|
Class 2 Device Recall Dash 3000/4000/5000 Patient Monitor |
|
Date Initiated by Firm |
April 03, 2013 |
Date Posted |
August 28, 2013 |
Recall Status1 |
Terminated 3 on July 25, 2016 |
Recall Number |
Z-2115-2013 |
Recall Event ID |
66016 |
510(K)Number |
K073462
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
GE Healthcare, Dash 3000, Dash 4000 and Dash 5000.
Product Usage: The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. |
Code Information |
Mfg Lot or Serial # AAB04397901GR SD007210479GR SD007231230GR SD007241751GR SD007366729GR SD007387285GR SD007387288GR SD007387291GR SD007387292GR SD007387294GR SD007439614GR SD008056174GR SD008076950GR SD008087478GR SD008138396GR SD008168769GR SD008219818GR SD008219821GR SD008356627GR SD008388272GR SD008388551GR SD008419461GR SD008419463GR SD008419495GR SD008421124GR SD008429787GR SD008429950GR SD008452355GR SD008452399GR SD008463177GR SD008516155GR SD008516281GR SD009037533GR SD009047690GR SD009088931GR SD009088944GR SD009099025GR SD009119669GR SD009121032GR SD009121171GR SD009129978GR SD009141371GA SD009141699GR SD009151940GA SD009182687GR SD009182701GR SD009182715GR SD009193075GR SD009193076GR SD009193084GR SD009193086GR SD009223767GR SD009223886GR SD009234001GR SD009234021GR SD009244557GR SD009244558GR SD009244559GR SD009244560GR SD009244564GR SD009244732GR SD009244820GR SD009255010GR SD009316784GR SD009316799GR SD009316886GR SD009316887GR SD009316899GR SD009316900GR SD009316901GR SD009316906GR SD009316909GR SD009316914GR SD009316918GR SD009316928GR SD009316933GR SD009317060GR SD009327366GR SD009327369GR SD009327371GR SD009327375GR SD009327376GR SD009358078GR SD009358079GR SD009368552GR SD009368640GR SD009400079GR SD009400082GR SD009400090GR SD009400096GR SD009400105GR SD009400109GR SD009409707GR SD009409720GR SD009409722GR SD009409748GR SD009409757GR SD009409767GR SD009409773GR SD009420737GR SD009472159GR SD009472167GR SD009513762GA SD010054455GR SD010074910GA SD010085194GR SD010095383GR SD010105782GR SD010116508GR SD010127167GR SD010127275GR SD010127663GR SD010127727GR SD010127744GA SD010127745GR SD010127802GR SD010127803GR SD010127804GR SD010127808GR SD010127809GR SD010127814GR SD010127815GR SD010138406GA SD010138409GA SD010138435GR SD010138442GA SD010138469GR SD010138494GR SD010148544GA SD010148546GA SD010148550GA SD010148560GR SD010148561GR SD010148656GR SD010148660GR SD010148692GR SD010148707GR SD010148711GR SD010148712GA SD010148719GR SD010148722GR SD010158817GR SD010158820GR SD010158981GR SD010159059GR SD010159065GR SD010179591GA SD010189668GR SD010189669GR SD010199870GA SD010199872GA SD010209915GA SD010209920GA SD010209972GA SD010210554GA SD010220665GR SD010220672GR SD010220683GR SD010241191GR SD010241193GR SD010251365GA SD010251520GA SD010261863GA SD010261884GR SD010261920GA SD010262032GR SD010262037GR SD010262052GR SD010262054GR SD010262058GR SD010272265GA SD010272311GA SD010272336GA SD010272351GA SD010272356GA SD010272361GA SD010272362GA SD010272435GA SD010272523GA SD010292604GA SD010292606GA SD010292653GA SD010292710GA SD010292766GA SD010292798GA SD010292838GA SD010292847GR SD010292927GA SD010292934GA SD010292976GA SD010293096GA SD010293101GA SD010293105GA SD010293130GA SD010293137GA SD010293140GA SD010303154GA SD010303318GA SD010303373GA SD010303415GA SD010303438GA SD010303453GA SD010303518GA SD010303527GA SD010303547GA SD010303620GA SD010313737GA SD010313739GA SD010313744GA SD010313748GA SD010313750GA SD010313752GA SD010313757GA SD010313758GA SD010313767GA SD010313768GA SD010313781GA SD010313794GA SD010313796GA SD010313798GA SD010313826GA SD010313828GA SD010313844GA SD010313845GA SD010313852GA SD010313854GA SD010313856GA SD010313863GA SD010313867GA SD010313875GA SD010313878GA SD010313886GA SD010313920GA SD010313948GA SD010313953GA SD010313992GA SD010313997GA SD010314032GA SD010314112GA SD010314117GA SD010314123GA SD010314149GA SD010314150GA SD010314159GA SD010314161GA SD010314168GA SD010314172GA SD010314256GA SD010314259GA SD010314262GA SD010314293GA SD010314305GA SD010314357GA SD010314359GA SD010314360GA SD010314361GA SD010314362GA SD010314364GA SD010314367GA SD010314369GA SD010314401 SD010314407GA SD010314418GA SD010314419GA SD010314421GA SD010314423GA SD010314426GA SD010314427GA SD010314428GA SD010324430GA SD010324433GA SD010324434GA SD010324435GA SD010324476GA SD010324484GA SD010324485GA SD010324495GA SD010324497GA SD010324515GA SD010324527GA SD010324561GA SD010324577GA SD010324593GA SD010324611GA SD010324632GA SD010324633GA SD010324636GA SD010324637GA SD010324646GA SD010324705GA SD010324706GA SD010324707GA SD010324708GA SD010324710GA SD010324711GA SD010324712GA SD010324713GA SD010324719GA SD010324721GA SD010324722GA SD010324723GA SD010324724GA SD010324726GA SD010324727GA SD010324735GR SD010324736GA SD010324737GA SD010324739GA SD010324745GA SD010324750GA SD010334811GA SD010334813GA SD010334814GA SD010334815GA SD010334816GA SD010334818GA SD010334824GA SD010334844GA SD010334862GA SD010334864GA SD010334879GA SD010334884GA SD010334914GA SD010334917GR SD010334918GR SD010334919GR SD010334921GR SD010334922GR SD010334923GR SD010334927GA SD010334928GR SD010335107GA SD010335109GA SD010335110GA SD010335111GA SD010335113GA SD010335117GA SD010335134GA SD010335139GA SD010335141GA SD010335142GA SD010335144GA SD010335147GA SD010335150GAMKE SD010335151GA SD010335152GAMKE SD010335154GA SD010335155GA SD010335157GA SD010335158GA SD010335162GA SD010335175GA SD010335176GA SD010335177GA SD010335178GA SD010335179GA SD010335182GA SD010335183GA SD010335185GA SD010335188GA SD010335194GA SD010345212GA SD010345213GA SD010345215GA SD010345217GA SD010345219GA SD010345220GA SD010345235GA SD010345240GA SD010345244GA SD010345246GA SD010345248GA SD010345272GA SD010345278GA SD010345279GA SD010345280GA SD010345281GA SD010345282GA SD010345284GA SD010345285GA SD010345286GA SD010345288GA SD010345290GA SD010345291GA SD010345292GA SD010345295GA SD010345300GA SD010345317GA SD010345318GA SD010345319GA SD010345345GA SD010345351GA SD010345352GA SD010345356GA SD010345358GA SD010345359GA SD010345362GA SD010345378GA SD010345385GA SD010345390GA SD010345391GA SD010345397GA SD010345408GA SD010345413GA SD010345415GA SD010345417GA SD010345418GA SD010345420GA SD010345423GA SD010345433GA SD010345450GA SD010345457GA SD010345459GA SD010345462GA SD010345463GA SD010345464GA SD010345466GA SD010345467GA SD010345468GA SD010345470GA SD010345471GA SD010345484GA SD010345486GA SD010345490GA SD010345494GA SD010345497GAMKE SD010345498GA SD010345499GA SD010345500GA SD010345501GA SD010345505GA SD010345507GA SD010345518GA SD010345526GA SD010345530GA SD010345532GA SD010345536GA SD010345543GA SD010345549GA SD010355564GA SD010355577GA SD010355720GA SD010365928GA SD010365975GA SD010365976GR SD010365981GA SD010366046GA SD010366058GA SD010366059GA SD010366060GR SD010366062GR SD010366064GR SD010366066GA SD010366112GA SD010366135GA SD010366163GA SD010366164GA SD010366183GA SD010366184GA SD010366186GA SD010366187GA SD010366188GA SD010366255GA SD010366289GA SD010366290GA SD010366292GA SD010366294GA SD010366296GA SD010366301GA SD010366318GR SD010366397GA SD010366413GA SD010366421GA SD010366434GA SD010366462GA SD010366464GA SD010366465GA SD010366466GA SD010366467GA SD010366469GA SD010366470GA SD010366471GA SD010366472GA SD010366473GA SD010366508GA SD010366511GA SD010366514GA SD010366522GA SD010366525GA SD010366533GA SD010376553GA SD010376565GA SD010376573GA SD010376588GA SD010376606GA SD010376613GA SD010376617GA SD010376618GA SD010376619GA SD010376620GA SD010376621GA SD010376622GA SD010376631GA SD010376632GA SD010376635GA SD010376636GA SD010376637GA SD010376640GA SD010376647GA SD010376648GA SD010376651GA SD010376669GA SD010376677GA SD010376681GA SD010376709GA SD010376710GA SD010376711GA SD010376715GA SD010376716GA SD010376717GA SD010376718GA SD010376722GA SD010376723GA SD010376726GA SD010376727GA SD010376728GA SD010376729GA SD010376734GA SD010376738GA SD010376787GA SD010376806GA SD010376809GA SD010376814GA SD010376817GA SD010376818GA SD010376819GA SD010376820GA SD010376821GA SD010376824GA SD010376826GA SD010376827GA SD010376951GA SD010376952GA SD010376953GA SD010376955GA SD010376956GA SD010376957GA SD010376960GA SD010376961GA SD010376965GA SD010376966GA SD010376967GA SD010376968GA SD010376969GA SD010376972GA SD010376973GA SD010376974GA SD010376976GA SD010376977GA SD010377021GA SD010377023GA SD010377024GA SD010377025GA SD010377028GA SD010377029GA SD010377031GA SD010377034GA SD010377035GA SD010377038GA SD010377039GA SD010377041GA SD010377044GA SD010377046GA SD010377048GA SD010377049GA SD010377050GA SD010377051GA SD010377052GA SD010377054GA SD010377055GA SD010377056GA SD010377057GA SD010377066GA SD010377072GA SD010377119GA SD010377136GA SD010377137GA SD010377155GA SD010377157GA SD010377158GA SD010377159GA SD010377160GA SD010377161GA SD010377162GA SD010377168GA SD010377169GA SD010377172GA SD010377174GA SD010377176GA SD010377178GA SD010377181GA SD010387182GA SD010387183GA SD010387184GA SD010387185GA SD010387186GA SD010387190GA SD010387191GA SD010387194GA SD010387196GA SD010387199GA SD010387201GA SD010387202GA SD010387203GA SD010387204GA SD010387205GA SD010387211GA SD010387218GA SD010387219GA SD010387220GA SD010387223GA SD010387226GA SD010387229GA SD010387231GA SD010387236GA SD010387237GA SD010387302GA SD010387307GA SD010387308GA SD010387310GA SD010387311GA SD010387314GA SD010387315GA SD010387316GA SD010387322GA SD010387334GA SD010387335GA SD010387336GA SD010387337GA SD010387338GA SD010387339GA SD010387340GA SD010387341GA SD010387342GA SD010387343GA SD010387344GA SD010387345GA SD010387346GA SD010387347GA SD010387349GA SD010387350GA SD010387351GA SD010387352GA SD010387353GA SD010387354GA SD010387355GA SD010387356GA SD010387357GA SD010387359GA SD010387362GA SD010387365GA SD010387366GA SD010387380GA SD010387381GA SD010387383GA SD010387403GA SD010387417GA SD010387418GA SD010387421GA SD010387422GA SD010387428GA SD010387433GA SD010387442GA SD010387443GA SD010387444GA SD010387446GA SD010387447GA SD010387448GA SD010387449GA SD010387450GA SD010387451GA SD010387480GA SD010387493GA SD010387515GA SD010387519GA SD010387525GA SD010387527GA SD010387533GA SD010387541GA SD010387542GA SD010387549GA SD010387552GA SD010387570GA SD010387571GA SD010387572GA SD010387591GA SD010387592GA SD010387595GA SD010387597GA SD010387598GA SD010387599GA SD010387601GA SD010387602GA SD010387603GA SD010387604GA SD010387605GA SD010387607GA SD010387608GA SD010387611GA SD010387613GA SD010387616GA SD010387617GA SD010387630GA SD010387631GA SD010387634GA SD010387635GA SD010387641GA SD010387643GA SD010387644GA SD010387646GA SD010387647GA SD010387649GA SD010387650GA SD010387653GA SD010387658GA SD010387659GA SD010387662GA SD010387663GA SD010387665GA SD010387666GA SD010387667GA SD010387668GA SD010387669GA SD010387670GA SD010387690GA SD010387692GA SD010387693GA SD010387694GA SHQ12319909SA SD010387697GA SD010387700GA SD010387701GA SD010387703GA SD010387719GA SD010387720GA SD010387721GA SD010387722GA SD010387723GA SD010387726GA SD010387727GA SD010387730GA SD010387731GA SD010387733GA SD010387735GA SD010387740GA SD010387744GA SD010387745GA SD010387748GA SD010387749GA SD010387752GA SD010387753GA SD010387755GA SD010387756GA SD010387757GA SD010387758GA SD010387759GA SD010387760GA SD010387762GA SD010387766GA SD010387768GA SD010387771GA SD010387772GA SD010387773GA SD010387775GA SD010387778GA SD010387779GA SD010387782GA SD010387783GR SD010387785GA SD010387786GA SD010387787GA SD010387816GA SD010387825GA SD010387827GA SD010387830GA SD010387831GA SD010387834GA SD010387841GA SD010387844GA SD010387846GA SD010387847GA SD010387850GA SD010387851GA SD010387856GA SD010397918GA SD010397919GA SD010397920GA SD010397921GA SD010397922GA SD010397923GA SD010397925GA SD010397926GA SD010397927GA SD010397928GA SD010397929GA SD010397930GA SD010397931GA SD010397933GA SD010397934GA SD010397936GA SD010397937GA SD010397939GA SD010397940GA SD010397941GA SD010397942GA SD010397943GA SD010397945GA SD010397946GA SD010397947GA SD010397990GA SD010397996GA SD010397997GA SD010397998GA SD010398001GA SD010398002GA SD010398003GA SD010398006GA SD010398007GA SD010398008GA SD010398009GA SD010398010GA SD010398011GA SD010398012GA SD010398013GA SD010398014GA SD010398015GA SD010398016GA SD010398017GA SD010398018GA SD010398020GA SD010398021GA SD010398025GA SD010398026GA SD010398030GA SD010408033GA SD010408034GA SD010408035GA SD010408036GA SD010408037GA SD010408039GA SD010408040GA SD010408043GA SD010408045GA SD010408046GA SD010408063GA SD010408064GA SD010408065GA SD010408069GA SD010408070GA SD010408071GA SD010408074GA SD010408079GA SD010408081GA SD010408082GA SD010408085GA SD010408086GA SD010408091GA SD010408093GA SD010408094GA SD010408095GA SD010408099GA SD010408100GA SD010408101GA SD010408102GA SD010408104GA SD010408106GA SD010408108GA SD010408109GA SD010408110GA SD010408112GA SD010408113GA SD010408115GA SD010408117GA SD010408122GA SD010408123GA SD010408124GA SD010408127GA SD010408128GA SD010408129GA SD010408153GA SD010408154GA SD010408155GA SD010408156GA SD010408157GA SD010408158GA SD010408159GA SD010408160GA SD010408161GA SD010408162GA SD010408195GA SD010408198GA SD010408201GA SD010408202GR SD010408203GA SD010408204GA SD010408205GA SD010408206GA SD010408208GA SD010408209GA SD010408213GR SD010408216GR SD010408218GR SD010408220GA SD010408223GA SD010408224GR SD010408228GA SD010408232GA SD010408234GA SD010408236GA SD010408237GA SD010408238GA SD010408242GA SD010408248GA SD010408262GA SD010408266GA SD010408268GA SD010408270GA SD010408272GA SD010408274GA SD010408275GA SD010408276GA SD010408277GA SD010408278GA SD010408279GA SD010408280GA SD010408282GA SD010408284GA SD010408285GA SD010408286GA SD010408289GA SD010408292GA SD010408293GA SD010408294GA SD010408295GA SD010408296GA SD010408297GA SD010408298GA SD010408308GA SD010408323GA SD010408337GA SD010408343GA SD010408344GA SD010408346GA SD010408350GA SD010408352GA SD010408354GA SD010408355GA SD010408358GA SD010408361GA SD010408362GA SD010408363GA SD010408364GA SD010408365GA SD010408366GA SD010408367GA SD010408369GA SD010408370GA SD010408371GA SD010408372GA SD010408373GA SD010408374GA SD010408375GA SD010408377GA SD010408378GA SD010408379GA SD010408380GA SD010408383GA SD010408384GA SD010408385GA SD010408387GA SD010408388GA SD010408390GA SD010408391GA SD010408392GA SD010408393GA SD010408394GA SD010408395GA SD010408396GA SD010408407GA SD010408410GA SD010408414GA SD010408415GA SD010418447GA SD010418451GA SD010418453GA SD010418454GA SD010418456GA SD010418506GA SD010418510GA SD010418519GA SD010418522GA SD010418523GA SD010418524GA SD010418525GA SD010418526GA SD010418527GA SD010418528GA SD010418529GA SD010418531GA SD010418532GA SD010418533GA SD010418535GA SD010418536GA SD010418538GA SD010418539GA SD010418541GA SD010418543GA SD010418555GA SD010418561GA SD010418585GA SD010418592GA SD010418597GA SD010418607GA SD010418608GA SD010418609GA SD010418610GA SD010418612GA SD010418614GA SD010418616GA SD010418618GA SD010418620GA SD010418621GA SD010418622GR SD010418624GR SD010418625GA SD010418626GA SD010418627GA SD010418630GA SD010418631GA SD010418632GA SD010418633GA SD010418634GA SD010418635GA SD010418636GA SD010418637GA SD010418638GA SD010418639GA SD010418640GA SD010418641GA SD010418642GA SD010418643GA SD010418644GA SD010418645GR SD010418647GA SD010418650GA SD010418651GR SD010418652GA SD010418653GA SD010418654GA SD010418655GA SD010418656GA SD010418657GA SD010418659GA SD010418688GA SD010418694GA SD010418695GA SD010418696GA SD010418697GA SD010418699GA SD010418700GA SD010418704GA SD010418705GA SD010418706GA SD010418708GA SD010418709GA SD010418710GA SD010418711GA SD010418712GA SD010418713GA SD010418714GA SD010418715GA SD010418717GA SD010418718GA SD010418719GA SD010418720GA SD010418744GA SD010418754GA SD010418758GR SD010418759GA SD010418760GA SD010418761GA SD010418762GA SD010418763GR SD010418764GA SD010418765GA SD010418766GA SD010418767GA SD010418768GR SD010418770GA SD010418771GA SD010418772GR SD010418773GA SD010418774GA SD010418775GA SD010418776GA SD010418778GA SD010418781GA SD010418786GA SD010418787GA SD010418788GA SD010418789GA SD010418790GA SD010418791GA SD010418796GA SD010418798GA SD010418800GA SD010418850GA SD010418858GA SD010418873GA SD010418879GA SD010418881GA SD010418882GA SD010418885GA SD010418886GA SD010418887GA SD010418888GA SD010418889GA SD010418890GA SD010418891GA SD010418892GA SD010418893GA SD010418894GA SD010418900GA SD010418901GA SD010418904GA SD010418906GA SD010418907GA SD010418908GA SD010418909GA SD010418910GA SD010418911GA SD010418913GA SD010428982GA SD010428987GA SD010428988GA SD010428991GA SD010428994GA SD010428995GA SD010428996GA SD010428997GA SD010428999GA SD010429000GA SD010429002GR SD010429003GA SD010429004GA SD010429005GA SD010429007GA SD010429010GA SD010429011GA SD010429012GA SD010429014GR SD010429015GA SD010429016GA SD010429017GA SD010429018GA SD010429040GA SD010429041GA SD010429042GA SD010429043GA SD010429044GA SD010429045GA SD010429046GA SD010429047GA SD010429048GA SD010429049GA SD010429050GA SD010429051GA SD010429091GA SD010429092GA SD010429093GA SD010429094GA SD010429095GA SD010429096GA SD010429104GA SD010429106GA SD010429108GA SD010429110GA SD010429116GA SD010429117GA SD010429119GA SD010429120GA SD010429122GA SD010429123GA SD010429126GA SD010429132GA SD010429134GA SD010429136GA SD010429137GA SD010429139GA SD010429140GA SD010429141GA SD010429142GA SD010429143GA SD010429145GA SD010429146GA SD010429147GA SD010429148GA SD010429149GA SD010429152GA SD010429153GA SD010429156GA SD010429157GA SD010429158GA SD010429159GA SD010429160GA SD010429161GA SD010429162GA SD010429163GA SD010429211GA SD010429218GA SD010429219GA SD010429221GA SD010429223GA SD010429225GA SD010429227GA SD010429228GA SD010429229GA SD010429233GA SD010429234GA SD010429235GA SD010429236GA SD010429237GA SD010429238GA SD010429239GA SD010429241GA SD010429302GA SD010429303GA SD010429304GA SD010430002SA SD010430004SA SD010430005SA SD010430008SA SD010430009SA SD010430012SA SD010430013SA SD010430014SA SD010430015SA SD010430017SA SD010430018SA SD010430019SA SD010430020SA SD010430021SA SD010430023SA SD010430024SA SD010430025SA SD010430027SA SD010430030SA SD010430031SA SD010430044SA SD010430045SA SD010430046SA SD010430047SA SD010430048SA SD010430049SA SD010430051SA SD010430052SA SD010430054SA SD010430055SA SD010430056SA SD010430056SR SD010430057SA SD010430057SR SD010430058SA SD010430059SA SD010430060SA SD010430061SA SD010430065SA SD010430067SA SD010430072SA SD010430073SA SD010430074SA SD010430075SA SD010430078SA SD010430079SA SD010430080SA SD010430081SA SD010449437GA SD010449438GA SD010449439GA SD010449440GA SD010449441GA SD010449442GA SD010449443GA SD010449444GA SD010449445GA SD010449447GA SD010449507GA SD010449508GA SD010449510GA SD010449515GA SD010449541GA SD010449542GA SD010449544GA SD010449546GA SD010449550GA SD010449551GA SD010449552GA SD010449557GA SD010449558GA SD010449559GA SD010449566GA SD010449567GA SD010449568GA SD010449580GA SD010449581GA SD010449583GA SD010449587GA SD010449616GA SD010449617GA SD010449618GA SD010449619GA SD010449620GA SD010449621GA SD010449622GA SD010449623GA SD010449624GA SD010449625GA SD010449626GA SD010449627GA SD010449628GA SD010449629GA SD010449630GA SD010449631GA SD010449633GA SD010449646GA SD010449647GA SD010449658GA SD010449661GA SD010449663GA SD010449668GA SD010449671GA SD010449672GA SD010449675GA SD010449677GA SD010449680GA SD010449685GA SD010449690GA SD010449691GA SD010449694GA SD010449696GA SD010449697GA SD010449698GA SD010449699GA SD010449703GA SD010449704GA SD010450002GA SD010450003GA SD010450004GA SD010450012GA SD010450014GA SD010450015GA SD010450016GA SD010450017GA SD010450023GA SD010450025GA SD010450026GA SD010450027GA SD010450028GA SD010450031GA SD010450037GA SD010450038GA SD010450039GA