| | Class 2 Recall Maquet Critical Care Ventilator System |  |
| Date Posted |
February 10, 2007 |
| Recall Number |
Z-0379-2007 |
| Product |
Ventilator System Servo-i V3.1
Maquet Critical Care
PC1771 |
| Code Information |
Article number 64-87-800---Serial Numbers 14000-22772
|
Recalling Firm/ Manufacturer |
MAQUET Inc. 1140 US Highway 22 Bridgewater, New Jersey 08807-2958 |
| For Additional Information Contact |
Jamie Yieh 908-947-2311
|
Reason for Recall |
The Connection of certain nebulizers to the built in driver has lead to reduced or no nebulization effect on the Servo-i Ventilator
|
| Action |
Maquet Critical Care issued an ''Extended Warranty'' letter on 5/23/2006. On March 13, 2007, Maquet modified its recall strategy and issued a Recall Notification letter accompanied by a Safety Alert letter to all customers with Servo-i ventilators, serial numbers in the 14000-22772 range, with Servo nebulizer installed. The Safety Alert letter provides customers with information on how to identify if the unit is affected, and how to identify if the unit has shown any failures, as well as, any precautions. Firm initiated recall is on-going. |
| Quantity in Commerce |
2,436 units |
| Distribution |
Worldwide |
| |
|
|
|
|