070; 103111, (01) 07332414124359 (21) 103111; 103112, (01) 07332414124359 (21) 103112; 103113, (01) 07332414124359 (21) 103113; 103114, (01) 07332414124359 (21) 103114; 103115, (01) 07332414124359 (21) 103115; 103116, (01) 07332414124359 (21) 103116; 103117, (01) 07332414124359 (21) 103117; 103118, (01) 07332414124359 (21) 103118; 103119, (01) 07332414124359 (21) 103119; 103120, (01) 07332414124359 (21) 103120; 103121, (01) 07332414124359 (21) 103121; 103122, (01) 07332414124359 (21) 103122; 103123, (01) 07332414124359 (21) 103123; 103124, (01) 07332414124359 (21) 103124; 103125, (01) 07332414124359 (21) 103125; 103146, (01) 07332414124359 (21) 103146; 103147, (01) 07332414124359 (21) 103147; 103148, (01) 07332414124359 (21) 103148; 103149, (01) 07332414124359 (21) 103149; 103150, (01) 07332414124359 (21) 103150; 103151, (01) 07332414124359 (21) 103151; 103152, (01) 07332414124359 (21) 103152; 103153, (01) 07332414124359 (21) 103153; 103154, (01) 07332414124359 (21) 103154; 103155, (01) 07332414124359 (21) 103155; 103156, (01) 07332414124359 (21) 103156; 103157, (01) 07332414124359 (21) 103157; 103158, (01) 07332414124359 (21) 103158; 103159, (01) 07332414124359 (21) 103159; 103160, (01) 07332414124359 (21) 103160; 103161, (01) 07332414124359 (21) 103161; 103162, (01) 07332414124359 (21) 103162; 103163, (01) 07332414124359 (21) 103163; 103164, (01) 07332414124359 (21) 103164; 103165, (01) 07332414124359 (21) 103165; 103166, (01) 07332414124359 (21) 103166; 103167, (01) 07332414124359 (21) 103167; 103168, (01) 07332414124359 (21) 103168; 103169, (01) 07332414124359 (21) 103169; 103170, (01) 07332414124359 (21) 103170; 103171, (01) 07332414124359 (21) 103171; 103172, (01) 07332414124359 (21) 103172; 103173, (01) 07332414124359 (21) 103173; 103174, (01) 07332414124359 (21) 103174; 103175, (01) 07332414124359 (21) 103175; 103236, (01) 07332414124359 (21) 103236; 103237, (01) 07332414124359 (21) 103237; 103238, (01) 07332414124359 (21) 103238; 103239, (01) 07332414124359 (21) 103239; 103240, (01) 07332414124359 (21) 103240; 103241, (01) 07332414124359 (21) 103241; 103242, (01) 07332414124359 (21) 103242; 103243, (01) 07332414124359 (21) 103243; 103244, (01) 07332414124359 (21) 103244; 103245, (01) 07332414124359 (21) 103245; 103246, (01) 07332414124359 (21) 103246; 103247, (01) 07332414124359 (21) 103247; 103248, (01) 07332414124359 (21) 103248; 103249, (01) 07332414124359 (21) 103249; 103250, (01) 07332414124359 (21) 103250; 103251, (01) 07332414124359 (21) 103251; 103252, (01) 07332414124359 (21) 103252; 103253, (01) 07332414124359 (21) 103253; 103254, (01) 07332414124359 (21) 103254; 103255, (01) 07332414124359 (21) 103255; 103256, (01) 07332414124359 (21) 103256; 103257, (01) 07332414124359 (21) 103257; 103258, (01) 07332414124359 (21) 103258; 103259, (01) 07332414124359 (21) 103259; 103260, (01) 07332414124359 (21) 103260; 103261, (01) 07332414124359 (21) 103261; 103262, (01) 07332414124359 (21) 103262; 103263, (01) 07332414124359 (21) 103263; 103264, (01) 07332414124359 (21) 103264; 103265, (01) 07332414124359 (21) 103265; 103266, (01) 07332414124359 (21) 103266; 103267, (01) 07332414124359 (21) 103267; 103268, (01) 07332414124359 (21) 103268; 103269, (01) 07332414124359 (21) 103269; 103270, (01) 07332414124359 (21) 103270; 103271, (01) 07332414124359 (21) 103271; 103272, (01) 07332414124359 (21) 103272; 103273, (01) 07332414124359 (21) 103273; 103274, (01) 07332414124359 (21) 103274; 103275, (01) 07332414124359 (21) 103275; 103276, (01) 07332414124359 (21) 103276; 103277, (01) 07332414124359 (21) 103277; 103278, (01) 07332414124359 (21) 103278; 103279, (01) 07332414124359 (21) 103279; 103280, (01) 07332414124359 (21) 103280; 103281, (01) 07332414124359 (21) 103281; 103282, (01) 07332414124359 (21) 103282; 103283, (01) 07332414124359 (21) 103283; 103284, (01) 07332414124359 (21) 103284; 103285, (01) 07332414124359 (21) 103285; 103296, (01) 07332414124359 (21) 103296; 103297, (01) 07332414124359 (21) 103297; 103298, (01) 07332414124359 (21) 103298; 103299, (01) 07332414124359 (21) 103299; 103300, (01) 07332414124359 (21) 103300; 103301, (01) 07332414124359 (21) 103301; 103302, (01) 07332414124359 (21) 103302; 103303, (01) 07332414124359 (21) 103303; 103304, (01) 07332414124359 (21) 103304; 103305, (01) 07332414124359 (21) 103305; 103306, (01) 07332414124359 (21) 103306; 103307, (01) 07332414124359 (21) 103307; 103308, (01) 07332414124359 (21) 103308; 103309, (01) 07332414124359 (21) 103309; 103310, (01) 07332414124359 (21) 103310; 103311, (01) 07332414124359 (21) 103311; 103312, (01) 07332414124359 (21) 103312; 103313, (01) 07332414124359 (21) 103313; 103314, (01) 07332414124359 (21) 103314; 103315, (01) 07332414124359 (21) 103315; 103316, (01) 07332414124359 (21) 103316; 103317, (01) 07332414124359 (21) 103317; 103318, (01) 07332414124359 (21) 103318; 103319, (01) 07332414124359 (21) 103319; 103320, (01) 07332414124359 (21) 103320; 103321, (01) 07332414124359 (21) 103321; 103322, (01) 07332414124359 (21) 103322; 103323, (01) 07332414124359 (21) 103323; 103324, (01) 07332414124359 (21) 103324; 103325, (01) 07332414124359 (21) 103325; 103376, (01) 07332414124359 (21) 103376; 103377, (01) 07332414124359 (21) 103377; 103378, (01) 07332414124359 (21) 103378; 103379, (01) 07332414124359 (21) 103379; 103380, (01) 07332414124359 (21) 103380; 103381, (01) 07332414124359 (21) 103381; 103382, (01) 07332414124359 (21) 103382; 103383, (01) 07332414124359 (21) 103383; 103384, (01) 07332414124359 (21) 103384; 103385, (01) 07332414124359 (21) 103385; 103386, (01) 07332414124359 (21) 103386; 103387, (01) 07332414124359 (21) 103387; 103388, (01) 07332414124359 (21) 103388; 103389, (01) 07332414124359 (21) 103389; 103390, (01) 07332414124359 (21) 103390; 103391, (01) 07332414124359 (21) 103391; 103392, (01) 07332414124359 (21) 103392; 103393, (01) 07332414124359 (21) 103393; 103394, (01) 07332414124359 (21) 103394; 103395, (01) 07332414124359 (21) 103395; 103436, (01) 07332414124359 (21) 103436; 103437, (01) 07332414124359 (21) 103437; 103438, (01) 07332414124359 (21) 103438; 103439, (01) 07332414124359 (21) 103439; 103440, (01) 07332414124359 (21) 103440; 103441, (01) 07332414124359 (21) 103441; 103442, (01) 07332414124359 (21) 103442; 103443, (01) 07332414124359 (21) 103443; 103444, (01) 07332414124359 (21) 103444; 103445, (01) 07332414124359 (21) 103445; 103486, (01) 07332414124359 (21) 103486; 103487, (01) 07332414124359 (21) 103487; 103488, (01) 07332414124359 (21) 103488; 103489, (01) 07332414124359 (21) 103489; 103490, (01) 07332414124359 (21) 103490; 103491, (01) 07332414124359 (21) 103491; 103492, (01) 07332414124359 (21) 103492; 103493, (01) 07332414124359 (21) 103493; 103494, (01) 07332414124359 (21) 103494; 103495, (01) 07332414124359 (21) 103495; 103496, (01) 07332414124359 (21) 103496; 103497, (01) 07332414124359 (21) 103497; 103498, (01) 07332414124359 (21) 103498; 103499, (01) 07332414124359 (21) 103499; 103500, (01) 07332414124359 (21) 103500; 103501, (01) 07332414124359 (21) 103501; 103502, (01) 07332414124359 (21) 103502; 103503, (01) 07332414124359 (21) 103503; 103504, (01) 07332414124359 (21) 103504; 103505, (01) 07332414124359 (21) 103505; 103506, (01) 07332414124359 (21) 103506; 103507, (01) 07332414124359 (21) 103507; 103508, (01) 07332414124359 (21) 103508; 103509, (01) 07332414124359 (21) 103509; 103510, (01) 07332414124359 (21) 103510; 103511, (01) 07332414124359 (21) 103511; 103512, (01) 07332414124359 (21) 103512; 103513, (01) 07332414124359 (21) 103513; 103514, (01) 07332414124359 (21) 103514; 103515, (01) 07332414124359 (21) 103515; 103516, (01) 07332414124359 (21) 103516; 103517, (01) 07332414124359 (21) 103517; 103518, (01) 07332414124359 (21) 103518; 103519, (01) 07332414124359 (21) 103519; 103520, (01) 07332414124359 (21) 103520; 103521, (01) 07332414124359 (21) 103521; 103522, (01) 07332414124359 (21) 103522; 103523, (01) 07332414124359 (21) 103523; 103524, (01) 07332414124359 (21) 103524; 103525, (01) 07332414124359 (21) 103525; 103526, (01) 07332414124359 (21) 103526; 103527, (01) 07332414124359 (21) 103527; 103528, (01) 07332414124359 (21) 103528; 103529, (01) 07332414124359 (21) 103529; 103530, (01) 07332414124359 (21) 103530; 103531, (01) 07332414124359 (21) 103531; 103532, (01) 07332414124359 (21) 103532; 103533, (01) 07332414124359 (21) 103533; 103534, (01) 07332414124359 (21) 103534; 103535, (01) 07332414124359 (21) 103535; 103586, (01) 07332414124359 (21) 103586; 103587, (01) 07332414124359 (21) 103587; 103588, (01) 07332414124359 (21) 103588; 103589, (01) 07332414124359 (21) 103589; 103590, (01) 07332414124359 (21) 103590; 103591, (01) 07332414124359 (21) 103591; 103592, (01) 07332414124359 (21) 103592; 103593, (01) 07332414124359 (21) 103593; 103594, (01) 07332414124359 (21) 103594; 103595, (01) 07332414124359 (21) 103595; 103596, (01) 07332414124359 (21) 103596; 103597, (01) 07332414124359 (21) 103597; 103598, (01) 07332414124359 (21) 103598; 103599, (01) 07332414124359 (21) 103599; 103600, (01) 07332414124359 (21) 103600; 103601, (01) 07332414124359 (21) 103601; 103602, (01) 07332414124359 (21) 103602; 103603, (01) 07332414124359 (21) 103603; 103604, (01) 07332414124359 (21) 103604; 103605, (01) 07332414124359 (21) 103605; 103616, (01) 07332414124359 (21) 103616; 103617, (01) 07332414124359 (21) 103617; 103618, (01) 07332414124359 (21) 103618; 103619, (01) 07332414124359 (21) 103619; 103620, (01) 07332414124359 (21) 103620; 103621, (01) 07332414124359 (21) 103621; 103622, (01) 07332414124359 (21) 103622; 103623, (01) 07332414124359 (21) 103623; 103624, (01) 07332414124359 (21) 103624; 103625, (01) 07332414124359 (21) 103625; 103636, (01) 07332414124359 (21) 103636; 103637, (01) 07332414124359 (21) 103637; 103638, (01) 07332414124359 (21) 103638; 103639, (01) 07332414124359 (21) 103639; 103640, (01) 07332414124359 (21) 103640; 103641, (01) 07332414124359 (21) 103641; 103642, (01) 07332414124359 (21) 103642; 103643, (01) 07332414124359 (21) 103643; 103644, (01) 07332414124359 (21) 103644; 103645, (01) 07332414124359 (21) 103645; 103656, (01) 07332414124359 (21) 103656; 103657, (01) 07332414124359 (21) 103657; 103658, (01) 07332414124359 (21) 103658; 103659, (01) 07332414124359 (21) 103659; 103660, (01) 07332414124359 (21) 103660; 103661, (01) 07332414124359 (21) 103661; 103662, (01) 07332414124359 (21) 103662; 103663, (01) 07332414124359 (21) 103663; 103664, (01) 07332414124359 (21) 103664; 103665, (01) 07332414124359 (21) 103665; 103676, (01) 07332414124359 (21) 103676; 103677, (01) 07332414124359 (21) 103677; 103678, (01) 07332414124359 (21) 103678; 103679, (01) 07332414124359 (21) 103679; 103680, (01) 07332414124359 (21) 103680; 103681, (01) 07332414124359 (21) 103681; 103682, (01) 07332414124359 (21) 103682; 103683, (01) 07332414124359 (21) 103683; 103684, (01) 07332414124359 (21) 103684; 103685, (01) 07332414124359 (21) 103685; 103756, (01) 07332414124359 (21) 103756; 103757, (01) 07332414124359 (21) 103757; 103758, (01) 07332414124359 (21) 103758; 103759, (01) 07332414124359 (21) 103759; 103760, (01) 