|
Class 2 Device Recall Flexicair II |
|
Date Initiated by Firm |
May 12, 2009 |
Date Posted |
August 03, 2009 |
Recall Status1 |
Terminated 3 on April 03, 2012 |
Recall Number |
Z-1874-2009 |
Recall Event ID |
52032 |
510(K)Number |
K863047
|
Product Classification |
air therapy mattress - Product Code IOQ
|
Product |
Flexicair¿ II (model 13000) Low Airloss Therapy Units, for the prevention of pressure sores. |
Code Information |
All serial numbers. The devices have a nomenclature for each respective product model where the serial number is IBxxxxxx, where "xxxxxx" is a six digit number. The prefix "IB" is representative of a Flexicair¿ II. |
Recalling Firm/ Manufacturer |
Hill-Rom Manufacturing, Inc. 4349 Corporate Rd Charleston SC 29405-7445
|
For Additional Information Contact |
Joe Fogel 843-740-8586
|
Manufacturer Reason for Recall |
The devices may emit smoke from the blower box assembly.
|
FDA Determined Cause 2 |
Equipment maintenance |
Action |
Urgent Field Safety Notice letter dated May 11, 2009. Hill-Rom service centers were notified by letter on 05/13/2009. Action to be taken are as follows: Rental Customers; In the next 60 days, Hill Rom will be removing the Flexicair¿ II and Flexicair¿ MC3 low Airloss Therapy Units from service to correct the devices. During that time, Hill-Rom may provide an equivalent substitute therapy to accommodate the patients who continue to need a pressure reducing therapy device. If t his malfunction occurs before a substitution or a corrected device is provided, unplug the device, remove the patient from the surface and notify Hill-Rom immediately. Facility-Owned Customers: In the next 60 days, for all facility owned Flexicair II and Flexicair¿ MC3 Low Airloss Therapy Units, a Hill-Rom service technician will either replace the affected components in the blower assembly or deliver an equivalent substitute unit from Hill-Rom's rental fleet. If this malfunction occurs before the device is corrected or a substitute provided, do as follows: unplug the device, remove the patient from the surface, and immediately notify Hill-Rom Technical Support using the attached listing their geographical region. |
Quantity in Commerce |
57 units |
Distribution |
Nationwide and Canada |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IOQ and Original Applicant = SSI MEDICAL SERVICES, INC.
|
|
|
|