|
Class 2 Device Recall Plum A Hyperbaric Infusion Pump |
|
Date Initiated by Firm |
February 05, 2012 |
Date Posted |
November 14, 2013 |
Recall Status1 |
Terminated 3 on November 14, 2016 |
Recall Number |
Z-0203-2014 |
Recall Event ID |
64091 |
510(K)Number |
K081412
|
Product Classification |
Pump, infusion - Product Code FRN
|
Product |
Plum A+ Hyperbaric Infusion Pump;
List Number 11005;
Labeled in part: PLUM A+ HYPERBARIC;
Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
|
Code Information |
List No. 11005; Serial Numbers: 16792201, 16792202, 16792205, 16792206, 16792208, 16792209, 16792210, 16792211, 16792213, 16792216, 16792217, 16792219, 16792220, 16792222, 16792223, 16792224, 16792225, 16792226, 16792227, 16792228, 16792230, 16792232, 16792233, 16792234, 16792235, 16792236, 16792238, 16792239, 16792240, 16792241, 16792242, 16792244, 16792245, 16792246, 16792248, 16792249, 16792250, 16792251, 16792252, 16792253, 16792254, 16792255, 16792256, 16792257, 16792258, 16792259, 16792260, 16792261, 16792263, 16792264, 16792265, 16792266, 16792268, 16792269, 16792271, 16792272, 16792273, 18807402, 18807405, 18807406, 18807407, 18807408, 18807409, 18807410, 18807413, 18807414, 18807417, 18807418, 18807420, 18807421, 18807423, 18807427, 18807428, 18807429, 18807430, 18807431, 18807432, 18807433, 18807434, 18807439, 18807443, 18807444, 18807446, 18807449, 18807452, 18807453, 18807455, 18807456, 18807459, 18807463, 18807466, 18807472, 18807474, 18807478, 18807484, 18807485, 18807486, 18807490, 18807491, 18807494, 18807496, 18807499, 18807500, 18807501, 18807506, 18807509, 18807510, 18807511, 18807513, 18807514, 18807518, 18807523, 18807524, 18807529, 18807535, 18807536, 18807539, 18807540, 18807541, 18807542, 18807543, 18807545, 18807546, 18807556, 18807557, 18807568, 18807574, 18807575, 18807576, 18807578, 18807581, 18807584, 18807585, 18807586, 18807588, 18807589, 18807590, 18807591, 18807596, 18807597, 18807598, 18807599, 18807600, 18807601, 18807602, 18807604, 18807605, 18807607, 18807608, 18807609, 18807610, 18807611, 18807612, 18807613, 18807614, 18807617, 18807618, 18807619, 18807620, 18807623, 18807625, 18807626, 18807627, 18807628, 18807630, 18807631, 18807632, 18807633, 18807634, 18807636, 18807637, 18807639, 18807640, 18807695, 18807696, 18807697, 18807698, 18807699, 18807700 |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
For Additional Information Contact |
224-212-2000
|
Manufacturer Reason for Recall |
Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Hospira sent an URGENT DEVICE FIELD CORRECTION letter dated February 5, 2013 was sent to all direct accounts notifying them of the issue. For device users who experience these issues, Hospira recommends the following actions be taken:
1) Remove the pump from service and perform a Performance Verification Test (PVT) Distal Occlusion Test to determine if the distal pressure sensor is performing correctly.
2) If the device does not pass the PVT, it may indicate that the distal pressure sensor has drifted. Perform the appropraite troubleshooting and repair activities defined by the user's facility, whihc may include returning the device to Hospira for further diagnosis and servicing.
The remediation for this issue is to perform a recalibration of the distal pressure sensor. Hospira also intends to release in 2013 a recommendation of a yearly test to determine if the distal pressure sensor's calibration has drifted.
Customers who suspect that their device(s) may be affected by this issue are urged to contact Hospira at 1-800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST) to report the issue. Hospira will provide loaner devices if necessary.
Customers are requested to complete and return the reply form included with the notification letter per the instructions on the reply form.
For questions regarding this recall call 224-212-2000. |
Quantity in Commerce |
List No. 11005: 179 units |
Distribution |
Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Croatia, Dominican Republic, Turks and Caicos. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = HOSPIRA, INC.
|
|
|
|