SD010450041GA SD010450044GA SD010450046GA SD010450047GA SD010450049GA SD010450050GA SD010450052GA SD010450055GA SD010450056GA SD010450059GA SD010450060GA SD010450061GA SD010450062GA SD010450067GA SD010450071GA SD010450073GA SD010450077GA SD010450079GA SD010450081GA SD010450087GA SD010450088GA SD010459706GA SD010459710GA SD010459711GA SD010459713GA SD010459714GA SD010459716GA SD010459717GA SD010459718GA SD010459720GA SD010459721GA SD010459722GA SD010459723GA SD010459724GA SD010459728GA SD010459729GA SD010459730GA SD010459731GA SD010459732GA SD010459733GA SD010459735GA SD010459738GA SD010459740GA SD010459745GA SD010459746GA SD010459747GA SD010459751GA SD010459752GA SD010459754GA SD010459755GA SD010459757GA SD010459758GA SD010459762GA SD010459763GA SD010459764GA SD010459765GA SD010459766GA SD010459767GA SD010459768GA SD010459769GA SD010459771GA SD010459772GA SD010459779GA SD010459780GA SD010459781GA SD010459783GA SD010459784GA SD010459785GA SD010459786GA SD010459787GA SD010459788GA SD010459789GA SD010459791GA SD010459795GA SD010459796GA SD010459798GA SD010459800GA SD010459802GA SD010459806GA SD010459807GA SD010459808GA SD010459809GA SD010459810GA SD010459811GA SD010459812GA SD010459814GA SD010459815GA SD010459816GA SD010459817GA SD010459818GA SD010459819GA SD010459820GA SD010459822GA SD010459823GA SD010459824GA SD010459825GA SD010459826GA SD010459828GA SD010459830GA SD010459833GA SD010459834GA SD010459835GA SD010459836GA SD010459837GA SD010459838GA SD010459839GA SD010459840GA SD010459841GA SD010459842GA SD010459843GA SD010459844GA SD010459846GR SD010459847GA SD010459848GA SD010459849GA SD010459850GA SD010459852GA SD010459853GA SD010459856GA SD010459857GA SD010459858GA SD010459859GA SD010459860GA SD010459861GA SD010459862GA SD010459863GA SD010459864GA SD010459865GA SD010459866GA SD010459867GA SD010459868GA SD010459870GA SD010459874GA SD010459875GA SD010459876GA SD010459877GA SD010459878GA SD010459879GA SD010459880GA SD010459881GA SD010459882GA SD010459883GA SD010459884GA SD010459885GA SD010459888GA SD010459889GA SD010459891GA SD010459892GA SD010459893GA SD010459894GA SD010459895GA SD010459896GA SD010459897GA SD010459898GA SD010459899GA SD010459900GA SD010459901GA SD010459902GA SD010459903GA SD010459904GA SD010459905GA SD010459906GA SD010459907GA SD010459908GA SD010459909GA SD010459910GA SD010459911GA SD010459912GA SD010459913GA SD010459914GA SD010459916GA SD010459917GA SD010459918GA SD010459919GA SD010459924GA SD010459929GA SD010459935GA SD010459936GA SD010459938GA SD010459940GA SD010459941GA SD010459942GA SD010459943GA SD010459948GA SD010459949GA SD010459950GA SD010459952GA SD010459953GA SD010459954GA SD010459956GA SD010459958GA SD010459959GA SD010459960GA SD010459962GA SD010459964GA SD010459965GA SD010459967GA SD010459969GA SD010459970GA SD010459972GA SD010459973GA SD010459974GA SD010459977GA SD010459978GA SD010459980GA SD010459981GA SD010459983GA SD010460094GA SD010460095GA SD010460097GA SD010460099GA SD010460100GA SD010460103GA SD010460105GA SD010460106GA SD010460108GA SD010460109GA SD010460113GA SD010460114GA SD010460115GA SD010460116GA SD010460118GA SD010460119GA SD010460121GA SD010460122GA SD010460124GA SD010460125GA SD010460128GA SD010460129GA SD010460130GA SD010460131GA SD010460133GA SD010460134GA SD010460135GA SD010460136GA SD010460138GA SD010460139GA SD010460140GA SD010460141GA SD010460142GA SD010460143GA SD010460144GA SD010460145GA SD010460149GA SD010460150GA SD010460151GA SD010460153GA SD010460156GA SD010460158GA SD010460159GA SD010460160GA SD010460162GA SD010460164GA SD010460166GA SD010460168GA SD010460169GA SD010460170GA SD010460175GA SD010460177GA SD010460180GA SD010460181GA SD010460182GA SD010460183GA SD010460184GA SD010460185GA SD010460186GA SD010460187GA SD010460188GA SD010460189GA SD010460190GA SD010460191GA SD010460192GA SD010460193GA SD010460194GA SD010460195GA SD010460196GA SD010460197GA SD010460198GA SD010460199GA SD010460200GA SD010460201GA SD010460202GA SD010460203GA SD010460205GA SD010460206GA SD010460207GA SD010460208GA SD010460209GA SD010460210GA SD010460211GA SD010460212GA SD010460213GA SD010460217GA SD010460219GA SD010460220GA SD010460221GA SD010460222GA SD010460223GA SD010460224GA SD010460225GA SD010460226GA SD010460227GA SD010460229GA SD010460230GA SD010460231GA SD010460232GA SD010460233GA SD010460234GA SD010460236GA SD010460238GA SD010460239GA SD010460240GA SD010460241GA SD010460243GA SD010460244GA SD010460250GA SD010460252GA SD010460253GA SD010460254GA SD010460255GA SD010460256GA SD010460257GA SD010460258GA SD010460259GA SD010460260GA SD010460261GA SD010460262GA SD010460263GA SD010460264GA SD010460265GA SD010460266GA SD010460269GA SD010460271GA SD010460272GA SD010460273GA SD010460274GA SD010460275GA SD010460276GA SD010460277GA SD010460279GA SD010460280GA SD010460282GA SD010460283GA SD010460285GA SD010460286GA SD010460287GA SD010460288GA SD010460289GA SD010460290GAMKE SD010460291GA SD010460292GA SD010460295GA SD010460300GA SD010460301GA SD010460306GA SD010460307GA SD010460308GA SD010460309GA SD010460310GA SD010460311GA SD010460312GA SD010460314GA SD010460315GA SD010460316GA SD010460317GA SD010460318GA SD010460319GA SD010460320GA SD010460321GA SD010460323GA SD010470324GA SD010470325GA SD010470326GA SD010470327GA SD010470329GA SD010470331GA SD010470332GA SD010470334GA SD010470337GA SD010470338GA SD010470340GA SD010470341GA SD010470342GA SD010470343GA SD010470345GA SD010470346GAMKE SD010470348GA SD010470350GA SD010470351GA SD010470352GA SD010470353GA SD010470354GA SD010470355GA SD010470357GA SD010470358GA SD010470360GA SD010470361GA SD010470362GA SD010470363GA SD010470364GA SD010470365GA SD010470366GA SD010470367GA SD010470369GA SD010470370GA SD010470371GA SD010470372GA SD010470373GA SD010470374GA SD010470375GA SD010470376GA SD010470378GA SD010470379GA SD010470380GA SD010470381GA SD010470382GA SD010470383GA SD010470384GA SD010470385GA SD010470386GA SD010470388GA SD010470389GA SD010470390GA SD010470391GA SD010470392GA SD010470393GA SD010470394GA SD010470395GA SD010470396GA SD010470397GA SD010470400GA SD010470407GA SD010470408GA SD010470409GA SD010470411GA SD010470412GA SD010470413GA SD010470414GA SD010470415GA SD010470417GA SD010470418GA SD010470419GA SD010470420GA SD010470421GA SD010470422GA SD010470423GA SD010470424GA SD010470425GA SD010470426GA SD010470427GA SD010470428GA SD010470429GA SD010470430GA SD010470431GA SD010470433GA SD010470434GA SD010470435GA SD010470436GA SD010470437GA SD010470440GA SD010470441GR SD010470442GA SD010470443GA SD010470445GA SD010470446GA SD010470447GA SD010470448GA SD010470449GA SD010470450GA SD010470451GA SD010470452GA SD010470453GA SD010470454GA SD010470455GA SD010470456GA SD010470457GA SD010470458GA SD010470459GA SD010470460GA SD010470461GA SD010470462GA SD010470463GA SD010470464GA SD010470465GA SD010470466GA SD010470467GA SD010470468GA SD010470470GA SD010470472GA SD010470474GA SD010470475GA SD010470476GA SD010470477GA SD010470478GA SD010470479GA SD010470480GA SD010470481GA SD010470482GA SD010470484GA SD010470486GA SD010470487GA SD010470488GA SD010470489GA SD010470490GA SD010470492GA SD010470493GA SD010470494GA SD010470495GA SD010470496GA SD010470497GA SD010470498GA SD010470501GA SD010470503GA SD010470505GA SD010470507GA SD010470508GA SD010470509GA SD010470510GA SD010470511GA SD010470513GA SD010470514GA SD010470515GA SD010470516GA SD010470517GA SD010470520GA SD010470521GA SD010470524GA SD010470525GA SD010470527GA SD010470529GA SD010470530GA SD010470531GR SD010470533GA SD010470537GA SD010470540GA SD010470541GA SD010470546GA SD010470547GA SD010470548GA SD010470549GA SD010470550GA SD010470553GA SD010470555GA SD010470556GA SD010470557GA SD010470558GA SD010470559GA SD010470560GA SD010470561GA SD010470562GA SD010470563GA SD010470564GA SD010470565GA SD010470566GA SD010470567GA SD010470569GA SD010470570GA SD010470573GA SD010470575GA SD010470577GA SD010470578GA SD010470580GA SD010470582GA SD010470583GA SD010470585GA SD010470587GA SD010470589GA SD010470590GA SD010470591GA SD010470592GA SD010470593GA SD010470595GA SD010470596GA SD010470597GA SD010470606GA SD010470607GA SD010470608GA SD010470609GR SD010470610GA SD010470611GA SD010470612GA SD010470614GA SD010470615GA SD010470616GA SD010470617GA SD010470618GA SD010470619GA SD010470620GA SD010470622GA SD010470625GA SD010470626GA SD010470627GA SD010470628GA SD010470629GA SD010470630GA SD010470631GA SD010470633GA SD010470635GA SD010470637GA SD010470639GA SD010470640GA SD010470641GA SD010470642GA SD010470643GA SD010470644GA SD010470645GA SD010470646GA SD010470647GA SD010470649GA SD010470650GA SD010470651GA SD010470652GA SD010470653GA SD010470655GA SD010470656GA SD010470657GA SD010470659GA SD010470660GA SD010470661GA SD010470662GA SD010470663GA SD010470664GA SD010470665GA SD010470666GA SD010470667GA SD010470668GA SD010470669GA SD010470670GA SD010470671GA SD010470674GA SD010470675GA SD010470677GA SD010470678GA SD010470704GA SD010470727GA SD010470728GA SD010470729GA SD010470730GA SD010470731GA SD010470733GA SD010470735GA SD010470736GA SD010470737GA SD010470738GA SD010470739GA SD010470740GA SD010470742GA SD010470743GA SD010470745GA SD010470746GA SD010470747GA SD010470749GA SD010470750GA SD010470752GA SD010470756GA SD010470757GA SD010470759GA SD010470760GA SD010470761GA SD010470763GA SD010470764GA SD010470766GA SD010470767GA SD010470768GA SD010470769GA SD010470770GA SD010470772GA SD010470773GA SD010470774GA SD010470775GA SD010470776GA SD010470778GA SD010470779GA SD010470780GA SD010470781GA SD010470782GA SD010470783GA SD010470784GA SD010470785GA SD010470786GA SD010470801GA SD010480106SA SD010480117SA SD010480124SA SD010480804GA SD010480805GA SD010480806GA SD010480818GA SD010480819GA SD010480820GA SD010480823GA SD010480824GA SD010480825GA SD010480826GA SD010480827GA SD010480828GA SD010480829GA SD010480830GA SD010480831GA SD010480832GA SD010480833GA SD010480834GA SD010480835GA SD010480836GA SD010480837GA SD010480857GA SD010480858GA SD010480859GA SD010480860GA SD010480862GA SD010480863GA SD010480864GA SD010480865GA SD010480866GA SD010480867GA SD010480868GA SD010480869GA SD010480871GA SD010480872GA SD010480873GA SD010480874GA SD010480875GA SD010480882GA SD010480885GA SD010480888GA SD010480892GA SD010480895GA SD010480896GA SD010480897GA SD010480900GA SD010480903GA SD010480905GA SD010480911GA SD010480912GA SD010480913GA SD010480915GA SD010480916GA SD010480917GA SD010480918GA SD010480919GA SD010480920GA SD010480921GA SD010480922GA SD010480923GA SD010480924GA SD010480942GA SD010480943GA SD010480945GA SD010480948GA SD010480949GA SD010480950GA SD010480951GA SD010480952GA SD010480954GA SD010480955GA SD010480956GA SD010480957GA SD010480958GA SD010480959GA SD010480960GA SD010480961GA SD010480962GA SD010480963GA SD010480964GA SD010480965GA SD010480966GA SD010480967GA SD010480968GA SD010480969GA SD010480970GA SD010480971GA SD010480972GA SD010480973GA SD010480975GA SD010480977GA SD010480978GA SD010480979GA SD010480981GA SD010480984GA SD010480985GA SD010480986GA SD010480987GA SD010480988GA SD010480990GA SD010480991GA SD010480992GA SD010480994GA SD010480995GA SD010480996GA SD010480998GA SD010480999GA SD010481000GA SD010481001GA SD010481002GA SD010481003GA SD010481004GA SD010481005GR SD010481007GA SD010481009GA SD010481010GA SD010481011GA SD010481012GA SD010481014GA SD010481015GA SD010481016GR SD010481017GA SD010481019GA SD010481020GA SD010481023GA SD010481025GA SD010481026GA SD010481027GA SD010481028GA SD010481029GA SD010481031GA SD010481033GA SD010481034GA SD010481035GA SD010481037GA SD010481042GA SD010481044GA SD010481045GA SD010481046GA SD010481047GA SD010481048GA SD010481049GA SD010481050GA SD010481051GA SD010481053GA SD010481054GA SD010481055GA SD010481056GA SD010481057GA SD010481058GA SD010481059GA SD010481060GA SD010481061GA SD010481062GA SD010481063GA SD010481064GA SD010481065GA SD010481066GA SD010481068GA SD010481072GA SD010481073GA SD010481078GA SD010481083GA SD010481088GA SD010481090GA SD010481091GA SD010481092GA SD010481093GA SD010481094GA SD010481095GA SD010481097GA SD010481099GA SD010481100GA SD010481102GA SD010481103GA SD010481106GA SD010481107GA SD010481108GA SD010481109GA SD010481111GA SD010481112GA SD010481113GA SD010481115GA SD010481117GA SD010481118GA SD010481119GA SD010481120GA SD010481121GA SD010481123GA SD010481124GA SD010481125GA SD010481126GA SD010481128GA SD010481129GA SD010481133GA SD010481134GA SD010481135GA SD010481137GA SD010481220GA SD010481221GA SD010481222GA SD010481223GA SD010481225GA SD010481228GA SD010481232GA SD010481234GA SD010481235GA SD010481239GA SD010481241GA SD010481242GA SD010481243GA SD010481249GA SD010481250GA SD010481251GA SD010481252GA SD010481254GA SD010481255GA SD010481257GA SD010481258GA SD010481259GA SD010481260GA SD010481261GA SD010481262GA SD010481263GA SD010481264GA SD010481265GA SD010481266GA SD010481267GA SD010481268GA SD010481271GA SD010481273GA SD010481274GA SD010481277GA SD010481278GA SD010481280GA SD010481281GA SD010481282GA SD010481283GA SD010481284GA SD010481285GA SD010481286GA SD010481287GA SD010481288GA SD010481291GA SD010481292GA SD010481293GA SD010481294GA SD010481295GA SD010481296GA SD010481297GA SD010481299GA SD010481300GA SD010481301GA SD010481302GA SD010481303GA SD010481304GA SD010481305GA SD010481306GA SD010481307GA SD010481308GA SD010491309GA SD010491312GA SD010491314GA SD010491317GA SD010491318GA SD010491319GA SD010491321GA SD010491325GA SD010491327GA SD010491328GA SD010491330GA SD010491332GA SD010491333GA SD010491334GA SD010491337GA SD010491343GA SD010491344GA SD010491345GA SD010491346GA SD010491347GA SD010491348GA SD010491352GA SD010491356GA SD010491358GA SD010491364GA SD010491380GA SD010491381GA SD010491382GA SD010491383GA SD010491384GA SD010491388GA SD010491389GA SD010491391GA SD010491392GA SD010491393GA SD010491394GA SD010491395GA SD010491398GA SD010491399GA SD010491403GA SD010491406GA SD010491407GA SD010491408GA SD010491410GA SD010491411GA SD010491412GA SD010491413GA SD010491414GA SD010491415GA SD010491416GA SD010491417GA SD010491418GA SD010491419GA SD010491420GA SD010491421GA SD010491446GA SD010491450GA SD010491454GA SD010491455GA SD010491457GA SD010491461GA SD010491462GA SD010491463GA SD010491464GA SD010491465GA SD010491466GA SD010491469GA SD010491470GA SD010491471GA SD010491472GA SD010491473GA SD010491474GA SD010491475GA SD010491476GA SD010491477GA SD010491478GA SD010491479GA SD010491480GA SD010491481GACOMPLETE SD010491482GA SD010491483GA SD010491484GACOMPLETE SD010491485GA SD010491486GA SD010491487GA SD010491488GA SD010491489GA SD010491490GA SD010491491GA SD010491492GA SD010491493GA SD010491494GA SD010491495GA SD010491496GA SD010491498GA SD010491499GA SD010491502GA SD010491505GA SD010491509GA SD010491510GA SD010491513GA SD010491514GA SD010491515GA SD010491516GA SD010491517GA SD010491518GA SD010491519GA SD010491520GA SD010491522GA SD010491524GA SD010491525GA SD010491527GA SD010491528GA SD010491530GA SD010491531GA SD010491534GA SD010491536GA SD010491539GA SD010491547GA SD010491548GA SD010491549GA SD010491552GA SD010491555GA SD010491562GA SD010491565GA SD010491571GA SD010491572GA SD010491573GA SD010491590GA SD010491592GA SD010491598GA SD010491606GA SD010491611GA SD010491615GA SD010491619GR SD010491620GA SD010491625GA SD010491627GA SD010491628GA SD010491630GA SD010491632GA SD010491633GA SD010491636GA SD010491637GA SD010491638GA SD010491639GA SD010491640GA SD010491641GA SD010491642GA SD010491643GA SD010491644GA SD010491645GA SD010491646GA SD010491649GA SD010491650GA SD010491651GA SD010491652GA SD010491653GA SD010491654GA SD010491666GA SD010491667GA SD010491670GA SD010491671GA SD010491687GA SD010491688GA SD010491689GA SD010491695GA SD010491698GA SD010491700GA SD010491701GA SD010491702GA SD010491705GA SD010491706GR SD010500133SA SD010500135SR SD010500137SR SD010500152SA SD010500171SR SD010500173SR SD010500177SR SD010500192SA SD010500193SA SD010500193SR SD010500194SR SD010500201SR SD010500202SA SD010500203SA SD010500204SA SD010500205SA SD010500206SA SD010500207SA SD010500208SA SD010500209SA SD010500210SA SD010500211SA SD010500212SA SD010500220SA SD010500238SA SD010500244SA SD010500246SA SD010500249SA SD010500257SA SD010500267SA SD010500268SA SD010500269SA SD010500271SA SD010500272SA SD010500273SA SD010500274SA SD010500275SA SD010500276SA SD010500277SA SD010500278SA SD010500279SA SD010500280SA SD010500281SA SD010500282SA SD010500283SA SD010500284SA SD010500285SA SD010500286SA SD010500287SA SD010501726GA SD010501727GA SD010501729GA SD010501730GA SD010501731GA SD010501733GA SD010501734GA SD010501735GA SD010501737GA SD010501739GA SD010501770GA SD010501776GA SD010501778GA SD010501781GA SD010501785GA SD010501787GA SD010501788GA SD010501790GA SD010501794GA SD010501795GA SD010501796GA SD010501797GA SD010501798GA SD010501799GA SD010501800GA SD010501801GA SD010501802GA SD010501803GA SD010501804GA SD010501805GA SD010501807GA SD010501808GA SD010501809GA SD010501810GA SD010501812GA SD010501813GA SD010501815GA SD010501816GA SD010501817GA SD010501818GA SD010501819GA SD010501820GA SD010501821GA SD010501822GA SD010501823GA SD010501825GA SD010501827GA SD010501829GA SD010501830GA SD010501831GA SD010501832GA SD010501833GA SD010501834GA SD010501835GA SD010501836GA SD010501837GA SD010501838GA SD010501839GA SD010501840GA SD010501841GA SD010501842GA SD010501843GA SD010501844GA SD010501845GA SD010501846GA SD010501847GA SD010501848GA SD010501849GA SD010501850GA SD010501854GR SD010501856GA SD010501857GA SD010501859GA SD010501860GA SD010501861GA SD010501864GA SD010501865GA SD010501866GR SD010501867GA SD010501868GA SD010501869GA SD010501870GA SD010501871GA SD010501885GA SD010501890GR SD010501896GA SD010501897GA SD010501898GA SD010501899GA SD010501900GA SD010501901GA SD010501902GA SD010501903GA SD010501904GA SD010501905GA SD010501906GA SD010501907GA SD010501908GA SD010501909GA SD010501910GA SD010501911GA SD010501912GA SD010501913GA SD010501914GA SD010501915GA SD010501916GA SD010501917GA SD010501918GR SD010501921GA SD010501925GA SD010501929GA SD010501932GA SD010501934GA SD010501935GA SD010501939GA SD010501940GA SD010501944GA SD010501945GA SD010501946GA SD010501947GA SD010501948GA SD010501949GA SD010501950GA SD010501951GA SD010501952GA SD010501953GA SD010501954GA SD010501955GA SD010501957GA SD010501958GA SD010501959GA SD010501960GA SD010501961GA SD010501962GA SD010501963GA SD010501964GA SD010501965GA SD010501966GA SD010501967GA SD010501968GA SD010501970GA SD010501972GA SD010501973GA SD010501978GA SD010501980GA SD010501982GA SD010501986GA SD010501988GA SD010501990GA SD010501996GA SD010502005GA SD010502007GA SD010502009GA SD010502020GA SD010502021GA SD010502022GA SD010502023GA SD010502025GA SD010502026GA SD010502032GA SD010502035GA SD010502038GA SD010502039GA SD010502040GA SD010502041GA SD010502042GA SD010502043GA SD010502044GA SD010502045GA SD010502046GA SD010502047GA SD010502048GA SD010502049GA SD010502050GA SD010502051GA SD010502052GA SD010502053GA SD010502054GA SD010502055GA SD010502057GA SD010502058GA SD010502059GA SD010502060GA SD010502061GA SD010502062GA SD010502063GA SD010502064GA SD010502065GA SD010502066GA SD010502067GA SD010502068GA SD010502069GA SD010502070GA SD010502071GA SD010502072GA SD010502073GA SD010502074GA SD010502075GA SD010502077GA SD010502078GA SD010502079GA SD010502080GR SD010502081GA SD010502082GA SD010502083GA SD010502084GA SD010502085GA SD010502086GA SD010502087GA SD010502088GA SD010502090GA SD010502091GA SD010502092GA SD010502093GA SD010502094GA SD010502095GA SD010502097GA SD010502098GA SD010502099GA SD010502115GA SD010502116GA SD010502121GA SD010502126GA SD010502130GA SD010502132GA SD010502135GA SD010502138GA SD010502140GA SD010502143GA SD010502144GA SD010502145GA SD010502152GA SD010502153GA SD010502154GA SD010502155GA SD010502156GA SD010502157GA SD010502158GA SD010502159GA SD010502164GA