 Class 3 Recall
Anaerotest
see related information
Date Posted February 17, 2007
Recall Number Z-0499-2007
Product Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck IVD 64271 Darmstadt, Germany
Code Information Item # 1.15112.0001 Lot # OC553145
Recalling Firm/
Manufacturer
EMD Chemicals Inc
480 S Democrat Rd
Gibbstown, New Jersey 08027-1239
Consumer Instructions Contact the recalling firm for information
For Additional Information Contact Joseph Scheollkopff
856-423-6300
Reason for
Recall
Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present
Action A Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.
Quantity in Commerce 800 packages of 50 strips each.
Distribution Product was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.
 
-
-