Class 2 Device Recall BD Syringe Set 20 DP MIC TBG 0.2MF CKV AntiSiphon SmartSite VLV Sensing Disc |
|
Date Initiated by Firm |
August 02, 2021 |
Create Date |
September 13, 2021 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2457-2021 |
Recall Event ID |
88203 |
510(K)Number |
K811885
|
Product Classification |
Set, administration, intravascular - Product Code FPA
|
Product |
BD Syringe Set 20 DP MIC TBG 0.2MF CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798696; BD Syringe Set 20 DP MIC TBG CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798703;
|
Code Information |
Lots: 20026497 20025590 20076421 21015406 21015407 |
Recalling Firm/ Manufacturer |
CAREFUSION 10020 Pacific Mesa Blvd San Diego CA 92121-4386
|
Manufacturer Reason for Recall |
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
|
FDA Determined Cause 2 |
Process control |
Action |
On 08/02/21, Urgent Medical Device Correction notices were mailed and emailed to consignees who were asked to notify recalling firm if this issue was experienced at 1-844-823-5433, productcomplaints@bd.com . Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device. If during priming the clinician encounters an occluded needle-free connector, take the following actions: 1) Depress the syringe plunger while the syringe is still attached to the needle-free connector. 2) If unable to depress the plunger, pull back on the syringe plunger and depress the syringe plunger again. Repeating this step may assist in opening the needle-free connector valve. 3) If the valve still is not open; without disconnecting the syringe male luer tip from the BD SmartSite valve surface, carefully unthread and rethread the syringe to partially disengage and reengage the syringe. Repeating this step up to three times may assist in opening the needle-free valve. Note: To maintain sterility, do not allow the syringe and needle-free valve to completely disconnect. 4) If previous steps do not open the valve, then open a new needle-free connector to replace the occluded connector. 5) After the steps above, proceed with the existing instructions for use. 6) Share this correction notification with all users within your facility network of the product to ensure they are also aware of this correction. 7) Complete and return the Customer Response Form. Distributors were asked to identify customers within their distribution network that purchased the affected product and to email an excel file listing of all customers to the recalling firm within 72 hours of receipt of this letter so the recalling firm can initiate customer notification. See online tool to support the identification of impacted lot numbers https://www.bd.com/MDS-21-4144 |
Quantity in Commerce |
16,700 |
Distribution |
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = FPA and Original Applicant = Y
|
|
|
|