Date Initiated by Firm |
July 26, 2022 |
Date Posted |
September 13, 2022 |
Recall Status1 |
Terminated 3 on May 31, 2024 |
Recall Number |
Z-1739-2022 |
Recall Event ID |
90739 |
510(K)Number |
K120005
|
Product Classification |
Dialysate concentrate for hemodialysis (liquid or powder) - Product Code KPO
|
Product |
Acidified Concentrate Distribution and Storage-The Isopure ACDS (Acidified Concentrate Distribution and Storage) with remote fill is intended to be used in Hemodialysis facilities for the storage and distribution of acid concentrate to be used in the treatment of Hemodialysis patients
Model: ACDS MD704262 |
Code Information |
Unique Device Identifier (UDI-DI) - 00852456007092
Serial Numbers:
AC-869226
AC-869232
AC-869223
AC-867759
AC-867306
AC-862958
AC-862507
AC-867765
AC-866732
AC-865085
AC-857327
AC-825341
AC-02041204
AC-828320
AC-843375
AC-866726
AC-121340
AC21041212
AC-835763
AC-847755
AC-119356
AC-864675
AC-867939
AC-852751
AC-118472
AC06241504
AC-844137
AC-830206
AC-123056
AC-857333
AC-123169
AC-124590
AC-117928
AC-845510
AC-841313
AC-851496
AC-851502
AC-849420
AC-846640
AC-848995
AC-861609
AC-114598
N/A
AC-117041
AC-867933
AC-825253
AC-834163
AC-849423
AC-854222
AC-854213
AC03041204
AC-846217
AC853485
AC-852090
AC-841600
AC-853559
AC-838743
AC-836202
AC-852745
AC-862513
AC-853730
AC-853535
AC-113494
AC-123880
AC-847434
AC-860585
AC-842280
AC-847758
AC-849828
AC10301410
AC-848634
AC-116620
AC-860205
AC-118659
AC-839632
AC-847177
AC-825308
AC-861274
AC-837654
AC-831676
AC-857330
AC-866735
AC-860588
AC-863236
AC-863239
AC-850162
AC-828331
AC-848358
AC-847888
AC12291504
AC-836714
AC-836724
AC-864696
AC-844134
AC-852535
AC-115347
AC-113451
AC122440
AC-117799
AC14311205
AC-121318
AC-857992
AC-853529
AC-848253
AC-837561
AC-113027
AC-866165
AC-841319
AC-843858
AC-124395
AC-846631
AC-860241
AC-121277
AC-859415
AC-837874
AC-842723
AC114761
AC-119466
AC-866035
AC-116609
AC846655
AC-845308
AC-121233
AC-855936
AC07271504
AC-854886
AC-839635
AC-120802
AC-853547
AC-121266
AC-834167
AC-840902
AC-115839
AC-850044
AC-124634
AC-855688
AC-840896
AC-843867
AC01061504
AC-116415
AC-119433
AC-830862
AC-869208
AC-846664
AC-858010
AC-838173
AC-843324
AC-837419
AC--852733
AC-854880
AC-838143
AC-849825
AC-121255
AC-838740
AC- 850159
AC-835829
AC-123891
AC-843372
AC-830688
AC-123147
AC-847891
AC-118181
AC-853538
AC-848361
AC-856367
AC-844970
AC02091504
AC-838020
AC-117342
AC-864687
AC-858844
AC-857048
AC-857060
AC-828689
AC-843139
AC-122217
AC-862976
AC-864678
AC-123847
AC-846634
AC-829065
AC03051311
AC-118980
AC07061406
AC-841074
AC-834162
AC-124235
AC-843369
AC-843360
AC-869217
AC-117052
AC-837664
AC-838050
AC-122765
AC-860253
AC-844425
AC-859840
AC-122743
AC07031205
AC09111205
AC10281504
AC-854228
AC-862210
AC03101404
AC-836744
AC-862522
AC-852541
AC-115794
AC-851508
AC-867954
AC-848640
AC-862222
AC-830802
AC-867300
AC-846658
AC-836209
AC-113554
AC-864699
AC16181206
AC-844976
AC-842283
AC-854216
AC-842143
AC-116271
AC-835851
AC-861615
AC-839310
AC-839626
AC-865549
AC-865546
AC-849209
AC-838749
AC-841310
AC-834171
AC-856370
AC-840412
AC-856361
AC-825275
AC04241504
AC-862973
AC01121402
AC-851212
AC-121307
AC-830812
AC-836223
AC-841594
AC-848998
AC-836216
AC-834645
AC-835752
AC-844972
AC-842017
AC-841080
AC-841304
AC-825264
AC-867957
AC-115726
AC-829240
AC-858013
AC-118569
AC-121755
AC-125017
AC-116282
AC-862219
AC-858004
AC-867630
AC-861277
AC-867945
AC05271403
AC-830842
AC-859834
AC-857066
AC-854895
AC-839212
AC-123924
AC-862961
AC-839484
AC-120791
AC-847183
AC08271504
AC-867942
AC-867948
AC-846637
AC-851505
AC-843363
AC05051204
AC-834651
AC-123274
AC-830188
AC-861603
AC-119444
AC-867309
AC-866162
AC-114609
AC-853482
AC-857054
AC-855697
AC-852748
N/A
AC-117144
AC-835840
AC-852095
AC-857051
AC-852544
AC-835741
AC-835785
AC-853553
AC-846661
AC-843061
AC-843870
AC-845010
AC-843096
AC-845516
AC-869235
AC-844422
AC-862531
AC-862534
AC-859837
AC-854892
AC-828700
AC-849160
AC-864693
AC-869211
AC-869205
AC-869229
AC-867303
AC-862516
AC-757936
AC-866168
AC-865555
AC-857324
AC-828342
AC-828353
AC-843861
AC-866038
AC-121753
AC22041212
AC-836764
AC-847761
AC-118580
AC-764672
AC-852736
AC-123902
AC-123935
AC-117331
AC-845513
AC-841316
