| | Class 2 Recall Symbia T Series SPECT-CT System with a Pinhole Collimator |  |
| Date Posted |
February 23, 2007 |
| Recall Number |
Z-0549-2007 |
| Product |
Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System with a Pinhole Collimator; Siemens Medical Solutions USA, Inc., Molecular Imaging, 2501 N. Barrington Road, Hoffman Estates, IL 60195-5203 USA; part number 08717733 |
| Code Information |
All systems with pinhole collimators and e.soft software versions 5.5 or earlier
part number 08717733, serial numbers 01015, 01032, 01037, 01045, 01061, 01065, 01068, 01072, 01074, 01076, 01077
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc. 2501 Barrington Rd Hoffman Estates, Illinois 60195-2061 |
| For Additional Information Contact |
Mr. Ron Nolte 800-767-2313
|
Reason for Recall |
Symbia systems, running on e.soft version 5.5 or earlier, could allow a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage.
|
| Action |
Siemens sent Customer Safety Advisory Notice/Urgent Field Correction Recall letters dated 1/30/07 to all users of Symbia systems with a pinhole collimator and e.soft software versions 5.5 or earlier. The accounts were informed of the potential for a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage. The customers were advised that a Siemens Customer Service Representative would be sent to their site to upgrade their system software to version syngeo MI Applications 2006A within 30 days. The upgrade will be done free of charge. |
| Quantity in Commerce |
11 units |
| Distribution |
Worldwide distribution --- including USA states of Alabama, California, Georgia, Indiana, Kentucky, Michigan, Ohio, Pennsylvania, Texas, Virginia and countries of Australia, Bolivia, Canada, China, France, Germany, Hong Kong and Spain. |
| |
|
|
|
|