SD010502165GA SD010502170GA SD010502171GA SD010502174GA SD010502175GA SD010502176GA SD010502181GA SD010502186GA SD010502187GA SD010502189GA SD010502190GA SD010502192GA SD010502193GA SD010502195GA SD010502196GA SD010502197GA SD010502203GA SD010502204GA SD010502205GA SD010502206GA SD010502207GA SD010502208GA SD010502211GA SD010502213GA SD010502214GA SD010502215GA SD010502216GA SD010502217GA SD010502218GA SD010502219GA SD010502220GA SD010502221GA SD010502222GA SD010502223GA SD010502224GA SD010502225GA SD010502228GA SD010502229GA SD010502230GR SD010502231GA SD010502232GR SD010502233GA SD010502235GA SD010502236GA SD010502237GA SD010502239GR SD010502240GA SD010502242GA SD010502243GA SD010502246GA SD010502247GA SD010502252GA SD010502256GA SD010502257GA SD010502258GA SD010502259GA SD010502261GA SD010502262GA SD010502263GA SD010502264GA SD010502265GA SD010502266GA SD010502267GA SD010502268GA SD010502269GA SD010502272GA SD010502273GA SD010502276GA SD010502279GA SD010502280GA SD010502281GA SD010502282GA SD010502283GA SD010502284GA SD010502286GA SD010502287GA SD010502288GA SD010502289GA SD010502290GA SD010502291GA SD010502292GA SD010502293GA SD010502294GA SD010502296GA SD010502297GA SD010502298GA SD010502299GA SD010502300GA SD010502321GA SD010510294SA SD010510295SA SD010510296SA SD010510297SA SD010510308SA SD010510314SA SD010510314SR SD010510316SA SD010510317SA SD010510318SA SD010510319SA SD010510320SA SD010510321SA SD010510322SA SD010510323SA SD010510324SA SD010510325SA SD010510326SA SD010510327SA SD010510328SA SD010510329SA SD010510330SA SD010510351SA SD010510352SA SD010510353SA SD010510354SA SD010510355SA SD010510358SA SD010510360SA SD010510361SA SD010510362SA SD010510363SA SD010510364SA SD010510366SA SD010512325GA SD010512327GA SD010512328GA SD010512329GA SD010512332GA SD010512335GA SD010512336GA SD010512338GA SD010512343GA SD010512361GA SD010512366GA SD010512370GA SD010512372GA SD010512374GA SD010512385GA SD010512401GA SD010512404GA SD010512407GA SD010512412GA SD010512421GA SD010520387SA SD010520388SA SD010520389SA SD010520390SA SD010520391SA SD010520392SA SD010520393SA SD010520394SA SD010520395SA SD010520396SA SD010520397SA SD010520398SA SD010520399SA SD010520400SA SD010520401SA SD010520402SA SD010520403SA SD010520406SA SD010520406SR SD010520409SA SD010520413SA SD010520414SA SD010520415SA SD010520416SA SD010520417SA SD010520418SA SD010520419SA SD010520420SA SD010520421SA SD010520422SA SD010520423SA SD010520424SA SD010520425SA SD010520426SA SD010520427SA SD010520428SA SD010520429SA SD010520431SA SD010520432SA SD010520433SA SD010520434SA SD010520438SA SD010520439SA SD010520440SA SD010520441SA SD010520442SA SD010520443SA SD010520444SA SD010520445SA SD010520446SA SD010520447SA SD010520448SA SD010520449SA SD010520450SA SD010520451SA SD010520452SA SD010520453SA SD010520454SA SD010520455SA SD010520456SA SD010520457SA SD010520458SA SD010520459SA SD010520460SA SD010520461SA SD010520462SA SD010520463SA SD010520464SA SD010520465SA SD010522422GA SD010522424GA SD010522425GA SD010522426GA SD010522427GA SD010522428GA SD010522429GA SD010522430GA SD010522431GA SD010522432GA SD010522433GA SD010522434GA SD010522435GA SD010522436GA SD010522437GA SD010522438GA SD010522439GA SD010522440GA SD010522442GA SD010522444GA SD010522445GA SD010522446GA SD010522451GA SD010522452GA SD010522454GA SD010522455GA SD010522456GA SD010522457GA SD010522458GA SD010522460GA SD010522463GA SD010522464GA SD010522465GA SD010522466GA SD010522469GA SD010522470GA SD010522472GA SD010522475GA SD010522477GA SD010522478GA SD010522479GA SD010522481GA SD010522482GA SD010522483GA SD010522484GA SD010522494GA SD010522495GA SD010522497GA SD010522498GA SD010522500GA SD010522501GA SD010522502GA SD010522504GA SD010522508GA SD010522510GA SD010522512GA SD010522514GA SD010522516GA SD010522517GA SD010522527GA SD010522532GA SD010522534GA SD010522543GA SD010522544GA SD010522546GA SD010522549GA SD010522550GA SD010522551GA SD010522554GA SD010522555GA SD010522556GA SD010522560GA SD010522562GA SD010522566GA SD010522567GA SD010522568GA SD010522569GA SD010522571GA SD010522572GA SD010522573GA SD010522575GA SD010522576GA SD010522577GA SD010522578GA SD010522579GA SD010522580GA SD010522581GA SD010522582GA SD010522583GA SD010522585GA SD010522586GA SD010522587GA SD010522588GA SD010522589GA SD010522590GA SD010522591GA SD010522592GA SD010522593GA SD010522594GA SD010522595GA SD010522596GA SD010522597GA SD010522598GA SD010522599GA SD010522600GA SD010522602GA SD010522603GA SD010522604GA SD010522605GA SD010522606GA SD010522607GA SD010522609GA SD010522610GA SD010522614GA SD010522617GA SD010522620GA SD010522621GA SD010522622GA SD010522625GA SD010522626GA SD010522627GA SD010522628GA SD010522629GA SD010522630GA SD010522631GA SD010522633GA SD010522635GA SD010522636GR SD010522637GA SD010522638GA SD010522639GA SD010522640GA SD010522641GA SD010522645GA SD010522650GA SD010522652GA SD010522653GA SD010522655GA SD010522657GA SD010522658GA SD010532752GA SD010532753GA SD010532754GA SD010532755GA SD010532757GA SD010532758GA SD010532761GA SD010532763GA SD010532764GA SD010532765GA SD010532766GA SD010532771GA SD010532776GA SD010532779GA SD010532780GA SD010532781GA SD010532782GA SD010532783GA SD010532785GA SD010532786GA SD010532790GA SD010532791GA SD010532804GA SD010532819GA SD010532821GA SD010532837GA SD010532838GA SD010532842GA SD010532847GA SD011020467SA SD011020468SA SD011020469SA SD011020470SA SD011020471SA SD011020472SA SD011020473SA SD011020474SA SD011020475SA SD011020476SA SD011020477SA SD011020478SA SD011020479SA SD011020480SA SD011020481SA SD011020482SA SD011020483SA SD011020484SA SD011020485SA SD011020486SA SD011020487SA SD011020488SA SD011020489SA SD011020490SA SD011020491SA SD011020495SA SD011020496SA SD011020497SA SD011020498SA SD011020499SA SD011020500SA SD011020501SA SD011020502SA SD011020504SA SD011020505SA SD011020506SA SD011020507SA SD011020508SA SD011020509SA SD011020510SA SD011020511SA SD011020512SA SD011020513SA SD011020514SA SD011020515SA SD011020518SA SD011020520SA SD011020521SA SD011020522SA SD011020523SA SD011020524SA SD011020525SA SD011020526SA SD011020528SA SD011020529SA SD011020530SA SD011030553SA SD011030554SA SD011030555SA SD011030556SA SD011030559SA SD011030560SA SD011030562SA SD011030564SA SD011030565SA SD011030566SA SD011030568SA SD011030569SA SD011030570SA SD011030574SA SD011030576SA SD011030578SA SD011030579SA SD011030580SA SD011030581SA SD011030582SA SD011030584SA SD011030585SA SD011030586SA SD011030588SA SD011030589SA SD011030590SA SD011030591SA SD011030592SA SD011030593SA SD011030594SA SD011030595SA SD011030596SA SD011030597SA SD011030598SA SD011030599SA SD011030600SA SD011030601SA SD011030602SA SD011030605SA SD011030607SA SD011030609SA SD011030610SA SD011030611SA SD011030612SA SD011030613SA SD011030614SA SD011030615SA SD011030616SA SD011030617SA SD011030619SA SD011030620SA SD011030621SA SD011030622SA SD011030623SA SD011030625SA SD011030626SA SD011030627SA SD011030630SA SD011030631SA SD011030632SA SD011030633SA SD011030633SR SD011030635SA SD011030637SA SD011030640SA SD011030641SA SD011030646SA SD011030647SA SD011030649SA SD011030650SA SD011030651SA SD011030652SA SD011030653SA SD011030654SA SD011030655SA SD011030656SA SD011030657SA SD011030658SA SD011030659SA SD011030660SA SD011030661SA SD011030662SA SD011030663SA SD011030664SA SD011030665SA SD011030666SA SD011030667SA SD011030668SA SD011030669SA SD011030670SA SD011030671SA SD011030672SA SD011030683SA SD011030684SA SD011030685SA SD011030686SA SD011030687SA SD011030688SA SD011030689SA SD011030690SA SD011030691SA SD011030692SA SD011030693SA SD011030694SA SD011030695SA SD011030696SA SD011030697SA SD011032920GA SD011032925GA SD011032926GA SD011032929GA SD011032931GA SD011032932GA SD011032935GA SD011032937GA SD011032940GA SD011032942GA SD011032943GA SD011032944GA SD011032945GA SD011032946GA SD011032947GA SD011032949GA SD011032950GA SD011032951GA SD011032952GA SD011032953GA SD011032954GA SD011032955GA SD011032956GA SD011032957GA SD011032958GA SD011032960GA SD011032961GA SD011032962GA SD011032963GA SD011032964GA SD011032965GA SD011032966GA SD011032967GA SD011032994GA SD011033013GA SD011033017GA SD011033018GA SD011033020GA SD011033022GA SD011033027GA SD011033029GA SD011033030GA SD011033035GA SD011033052GA SD011033060GA SD011033071GA SD011033073GA SD011033074GA SD011033078GA SD011033081GA SD011033091GA SD011033092GA SD011033110GA SD011033111GA SD011033114GA SD011033143GA SD011033156GA SD011033159GA SD011040703SA SD011040704SA SD011040705SA SD011040707SA SD011040708SA SD011040711SA SD011040712SA SD011040716SA SD011040717SA SD011040719SA SD011040720SA SD011040721SA SD011040722SA SD011040723SA SD011040724SA SD011040725SA SD011040726SA SD011040727SA SD011040729SA SD011040730SA SD011040731SA SD011040733SA SD011040734SA SD011040735SA SD011040736SA SD011040737SA SD011040738SA SD011040739SA SD011040740SA SD011040741SA SD011040741SR SD011040742SA SD011040743SA SD011040744SA SD011040744SR SD011040745SA SD011040746SA SD011040747SA SD011040748SA SD011040749SA SD011040749SR SD011040750SA SD011040751SA SD011040751SR SD011040752SA SD011040753SA SD011040753SR SD011040754SA SD011040756SA SD011040757SA SD011040758SA SD011040758SR SD011040759SA SD011040760SA SD011040761SA SD011040762SA SD011040763SA SD011040764SA SD011040765SA SD011040766SA SD011040767SA SD011040768SA SD011040769SA SD011040770SA SD011040771SA SD011040772SA SD011040773SA SD011040774SA SD011040775SA SD011040776SA SD011040777SA SD011040778SA SD011040779SA SD011040780SA SD011040781SA SD011040783SA SD011040784SA SD011040785SA SD011040786SA SD011040787SA SD011040788SA SD011040789SA SD011040790SA SD011040791SA SD011040792SA SD011040793SA SD011040794SA SD011040796SA SD011040797SA SD011040798SA SD011040799SA SD011040800SA SD011040801SA SD011040802SA SD011040803SA SD011040804SA SD011040805SA SD011040806SA SD011040807SA SD011040808SA SD011040809SA SD011040810SA SD011040811SA SD011040813SA SD011040814SA SD011040818SA SD011040819SA SD011040820SA SD011040821SA SD011040822SA SD011040823SA SD011040824SA SD011040825SA SD011040826SA SD011040827SA SD011040828SA SD011040829SA SD011040830SA SD011040831SA SD011040832SA SD011040833SA SD011040834SA SD011040835SA SD011040836SA SD011040837SA SD011040838SA SD011040839SA SD011040840SA SD011040841SA SD011040842SA SD011040843SA SD011040844SA SD011040845SA SD011040846SA SD011040847SA SD011040848SA SD011040849SA SD011040850SA SD011040851SA SD011040852SA SD011040853SA SD011040854SA SD011040855SA SD011040856SA SD011040857SA SD011040858SA SD011040859SA SD011040860SA SD011040861SA SD011040862SA SD011040863SA SD011040864SA SD011040865SA SD011040866SA SD011040867SA SD011040868SA SD011040869SA SD011040870SA SD011040871SA SD011040872SA SD011040874SA SD011040875SA SD011040876SA SD011040877SA SD011040878SA SD011040879SA SD011040880SA SD011040883SA SD011040884SA SD011040885SA SD011040886SA SD011040887SA SD011040888SA SD011040889SA SD011040890SA SD011040891SA SD011040892SA SD011040893SA SD011040894SA SD011040896SA SD011040897SA SD011040899SA SD011040901SA SD011040903SA SD011040903SR SD011040904SA SD011040905SA SD011043163GA SD011043177GA SD011050906SA SD011050907SA SD011050908SA SD011050909SA SD011050910SA SD011050911SA SD011050912SA SD011050913SA SD011050914SA SD011050915SA SD011050916SA SD011050917SA SD011050920SA SD011050921SA SD011050922SA SD011050923SA SD011050926SA SD011050927SA SD011050928SA SD011050929SA SD011050931SA SD011050931SR SD011050932SA SD011050933SA SD011050935SA SD011050936SA SD011050936SR SD011050937SA SD011050939SA SD011050941SA SD011050942SA SD011050943SA SD011050945SA SD011050946SA SD011050947SA SD011050950SA SD011050951SA SD011050952SA SD011050954SA SD011050955SA SD011050956SA SD011050957SA SD011050958SA SD011050959SA SD011050959SR SD011050960SA SD011050961SA SD011050962SA SD011050964SA SD011050965SA SD011050966GR SD011050966SA SD011050967SA SD011050968SA SD011050970SA SD011050971SA SD011050972SA SD011050973SA SD011050974SA SD011050975SA SD011050977SA SD011050978SA SD011050979SA SD011050980SA SD011050981SA SD011050982SA SD011050983SA SD011050984SA SD011050985SA SD011050986SA SD011050987SA SD011050988SA SD011050989SA SD011050990SA SD011050991SA SD011050992SA SD011050993SA SD011050994SA SD011050995SA SD011050996SA SD011050999SA SD011051000SA SD011051001SA SD011051002SA SD011051003SA SD011051004SA SD011051005SA SD011051006SA SD011051007SA SD011051008SA SD011051009SA SD011051010SA SD011051011SA SD011051012SA SD011051013SA SD011051014SA SD011051016SA SD011051017SA SD011051018SA SD011051019SA SD011051020SA SD011051024SA SD011051025SA SD011051026SA SD011051027SA SD011051028SA SD011051029SA SD011051032SA SD011051033SA SD011051034SA SD011051035SA SD011051037SA SD011051038SA SD011051039SA SD011051041SA SD011051042SA SD011051043SA SD011051044SA SD011051045SA SD011051046SA SD011051047SA SD011051048SA SD011051050SA SD011051052SA SD011051053SA SD011051056SA SD011051057SA SD011051058SA SD011051061SA SD011051063SA SD011051065SA SD011051066SA SD011051067SA SD011051068SA SD011051069SA SD011051070SA SD011051071SA SD011051074SA SD011051077SA SD011051080SA SD011051081SA SD011051083SA SD011051084SA SD011051087SA SD011051088SA SD011051089SA SD011051091SA SD011051092SA SD011051093SA SD011051095SA SD011051096SA SD011051097SA SD011051098SA SD011051099SA SD011051100SA SD011051101SA SD011051102SA SD011051103SA SD011051104SA SD011051105SA SD011051106SA SD011051107SA SD011051108SA SD011051109SA SD011051110SA SD011051111SA SD011051112SA SD011051113SA SD011051114SA SD011051115SA SD011051116SA SD011051117SA SD011051118SA SD011051119SA SD011051124SA SD011051124SR SD011051126SA SD011051126SR SD011051128SA SD011051128SR SD011051129SA SD011051130SA SD011051131SA SD011051132SA SD011051133SA SD011051134SA SD011051135SA SD011051136SA SD011051137SA SD011051138SA SD011051139SA SD011051140SA SD011051141SA SD011051142SA SD011051143SA SD011051144SA SD011051145SA SD011051146SA SD011051147SA SD011051148SA SD011051149SA SD011051150SA SD011051151SA SD011051152SA SD011051153SA SD011051154SA SD011051155SA SD011051156SA SD011051157SA SD011051158SA SD011051159SA SD011051160SA SD011051161SA SD011051162SA SD011051164SA SD011051166SA SD011051167SA SD011051168SA SD011051169SA SD011051170SA SD011051171SA SD011051172SA SD011051173SA SD011051174SA SD011051175SA SD011051176SA SD011051177SA SD011051178SA SD011051180SA SD011051181SA SD011051184SA SD011051185SA SD011051186SA SD011051188SA SD011051189SA SD011051192SA SD011051194SA SD011051194SR SD011051195SA SD011051211SA SD011053263GA SD011053264GA SD011053265GA SD011053268GA SD011053269GA SD011053270GA SD011053271GA SD011061216SA SD011061218SA SD011061220SA SD011061223SA SD011061224SA SD011061225SA SD011061227SA SD011061228SA SD011061229SA SD011061230SA SD011061231SA SD011061232SA SD011061233SA SD011061234SA SD011061236SA SD011061237SA SD011061238SA SD011061239SA SD011061240SA SD011061241SA SD011061242SA SD011061243SA SD011061244SA SD011061245SA SD011061246SA SD011061247SA SD011061248SA SD011061249SA SD011061250SA SD011061251SA SD011061252SA SD011061253SA SD011061254SA SD011061255SA SD011061256SA SD011061257SA SD011061258SA SD011061261SA SD011061262SA SD011061263SA SD011061264SA SD011061265SA SD011061266SA SD011061268SA SD011061269SA SD011061270SA SD011061271SA SD011061272SA SD011061273SA SD011061274SA SD011061275SA SD011061276SA SD011061279SA SD011061281SA SD011061282SA SD011061287SA SD011061288SA SD011061289SA SD011061290SA SD011061292SA SD011061295SA SD011061296SA SD011061301SA SD011061307SA SD011061312SA SD011061313SA SD011061313SR SD011061340SA SD011061342SA SD011061346SA SD011061349SA SD011061350SA SD011061355SA SD011061356SA SD011061358SA SD011061360SA SD011061362SA SD011061363SA SD011061364SA SD011061365SA SD011061367SA SD011061368SA SD011061371SA SD011061373SA SD011061374SA SD011061378SA SD011061379SA SD011061381SA SD011061383SA SD011061384SA SD011061402SA SD011061407SA SD011061408SA SD011061414SA SD011061424SA SD011061429SA SD011061430SA SD011061431SA SD011061434SA SD011061446SA SD011061451SA SD011061457SA SD011061461SA SD011061465SA SD011061467SA SD011061469SA SD011061472SA SD011061474SA SD011061477SA SD011061478SA SD011061482SA SD011061484SA SD011061485SA SD011061486SA SD011061487SA SD011061489SA SD011061495SA SD011061496SA SD011061497SA SD011061498SA SD011061499SA SD011061500SA SD011061501SA SD011061502SA SD011061503SA SD011061504SA SD011061505SA SD011061507SA SD011061509SA SD011061510SA SD011061512SA SD011061513SA SD011061514SA SD011061515SA SD011061516SA SD011061517SA SD011061518SA SD011061520SA SD011061521SA SD011063241GA SD011063242GA SD011063243GA SD011063245GA SD011063248GA SD011063251GA SD011063257GA SD011063307GA SD011071526SA SD011071531SA SD011071539SA SD011071541SA SD011071544SA SD011071545SA SD011071547SA SD011071548SA SD011071550SA SD011071551SA SD011071555SA SD011071557SA SD011071562SA SD011071564SA SD011071565SA SD011071577SA SD011071580SA SD011071581SA SD011071588SA SD011071589SA SD011071593SA SD011071594SA SD011071595SA SD011071608SA SD011071612SA SD011071613SA SD011071622SA SD011071625SA SD011071626SA SD011071633SA SD011071635SA SD011071638SA SD011071640SA SD011071641SA SD011071643SA SD011071645SA SD011071647SA SD011071648SA SD011071659SA SD011071660SA SD011071661SA SD011071662SA SD011071663SA SD011071664SA SD011071665SA SD011071666SA SD011071667SA SD011071668SA SD011071669SA SD011071671SA SD011071672SA SD011071674SA SD011071675SA SD011071677SA SD011071679SA SD011071680SA SD011071681SA SD011071682SA SD011071683SA SD011071684SA SD011071689SA SD011071692SA SD011071694SA SD011071696SA SD011071697SA SD011071698SR SD011071699SA SD011071700SA SD011071701SA SD011071702SA SD011071704SA SD011071705SA SD011071706SA SD011071707SA SD011071708SA SD011071709SA SD011071710SA SD011071711SA SD011071712SA SD011071713SA SD011071714SA SD011071715SA SD011071716SA SD011071717SA SD011071718SA SD011071719SA SD011071720SA SD011071721SA SD011071722SA SD011071723SA SD011071724SA SD011071725SA SD011071726SA SD011071728SA SD011071734SA SD011071739SA SD011071740SA SD011071741SA SD011071742SA SD011071743SA SD011071744SA SD011071745SA SD011071746SA SD011071747SA SD011071748SA SD011071749SA SD011071750SA SD011071751SA SD011071752SA SD011071753SA SD011071754SA SD011071755SA SD011071757SA SD011071758SA SD011071759SA SD011071760SA SD011071761SA SD011071762SA SD011071763SA SD011071764SA SD011071765SA SD011071766SA SD011071767SA SD011071768SA SD011071768SR SD011071769SA SD011071770SA SD011071771SA SD011071772SA SD011071773SA SD011071774SA SD011071775SA SD011071776SA SD011071777SA SD011071778SA SD011071779SA SD011071780SA SD011071781SA SD011071782SA SD011071783SA SD011071784SA SD011071785SA SD011071786SA SD011071787SA SD011071788SA SD011071789SA SD011071790SA SD011071791SA SD011071792SA SD011071793SA SD011071794SA SD011071795SA SD011071796SA SD011071797SA SD011071798SA SD011071799SA SD011071800SA SD011071801SA SD011071802SA SD011071803SA SD011071804SA SD011071805SA SD011071806SA SD011071807SA SD011071808SA SD011071810SA SD011071811SA SD011071812SA SD011071813SA SD011071814SA SD011071815SA SD011071818SA SD011071820SA SD011071821SA SD011071822SA SD011071824SA SD011071826SA SD011071829SA SD011071830SA SD011071832SA SD011071835SA SD011071837SA SD011071838SA SD011071839SA SD011071841SA SD011071842SA SD011071843SA SD011071844SA SD011071845SA SD011071846SA SD011071847SA SD011071848SA SD011071849SA SD011071850SA SD011071851SA SD011071852SA SD011071853SA SD011071854SA SD011071855SA SD011071856SA SD011071857SA SD011071858SA SD011071859SR SD011071860SA SD011071861SA SD011071862SA SD011071863SA SD011071864SR SD011071865SA SD011071866SA SD011071867SA SD011071868SA SD011071869SA SD011071870SA SD011071871SA SD011071872SA SD011071873SA SD011071875SA SD011071876SA SD011071877SA SD011071878SA SD011071879SA