|
Class 2 Device Recall CardioMEMSTM HF System Patient Electronics System (PES) |
|
Date Initiated by Firm |
February 07, 2023 |
Date Posted |
March 03, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1219-2023 |
Recall Event ID |
91610 |
PMA Number |
P100045 |
Product Classification |
System, hemodynamic, implantable - Product Code MOM
|
Product |
CardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment. |
Code Information |
UDI/DI 05414734509176, Serial Numbers: M140300439, M140300440, M140300441, M140300445, M140300446, M140300447, M140300449, M140300450, M140300451, M140300454, M140300458, M140300459, M140300460, M140300463, M140500480, M140500486, M140800987, M140800988, M140800992, M140800995, M140801002, M140801003, M140801004, M140801007, M140801010, M140801011, M141001528, M141001529, M141001530, M141001531, M141001536, M141001539, M141001541, M141001544, M141001545, M141001546, M141101678, M141101681, M141101682, M141101683, M141101684, M141101685, M141101686, M141101693, M141101696, M141101697, M141101700, M141101701, M141101703, M141101705, M141101706, M150102038, M150102039, M150102049, M150102052, M150102054, M150102055, M150102059, M150102061, M150102062, M150102064, M150102066, M150102399, M150102402, M150102404, M150102409, M150102410, M150102414, M150102416, M150202537, M150202538, M150202541, M150202542, M150202544, M150202545, M150302658, M150302660, M150302662, M150302664, M150302666, M150302667, M150302669, M150302673, M150302680, M150302682, M150302683, M150302687, M150302689, M150302690, M150302692, M150302694, M150302695, M150302702, M150302703, M150302707, M150302711, M150302712, M150302713, M150302716, M150302721, M150302723, M150302724, M150302725, M150302729, M150302737, M150302739, M150302742, M150302747, M150302748, M150302751, M150302755, M150302757, M150302758, M150302759, M150302762, M150302769, M150302773, M150302777, M150302778, M150302779, M150302780, M150302782, M150302783, M150302786, M150403048, M150403050, M150403051, M150403057, M150403066, M150805139, M150805143, M150805151, M150805154, M150805155, M150805156, M150805158, M150805160, M150805161, M150805162, M150805166, M150905677, M150905680, M150905682, M150905683, M150905685, M150905688, M150905689, M150905691, M150905692, M150905697, M150905698, M150905699, M150905702, M150905703, M150905704, M150905706, M150905707, M150905708, M150905709, M150905711, M150905714, M151206628, M151206631, M151206633, M151206636, M151206637, M151206641, M151206644, M151206645, M151206647, M151206648, M151206651, M151206655, M151206659, M151206662, M151206665, M160206849, M160206853, M160206854, M160206857, M160206861, M160206867, M160206869, M160206872, M160206876, M160206879, M160206880, M160206882, M160206885, M160206888, M160206896, M160206899, M160206900, M160206902, M160206903, M160206906, M160807699, M160807700, M160807701, M160807772. |
Recalling Firm/ Manufacturer |
St. Jude Medical 387 Technology Cir Nw Ste 500 Atlanta GA 30313-2424
|
For Additional Information Contact |
Mr. Justin Paquette 651-756-6293
|
Manufacturer Reason for Recall |
Radiated Radiofrequency (RF) emissions of the
CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
|
FDA Determined Cause 2 |
Device Design |
Action |
Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to continue to follow CardioMEMS Instructions for Use (IFU). Abbott will notify patients of the emissions issue. If you suspect a CardioMEMS Electronics System has interfered with another medical device during use, report the event to Abbott Remote Care Technical Support at 1-844-MYCMEMS (692-6367) (US). Customers were also instructed to complete and return the acknowledgement and receipt form.
Abbott notified patients on about 02/07/2023 via mailed letter. Customers were instructed that they could continue use of their CardioMEMS System and that their doctor has been notified of the issue as well. If there is suspicion that interference may be occurring during a sensor reading or any other quality issues, please report it to: Abbott Technical Support at 1-844-MYCMEMS (692-6367) (US); Monday-Friday: 8am-8pm EST |
Quantity in Commerce |
200 units |
Distribution |
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
PMA Database |
PMAs with Product Code = MOM and Original Applicant = Abbott Medical
|
|
|
|