• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall B. Braun Infusomat Space Pump US Version

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall B. Braun Infusomat Space Pump US Version see related information
Date Initiated by Firm July 12, 2011
Date Posted February 29, 2012
Recall Status1 Terminated 3 on November 03, 2014
Recall Number Z-1015-2012
Recall Event ID 60995
510(K)Number K062700  K083689  
Product Classification Pump, Infusion - Product Code FRN
Product B. Braun Infusomat Space Pump - US Version Infusion pump. Model number 8713050U.
Code Information Serial numbers  108288 108290 108291 108292 108293 108294 108296 108297 108298 108299 108300 108302 108303 108304 108305 108306 108307 108309 108310 108311 108312 108313 108315 108318 108319 108321 108322 108323 108324 108326 108327 108331 108332 108334 108335 108337 108338 108339 108342 108343 108344 108347 108348 108349 108350 108351 108352 108353 108354 108359 108360 108362 108363 108364 108365 108366 108367 108368 108377 108378 108380 108381 108383 108385 108388 108389 108390 108395 108399 108400 108401 108402 108403 108405 108406 108407 108408 108409 108412 108418 108420 108422 108423 108424 108427 108428 108429 108430 108439 108442 108443 108445 108448 108450 108451 108453 108454 108455 108456 108457 108465 108469 108470 108471 108473 108474 108477 108478 108486 108490 108493 108495 108497 108500 108501 108508 108510 108511 108513 108517 108520 108536 108537 108538 108541 108542 108548 108549 108550 108552 108553 108556 108557 108560 108561 108563 108569 108570 108572 108577 108578 108580 108581 108582 108584 108585 108586 108588 108589 108592 108596 108602 108604 108606 108607 108608 108609 108610 108611 108612 108614 108615 108617 108618 108619 108620 108622 108623 108625 108626 108628 108631 108633 108634 108635 108638 108640 108641 108642 108643 108645 108647 108648 108650 108651 108653 108655 108656 108659 108660 108661 108662 108663 108664 108665 108666 108667 108668 108669 108670 108671 108672 108673 108674 108676 108678 108680 108682 108683 108685 108687 108688 108689 108693 108696 108697 108700 108701 108702 108703 108706 108707 108708 108709 108711 108712 108714 108715 108720 108721 108724 108726 108727 108728 108729 108731 108732 108733 108734 108735 108737 108739 108740 108743 108744 108746 108747 108748 108750 108751 108752 108753 108754 108755 108756 108758 108759 108761 108762 108769 108770 108771 108772 108773 108774 108775 108776 108777 108778 108779 108780 108781 108783 108784 108785 108786 108788 108789 108790 108791 108793 108794 108795 108796 108797 108798 108799 108800 108801 108802 108803 108804 108805 108806 108808 108809 108810 108811 108812 108813 108814 108815 108816 108817 108818 108819 108820 108821 108822 108823 108824 108825 108826 108827 108828 108829 108830 108831 108832 108833 108834 108835 108836 108837 108838 108840 108842 108844 108846 108847 108848 108849 108850 108851 108852 108853 108854 108855 108856 108857 108859 108860 108862 108863 108864 108865 108866 108867 108868 108869 108870 108871 108874 108879 108883 108885 108886 108887 108888 108889 108890 108891 108897 108899 108902 108906 108907 108908 108912 108914 108915 108916 108917 108918 108919 108920 108921 108922 108923 108924 108925 108926 108927 108928 108930 108934 108936 108937 108941 108942 108943 108945 108948 108949 108950 108951 108952 108954 108956 108957 108958 108961 108963 108967 108968 108971 108972 108973 108974 108975 108976 108977 108978 108979 108980 108981 108982 108984 108985 108986 108987 108988 108989 108990 108991 108992 108993 108994 108995 108996 108997 108998 108999 109000 109001 109002 109003 109004 109005 109006 109009 109010 109011 109013 109014 109015 109016 109017 109018 109019 109020 109022 109023 109024 109025 109026 109027 109028 109029 109030 109031 109032 109033 109034 109035 109036 109037 109038 109039 109040 109042 109043 109044 109045 109047 109048 109049 109050 109051 109052 109053 109054 109055 109056 109057 109059 109060 109061 109062 109063 109064 109065 109066 109067 109069 109070 109071 109073 109074 109075 109076 109077 109078 109079 109080 109081 109082 109084 109085 109086 109087 109088 109089 109090 109091 109092 109093 109094 109095 109096 109097 109098 109099 109100 109101 109102 109103 109104 109105 109106 109107 109108 109109 109110 109111 109112 109113 109114 109115 109117 109118 109119 109120 109121 109122 109123 109124 109125 109126 109127 109128 109129 109130 109131 109132 109133 109134 109136 109137 109138 109139 109140 109141 109142 109143 109144 109145 109146 109148 109149 109151 109152 109153 109154 109156 109157 109158 109159 109160 109161 109162 109164 109165 109166 109167 109168 109170 109171 109173 109175 109176 109178 109180 109181 109182 109183 109184 109185 109187 132785 132795 132808 132809 132835 132839 132892 132893 132922 132987 133004 133005 133006 133018 133020 133026 133029 133030 133035 133037 133039 133043 133055 133058 133064 133065 133067 133068 133077 133092 133095 133096 133099 133100 133101 133104 133106 133124 133127 133129 133130 133131 133132 133133 133134 133135 133136 133155 133156 133158 147597 147683 147818 35198 35201 35202 35205 35209 35211 35212 35216 35217 35218 35219 35220 35221 35222 35223 35224 35226 35229 35231 35232 35235 35236 35237 35238 35240 35241 35242 35243 35244 35246 35247 35248 35249 35250 35251 35252 35254 35255 35256 35257 35259 35261 35263 35265 35267 35270 35271 35272 35279 35284 35285 35287 35288 35300 39868 39870 39875 39881 39884 39890 39903 39912 39914 39922 39923 39933 39942 39959 39965 39974 39978 39980 39989 39994 39997 40007 40008 40010 40011 40012 40013 40014 40018 40025 40038 40039 40043 40044 40045 40047 40889 40895 40909 40912 40927 40928 40929 40938 40947 40951 40952 40954 40955 40959 40961 40972 40977 40999 41000 41003 41017 41032 41035 41045 41053 41071 41082 41099 41105 41109 41117 41130 41132 41133 41135 41144 41148 41153 41168 41174 41185 41191 41209 41213 41224 41227 41229 41237 41249 41261 41266 41272 41273 41275 41277 41279 41297 41305 41325 41326 41328 41329 41330 41343 41364 41367 41369 41370 41371 41373 41381 41389 41393 41399 41412 41415 41421 41435 41437 41449 41450 41459 41484 41515 41517 41520 41530 41565 41598 41629 41641 47711 47722 47735 47736 47742 47753 47766 47779 47784 47789 47800 47801 47804 47806 47808 47809 47810 47811 47813 47815 47827 47835 47836 47837 47843 47845 47846 47847 47855 47874 47878 47880 47889 47892 47900 47932 47948 47954 47955 47970 47973 47988 47994 48003 48004 48005 48668 48669 48672 48673 48674 48676 48677 48678 48679 48680 48681 48682 48683 48684 48685 48686 48687 48688 48690 48691 48692 48693 48694 48695 48697 48699 48703 48704 48705 48706 48707 48708 48709 48711 48712 48713 48715 48716 48717 48718 48719 48720 48721 48722 48723 48724 48725 48727 48728 48729 48730 48731 48732 48734 48735 48736 48737 48739 48740 48741 48743 48744 48745 48746 51253 51254 51255 51256 51257 51259 51260 51262 51264 51265 51267 51268 51271 51272 51273 51274 51276 51278 51279 51280 51281 51283 51284 51285 51288 51289 51290 51291 51292 51293 51294 51295 51296 51297 51298 51299 51300 51301 51302 51303 51305 51308 51309 51310 51312 51313 51314 51315 51316 51317 51318 51320 51321 51322 51323 51325 51327 51328 51329 51330 51331 51333 51335 51336 51337 51338 51339 51340 51341 51342 51343 51344 51349 51350 51351 51352 51353 51354 51355 51356 51357 51360 51361 51363 51364 51365 51366 51367 51368 51369 51370 51371 51372 51628 51629 51630 51631 51632 51638 51639 51640 51641 51642 51643 51645 51646 51648 51649 51650 51651 51652 51654 51656 51657 51658 51659 51660 51661 51662 51663 51664 51665 51666 51667 51672 51678 51679 51681 51682 51683 51685 51686 51687 51688 51690 51691 51692 51693 51694 51695 51696 51697 51698 51699 51700 51701 51702 51704 51705 51707 51708 51709 51710 51711 51714 51715 51716 51717 51718 51719 51720 51721 51722 51723 51724 51725 51727 51728 51729 51730 51732 51733 51734 51735 51736 51737 51739 51740 51741 57955 60611 60757 62155 62184 62185 86327 86726 86803 87090 87093 87104 106327 106942 107360 107399 107446 107553 107651 107768 144569 144570 144590 144616 144648 144655 148611 148675 148883 148894 148914 148918 148960 148962 148978 148980 148989 149020 149021 149024 149027 149028 149033 149034 149054 149061 35204 35207 35225 35233 35234 35253 35258 35273 39205 39232 39356 39360 39395 39404 39455 39475 39548 39949 39950 40005 41666 41679 41685 41692 41714 41745 41746 41767 41772 41773 41789 41805 41811 41821 41835 41839 41873 41876 41890 41891 41928 41934 41938 41944 41947 41948 41964 41969 41970 41976 41981 41985 41993 42000 42003 42036 42037 48671 48675 48696 48726 51263 51266 51304 51306 51307 51311 51347 51359 51362 51623 51624 51625 51626 