SD011071881SA SD011071882SA SD011071883SA SD011071884SA SD011071886SA SD011071887SA SD011071889SA SD011071892SA SD011071893SA SD011071894SA SD011071896SA SD011071897SA SD011071898SA SD011071900SA SD011071901SA SD011071902SA SD011071904SA SD011071905SA SD011071906SA SD011071907SA SD011071908SA SD011071910SA SD011071912SA SD011071913SA SD011071915SA SD011071916SA SD011071917SA SD011071918SA SD011071919SA SD011071920SA SD011071921SA SD011071922SA SD011071923SA SD011071924SA SD011071925SA SD011071926SA SD011071927SA SD011071928SA SD011071929SA SD011071930SA SD011071931SA SD011071932SA SD011071933SA SD011071934SA SD011071935SA SD011071936SA SD011071937SA SD011071938SA SD011071939SA SD011071940SA SD011071941SA SD011071942SA SD011071943SA SD011071944SA SD011071945SA SD011071946SA SD011071947SA SD011071948SA SD011071949SA SD011071950SA SD011071951SA SD011071952SA SD011071953SA SD011071954SA SD011071955SA SD011071956SA SD011071958SA SD011071959SA SD011071960SA SD011071961SA SD011071962SA SD011071963SA SD011071965SA SD011071966SA SD011071967SA SD011071968SA SD011071969SA SD011071970SA SD011071971SA SD011071972SA SD011071973SA SD011071974SA SD011071975SA SD011071976SA SD011071977SA SD011071977SR SD011071978SA SD011071979SA SD011071980SA SD011071981SA SD011071982SA SD011071983SA SD011071989SA SD011071991SA SD011071992SA SD011071993SA SD011071994SA SD011071998SA SD011072003SA SD011072004SA SD011072005SA SD011072007SA SD011072009SA SD011072010SA SD011072012SA SD011072015SA SD011072016SA SD011072018SA SD011072020SA SD011072021SA SD011072022SA SD011072023SA SD011072025SA SD011072026SA SD011072028SA SD011072029SA SD011072030SA SD011072033SA SD011072034SA SD011072043SA SD011072044SA SD011072046SA SD011072047SA SD011072048SA SD011072049SA SD011072053SA SD011072055SA SD011072056SA SD011072057SA SD011072058SA SD011072060SA SD011072064SA SD011072065SA SD011072066SA SD011072071SA SD011072072SA SD011072073SA SD011072074SA SD011072076SA SD011072078SA SD011072082SA SD011072084SA SD011072086SA SD011072087SA SD011072088SA SD011072090SA SD011072091SA SD011072092SA SD011072102SA SD011072120SA SD011072121SA SD011072123SA SD011072125SA SD011072126SA SD011072130SA SD011072135SA SD011072136SA SD011072141SA SD011072143SA SD011072144SA SD011072147SA SD011072148SA SD011072149SA SD011072150SA SD011072151SA SD011072152SA SD011072153SA SD011072154SA SD011072156SA SD011072158SA SD011072161SA SD011073443GA SD011082166SA SD011082167SA SD011082168SA SD011082173SA SD011082174SA SD011082176SA SD011082177SA SD011082178SA SD011082179SA SD011082179SR SD011082181SA SD011082184SA SD011082187SA SD011082189SA SD011082190SA SD011082191SA SD011082192SA SD011082194SA SD011082195SA SD011082197SA SD011082200SA SD011082201SA SD011082202SA SD011082206SA SD011082212SA SD011082217SA SD011082218SA SD011082223SA SD011082226SA SD011082227SA SD011082234SA SD011082236SA SD011082238SA SD011082243SA SD011082244SA SD011082245SA SD011082246SA SD011082247SA SD011082248SA SD011082249SA SD011082250SA SD011082251SA SD011082253SA SD011082254SA SD011082255SA SD011082256SA SD011082257SA SD011082258SA SD011082260SA SD011082261SA SD011082262SA SD011082263SA SD011082264SA SD011082265SA SD011082266SA SD011082268SA SD011082269SA SD011082270SA SD011082271SA SD011082272SA SD011082273SA SD011082275SA SD011082276SA SD011082277SA SD011082278SA SD011082279SA SD011082280SA SD011082281SA SD011082282SA SD011082283SA SD011082285SA SD011082287SA SD011082288SA SD011082289SA SD011082290SA SD011082291SA SD011082292SA SD011082293SA SD011082294SA SD011082295SA SD011082297SA SD011082298SA SD011082299SA SD011082300SA SD011082301SA SD011082317SA SD011082318SA SD011082319SA SD011082320SA SD011082321SA SD011082322SA SD011082323SA SD011082324SA SD011082325SA SD011082326SA SD011082327SA SD011082328SA SD011082330SA SD011082331SA SD011082332SA SD011082333SA SD011082335SA SD011082337SA SD011082338SA SD011082339SA SD011082340SA SD011082341SA SD011082343SA SD011082344SA SD011082345SA SD011082346SA SD011082347SA SD011082348SA SD011082349SA SD011082350SA SD011082351SA SD011082352SA SD011082353SA SD011082354SA SD011082355SA SD011082356SA SD011082357SA SD011082359SA SD011082361SA SD011082362SA SD011082363SA SD011082364SA SD011082365SA SD011082366SA SD011082367SA SD011082368SA SD011082370SA SD011082371SA SD011082372SA SD011082373SA SD011082374SA SD011082375SA SD011082376SA SD011082377SA SD011082378SA SD011082379SA SD011082380SA SD011082381SA SD011082382SA SD011082384SA SD011082385SA SD011082386SA SD011082387SA SD011082388SA SD011082389SA SD011082390SA SD011082391SA SD011082392SA SD011083577GA SD011092393SA SD011092394SA SD011092395SA SD011092396SA SD011092397SA SD011092398SA SD011092399SA SD011092400SA SD011092401SA SD011092402SA SD011092403SA SD011092405SA SD011092406SA SD011092407SA SD011092408SA SD011092409SA SD011092410SA SD011092411SA SD011092412SA SD011092413SA SD011092414SA SD011092415SA SD011092416SA SD011092417SA SD011092418SA SD011092419SA SD011092420SA SD011092421SA SD011092422SA SD011092423SA SD011092424SA SD011092425SA SD011092426SA SD011092427SA SD011092428SA SD011092430SA SD011092433SA SD011092433SR SD011092434SA SD011092435SA SD011092437SA SD011092438SA SD011092438SR SD011092440SA SD011092442SA SD011092443SA SD011092444SA SD011092445SA SD011092446SA SD011092447SA SD011092448SA SD011092449SA SD011092450SA SD011092451SA SD011092452SA SD011092453SA SD011092455SA SD011092456SA SD011092457SA SD011092458SA SD011092459SA SD011092461SA SD011092462SA SD011092463SA SD011092465SA SD011092466SA SD011092467SA SD011092469SA SD011092471SA SD011092474SA SD011092478SA SD011092479SA SD011092483SA SD011092487SA SD011092489SA SD011092490SA SD011092491SA SD011092492SA SD011092493SA SD011092494SA SD011092495SA SD011092496SA SD011092497SA SD011092498SA SD011092499SA SD011092500SA SD011092501SA SD011092502SA SD011092503SA SD011092504SA SD011092505SA SD011092506SA SD011092507SA SD011092508SA SD011092509SA SD011092510SA SD011093700GA SD011093705GA SD011093709GA SD011093725GA SD011102511SA SD011102515SA SD011102516SA SD011102517SA SD011102519SA SD011102520SA SD011102521SA SD011102522SA SD011102523SA SD011102524SA SD011102525SA SD011102527SA SD011102528SA SD011102529SA SD011102531SA SD011102533SA SD011102534SA SD011102535SA SD011102536SA SD011103759GA SD011103760GA SD011103762GA SD011103763GA SD011103766GA SD011103767GA SD011103768GA SD011103770GA SD011103780GA SD011103792GA SD011103793GA SD011103795GA SD011103796GA SD011103805GA SD011103806GA SD011103807GA SD011103809GA SD011103810GA SD011103811GA SD011103812GA SD011103813GA SD011103814GA SD011103815GA SD011103816GA SD011103817GA SD011112540SA SD011112541SA SD011112542SA SD011112543SA SD011112544SA SD011112545SA SD011112546SA SD011112547SA SD011112549SA SD011112550SA SD011112551SA SD011112552SA SD011112553SA SD011112557SA SD011112563SA SD011112566SA SD011112567SA SD011112569SA SD011112571SA SD011112572SA SD011112573SA SD011112574SA SD011112575SA SD011112576SA SD011112577SA SD011112578SA SD011112579SA SD011112580SA SD011112581SA SD011112582SA SD011112583SA SD011112584SA SD011112586SA SD011112588SA SD011112589SA SD011112590SA SD011112591SA SD011112592SA SD011112593SA SD011112594SA SD011112595SA SD011112596SA SD011112599SA SD011112601SA SD011112602SA SD011112603SA SD011112607SA SD011112608SA SD011112609SA SD011112610SA SD011112611SA SD011112612SA SD011112615SA SD011112617SA SD011112618SA SD011112619SA SD011112622SA SD011112623SA SD011112624SA SD011112625SA SD011112626SA SD011112628SA SD011112629SA SD011112632SA SD011112633SA SD011112636SA SD011112637SA SD011112639SA SD011112640SA SD011112642SA SD011112643SA SD011112644SA SD011112645SA SD011112646SA SD011112647SA SD011112648SA SD011112649SA SD011112650SA SD011112651SA SD011112652SA SD011112653SA SD011112654SA SD011112656SA SD011112657SA SD011112658SA SD011112659SA SD011112660SA SD011112661SA SD011112662SA SD011112663SA SD011112664SA SD011112665SA SD011112666SA SD011112667SA SD011112668SA SD011112669SA SD011112670SA SD011112671SA SD011112672SA SD011112673SA SD011112674SA SD011112675SA SD011112676SA SD011112677SA SD011112678SA SD011112679SA SD011112683SA SD011112684SA SD011112686SA SD011112689SA SD011112694SA SD011112696SA SD011112697SA SD011112698SA SD011112699SA SD011112700SA SD011112702SA SD011112704SA SD011112704SR SD011112706SA SD011112707SA SD011112708SA SD011112709SA SD011112711SA SD011112712SA SD011112713SA SD011112714SA SD011112715SA SD011112716SA SD011112717SA SD011112718SA SD011112719SA SD011112720SA SD011112721SA SD011112722SA SD011112723SA SD011112725SA SD011112725SR SD011112726SA SD011112727SA SD011112732SA SD011112733SA SD011112734SA SD011112735SA SD011112737SA SD011112738SA SD011112739SA SD011112743SA SD011112745SA SD011112746SA SD011112747SA SD011112750SA SD011112751SA SD011112752SA SD011112754SA SD011112755SA SD011112755SR SD011112756SA SD011112762SA SD011112764SA SD011112764SR SD011112767SA SD011112768SA SD011112769SA SD011112769SR SD011112770SA SD011112771SA SD011112773SA SD011112776SA SD011112777SA SD011112778SA SD011112781SA SD011112788SA SD011112791SA SD011112794SA SD011112801SA SD011113818GA SD011113819GA SD011113820GA SD011113821GA SD011113823GA SD011113824GA SD011113825GA SD011113826GA SD011113827GA SD011113828GA SD011113830GA SD011113832GA SD011113834GA SD011113836GA SD011113837GA SD011113839GA SD011113840GA SD011113844GA SD011113846GA SD011113848GA SD011113853GA SD011113855GA SD011113857GA SD011113859GA SD011113861GA SD011113862GA SD011113870GA SD011113872GA SD011113874GA SD011113875GA SD011113878GA SD011113879GA SD011113881GA SD011113882GA SD011113883GA SD011113884GA SD011113885GA SD011113886GA SD011113888GA SD011113889GA SD011113890GA SD011113891GA SD011113916GA SD011113917GA SD011113919GA SD011113920GA SD011113921GA SD011113926GA SD011113927GA SD011113928GA SD011113930GA SD011113931GA SD011113932GA SD011113936GA SD011113950GA SD011113952GA SD011113992GA SD011113994GA SD011122807SA SD011122808SA SD011122811SA SD011122812SA SD011122813SA SD011122815SA SD011122817SA SD011122819SA SD011122820SA SD011122821SA SD011122822SA SD011122823SA SD011122824SA SD011122825SA SD011122826SA SD011122827SA SD011122828SA SD011122829SA SD011122830SA SD011122831SA SD011122832SA SD011122833SA SD011122834SA SD011122835SA SD011122837SA SD011122838SA SD011122839SA SD011122840SA SD011122841SA SD011122842SA SD011122843SA SD011122844SA SD011122845SA SD011122846SA SD011122847SA SD011122848SA SD011122849SA SD011122850SA SD011122851SA SD011122852SA SD011122853SA SD011122855SA SD011122856SA SD011122857SA SD011122858SA SD011122859SA SD011122860SA SD011122861SA SD011122862SA SD011122863SA SD011122864SA SD011122865SA SD011122866SA SD011122867SA SD011122867SR SD011122868SA SD011122869SA SD011122870SA SD011122871SA SD011122872SA SD011122873SA SD011122874SA SD011122875SA SD011122876SA SD011122877SA SD011122878SA SD011122879SA SD011122880SA SD011122882SA SD011122884SA SD011122885SA SD011122886SA SD011122887SA SD011122889SA SD011122890SA SD011122891SA SD011122892SA SD011122893SA SD011122894SA SD011122895SA SD011122896SA SD011122897SA SD011122900SA SD011122901SA SD011122902SA SD011122903SA SD011122904SA SD011122905SA SD011122906SA SD011122907SA SD011122908SA SD011122909SA SD011122910SA SD011122911SA SD011122912SA SD011122913SA SD011122914SA SD011122917SA SD011122918SA SD011122919SA SD011122920SA SD011122921SA SD011122922SA SD011122927SA SD011122928SA SD011122929SA SD011122931SA SD011122932SA SD011122935SA SD011122937SA SD011122938SA SD011122939SA SD011122940SA SD011122941SA SD011122944SA SD011122947SA SD011122948SA SD011122953SA SD011122958SA SD011122959SA SD011122961SA SD011122966SA SD011122968SA SD011122969SA SD011122971SA SD011122972SA SD011122973SA SD011122974SA SD011122975SA SD011122976SA SD011122977SA SD011122978SA SD011122979SA SD011122980SA SD011122981SA SD011122983SA SD011122984SA SD011122985SA SD011122987SA SD011122988SA SD011122989SA SD011122993SA SD011122994SA SD011122997SA SD011122998SA SD011122999SA SD011123000SA SD011123001SA SD011123002SA SD011123003SA SD011123003SABOP SD011123006SA SD011123007SA SD011123009SA SD011123010SA SD011123011SA SD011123012SA SD011123013SA SD011123014SA SD011123016SA SD011123017SA SD011123019SA SD011123020SA SD011123021SA SD011123024SA SD011123025SA SD011123026SA SD011123028SA SD011123029SA SD011123030SA SD011123031SA SD011123032SA SD011123033SA SD011123036SA SD011123037SA SD011123038SA SD011123039SA SD011123040SA SD011123041SA SD011123042SA SD011123043SA SD011123044SA SD011123045SA SD011123047SA SD011123049SA SD011123053SA SD011123054SA SD011123055SA SD011123056SA SD011123057SA SD011123058SA SD011123059SA SD011123061SA SD011123064SA SD011123066SA SD011123070SA SD011123073SA SD011123074SA SD011123079SA SD011123084SA SD011123085SA SD011123086SA SD011123087SA SD011123088SA SD011123089SA SD011123090SA SD011123091SA SD011123092SA SD011123093SA SD011123094SA SD011123096SA SD011123097SA SD011123098SA SD011123099SA SD011123100SA SD011123101SA SD011123102SA SD011123103SA SD011123104SA SD011123105SA SD011123106SA SD011123107SA SD011123108SA SD011123111SA SD011123112SA SD011123113SA SD011123114SA SD011123115SA SD011123116SA SD011123118SA SD011123119SA SD011123120SA SD011123121SA SD011123122SA SD011123123SA SD011123124SA SD011123125SA SD011123126SA SD011123127SA SD011123128SA SD011123129SA SD011123130SA SD011123131SA SD011123133SA SD011123134SA SD011123135SA SD011123136SA SD011123137SA SD011123139SA SD011123141SA SD011123142SA SD011123143SA SD011123144SA SD011123145SA SD011123146SA SD011123147SA SD011123148SA SD011123149SA SD011123150SA SD011123151SA SD011123152SA SD011123153SA SD011123154SA SD011123155SA SD011123156SA SD011123157SA SD011123158SA SD011123159SA SD011123160SA SD011123161SA SD011123162SA SD011123163SA SD011123164SA SD011123165SA SD011123166SA SD011123167SA SD011123168SA SD011123169SA SD011123170SA SD011123171SA SD011123172SA SD011123173SA SD011123174SA SD011123176SA SD011123177SA SD011123179SA SD011123180SA SD011123181SA SD011123182SA SD011123183SA SD011123184SA SD011123185SA SD011123186SA SD011123187SA SD011123188SA SD011123189SA SD011123190SA SD011123192SA SD011123193SA SD011123194SA SD011123195SA SD011123196SA SD011123197SA SD011123198SA SD011123199SA SD011123200SA SD011123201SA SD011123202SA SD011123203SA SD011123204SA SD011123205SA SD011123206SA SD011123207SA SD011123208SA SD011123209SA SD011123210SA SD011123211SA SD011123212SA SD011123213SA SD011123214SA SD011123215SA SD011123216SA SD011123217SA SD011123218SA SD011123219SA SD011123220SA SD011123221SA SD011123222SA SD011123223SA SD011123224SA SD011123225SA SD011123226SA SD011123227SA SD011123228SA SD011123229SA SD011123230SA SD011123231SA SD011123232SA SD011123233SA SD011123234SA SD011123235SA SD011123236SA SD011123238SA SD011123239SA SD011123239SR SD011123240SA SD011123241SA SD011123242SA SD011123254SA SD011123272SA SD011123278SA SD011123285SA SD011123288SA SD011123289SA SD011123290SA SD011123293SA SD011123294SA SD011123296SA SD011123297SA SD011123298SA SD011123302SA SD011123303SA SD011123304SA SD011123306SA SD011123308SA SD011123309SA SD011123311SA SD011123313SA SD011123314SA SD011123321SA SD011123322SA SD011123323SA SD011123334SA SD011123336SA SD011123337SA SD011123338SA SD011123342SA SD011123344SA SD011123345SA SD011123346SA SD011123347SA SD011123348SA SD011123349SA SD011123353SA SD011123354SA SD011123358SA SD011123360SA SD011123361SA SD011123363SA SD011124030GA SD011124031GA SD011124033GA SD011124034GA SD011124037GA SD011124038GA SD011124039GA SD011124040GA SD011124041GA SD011124042GA SD011124043GA SD011124045GA SD011124046GA SD011124047GA SD011124049GA SD011124050GA SD011124051GA SD011124052GA SD011124053GA SD011124078GA SD011124103GA SD011124132GA SD011124134GA SD011124136GA SD011124174GA SD011124178GA SD011124193GA SD011124194GA SD011124199GA SD011124205GA SD011124206GA SD011124207GA SD011124209GA SD011124210GA SD011124212GA SD011124213GA SD011124214GA SD011124215GR SD011124217GA SD011124218GA SD011124219GA SD011124228GA SD011124231GA SD011124233GA SD011133364SA SD011133365SA SD011133366SA SD011133367SA SD011133369SA SD011133370SA SD011133371SA SD011133372SA SD011133373SA SD011133374SA SD011133375SA SD011133377SA SD011133378SA SD011133379SA SD011133380SA SD011133381SA SD011133384SA SD011133385SA SD011133387SA SD011133388SA SD011133389SA SD011133391SA SD011133393SA SD011133394SA SD011133397SA SD011133398SA SD011133399SA SD011133400SA SD011133402SA SD011133405SA SD011133406SA SD011133407SA SD011133408SA SD011133409SA SD011133410SA SD011133411SA SD011133412SA SD011133413SA SD011133414SA SD011133416SA SD011133417SA SD011133419SA SD011133420SA SD011133421SA SD011133422SA SD011133423SA SD011133424SA SD011133425SA SD011133426SA SD011133427SA SD011133428SA SD011133429SA SD011133430SA SD011133431SA SD011133432SA SD011133433SA SD011133435SA SD011133436SA SD011133439SA SD011133440SA SD011133442SA SD011133443SA SD011133444SA SD011133445SA SD011133446SA SD011133447SA SD011133448SA SD011133449SA SD011133450SA SD011133451SA SD011133452SA SD011133453SA SD011133454SA SD011133455SA SD011133456SA SD011133457SA SD011133459SA SD011133460SA SD011133461SA SD011133462SA SD011133463SA SD011133464SA SD011133465SA SD011133466SA SD011133467SA SD011133468SA SD011133469SA SD011133470SA SD011133471SA SD011133472SA SD011133473SA SD011133475SA SD011133476SA SD011133479SA SD011133480SA SD011133482SA SD011133483SA SD011133484SA SD011133485SA SD011133486SA SD011133487SA SD011133488SA SD011133489SA SD011133490SA SD011133491SA SD011133492SA SD011133493SA SD011133494SA SD011133495SA SD011133496SA SD011133497SA SD011133498SA SD011133499SA SD011133500SA SD011133501SA SD011133502SA SD011133503SA SD011133504SA SD011133505SA SD011133506SA SD011133507SA SD011133508SA SD011133509SA SD011133510SA SD011133511SA SD011133512SA SD011133513SA SD011133514SA SD011133515SA SD011133516SA SD011133517SA SD011133518SA SD011133519SA SD011133520SA SD011133521SA SD011133522SA SD011133523SA SD011133524SA SD011133525SA SD011133526SA SD011133527SA SD011133528SA SD011133529SA SD011133530SA SD011133531SA SD011133532SA SD011133534SA SD011133536SA SD011133538SA SD011133539SA SD011133540SA SD011133541SA SD011133542SA SD011133543SA SD011133544SA SD011133546SA SD011133547SA SD011133549SA SD011133550SA SD011133551SA SD011133552SA SD011133553SA SD011133554SA SD011133555SA SD011133556SA SD011133557SA SD011133558SA SD011133559SA SD011133560SA SD011133561SA SD011133562SA SD011133564SA SD011133565SA SD011133566SA SD011133567SA SD011133568SA SD011133569SA SD011133570SA SD011133571SA SD011133572SA SD011133573SA SD011133574SA SD011133575SA SD011133576SA SD011133577SA SD011133578SA SD011133579SA SD011133580SA SD011133581SA SD011133582SA SD011133583SA SD011133584SA SD011133585SA SD011133586SA SD011133587SA SD011133588SA SD011133589SA SD011133590SA SD011133591SA SD011133592SA SD011133593SA SD011133595SA SD011133596SA SD011133597SA SD011133598SA SD011133599SA SD011133600SA SD011133601SA SD011133602SA SD011133603SA SD011133604SA SD011133605SA SD011133606SA SD011133608SA SD011133610SA SD011133611SA SD011133612SA SD011133613SA SD011133614SA SD011133615SA SD011133616SA SD011133617SA SD011133618SA SD011133619SA SD011133621SA SD011133623SA SD011133624SA SD011133625SA SD011133626SA SD011133627SA SD011133628SA SD011133629SA SD011133630SA SD011133631SA SD011133632SA SD011133633SA SD011133634SA SD011133635SA SD011133636SA SD011133637SA SD011133638SA SD011133639SA SD011133640SA SD011133641SA SD011133642SA SD011133643SA SD011133645SA SD011133646SA SD011133647SA SD011133649SA SD011133650SA SD011133651SA SD011133652SA SD011133653SA SD011133653SR SD011133654SA SD011133655SA SD011133656SA SD011133657SA SD011133660SA SD011133661SA SD011133662SA SD011133663SA SD011133664SA SD011133665SA SD011133666SA SD011133668SA SD011133669SA SD011133670SA SD011133671SA