| | Class 2 Recall Zimmer Pulsavac Wound Debridement System high capacity intramedullary brush tip |  |
| Date Posted |
July 23, 2009 |
| Recall Number |
Z-1698-2009 |
| Product |
Zimmer Pulsavac Wound Debridement System high capacity intramedullary brush tip, sterile, Zimmer, Dover, Ohio; Catatog Number: 00515018300.
Used in conjunction with the Pulsavac Plus Wound Debridement System for pulsatile lavage/debridement. |
| Code Information |
Lot Number: 61229259.
|
Recalling Firm/ Manufacturer |
Zimmer Inc. 345 E Main St Warsaw, Indiana 46580-2746 |
Reason for Recall |
The seal on the sterile barrier pouch may be inadequate, resulting in lack of assurance of sterility.
|
| Action |
Zimmer Orthopedic Surgical Products issued an "Urgent: Medical Device Recall" notice dated June 8, 2009 informing users of the affected devices and actions to be taken including the return of product and acknowledgement of receipt of notice by returning the Recall Certification Form by fax to 1-330-364-0974. |
| Quantity in Commerce |
15 |
| Distribution |
Worldwide Distribution -- United States, Canada and Japan. |
| |
|
|
|
|