51627 51634 51635 51636 51647 51669 51674 51675 51676 51677 51680 51712 51743 51746 51749 51750 51752 51753 51754 51755 51757 51758 51759 51760 51761 51762 51764 51766 51767 51768 51771 51772 52223 52227 52229 52232 52233 52235 52236 52237 52244 52246 52248 52250 52252 52254 52255 52257 52258 52260 52261 52265 52267 52268 52269 52270 52274 52278 52279 52280 52281 52282 52284 52286 52287 52288 52289 52290 52291 52293 52294 52295 52296 52297 52298 52300 52301 52311 52312 52313 52314 52315 52316 52317 52319 52321 52323 52328 52329 52333 52334 52335 52336 52337 52338 52342 52346 52347 52348 52349 52352 52353 52355 52359 52360 52361 52363 52364 52366 52372 52856 52874 52941 53081 53089 53113 53186 53259 53399 53415 53416 53427 53430 53434 53439 53445 53468 53475 53482 53489 53498 53535 53563 53594 53604 53607 53612 53613 53620 53627 53633 53637 53652 53736 53737 53806 53808 53841 53843 53957 55729 55730 55740 55762 55787 55881 55888 55889 55936 55944 55961 55969 55982 55985 55988 55993 55995 56005 56017 56021 56037 56063 56076 56078 56086 56093 56104 56124 56130 56132 56142 56143 56145 56150 56153 56159 56619 56626 56628 56640 56647 56657 56679 56682 56686 56689 56690 56713 56717 56981 56997 56999 57009 57013 57015 57021 57035 57036 57044 57045 57064 57070 57073 57074 57092 57094 57102 57110 57120 57138 57139 57144 57157 57158 57168 57170 57171 57184 57186 57198 57205 57207 57209 57219 57224 57226 57227 57228 57230 57232 57233 57240 57242 57251 57260 57262 57264 57281 57292 57294 57299 57301 57302 57308 57310 57311 57312 57315 57350 57359 57381 57387 57399 57401 57402 57410 57416 57422 57440 57447 57457 57462 57463 57465 57471 57873 57880 57889 57892 57893 57909 57913 57914 57915 57918 57923 57924 57930 57940 57941 57948 57953 57962 57980 57992 57994 58000 58003 58004 58007 58012 58024 58026 58027 58028 58029 58030 58043 58053 58070 58079 58083 58087 58091 58109 58111 58112 58126 58134 58137 58147 58148 58159 58176 58179 58190 58195 58198 58210 58218 58231 58232 58236 58250 58257 58263 58265 58273 58277 58291 58296 58312 58316 58318 58346 58375 58388 58391 58424 58443 58444 58450 58458 58461 58475 58476 58508 58513 58533 58539 58543 58545 58550 58552 58581 58586 58587 58591 58593 58610 60511 60523 60540 60552 60584 60588 60591 60592 60596 60603 60620 60650 60652 60657 60671 60684 60691 60731 60733 60734 60738 60745 60750 60753 60766 60771 60782 60788 60800 60804 60806 60807 60818 60822 60826 60829 60830 60833 60836 60838 60839 60841 60843 60844 60846 60847 60848 60851 60857 60859 60862 60864 60871 60874 60876 60882 60888 60890 60894 60895 60897 60898 60902 60904 60906 60907 60910 60915 60916 60918 60919 60923 60925 60927 60928 60929 60930 60931 60932 60933 60936 60938 60943 60947 60949 60950 60953 60954 60955 60958 60959 60961 60962 60965 60968 60971 60973 60974 60976 60977 60978 60979 60980 60983 60988 60989 60991 61997 61998 62002 62007 62008 62009 62015 62016 62017 62018 62019 62020 62025 62028 62031 62034 62036 62038 62039 62041 62046 62049 62055 62057 62060 62062 62063 62065 62066 62078 62080 62088 62090 62093 62094 62096 62099 62100 62103 62111 62114 62117 62123 62127 62128 62129 62134 62140 62141 62144 62146 62147 62149 62150 62152 62153 62157 62159 62162 62163 62164 62168 62174 62176 62177 62186 62192 62198 62200 62204 62208 62210 62216 62217 62218 62220 62222 62223 62234 62237 62238 62239 62243 62248 62258 62265 62270 62275 62277 62279 62280 62281 62289 62291 62292 62295 62302 62305 62308 62309 62312 62314 62318 62333 62339 62343 62344 62347 62348 62351 62358 62362 62366 62368 62381 62385 62390 64459 64554 64615 64670 64688 64702 64717 64720 64792 64805 64809 64810 64827 64892 109190 109194 109207 109209 109212 109215 109218 109230 109238 109239 109282 109293 109295 109300 109326 109345 109355 109366 109381 109418 109423 109432 109436 109438 109493 109506 109507 109532 109540 109602 109630 109640 109645 109663 109667 109681 109707 109708 109715 109719 109723 109727 109732 109747 109758 109764 109766 109808 109810 109811 109812 109813 109814 109816 109818 109835 109845 109850 109851 109857 109863 109867 109868 109877 109884 109886 109889 109900 109903 109908 109909 109910 109913 109920 109921 109929 109932 109941 109943 109945 109952 109959 109964 109967 109973 109974 109981 109983 109984 109986 109989 109991 109995 109996 109999 110002 110006 110007 110009 110010 110026 110030 110034 110045 110052 110068 110070 110075 110081 111291 111308 111317 111319 111323 111334 111346 111355 111363 111364 111370 111376 111389 111393 111402 111404 111428 111433 111437 111439 111454 111456 111457 111458 111460 111461 111465 111484 111485 111488 111508 111509 111510 111514 111543 111546 111565 111570 111573 111575 111576 111588 111589 111591 111596 111597 111602 111604 111605 111606 111609 111611 111612 111613 111616 111617 111618 111620 111624 111629 111630 111637 111639 111640 111641 111642 111644 111651 111663 111666 111668 111673 111675 111676 111678 111679 111682 111685 111691 111693 111698 111702 111703 111707 111710 111718 111727 111736 111741 111743 111744 111747 111748 111751 111754 111755 111756 111761 111765 111769 111770 111772 111775 111776 111783 111784 111785 111789 111791 111792 111794 111795 111797 111798 111803 111805 111811 111813 111814 111815 111822 111825 111832 111833 111835 111836 111842 111843 111845 111846 111848 111850 111852 111854 111857 111858 111859 111863 111865 111866 111868 111869 111876 111879 111880 111882 111883 111884 111885 111887 111889 111893 111897 111898 111901 111902 111905 111906 111907 111908 111912 111916 111918 111919 111920 111923 111926 111927 111928 111931 111933 111934 111937 111939 111940 111948 111950 111951 111952 111954 111955 111958 111959 111960 111961 111962 111965 111966 111968 111969 111971 111973 111974 111975 111977 111978 111979 111981 111984 111988 111989 111990 111991 111992 111997 111998 112000 112001 112002 112003 112005 112006 112007 112012 112014 112018 112019 112020 112021 112022 112024 112025 112026 112027 112028 112030 112032 112033 112034 112037 112038 112039 112040 112041 112043 112044 112046 112047 112048 112049 112050 112052 112056 112059 112061 112063 112064 112065 112077 112078 112079 112080 112082 112083 112084 112085 112091 112096 112097 112098 112108 112109 112110 112111 112115 112116 112120 112121 112125 112127 112128 112130 112133 112134 112137 112141 112143 112144 112147 112152 112154 112155 112156 112163 112164 112165 112166 112167 112170 112172 112174 112175 112176 112177 112178 112182 112183 112184 112185 112188 112191 112192 112193 112196 112198 112199 112203 112205 112206 112212 112219 112224 112225 112227 112228 112230 112232 112233 112234 112241 112242 112243 112251 112252 112256 112269 112272 112275 112276 112277 112278 112283 112289 112291 112295 112298 112299 112304 112306 112314 112316 112318 112319 112331 112332 112338 112340 112342 112346 112348 112351 112358 112362 112364 112368 112378 112381 112385 112386 112393 112394 112397 112398 112401 112413 112416 112417 112418 112419 112420 112421 112422 112423 112424 112428 112430 112431 112432 112434 112442 112445 112448 112456 112457 112458 112463 112464 112467 112471 112472 112480 139450 139483 139568 139605 139703 111294 111316 111347 111353 111409 111415 111426 111441 111442 111443 111449 111450 111464 111470 111477 111486 111489 111492 111516 111526 111534 111537 111539 111540 111544 111555 111557 111560 111561 111562 111572 111592 111593 111594 111595 111599 111600 111601 111610 111634 111635 111645 111660 111665 111669 111670 111671 111684 111696 111697 111704 111705 111722 111728 111729 111730 111738 111745 111746 111760 111762 111764 111774 111780 111781 111782 111786 111788 111793 111823 111831 111839 111851 111856 111861 111864 111867 111873 111874 111888 111896 111904 111913 111914 111922 111924 111935 111942 111946 111949 111956 111963 111964 111967 111970 111980 111982 111983 111985 111986 111987 111993 111995 112004 112008 112011 112015 112017 112023 112031 112036 112051 112054 112055 112057 112058 112068 112075 112081 112088 112089 112093 112094 112095 112099 112104 112105 112106 112107 112112 112113 112114 112117 112119 112122 112126 112132 112136 112145 112148 112151 112161 112171 112179 112180 112181 112186 112187 112189 112190 112194 112197 112202 112209 112210 112211 112215 112216 112217 112221 112223 112229 112231 112236 112248 112250 112255 112260 112286 112287 112301 112310 112311 112313 112315 112322 112327 112341 112343 112350 112361 112363 112371 112372 112374 112377 112380 112387 112389 112396 112399 112400 112405 112406 112407 112410 112414 112415 112427 112436 112446 112447 112454 112455 112459 112462 112469 112470 112478 112485 133550 133693 30443 30453 30463 30474 30499 30503 30509 30536 30538 30540 