SD011133672SA SD011133673SA SD011133675SA SD011133676SA SD011133678SA SD011133679SA SD011133680SA SD011133682SA SD011133683SA SD011133691SA SD011133692SA SD011133693SA SD011133694SA SD011133695SA SD011133697SA SD011133699SA SD011133703SA SD011133705SA SD011133742SA SD011133744SA SD011134235GA SD011134236GR SD011134237GR SD011134238GA SD011134239GA SD011134241GA SD011134243GA SD011134244GA SD011134245GA SD011134246GR SD011134250GA SD011134251GA SD011134252GA SD011134253GA SD011134254GA SD011134256GA SD011134257GA SD011134259GA SD011134260GA SD011134274GA SD011134277GA SD011134278GA SD011134286GA SD011134287GA SD011134291GA SD011134294GA SD011134298GA SD011134299GA SD011134300GA SD011134302GA SD011134303GA SD011134304GA SD011134305GA SD011134307GA SD011134308GA SD011134309GA SD011134311GA SD011134312GA SD011134313GA SD011134316GA SD011134317GA SD011134320GA SD011134321GA SD011134322GA SD011134323GA SD011134325GA SD011134326GA SD011134327GA SD011134328GA SD011134329GA SD011134330GA SD011134332GA SD011134333GA SD011134334GA SD011134335GA SD011134336GA SD011134337GA SD011134338GA SD011134340GA SD011134345GA SD011134346GA SD011134348GR SD011134349GA SD011134351GA SD011134353GA SD011134355GA SD011134362GA SD011134378GA SD011134379GA SD011134381GA SD011134382GA SD011134395GA SD011134396GA SD011143745SA SD011143746SA SD011143747SA SD011143748SA SD011143750SA SD011143751SA SD011143752SA SD011143753SA SD011143754SA SD011143755SA SD011143756SA SD011143757SA SD011143758SA SD011143760SA SD011143762SA SD011143763SA SD011143764SA SD011143765SA SD011143767SA SD011143768SA SD011143769SA SD011143770SA SD011143771SA SD011143772SA SD011143773SA SD011143774SA SD011143775SA SD011143776SA SD011143777SA SD011143778SA SD011143779SA SD011143780SA SD011143781SA SD011143782SA SD011143784SA SD011143786SA SD011143787SA SD011143788SA SD011143790SA SD011143792SA SD011143793SA SD011143794SA SD011143795SA SD011143796SA SD011143797SA SD011143798SA SD011143799SA SD011143801SA SD011143802SA SD011143803SA SD011143804SA SD011143805SA SD011143806SA SD011143807SA SD011143808SA SD011143811SA SD011143812SA SD011143815SA SD011143817SA SD011143819SA SD011143820SA SD011143823SA SD011143824SA SD011143825SA SD011143826SA SD011143827SA SD011143828SA SD011143829SA SD011143830SA SD011143831SA SD011143832SA SD011143833SA SD011143834SA SD011143835SA SD011143836SA SD011143837SA SD011143838SA SD011143839SA SD011143840SA SD011143841SA SD011143842SA SD011143843SA SD011143844SA SD011143845SA SD011143846SA SD011143847SA SD011143848SA SD011143849SA SD011143850SA SD011143851SA SD011143852SA SD011143853SA SD011143854SA SD011143856SA SD011143857SA SD011143858SA SD011143859SA SD011143860SA SD011143862SA SD011143867SA SD011143869SA SD011143871SA SD011143873SA SD011143874SA SD011143894SA SD011143910SA SD011143911SA SD011143912SA SD011143913SA SD011143914SA SD011143915SA SD011143916SA SD011143921SA SD011143926SA SD011143929SA SD011143929SR SD011143930SA SD011143931SA SD011143931SR SD011143932SA SD011143934SA SD011143935SA SD011143936SA SD011143936SR SD011143937SA SD011143938SA SD011143938SR SD011143942SA SD011143942SR SD011143945SA SD011143945SR SD011143946SA SD011143950SA SD011143950SR SD011143951SA SD011143957SA SD011143959SA SD011143959SR SD011143964SA SD011143967SA SD011143972SA SD011143972SR SD011143974SA SD011143975SA SD011143975SR SD011143976SA SD011143977SA SD011143977SR SD011143978SA SD011143978SR SD011143979SA SD011143979SR SD011143980SA SD011143980SR SD011143981SA SD011143983SA SD011143984SA SD011143986SA SD011143986SR SD011144417GA SD011144419GA SD011144421GA SD011144423GA SD011144424GA SD011144425GA SD011144426GA SD011144427GA SD011144429GA SD011144431GA SD011144432GA SD011144433GA SD011144434GA SD011144435GA SD011144436GA SD011144444GA SD011144448GA SD011144457GA SD011144458GR SD011144459GR SD011144460GA SD011144464GA SD011144469GA SD011144472GR SD011144473GR SD011144474GR SD011144475GR SD011144476GR SD011144478GA SD011144479GA SD011144480GA SD011144482GA SD011144483GA SD011144484GA SD011144485GA SD011144486GA SD011153987SA SD011153988SA SD011153989SA SD011153990SA SD011153991SA SD011153992SA SD011153993SA SD011153994SA SD011153995SA SD011153996SA SD011154012SA SD011154014SA SD011154017SA SD011154021SA SD011154022SA SD011154023SA SD011154024SA SD011154025SA SD011154026SA SD011154027SA SD011154028SA SD011154029SA SD011154030SA SD011154031SA SD011154032SA SD011154033SA SD011154034SA SD011154035SA SD011154036SA SD011154037SA SD011154038SA SD011154039SA SD011154040SA SD011154041SA SD011154042SA SD011154043SA SD011154044SA SD011154045SA SD011154046SA SD011154047SA SD011154048SA SD011154049SA SD011154050SA SD011154051SA SD011154052SA SD011154053SA SD011154054SA SD011154055SA SD011154056SA SD011154057SA SD011154058SA SD011154059SA SD011154060SA SD011154061SA SD011154062SA SD011154063SA SD011154064SA SD011154065SA SD011154066SA SD011154067SA SD011154068SA SD011154069SA SD011154070SA SD011154071SA SD011154535GA SD011154538GA SD011154541GA SD011154544GA SD011154547GA SD011154549GA SD011154551GA SD011154584GA SD011154585GA SD011154586GA SD011154591GA SD011164073SA SD011164075SA SD011164077SA SD011164078SA SD011164079SA SD011164080SA SD011164081SA SD011164082SA SD011164083SA SD011164085SA SD011164087SA SD011164088SA SD011164089SA SD011164090SA SD011164091SA SD011164092SA SD011164093SA SD011164094SA SD011164095SA SD011164096SA SD011164097SA SD011164098SA SD011164100SA SD011164102SA SD011164103SA SD011164106SA SD011164107SA SD011164110SA SD011164115SA SD011164116SA SD011164118SA SD011164119SA SD011164120SA SD011164121SA SD011164122SA SD011164126SA SD011164128SA SD011164130SA SD011164131SA SD011164132SA SD011164133SA SD011164134SA SD011164135SA SD011164136SA SD011164137SA SD011164138SA SD011164139SA SD011164140SA SD011164141SA SD011164142SA SD011164143SA SD011164144SA SD011164145SA SD011164146SA SD011164147SA SD011164148SA SD011164149SA SD011164150SA SD011164151SA SD011174152SA SD011174159SA SD011174161SA SD011174164SA SD011174165SA SD011174166SA SD011174167SA SD011174168SA SD011174169SA SD011174171SA SD011174172SA SD011174173SA SD011174176SA SD011174177SA SD011174183SA SD011174184SA SD011174185SA SD011174186SA SD011174187SA SD011174188SA SD011174190SA SD011174191SA SD011174192SA SD011174194SA SD011174195SA SD011174196SA SD011174197SA SD011174198SA SD011174199SA SD011174200SA SD011174201SA SD011174202SA SD011174204SA SD011174205SA SD011174206SA SD011174207SA SD011174208SA SD011174209SA SD011174210SA SD011174211SA SD011174212SA SD011174213SA SD011174214SA SD011174215SA SD011174216SA SD011174217SA SD011174218SA SD011174219SA SD011174220SA SD011174221SA SD011174222SA SD011174223SA SD011174226SA SD011174227SA SD011174228SA SD011174230SA SD011174235SA SD011174238SA SD011174240SA SD011174241SA SD011174244SA SD011174245SA SD011174249SA SD011174251SA SD011174253SA SD011174263SA SD011174266SA SD011174269SA SD011174271SA SD011174274SA SD011174275SA SD011174276SA SD011174278SA SD011174280SA SD011174281SA SD011174285SA SD011174286SA SD011174288SA SD011174289SA SD011174291SA SD011174294SA SD011174295SA SD011174298SA SD011174302SA SD011174303SA SD011174304SA SD011174305SA SD011174307SA SD011174309SA SD011174310SA SD011174311SA SD011174749GA SD011174750GA SD011174752GA SD011174753GA SD011174754GA SD011174755GA SD011174756GA SD011174757GA SD011174758GA SD011174759GA SD011174760GA SD011174761GA SD011174762GA SD011174765GA SD011174778GA SD011174781GA SD011174794GA SD011174797GA SD011174799GA SD011174801GA SD011174803GA SD011174804GA SD011174805GA SD011184332SA SD011184333SA SD011184334SA SD011184335SA SD011184336SA SD011184337SA SD011184338SA SD011184339SA SD011184340SR SD011184341SA SD011184342SA SD011184343SA SD011184344SA SD011184345SA SD011184346SA SD011184347SA SD011184349SA SD011184355SA SD011184362SA SD011184363SA SD011184364SA SD011184365SA SD011184365SR SD011184366SA SD011184367SA SD011184368SA SD011184369SA SD011184370SA SD011184373SA SD011184376SA SD011184379SA SD011184381SA SD011184382SA SD011184383SA SD011184384SA SD011184386SA SD011184387SA SD011184388SA SD011184389SA SD011184401SA SD011184406SA SD011184407SA SD011184423SA SD011184424SA SD011184425SA SD011184426SA SD011184427SA SD011184428SA SD011184429SA SD011184430SA SD011184431SA SD011184432SA SD011184433SA SD011184434SA SD011184435SA SD011184436SA SD011184437SA SD011184438SA SD011184439SA SD011184440SA SD011184441SA SD011184442SA SD011184443SR SD011184444SA SD011184445SA SD011184447SA SD011184448SABOP SD011184449SA SD011184450SA SD011184451SA SD011184453SR SD011184454SA SD011184455SA SD011184455SR SD011184456SA SD011184459SA SD011184462SA SD011184462SABOP SD011184464SA SD011184465SA SD011184466SA SD011184467SA SD011184468SA SD011184469SA SD011184470SA SD011184471SA SD011184471SR SD011184472SA SD011184473SA SD011184474SA SD011184475SA SD011184476SA SD011184891GA SD011184892GA SD011184894GA SD011184895GA SD011184896GA SD011184899GA SD011184901GA SD011184902GA SD011184903GA SD011184908GA SD011184909GA SD011194477SA SD011194478SA SD011194479SA SD011194480SA SD011194481SA SD011194482SA SD011194484SA SD011194485SA SD011194487SA SD011194488SA SD011194489SA SD011194490SA SD011194492SA SD011194493SA SD011194495SA SD011194496SA SD011194497SA SD011194498SA SD011194499SA SD011194500SA SD011194501SA SD011194502SA SD011194503SA SD011194504SA SD011194505SA SD011194506SA SD011194508SA SD011194510SA SD011194511SA SD011194512SA SD011194513SA SD011194514SA SD011194515SA SD011194519SA SD011194521SA SD011194523SA SD011194526SA SD011194527SA SD011194529SA SD011194530SA SD011194531SA SD011194532SA SD011194534SA SD011194535SA SD011194536SA SD011194537SA SD011194538SA SD011194539SA SD011194540SA SD011194541SA SD011194542SA SD011194543SA SD011194544SA SD011194545SA SD011194546SA SD011194547SA SD011194548SA SD011194550SA SD011194551SA SD011194552SA SD011194553SA SD011194554SA SD011194555SA SD011194556SA SD011194557SA SD011194558SA SD011194559SA SD011194560SA SD011194561SA SD011194562SA SD011194563SA SD011194564SA SD011194565SA SD011194567SA SD011194568SA SD011194571SA SD011194572SA SD011194574SA SD011194575SA SD011194576SA SD011194577SA SD011194578SA SD011194579SA SD011194580SA SD011194581SA SD011194582SA SD011194585SA SD011194588SA SD011194589SA SD011194591SA SD011194592SA SD011194593SA SD011194594SA SD011194595SA SD011194597SA SD011194598SA SD011194600SA SD011194602SA SD011194603SA SD011194606SA SD011194610SA SD011194611SA SD011194613SA SD011194615SA SD011194616SA SD011194617SA SD011194620SA SD011194621SA SD011194622SA SD011194623SA SD011194624SA SD011194625SA SD011194626SA SD011194627SA SD011194628SA SD011194629SA SD011194630SA SD011194631SA SD011194632SA SD011194633SA SD011194634SA SD011194636SA SD011194637SA SD011194638SA SD011194639SA SD011194640SA SD011194641SA SD011194642SA SD011194643SA SD011194644SA SD011194645SA SD011194646SA SD011194647SA SD011194648SA SD011194649SA SD011194650SA SD011194651SA SD011204652SA SD011204653SA SD011204655SA SD011204657SA SD011204658SA SD011204659SA SD011204660SA SD011204661SA SD011204663SA SD011204664SA SD011204665SA SD011204666SA SD011204667SA SD011204668SA SD011204670SA SD011204672SA SD011204673SA SD011204674SA SD011204676SA SD011204678SA SD011204679SA SD011204680SA SD011204681SA SD011204682SA SD011204683SA SD011204684SA SD011204685SA SD011204687SA SD011204688SA SD011204689SA SD011204690SA SD011204692SA SD011204693SA SD011204694SA SD011204695SA SD011204696SA SD011204697SA SD011204698SA SD011204699SA SD011204700SA SD011204702SA SD011204703SA SD011204704SA SD011204705SA SD011204706SA SD011204707SA SD011204708SA SD011204710SA SD011204713SA SD011204714SA SD011204715SA SD011204716SA SD011204717SA SD011204718SA SD011204719SA SD011204720SA SD011204722SA SD011204723SA SD011204724SA SD011204725SA SD011204726SA SD011204727SA SD011204728SA SD011204729SA SD011204731SA SD011204732SA SD011204733SA SD011204734SA SD011204736SA SD011204737SA SD011204740SA SD011204742SA SD011204744SA SD011204745SA SD011204749SA SD011204756SA SD011204763SA SD011204767SA SD011204768SA SD011204769SA SD011204770SA SD011204771SA SD011214772SA SD011214773SA SD011214774SA SD011214775SA SD011214776SA SD011214778SA SD011214779SA SD011214780SA SD011214781SA SD011214782SA SD011214783SA SD011214785SA SD011214786SA SD011214787SA SD011214788SA SD011214790SA SD011214791SA SD011214792SA SD011214793SA SD011214794SA SD011214795SA SD011214796SA SD011214797SA SD011214799SA SD011214800SA SD011214801SA SD011214802SA SD011214803SA SD011214805SA SD011214806SA SD011214807SA SD011214808SA SD011214809SA SD011214810SA SD011214811SA SD011214812SA SD011214813SA SD011214814SA SD011214815SA SD011214817SA SD011214818SA SD011214819SA SD011214820SA SD011214821SA SD011214822SA SD011214823SA SD011214824SA SD011214825SA SD011214826SA SD011214827SA SD011214828SA SD011214829SA SD011214830SA SD011214831SA SD011214832SA SD011214833SA SD011214834SA SD011214835SA SD011214836SA SD011214837SA SD011214838SA SD011214839SA SD011214840SA SD011214841SA SD011214842SA SD011214843SA SD011214844SA SD011214845SA SD011214846SA SD011214847SA SD011214848SA SD011214849SA SD011214850SA SD011214851SA SD011214852SA SD011214853SA SD011214855SA SD011214856SA SD011214857SA SD011214858SA SD011214859SA SD011214860SA SD011214862SA SD011214863SA SD011214864SA SD011214865SA SD011214866SA SD011214878SA SD011214879SA SD011214880SA SD011214881SA SD011214882SA SD011214884SA SD011214896SA SD011214897SA SD011214898SA SD011214902SA SD011214903SA SD011214904SA SD011214905SA SD011214906SA SD011214908SA SD011214909SA SD011214910SA SD011214911SA SD011214912SA SD011214915SA SD011214917SA SD011214919SA SD011214920SA SD011214921SA SD011224924SA SD011224925SA SD011224926SA SD011224929SA SD011224930SA SD011224931SA SD011224933SA SD011224934SA SD011224935SA SD011224936SA SD011224937SA SD011224938SA SD011224939SA SD011224940SA SD011224941SA SD011224942SA SD011224943SA SD011224944SA SD011224945SA SD011224946SA SD011224947SA SD011224948SA SD011224949SA SD011224951SA SD011224952SA SD011224953SA SD011224954SA SD011224955SA SD011224956SA SD011224957SA SD011224958SA SD011224959SA SD011224960SA SD011224961SR SD011224962SA SD011224963SA SD011224964SA SD011224965SA SD011224966SA SD011224967SA SD011224968SA SD011224969SA SD011224970SA SD011224971SA SD011224978SA SD011224978SR SD011224985SA SD011224991SA SD011224991SR SD011224997SA SD011224998SA SD011224999SA SD011225000SA SD011225003SA SD011225004SA SD011225005SA SD011225006SA SD011225007SA SD011225010SA SD011225011SA SD011225012SA SD011225014SA SD011225015SA SD011225016SA SD011225017SA SD011225018SA SD011225019SA SD011225020SA SD011225021SA SD011225024SA SD011225025SA SD011225025SABOP SD011225026SA SD011225028SA SD011225029SA SD011225030SA SD011225034SA SD011225035SA SD011225036SA SD011225038SA SD011225039SA SD011225041SA SD011225043SA SD011225044SA SD011225045SA SD011225047SA SD011225048SA SD011225049SA SD011225050SA SD011225051SA SD011225052SA SD011225059SA SD011225060SA SD011225061SA SD011225062SA SD011225062SR SD011225063SA SD011225065SA SD011225066SA SD011225071SA SD011235072SA SD011235073SA SD011235074SA SD011235075SA SD011235076SA SD011235077SA SD011235078SA SD011235079SA SD011235080SA SD011235081SA SD011235082SA SD011235083SA SD011235084SA SD011235086SA SD011235087SA SD011235088SA SD011235089SA SD011235090SA SD011235091SA SD011235092SA SD011235093SA SD011235094SA SD011235095SA SD011235096SA SD011235097SA SD011235098SA SD011235099SA SD011235100SA SD011235101SA SD011235102SA SD011235103SA SD011235104SA SD011235105SA SD011235106SA SD011235109SA SD011235111SA SD011235112SA SD011235113SA SD011235114SA SD011235115SA SD011235116SA SD011235116SABOP SD011235117SA SD011235119SA SD011235120SA SD011235123SA SD011235125SA SD011235127SA SD011235128SA SD011236093SA SD011236097SA SHQ11072160SA SHQ11072160SR SHQ11143785SA SHQ11190002SA SHQ11190003SA SHQ11190022SA SHQ11190023SA SHQ11190024SA SHQ11190025SA SHQ11190027SA SHQ11190028SA SHQ11190029SA SHQ11190031SA SHQ11190035GA SHQ11190036GA SHQ11190037GA SHQ11190038GA SHQ11190039GA SHQ11230032SA SHQ11230034SA SHQ11230035SA SHQ11230036SA SHQ11230037SA SHQ11230039SA SHQ11230040SA SHQ11230041SA SHQ11230042SA SHQ11230044SA SHQ11230046SA SHQ11230047SA SHQ11230048SA SHQ11230049SA SHQ11230050SA SHQ11230051SA SHQ11230052SA SHQ11230054SA SHQ11230061SA SHQ11230071SA SHQ11230078SA SHQ11230091SA SHQ11230092SA SHQ11230092SR SHQ11230095SA SHQ11230096SA SHQ11230097SA SHQ11230099SA SHQ11230102SA SHQ11230103SA SHQ11230105SA SHQ11230107SA SHQ11230108SA SHQ11230109SA SHQ11230110SA SHQ11230111SA SHQ11230112SA SHQ11230113SA SHQ11230115SA SHQ11230117SA SHQ11230119SA SHQ11230120SA SHQ11230122SR SHQ11235129SA SHQ11235130SA SHQ11235131SA SHQ11235132SA SHQ11235133SA SHQ11235134SA SHQ11235135SA SHQ11235136SA SHQ11235138SA SHQ11240124SA SHQ11240126SA SHQ11240128SA SHQ11240129SA SHQ11240130SA SHQ11240131SA SHQ11240132SA SHQ11240133SA SHQ11240134SA SHQ11240135SA SHQ11240136SA SHQ11240137SA SHQ11240138SA SHQ11240139SA SHQ11240140SA SHQ11240143SA SHQ11240144SA SHQ11240145SA SHQ11240146SA SHQ11240148SA SHQ11240149SA SHQ11240152SA SHQ11240153SA SHQ11240154SA SHQ11240155SA SHQ11240156SA SHQ11240157SA SHQ11240159SA SHQ11240160SA SHQ11240161SA SHQ11240162SA SHQ11240163SA SHQ11240165SA SHQ11240166SA SHQ11240167SA SHQ11240168SA SHQ11240169SA SHQ11240170SA SHQ11240171SA SHQ11240172SA SHQ11240173SA SHQ11240174SA SHQ11240175SA SHQ11240180SA SHQ11240181SA SHQ11240182SA SHQ11240183SA SHQ11240184SA SHQ11240187SA SHQ11240189SA SHQ11240192SA SHQ11240192SR SHQ11240193SA SHQ11240196SA SHQ11240197SA SHQ11240199SA SHQ11240201SA SHQ11240202SA SHQ11240203SA SHQ11240204SA SHQ11240205SA SHQ11240206SA SHQ11240208SA SHQ11240210SA SHQ11240213SA SHQ11240215SA SHQ11240216SA SHQ11240217SA SHQ11240218SA SHQ11240219SA SHQ11240220SA SHQ11240222SA SHQ11240223SA SHQ11240225SA SHQ11240226SA SHQ11240227SA SHQ11240229SA SHQ11240230SA SHQ11240231SA SHQ11240232SA SHQ11240233SA SHQ11240234SA SHQ11240236SA SHQ11240239SA SHQ11240240SA SHQ11240241SA SHQ11240242SA SHQ11240243SA SHQ11240244SA SHQ11240245SA SHQ11240246SA SHQ11240248SA SHQ11240249SA SHQ11240252SA SHQ11240254SA SHQ11240257SA SHQ11240258SA SHQ11250278SA SHQ11250284SA SHQ11250285SA SHQ11250287SA SHQ11250288SA SHQ11250291SA SHQ11250299SA SHQ11250300SA SHQ11250301SA SHQ11250304SA SHQ11250306SA SHQ11250308SA SHQ11250309SA SHQ11250310SA SHQ11250311SA SHQ11250315SA SHQ11250316SA SHQ11250317SA SHQ11250318SA SHQ11250319SA SHQ11250320SA SHQ11250321SA SHQ11250321SR SHQ11250322SA SHQ11250324SA SHQ11250326SA SHQ11250328SA SHQ11250329SA SHQ11250332SA SHQ11250333SA SHQ11250336SA SHQ11250338SA SHQ11250339SA SHQ11250340SA SHQ11250341SA SHQ11250342SA SHQ11250343SA SHQ11250344SA SHQ11250345SA SHQ11250346SA SHQ11250347SA SHQ11250348SA SHQ11250385SA SHQ11250386SA SHQ11250389SA SHQ11250390SA SHQ11250393SA SHQ11250394SA SHQ11250395SA SHQ11250397SA SHQ11250398SA SHQ11250399SA SHQ11250401SA SHQ11250402SA SHQ11250403SA SHQ11250404SA SHQ11250405SA SHQ11250406SA SHQ11250407SA SHQ11250408SA SHQ11250409SA SHQ11250411SA SHQ11250412SA SHQ11250413SA SHQ11250414SA SHQ11250415SA SHQ11250416SA SHQ11250417SA SHQ11250418SA SHQ11250419SA SHQ11250420SA SHQ11250421SA SHQ11250422SA SHQ11250423SA SHQ11250424SA SHQ11250425SA SHQ11250426SA SHQ11250427SA SHQ11250428SA SHQ11250429SA SHQ11250430SA SHQ11250431SA SHQ11250432SA SHQ11250433SA SHQ11250434SA SHQ11250435SA SHQ11250436SA SHQ11250437SA SHQ11250439SA SHQ11250440SA SHQ11250441SA SHQ11250442SA SHQ11250443SA SHQ11250445SA SHQ11250446SA SHQ11250447SA SHQ11250450SA SHQ11250451SA SHQ11250452SA SHQ11250453SA SHQ11250454SA SHQ11250456SA SHQ11250457SA SHQ11250458SA SHQ11250459SA SHQ11250460SA SHQ11250464SA SHQ11250465SA SHQ11250467SA SHQ11250471SA SHQ11250474SA SHQ11250476SA SHQ11250478SA SHQ11250479SA SHQ11250480SA SHQ11250481SA SHQ11250482SA SHQ11250483SA SHQ11250484SA