| | Class 2 Recall SMS |  |
| Date Posted |
March 01, 2007 |
| Recall Number |
Z-0545-2007 |
| Product |
Sample Management System software for in vitro diagnostic testing
Product # 030102-03 |
| Code Information |
SMS Product # 030102-03, Serial #s D0314, D0280, E0545, C0081, C0026, D0384, D0244, E0515, E0511, E0550, C0222, C0223, C0207, D0355, E0517, C0219R,C0221, B0021, B0022, D0353, E0503, E0446, C0193, E0471, E0566, D0299, E0513, D0281, D0356, C0101, C0203, C0205, B0027, E0426, E0440, C0176, C0182, E0504, C0220, E0495, D0272, E0490, E0534, C0208, C0218, C0191, E0473, D0296, C0168, E0470, C0199, D0330, C0206, D0313, C0192, E0509, D0369, E0492, E0493, D0304, D0310, D0311, E0563, D0331, C0204, E0589, D0327, D0275, E0514, E0443, E0444, D0243, C0216, C0217, E0546, E0548, C0209, E0450, D0291, D0276, D0273, D0332, E0508, B0033, E0449, E0516, D0357
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions Diagnostics 62 Flanders-Bartley Rd Flanders, New Jersey 07836-4715 |
| For Additional Information Contact |
Christopher clegg 973-927-2828 Ext. 4336
|
Reason for Recall |
Under limited circumstances, an incorrect patient result could be printed on the optional chartable patient report.
|
| Action |
A recall communication was sent on 1/15/2007 to all customers by fax. |
| Quantity in Commerce |
87 units |
| Distribution |
Worldwide, including USA, Australia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Russia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE and United Kingdom. |
| |
|
|
|
|