AC-851499
AC-849426
AC-846649
AC-849157
AC-861606
AC-114772
AC-117155
AC-867930
AC-830197
AC-849831
AC-853556
AC04041204
AC-846214
AC853523
AC-841597
AC-853544
AC-837571
AC-852739
AC-862528
AC-853194
AC-854219
AC114909
AC-123957
AC-847437
AC-860579
AC-842726
AC-847752
AC-850038
AC11301410
AC-848643
AC-116342
AC-860208
AC-118670
AC-839629
AC-847180
AC-825330
AC-860582
AC-855691
AC-867621
AC-860576
AC-862949
AC-862964
AC-850047
AC-825286
AC-848355
AC-847882
AC13301504
AC-836754
AC-836734
AC-862967
AC-845317
AC-852100
AC-115737
AC-117133
AC-122776
AC15311205
AC-121244
AC-853532
AC-847885
AC-114586
AC-866729
AC-843855
AC-825242
AC-846646
AC-860250
AC-121288
AC-841825
AC-837864
AC-842729
AC114574
AC-119455
AC-865543
AC-117069
N/A
AC-845311
AC-121764
AC11291504
AC-120822
AC-854889
AC-834161
AC-841077
AC-115850
AC-850041
AC-825319
AC-855700
AC-840905
AC-843864
AC-119477
AC-8231720
AC-867771
AC-847186
AC-838040
AC-843142
AC-853200
AC-854883
AC-838153
AC-849834
AC-838260
AC- 848370
AC-835796
AC-123913
AC-843357
AC-830677
AC-123968
AC-848256
AC-117917
AC-853526
AC-848367
AC-845016
AC03201504
AC-838030
AC-117535
AC-864681
AC-858841
AC-855694
AC-829073
AC-83188
AC- 117546
AC-862952
AC-863233
AC-124579
AC-846652
AC-829057
AC04051311
AC-118991
AC-841071
AC-834168
AC-123092
AC-843354
AC-843366
AC-117166
AC-838270
AC-124236
AC-857911
AC-845314
AC-859831
AC-122754
AC08031205
AC10111205
AC14301504
AC-854225
AC-862216
AC04101404
AC-837425
AC-862510
AC-115358
AC-850598
AC-867774
AC-848637
AC-861612
AC-830872
AC-846643
AC-864684
AC17181206
AC-844968
AC-842720
AC-853541
AC-842140
AC-835818
AC-862225
AC-839313
AC-840409
AC-865552
AC-849206
AC-834188
AC-838746
AC-856364
AC-825363
AC05241504
AC-863242
AC02121402
AC-830832
AC-836195
AC-841819
AC-836188
AC-834638
AC-835730
AC-844974
AC-841822
AC-840899
AC-841307
AC-825297
AC-867951
AC115783
AC-829229
AC-857995
AC-122774
AC-124601
AC-116404
AC-862213
AC-858001
AC-867624
AC-865088
AC-861594
AC-831698
AC-857063
AC-839319
AC-124428
AC-862955
AC-839487
AC-120813
AC09281504
AC-867762
AC-867768
AC-846211
AC-848364
AC06051204
AC-834658
AC-123136
AC-861600
AC-119488
AC-867486
AC-866723
AC-116353
AC-853197
AC-857045
AC-852742
AC-118170
AC-835807
AC-852085
AC-857057
AC-852538
AC-835774
AC-853550
AC-846220
AC-843099
AC-844131
AC-845019
AC-845519
AC-844140
AC-862519
AC-862525
AC-860244
AC-856085
AC-864690
|
Recalling Firm/ Manufacturer |
Isopure Corp 11851 Plantside Dr Louisville KY 40299-6328
|
For Additional Information Contact |
Tracy Hardin 502-907-6652
|
Manufacturer Reason for Recall |
Acidified Concentrate Distribution System potential electrical safety hazard may result in electric shock, arching
|
FDA Determined Cause 2 |
Device Design |
Action |
Isopure initiated contact via telephone on 7/26/22 followed by Emails to consignees. .Letter states reason for recall, health risk and action to take:
1. Refer to Distribution List of ACDS units installed.
2. Perform an inspection of the ACDS device, associated wiring, cabling and cord. Provide a photograph if damage is found during inspection.
3. Video links are included in the inspection packet of the Isopure ACDS Instructions for Electrical Review and Preventive Maintenance of the ACDS and proper use of a torque screwdriver as a guide to inspecting electrical connections WI-50602. Please follow these guidance documents. Adhere to all safety information provided.
4. Fill out the Acknowledgement and Inspection Form upon completion of the inspection. Return the Acknowledgment form back to Isopure to Quality@isopure.com by September 30, 2022.
5. Always replace any depleted or blown fuses with the correct type and correct rating.
6. Follow the recommended preventive maintenance schedule for the device (refer to LA704262 ACDS Acidified Concentrate Distribution and Storage System User Manual).
7. Contact Isopure Customer/Technical Support at 502-967-7873 with any questions or concerns.
Refer to: How -To-Video- Instructions ACDS Preventive Maintenance Instructions
https://vimeo.com/732224767
|
Quantity in Commerce |
595 units |
Distribution |
US Nationwide distribution.
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPO and Original Applicant = ISOPURE CORP.
|