SHQ11250485SA SHQ11250486SA SHQ11250487SA SHQ11250489SA SHQ11250490SA SHQ11250491SA SHQ11250493SA SHQ11250494SA SHQ11250496SA SHQ11250497SA SHQ11250498SA SHQ11250500SA SHQ11250502SA SHQ11260508SA SHQ11260513SA SHQ11260514SA SHQ11260518SA SHQ11260519SA SHQ11260520SA SHQ11260521SA SHQ11260522SA SHQ11260523SA SHQ11260524SA SHQ11260525SA SHQ11260526SA SHQ11260527SA SHQ11260530SA SHQ11260534SA SHQ11260537SA SHQ11260544SA SHQ11260545SA SHQ11280242GA SHQ11280548SA SHQ11280549SA SHQ11280550SA SHQ11280551SA SHQ11280552SA SHQ11280553SA SHQ11280554SA SHQ11280555SA SHQ11280556SA SHQ11280557SA SHQ11280558SA SHQ11280560SA SHQ11280561SA SHQ11280562SA SHQ11280563SA SHQ11280564SA SHQ11280565SA SHQ11280566SA SHQ11280568SA SHQ11280569SA SHQ11280571SA SHQ11280572SA SHQ11280573SA SHQ11280574SA SHQ11280575SA SHQ11280576SA SHQ11280577SA SHQ11280578SA SHQ11280579SA SHQ11280581SA SHQ11280582SA SHQ11280584SA SHQ11280585SA SHQ11280586SA SHQ11280587SA SHQ11280589SA SHQ11280590SA SHQ11280593SA SHQ11280594SA SHQ11280595SA SHQ11280596SA SHQ11280597SA SHQ11280598SA SHQ11280603SA SHQ11280604SA SHQ11280605SA SHQ11280606SA SHQ11280607SA SHQ11280608SA SHQ11280610SA SHQ11280611SA SHQ11280612SA SHQ11280613SA SHQ11280614SA SHQ11280616SA SHQ11280617SA SHQ11280620SA SHQ11280623SA SHQ11280625SA SHQ11280628SA SHQ11280631SA SHQ11280632SA SHQ11280634SA SHQ11280640SA SHQ11280641SA SHQ11280642SA SHQ11280644SA SHQ11280647SA SHQ11280648SA SHQ11280649SA SHQ11280650SA SHQ11280651SA SHQ11280652SA SHQ11280653SA SHQ11280658SA SHQ11280659SA SHQ11280661SA SHQ11280662SA SHQ11280663SA SHQ11280666SA SHQ11280669SA SHQ11280670SA SHQ11280671SA SHQ11280674SA SHQ11280676SA SHQ11280677SA SHQ11280678SA SHQ11280679SA SHQ11280680SA SHQ11280681SA SHQ11280683SA SHQ11280685SA SHQ11280686SA SHQ11280687SA SHQ11280688SA SHQ11280689SA SHQ11280690SR SHQ11280691SA SHQ11280692SA SHQ11280693SA SHQ11280694SA SHQ11280696SA SHQ11280700SA SHQ11280701SA SHQ11280702SA SHQ11280703SA SHQ11280704SA SHQ11280705SA SHQ11280709SA SHQ11280710SA SHQ11280711SA SHQ11280713SA SHQ11280714SA SHQ11280715SA SHQ11280716SA SHQ11280717SA SHQ11280718SA SHQ11280721SA SHQ11280723SA SHQ11280724SA SHQ11280725SA SHQ11280726SA SHQ11280728SA SHQ11280729SA SHQ11280730SA SHQ11280732SA SHQ11280733SA SHQ11280734SA SHQ11280737SA SHQ11280738SA SHQ11280739SA SHQ11280740SA SHQ11280748SA SHQ11280749SA SHQ11280750SA SHQ11280751SA SHQ11280754SA SHQ11280755SA SHQ11280758SA SHQ11280761SA SHQ11280765SA SHQ11280768SA SHQ11280771SA SHQ11280773SA SHQ11280775SA SHQ11280778SA SHQ11280785SA SHQ11280787SA SHQ11280789SA SHQ11280790SA SHQ11280793SA SHQ11280795SA SHQ11280799SA SHQ11280800SA SHQ11280801SA SHQ11280802SA SHQ11280803SA SHQ11280805SA SHQ11280808SA SHQ11280811SA SHQ11280814SA SHQ11280816SA SHQ11280817SA SHQ11280818SA SHQ11280819SA SHQ11280820SA SHQ11280823SA SHQ11280824SA SHQ11280825SA SHQ11290343GA SHQ11290344GA SHQ11290345GA SHQ11290346GA SHQ11290826SA SHQ11290827SA SHQ11290836SA SHQ11290837SA SHQ11290843SA SHQ11290847SA SHQ11290852SA SHQ11290853SA SHQ11290854SA SHQ11290855SA SHQ11290856SA SHQ11290858SA SHQ11290861SA SHQ11290866SA SHQ11290867SA SHQ11290868SA SHQ11290869SA SHQ11290870SA SHQ11290871SA SHQ11290872SA SHQ11290873SA SHQ11290874SA SHQ11290875SA SHQ11290876SA SHQ11290877SA SHQ11290878SA SHQ11290879SA SHQ11290880SA SHQ11290881SA SHQ11290882SA SHQ11290883SA SHQ11290885SA SHQ11290886SA SHQ11290887SA SHQ11290888SA SHQ11290889SA SHQ11290890SA SHQ11290892SA SHQ11290893SA SHQ11290894SA SHQ11290895SA SHQ11290896SA SHQ11290897SA SHQ11290898SA SHQ11290899SA SHQ11290900SA SHQ11290901SA SHQ11290902SA SHQ11290903SA SHQ11290905SA SHQ11290906SA SHQ11290907SA SHQ11290908SA SHQ11290909SA SHQ11290910SA SHQ11290911SA SHQ11290912SA SHQ11290913SA SHQ11290914SA SHQ11290915SA SHQ11290916SA SHQ11290917SA SHQ11290918SA SHQ11290919SA SHQ11290920SA SHQ11290921SA SHQ11290922SA SHQ11290923SA SHQ11290924SA SHQ11290925SA SHQ11290926SA SHQ11290927SA SHQ11290928SA SHQ11290929SA SHQ11290930SA SHQ11290931SA SHQ11290932SA SHQ11290933SA SHQ11290934SA SHQ11290935SA SHQ11290936SA SHQ11290937SA SHQ11290939SA SHQ11290940SA SHQ11290941SA SHQ11290942SA SHQ11290943SA SHQ11290944SA SHQ11290945SA SHQ11290947SA SHQ11290949SA SHQ11290950SA SHQ11290952SA SHQ11290953SA SHQ11290954SA SHQ11290955SA SHQ11290956SA SHQ11290957SA SHQ11290959SA SHQ11290960SA SHQ11290961SA SHQ11290963SA SHQ11290964SA SHQ11290968SA SHQ11290969SA SHQ11290970SA SHQ11290971SA SHQ11290973SA SHQ11290974SA SHQ11290975SA SHQ11290976SA SHQ11290977SA SHQ11290979SA SHQ11290980SA SHQ11290981SA SHQ11290982SA SHQ11290983SA SHQ11290984SA SHQ11290986SA SHQ11290987SA SHQ11290988SA SHQ11290990SA SHQ11290991SA SHQ11290992SA SHQ11290993SA SHQ11290994SA SHQ11290995SA SHQ11290997SA SHQ11290999SA SHQ11291000SA SHQ11291002SA SHQ11291004SA SHQ11291005SA SHQ11291006SA SHQ11291007SA SHQ11291008SA SHQ11291012SA SHQ11291013SA SHQ11291014SA SHQ11291017SA SHQ11291018SA SHQ11291019SA SHQ11291022SA SHQ11291026SA SHQ11291027SA SHQ11291030SA SHQ11291031SA SHQ11291032SA SHQ11291033SA SHQ11291034SA SHQ11291035SA SHQ11291040SA SHQ11291041SA SHQ11291042SA SHQ11291043SA SHQ11291045SA SHQ11291046SA SHQ11291047SA SHQ11291048SA SHQ11291050SA SHQ11291051SA SHQ11291052SA SHQ11291053SA SHQ11291054SA SHQ11291058SA SHQ11291059SA SHQ11291062SA SHQ11291064SA SHQ11291065SA SHQ11291066SA SHQ11291067SA SHQ11291068SA SHQ11291071SA SHQ11291072SA SHQ11291073SA SHQ11291074SA SHQ11291077SA SHQ11291077SR SHQ11291078SA SHQ11291079SA SHQ11291080SA SHQ11291081SA SHQ11291082SA SHQ11291083SA SHQ11291084SA SHQ11291085SA SHQ11291087SA SHQ11291088SA SHQ11291089SA SHQ11291090SA SHQ11291091SA SHQ11291092SA SHQ11291093SA SHQ11291094SA SHQ11291095SA SHQ11300389GA SHQ11300406GA SHQ11300410GA SHQ11300411GA SHQ11300414GA SHQ11300416GA SHQ11300417GA SHQ11301096SA SHQ11301097SA SHQ11301098SA SHQ11301099SA SHQ11301100SA SHQ11301101SA SHQ11301102SA SHQ11301103SA SHQ11301104SA SHQ11301105SA SHQ11301106SA SHQ11301107SA SHQ11301108SA SHQ11301109SA SHQ11301110SA SHQ11301111SA SHQ11301112SA SHQ11301113SA SHQ11301114SA SHQ11301115SA SHQ11301116SA SHQ11301117SA SHQ11301118SA SHQ11301120SA SHQ11301122SA SHQ11301123SA SHQ11301124SA SHQ11301125SA SHQ11301126SA SHQ11301127SA SHQ11301129SA SHQ11301130SA SHQ11301133SA SHQ11301134SA SHQ11301135SA SHQ11301144SA SHQ11301145SA SHQ11301150SA SHQ11301155SA SHQ11301157SR SHQ11301158SA SHQ11301159SR SHQ11301160SA SHQ11301162SA SHQ11301163SA SHQ11301165SA SHQ11301166SA SHQ11301169SA SHQ11301170SA SHQ11301173SR SHQ11301174SA SHQ11301175SA SHQ11301177SA SHQ11301178SA SHQ11301179SA SHQ11301181SA SHQ11301182SA SHQ11301183SA SHQ11301189SA SHQ11301190SA SHQ11301194SA SHQ11301195SA SHQ11301197SA SHQ11301198SA SHQ11301201SA SHQ11301202SA SHQ11301204SA SHQ11301206SA SHQ11301208SA SHQ11301211SA SHQ11301212SA SHQ11301213SA SHQ11301214SA SHQ11301215SA SHQ11301217SA SHQ11301219SA SHQ11301220SA SHQ11301222SA SHQ11301224SA SHQ11301225SA SHQ11301226SA SHQ11301227SA SHQ11301228SA SHQ11301229SA SHQ11301230SA SHQ11301231SA SHQ11301234SA SHQ11301235SA SHQ11301236SA SHQ11301237SA SHQ11301238SA SHQ11301241SA SHQ11301242SA SHQ11301243SA SHQ11301244SA SHQ11301245SA SHQ11301246SA SHQ11301247SA SHQ11301248SA SHQ11301249SA SHQ11301250SA SHQ11301251SA SHQ11301254SA SHQ11301255SA SHQ11301256SA SHQ11301257SA SHQ11301259SA SHQ11301260SA SHQ11301261SA SHQ11301262SA SHQ11301263SA SHQ11301264SA SHQ11301265SA SHQ11301266SA SHQ11301267SA SHQ11301268SA SHQ11301269SA SHQ11301270SA SHQ11301271SA SHQ11301272SA SHQ11301273SA SHQ11301274SA SHQ11301275SA SHQ11301276SA SHQ11301277SA SHQ11301278SA SHQ11301279SA SHQ11301283SA SHQ11301285SA SHQ11301288SA SHQ11301289SA SHQ11301291SA SHQ11301295SA SHQ11301296SA SHQ11301297SA SHQ11301298SA SHQ11301299SA SHQ11301300SA SHQ11301301SA SHQ11301303SA SHQ11301304SA SHQ11301305SA SHQ11301306SA SHQ11301307SA SHQ11301309SA SHQ11301310SA SHQ11301311SA SHQ11301313SA SHQ11301314SA SHQ11301315SA SHQ11301316SA SHQ11301317SA SHQ11301318SA SHQ11301319SA SHQ11301320SA SHQ11301321SA SHQ11301322SA SHQ11301323SA SHQ11301324SA SHQ11301325SA SHQ11301326SA SHQ11301328SA SHQ11301330SA SHQ11301331SA SHQ11301332SA SHQ11301333SA SHQ11301334SA SHQ11301335SA SHQ11301336SA SHQ11301337SA SHQ11301338SA SHQ11301339SA SHQ11301340SA SHQ11301341SA SHQ11301342SA SHQ11301343SA SHQ11301344SA SHQ11301345SA SHQ11301346SA SHQ11301347SA SHQ11301348SA SHQ11301349SA SHQ11301351SA SHQ11301352SA SHQ11301353SA SHQ11301354SA SHQ11301355SA SHQ11301356SA SHQ11301357SA SHQ11301358SA SHQ11301359SA SHQ11301360SA SHQ11301361SA SHQ11301362SA SHQ11301363SA SHQ11301364SA SHQ11301365SA SHQ11301366SA SHQ11301367SA SHQ11301368SA SHQ11301369SA SHQ11311370SA SHQ11311371SA SHQ11311373SA SHQ11311374SA SHQ11311375SA SHQ11311376SA SHQ11311377SA SHQ11311378SA SHQ11311379SA SHQ11311380SA SHQ11311381SA SHQ11311382SA SHQ11311386SA SHQ11311387SA SHQ11311388SA SHQ11311390SA SHQ11311391SA SHQ11311392SA SHQ11311393SA SHQ11311394SA SHQ11311395SA SHQ11311396SA SHQ11311397SA SHQ11311398SA SHQ11311399SA SHQ11311400SA SHQ11311401SA SHQ11311402SA SHQ11311404SA SHQ11311405SA SHQ11311406SA SHQ11311407SA SHQ11311408SA SHQ11311411SA SHQ11311412SA SHQ11311413SA SHQ11311414SA SHQ11311416SA SHQ11311417SA SHQ11311418SA SHQ11311419SA SHQ11311420SA SHQ11311423SA SHQ11311430SA SHQ11311435SA SHQ11311436SA SHQ11311438SA SHQ11311441SA SHQ11311443SA SHQ11311444SA SHQ11311445SA SHQ11311446SA SHQ11311449SA SHQ11311450SA SHQ11311452SA SHQ11311453SA SHQ11311454SA SHQ11311456SA SHQ11311457SA SHQ11311458SA SHQ11311459SA SHQ11311460SA SHQ11311461SA SHQ11311463SA SHQ11311464SA SHQ11311465SA SHQ11311466SA SHQ11311467SA SHQ11311468SA SHQ11311469SA SHQ11311470SA SHQ11311471SA SHQ11311472SA SHQ11311473SA SHQ11311474SA SHQ11311475SA SHQ11311476SA SHQ11311477SA SHQ11311478SA SHQ11311479SA SHQ11311480SA SHQ11311481SA SHQ11311482SA SHQ11311483SA SHQ11311484SA SHQ11311485SA SHQ11311486SA SHQ11311487SA SHQ11311488SA SHQ11311489SA SHQ11311490SA SHQ11311491SA SHQ11311492SA SHQ11311493SA SHQ11311494SA SHQ11311495SA SHQ11311496SA SHQ11311497SA SHQ11311498SA SHQ11311499SA SHQ11311500SA SHQ11311501SA SHQ11311502SA SHQ11311503SA SHQ11311504SA SHQ11311506SA SHQ11311507SA SHQ11311508SA SHQ11311509SA SHQ11311510SA SHQ11311511SA SHQ11311512SA SHQ11311513SA SHQ11311514SA SHQ11311515SA SHQ11311516SA SHQ11311517SA SHQ11311518SA SHQ11311519SA SHQ11311520SA SHQ11311521SA SHQ11311522SA SHQ11311523SA SHQ11311524SA SHQ11311525SA SHQ11311526SA SHQ11311527SA SHQ11311528SA SHQ11311529SA SHQ11311531SA SHQ11311532SA SHQ11311533SA SHQ11311535SA SHQ11311536SA SHQ11311537SA SHQ11311538SA SHQ11311539SA SHQ11311540SA SHQ11311541SA SHQ11311542SA SHQ11311543SA SHQ11311544SA SHQ11311545SA SHQ11311546SA SHQ11311547SA SHQ11311548SA SHQ11311549SA SHQ11311550SA SHQ11311551SA SHQ11311552SA SHQ11311553SA SHQ11311554SA SHQ11311555SA SHQ11311556SA SHQ11311557SA SHQ11311558SA SHQ11311559SA SHQ11311561SA SHQ11311562SA SHQ11311564SA SHQ11311565SA SHQ11311567SA SHQ11311568SA SHQ11311569SA SHQ11311570SA SHQ11311571SA SHQ11311572SA SHQ11311573SA SHQ11311574SA SHQ11311575SA SHQ11311576SA SHQ11311577SA SHQ11311578SA SHQ11311579SA SHQ11311580SA SHQ11311581SA SHQ11311582SA SHQ11311583SA SHQ11311585SA SHQ11311587SA SHQ11311588SA SHQ11311589SA SHQ11311591SA SHQ11311592SA SHQ11311593SA SHQ11311594SA SHQ11311597SA SHQ11311598SA SHQ11311599SA SHQ11311600SA SHQ11311602SA SHQ11311603SA SHQ11311604SA SHQ11311610SA SHQ11311614SA SHQ11311615SA SHQ11311616SA SHQ11311617SA SHQ11311617SR SHQ11311619SA SHQ11311620SA SHQ11311621SA SHQ11311622SA SHQ11311624SA SHQ11311625SA SHQ11311626SA SHQ11311627SA SHQ11311629SA SHQ11311630SA SHQ11311631SA SHQ11311633SA SHQ11311642SA SHQ11311645SA SHQ11311647SA SHQ11311648SA SHQ11320558GA SHQ11320577GA SHQ11320578GA SHQ11320579GA SHQ11320580GA SHQ11320581GA SHQ11320584GA SHQ11320586GA SHQ11320587GA SHQ11320588GA SHQ11321650SA SHQ11321651SA SHQ11321652SA SHQ11321654SA SHQ11321657SA SHQ11321658SA SHQ11321662SA SHQ11321664SA SHQ11321665SA SHQ11321668SA SHQ11321669SA SHQ11321670SA SHQ11321673SA SHQ11321674SA SHQ11321675SA SHQ11321676SA SHQ11321677SA SHQ11321678SA SHQ11321679SA SHQ11321680SA SHQ11321681SA SHQ11321682SA SHQ11321683SA SHQ11321684SA SHQ11321685SA SHQ11321686SA SHQ11321687SA SHQ11321688SA SHQ11321689SA SHQ11321690SA SHQ11321691SA SHQ11321692SA SHQ11321693SA SHQ11321694SA SHQ11321695SA SHQ11321696SA SHQ11321698SA SHQ11321700SA SHQ11321701SA SHQ11321702SA SHQ11321703SA SHQ11321705SA SHQ11321708SA SHQ11321712SA SHQ11321713SA SHQ11321714SA SHQ11321715SA SHQ11321716SA SHQ11321719SA SHQ11321720SA SHQ11321721SA SHQ11321722SA SHQ11321725SA SHQ11321726SA SHQ11321727SA SHQ11321728SA SHQ11321729SA SHQ11321730SA SHQ11321731SA SHQ11321732SA SHQ11321733SA SHQ11321734SA SHQ11321735SA SHQ11321736SA SHQ11321737SA SHQ11321739SA SHQ11321740SA SHQ11321741SA SHQ11321742SA SHQ11321743SA SHQ11321744SA SHQ11321745SA SHQ11321746SA SHQ11321747SA SHQ11321748SA SHQ11321750SA SHQ11321752SA SHQ11321754SA SHQ11321755SA SHQ11321757SA SHQ11321758SA SHQ11321761SA SHQ11321763SA SHQ11321765SA SHQ11321767SA SHQ11321768SA SHQ11321769SA SHQ11321770SA SHQ11321772SA SHQ11321773SA SHQ11321775SA SHQ11321776SA SHQ11321777SA SHQ11321778SA SHQ11321779SA SHQ11321780SA SHQ11321781SA SHQ11321782SA SHQ11321783SA SHQ11321784SA SHQ11321785SA SHQ11321786SA SHQ11321787SA SHQ11321788SA SHQ11321789SA SHQ11321790SA SHQ11321791SA SHQ11321792SA SHQ11321793SA SHQ11321794SA SHQ11321795SA SHQ11321796SA SHQ11321797SA SHQ11321798SA SHQ11321799SA SHQ11321800SA SHQ11321801SA SHQ11321803SA SHQ11321804SA SHQ11321805SA SHQ11321806SA SHQ11321808SA SHQ11321809SA SHQ11321811SA SHQ11321812SA SHQ11321813SA SHQ11321814SA SHQ11321816SA SHQ11321818SA SHQ11321819SA SHQ11321820SA SHQ11321821SA SHQ11321822SA SHQ11321823SA SHQ11321824SA SHQ11321825SA SHQ11321826SA SHQ11321827SA SHQ11321828SA SHQ11321829SA SHQ11321830SA SHQ11321831SA SHQ11321832SA SHQ11321833SA SHQ11321834SA SHQ11321835SA SHQ11321836SA SHQ11321837SA SHQ11321838SA SHQ11321839SA SHQ11321840SA SHQ11321841SA SHQ11321843SA SHQ11321844SA SHQ11321845SA SHQ11321846SA SHQ11321848SA SHQ11321850SA SHQ11321851SA SHQ11321852SA SHQ11321853SA SHQ11321855SA SHQ11321856SA SHQ11321858SA SHQ11321860SA SHQ11321863SA SHQ11321865SA SHQ11321869SA SHQ11321870SA SHQ11321871SA SHQ11321873SA SHQ11321876SA SHQ11321878SA SHQ11321881SA SHQ11331885SA SHQ11331886SA SHQ11331887SA SHQ11331888SA SHQ11331889SA SHQ11331890SA SHQ11331891SA SHQ11331892SA SHQ11331893SA SHQ11331894SA SHQ11331895SA SHQ11331896SA SHQ11331897SA SHQ11331898SA SHQ11331899SA SHQ11331900SA SHQ11331901SA SHQ11331902SA SHQ11331903SA SHQ11331904SA SHQ11331905SA SHQ11331906SA SHQ11331907SA SHQ11331908SA SHQ11331909SA SHQ11331910SA SHQ11331911SA SHQ11331912SA SHQ11331913SA SHQ11331914SA SHQ11331927SA SHQ11331928SA SHQ11331930SA SHQ11331931SA SHQ11331932SA SHQ11331933SA SHQ11331934SA SHQ11331937SA SHQ11331938SA SHQ11331940SA SHQ11331941SA SHQ11331942SA SHQ11331944SA SHQ11331945SA SHQ11331946SA SHQ11331949SA SHQ11331950SA SHQ11331951SA SHQ11331952SA SHQ11331953SA SHQ11331955SA SHQ11331956SA SHQ11331957SA SHQ11331958SA SHQ11331959SA SHQ11331960SA SHQ11331961SA SHQ11331962SA SHQ11331963SA SHQ11331964SA SHQ11331965SA SHQ11331967SA SHQ11331968SA SHQ11331969SA SHQ11331970SA SHQ11331971SA SHQ11331972SA SHQ11331973SA SHQ11331974SA SHQ11331975SA SHQ11331976SA SHQ11331977SA SHQ11331979SA SHQ11331980SA SHQ11331982SA SHQ11331983SA SHQ11331985SA SHQ11331987SA SHQ11331988SA SHQ11331989SA SHQ11331991SA SHQ11331992SA SHQ11331993SA SHQ11331994SA SHQ11331995SA SHQ11331997SA SHQ11331998SA SHQ11331999SA SHQ11332000SA SHQ11332001SA SHQ11332002SA SHQ11332003SA SHQ11332004SA SHQ11332005SA SHQ11332006SA SHQ11332007SA SHQ11332008SA SHQ11332010SA SHQ11332012SA SHQ11332013SA SHQ11332014SA SHQ11332015SA SHQ11332016SA SHQ11332017SA SHQ11332018SA SHQ11332019SA SHQ11332022SA SHQ11332023SA SHQ11332025SA SHQ11332026SA SHQ11332027SA SHQ11332028SA SHQ11332029SA SHQ11332031SA SHQ11332032SA SHQ11332033SA SHQ11332034SA SHQ11332036SA SHQ11332038SA SHQ11332039SA SHQ11332040SA SHQ11332041SA SHQ11332042SA SHQ11332043SA SHQ11332044SA SHQ11332045SA SHQ11332046SA SHQ11332047SA SHQ11332048SA SHQ11332052SA SHQ11332055SA SHQ11332056SA SHQ11332058SA SHQ11332059SA SHQ11332060SA SHQ11332061SA SHQ11332062SA SHQ11332063SA SHQ11332064SA SHQ11332065SA SHQ11332066SA SHQ11332067SA SHQ11332068SA SHQ11332069SA SHQ11332070SA SHQ11332071SA SHQ11332072SA SHQ11332073SA SHQ11332074SA SHQ11332075SA SHQ11332076SA SHQ11332077SA SHQ11332078SA SHQ11332079SA SHQ11332080SA SHQ11332081SA SHQ11332082SA SHQ11332083SA SHQ11332084SA SHQ11332085SA SHQ11332086SA SHQ11332087SA SHQ11332088SA SHQ11332089SA SHQ11332090SA SHQ11332091SA SHQ11332092SA SHQ11332093SA SHQ11332094SA SHQ11332095SA SHQ11332096SA SHQ11332097SA SHQ11332098SA SHQ11332099SA SHQ11332100SA SHQ11332102SA SHQ11332103SA SHQ11332104SA SHQ11332105SA SHQ11332107SA SHQ11332108SA SHQ11332109SA SHQ11332111SA SHQ11332112SA SHQ11332114SA SHQ11342116SA SHQ11342118SA SHQ11342119SA SHQ11342120SA SHQ11342121SA SHQ11342122SA SHQ11342125SA SHQ11342126SA SHQ11342129SA SHQ11342130SA SHQ11342131SA SHQ11342132SA SHQ11342135SA SHQ11342137SA SHQ11342138SA SHQ11342140SA SHQ11342142SA SHQ11342143SA SHQ11342144SA SHQ11342145SA SHQ11342146SA SHQ11342147SA SHQ11342148SA SHQ11342150SA SHQ11342151SA SHQ11342153SA SHQ11342154SA SHQ11342156SA SHQ11342157SA SHQ11342158SA SHQ11342159SA SHQ11342160SA SHQ11342161SA SHQ11342163SA SHQ11342164SA SHQ11342165SA SHQ11342168SA SHQ11342169SA SHQ11342171SA SHQ11342172SA SHQ11342173SA SHQ11342175SA SHQ11342176SA SHQ11342177SA SHQ11342178SA SHQ11342179SA SHQ11342180SA SHQ11342181SA SHQ11342183SA SHQ11342184SA SHQ11342186SA SHQ11342188SA SHQ11342189SA SHQ11342190SA SHQ11342191SA SHQ11342192SA SHQ11342193SA SHQ11342197SA SHQ11342198SA SHQ11342199SA SHQ11342200SA SHQ11342201SA SHQ11342202SA SHQ11342203SA SHQ11342204SA SHQ11342205SA SHQ11342206SA SHQ11342207SA SHQ11342208SA SHQ11342209SA SHQ11342210SA SHQ11342211SA SHQ11342212SA SHQ11342214SA SHQ11342216SA SHQ11342217SA SHQ11342218SA SHQ11342219SA SHQ11342220SA SHQ11342223SA SHQ11342224SA SHQ11342226SA SHQ11342227SA SHQ11342228SA SHQ11342229SA SHQ11342230SA SHQ11342231SA SHQ11342232SA SHQ11342233SA SHQ11342234SA SHQ11342235SA SHQ11342236SA SHQ11342237SA SHQ11342238SA SHQ11342239SA SHQ11342240SA SHQ11342241SA SHQ11342242SA SHQ11342243SA SHQ11342244SA SHQ11342245SA SHQ11342246SA SHQ11342247SA SHQ11342248SA SHQ11342250SA SHQ11342251SA SHQ11342252SA SHQ11342253SA SHQ11342255SA SHQ11342256SA SHQ11342257SA SHQ11342258SA SHQ11342259SA SHQ11342260SA SHQ11342261SA SHQ11342270SA SHQ11342276SA SHQ11342277SA SHQ11342278SA SHQ11342279SA SHQ11342280SA SHQ11342281SA SHQ11342282SA SHQ11342283SA SHQ11342284SA SHQ11342287SA SHQ11342288SA SHQ11342289SA SHQ11342290SA SHQ11342292SA SHQ11342293SA SHQ11342294SA SHQ11342295SA SHQ11342296SA SHQ11342297SA SHQ11342298SA SHQ11342300SA SHQ11342301SA SHQ11342302SA SHQ11342303SA SHQ11342304SA SHQ11342305SA SHQ11342306SA SHQ11342308SA SHQ11342310SA SHQ11342311SA SHQ11342313SA SHQ11342314SA SHQ11342320SA SHQ11342331SA SHQ11352332SA SHQ11352334SA SHQ11352335SA SHQ11352336SA SHQ11352337SA SHQ11352338SA SHQ11352339SA SHQ11352340SA SHQ11352341SA SHQ11352342SA SHQ11352343SA SHQ11352344SA SHQ11352345SA SHQ11352346SA SHQ11352347SA SHQ11352348SA SHQ11352349SA SHQ11352350SA SHQ11352352SA SHQ11352353SA SHQ11352355SA SHQ11352356SA SHQ11352357SA SHQ11352358SA SHQ11352358SR SHQ11352359SA SHQ11352359SR SHQ11352360SA SHQ11352361SA SHQ11352362SA SHQ11352363SA SHQ11352364SA SHQ11352365SA SHQ11352367SA SHQ11352369SA SHQ11352370SA SHQ11352371SA SHQ11352372SA SHQ11352373SA SHQ11352374SA SHQ11352375SA SHQ11352376SA SHQ11352379SA SHQ11352380SA SHQ11352381SA SHQ11352382SA SHQ11352383SA SHQ11352386SA SHQ11352390SA SHQ11352391SA SHQ11352392SA SHQ11352394SA SHQ11352395SA SHQ11352396SA SHQ11352399SA SHQ11352401SA SHQ11352403SA SHQ11352404SA SHQ11352406SA SHQ11352410SA SHQ11352411SA SHQ11352413SA SHQ11352414SA SHQ11352415SA SHQ11352416SA SHQ11352417SA