33045 33048 33065 33085 33087 33092 33093 33099 33100 33111 33119 39024 39189 39202 39382 39385 39389 39405 39408 39454 39514 39518 39550 39560 39874 39887 39888 39900 39910 39927 39929 39930 39934 39938 39957 39961 39966 39970 39979 40030 40033 40036 40041 40057 40123 40142 40892 40936 40940 40996 41098 41344 41513 41519 41534 41620 41673 41683 41693 41713 41726 41734 41747 41800 41810 41877 41898 41920 41923 41941 41943 41953 41968 41998 42006 42031 42038 42048 42052 42057 43991 47849 48905 52827 52835 52847 52849 52854 52867 52868 52869 52875 52878 52881 52888 52901 52902 52910 52928 52949 52958 53053 53062 53065 53066 53082 53084 53088 53124 53129 53130 53134 53139 53141 53142 53143 53147 53155 53168 53185 53195 53220 53222 53225 53230 53233 53245 53247 53251 53252 53255 53263 53268 53269 53270 53375 53385 53400 53401 53402 53404 53409 53421 53423 53431 53432 53433 53440 53454 53457 53466 53467 53470 53472 53478 53480 53483 53495 53501 53505 53507 53509 53510 53518 53520 53521 53524 53531 53549 53550 53551 53555 53561 53569 53570 53572 53585 53589 53591 53603 53606 53608 53611 53615 53617 53619 53625 53626 53632 53634 53641 53657 53659 53662 53668 53671 53676 53682 53683 53685 53690 53704 53712 53713 53714 53716 53719 53734 53735 53740 53754 53764 53767 53778 53780 53791 53814 53816 53820 53844 53853 53856 53873 53900 53931 54021 54045 54065 55668 55673 55674 55675 55678 55695 55704 55709 55720 55725 55726 55734 55735 55739 55744 55745 55750 55751 55756 55757 55776 55778 55803 55805 55809 55815 55830 55840 55841 55848 55867 55879 55880 55892 55895 55905 55910 55920 55921 55928 55934 55937 55939 55942 55947 55964 55966 55967 55976 55977 55979 55980 55986 55989 55994 56001 56002 56010 56013 56022 56028 56033 56036 56043 56046 56047 56059 56069 56072 56073 56079 56084 56095 56096 56102 56111 56117 56119 56138 56156 56625 56629 56641 56642 56646 56661 56667 56671 56672 56677 56678 56687 56691 56692 56694 56696 56701 56704 56982 57022 57039 57041 57048 57079 57082 57085 57114 57142 57178 57216 57285 57335 57411 57449 57453 57458 57899 57902 57928 57961 57965 57988 58006 58021 58031 58032 58037 58039 58050 58062 58075 58103 58118 58119 58175 58188 58194 58219 58262 58268 58299 58301 58302 58307 58333 58337 58341 58376 58402 58415 58417 58431 58433 58437 58442 58449 58457 58460 58469 58482 58488 58490 58503 58504 58517 58523 58529 58542 58565 58571 58574 58582 58584 58585 58598 58618 58621 82139 82285 82294 82365 84278 84287 84291 84299 84303 84310 84323 84343 84346 84347 84350 84358 84361 84372 84439 84460 84472 84500 84512 84530 84535 84655 84684 84775 86356 86455 144498 144499 144519 144520 144525 144626 144652 146264 146308 146336 146343 147703 147707 147708 147748 147760 147761 147808 111293 111300 111302 111313 111329 111332 111333 111348 111354 111356 111366 111377 111382 111383 111384 111386 111391 111394 111397 111403 111408 111410 111411 111414 111418 111419 111422 111427 111429 111431 111432 111438 111440 111444 111446 111447 111452 111453 111455 111466 111467 111468 111469 111473 111474 111475 111476 111478 111480 111487 111490 111491 111493 111494 111495 111496 111497 111498 111500 111501 111502 111503 111511 111512 111513 111515 111517 111518 111519 111520 111521 111522 111523 111524 111527 111529 111531 111533 111535 111536 111538 111541 111545 111547 111549 111550 111551 111552 111553 111556 111558 111563 111564 111566 111568 111571 111577 111578 111579 111580 111581 111583 111584 111585 111587 111598 111614 111615 111621 111622 111625 111643 111647 111649 111650 111656 111657 111658 111659 111661 111667 111672 111677 111683 111686 111687 111688 111694 111700 111701 111706 111708 111709 111711 111715 111716 111717 111719 111720 111725 111732 111737 111742 111752 111758 111759 111766 111767 111779 111796 111799 111804 111809 111812 111816 111821 111824 111828 111840 111844 111847 111849 111860 111871 111872 111877 111890 111894 111895 111900 111909 111915 111917 111943 111972 112042 112312 112347 112349 112369 112375 112376 112388 112390 112392 112395 112402 112403 112404 112408 112409 112411 112412 112425 112426 112429 112433 112435 112437 112438 112439 112440 112441 112443 112449 112451 112452 112460 112461 112465 112466 112468 112473 112474 112475 112476 112477 112479 112481 112482 112483 112484 112486 112487 136169 136539 30438 30449 30486 33146 35210 35213 35214 35260 35266 35269 35274 35289 35299 38850 38930 39234 40165 41699 41701 41711 41722 41759 41766 42030 42046 47778 48670 48710 48733 48738 51261 51275 51287 51346 51348 51358 51644 51668 51671 51703 51726 51731 51742 51763 52239 52283 52285 52292 52302 52303 52307 52308 52310 52320 52327 52339 52340 52351 52362 52369 52370 52832 52833 52844 52877 53093 53254 53373 53425 53442 53444 53459 53462 53479 53496 53526 53528 53529 53584 53593 53598 53616 53624 53624 53647 53648 53669 53691 53771 53789 53823 53862 53993 54051 55818 56026 56027 56065 56110 56617 56676 56681 57127 57185 57187 57936 58160 58206 58252 58453 58526 60501 60513 60593 60604 60636 60694 60715 60760 60761 60774 60792 60815 60855 60858 60861 60868 60881 60903 60922 60937 60951 60982 60985 60986 62005 62012 62045 62059 62083 62178 62229 62254 62317 62319 62380 64492 64553 64579 64590 64599 64618 64621 64671 64694 64741 64742 64756 64774 64871 64874 64901 73682 73698 73700 73707 73708 73711 73712 73714 73730 73753 73755 73756 73761 73764 73776 73778 73786 73789 73791 73795 73796 73802 73803 73804 73805 73813 73819 73824 73826 73827 73830 73833 73838 73841 73842 73851 73856 73859 73862 73866 73873 73878 73881 73884 73896 73905 73916 73917 73926 73932 73934 73945 73946 73961 73970 73983 73988 73995 74005 74018 74024 74038 74039 74045 74058 74059 74065 74070 74094 74099 74116 74134 74140 74145 74169 74170 74188 74201 74207 74210 74212 74228 74258 74264 74277 74296 74305 74308 74311 74312 74316 74321 75134 75145 75149 75157 75159 75164 75168 75169 75170 75179 75180 75186 75189 75190 75202 75203 75213 75216 75226 75228 75232 75237 75240 75241 75244 75245 75247 75248 75253 75264 75265 75282 75283 75286 75289 75292 75293 75294 75295 75299 75302 75311 75312 75317 75326 75327 75328 75333 75346 75351 75353 75360 75365 75366 75371 75379 75380 75382 75402 75405 75406 75408 75413 75415 75416 75417 75418 75420 75423 75424 75426 75431 75434 75435 75448 75449 75459 75460 75462 75469 75474 75480 75481 75483 75487 75489 75497 75499 75513 75517 75523 75527 75532 75533 75535 75537 75547 75549 75550 75558 75560 75567 75591 75596 75597 75599 75600 75601 75602 75609 75614 75615 75616 75625 75629 75630 75633 75634 75635 75641 75642 75658 75659 75669 75670 75671 75672 75690 75698 75702 75703 75704 75706 75707 75708 75709 75710 75715 75718 75719 75720 75721 75723 75724 75725 75728 75729 75731 75732 75734 75735 75736 75737 75741 75742 75743 75744 75745 75747 75749 75750 75751 75753 75754 75755 75756 75758 75761 75762 75763 75764 75768 75769 75770 75771 75772 75773 75774 75775 75776 77228 77229 77231 77234 77236 77237 77238 77239 77240 77242 77243 77244 77245 77246 77248 77249 77250 77251 77252 77253 77254 77255 77256 77257 77258 77260 77262 77263 77264 77265 77266 77267 77270 77271 77272 77274 77275 77276 77277 77278 77279 77280 77281 77283 77284 77285 77286 77287 77288 77289 77290 77292 77293 77294 77297 77298 77300 77301 77302 77303 77306 77307 77309 77311 77312 77313 77314 77316 77317 77318 77319 77320 77321 77322 77323 77324 77325 77326 77327 77328 77329 77330 77333 77334 77336 77337 77338 77339 77341 77342 77343 77344 77345 77347 77349 77350 77352 77353 77354 77357 77358 77359 77360 77361 77362 77363 77364 77365 77369 77370 77371 77372 77374 77375 77376 77377 77378 77379 77381 77383 77385 77386 77388 77389 77390 77392 77393 77397 77398 77399 77400 77401 77403 77404 77409 77410 77421 77427 77428 77430 77431 77434 77436 77441 77442 77444 77445 77447 77448 77450 77452 77458 77459 77462 77464 77465 77466 77467 77468 77469 77470 77471 77472 77473 77474 77475 77476 77477 77478 77480 77481 77482 77483 77485 77487 77489 77491 77492 77493 77494 77495 77496 77500 77503 77507 77509 77512 77513 77514 77517 77518 77519 77525 77526 77529 77533 77535 77540 77541 77542 77543 77546 77561 77563 77565 77572 77575 77577 77578 77586 77589 77591 77598 77602 77603 77605 77608 77612 77626 77631 77640 77641 77644 77646 77648 77653 77655 77657 77670 77678 77679 77681 77684 77685 77688 77691 77692 77697 77703 77710 77726 77728 77730 77743 77744 77752 77753 77762 77765 77766 77767 77773 77782 77788 77805 77809 77814 77820 77827 77830 77838 77840 77853 77854 77855 77864 77871 77877 77878 77880 77881 77934 77943 146071 146074 146239 146259 146261 146274 146275 146285 146286 146298 146301 146306 146310 146311 146312 146324 146349 146350 146430 146515 146542 146661 146914 146979 147037 