07332414124359 (21) 103760; 103761, (01) 07332414124359 (21) 103761; 103762, (01) 07332414124359 (21) 103762; 103763, (01) 07332414124359 (21) 103763; 103764, (01) 07332414124359 (21) 103764; 103765, (01) 07332414124359 (21) 103765; 103776, (01) 07332414124359 (21) 103776; 103777, (01) 07332414124359 (21) 103777; 103778, (01) 07332414124359 (21) 103778; 103779, (01) 07332414124359 (21) 103779; 103780, (01) 07332414124359 (21) 103780; 103781, (01) 07332414124359 (21) 103781; 103782, (01) 07332414124359 (21) 103782; 103783, (01) 07332414124359 (21) 103783; 103784, (01) 07332414124359 (21) 103784; 103785, (01) 07332414124359 (21) 103785; 103796, (01) 07332414124359 (21) 103796; 103797, (01) 07332414124359 (21) 103797; 103798, (01) 07332414124359 (21) 103798; 103799, (01) 07332414124359 (21) 103799; 103800, (01) 07332414124359 (21) 103800; 103801, (01) 07332414124359 (21) 103801; 103802, (01) 07332414124359 (21) 103802; 103803, (01) 07332414124359 (21) 103803; 103804, (01) 07332414124359 (21) 103804; 103805, (01) 07332414124359 (21) 103805; 103816, (01) 07332414124359 (21) 103816; 103817, (01) 07332414124359 (21) 103817; 103818, (01) 07332414124359 (21) 103818; 103819, (01) 07332414124359 (21) 103819; 103820, (01) 07332414124359 (21) 103820; 103821, (01) 07332414124359 (21) 103821; 103822, (01) 07332414124359 (21) 103822; 103823, (01) 07332414124359 (21) 103823; 103824, (01) 07332414124359 (21) 103824; 103825, (01) 07332414124359 (21) 103825; 103836, (01) 07332414124359 (21) 103836; 103837, (01) 07332414124359 (21) 103837; 103838, (01) 07332414124359 (21) 103838; 103839, (01) 07332414124359 (21) 103839; 103840, (01) 07332414124359 (21) 103840; 103841, (01) 07332414124359 (21) 103841; 103842, (01) 07332414124359 (21) 103842; 103843, (01) 07332414124359 (21) 103843; 103844, (01) 07332414124359 (21) 103844; 103845, (01) 07332414124359 (21) 103845; 103996, (01) 07332414124359 (21) 103996; 103997, (01) 07332414124359 (21) 103997; 103998, (01) 07332414124359 (21) 103998; 103999, (01) 07332414124359 (21) 103999; 104000, (01) 07332414124359 (21) 104000; 104001, (01) 07332414124359 (21) 104001; 104002, (01) 07332414124359 (21) 104002; 104003, (01) 07332414124359 (21) 104003; 104004, (01) 07332414124359 (21) 104004; 104005, (01) 07332414124359 (21) 104005; 104016, (01) 07332414124359 (21) 104016; 104017, (01) 07332414124359 (21) 104017; 104018, (01) 07332414124359 (21) 104018; 104019, (01) 07332414124359 (21) 104019; 104020, (01) 07332414124359 (21) 104020; 104021, (01) 07332414124359 (21) 104021; 104022, (01) 07332414124359 (21) 104022; 104023, (01) 07332414124359 (21) 104023; 104024, (01) 07332414124359 (21) 104024; 104025, (01) 07332414124359 (21) 104025; 104036, (01) 07332414124359 (21) 104036; 104037, (01) 07332414124359 (21) 104037; 104038, (01) 07332414124359 (21) 104038; 104039, (01) 07332414124359 (21) 104039; 104040, (01) 07332414124359 (21) 104040; 104041, (01) 07332414124359 (21) 104041; 104042, (01) 07332414124359 (21) 104042; 104043, (01) 07332414124359 (21) 104043; 104044, (01) 07332414124359 (21) 104044; 104045, (01) 07332414124359 (21) 104045; 104056, (01) 07332414124359 (21) 104056; 104057, (01) 07332414124359 (21) 104057; 104058, (01) 07332414124359 (21) 104058; 104059, (01) 07332414124359 (21) 104059; 104060, (01) 07332414124359 (21) 104060; 104061, (01) 07332414124359 (21) 104061; 104062, (01) 07332414124359 (21) 104062; 104063, (01) 07332414124359 (21) 104063; 104064, (01) 07332414124359 (21) 104064; 104065, (01) 07332414124359 (21) 104065; 104266, (01) 07332414124359 (21) 104266; 104267, (01) 07332414124359 (21) 104267; 104268, (01) 07332414124359 (21) 104268; 104269, (01) 07332414124359 (21) 104269; 104270, (01) 07332414124359 (21) 104270; 104271, (01) 07332414124359 (21) 104271; 104272, (01) 07332414124359 (21) 104272; 104273, (01) 07332414124359 (21) 104273; 104274, (01) 07332414124359 (21) 104274; 104275, (01) 07332414124359 (21) 104275; 104286, (01) 07332414124359 (21) 104286; 104287, (01) 07332414124359 (21) 104287; 104288, (01) 07332414124359 (21) 104288; 104289, (01) 07332414124359 (21) 104289; 104290, (01) 07332414124359 (21) 104290; 104291, (01) 07332414124359 (21) 104291; 104292, (01) 07332414124359 (21) 104292; 104293, (01) 07332414124359 (21) 104293; 104294, (01) 07332414124359 (21) 104294; 104295, (01) 07332414124359 (21) 104295; 104306, (01) 07332414124359 (21) 104306; 104307, (01) 07332414124359 (21) 104307; 104308, (01) 07332414124359 (21) 104308; 104309, (01) 07332414124359 (21) 104309; 104310, (01) 07332414124359 (21) 104310; 104311, (01) 07332414124359 (21) 104311; 104312, (01) 07332414124359 (21) 104312; 104313, (01) 07332414124359 (21) 104313; 104314, (01) 07332414124359 (21) 104314; 104315, (01) 07332414124359 (21) 104315; 104396, (01) 07332414124359 (21) 104396; 104397, (01) 07332414124359 (21) 104397; 104398, (01) 07332414124359 (21) 104398; 104399, (01) 07332414124359 (21) 104399; 104400, (01) 07332414124359 (21) 104400; 104401, (01) 07332414124359 (21) 104401; 104402, (01) 07332414124359 (21) 104402; 104403, (01) 07332414124359 (21) 104403; 104404, (01) 07332414124359 (21) 104404; 104405, (01) 07332414124359 (21) 104405; 104436, (01) 07332414124359 (21) 104436; 104437, (01) 07332414124359 (21) 104437; 104438, (01) 07332414124359 (21) 104438; 104439, (01) 07332414124359 (21) 104439; 104440, (01) 07332414124359 (21) 104440; 104441, (01) 07332414124359 (21) 104441; 104442, (01) 07332414124359 (21) 104442; 104443, (01) 07332414124359 (21) 104443; 104444, (01) 07332414124359 (21) 104444; 