SHQ11352418SA SHQ11352419SA SHQ11352420SA SHQ11352421SA SHQ11352422SA SHQ11352423SA SHQ11352424SA SHQ11352425SA SHQ11352426SA SHQ11352427SA SHQ11352428SA SHQ11352429SA SHQ11352430SA SHQ11352431SA SHQ11352432SA SHQ11352433SA SHQ11352434SA SHQ11352435SA SHQ11352436SA SHQ11352437SA SHQ11352438SA SHQ11352439SA SHQ11352440SA SHQ11352441SA SHQ11352442SA SHQ11352443SA SHQ11352444SA SHQ11352445SA SHQ11352446SA SHQ11352447SA SHQ11352448SA SHQ11352449SA SHQ11352450SA SHQ11352451SA SHQ11352452SA SHQ11352453SA SHQ11352454SA SHQ11352455SA SHQ11352456SA SHQ11352457SA SHQ11352458SA SHQ11352459SA SHQ11352460SA SHQ11352461SA SHQ11352462SA SHQ11352463SA SHQ11352464SA SHQ11352465SA SHQ11352466SA SHQ11352467SA SHQ11352468SA SHQ11352470SA SHQ11352471SA SHQ11352472SA SHQ11352473SA SHQ11352475SA SHQ11352476SA SHQ11352477SA SHQ11352478SA SHQ11352480SA SHQ11352481SA SHQ11352482SA SHQ11352484SA SHQ11352485SA SHQ11352486SA SHQ11352488SA SHQ11352490SA SHQ11352491SA SHQ11352493SA SHQ11352494SA SHQ11352497SA SHQ11352498SA SHQ11352499SA SHQ11352504SA SHQ11352506SA SHQ11352507SA SHQ11352509SA SHQ11352511SA SHQ11352513SA SHQ11352514SA SHQ11362520SA SHQ11362521SA SHQ11362528SA SHQ11362529SA SHQ11362531SA SHQ11362532SA SHQ11362533SA SHQ11362534SA SHQ11362535SA SHQ11362536SA SHQ11362537SA SHQ11362538SA SHQ11362539SA SHQ11362540SA SHQ11362541SA SHQ11362542SA SHQ11362543SA SHQ11362544SA SHQ11362545SA SHQ11362546SA SHQ11362547SA SHQ11362548SA SHQ11362549SA SHQ11362550SA SHQ11362551SA SHQ11362552SA SHQ11362553SA SHQ11362554SA SHQ11362555SA SHQ11362556SA SHQ11362557SA SHQ11362558SA SHQ11362559SA SHQ11362560SA SHQ11362561SA SHQ11362562SA SHQ11362563SA SHQ11362564SA SHQ11362565SA SHQ11362566SA SHQ11362567SA SHQ11362568SA SHQ11362569SA SHQ11362570SA SHQ11362571SA SHQ11362572SA SHQ11362573SA SHQ11362574SA SHQ11362575SA SHQ11362576SA SHQ11362581SA SHQ11362582SA SHQ11362584SA SHQ11362585SA SHQ11362587SA SHQ11362588SA SHQ11362589SA SHQ11362590SA SHQ11362591SA SHQ11362592SA SHQ11362593SA SHQ11362594SA SHQ11362595SA SHQ11362596SA SHQ11362597SA SHQ11362598SA SHQ11362599SA SHQ11362600SA SHQ11362601SA SHQ11362602SA SHQ11362603SA SHQ11362604SA SHQ11362605SA SHQ11362606SA SHQ11362607SA SHQ11362617SA SHQ11362622SA SHQ11362624SA SHQ11362625SA SHQ11362626SA SHQ11362628SA SHQ11362629SA SHQ11362632SA SHQ11362633SA SHQ11362635SA SHQ11362637SA SHQ11362639SA SHQ11362640SA SHQ11362641SA SHQ11362642SA SHQ11362644SA SHQ11362645SA SHQ11362646SA SHQ11362648SA SHQ11362649SA SHQ11362650SA SHQ11362651SA SHQ11362653SA SHQ11362657SA SHQ11362658SA SHQ11362660SA SHQ11362662SA SHQ11362663SA SHQ11362668SA SHQ11362669SA SHQ11362670SA SHQ11362672SA SHQ11362674SA SHQ11362675SA SHQ11362680SA SHQ11362681SA SHQ11362682SA SHQ11362683SA SHQ11362684SA SHQ11362685SA SHQ11362686SA SHQ11362687SA SHQ11362688SA SHQ11362689SA SHQ11362690SA SHQ11362691SA SHQ11362693SA SHQ11362695SA SHQ11362697SA SHQ11362702SA SHQ11362703SA SHQ11372707SA SHQ11372708SA SHQ11372709SA SHQ11372710SA SHQ11372711SA SHQ11372712SA SHQ11372713SA SHQ11372714SA SHQ11372715SA SHQ11372716SA SHQ11372717SA SHQ11372718SA SHQ11372719SA SHQ11372720SA SHQ11372721SA SHQ11372722SA SHQ11372723SA SHQ11372724SA SHQ11372725SA SHQ11372726SA SHQ11372727SA SHQ11372728SA SHQ11372729SA SHQ11372730SA SHQ11372731SA SHQ11372732SA SHQ11372733SA SHQ11372734SA SHQ11372736SA SHQ11372737SA SHQ11372738SA SHQ11372739SA SHQ11372740SA SHQ11372741SA SHQ11372742SA SHQ11372743SA SHQ11372744SA SHQ11372745SA SHQ11372746SA SHQ11372747SA SHQ11372749SA SHQ11372750SA SHQ11372751SA SHQ11372752SA SHQ11372753SA SHQ11372754SA SHQ11372755SA SHQ11372756SA SHQ11372757SA SHQ11372758SA SHQ11372759SA SHQ11372760SA SHQ11372761SA SHQ11372762SA SHQ11372763SA SHQ11372764SA SHQ11372765SA SHQ11372766SA SHQ11372767SA SHQ11372768SA SHQ11372769SA SHQ11372770SA SHQ11372773SA SHQ11372774SA SHQ11372776SA SHQ11372777SA SHQ11372778SA SHQ11372780SA SHQ11372781SA SHQ11372783SA SHQ11372788SA SHQ11372789SA SHQ11372792SA SHQ11372793SA SHQ11372794SA SHQ11372795SA SHQ11372799SA SHQ11372800SA SHQ11372801SA SHQ11372804SA SHQ11372807SA SHQ11372808SA SHQ11372809SA SHQ11372810SA SHQ11372812SA SHQ11372813SA SHQ11372814SA SHQ11372815SA SHQ11372816SA SHQ11372817SA SHQ11372818SA SHQ11372819SA SHQ11372821SA SHQ11372822SA SHQ11372823SA SHQ11372824SA SHQ11372825SA SHQ11372826SA SHQ11372827SA SHQ11372828SA SHQ11372831SA SHQ11372832SA SHQ11372833SA SHQ11372834SA SHQ11372835SA SHQ11372836SA SHQ11372839SA SHQ11372840SA SHQ11372841SA SHQ11372842SA SHQ11372843SA SHQ11372844SA SHQ11372845SA SHQ11372846SA SHQ11372847SA SHQ11372848SA SHQ11372849SA SHQ11372850SA SHQ11372851SA SHQ11372852SA SHQ11372853SA SHQ11372854SA SHQ11372855SA SHQ11372856SA SHQ11372857SA SHQ11372858SA SHQ11372859SA SHQ11372860SA SHQ11372861SA SHQ11372862SA SHQ11372863SA SHQ11372864SA SHQ11372865SA SHQ11372866SA SHQ11372867SA SHQ11372868SA SHQ11372869SA SHQ11372870SA SHQ11372871SA SHQ11372872SA SHQ11372873SA SHQ11372874SA SHQ11372875SA SHQ11372876SA SHQ11372877SA SHQ11372878SA SHQ11372879SA SHQ11372880SA SHQ11372881SA SHQ11372882SA SHQ11372883SA SHQ11372884SA SHQ11372885SA SHQ11372886SA SHQ11372887SA SHQ11372888SA SHQ11372889SA SHQ11372897SA SHQ11372898SA SHQ11372899SA SHQ11372900SA SHQ11372901SA SHQ11372902SA SHQ11372903SA SHQ11372904SA SHQ11372905SA SHQ11372906SA SHQ11372908SA SHQ11372909SA SHQ11372910SA SHQ11372911SA SHQ11372912SA SHQ11372913SA SHQ11372914SA SHQ11372915SA SHQ11372916SA SHQ11372917SA SHQ11372918SA SHQ11372919SA SHQ11372920SA SHQ11372921SA SHQ11372922SA SHQ11372923SA SHQ11372924SA SHQ11372925SA SHQ11372926SA SHQ11372927SA SHQ11372928SA SHQ11372929SA SHQ11372930SA SHQ11372931SA SHQ11372932SA SHQ11372933SA SHQ11372934SA SHQ11372935SA SHQ11372936SA SHQ11372937SA SHQ11372938SA SHQ11372939SA SHQ11372940SA SHQ11372941SA SHQ11372942SA SHQ11372943SA SHQ11372944SA SHQ11372946SA SHQ11372947SA SHQ11372948SA SHQ11372952SA SHQ11372956SA SHQ11372960SA SHQ11372961SA SHQ11372964SA SHQ11372966SA SHQ11372967SA SHQ11372968SA SHQ11372969SA SHQ11372970SA SHQ11372971SA SHQ11372972SA SHQ11372973SA SHQ11372974SA SHQ11372975SA SHQ11372976SA SHQ11382977SA SHQ11382978SA SHQ11382979SA SHQ11382980SA SHQ11382995SA SHQ11383002SA SHQ11383007SA SHQ11383012SA SHQ11383013SA SHQ11383014SA SHQ11383016SA SHQ11383018SA SHQ11383019SA SHQ11383020SA SHQ11383021SA SHQ11383022SA SHQ11383023SA SHQ11383024SA SHQ11383026SA SHQ11383027SA SHQ11383028SA SHQ11383030SA SHQ11383031SA SHQ11383033SA SHQ11383034SA SHQ11383035SA SHQ11383037SA SHQ11383039SA SHQ11383041SA SHQ11383042SA SHQ11383043SA SHQ11383045SA SHQ11383047SA SHQ11383048SA SHQ11383049SA SHQ11383050SA SHQ11383051SA SHQ11383052SA SHQ11383053SA SHQ11383055SA SHQ11383056SA SHQ11383057SA SHQ11383058SA SHQ11383059SA SHQ11383060SA SHQ11383061SA SHQ11383062SA SHQ11383063SA SHQ11383064SA SHQ11383065SA SHQ11383066SA SHQ11383068SA SHQ11383069SA SHQ11383070SA SHQ11383071SA SHQ11383072SA SHQ11383073SA SHQ11383074SA SHQ11383075SA SHQ11383076SA SHQ11383077SA SHQ11383078SA SHQ11383079SA SHQ11383080SA SHQ11383081SA SHQ11391099GA SHQ11391100GA SHQ11391101GA SHQ11391102GA SHQ11391103GA SHQ11391104GA SHQ11391105GA SHQ11391106GA SHQ11391107GA SHQ11391108GA SHQ11391109GA SHQ11391110GA SHQ11391111GA SHQ11391112GA SHQ11391113GA SHQ11391114GA SHQ11391115GA SHQ11391116GA SHQ11391117GA SHQ11391118GA SHQ11391119GA SHQ11391120GA SHQ11391121GA SHQ11391122GA SHQ11391123GA SHQ11391124GA SHQ11391125GA SHQ11391127GA SHQ11391139GA SHQ11391140GA SHQ11391141GA SHQ11391142GA SHQ11391143GA SHQ11391144GA SHQ11391145GA SHQ11391147GA SHQ11391148GA SHQ11391149GA SHQ11391150GA SHQ11391151GA SHQ11391152GA SHQ11391154GA SHQ11391155GA SHQ11391156GA SHQ11403089SA SHQ11403090SA SHQ11403094SA SHQ11403108SA SHQ11421362GA SHQ11421363GA SHQ11423152SA SHQ11423154SA SHQ11423160SA SHQ11423161SA SHQ11423164SA SHQ11423166SA SHQ11423169SA SHQ11423170SA SHQ11423200SA SHQ11423213SA SHQ11423216SA SHQ11423218SA SHQ11423220SA SHQ11423222SA SHQ11423226SA SHQ11423228SA SHQ11423229SA SHQ11423230SA SHQ11423232SA SHQ11423235SA SHQ11423236SA SHQ11423238SA SHQ11423239SA SHQ11423240SA SHQ11423241SA SHQ11423242SA SHQ11423243SA SHQ11423244SA SHQ11423245SA SHQ11423246SA SHQ11423247SA SHQ11423248SA SHQ11423249SA SHQ11423250SA SHQ11423251SA SHQ11423252SA SHQ11423253SA SHQ11423254SA SHQ11423255SA SHQ11423256SA SHQ11423257SA SHQ11423258SA SHQ11423259SA SHQ11423260SA SHQ11423261SA SHQ11423263SA SHQ11423264SA SHQ11423265SA SHQ11423266SA SHQ11423267SA SHQ11423269SA SHQ11423270SA SHQ11423271SA SHQ11423272SA SHQ11423273SA SHQ11423274SA SHQ11423275SA SHQ11423276SA SHQ11423277SA SHQ11423278SA SHQ11423279SA SHQ11423283SA SHQ11423286SA SHQ11423287SA SHQ11423288SA SHQ11423289SA SHQ11423290SA SHQ11423291SA SHQ11423292SA SHQ11423293SA SHQ11423294SA SHQ11423295SA SHQ11423296SA SHQ11423297SA SHQ11423298SA SHQ11423299SA SHQ11423300SA SHQ11423302SA SHQ11423303SA SHQ11423304SA SHQ11423305SA SHQ11423306SA SHQ11423308SA SHQ11423309SA SHQ11423310SA SHQ11423311SA SHQ11423312SA SHQ11423314SA SHQ11423316SA SHQ11423317SA SHQ11423319SA SHQ11423320SA SHQ11423322SA SHQ11423323SA SHQ11423325SA SHQ11423331SA SHQ11423336SA SHQ11423339SA SHQ11423340SA SHQ11423351SA SHQ11423353SA SHQ11423359SA SHQ11423362SA SHQ11423363SA SHQ11423364SA SHQ11423365SA SHQ11423366SA SHQ11423367SA SHQ11423368SA SHQ11423369SA SHQ11423370SA SHQ11423371SA SHQ11423372SA SHQ11423373SA SHQ11423374SA SHQ11423375SA SHQ11423377SA SHQ11423378SA SHQ11423379SA SHQ11423380SA SHQ11423381SA SHQ11423382SA SHQ11423383SA SHQ11423384SA SHQ11423385SA SHQ11423386SA SHQ11423387SA SHQ11423388SA SHQ11423389SA SHQ11423390SA SHQ11423391SA SHQ11423392SA SHQ11423393SA SHQ11423394SA SHQ11433396SA SHQ11433397SA SHQ11433398SA SHQ11433399SA SHQ11433400SA SHQ11433403SA SHQ11433404SA SHQ11433405SA SHQ11433406SA SHQ11433407SA SHQ11433408SA SHQ11433409SA SHQ11433410SA SHQ11433411SA SHQ11433412SA SHQ11433413SA SHQ11433414SA SHQ11433416SA SHQ11433417SA SHQ11433418SA SHQ11433419SA SHQ11433420SA SHQ11433421SA SHQ11433423SA SHQ11433426SA SHQ11433429SA SHQ11433430SA SHQ11433433SA SHQ11433436SA SHQ11433439SA SHQ11433445SA SHQ11433449SA SHQ11433453SA SHQ11433455SA SHQ11433456SA SHQ11433458SA SHQ11433459SA SHQ11433461SA SHQ11433462SA SHQ11433464SA SHQ11433465SA SHQ11433466SA SHQ11433467SA SHQ11433468SA SHQ11433469SA SHQ11433470SA SHQ11433471SA SHQ11433473SA SHQ11433474SA SHQ11433476SA SHQ11433477SA SHQ11433478SA SHQ11433480SA SHQ11433481SA SHQ11433482SA SHQ11433483SA SHQ11433484SA SHQ11433485SA SHQ11433486SA SHQ11433487SA SHQ11433488SA SHQ11433489SA SHQ11433490SA SHQ11433491SA SHQ11433492SA SHQ11433493SA SHQ11433494SA SHQ11433495SA SHQ11433496SA SHQ11433497SA SHQ11433498SA SHQ11433499SA SHQ11433500SA SHQ11433501SA SHQ11433502SA SHQ11433503SA SHQ11433504SA SHQ11433505SA SHQ11433506SA SHQ11433507SA SHQ11433508SA SHQ11433509SA SHQ11433510SA SHQ11433511SA SHQ11433512SA SHQ11433513SA SHQ11433514SA SHQ11433515SA SHQ11433516SA SHQ11433517SA SHQ11433519SA SHQ11433520SA SHQ11433521SA SHQ11433522SA SHQ11433524SA SHQ11433526SA SHQ11433527SA SHQ11433528SA SHQ11433529SA SHQ11433530SA SHQ11433531SA SHQ11433532SA SHQ11441410GA SHQ11441411GA SHQ11441419GA SHQ11443544SA SHQ11443545SA SHQ11443546SA SHQ11443548SA SHQ11443554SA SHQ11443555SA SHQ11443558SA SHQ11443559SA SHQ11443560SA SHQ11443561SA SHQ11443562SA SHQ11443563SA SHQ11443564SA SHQ11443565SA SHQ11443566SA SHQ11443567SA SHQ11443568SA SHQ11443569SA SHQ11443570SA SHQ11443571SA SHQ11443572SA SHQ11443573SA SHQ11443574SA SHQ11443575SA SHQ11443576SA SHQ11443577SA SHQ11443578SA SHQ11443579SA SHQ11443580SA SHQ11443581SA SHQ11443582SA SHQ11443584SA SHQ11443585SA SHQ11443586SA SHQ11443587SA SHQ11443588SA SHQ11443589SA SHQ11443590SA SHQ11443591SA SHQ11443592SA SHQ11443593SA SHQ11443594SA SHQ11443595SA SHQ11443601SA SHQ11443602SA SHQ11443603SA SHQ11443604SA SHQ11443605SA SHQ11443606SA SHQ11443608SA SHQ11443609SA SHQ11443610SA SHQ11443611SA SHQ11443612SA SHQ11443613SA SHQ11443614SA SHQ11443615SA SHQ11443616SA SHQ11443617SA SHQ11443618SA SHQ11443621SA SHQ11443622SA SHQ11443623SA SHQ11443624SA SHQ11443625SA SHQ11443626SA SHQ11443627SA SHQ11443629SA SHQ11443630SA SHQ11443631SA SHQ11443632SA SHQ11443634SA SHQ11443635SA SHQ11443636SA SHQ11443637SA SHQ11443638SA SHQ11443640SA SHQ11443641SA SHQ11443642SA SHQ11443643SA SHQ11443644SA SHQ11443645SA SHQ11443646SA SHQ11443647SA SHQ11443648SA SHQ11443649SA SHQ11443650SA SHQ11443651SA SHQ11443652SA SHQ11443653SA SHQ11443654SA SHQ11443655SA SHQ11443656SA SHQ11443657SA SHQ11443659SA SHQ11443660SA SHQ11443661SA SHQ11443662SA SHQ11443663SA SHQ11443664SA SHQ11443665SA SHQ11443666SA SHQ11443667SA SHQ11443668SA SHQ11443669SA SHQ11443670SA SHQ11443671SA SHQ11443672SA SHQ11443673SA SHQ11443674SA SHQ11443675SA SHQ11443676SA SHQ11443677SA SHQ11443678SA SHQ11443679SA SHQ11443680SA SHQ11443681SA SHQ11443683SA SHQ11443685SA SHQ11443686SA SHQ11443687SA SHQ11443688SA SHQ11443689SA SHQ11443690SA SHQ11443691SA SHQ11443692SA SHQ11443693SA SHQ11443696SA SHQ11443697SA SHQ11443702SA SHQ11443703SA SHQ11443705SA SHQ11443706SA SHQ11443707SA SHQ11443709SA SHQ11443710SA SHQ11443711SA SHQ11443712SA SHQ11443713SA SHQ11443714SA SHQ11443715SA SHQ11443716SA SHQ11443717SA SHQ11443718SA SHQ11443719SA SHQ11443720SA SHQ11443721SA SHQ11443722SA SHQ11443723SA SHQ11443725SA SHQ11443726SA SHQ11443727SA SHQ11443728SA SHQ11443729SA SHQ11443730SA SHQ11443731SA SHQ11443732SA SHQ11443733SA SHQ11443734SA SHQ11443735SA SHQ11443736SA SHQ11443737SA SHQ11443738SA SHQ11443740SA SHQ11443741SA SHQ11443744SA SHQ11443746SA SHQ11443748SA SHQ11443749SA SHQ11443759SA SHQ11443761SA SHQ11443766SA SHQ11443771SA SHQ11443773SA SHQ11451426GA SHQ11451427GA SHQ11451428GA SHQ11451431GA SHQ11451432GA SHQ11453782SA SHQ11453783SA SHQ11453794SA SHQ11453800SA SHQ11453801SA SHQ11453803SA SHQ11453804SA SHQ11453806SA SHQ11453807SA SHQ11453812SA SHQ11453813SA SHQ11453814SA SHQ11453815SA SHQ11453817SA SHQ11453819SA SHQ11453820SA SHQ11453821SA SHQ11453823SA SHQ11453826SA SHQ11453827SA SHQ11453828SA SHQ11453829SA SHQ11453830SA SHQ11453831SA SHQ11453832SA SHQ11453833SA SHQ11453835SA SHQ11453837SA SHQ11453838SA SHQ11453842SA SHQ11453844SA SHQ11453845SA SHQ11453846SA SHQ11453847SA SHQ11453850SA SHQ11453851SA SHQ11453852SA SHQ11453853SA SHQ11453856SA SHQ11453857SA SHQ11453859SA SHQ11453861SA SHQ11453862SA SHQ11453863SA SHQ11453864SA SHQ11453865SA SHQ11453866SA SHQ11453867SA SHQ11453868SA SHQ11453869SA SHQ11453870SA SHQ11453871SA SHQ11453872SA SHQ11453873SA SHQ11453874SA SHQ11453877SA SHQ11453880SA SHQ11453881SA SHQ11453882SA SHQ11453883SA SHQ11453884SA SHQ11453885SA SHQ11453888SA SHQ11453889SA SHQ11453893SA SHQ11453896SA SHQ11453902SA SHQ11453905SA SHQ11453906SA SHQ11453907SA SHQ11453908SA SHQ11453909SA SHQ11453910SA SHQ11453911SA SHQ11453912SA SHQ11453913SA SHQ11453914SA SHQ11453915SA SHQ11453916SA SHQ11453917SA SHQ11453918SA SHQ11453919SA SHQ11453920SA SHQ11453922SA SHQ11453924SA SHQ11453925SA SHQ11453926SA SHQ11453927SA SHQ11453928SA SHQ11453929SA SHQ11453930SA SHQ11453931SA SHQ11453932SA SHQ11453933SA SHQ11453934SA SHQ11453935SA SHQ11453936SA SHQ11453937SA SHQ11453939SA SHQ11453940SA SHQ11453941SA SHQ11453942SA SHQ11453943SA SHQ11453944SA SHQ11453945SA SHQ11453946SA SHQ11453948SA SHQ11453949SA SHQ11453954SA SHQ11453955SA SHQ11453957SA SHQ11453964SA SHQ11453966SA SHQ11453970SA SHQ11453971SA SHQ11453974SA SHQ11453979SA SHQ11453980SA SHQ11453985SA SHQ11453986SA SHQ11453987SA SHQ11453988SA SHQ11453991SA SHQ11453992SA SHQ11453993SA SHQ11453994SA SHQ11453995SA SHQ11453997SA SHQ11453998SA SHQ11454010SA SHQ11454011SA SHQ11454013SA SHQ11454014SA SHQ11454015SA SHQ11454016SA SHQ11454017SA SHQ11454018SA SHQ11454022SA SHQ11454024SA SHQ11454025SA SHQ11454027SA SHQ11454028SA SHQ11454029SA SHQ11454035SA SHQ11454039SA SHQ11454042SA SHQ11454045SA SHQ11454046SA SHQ11454048SA SHQ11454051SA SHQ11454053SA SHQ11454054SA SHQ11454055SA SHQ11454056SA SHQ11454057SA SHQ11454060SA SHQ11454061SA SHQ11454063SA SHQ11454065SA SHQ11454068SA SHQ11454069SA SHQ11454070SA SHQ11454074SA SHQ11454075SA SHQ11454076SA SHQ11461479GA SHQ11461480GA SHQ11461481GA SHQ11461482GA SHQ11461483GA SHQ11461484GA SHQ11461485GA SHQ11461486GA SHQ11461487GA SHQ11461488GA SHQ11461489GA SHQ11461490GA SHQ11461491GA SHQ11461492GA SHQ11461493GA SHQ11461494GA SHQ11461495GA SHQ11461496GA SHQ11461497GA SHQ11461498GA SHQ11461499GA SHQ11461500GA SHQ11461501GA SHQ11461502GA SHQ11461503GA SHQ11461504GA SHQ11461505GA SHQ11461506GA SHQ11461507GA SHQ11461508GA SHQ11461509GA SHQ11461510GA SHQ11461511GA SHQ11461512GA SHQ11461513GA SHQ11461514GA SHQ11461515GA SHQ11461516GA SHQ11461517GA SHQ11461518GA SHQ11461519GA SHQ11461520GA SHQ11461521GA SHQ11461522GA SHQ11461523GA SHQ11461524GA SHQ11461525GA SHQ11461526GA SHQ11461527GA SHQ11461528GA SHQ11461529GA SHQ11461530GA SHQ11461531GA SHQ11461532GA SHQ11461533GA SHQ11461534GA SHQ11461535GA SHQ11461536GA SHQ11461537GA SHQ11461538GA SHQ11461539GA SHQ11461540GA SHQ11461541GA SHQ11461542GA SHQ11461543GA SHQ11461544GA SHQ11461545GA SHQ11461546GA SHQ11461547GA SHQ11461548GA SHQ11461549GA SHQ11461550GA SHQ11461551GA SHQ11461552GA SHQ11461553GA SHQ11461554GA SHQ11461555GA SHQ11461556GA SHQ11461557GA SHQ11461558GA SHQ11464078SA SHQ11464079SA SHQ11464080SA SHQ11464082SA SHQ11464085SA SHQ11464086SA SHQ11464088SA SHQ11464090SA SHQ11464091SA SHQ11464092SA SHQ11464093SA SHQ11464094SA SHQ11464095SA SHQ11464096SA SHQ11464097SA SHQ11464098SA SHQ11464099SA SHQ11464100SA SHQ11464104SA SHQ11464107SA SHQ11464109SA SHQ11464110SA SHQ11464111SA SHQ11464113SA SHQ11464115SA SHQ11464116SA SHQ11464117SA SHQ11464119SA SHQ11464120SA SHQ11464121SA SHQ11464124SA SHQ11464125SA SHQ11464131SA SHQ11464132SA SHQ11464133SA SHQ11464135SA SHQ11464136SA SHQ11464137SA SHQ11464139SA SHQ11464141SA SHQ11464142SA SHQ11464144SA SHQ11464145SA SHQ11464147SA SHQ11464148SA SHQ11464165SA SHQ11464169SA SHQ11464170SA SHQ11464176SA SHQ11464177SA SHQ11464179SA SHQ11464180SA SHQ11464181SA SHQ11464182SA SHQ11464183SA SHQ11464194SA SHQ11464195SA SHQ11464196SA SHQ11464197SA SHQ11464198SA SHQ11464199SA SHQ11464200SA SHQ11464201SA SHQ11464202SA SHQ11464203SA SHQ11464205SA SHQ11464206SA SHQ11464207SA SHQ11464208SA SHQ11464209SA SHQ11464212SA SHQ11464217SA SHQ11464218SA SHQ11464220SA SHQ11464221SA SHQ11464223SA SHQ11464225SA SHQ11464226SA SHQ11464228SA SHQ11464229SA SHQ11464230SA SHQ11464231SA SHQ11464233SA SHQ11464234SA SHQ11464235SA SHQ11464236SA SHQ11464237SA SHQ11464238SA SHQ11464239SA SHQ11464240SA SHQ11464241SA SHQ11464242SA SHQ11464243SA SHQ11464244SA SHQ11464245SA SHQ11464246SA SHQ11464247SA SHQ11464249SA SHQ11464250SA SHQ11464251SA SHQ11464252SA SHQ11464253SA SHQ11464254SA SHQ11464255SA SHQ11464256SA SHQ11464257SA SHQ11464258SA SHQ11464259SA SHQ11464261SA SHQ11464262SA SHQ11464263SA SHQ11464264SA SHQ11464265SA SHQ11464266SA SHQ11464267SA SHQ11464268SA SHQ11464269SA SHQ11464270SA SHQ11464271SA SHQ11464272SA SHQ11464273SA SHQ11464274SA SHQ11464275SA SHQ11464276SA SHQ11464277SA SHQ11464278SA SHQ11464279SA SHQ11464280SA SHQ11464281SA SHQ11464283SA SHQ11464284SA SHQ11464285SA SHQ11464286SA SHQ11464287SA SHQ11464288SA SHQ11464291SA SHQ11464294SA SHQ11464295SA SHQ11464296SA SHQ11464297SA SHQ11464298SA SHQ11464299SA SHQ11464300SA SHQ11464302SA SHQ11464304SA SHQ11464305SA SHQ11464307SA SHQ11464308SA SHQ11464314SA SHQ11464315SA SHQ11464318SA SHQ11464320SA SHQ11464323SA SHQ11464324SA SHQ11464325SA SHQ11464327SA SHQ11464328SA SHQ11464329SA SHQ11464330SA SHQ11464331SA SHQ11464332SA SHQ11464333SA SHQ11464334SA SHQ11464335SA SHQ11464337SA SHQ11464339SA SHQ11464349SA SHQ11464350SA SHQ11464351SA SHQ11464353SA SHQ11464354SA SHQ11464358SA SHQ11464359SA SHQ11464361SA SHQ11464362SA SHQ11464367SA SHQ11464376SA SHQ11464398SA SHQ11464399SA SHQ11464400SA SHQ11464401SA SHQ11464402SA SHQ11464403SA SHQ11464404SA SHQ11464406SA SHQ11464407SA SHQ11464409SA SHQ11464411SA SHQ11464412SA SHQ11464414SA SHQ11464415SA SHQ11464416SA SHQ11464417SA SHQ11471595GA SHQ11471596GA SHQ11471597GA SHQ11471598GA SHQ11471599GA SHQ11474419SA SHQ11474427SA