147042 147095 147096 147126 147138 147140 147819 147849 147852 147853 147862 147867 147894 147895 147929 147933 147966 147968 35282 48689 48701 51269 51277 51282 51332 51706 52299 52318 58306 58518 78078 81486 81601 81621 81631 81651 81698 81704 81838 81842 81902 82080 82103 82119 82129 82229 82276 82281 82283 82287 82327 82354 82379 82408 82416 82463 82471 82488 82505 82508 82514 82519 82540 82542 82546 82554 82565 84275 84280 84281 84286 84293 84295 84297 84311 84312 84317 84320 84321 84322 84325 84330 84331 84333 84334 84338 84342 84348 84357 84365 84368 84377 84378 84382 84385 84388 84398 84399 84400 84401 84404 84409 84411 84413 84414 84420 84421 84429 84436 84438 84440 84441 84442 84443 84444 84445 84448 84452 84456 84468 84471 84474 84478 84479 84480 84483 84485 84488 84489 84490 84491 84498 84502 84504 84513 84518 84521 84525 84526 84527 84536 84537 84539 84548 84549 84551 84556 84561 84569 84571 84572 84574 84575 84580 84581 84590 84592 84593 84594 84598 84599 84602 84610 84611 84613 84614 84616 84618 84619 84622 84623 84628 84630 84631 84634 84637 84638 84639 84640 84643 84650 84651 84653 84654 84656 84658 84661 84662 84665 84667 84668 84669 84670 84681 84682 84683 84686 84688 84690 84691 84693 84694 84695 84698 84701 84702 84704 84707 84708 84712 84713 84720 84721 84723 84727 84728 84731 84733 84734 84738 84740 84742 84749 84750 84752 84754 84756 84758 84764 84765 84772 84773 84774 84776 84778 84779 84787 84788 84789 84791 84792 84797 84799 84805 84807 84809 84813 84816 84819 84820 84824 84826 84827 84828 84829 84834 84835 84837 84838 84839 84840 84842 84843 84847 84848 84850 84853 84855 84857 84863 84865 84868 84870 84871 84901 84902 84906 84907 84908 84911 84912 84913 84914 84916 84917 84919 84920 84922 84923 84932 84933 84936 84942 84954 84957 84965 84967 84973 84975 84977 84978 84982 84984 84986 84987 84993 84994 84996 73743 77637 77716 79426 79429 79524 79625 79699 79838 32975 33103 33115 33143 47786 52969 53622 58434 60765 60842 60963 62138 62158 62180 62212 62266 73850 73875 73902 73913 74041 74085 74123 74182 74229 74253 74286 75160 75188 75263 75281 75304 75342 75345 75383 75393 75443 75466 75479 75501 75510 75530 75548 75578 75585 75586 75592 75626 75637 75656 75668 75688 75694 75699 75712 75722 75730 75757 75777 77230 77235 77259 77273 77295 77299 77315 77368 77373 77380 77384 77395 77406 77411 77413 77414 77416 77423 77424 77425 77443 77451 77454 77456 77461 77486 77497 77499 77511 77515 77528 77530 77544 77545 77550 77551 77560 77562 77566 77569 77570 77579 77593 77594 77595 77596 77601 77607 77609 77610 77613 77614 77616 77623 77630 77635 77636 77639 77642 77643 77645 77658 77659 77662 77664 77671 77674 77677 77680 77686 77687 77695 77696 77698 77699 77706 77707 77715 77731 77734 77735 77738 77739 77740 77749 77751 77754 77755 77758 77759 77760 77763 77770 77772 77778 77784 77785 77786 77787 77789 77790 77792 77794 77797 77800 77807 77808 77810 77811 77812 77813 77816 77819 77821 77823 77824 77833 77835 77836 77839 77844 77845 77850 77851 77852 77857 77858 77860 77862 77863 77865 77867 77870 77873 77876 77879 77893 77906 77913 77923 77924 77925 77927 77936 77938 77939 77946 77948 77949 77952 77960 77966 77971 78001 78019 78694 75506 78597 78628 78674 78704 79380 79587 60926 62322 64468 64524 64577 64606 64622 64642 64677 64697 64753 64862 64873 73986 74010 74052 74053 74055 74087 74118 74125 74175 75748 77247 77268 77269 77340 77437 77571 77719 77764 77799 77803 77944 78003 78091 78107 78140 78152 78172 78209 78235 78236 78238 78267 78363 78379 78391 78417 78482 78513 78546 78554 78566 78583 78595 78612 78627 78632 78634 78636 78653 78666 78668 78670 78671 78672 78679 78710 78713 78726 78768 78770 78794 78806 78811 78815 78842 78843 78859 78860 78901 78906 78921 78924 78926 79352 79355 79364 79365 79367 79371 79374 79381 79382 79386 79391 79393 79395 79399 79402 79406 79407 79408 79415 79417 79419 79421 79430 79433 79436 79437 79438 79440 79441 79442 79443 79445 79446 79447 79451 79459 79460 79465 79468 79470 79471 79472 79473 79476 79477 79478 79483 79485 79488 79500 79502 79503 79504 79506 79509 79512 79514 79517 79521 79527 79529 79530 79534 79537 79538 79539 79540 79541 79542 79543 79544 79546 79547 79548 79549 79556 79558 79559 79560 79566 79567 79569 79572 79576 79580 79586 79588 79593 79595 79597 79599 79600 79601 79602 79613 79615 79619 79621 79627 79629 79630 79631 79633 79641 79644 79645 79652 79655 79656 79660 79664 79667 79670 79671 79674 79677 79682 79683 79684 79687 79690 79692 79696 79698 79712 79720 79721 79728 79729 79730 79744 79745 79752 79755 79761 79764 79767 79770 79774 79776 79782 79783 79785 79786 79787 79792 79795 79802 79807 79808 79815 79825 79832 79834 79835 79836 79840 79842 79846 79848 79849 79852 79875 79877 79879 79883 79890 79898 79925 79926 79927 79928 79931 79935 79945 79958 79964 79970 79972 79976 79979 79982 79985 80011 80025 80026 80029 80037 80046 80048 80062 80122 80123 80124 80126 80129 80141 80142 80143 80146 80149 80168 80172 80185 80195 80196 80199 80224 80238 80239 80247 80260 80268 80269 80275 80280 80308 41754 41813 41932 41950 53240 53438 53465 53486 53541 53644 53681 53698 53793 53992 54014 54028 54078 55788 55878 55932 55987 55991 56070 56160 56161 56620 56653 56714 57095 57108 57239 57984 58426 58427 60545 60570 60578 60653 60752 60755 60758 60759 60769 60772 60776 60812 60820 60835 60853 60865 60875 60899 60901 60935 60942 60948 60957 60964 60987 61995 62032 62051 62125 62160 62165 62167 62173 62181 62195 62201 62211 62219 62228 62264 62271 62282 62286 62323 62373 64434 64437 64444 64474 64485 64502 64504 64514 64522 64526 64530 64548 64558 64583 64611 64655 64669 64672 64679 64705 64715 64723 64731 64743 64767 64772 64776 64778 64808 64818 64832 64834 64841 64864 64879 64889 64891 64896 64903 64904 64911 73685 73686 73691 73695 73697 73699 73702 73715 73721 73724 73726 73729 73731 73732 73734 73735 73736 73738 73744 73745 73749 73752 73754 73759 73765 73766 73767 73768 73773 73782 73783 73785 73793 73794 73808 73823 73843 73845 73848 73849 73853 73854 73858 73865 73869 73870 73871 73872 73879 73880 73883 73886 73891 73892 73893 73894 73899 73903 73907 73915 73918 73920 73922 73927 73933 73949 73951 73954 73957 73958 73960 73962 73969 73971 73976 73977 73991 73998 73999 74006 74008 74011 74012 74014 74019 74021 74026 74027 74029 74033 74034 74037 74040 74042 74044 74048 74051 74054 74057 74062 74063 74074 74079 74082 74088 74091 74096 74098 74101 74102 74105 74106 74113 74115 74117 74120 74121 74122 74124 74128 74129 74131 74133 74135 74136 74137 74139 74146 74151 74154 74155 74157 74158 74159 74160 74163 74164 74166 74167 74171 74179 74184 74192 74198 74200 74206 74208 74211 74213 74214 74219 74220 74221 74223 74224 74230 74233 74237 74240 74241 74242 74243 74245 74246 74252 74254 74255 74259 74262 74263 74265 74269 74270 74273 74275 74280 74282 74283 74284 74285 74288 74290 74291 74294 74295 74298 74299 74306 74310 74313 74314 74315 74318 74319 74320 74323 74326 74327 75128 75129 75130 75131 75132 75133 75135 75136 75137 75138 75139 75141 75143 75146 75148 75150 75152 75153 75154 75155 75156 75158 75161 75162 75163 75165 75166 75171 75172 75175 75177 75182 75183 75184 75185 75187 75191 75193 75194 75195 75196 75197 75198 75199 75200 75201 75207 75208 75210 75211 75212 75214 75215 75218 75219 75220 75221 75222 75223 75224 75225 75227 75229 75230 75231 75233 75234 75235 75236 75238 75239 75246 75249 75250 75251 75255 75256 75258 75259 75260 75261 75262 75267 75270 75271 75272 75273 75274 75275 75276 75277 75278 75279 75280 75284 75290 75291 75297 75298 75303 75305 75306 75307 75308 75309 75313 75314 75315 75315 75316 75318 75319 75321 75322 75324 75330 75334 75335 75336 75340 75341 75344 75348 75349 75352 75354 75357 75358 75361 75362 75364 75367 75369 75370 75372 75373 75374 75375 75376 75378 75384 75385 75386 75387 75388 75389 75390 75391 75392 75395 75397 75403 75404 75407 75411 75412 75414 75421 75425 75427 75429 75432 75436 75438 75439 75440 75441 75442 75444 75445 75446 75447 75451 75452 75453 75454 75455 75456 75457 75458 75461 75463 75464 75468 75470 75471 75472 75475 75477 75478 75482 75484 75485 75486 75488 75490 75491 75492 75493 75494 75495 75496 75498 75500 75502 75503 75505 75507 75508 75511 75515 75518 75522 75524 75528 75529 75534 75536 75538 75539 75541 75543 75545 75551 75553 75554 75555 75556 75557 75559 75561 75562 75564 75565 75566 75569 75570 75572 75574 75576 75577 75579 75581 75584 75587 75589 75593 75594 75603 75604 75606 75607 75610 75611 75612 75613 75619 75622 75627 75628 75639 75640 75643 75644 75645 75648 75651 75652 75653 75654 75657 75660 75661 75662 75663 75664 75665 75667 75674 75675 75676 75677 75678 75680 75681 75683 75684 75685 75687 75689 75691 75693 