104445, (01) 07332414124359 (21) 104445; 104496, (01) 07332414124359 (21) 104496; 104497, (01) 07332414124359 (21) 104497; 104498, (01) 07332414124359 (21) 104498; 104499, (01) 07332414124359 (21) 104499; 104500, (01) 07332414124359 (21) 104500; 104501, (01) 07332414124359 (21) 104501; 104502, (01) 07332414124359 (21) 104502; 104503, (01) 07332414124359 (21) 104503; 104504, (01) 07332414124359 (21) 104504; 104505, (01) 07332414124359 (21) 104505; 104546, (01) 07332414124359 (21) 104546; 104547, (01) 07332414124359 (21) 104547; 104548, (01) 07332414124359 (21) 104548; 104549, (01) 07332414124359 (21) 104549; 104550, (01) 07332414124359 (21) 104550; 104551, (01) 07332414124359 (21) 104551; 104552, (01) 07332414124359 (21) 104552; 104553, (01) 07332414124359 (21) 104553; 104554, (01) 07332414124359 (21) 104554; 104555, (01) 07332414124359 (21) 104555; 104576, (01) 07332414124359 (21) 104576; 104577, (01) 07332414124359 (21) 104577; 104578, (01) 07332414124359 (21) 104578; 104579, (01) 07332414124359 (21) 104579; 104580, (01) 07332414124359 (21) 104580; 104581, (01) 07332414124359 (21) 104581; 104582, (01) 07332414124359 (21) 104582; 104583, (01) 07332414124359 (21) 104583; 104584, (01) 07332414124359 (21) 104584; 104585, (01) 07332414124359 (21) 104585; 104596, (01) 07332414124359 (21) 104596; 104597, (01) 07332414124359 (21) 104597; 104598, (01) 07332414124359 (21) 104598; 104599, (01) 07332414124359 (21) 104599; 104600, (01) 07332414124359 (21) 104600; 104601, (01) 07332414124359 (21) 104601; 104602, (01) 07332414124359 (21) 104602; 104603, (01) 07332414124359 (21) 104603; 104604, (01) 07332414124359 (21) 104604; 104605, (01) 07332414124359 (21) 104605; 104626, (01) 07332414124359 (21) 104626; 104627, (01) 07332414124359 (21) 104627; 104628, (01) 07332414124359 (21) 104628; 104629, (01) 07332414124359 (21) 104629; 104630, (01) 07332414124359 (21) 104630; 104631, (01) 07332414124359 (21) 104631; 104632, (01) 07332414124359 (21) 104632; 104633, (01) 07332414124359 (21) 104633; 104634, (01) 07332414124359 (21) 104634; 104635, (01) 07332414124359 (21) 104635; 104766, (01) 07332414124359 (21) 104766; 104767, (01) 07332414124359 (21) 104767; 104768, (01) 07332414124359 (21) 104768; 104769, (01) 07332414124359 (21) 104769; 104770, (01) 07332414124359 (21) 104770; 104771, (01) 07332414124359 (21) 104771; 104772, (01) 07332414124359 (21) 104772; 104773, (01) 07332414124359 (21) 104773; 104774, (01) 07332414124359 (21) 104774; 104775, (01) 07332414124359 (21) 104775; 104786, (01) 07332414124359 (21) 104786; 104787, (01) 07332414124359 (21) 104787; 104788, (01) 07332414124359 (21) 104788; 104789, (01) 07332414124359 (21) 104789; 104790, (01) 07332414124359 (21) 104790; 104791, (01) 07332414124359 (21) 104791; 104792, (01) 07332414124359 (21) 104792; 104793, (01) 07332414124359 (21) 104793; 104794, (01) 07332414124359 (21) 104794; 104795, (01) 07332414124359 (21) 104795; 104796, (01) 07332414124359 (21) 104796; 104797, (01) 07332414124359 (21) 104797; 104798, (01) 07332414124359 (21) 104798; 104799, (01) 07332414124359 (21) 104799; 104800, (01) 07332414124359 (21) 104800; 104801, (01) 07332414124359 (21) 104801; 104802, (01) 07332414124359 (21) 104802; 104803, (01) 07332414124359 (21) 104803; 104804, (01) 07332414124359 (21) 104804; 104805, (01) 07332414124359 (21) 104805; 104806, (01) 07332414124359 (21) 104806; 104807, (01) 07332414124359 (21) 104807; 104808, (01) 07332414124359 (21) 104808; 104809, (01) 07332414124359 (21) 104809; 104810, (01) 07332414124359 (21) 104810; 104811, (01) 07332414124359 (21) 104811; 104812, (01) 07332414124359 (21) 104812; 104813, (01) 07332414124359 (21) 104813; 104814, (01) 07332414124359 (21) 104814; 104815, (01) 07332414124359 (21) 104815; 104826, (01) 07332414124359 (21) 104826; 104827, (01) 07332414124359 (21) 104827; 104828, (01) 07332414124359 (21) 104828; 104829, (01) 07332414124359 (21) 104829; 104830, (01) 07332414124359 (21) 104830; 104831, (01) 07332414124359 (21) 104831; 104832, (01) 07332414124359 (21) 104832; 104833, (01) 07332414124359 (21) 104833; 104834, (01) 07332414124359 (21) 104834; 104835, (01) 07332414124359 (21) 104835; 104906, (01) 07332414124359 (21) 104906; 104907, (01) 07332414124359 (21) 104907; 104908, (01) 07332414124359 (21) 104908; 104909, (01) 07332414124359 (21) 104909; 104910, (01) 07332414124359 (21) 104910; 104911, (01) 07332414124359 (21) 104911; 104912, (01) 07332414124359 (21) 104912; 104913, (01) 07332414124359 (21) 104913; 104914, (01) 07332414124359 (21) 104914; 104915, (01) 07332414124359 (21) 104915; 104936, (01) 07332414124359 (21) 104936; 104937, (01) 07332414124359 (21) 104937; 104938, (01) 07332414124359 (21) 104938; 104939, (01) 07332414124359 (21) 104939; 104940, (01) 07332414124359 (21) 104940; 104941, (01) 07332414124359 (21) 104941; 104942, (01) 07332414124359 (21) 104942; 104943, (01) 07332414124359 (21) 104943; 104944, (01) 07332414124359 (21) 104944; 104945, (01) 07332414124359 (21) 104945; 104966, (01) 07332414124359 (21) 104966; 104967, (01) 07332414124359 (21) 104967; 104968, (01) 07332414124359 (21) 104968; 104969, (01) 07332414124359 (21) 104969; 104970, (01) 07332414124359 (21) 104970; 104971, (01) 07332414124359 (21) 104971; 104972, (01) 07332414124359 (21) 104972; 104973, (01) 07332414124359 (21) 104973; 104974, (01) 07332414124359 (21) 104974; 104975, (01) 07332414124359 (21) 104975; 104986, (01) 07332414124359 (21) 104986; 104987, (01) 07332414124359 (21) 104987; 104988, (01) 07332414124359 (21) 104988; 104989, (01) 07332414124359 (21) 104989; 104990, (01) 07332414124359 (21) 104990; 104991, (01) 07332414124359 (21) 104991; 104992, (01) 07332414124359 (21) 104992; 104993, (01) 07332414124359 (21) 104993; 104994, (01) 07332414124359 (21) 104994; 104995, (01) 07332414124359 (21) 104995; 105056, (01) 07332414124359 (21) 105056; 105057, (01) 07332414124359 (21) 105057; 105058, (01) 07332414124359 (21) 105058; 105059, (01) 07332414124359 (21) 105059; 105060, (01) 07332414124359 (21) 105060; 105061, (01) 07332414124359 (21) 105061; 105062, (01) 07332414124359 (21) 105062; 105063, (01) 07332414124359 (21) 105063; 105064, (01) 07332414124359 (21) 105064; 105065, (01) 07332414124359 (21) 105065; 105076, (01) 07332414124359 (21) 105076; 105077, (01) 07332414124359 (21) 105077; 105079, (01) 07332414124359 (21) 105079; 105080, (01) 07332414124359 (21) 105080; 105081, (01) 07332414124359 (21) 105081; 105082, (01) 07332414124359 (21) 105082; 105083, (01) 07332414124359 (21) 105083; 105084, (01) 07332414124359 (21) 105084; 105085, (01) 07332414124359 (21) 105085; 105086, (01) 07332414124359 (21) 105086; 105087, (01) 07332414124359 (21) 105087; 105088, (01) 07332414124359 (21) 105088; 105089, (01) 07332414124359 (21) 105089; 105090, (01) 07332414124359 (21) 105090; 105091, (01) 07332414124359 (21) 105091; 105092, (01) 07332414124359 (21) 105092; 105093, (01) 07332414124359 (21) 105093; 105094, (01) 07332414124359 (21) 105094; 105095, (01) 07332414124359 (21) 105095; 105106, (01) 07332414124359 (21) 105106; 105107, (01) 07332414124359 (21) 105107; 105108, (01) 07332414124359 (21) 105108; 105109, (01) 07332414124359 (21) 105109; 105110, (01) 07332414124359 (21) 105110; 105111, (01) 07332414124359 (21) 105111; 105112, (01) 07332414124359 (21) 105112; 105113, (01) 07332414124359 (21) 105113; 105114, (01) 07332414124359 (21) 105114; 105115, (01) 07332414124359 (21) 105115; 105126, (01) 07332414124359 (21) 105126; 105127, (01) 07332414124359 (21) 105127; 105128, (01) 07332414124359 (21) 105128; 105129, (01) 07332414124359 (21) 105129; 105130, (01) 07332414124359 (21) 105130; 105131, (01) 07332414124359 (21) 105131; 105132, (01) 07332414124359 (21) 105132; 105133, (01) 07332414124359 (21) 105133; 105134, (01) 07332414124359 (21) 105134; 105135, (01) 07332414124359 (21) 105135; 105146, (01) 07332414124359 (21) 105146; 105147, (01) 07332414124359 (21) 105147; 105148, (01) 07332414124359 (21) 105148; 105149, (01) 07332414124359 (21) 105149; 105150, (01) 07332414124359 (21) 105150; 105151, (01) 07332414124359 (21) 105151; 105152, (01) 07332414124359 (21) 105152; 105153, (01) 07332414124359 (21) 105153; 105154, (01) 07332414124359 (21) 105154; 105155, (01) 07332414124359 (21) 105155; 105156, (01) 07332414124359 (21) 105156; 105157, (01) 07332414124359 (21) 105157; 105158, (01) 07332414124359 (21) 105158; 105159, (01) 07332414124359 (21) 105159; 105160, (01) 07332414124359 (21) 105160; 105161, (01) 07332414124359 (21) 105161; 105162, (01) 07332414124359 (21) 105162; 105163, (01) 07332414124359 (21) 105163; 105164, (01) 07332414124359 (21) 105164; 105165, (01) 07332414124359 (21) 105165; 105176, (01) 07332414124359 (21) 105176; 105177, (01) 07332414124359 (21) 105177; 105178, (01) 07332414124359 (21) 105178; 105179, (01) 07332414124359 (21) 105179; 105180, (01) 07332414124359 (21) 105180; 105181, (01) 07332414124359 (21) 105181; 105182, (01) 07332414124359 (21) 105182; 105183, (01) 07332414124359 (21) 105183; 105184, (01) 07332414124359 (21) 105184; 105185, (01) 07332414124359 (21) 105185; 105206, (01) 07332414124359 (21) 105206; 105207, (01) 07332414124359 (21) 105207; 105208, (01) 07332414124359 (21) 105208; 105209, (01) 07332414124359 (21) 105209; 105210, (01) 07332414124359 (21) 105210; 105211, (01) 07332414124359 (21) 105211; 105212, (01) 07332414124359 (21) 105212; 105213, (01) 07332414124359 (21) 105213; 105214, (01) 07332414124359 (21) 105214; 105215, (01) 07332414124359 (21) 105215; 105226, (01) 07332414124359 (21) 105226; 105227, (01) 07332414124359 (21) 105227; 105228, (01) 07332414124359 (21) 105228; 105229, (01) 07332414124359 (21) 105229; 105230, (01) 07332414124359 (21) 105230; 105231, (01) 07332414124359 (21) 105231; 105232, (01) 07332414124359 (21) 105232; 105233, (01) 07332414124359 (21) 105233; 105234, (01) 07332414124359 (21) 105234; 105235, (01) 07332414124359 (21) 105235; 105236, (01) 07332414124359 (21) 105236; 105237, (01) 07332414124359 (21) 105237; 105238, (01) 07332414124359 (21) 105238; 105239, (01) 07332414124359 (21) 105239; 105240, (01) 07332414124359 (21) 105240; 105241, (01) 07332414124359 (21) 105241; 105242, (01) 07332414124359 (21) 105242; 105243, (01) 07332414124359 (21) 105243; 105244, (01) 07332414124359 (21) 105244; 105245, (01) 07332414124359 (21) 105245; 105286, (01) 07332414124359 (21) 105286; 105287, (01) 07332414124359 (21) 105287; 105288, (01) 07332414124359 (21) 105288; 105289, (01) 07332414124359 (21) 105289; 105290, (01) 07332414124359 (21) 105290; 105291, (01) 07332414124359 (21) 105291; 105292, (01) 07332414124359 (21) 105292; 105293, (01) 07332414124359 (21) 105293; 105294, (01) 07332414124359 (21) 105294; 105295, (01) 07332414124359 (21) 105295; 105306, (01) 07332414124359 (21) 105306; 105307, (01) 07332414124359 (21) 105307; 105308, (01) 07332414124359 (21) 105308; 105309, (01) 07332414124359 (21) 105309; 105310, (01) 07332414124359 (21) 105310; 105311, (01) 07332414124359 (21) 105311; 105312, (01) 07332414124359 (21) 105312; 105313, (01) 07332414124359 (21) 105313; 105314, (01) 07332414124359 (21) 105314; 105315, (01) 