| | Class 2 Recall ApexPro Telemetry System |  |
| Date Posted |
October 20, 2006 |
| Recall Number |
Z-0055-2007 |
| Product |
GE ApexPro CH Telemetry System. The ApexPro Telemetry System is composed of 6 major components. -The patient worn data acquisition transmitters; -The receiver antenna system infrastructure; -The receivers; -The receiver subsystem; -A computer platform running the Apex Pro Telemetry Application; -A computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application) |
| Code Information |
Software version 3.8 and earlier
|
Recalling Firm/ Manufacturer |
Ge Healthcare 9900 W Innovation Dr Wauwatosa, Wisconsin 53226-4856 |
Reason for Recall |
When a patient being monitored with ApexPro/Apex Pro CH Telemetry with ST monitoring disabled is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC (Clinical Information Center) does not occur.
|
| Action |
An Urgent Medical Device Correction letter dated August 04, 2006 was sent to customers with affected devices. The letter described the issue affecting devices containing software versions v3.8 and earlier, provides short term recommendations and a long term solution of updating the software. |
| Quantity in Commerce |
5,203 |
| Distribution |
Worldwide-USA and Puerto Rico and countries of ,Australia, Belgium, Canada, China, Croatia, Czech Republic, Estonia, Finland, France, Germany, Great Britain, Greece, Guam, Honduras, Hong Kong, Hunary, India, Indonesia, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Malaysia, Malta, Mexico, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerlant, Syria, Taiwan, Thailand, Tunisia, and Venezuela. |
| |
|
|
|
|