Class 2 Device Recall Sarns |
|
Date Initiated by Firm |
May 09, 2012 |
Date Posted |
June 27, 2012 |
Recall Status1 |
Terminated 3 on April 24, 2013 |
Recall Number |
Z-1891-2012 |
Recall Event ID |
61731 |
510(K)Number |
K873212
|
Product Classification |
Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
|
Product |
CPB CATHETER KIT- DR HENDRICKSO
Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary bypass procedures.
|
Code Information |
Catalog number: 80003-02 and lot number: MH12, MK30, MP06, NA31, NE18, NF23, NM31, NN28, NP19, and PD05. |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor MI 48103-9586
|
For Additional Information Contact |
Tracy Bellanca 734-741-6173
|
Manufacturer Reason for Recall |
During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips.
Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the
|
FDA Determined Cause 2 |
Material/Component Contamination |
Action |
Terumo sent an Urgent Medical Device Recall letter dated May 9, 2012 to All US first consignees will receive a removal notification letter via Federal Express. The letter identified the affected products, reason for correction and actions to be taken. Customers were instructed to discontinue use of affected products and return for replacement. The letter advised customers to assure that all users are aware of this notice and to fill out and return the attached Response Form. For questions call Terumo CVS Customer Service at 1-800-521-2818 or by fax at 1-800-292-6551. |
Quantity in Commerce |
53,139 (each) in total |
Distribution |
Worldwide Distribution - US (nationwide) including the states of: AL, AR , AZ , CA, CT , FL, IA , IL, IN , KS , LA , MI , NM, MO , NC, NE , NJ , NV, NY , OH, OK , PA, RI , SC , TN, TX, WA , WI, WV and the countries of: BELGIUM, BRAZIL, Byelorussian SSR, CAMBODIA , Dominican Republic, Honduras, Japan, Malaysia, Mexico, Nicaragua, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, UNITED ARAB EMIRATES (UAE), and Vietnam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DWF and Original Applicant = 3M HEALTH CARE, SARNS
|
|
|
|