07332414124359 (21) 105315; 105316, (01) 07332414124359 (21) 105316; 105317, (01) 07332414124359 (21) 105317; 105318, (01) 07332414124359 (21) 105318; 105319, (01) 07332414124359 (21) 105319; 105320, (01) 07332414124359 (21) 105320; 105321, (01) 07332414124359 (21) 105321; 105322, (01) 07332414124359 (21) 105322; 105323, (01) 07332414124359 (21) 105323; 105324, (01) 07332414124359 (21) 105324; 105325, (01) 07332414124359 (21) 105325; 105336, (01) 07332414124359 (21) 105336; 105337, (01) 07332414124359 (21) 105337; 105338, (01) 07332414124359 (21) 105338; 105339, (01) 07332414124359 (21) 105339; 105340, (01) 07332414124359 (21) 105340; 105341, (01) 07332414124359 (21) 105341; 105342, (01) 07332414124359 (21) 105342; 105343, (01) 07332414124359 (21) 105343; 105344, (01) 07332414124359 (21) 105344; 105345, (01) 07332414124359 (21) 105345; 105356, (01) 07332414124359 (21) 105356; 105357, (01) 07332414124359 (21) 105357; 105358, (01) 07332414124359 (21) 105358; 105359, (01) 07332414124359 (21) 105359; 105360, (01) 07332414124359 (21) 105360; 105361, (01) 07332414124359 (21) 105361; 105362, (01) 07332414124359 (21) 105362; 105363, (01) 07332414124359 (21) 105363; 105364, (01) 07332414124359 (21) 105364; 105365, (01) 07332414124359 (21) 105365; 105366, (01) 07332414124359 (21) 105366; 105367, (01) 07332414124359 (21) 105367; 105368, (01) 07332414124359 (21) 105368; 105369, (01) 07332414124359 (21) 105369; 105370, (01) 07332414124359 (21) 105370; 105371, (01) 07332414124359 (21) 105371; 105372, (01) 07332414124359 (21) 105372; 105373, (01) 07332414124359 (21) 105373; 105374, (01) 07332414124359 (21) 105374; 105375, (01) 07332414124359 (21) 105375; 105376, (01) 07332414124359 (21) 105376; 105378, (01) 07332414124359 (21) 105378; 105379, (01) 07332414124359 (21) 105379; 105380, (01) 07332414124359 (21) 105380; 105381, (01) 07332414124359 (21) 105381; 105382, (01) 07332414124359 (21) 105382; 105383, (01) 07332414124359 (21) 105383; 105384, (01) 07332414124359 (21) 105384; 105385, (01) 07332414124359 (21) 105385; 105386, (01) 07332414124359 (21) 105386; 105387, (01) 07332414124359 (21) 105387; 105388, (01) 07332414124359 (21) 105388; 105389, (01) 07332414124359 (21) 105389; 105390, (01) 07332414124359 (21) 105390; 105391, (01) 07332414124359 (21) 105391; 105392, (01) 07332414124359 (21) 105392; 105393, (01) 07332414124359 (21) 105393; 105394, (01) 07332414124359 (21) 105394; 105395, (01) 07332414124359 (21) 105395; 105396, (01) 07332414124359 (21) 105396; 105397, (01) 07332414124359 (21) 105397; 105398, (01) 07332414124359 (21) 105398; 105399, (01) 07332414124359 (21) 105399; 105400, (01) 07332414124359 (21) 105400; 105401, (01) 07332414124359 (21) 105401; 105402, (01) 07332414124359 (21) 105402; 105403, (01) 07332414124359 (21) 105403; 105404, (01) 07332414124359 (21) 105404; 105405, (01) 07332414124359 (21) 105405; 105406, (01) 07332414124359 (21) 105406; 105407, (01) 07332414124359 (21) 105407; 105408, (01) 07332414124359 (21) 105408; 105409, (01) 07332414124359 (21) 105409; 105410, (01) 07332414124359 (21) 105410; 105411, (01) 07332414124359 (21) 105411; 105412, (01) 07332414124359 (21) 105412; 105413, (01) 07332414124359 (21) 105413; 105414, (01) 07332414124359 (21) 105414; 105415, (01) 07332414124359 (21) 105415; 105426, (01) 07332414124359 (21) 105426; 105427, (01) 07332414124359 (21) 105427; 105428, (01) 07332414124359 (21) 105428; 105429, (01) 07332414124359 (21) 105429; 105430, (01) 07332414124359 (21) 105430; 105431, (01) 07332414124359 (21) 105431; 105432, (01) 07332414124359 (21) 105432; 105433, (01) 07332414124359 (21) 105433; 105434, (01) 07332414124359 (21) 105434; 105435, (01) 07332414124359 (21) 105435; 105446, (01) 07332414124359 (21) 105446; 105447, (01) 07332414124359 (21) 105447; 105448, (01) 07332414124359 (21) 105448; 105449, (01) 07332414124359 (21) 105449; 105450, (01) 07332414124359 (21) 105450; 105451, (01) 07332414124359 (21) 105451; 105452, (01) 07332414124359 (21) 105452; 105453, (01) 07332414124359 (21) 105453; 105454, (01) 07332414124359 (21) 105454; 105455, (01) 07332414124359 (21) 105455; 105546, (01) 07332414124359 (21) 105546; 105547, (01) 07332414124359 (21) 105547; 105548, (01) 07332414124359 (21) 105548; 105549, (01) 07332414124359 (21) 105549; 105550, (01) 07332414124359 (21) 105550; 105551, (01) 07332414124359 (21) 105551; 105552, (01) 07332414124359 (21) 105552; 105553, (01) 07332414124359 (21) 105553; 105554, (01) 07332414124359 (21) 105554; 105555, (01) 07332414124359 (21) 105555; 105566, (01) 07332414124359 (21) 105566; 105567, (01) 07332414124359 (21) 105567; 105568, (01) 07332414124359 (21) 105568; 105569, (01) 07332414124359 (21) 105569; 105570, (01) 07332414124359 (21) 105570; 105571, (01) 07332414124359 (21) 105571; 105572, (01) 07332414124359 (21) 105572; 105573, (01) 07332414124359 (21) 105573; 105574, (01) 07332414124359 (21) 105574; 105575, (01) 07332414124359 (21) 105575; 105586, (01) 07332414124359 (21) 105586; 105587, (01) 07332414124359 (21) 105587; 105588, (01) 07332414124359 (21) 105588; 105589, (01) 07332414124359 (21) 105589; 105590, (01) 07332414124359 (21) 105590; 105591, (01) 07332414124359 (21) 105591; 105592, (01) 07332414124359 (21) 105592; 105593, (01) 07332414124359 (21) 105593; 105594, (01) 07332414124359 (21) 105594; 105595, (01) 07332414124359 (21) 105595; 105606, (01) 07332414124359 (21) 105606; 105607, (01) 07332414124359 (21) 105607; 105608, (01) 07332414124359 (21) 105608; 105609, (01) 07332414124359 (21) 105609; 105610, (01) 07332414124359 (21) 105610; 