|
Class 2 Device Recall NeuViz 16 MultiSliced CT Scanner System and NeuViz Dual series CT Scanner System |
|
Date Initiated by Firm |
October 18, 2012 |
Date Posted |
December 04, 2012 |
Recall Status1 |
Terminated 3 on July 08, 2013 |
Recall Number |
Z-0466-2013 |
Recall Event ID |
63553 |
510(K)Number |
K090173 K092742 K062451 K071308
|
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product |
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321
The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. |
Code Information |
NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013. |
Recalling Firm/ Manufacturer |
Philips And Neusoft Medical Systems Co., Ltd. 16 Century Road, Neusoft Park, Hun Nan Industrial Area Shenyang, Liaoning China
|
Manufacturer Reason for Recall |
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
|
FDA Determined Cause 2 |
Device Design |
Action |
Philips and Neusoft Medical Systems Co., Ltd. sent an Urgent Device Correction NeuViz Dual, NeuViz 16 letter, dated Oct. 18, 2012, to all affected US customers and distributors.
The letter described the product, problem, and actions to be taken by the customers. Customers are advised to:
1. Have a mechanic check the patient table top screws fixing. Detailed
instructions are attached as Annex with the letter.
2. If it is found that table top screws are improperly assembled, stop using
the product and contact local Service personnel for repair.
Philips and Neusoft Medical Systems Co., Ltd is in the process of upgrading devices with aforementioned kits and will contact customers to arrange for an update to affected systems.
If you have any questions about this corrective action program, please contact the Service Support Department, Neusoft Medical Systems Co. Ltd. at nms-service@neusoft.com
For questions regarding this recall call 425-487-7665. |
Quantity in Commerce |
20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series |
Distribution |
Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = NEUSOFT MEDICAL SYSTEMS CO., LTD.
|
|
|
|