75695 75697 75701 75705 75713 75714 75717 75727 75738 75740 108290 108292 108294 108297 108298 108299 108302 108307 108310 108313 108315 108319 108321 108326 108331 108332 108334 108335 108338 108343 108344 108347 108348 108349 108351 108352 108354 108360 108364 108365 108366 108367 108368 108377 108378 108380 108388 108389 108390 108399 108400 108401 108402 108403 108405 108406 108408 108409 108424 108429 108439 108442 108448 108455 108456 108470 108473 108474 108477 108478 108490 108497 108508 108510 108511 108513 108520 108536 108538 108541 108542 108549 108553 108556 108557 108560 108563 108578 108581 108582 108585 108589 108592 108596 108602 108606 108610 108611 108614 108615 108617 108618 108626 108628 108631 108633 108634 108638 108641 108642 108647 108648 108650 108651 108659 108660 108664 108669 108670 108673 108674 108682 108685 108688 108693 108696 108697 108700 108701 108706 108707 108708 108709 108712 108714 108720 108721 108726 108727 108729 108731 108732 108733 108734 108735 108739 108743 108746 108748 108750 108754 108756 108762 108770 108773 108779 108780 108783 108786 108788 108789 108790 108795 108798 108799 108800 108801 108802 108803 108806 108808 108811 108813 108814 108815 108817 108818 108819 108821 108823 108824 108825 108828 108829 108833 108840 108846 108849 108851 108855 108857 108859 108863 108864 108866 108868 108869 108874 108879 108885 108887 108888 108889 108899 108908 108915 108916 108917 108918 108919 108920 108922 108926 108930 108936 108943 108950 108951 108957 108958 108961 108971 108974 108975 108976 108977 108979 108980 108982 108984 108985 108986 108988 108989 108990 108992 108999 109000 109001 109004 109005 109006 109010 109011 109014 109015 109020 109022 109023 109025 109026 109029 109033 109034 109035 109036 109037 109039 109047 109048 109049 109050 109051 109054 109055 109056 109059 109060 109062 109063 109064 109065 109066 109069 109070 109071 109074 109075 109076 109078 109079 109081 109084 109085 109086 109087 109088 109089 109094 109095 109096 109098 109099 109100 109101 109102 109104 109105 109106 109107 109108 109109 109111 109112 109114 109117 109119 109122 109123 109125 109126 109127 109128 109130 109131 109132 109134 109137 109138 109139 109140 109142 109144 109146 109148 109149 109151 109152 109153 109154 109157 109159 109160 109161 109165 109166 109168 109170 109171 109175 109176 109178 109180 108350 108744 108753 109073 109143 108902 109061 108428 108607 108623 108577 108967 108337 108353 108453 108822 108954 109002 108805 108847 108891 108288 108493 108584 108620 108856 108862 108998 108835 109121 108311 108569 108324 108776 108342 108663 108385 109019 109097 132785 132795 132809 132835 132839 132922 132987 133004 133005 133006 133018 133020 133026 133029 133030 133035 133037 133039 133043 133055 133058 133064 133065 133077 133092 133095 133096 133101 133104 133106 133127 133129 133130 133131 133132 133133 133135 133136 133155 133156 133158 109103 108427 108454 108643 108761 108844 108608 108758 108785 108827 108838 108867 109043 109017 109038 108443 108752 108755 108810 108830 108848 109067 109181 108327 108339 108407 108552 108561 108820 108852 108927 108948 109003 109030 109040 109044 109133 108672 108963 108548 108588 108949 108972 109162 109183 109158 108304 108359 108362 108469 108834 108381 108809 109185 108751 108791 108430 108702 109145 109156 108604 108890 108897 108303 108537 108580 108687 108703 109118 108666 109053 108363 108826 108924 108450 109182 108451 108796 109031 109045 109080 108836 109028 109090 108486 108572 108769 108794 108912 108973 108997 133100 109091 109173 108305 109082 109164 108300 108395 108612 108689 108831 108934 108941 108942 108994 109027 133134 108586 108832 133068 108312 132808 108837 147683 147818 108676 108781 108842 133067 109124 108945 108322 108412 108420 108728 108777 108797 108854 108914 108952 108978 109052 109110 109141 109184 132892 108418 108737 108968 109018 108619 108711 109115 108457 108678 108724 108740 108870 108923 108871 109009 109013 109077 108653 108661 108886 108293 108422 108883 108956 108640 108656 108668 108784 108804 108850 108907 108993 109093 109120 108423 108445 108759 108775 109129 35198 35201 35202 35205 35209 35211 35212 35216 35217 35218 35219 35220 35221 35222 35223 35224 35226 35229 35231 35232 35235 35236 35237 35238 35240 35241 35242 35243 35244 35246 35247 35248 35249 35250 35251 35252 35254 35255 35256 35257 35259 35261 35263 35265 35267 35270 35271 35272 35279 35284 35285 35287 35288 35300 39868 39870 39875 39881 39884 39890 39903 39912 39914 39922 39923 39933 39942 39959 39965 39974 39978 39980 39989 39994 39997 40007 40008 40010 40011 40012 40013 40014 40018 40025 40038 40039 40043 40044 40045 40047 40889 40895 40912 40927 40928 40929 40938 40947 40951 40952 40954 40955 40959 40961 40972 40977 40999 41000 41003 41017 41032 41035 41045 41053 41071 41082 41099 41105 41109 41117 41130 41132 41133 41135 41144 41148 41153 41168 41174 41185 41191 41209 41213 41224 41227 41229 41237 41249 41261 41266 41272 41273 41275 41277 41279 41297 41305 41325 41326 41328 41329 41330 41343 41364 41367 41369 41370 41371 41373 41381 41389 41393 41399 41412 41415 41421 41435 41437 41449 41450 41459 41484 41515 41517 41520 41530 41565 41598 41629 41641 47711 47722 47735 47736 47742 47753 47766 47779 47784 47800 47801 47804 47806 47808 47809 47810 47811 47813 47815 47827 47835 47836 47837 47843 47845 47846 47847 47855 47874 47878 47880 47889 47892 47900 47932 47948 47955 47970 47973 47988 47994 48003 48004 48005 48668 48669 48672 48673 48674 48676 48677 48678 48679 48680 48681 48682 48683 48684 48685 48686 48687 48688 48690 48691 48692 48693 48695 48697 48699 48703 48704 48705 48706 48707 48708 48709 48711 48712 48713 48715 48716 48717 48718 48719 48720 48721 48722 48723 48724 48725 48727 48728 48729 48730 48731 48732 48734 48735 48736 48737 48739 48740 48741 48743 48744 48745 48746 51253 51254 51255 51256 51259 51260 51262 51264 51265 51267 51268 51271 51272 51273 51274 51276 51278 51279 51280 51281 51283 51284 51285 51288 51289 51290 51292 51293 51294 51295 51296 51297 51298 51299 51300 51301 51302 51303 51305 51308 51309 51310 51312 51313 51314 51315 51316 51317 51318 51320 51321 51322 51323 51325 51327 51328 51329 51330 51331 51333 51335 51336 51337 51338 51339 51340 51341 51342 51343 51344 51349 51350 51351 51352 51353 51354 51356 51357 51360 51361 51363 51364 51365 51366 51367 51368 51369 51370 51371 51372 51628 51629 51630 51631 51632 51638 51639 51640 51641 51642 51643 51645 51646 51648 51649 51650 51651 51652 51654 51656 51657 51658 51659 51660 51661 51662 51663 51664 51665 51666 51667 51672 51678 51679 51681 51682 51683 51685 51686 51687 51688 51690 51691 51692 51693 51694 51695 51696 51697 51698 51699 51700 51701 51702 51704 51705 51707 51708 51709 51710 51711 51714 51715 51716 51717 51718 51719 51720 51721 51722 51723 51724 51725 51727 51728 51729 51730 51732 51733 51734 51735 51737 51739 51740 51741 57955 60611 60757 62155 62184 62185 86327 86726 86803 87090 87093 87104 106327 106942 107360 107399 107446 107553 107651 107768 144569 144570 144590 144616 144648 144655 35204 35207 35225 35233 35234 35253 35258 35273 39205 39232 39356 39360 39395 39404 39455 39475 39548 39949 39950 40005 41679 41692 41714 41745 41746 41767 41772 41773 41789 41805 41811 41835 41839 41873 41876 41891 41928 41934 41938 41944 41947 41948 41964 41969 41970 41976 41985 42000 42003 42036 42037 48671 48675 48696 48726 51263 51266 51304 51306 51307 51311 51347 51359 51362 51623 51624 51625 51626 51627 51634 51635 51636 51647 51669 51674 51675 51676 51680 51712 51743 51746 51749 51750 51752 51753 51754 51757 51758 51759 51760 51761 51762 51764 51766 51768 51771 51772 52223 52227 52232 52233 52235 52236 52237 52244 52246 52248 52250 52252 52254 52255 52257 52258 52260 52261 52265 52267 52268 52269 52270 52274 52278 52279 52280 52281 52282 52286 52287 52288 52289 52290 52291 52293 52294 52295 52296 52297 52298 52301 52311 52312 52313 52314 52315 52316 52317 52319 52321 52328 52329 52333 52337 52338 52342 52346 52347 52348 52349 52352 52353 52355 52359 52361 52363 52364 52366 52372 52856 52874 52941 53089 53113 53186 53259 53399 53415 53416 53427 53430 53434 53445 53475 53482 53489 53498 53535 53563 53594 53604 53607 53612 53613 53620 53627 53633 53637 53652 53736 53737 53806 53808 53841 53843 53957 55729 55730 55740 55762 55787 55889 55936 55944 55961 55969 55982 55985 55988 55993 56005 56017 56021 56037 56063 56076 56078 56086 56093 56104 56124 56130 56132 56142 56145 56150 56153 56159 56619 56626 56628 56640 56647 56657 56679 56682 56686 56689 56690 56713 56717 56981 56997 56999 57009 57013 57015 57021 57035 57036 57044 57045 57064 57070 57073 57074 57092 57094 57102 57110 