| | Class 2 Recall Image-intensified fluoroscopic x-ray system. |  |
| Date Posted |
February 15, 2007 |
| Recall Number |
Z-0514-2007 |
| Product |
OEC Uroview 2800 with Dual Monitor Option, fluoroscopic x-ray system, Model Numbers: 882082-01, 882082-02, 886765-01, GE Healthcare, Surgery, Salt Lake City, UT. |
| Code Information |
Model Numbers: 882082-01, 882082-02, 886765-01. Units with manufactured dates between 03/2003 and 08/2006.
|
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc 384 Wright Brothers Dr Salt Lake City, Utah 84116-2862 |
| For Additional Information Contact |
Ms. Karen Tolson 847-501-1884
|
Reason for Recall |
Uroview 2800 system with dual monitors may pose a safety risk to patients if the monitors become disconnected and fall.
|
| Action |
Consignees were notified by letter on 01/15/2007. Instructions were given to position and not move the dual monitors until inspection and correction by GE personnel. For the affected units with a 3 phase power distribution box, users were cautioned to ensure that the access door to the power box is completely closed when the unit is powered on to prevent a shock hazard. |
| Quantity in Commerce |
425 units |
| Distribution |
Worldwide, including USA, Canada, China, France, Germany, Hong Kong, Italy, Japan, Singapore, South Africa, Switzerland, and Venezuela. |
| |
|
|
|
|