105611, (01) 07332414124359 (21) 105611; 105612, (01) 07332414124359 (21) 105612; 105613, (01) 07332414124359 (21) 105613; 105614, (01) 07332414124359 (21) 105614; 105615, (01) 07332414124359 (21) 105615; 105712, (01) 07332414124359 (21) 105712; 105713, (01) 07332414124359 (21) 105713; 105714, (01) 07332414124359 (21) 105714; 105715, (01) 07332414124359 (21) 105715; 105716, (01) 07332414124359 (21) 105716; 105717, (01) 07332414124359 (21) 105717; 105718, (01) 07332414124359 (21) 105718; 105719, (01) 07332414124359 (21) 105719; 105720, (01) 07332414124359 (21) 105720; 105721, (01) 07332414124359 (21) 105721; 105722, (01) 07332414124359 (21) 105722; 105723, (01) 07332414124359 (21) 105723; 105724, (01) 07332414124359 (21) 105724; 105725, (01) 07332414124359 (21) 105725; 105726, (01) 07332414124359 (21) 105726; 105727, (01) 07332414124359 (21) 105727; 105728, (01) 07332414124359 (21) 105728; 105729, (01) 07332414124359 (21) 105729; 105730, (01) 07332414124359 (21) 105730; 105731, (01) 07332414124359 (21) 105731; 105742, (01) 07332414124359 (21) 105742; 105743, (01) 07332414124359 (21) 105743; 105744, (01) 07332414124359 (21) 105744; 105745, (01) 07332414124359 (21) 105745; 105746, (01) 07332414124359 (21) 105746; 105747, (01) 07332414124359 (21) 105747; 105748, (01) 07332414124359 (21) 105748; 105749, (01) 07332414124359 (21) 105749; 105750, (01) 07332414124359 (21) 105750; 105751, (01) 07332414124359 (21) 105751; 105752, (01) 07332414124359 (21) 105752; 105753, (01) 07332414124359 (21) 105753; 105754, (01) 07332414124359 (21) 105754; 105755, (01) 07332414124359 (21) 105755; 105756, (01) 07332414124359 (21) 105756; 105757, (01) 07332414124359 (21) 105757; 105758, (01) 07332414124359 (21) 105758; 105759, (01) 07332414124359 (21) 105759; 105760, (01) 07332414124359 (21) 105760; 105761, (01) 07332414124359 (21) 105761; 105792, (01) 07332414124359 (21) 105792; 105793, (01) 07332414124359 (21) 105793; 105794, (01) 07332414124359 (21) 105794; 105795, (01) 07332414124359 (21) 105795; 105796, (01) 07332414124359 (21) 105796; 105797, (01) 07332414124359 (21) 105797; 105798, (01) 07332414124359 (21) 105798; 105799, (01) 07332414124359 (21) 105799; 105800, (01) 07332414124359 (21) 105800; 105801, (01) 07332414124359 (21) 105801; 105832, (01) 07332414124359 (21) 105832; 105833, (01) 07332414124359 (21) 105833; 105834, (01) 07332414124359 (21) 105834; 105835, (01) 07332414124359 (21) 105835; 105836, (01) 07332414124359 (21) 105836; 105837, (01) 07332414124359 (21) 105837; 105838, (01) 07332414124359 (21) 105838; 105839, (01) 07332414124359 (21) 105839; 105840, (01) 07332414124359 (21) 105840; 105841, (01) 07332414124359 (21) 105841; 105842, (01) 07332414124359 (21) 105842; 105843, (01) 07332414124359 (21) 105843; 105844, (01) 07332414124359 (21) 105844; 105845, (01) 07332414124359 (21) 105845; 105846, (01) 07332414124359 (21) 105846; 105847, (01) 07332414124359 (21) 105847; 105848, (01) 07332414124359 (21) 105848; 105849, (01) 07332414124359 (21) 105849; 105850, (01) 07332414124359 (21) 105850; 105851, (01) 07332414124359 (21) 105851; 105882, (01) 07332414124359 (21) 105882; 105883, (01) 07332414124359 (21) 105883; 105884, (01) 07332414124359 (21) 105884; 105885, (01) 07332414124359 (21) 105885; 105886, (01) 07332414124359 (21) 105886; 105887, (01) 07332414124359 (21) 105887; 105889, (01) 07332414124359 (21) 105889; 105890, (01) 07332414124359 (21) 105890; 105891, (01) 07332414124359 (21) 105891; 105892, (01) 07332414124359 (21) 105892; 105893, (01) 07332414124359 (21) 105893; 105894, (01) 07332414124359 (21) 105894; 105895, (01) 07332414124359 (21) 105895; 105896, (01) 07332414124359 (21) 105896; 105897, (01) 07332414124359 (21) 105897; 105898, (01) 07332414124359 (21) 105898; 105899, (01) 07332414124359 (21) 105899; 105900, (01) 07332414124359 (21) 105900; 105901, (01) 07332414124359 (21) 105901; 105902, (01) 07332414124359 (21) 105902; 105903, (01) 07332414124359 (21) 105903; 105904, (01) 07332414124359 (21) 105904; 105905, (01) 07332414124359 (21) 105905; 105906, (01) 07332414124359 (21) 105906; 105907, (01) 07332414124359 (21) 105907; 105908, (01) 07332414124359 (21) 105908; 105909, (01) 07332414124359 (21) 105909; 105910, (01) 07332414124359 (21) 105910; 105911, (01) 07332414124359 (21) 105911; 105932, (01) 07332414124359 (21) 105932; 105933, (01) 07332414124359 (21) 105933; 105934, (01) 07332414124359 (21) 105934; 105935, (01) 07332414124359 (21) 105935; 105936, (01) 07332414124359 (21) 105936; 105938, (01) 07332414124359 (21) 105938; 105939, (01) 07332414124359 (21) 105939; 105940, (01) 07332414124359 (21) 105940; 105941, (01) 07332414124359 (21) 105941; 105942, (01) 07332414124359 (21) 105942; 105943, (01) 07332414124359 (21) 105943; 105944, (01) 07332414124359 (21) 105944; 105945, (01) 07332414124359 (21) 105945; 105946, (01) 07332414124359 (21) 105946; 105948, (01) 07332414124359 (21) 105948; 105949, (01) 07332414124359 (21) 105949; 105950, (01) 07332414124359 (21) 105950; 105951, (01) 07332414124359 (21) 105951; 105992, (01) 07332414124359 (21) 105992; 105993, (01) 07332414124359 (21) 105993; 105994, (01) 07332414124359 (21) 105994; 105995, (01) 07332414124359 (21) 105995; 105996, (01) 07332414124359 (21) 105996; 105997, (01) 07332414124359 (21) 105997; 105998, (01) 07332414124359 (21) 105998; 105999, (01) 07332414124359 (21) 105999; 106000, (01) 07332414124359 (21) 106000; 106001, (01) 07332414124359 (21) 106001; 106002, (01) 07332414124359 (21) 106002; 106003, (01) 07332414124359 (21) 106003; 106004, (01) 07332414124359 (21) 