57120 57138 57139 57144 57158 57168 57170 57171 57186 57198 57205 57207 57209 57219 57224 57226 57227 57228 57230 57240 57242 57251 57260 57262 57264 57281 57292 57299 57301 57302 57308 57310 57311 57312 57315 57350 57359 57381 57387 57399 57401 57402 57410 57416 57422 57440 57447 57457 57462 57463 57465 57471 57873 57880 57889 57892 57893 57909 57913 57914 57915 57918 57923 57924 57930 57940 57941 57948 57953 57962 57980 57992 57994 58000 58003 58004 58007 58012 58024 58026 58027 58028 58030 58053 58070 58079 58083 58087 58091 58109 58111 58112 58126 58134 58137 58147 58159 58176 58179 58190 58195 58198 58210 58218 58231 58232 58236 58257 58263 58265 58273 58277 58291 58296 58312 58316 58318 58346 58375 58388 58391 58424 58444 58450 58458 58461 58475 58508 58513 58533 58539 58543 58545 58550 58552 58581 58586 58587 58591 58593 58610 60511 60523 60540 60552 60584 60588 60591 60592 60596 60603 60620 60650 60652 60657 60671 60684 60691 60731 60734 60738 60745 60750 60753 60766 60771 60782 60788 60800 60804 60806 60807 60818 60822 60826 60829 60830 60833 60836 60838 60839 60841 60843 60844 60846 60847 60848 60851 60857 60859 60871 60874 60876 60882 60888 60890 60897 60898 60902 60904 60906 60907 60910 60916 60918 60919 60923 60925 60927 60928 60929 60930 60931 60932 60933 60936 60938 60943 60947 60949 60950 60953 60954 60955 60958 60959 60961 60962 60965 60968 60973 60976 60977 60978 60979 60980 60983 60988 60989 60991 61997 61998 62002 62007 62008 62009 62015 62016 62017 62018 62019 62020 62025 62028 62031 62034 62036 62038 62039 62041 62046 62049 62055 62057 62062 62063 62065 62066 62078 62080 62088 62090 62093 62094 62096 62099 62100 62103 62111 62114 62117 62123 62127 62128 62129 62134 62140 62141 62144 62146 62147 62149 62150 62152 62153 62157 62159 62162 62163 62168 62174 62176 62177 62192 62198 62200 62204 62208 62210 62216 62218 62220 62222 62223 62234 62237 62238 62239 62243 62248 62258 62270 62275 62277 62279 62280 62281 62289 62291 62292 62295 62302 62305 62308 62309 62312 62314 62318 62333 62339 62343 62344 62347 62348 62351 62358 62362 62366 62368 62381 62385 62390 64459 64615 64670 64688 64702 64717 64720 64792 64805 64809 64810 64827 64892 109207 109209 109212 109215 109218 109230 109238 109239 109282 109293 109295 109300 109326 109345 109355 109366 109381 109418 109423 109432 109436 109438 109493 109506 109532 109540 109602 109630 109640 109645 109663 109681 109707 109708 109719 109723 109727 109732 109747 109758 109766 109808 109810 109811 109812 109814 109818 109835 109845 109850 109851 109857 109863 109867 109868 109877 109886 109889 109900 109903 109908 109909 109910 109913 109920 109921 109929 109932 109943 109945 109952 109964 109967 109973 109974 109981 109983 109984 109986 109989 109991 109995 109999 110002 110007 110010 110026 110030 110034 110045 110052 110068 110070 110075 110081 111308 111317 111319 111323 111334 111346 111355 111364 111370 111376 111389 111393 111402 111404 111433 111439 111454 111456 111457 111458 111460 111461 111465 111484 111485 111488 111508 111509 111514 111546 111565 111570 111573 111575 111576 111588 111591 111596 111597 111602 111604 111605 111606 111609 111611 111612 111613 111616 111620 111624 111629 111630 111637 111639 111640 111641 111642 111644 111663 111666 111668 111673 111675 111678 111679 111682 111685 111691 111693 111698 111702 111703 111707 111710 111727 111741 111743 111744 111747 111748 111751 111754 111755 111756 111761 111765 111769 111770 111772 111775 111776 111783 111784 111785 111789 111791 111792 111794 111795 111797 111798 111803 111805 111811 111813 111814 111815 111822 111825 111832 111833 111836 111842 111843 111845 111846 111848 111850 111852 111854 111857 111863 111865 111866 111868 111869 111876 111879 111880 111883 111884 111885 111889 111893 111897 111898 111905 111906 111907 111908 111912 111916 111918 111919 111920 111923 111926 111927 111928 111931 111933 111934 111937 111940 111948 111950 111951 111952 111954 111955 111958 111959 111960 111961 111962 111965 111966 111968 111969 111971 111973 111974 111975 111978 111979 111981 111984 111988 111989 111990 111991 111992 111997 111998 112000 112001 112003 112005 112006 112007 112012 112014 112018 112019 112020 112021 112022 112024 112025 112026 112027 112028 112030 112033 112038 112039 112043 112044 112047 112048 112049 112050 112052 112056 112059 112061 112063 112064 112065 112077 112078 112079 112080 112082 112083 112084 112085 112091 112096 112097 112098 112108 112109 112110 112111 112115 112116 112120 112121 112125 112127 112128 112130 112133 112134 112137 112141 112144 112147 112154 112155 112163 112164 112165 112166 112167 112170 112172 112174 112175 112176 112177 112178 112182 112183 112184 112185 112188 112191 112192 112193 112196 112198 112199 112203 112205 112206 112212 112219 112224 112225 112227 112228 112230 112232 112233 112234 112241 112242 112243 112251 112269 112272 112275 112276 112277 112278 112283 112289 112291 112295 112298 112299 112304 112314 112316 112318 112319 112331 112332 112340 112342 112346 112348 112351 112358 112362 112364 112368 112378 112381 112385 112386 112393 112394 112397 112413 112416 112417 112418 112419 112420 112421 112422 112423 112424 112431 112432 112434 112442 112445 112448 112456 112457 112458 112463 112471 112472 112480 139450 139483 139568 139605 139703 111294 111316 111347 111353 111409 111415 111426 111441 111449 111464 111470 111477 111486 111489 111492 111526 111537 111539 111544 111555 111557 111560 111561 111562 111572 111592 111593 111595 111599 111600 111601 111610 111634 111635 111645 111660 111665 111671 111696 111697 111704 111705 111722 111728 111730 111738 111745 111746 111760 111762 111774 111780 111781 111782 111786 111788 111793 111823 111839 111851 111856 111861 111864 111874 111888 111904 111913 111914 111922 111924 111946 111949 111956 111963 111964 111967 111970 111980 111982 111983 111985 111986 111987 111993 111995 112008 112017 112023 112031 112051 112055 112057 112058 112068 112075 112081 112089 112093 112094 112095 112104 112106 112107 112113 112114 112117 112119 112122 112126 112132 112136 112145 112148 112161 112171 112179 112180 112181 112186 112187 112190 112194 112197 112209 112210 112211 112216 112217 112221 112223 112229 112231 112255 112260 112286 112301 112310 112311 112313 112322 112327 112343 112350 112363 112371 112374 112380 112387 112389 112396 112399 112400 112405 112407 112410 112414 112427 112447 112455 112459 112462 112469 112470 112485 133550 133693 111831 111729 112011 111764 112446 111935 111873 112377 112436 112478 111443 111534 111450 111540 112202 112361 112372 112406 111442 112248 112099 112287 112454 111669 111288 111290 111295 111297 111303 111305 111314 111315 111318 111321 111322 111324 111330 111341 111342 111343 111344 111345 111349 111350 111351 111352 111358 111359 111365 111367 111369 111371 111372 111373 111374 111375 111378 111379 111380 111381 111385 111387 111392 111395 111399 111401 111406 111412 111413 111416 111417 111420 111424 111430 111434 111435 111436 111445 111448 111451 111459 111463 111471 111472 111479 111481 111482 111499 111505 111528 111530 111532 111548 111554 111559 111567 111569 111574 111582 111586 111590 111603 111607 111608 111619 111626 111627 111628 111631 111632 111633 111636 111638 111646 111648 111652 111653 111654 111655 111662 111664 111674 111680 111681 111689 111690 111692 111695 111699 111712 111713 111714 111724 111726 111731 111733 111734 111735 111739 111740 111749 111750 111753 111757 111763 111768 111771 111773 111777 111790 111800 111801 111806 111807 111808 111810 111817 111818 111819 111826 111829 111830 111838 111841 111855 111862 111870 111875 111886 111891 111892 111899 111921 111929 111930 111932 111938 111944 111947 111996 111999 112009 112035 112045 112062 112066 112067 112071 112072 112073 112074 112076 112086 112087 112090 112092 112100 112118 112129 112139 112140 112146 112150 112153 112157 112158 112159 112160 112162 112169 112173 112195 112200 112201 112204 112207 112208 112213 112214 112218 112220 112222 112226 112235 112238 112239 112240 112245 112246 112247 112249 112253 112254 112257 112258 112259 112261 112262 112263 112264 112265 112266 112267 112268 112271 112273 112274 112279 112280 112281 112282 112284 112285 112288 112290 112292 112293 112296 112297 112300 112302 112303 112305 112307 112308 112309 112317 112320 112321 112323 112324 112325 112326 112328 112329 112330 112333 112334 112335 112336 112337 112339 112344 112345 112352 112354 112355 112357 112359 112360 112365 112367 112370 112373 112379 112382 112383 112384 112391 112444 112453 132799 132847 132853 132854 132857 132873 132874 132886 132889 132902 132903 132904 132906 132908 132912 132919 132920 132924 132927 