Class 2 Device Recall Skytron Infinity Series surgical light

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards | Inspections
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Recall
Skytron Infinity Series surgical light
see related information
Date Posted February 22, 2007
Recall Status1 Terminated on November 18, 2009
Recall Number Z-0533-2007
Recall Event ID 37120
Premarket Notification
510(K) Number
K915285 
Product Classification Lamp, Surgical - Product Code FTD
Product Skytron Infinity Series surgical light with dual lightheads, 8 bulbs in 1 - 30' diameter lighthead and 5 bulbs in 1 - 22' diameter lighthead, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Models IN3022EL and IF3022EL. (note only the 22'' lighthead is affected).
Code Information Each lighthead bears a unique serial number. For units with model numbers containing IF and 22, all serial numbers with the following numbers in the third and fourth position of the numerical series: 95, 96, 97, 98, 99 or 00 (examples: xx97xxxx or xx00xxxB). For units with model numbers containing IN and 22 , all serial numbers with the following numbers in the third and fourth positions of the numerical series: 92, 93 or 94 (example: xx93xxxx).
Recalling Firm/
Manufacturer
Skytron, Div. The KMW Group, Inc
5000 36th St Se
Grand Rapids, Michigan 49512-2008
For Additional Information Contact Mary Preston-Murphy
800-759-8766 Ext. 3048
Manufacturer Reason
for Recall
If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.
FDA Determined
Cause 2
DESIGN: Device Design
Action The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
Quantity in Commerce 5,000 for all models.
Distribution Nationwide and Canada.