106004; 106005, (01) 07332414124359 (21) 106005; 106006, (01) 07332414124359 (21) 106006; 106007, (01) 07332414124359 (21) 106007; 106008, (01) 07332414124359 (21) 106008; 106009, (01) 07332414124359 (21) 106009; 106010, (01) 07332414124359 (21) 106010; 106011, (01) 07332414124359 (21) 106011; 106012, (01) 07332414124359 (21) 106012; 106013, (01) 07332414124359 (21) 106013; 106014, (01) 07332414124359 (21) 106014; 106015, (01) 07332414124359 (21) 106015; 106016, (01) 07332414124359 (21) 106016; 106017, (01) 07332414124359 (21) 106017; 106018, (01) 07332414124359 (21) 106018; 106020, (01) 07332414124359 (21) 106020; 106021, (01) 07332414124359 (21) 106021; 106052, (01) 07332414124359 (21) 106052; 106053, (01) 07332414124359 (21) 106053; 106054, (01) 07332414124359 (21) 106054; 106055, (01) 07332414124359 (21) 106055; 106056, (01) 07332414124359 (21) 106056; 106057, (01) 07332414124359 (21) 106057; 106058, (01) 07332414124359 (21) 106058; 106059, (01) 07332414124359 (21) 106059; 106060, (01) 07332414124359 (21) 106060; 106061, (01) 07332414124359 (21) 106061; 106122, (01) 07332414124359 (21) 106122; 106123, (01) 07332414124359 (21) 106123; 106125, (01) 07332414124359 (21) 106125; 106126, (01) 07332414124359 (21) 106126; 106127, (01) 07332414124359 (21) 106127; 106129, (01) 07332414124359 (21) 106129; 106130, (01) 07332414124359 (21) 106130; 106131, (01) 07332414124359 (21) 106131; 106132, (01) 07332414124359 (21) 106132; 106133, (01) 07332414124359 (21) 106133; 106134, (01) 07332414124359 (21) 106134; 106135, (01) 07332414124359 (21) 106135; 106137, (01) 07332414124359 (21) 106137; 106138, (01) 07332414124359 (21) 106138; 106140, (01) 07332414124359 (21) 106140; 106141, (01) 07332414124359 (21) 106141; 106143, (01) 07332414124359 (21) 106143; 106144, (01) 07332414124359 (21) 106144; 106146, (01) 07332414124359 (21) 106146; 106147, (01) 07332414124359 (21) 106147; 106148, (01) 07332414124359 (21) 106148; 106149, (01) 07332414124359 (21) 106149; 106150, (01) 07332414124359 (21) 106150; 106151, (01) 07332414124359 (21) 106151; 106202, (01) 07332414124359 (21) 106202; 106203, (01) 07332414124359 (21) 106203; 106204, (01) 07332414124359 (21) 106204; 106206, (01) 07332414124359 (21) 106206; 106208, (01) 07332414124359 (21) 106208; 106210, (01) 07332414124359 (21) 106210; 106220, (01) 07332414124359 (21) 106220 (1683 units) c) 955701: 104446, (01) 07332414125844 (21) 104446; 104447, (01) 07332414125844 (21) 104447; 104448, (01) 07332414125844 (21) 104448; 104449, (01) 07332414125844 (21) 104449; 104450, (01) 07332414125844 (21) 104450; 104451, (01) 07332414125844 (21) 104451; 104452, (01) 07332414125844 (21) 104452; 104453, (01) 07332414125844 (21) 104453; 104454, (01) 07332414125844 (21) 104454; 104455, (01) 07332414125844 (21) 104455; 104466, (01) 07332414125844 (21) 104466; 104467, (01) 07332414125844 (21) 104467; 104468, (01) 07332414125844 (21) 104468; 104469, (01) 07332414125844 (21) 104469; 104470, (01) 07332414125844 (21) 104470; 104471, (01) 07332414125844 (21) 104471; 104472, (01) 07332414125844 (21) 104472; 104473, (01) 07332414125844 (21) 104473; 104474, (01) 07332414125844 (21) 104474; 104475, (01) 07332414125844 (21) 104475; 105646, (01) 07332414125844 (21) 105646; 105647, (01) 07332414125844 (21) 105647; 105648, (01) 07332414125844 (21) 105648; 105649, (01) 07332414125844 (21) 105649; 105650, (01) 07332414125844 (21) 105650; 105651, (01) 07332414125844 (21) 105651; 105652, (01) 07332414125844 (21) 105652; 105653, (01) 07332414125844 (21) 105653 (28 units) Product Code SC6169: Lot Numbers: 1286902, 1286903, 1286871, 1286870, 1286868, 1286869, 1286623, 1278425, 1281024, 1278909, 1278423, 1278424, 1277964, 1277966, 1277965, 1277743, 1276586, 1277559, 1275517 Product Code SC6035, Lot Numbers: 1287266, 1275298, 1275299 Product Code SC6053, Lot Numbers: 1277956, 1278671 |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
For Additional Information Contact |
Center for One Baxter 800-422-9837
|
Manufacturer Reason for Recall |
Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance. Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Baxter sent An Urgent Medical Device Correction letter dated January 7, 2021. To prevent potential alarm situations, the affected tubing in the ARPS Pump Assembly for the PrisMax devices will be replaced with improved tubing. Operators may continue to safely use the PrisMax System until the tubing is replaced within the ARPS Pump Assembly. If an alarm occurs, the PrisMax System will default to a safe state and the user should follow the on-screen instructions to end treatment and to contact service. A local Baxter service representative will contact each facility to determine the correction plan and schedule the replacement for the impacted devices only. Facilities involved will be receiving this replacement from Baxter at no charge. If you received this communication directly from Baxter, complete the enclosed Baxter Customer Reply Form and return it to Baxter by e-mailing it to fca@baxter.com. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the reply form. For general questions regarding this communication, contact Baxter Corporate |
Quantity in Commerce |
2314 units |
Distribution |
Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Israel, Italy, Japan, Jordan, Kenya, Korea (Republic Of), Kuwait, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Spain, Sweden, United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KDI and Original Applicant = Baxter Healthcare Corp
|
|
|
|