132946 132955 132988 132989 30443 30463 30474 30499 30503 30509 30536 30538 30540 33045 33048 33065 33085 33092 33093 33099 33100 33119 39024 39189 39202 39382 39385 39389 39405 39408 39454 39518 39550 39560 39874 39887 39888 39900 39910 39929 39930 39934 39938 39961 39966 39970 39979 40030 40033 40036 40041 40057 40123 40142 40892 40936 40940 40996 41344 41513 41519 41534 41673 41683 41693 41713 41726 41734 41747 41800 41810 41877 41898 41920 41923 41941 41943 41953 41998 42006 42048 42052 42057 43991 47849 48905 52827 52835 52847 52849 52854 52867 52868 52869 52875 52878 52881 52888 52902 52910 52928 52949 52958 53053 53062 53065 53066 53082 53084 53088 53124 53129 53130 53134 53139 53141 53142 53143 53147 53155 53168 53185 53195 53222 53230 53233 53245 53247 53251 53252 53255 53263 53268 53269 53270 53375 53385 53400 53401 53402 53404 53409 53421 53423 53431 53432 53433 53440 53454 53457 53466 53467 53470 53472 53478 53480 53483 53495 53501 53505 53507 53509 53518 53521 53524 53531 53549 53550 53551 53555 53561 53569 53572 53585 53589 53603 53606 53608 53615 53617 53619 53625 53626 53632 53634 53641 53657 53662 53668 53671 53676 53682 53683 53685 53704 53712 53713 53714 53716 53719 53734 53735 53740 53754 53764 53767 53778 53780 53791 53814 53816 53820 53844 53853 53856 53873 53900 53931 54021 54045 54065 55668 55673 55674 55675 55678 55695 55704 55709 55720 55725 55726 55734 55735 55739 55744 55745 55750 55751 55756 55757 55776 55778 55803 55805 55809 55815 55830 55840 55841 55848 55867 55879 55880 55892 55895 55905 55910 55920 55921 55928 55934 55937 55939 55942 55947 55964 55966 55967 55976 55977 55979 55980 55986 55989 55994 56001 56002 56010 56013 56022 56028 56033 56036 56043 56046 56047 56059 56069 56072 56073 56079 56084 56095 56096 56102 56111 56117 56119 56138 56156 56625 56629 56641 56642 56646 56661 56667 56671 56672 56677 56678 56687 56691 56692 56694 56696 56701 56704 56982 57022 57039 57041 57048 57079 57082 57085 57114 57178 57216 57285 57335 57411 57449 57453 57458 57899 57902 57928 57961 57965 57988 58006 58021 58031 58032 58037 58039 58050 58062 58075 58103 58118 58119 58175 58188 58194 58219 58262 58268 58299 58301 58302 58307 58333 58337 58341 58376 58402 58415 58417 58431 58433 58437 58442 58449 58457 58460 58469 58482 58488 58490 58503 58504 58517 58523 58529 58542 58565 58571 58574 58582 58584 58585 58598 58618 58621 82139 82285 82294 82365 84278 84287 84291 84303 84310 84323 84346 84347 84350 84358 84361 84439 84460 84472 84500 84512 84535 84655 84684 84775 86356 86455 144498 144499 144519 144520 144525 144626 144652 146264 146308 146336 146343 147703 147707 147708 147748 147760 147761 147808 111366 111503 111521 111533 111536 111549 111621 111643 111647 111650 111672 111711 112404 112411 111732 111467 111473 111584 111585 111700 111895 112412 112440 111502 111779 111804 136539 111300 111302 111313 111329 111332 111333 111348 111354 111356 111377 111382 111383 111386 111391 111397 111403 111408 111410 111411 111414 111418 111419 111422 111427 111429 111431 111432 111438 111440 111444 111446 111447 111452 111453 111455 111466 111468 111474 111475 111476 111478 111480 111487 111490 111491 111493 111494 111495 111496 111497 111498 111501 111511 111512 111513 111515 111517 111518 111519 111520 111523 111524 111527 111529 111531 111535 111538 111541 111545 111547 111550 111551 111552 111556 111558 111563 111564 111566 111568 111571 111577 111578 111579 111580 111583 111587 111598 111614 111622 111625 111649 111657 111658 111661 111667 111677 111683 111686 111687 111688 111694 111701 111706 111708 111709 111715 111716 111717 111719 111720 111725 111742 111752 111758 111766 111767 111796 111799 111809 111812 111816 111821 111824 111828 111840 111844 111847 111849 111860 111872 111877 111890 111894 111900 111909 111915 111917 111943 111972 112042 112312 112347 112349 112369 112375 112376 112388 112390 112392 112402 112403 112408 112409 112425 112426 112429 112433 112435 112437 112439 112441 112443 112451 112452 112460 112461 112465 112466 112468 112473 112476 112477 112481 112482 112484 112486 112487 30438 30449 30486 33146 35210 35213 35260 35266 35269 35274 35299 38850 38930 39234 40165 41699 41701 41711 41722 41759 41766 42030 42046 47778 48670 48710 48733 48738 51261 51275 51287 51346 51348 51358 51644 51668 51671 51703 51726 51731 51742 52239 52283 52285 52292 52302 52303 52307 52308 52310 52320 52327 52339 52340 52351 52362 52369 52370 52832 52833 52844 52877 53254 53373 53425 53442 53444 53459 53462 53479 53496 53526 53528 53529 53584 53593 53598 53616 53624 53647 53648 53691 53771 53789 53823 53862 53993 54051 55818 56026 56027 56065 56110 56617 56676 56681 57127 57185 57187 57936 58206 58252 58453 58526 60501 60513 60593 60604 60636 60694 60715 60760 60761 60774 60792 60815 60855 60858 60861 60868 60881 60903 60922 60937 60951 60982 60985 60986 62005 62012 62045 62059 62083 62178 62229 62254 62317 62319 62380 64492 64553 64579 64590 64599 64618 64621 64694 64741 64742 64756 64774 64871 64874 64901 73682 73698 73700 73707 73708 73711 73712 73714 73730 73753 73755 73756 73761 73764 73776 73778 73786 73789 73791 73795 73796 73802 73803 73804 73805 73813 73819 73824 73826 73827 73830 73833 73838 73841 73842 73851 73856 73859 73862 73866 73873 73878 73881 73884 73896 73905 73917 73926 73932 73934 73945 73946 73961 73970 73983 73988 73995 74005 74018 74024 74038 74039 74045 74058 74059 74065 74070 74094 74099 74116 74134 74140 74145 74169 74170 74188 74201 74207 74210 74212 74228 74258 74264 74277 74296 74308 74311 74312 74316 74321 75134 75145 75149 75157 75159 75164 75168 75170 75179 75180 75186 75189 75190 75202 75203 75213 75216 75226 75228 75232 75237 75240 75241 75244 75245 75247 75248 75253 75264 75265 75282 75283 75286 75289 75292 75294 75295 75299 75302 75311 75312 75317 75326 75327 75328 75333 75346 75351 75353 75360 75365 75366 75371 75379 75380 75382 75402 75405 75406 75408 75413 75415 75416 75417 75418 75420 75423 75424 75426 75431 75434 75435 75448 75449 75459 75460 75462 75469 75474 75480 75481 75483 75487 75489 75497 75499 75513 75517 75523 75527 75532 75533 75535 75537 75547 75549 75550 75558 75560 75567 75596 75597 75599 75600 75601 75602 75609 75614 75615 75616 75625 75629 75630 75633 75634 75635 75641 75642 75658 75659 75669 75670 75671 75672 75690 75698 75702 75703 75704 75706 75707 75708 75709 75710 75715 75718 75719 75720 75721 75723 75724 75725 75728 75729 75731 75732 75734 75735 75736 75737 75741 75742 75744 75745 75747 75749 75750 75751 75754 75755 75756 75758 75761 75762 75763 75764 75768 75769 75770 75771 75772 75773 75774 75775 75776 77229 77231 77234 77236 77237 77238 77239 77240 77242 77243 77244 77245 77246 77248 77249 77250 77251 77252 77253 77254 77255 77256 77257 77258 77260 77262 77263 77264 77265 77266 77267 77270 77271 77272 77274 77275 77276 77277 77278 77279 77280 77281 77283 77284 77285 77286 77287 77288 77289 77290 77292 77293 77294 77297 77298 77300 77301 77302 77303 77306 77307 77309 77311 77312 77313 77314 77316 77317 77318 77319 77320 77321 77322 77323 77326 77327 77328 77329 77330 77333 77334 77336 77337 77338 77339 77341 77342 77343 77344 77345 77347 77349 77350 77352 77353 77354 77357 77358 77359 77360 77361 77362 77363 77365 77369 77370 77371 77372 77374 77375 77376 77377 77378 77379 77381 77383 77385 77386 77388 77389 77390 77392 77393 77397 77398 77399 77400 77401 77403 77404 77409 77410 77421 77427 77428 77430 77431 77434 77436 77441 77442 77444 77445 77447 77448 77450 77452 77458 77459 77462 77464 77465 77466 77467 77468 77469 77470 77471 77472 77473 77474 77475 77476 77478 77480 77481 77482 77483 77485 77487 77489 77491 77492 77493 77494 77495 77496 77500 77503 77507 77509 77512 77513 77514 77517 77518 77519 77525 77526 77529 77533 77535 77540 77541 77542 77543 77546 77561 77563 77565 77572 77575 77577 77578 77586 77589 77591 77598 77602 77603 77605 77608 77612 77626 77631 77640 77641 77644 77646 77648 77653 77655 77657 77670 77678 77679 77681 77684 77685 77688 77691 77692 77697 77703 77710 77726 77728 77730 77743 77744 77752 77753 77762 77765 77767 77773 77782 77788 77805 77809 77820 77827 77830 77838 77840 77853 77854 77855 77864 77871 77877 77878 77880 77881 77934 77943 35282 48689 48701 51269 51277 51282 51706 52299 52318 58518 78078 81486 81601 81621 81631 81651 81698 81704 81838 81842 81902 82080 82103 82119 82229 82276 82281 82283 82327 82354 82379 82408 82416 82463 82471 82488 82505 82508 82514 82519 82540 82546 82554 82565 84280 84281 84286 84293 84295 84297 84311 84312 84317 84320 84321 84322 84325 84330 84331 84333 84338 84342 84348 84357 84368 84377 84382 84385 84388 84398 84399 84400 84401 84404 84409 84411 84413 84414 84429 84436 84440 84441 84442 84443 84445 84448 84452 84456 84468 84471 84474 84478 84479 84480 84485 84489 84490 84491 84498 84502 84504 84513 84518 84521 84525 84526 84527 84536 84537 84539 84548 84549 84551 84556 84561 84569 84571 84572 84574 84575 84580 84581 84592 84593 84594 84602 84610 84611 84613 84614 84616 84618 84619 84622 84623 84628 84630 84631 84634 84637 84638 84639 84640 84643 84650 84651 84653 84654 84656 84658 84661 84665 84667 84668 84669 84670 84681 84682 84683 84686 84688 84690 84691 84693 84694 84695 84698 84701 84702 84704 84707 84708 84712 84713 84720 84721 84728 84731 84733 84734 84738 84740 84742 84749 84750 84752 84754 84756 84758 84764 84765 84772 84773 84774 84776 84778 84779 84787 84788 84791 84792 84797 84799 84805 84807 84809 84813 84816 84819 84820 84824 84826 84827 84828 84829 84834 84835 84837 84838 84839 84840 84842 84843 84847 84848 84850 84853 84855 84857 84863 84865 84868 84870 84871 84901 84902 84906 84907 84908 84911 84912 84913 84914 84917 84919 84920 84922 84923 84932 84933 84936 84942 84954 84965 84967 84973 84975 84977 84978 84982 84984 84986 84987 84993 84994 84996 73743 77637 77716 79426 79524 79625 79699 79838 32975 33103 33115 33143 47786 52969 53622 60765 60842 60963 62138 62158 62180 62212 62266 73850 73875 73902 73913 74041 74085 74123 74229 74253 74286 75160 75188 75263 75281 75304 75342 75345 75383 75393 75443 75466 75479 75501 75510 75530 75548 75578 75585 75586 75592 75626 75637 75656 75668 75688 75694 75699 75712 75722 75730 75757 75777 77235 77259 77273 77295 77299 77315 77368 77373 77380 77384 77395 77406 77411 77413 77414 77416 77423 77424 77425 77443 77451 77454 77456 77461 77486 77497 77499 77511 77515 77528 77530 77544 77545 77551 77560 77562 77566 77569 77570 77579 77593 77594 77595 77596 77601 77607 77609 77610 77613 77616 77630 77635 77636 77639 77642 77643 77645 77658 77659 77664 77671 77674 77677 77680 77686 77695 77696 77698 77699 77706 77707 77715 77731 77734 77735 77738 77740 77749 77751 77754 77758 77759 77760 77763 77770 77772 77785 77786 77787 77789 77790 77794 77797 77800 77807 77808 77810 77811 77812 77816 77819 77821 77823 77824 77833 77835 77836 77839 77844 77845 77850 77851 77852 77857 77858 77862 77863 77865 77867 77870 77873 77876 77879 77893 77906 77913 77923 77924 77925 77927 77936 77938 77939 77946 77948 77949 77952 77960 77966 77971 78001 78019 78694 60926 62322 64468 64524 64577 64606 64622 64642 64677 64697 64753 64862 64873 73986 74010 74052 74053 74055 74087 74118 74125 74175 75748 77247 77268 77269 77340 77437 77571 77719 77764 77799 77803 77944 78003 78091 78107 78140 78152 78172 78209 78235 78236 78238 78267 78363 78379 78391 78417 78482 78513 78546 78554 78566 78583 78595 78612 78627 78632 78634 78636 78653 78666 78668 78670 78671 78672 78679 78710 78713 78726 78768 78770 78794 78806 78811 78815 78842 78843 78859 78860 78901 78906 78921 78924 78926 79352 79355 79364 79365 79367 79371 79374 79381 79382 79386 79391 79393 79395 79399 79402 79406 79407 79408 79415 79417 79419 79421 79430 79433 79436 79437 79438 79440 79441 79442 79443 79445 79446 79447 79451 79459 79460 79465 79468 79470 79471 79472 79473 79476 79477 79478 79483 79485 79488 79500 79502 79503 79504 79506 79509 79512 79514 79517 79521 79527 79529 79530 79534 79537 79538 79539 79540 79541 79542 79543 79544 79546 79547 79548 79549 79556 79558 79559 79560 79566 79567 79569 79572 79576 79580 79586 79588 79593 79595 79597 79599 79600 79601 79602 79613 79615 79619 79621 79627 79629 79630 79631 79633 79641 79644 79645 79652 79655 79656 79660 79664 79667 79670 79671 79674 79677 79682 79683 79684 79687 79690 79692 79696 79698 79712 79720 79721 79728 79729 79730 79744 79745 79752 79755 79761 79764 79767 79770 79774 79776 79782 79783 79785 79786 79787 79792 79795 79802 79807 79808 79815 79825 79832 79834 79835 79836 79840 79842 79846 79848 79849 79852 79875 79877 79879 79883 79890 79898 79925 79926 79927 79928 79931 79935 79945 79958 79964 79970 79972 79976 79979 79982 79985 80011 80025 80026 80029 80037 80046 80048 80062 80122 80123 80124 80126 80129 80141 80142 80143 80146 80149 80168 80172 80185 80195 80196 80199 80224 80238 80239 80247 80260 80268 80269 80275 80280 80308 41754 41813 41932 41950 53240 53438 53465 53486 53541 53644 53681 53793 53992 54014 54028 54078 55788 55878 55932 55987 55991 56070 56160 56161 56653 56714 57095 57108 57239 57984 58427 60545 60570 60578 60653 60752 60755 60758 60769 60772 60776 60812 60820 60835 60853 60865 60875 60899 60901 60935 60942 60948 60957 60964 60987 61995 62032 62051 62125 62160 62165 62167 62173 62181 62195 62201 62211 62219 62228 62264 62271 62282 62286 62323 62373 64434 64437 64444 64474 64485 64502 64504 64514 64522 64526 64548 64558 64583 64611 64655 64669 64672 64705 64715 64723 64731 64743 64767 64772 64776 64808 64818 64832 64834 64841 64864 64879 64889 64891 64896 64903 64904 64911 73685 73686 73691 73695 73697 73699 73702 73715 73721 73724 73726 73729 73731 73734 73736 73744 73745 73749 73752 73754 73759 73765 73766 73767 73768 73773 73782 73785 73793 73794 73823 73843 73845 73848 73849 73853 73854 73858 73865 73869 73870 73871 73872 73880 73883 73886 73891 73892 73893 73894 73899 73903 73907 73915 73918 73920 73922 73927 73933 73949 73951 73954 73957 73958 73960 73962 73971 73976 73977 73991 73998 73999 74006 74008 74011 74012 74014 74019 74021 74026 74027 74029 74033 74034 74037 74040 74042 74044 74048 74051 74054 74057 74062 74063 74079 74082 74088 74091 74096 74098 74101 74102 74105 74106 74113 74115 74117 74120 74121 74122 74124 74128 74129 74131 74133 74135 74136 74137 74139 74146 74154 74155 74157 74158 74159 74160 74163 74164 74166 74167 74171 74179 74184 74192 74198 74200 74206 74208 74211 74213 74214 74219 74220 74221 74223 74224 74230 74233 74237 74240 74241 74242 74243 74245 74246 74252 74254 74255 74259 74262 74263 74265 74269 74270 74275 74280 74282 74283 74285 74288 74290 74291 74294 74295 74298 74299 74306 74310 74313 74314 74315 74318 74319 74320 74323 74326 74327 75128 75129 75131 75132 75133 75135 75136 75137 75138 75139 75141 75143 75146 75148 75150 75152 75153 75154 75155 75156 75158 75161 75162 75163 75165 75166 75171 75172 75175 75182 75183 75184 75185 75191 75193 75194 75195 75196 75197 75198 75199 75200 75201 75207 75208 75210 75211 75212 75214 75215 75218 75219 75220 75221 75222 75223 75224 75225 75227 75229 75230 75231 75234 75235 75236 75238 75239 75246 75249 75250 75251 75255 75256 75258 75259 75260 75261 75262 75267 75270 75271 75272 75273 75274 75275 75276 75277 75278 75279 75280 75284 75290 75291 75297 75298 75303 75305 75306 75307 75308 75313 75314 75316 75318 75319 75321 75322 75324 75330 75334 75335 75336 75340 75341 75344 75348 75349 75352 75354 75357 75358 75361 75362 75364 75367 75369 75370 75372 75373 75374 75375 75376 75378 75384 75385 75386 75387 75388 75389 75390 75391 75392 75395 75397 75403 75404 75407 75411 75414 75421 75425 75427 75429 75432 75436 75438 75440 75441 75442 75444 75446 75447 75451 75452 75453 75454 75455 75456 75457 75458 75461 75463 75464 75468 75470 75471 75472 75475 75477 75478 75482 75484 75485 75486 75488 75490 75491 75492 75493 75494 75495 75496 75498 75500 75502 75503 75505 75507 75508 75511 75515 75518 75522 75524 75528 75529 75534 75538 75539 75541 75543 75545 75551 75553 75554 75555 75556 75557 75559 75561 75562 75564 75565 75566 75569 75570 75572 75574 75576 75577 75579 75581 75584 75587 75589 75593 75594 75603 75604 75606 75607 75610 75611 75612 75613 75619 75622 75627 75628 75639 75640 75643 75644 75645 75648 75651 75652 75653 75654 75657 75660 75662 75664 75665 75667 75674 75675 75676 75677 75678 75680 75681 75683 75684 75685 75687 75689 75691 75695 75697 75701 75705 75713 75714 75717 75727 75738 75740 
Recalling Firm/
Manufacturer
B. Braun Medical, Inc.
901 Marcon Boulevard
Allentown PA 18109-9512
For Additional Information Contact Kimberly Paris
610-266-0500 Ext. 2669
Manufacturer Reason
for Recall
If the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed, the clip catch on the inside of the pump door may break. If the clip catch is broken and the door opened, free flow protection is ensured. If IV set is then removed from pump without closing the roller clamp, the clip may not re-engage the tubing, leaving an open fluid path with free flow pote
FDA Determined
Cause 2
Device Design
Action B Braun sent a letter on August 16, 2011, to their customers, that detailed the proposed G03 implementation in response to reported customer complaints. The software upgrade with the door jam was agreed upon and the eight hospitals were upgraded in November 2011.
Quantity in Commerce 5769
Distribution Product was distributed to hospitals in CA and MO.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = FRN and Original Applicant = B. BRAUN MEDICAL, INC.
-
-