SHQ11474428SA SHQ11474468SA SHQ11474469SA SHQ11474470SA SHQ11474471SA SHQ11474472SA SHQ11474473SA SHQ11474474SA SHQ11474475SA SHQ11474476SA SHQ11474477SA SHQ11474478SA SHQ11474481SA SHQ11474482SA SHQ11474484SA SHQ11474485SA SHQ11474486SA SHQ11474487SA SHQ11474493SA SHQ11474497SA SHQ11474498SA SHQ11474499SA SHQ11474500SA SHQ11474501SA SHQ11474502SA SHQ11474503SA SHQ11474504SA SHQ11474505SA SHQ11474506SA SHQ11474507SA SHQ11474514SA SHQ11474522SA SHQ11474525SA SHQ11474525SABOP SHQ11474526SA SHQ11474531SA SHQ11474533SA SHQ11474536SA SHQ11474537SA SHQ11474539SA SHQ11474540SA SHQ11474545SA SHQ11474551SA SHQ11474558SA SHQ11474559SA SHQ11474560SA SHQ11474563SA SHQ11474564SA SHQ11474565SA SHQ11474566SA SHQ11474567SA SHQ11474568SA SHQ11474570SA SHQ11474571SA SHQ11474573SA SHQ11474574SA SHQ11474575SA SHQ11474576SA SHQ11474577SA SHQ11474578SA SHQ11474580SA SHQ11474581SA SHQ11474582SA SHQ11474588SA SHQ11474590SA SHQ11474591SA SHQ11474593SA SHQ11474594SA SHQ11474596SA SHQ11474597SA SHQ11474601SA SHQ11474602SA SHQ11474607SA SHQ11474614SA SHQ11474616SA SHQ11474618SA SHQ11474619SA SHQ11474620SA SHQ11474621SA SHQ11474622SA SHQ11474623SA SHQ11474624SA SHQ11474625SA SHQ11474627SA SHQ11474628SA SHQ11474629SA SHQ11474630SA SHQ11474631SA SHQ11474632SA SHQ11474633SA SHQ11474636SA SHQ11474639SA SHQ11474641SA SHQ11474642SA SHQ11474643SA SHQ11474646SA SHQ11474647SA SHQ11474SHQ11474480SA610SA SHQ11481623GA SHQ11484649SA SHQ11484654SA SHQ11484656SA SHQ11484660SA SHQ11484663SA SHQ11484678SA SHQ11484679SA SHQ11484680SA SHQ11484682SA SHQ11484683SA SHQ11484684SA SHQ11484685SA SHQ11484686SA SHQ11484687SA SHQ11484688SA SHQ11484690SA SHQ11484691SA SHQ11484692SA SHQ11484693SA SHQ11484694SA SHQ11484695SA SHQ11484696SA SHQ11484697SA SHQ11484698SA SHQ11484699SA SHQ11484700SA SHQ11484701SA SHQ11484702SA SHQ11484704SA SHQ11484709SA SHQ11484710SA SHQ11484715SA SHQ11484716SA SHQ11484717SA SHQ11484727SA SHQ11484737SA SHQ11484745SA SHQ11484747SA SHQ11484748SA SHQ11484749SA SHQ11484758SA SHQ11484759SA SHQ11484761SA SHQ11484764SA SHQ11484765SA SHQ11484767SA SHQ11484772SA SHQ11484773SA SHQ11484775SA SHQ11484779SA SHQ11484780SA SHQ11484786SA SHQ11484788SA SHQ11484793SA SHQ11484794SA SHQ11484797SA SHQ11484799SA SHQ11484800SA SHQ11484802SA SHQ11484804SA SHQ11484806SA SHQ11484808SA SHQ11484809SA SHQ11484810SA SHQ11484812SA SHQ11484813SA SHQ11484816SA SHQ11484817SA SHQ11484818SA SHQ11484819SA SHQ11484821SA SHQ11484826SA SHQ11484829SA SHQ11484830SA SHQ11484832SA SHQ11484833SA SHQ11484834SA SHQ11484835SA SHQ11484837SA SHQ11484839SA SHQ11484842SA SHQ11484844SA SHQ11484847SA SHQ11484848SA SHQ11484850SA SHQ11484851SA SHQ11484851SABOP SHQ11484852SA SHQ11484853SA SHQ11484854SA SHQ11484855SA SHQ11484856SA SHQ11484857SA SHQ11484861SA SHQ11484862SA SHQ11484863SA SHQ11484867SA SHQ11484868SA SHQ11484869SA SHQ11484870SA SHQ11484871SA SHQ11484873SA SHQ11484874SA SHQ11484876SA SHQ11484881SA SHQ11484882SA SHQ11484883SA SHQ11484884SA SHQ11484885SA SHQ11484888SA SHQ11484889SA SHQ11484892SA SHQ11484893SA SHQ11484894SA SHQ11484897SA SHQ11484898SA SHQ11484899SA SHQ11484900SA SHQ11484901SA SHQ11484902SA SHQ11484904SA SHQ11484905SA SHQ11484906SA SHQ11484907SA SHQ11484908SA SHQ11484909SA SHQ11484910SA SHQ11484911SA SHQ11484912SA SHQ11484913SA SHQ11484914SA SHQ11484915SA SHQ11484916SA SHQ11484921SA SHQ11484922SA SHQ11484924SA SHQ11484938SA SHQ11484939SA SHQ11484942SA SHQ11484954SA SHQ11484956SA SHQ11484957SA SHQ11484958SA SHQ11484960SA SHQ11484964SA SHQ11484965SA SHQ11484968SA SHQ11484969SA SHQ11484973SA SHQ11484975SA SHQ11484978SA SHQ11484982SA SHQ11484984SA SHQ11484989SA SHQ11491903GACOMPLETE SHQ11491918GA SHQ11491919GA SHQ11491920GA SHQ11491921GA SHQ11491922GA SHQ11491923GA SHQ11494992SA SHQ11494993SA SHQ11494994SA SHQ11494995SA SHQ11494996SA SHQ11494997SA SHQ11494998SA SHQ11494999SA SHQ11495000SA SHQ11495001SA SHQ11495002SA SHQ11495003SA SHQ11495004SA SHQ11495005SA SHQ11495006SA SHQ11495007SA SHQ11495008SA SHQ11495010SA SHQ11495011SA SHQ11495012SA SHQ11495013SA SHQ11495014SA SHQ11495015SA SHQ11495016SA SHQ11495017SA SHQ11495018SA SHQ11495019SA SHQ11495020SA SHQ11495021SA SHQ11495022SA SHQ11495023SA SHQ11495024SA SHQ11495025SA SHQ11495026SA SHQ11495027SA SHQ11495028SA SHQ11495029SA SHQ11495030SA SHQ11495031SA SHQ11495032SA SHQ11495033SA SHQ11495034SA SHQ11495035SA SHQ11495036SA SHQ11495037SA SHQ11495038SA SHQ11495039SA SHQ11495040SA SHQ11495041SA SHQ11495042SA SHQ11495043SA SHQ11495044SA SHQ11495045SA SHQ11495046SA SHQ11495047SA SHQ11495048SA SHQ11495049SA SHQ11495050SA SHQ11495051SA SHQ11495053SA SHQ11495054SA SHQ11495055SA SHQ11495057SA SHQ11495058SA SHQ11495059SA SHQ11495060SA SHQ11495061SA SHQ11495062SA SHQ11495063SA SHQ11495064SA SHQ11495066SA SHQ11495070SA SHQ11495071SA SHQ11495087SA SHQ11495088SA SHQ11495089SA SHQ11495090SA SHQ11495092SA SHQ11495093SA SHQ11495094SA SHQ11495095SA SHQ11495096SA SHQ11495097SA SHQ11495100SA SHQ11495101SA SHQ11495102SA SHQ11495103SA SHQ11495104SA SHQ11495105SA SHQ11495106SA SHQ11495107SA SHQ11495108SA SHQ11495109SA SHQ11495110SA SHQ11495111SA SHQ11495112SA SHQ11495113SA SHQ11495114SA SHQ11495115SA SHQ11495116SA SHQ11495117SA SHQ11495118SA SHQ11495119SA SHQ11495120SA SHQ11495121SA SHQ11495122SA SHQ11495123SA SHQ11495124SA SHQ11495125SA SHQ11495126SA SHQ11495127SA SHQ11495128SA SHQ11495129SA SHQ11495130SA SHQ11495131SA SHQ11495132SA SHQ11495133SA SHQ11495134SA SHQ11495135SA SHQ11495136SA SHQ11495137SA SHQ11495138SA SHQ11495139SA SHQ11495140SA SHQ11495141SA SHQ11495142SA SHQ11495143SA SHQ11495144SA SHQ11495145SA SHQ11495146SA SHQ11495147SA SHQ11495148SA SHQ11495149SA SHQ11495150SA SHQ11495151SA SHQ11495152SA SHQ11495153SA SHQ11495154SA SHQ11495155SA SHQ11495156SA SHQ11495157SA SHQ11495158SA SHQ11495159SA SHQ11495160SA SHQ11495162SA SHQ11495163SA SHQ11495164SA SHQ11495165SA SHQ11495166SA SHQ11495167SA SHQ11495171SA SHQ11495172SA SHQ11495174SA SHQ11495175SA SHQ11495176SA SHQ11495177SA SHQ11495178SA SHQ11495179SA SHQ11495180SA SHQ11495181SA SHQ11495182SA SHQ11495183SA SHQ11495189SA SHQ114SHQ11464146SA64152SA SHQ11502031GA SHQ11502032GA SHQ11502033GA SHQ11502034GA SHQ11502035GA SHQ11502036GA SHQ11502037GA SHQ11502038GA SHQ11502039GA SHQ11502040GA SHQ11502041GA SHQ11502042GA SHQ11502043GA SHQ11502044GA SHQ11502045GA SHQ11502046GA SHQ11502047GA SHQ11502048GA SHQ11502050GA SHQ11502051GA SHQ11502052GA SHQ11502053GA SHQ11502054GA SHQ11502055GA SHQ11502056GA SHQ11502057GA SHQ11502058GA SHQ11502059GA SHQ11502060GA SHQ11502061GA SHQ11502063GA SHQ11502064GA SHQ11502065GA SHQ11502066GA SHQ11502067GA SHQ11502068GA SHQ11502069GA SHQ11502070GA SHQ11502072GA SHQ11502073GA SHQ11502075GA SHQ11502082GA SHQ11502083GA SHQ11502084GA SHQ11502085GA SHQ11502086GA SHQ11502087GA SHQ11502088GA SHQ11502089GA SHQ11502090GA SHQ11502091GA SHQ11502092GA SHQ11502093GA SHQ11502094GA SHQ11502095GA SHQ11502096GA SHQ11502097GA SHQ11502098GA SHQ11502099GA SHQ11502100GA SHQ11502101GA SHQ11502103GA SHQ11502104GA SHQ11502105GA SHQ11502106GA SHQ11502107GA SHQ11502108GA SHQ11502109GA SHQ11502110GA SHQ11502111GA SHQ11502112GA SHQ11502113GA SHQ11502114GA SHQ11502115GA SHQ11502116GA SHQ11502117GA SHQ11502118GA SHQ11502119GA SHQ11502120GA SHQ11502121GA SHQ11502122GA SHQ11502123GA SHQ11502124GA SHQ11502125GA SHQ11502126GA SHQ11502127GA SHQ11502128GA SHQ11502129GA SHQ11502130GA SHQ11502131GA SHQ11502132GA SHQ11502133GA SHQ11502134GA SHQ11502135GA SHQ11502136GA SHQ11502137GA SHQ11502138GA SHQ11502139GA SHQ11502140GA SHQ11502141GA SHQ11502142GA SHQ11502143GA SHQ11502144GA SHQ11502145GA SHQ11502146GA SHQ11502147GA SHQ11502148GA SHQ11502149GA SHQ11502150GA SHQ11502151GA SHQ11502152GA SHQ11502153GA SHQ11502154GA SHQ11502155GA SHQ11502156GA SHQ11502157GA SHQ11502158GA SHQ11502159GA SHQ11502160GA SHQ11502161GA SHQ11502162GA SHQ11502163GA SHQ11502196GA SHQ11502197GA SHQ11502198GA SHQ11502199GA SHQ11502200GA SHQ11502201GA SHQ11502202GA SHQ11502203GA SHQ11502204GA SHQ11502205GA SHQ11502206GA SHQ11502207GA SHQ11502208GA SHQ11502209GA SHQ11502210GA SHQ11502211GA SHQ11502212GA SHQ11502213GA SHQ11502214GA SHQ11502215GA SHQ11502216GA SHQ11502217GA SHQ11502218GA SHQ11502219GA SHQ11502220GA SHQ11502221GA SHQ11502222GA SHQ11502223GA SHQ11502224GA SHQ11502225GA SHQ11505199SA SHQ11505200SA SHQ11505201SA SHQ11505202SA SHQ11505203SA SHQ11505204SA SHQ11505205SA SHQ11505206SA SHQ11505207SA SHQ11505209SA SHQ11505213SA SHQ11505214SA SHQ11505216SA SHQ11505224SA SHQ11505225SA SHQ11505226SA SHQ11505227SA SHQ11505228SA SHQ11505229SA SHQ11505230SA SHQ11505231SA SHQ11505232SA SHQ11505233SA SHQ11505234SA SHQ11505235SA SHQ11505236SA SHQ11505237SA SHQ11505238SA SHQ11505239SA SHQ11505240SA SHQ11505241SA SHQ11505242SA SHQ11505243SA SHQ11505244SA SHQ11505245SA SHQ11505246SA SHQ11505247SA SHQ11505248SA SHQ11505249SA SHQ11505250SA SHQ11505251SA SHQ11505252SA SHQ11505253SA SHQ11505254SA SHQ11505255SA SHQ11505256SA SHQ11505257SA SHQ11505258SA SHQ11505259SA SHQ11505260SA SHQ11505261SA SHQ11505262SA SHQ11505263SA SHQ11505264SA SHQ11505265SA SHQ11505266SA SHQ11505267SA SHQ11505268SA SHQ11505269SA SHQ11505270SA SHQ11505271SA SHQ11505272SA SHQ11505273SA SHQ11505274SA SHQ11505275SA SHQ11505276SA SHQ11505277SA SHQ11505279SA SHQ11505280SA SHQ11505281SA SHQ11505282SA SHQ11505283SA SHQ11505284SA SHQ11505285SA SHQ11505287SA SHQ11505288SA SHQ11505289SA SHQ11505290SA SHQ11505291SA SHQ11505292SA SHQ11505293SA SHQ11505294SA SHQ11505296SA SHQ11505297SA SHQ11505298SA SHQ11505299SA SHQ11505301SA SHQ11505303SA SHQ11505304SA SHQ11505325SA SHQ11505326SA SHQ11505328SA SHQ11505329SA SHQ11505330SA SHQ11505331SA SHQ11505332SA SHQ11505334SA SHQ11505335SA SHQ11505336SA SHQ11505337SA SHQ11505339SA SHQ11505341SA SHQ11505342SA SHQ11505343SA SHQ11505344SA SHQ11505345SA SHQ11505346SA SHQ11505349SA SHQ11505350SA SHQ11505351SA SHQ11505353SA SHQ11505354SA SHQ11505355SA SHQ11505357SA SHQ11505362SA SHQ11505366SA SHQ11505367SA SHQ11505368SA SHQ11505369SA SHQ11505370SA SHQ11505371SA SHQ11505373SA SHQ11505374SA SHQ11505375SA SHQ11505377SA SHQ11505378SA SHQ11505380SA SHQ11505381SA SHQ11505382SA SHQ11512226GA SHQ11512228GA SHQ11512229GA SHQ11512232GA SHQ12015385SA SHQ12015390SA SHQ12015391SA SHQ12015392SA SHQ12015393SA SHQ12015394SA SHQ12015397SA SHQ12015398SA SHQ12015400SA SHQ12015402SA SHQ12015403SA SHQ12015404SA SHQ12015406SA SHQ12015407SA SHQ12015409SA SHQ12015410SA SHQ12015411SA SHQ12015414SA SHQ12015415SA SHQ12015416SA SHQ12015417SA SHQ12015419SA SHQ12015421SA SHQ12015422SA SHQ12015423SA SHQ12015427SA SHQ12015428SA SHQ12015430SA SHQ12015432SA SHQ12015433SA SHQ12015437SA SHQ12015438SA SHQ12015440SA SHQ12015444SA SHQ12015445SA SHQ12015446SA SHQ12015450SA SHQ12015452SA SHQ12015459SA SHQ12015462SA SHQ12015463SA SHQ12015464SA SHQ12015467SA SHQ12015468SA SHQ12015469SA SHQ12015470SA SHQ12015474SA SHQ12025483SA SHQ12025484SA SHQ12025486SA SHQ12025487SA SHQ12025488SA SHQ12025489SA SHQ12025490SA SHQ12025492SA SHQ12025493SA SHQ12025494SA SHQ12025495SA SHQ12025496SA SHQ12025497SA SHQ12025498SA SHQ12025499SA SHQ12025500SA SHQ12025501SA SHQ12025502SA SHQ12025503SA SHQ12025504SA SHQ12025505SA SHQ12025506SA SHQ12025507SA SHQ12025509SA SHQ12025511SA SHQ12025512SA SHQ12025514SA SHQ12025515SA SHQ12025517SA SHQ12025520SA SHQ12025521SA SHQ12025522SA SHQ12025523SA SHQ12025526SA SHQ12025527SA SHQ12025530SA SHQ12025540SA SHQ12032468GA SHQ12032469GA SHQ12032470GA SHQ12032487GA SHQ12032488GA SHQ12032489GA SHQ12032491GA SHQ12032492GA SHQ12032493GA SHQ12032494GA SHQ12032495GA SHQ12032496GA SHQ12032497GA SHQ12032498GA SHQ12035542SA SHQ12035543SA SHQ12035544SA SHQ12035546SA SHQ12035547SA SHQ12035555SA SHQ12035561SA SHQ12035564SA SHQ12035565SA SHQ12035566SA SHQ12035567SA SHQ12035568SA SHQ12035569SA SHQ12035570SA SHQ12035572SA SHQ12035575SA SHQ12035576SA SHQ12035578SA SHQ12035584SA SHQ12035586SA SHQ12035588SA SHQ12035589SA SHQ12035590SA SHQ12035591SA SHQ12035592SA SHQ12035593SA SHQ12035594SA SHQ12035595SA SHQ12035596SA SHQ12035597SA SHQ12035598SA SHQ12035599SA SHQ12035600SA SHQ12035601SA SHQ12035602SA SHQ12035603SA SHQ12035605SA SHQ12035606SA SHQ12035607SA SHQ12035608SA SHQ12035611SA SHQ12035612SA SHQ12035613SA SHQ12035614SA SHQ12035615SA SHQ12035616SA SHQ12035617SA SHQ12035618SA SHQ12035619SA SHQ12035620SA SHQ12035621SA SHQ12035622SA SHQ12035623SA SHQ12035624SA SHQ12035625SA SHQ12035627SA SHQ12035628SA SHQ12035629SA SHQ12035630SA SHQ12035631SA SHQ12035632SA SHQ12035633SA SHQ12035634SA SHQ12035635SA SHQ12035636SA SHQ12035637SA SHQ12035638SA SHQ12035639SA SHQ12035640SA SHQ12035643SA SHQ12035644SA SHQ12035645SA SHQ12035646SA SHQ12035647SA SHQ12035648SA SHQ12035650SA SHQ12035651SA SHQ12035652SA SHQ12035653SA SHQ12035654SA SHQ12035656SA SHQ12035657SA SHQ12035658SA SHQ12035659SA SHQ12035660SA SHQ12035661SA SHQ12035662SA SHQ12035663SA SHQ12035664SA SHQ12035665SA SHQ12035666SA SHQ12035667SA SHQ12035668SA SHQ12035670SA SHQ12035671SA SHQ12035672SA SHQ12035675SA SHQ12035676SA SHQ12035677SA SHQ12035679SA SHQ12035680SA SHQ12035681SA SHQ12035685SA SHQ12035687SA SHQ12035692SA SHQ12035693SA SHQ12035701SA SHQ12035702SA SHQ12035703SA SHQ12035704SA SHQ12035705SA SHQ12035706SA SHQ12035707SA SHQ12035708SA SHQ12035709SA SHQ12035710SA SHQ12035711SA SHQ12035712SA SHQ12035713SA SHQ12035714SA SHQ12035715SA SHQ12035716SA SHQ12035717SA SHQ12035719SA SHQ12042508GA SHQ12042519GA SHQ12042520GA SHQ12042521GA SHQ12042522GA SHQ12042523GA SHQ12042524GA SHQ12042562GA SHQ12042563GA SHQ12042564GA SHQ12042565GA SHQ12042566GA SHQ12042567GA SHQ12042568GA SHQ12042570GA SHQ12042571GA SHQ12045754SA SHQ12045755SA SHQ12045757SA SHQ12045758SA SHQ12045759SA SHQ12045761SA SHQ12045762SA SHQ12045765SA SHQ12045767SA SHQ12045769SA SHQ12045770SA SHQ12045776SA SHQ12045777SA SHQ12045779SA SHQ12045780SA SHQ12045781SA SHQ12045784SA SHQ12045785SA SHQ12045786SA SHQ12045789SA SHQ12045790SA SHQ12045791SA SHQ12045792SA SHQ12045796SA SHQ12045797SA SHQ12045799SA SHQ12045800SA SHQ12045801SA SHQ12045802SA SHQ12045803SA SHQ12045804SA SHQ12045805SA SHQ12045806SA SHQ12045807SA SHQ12045808SA SHQ12045809SA SHQ12045810SA SHQ12045811SA SHQ12045812SA SHQ12045814SA SHQ12045815SA SHQ12045816SA SHQ12045817SA SHQ12045818SA SHQ12045819SA SHQ12045820SA SHQ12045821SA SHQ12045822SA SHQ12045823SA SHQ12045824SA SHQ12045825SA SHQ12045828SA SHQ12045831SA SHQ12045837SA SHQ12045838SA SHQ12045839SA SHQ12045840SA SHQ12045841SA SHQ12045842SA SHQ12045843SA SHQ12045845SA SHQ12045846SA SHQ12045847SA SHQ12045848SA SHQ12045849SA SHQ12045850SA SHQ12045852SA SHQ12045853SA SHQ12045854SA SHQ12045855SA SHQ12045856SA SHQ12045857SA SHQ12045858SA SHQ12045859SA SHQ12045860SA SHQ12045861SA SHQ12045862SA SHQ12045863SA SHQ12045864SA SHQ12045865SA SHQ12045866SA SHQ12045867SA SHQ12045868SA SHQ12045869SA SHQ12045870SA SHQ12045871SA SHQ12045872SA SHQ12045873SA SHQ12045875SA SHQ12045876SA SHQ12045877SA SHQ12045878SA SHQ12045879SA SHQ12045880SA SHQ12045881SA SHQ12045882SA SHQ12045883SA SHQ12045884SA SHQ12045885SA SHQ12045886SA SHQ12045887SA SHQ12045888SA SHQ12045889SA SHQ12045890SA SHQ12045891SA SHQ12045892SA SHQ12045893SA SHQ12045894SA SHQ12045895SA SHQ12045896SA SHQ12045897SA SHQ12045898SA SHQ12045899SA SHQ12045900SA SHQ12045901SA SHQ12052572GA SHQ12052585GA SHQ12052586GA SHQ12052587GA SHQ12052588GA SHQ12055902SA SHQ12055903SA SHQ12055904SA SHQ12055905SA SHQ12055906SA SHQ12055907SA SHQ12055908SA SHQ12055909SA SHQ12055910SA SHQ12055911SA SHQ12055912SA SHQ12055913SA SHQ12055914SA SHQ12055916SA SHQ12055919SA SHQ12055923SA SHQ12055926SA SHQ12055928SA SHQ12055929SA SHQ12055930SA SHQ12055951SA SHQ12055952SA SHQ12055953SA SHQ12055954SA SHQ12055955SA SHQ12055958SA SHQ12055960SA SHQ12055961SA SHQ12055963SA SHQ12055964SA SHQ12055965SA SHQ12055970SA SHQ12055973SA SHQ12055977SA SHQ12055978SA SHQ12055979SA SHQ12055981SA SHQ12055982SA SHQ12055983SA SHQ12055984SA SHQ12055986SA SHQ12055987SA SHQ12055988SA SHQ12055989SA SHQ12055991SA SHQ12055992SA SHQ12055993SA SHQ12055994SA SHQ12055995SA SHQ12055998SA SHQ12056001SA SHQ12062637GA SHQ12062639GA SHQ12062640GA SHQ12062641GA SHQ12062642GA SHQ12062643GA SHQ12062644GA SHQ12062645GA SHQ12062679GA SHQ12066023SA SHQ12066026SA SHQ12066027SA SHQ12066028SA SHQ12066034SA SHQ12066040SA SHQ12066041SA SHQ12066051SA SHQ12066053SA SHQ12066054SA SHQ12066055SA SHQ12066056SA SHQ12066057SA SHQ12066060SA SHQ12066064SA SHQ12066065SA SHQ12066066SA SHQ12066067SA SHQ12066068SA SHQ12066069SA SHQ12066070SA SHQ12066071SA SHQ12066072SA SHQ12066073SA SHQ12066074SA SHQ12066075SA SHQ12066076SA SHQ12066077SA SHQ12066078SA SHQ12066079SA SHQ12066080SA SHQ12066081SA SHQ12066082SA SHQ12066086SA SHQ12066087SA SHQ12066088SA SHQ12066090SA SHQ12066094SA SHQ12066096SA SHQ12066097SA SHQ12066099SA SHQ12066100SA SHQ12066101SA SHQ12066102SA SHQ12066103SA SHQ12066104SA SHQ12066109SA SHQ12066122SA SHQ12066128SA SHQ12066129SA SHQ12066130SA SHQ12066131SA SHQ12066132SA SHQ12066133SA SHQ12066134SA SHQ12066135SA SHQ12066136SA SHQ12066137SA SHQ12066138SA SHQ12066139SA SHQ12066140SA SHQ12066141SA SHQ12066142SA SHQ12066143SA SHQ12066144SA SHQ12066145SA SHQ12066146SA SHQ12066147SA SHQ12066148SA SHQ12066149SA SHQ12066151SA SHQ12066152SA SHQ12066154SA SHQ12066155SA SHQ12066156SA SHQ12066157SA SHQ12066158SA SHQ12066159SA SHQ12066160SA SHQ12066161SA SHQ12066184SA SHQ12066185SA SHQ12066188SA SHQ12066194SA SHQ12066195SA SHQ12066196SA SHQ12066197SA SHQ12072680GA SHQ12072681GA SHQ12072682GA SHQ12072683GA SHQ12072684GA SHQ12072685GA SHQ12072686GA SHQ12072693GA SHQ12072694GA SHQ12072696GA SHQ12072697GA SHQ12072699GA SHQ12072700GA SHQ12076198SA SHQ12076199SA SHQ12076200SA SHQ12076201SA SHQ12076202SA SHQ12076203SA SHQ12076204SA SHQ12076205SA SHQ12076206SA SHQ12076207SA SHQ12076208SA SHQ12076210SA SHQ12076211SA SHQ12076212SA SHQ12076213SA SHQ12076214SA SHQ12076215SA SHQ12076216SA SHQ12076217SA SHQ12076218SA SHQ12076219SA SHQ12076220SA SHQ12076221SA SHQ12076222SA SHQ12076223SA SHQ12076224SA SHQ12076225SA SHQ12076227SA SHQ12076228SA SHQ12076229SA SHQ12076236SA SHQ12076239SA SHQ12076254SA SHQ12076255SA SHQ12076259SA SHQ12076270SA SHQ12076271SA SHQ12076272SA SHQ12076273SA SHQ12076274SA SHQ12076275SA SHQ12076276SA SHQ12076277SA SHQ12076278SA SHQ12076279SA SHQ12076280SA SHQ12076281SA SHQ12076282SA SHQ12076283SA SHQ12076284SA SHQ12076285SA SHQ12076286SA SHQ12076287SA SHQ12076288SA SHQ12076289SA SHQ12076290SA SHQ12076291SA SHQ12076292SA SHQ12076293SA SHQ12076294SA SHQ12076295SA SHQ12076296SA SHQ12076297SA SHQ12076298SA SHQ12076299SA SHQ12076300SA SHQ12076301SA SHQ12076302SA SHQ12076304SA SHQ12076305SA SHQ12076307SA SHQ12076308SA SHQ12076309SA SHQ12076311SA SHQ12076313SA SHQ12082707GA SHQ12082708GA SHQ12082709GA SHQ12082710GA SHQ12082713GA SHQ12082714GA SHQ12082715GA SHQ12082716GA SHQ12082717GA SHQ12082718GA SHQ12082719GA SHQ12082720GA SHQ12082721GA SHQ12082722GA SHQ12082723GA SHQ12082725GA SHQ12082747GA SHQ12082748GA SHQ12082749GA SHQ12082750GA SHQ12082752GA SHQ12082753GA SHQ12082754GA SHQ12082755GA SHQ12082779GA SHQ12086315SA SHQ12086316SA SHQ12086317SA SHQ12086318SA SHQ12086319SA SHQ12086320SA SHQ12086321SA SHQ12086322SA SHQ12086323SA SHQ12086324SA SHQ12086325SA SHQ12086326SA SHQ12086327SA SHQ12086328SA SHQ12086329SA SHQ12086330SA SHQ12086331SA SHQ12086332SA SHQ12086333SA SHQ12086334SA SHQ12086335SA SHQ12086336SA SHQ12086337SA SHQ12086338SA SHQ12086339SA SHQ12086340SA SHQ12086341SA SHQ12086342SA SHQ12086343SA SHQ12086344SA SHQ12092819GA SHQ12092822GA SHQ12092824GA SHQ12092826GA SHQ12092827GA SHQ12096345SA SHQ12096346SA SHQ12096347SA SHQ12096348SA SHQ12096349SA SHQ12096350SA SHQ12096351SA SHQ12096354SA SHQ12096358SA SHQ12096370SA SHQ12096375SA SHQ12096376SA SHQ12096378SA SHQ12096381SA SHQ12096383SA SHQ12096390SA SHQ12096394SA SHQ12096395SA SHQ12096396SA SHQ12096399SA SHQ12096400SA SHQ12096401SA SHQ12096402SA SHQ12096403SA SHQ12096404SA SHQ12096406SA SHQ12096408SA SHQ12096409SA SHQ12096410SA SHQ12096412SA SHQ12096414SA SHQ12096417SA SHQ12096419SA SHQ12096421SA SHQ12096423SA SHQ12096426SA SHQ12096427SA SHQ12096430SA SHQ12096441SA SHQ12096447SA SHQ12096450SA SHQ12096452SA SHQ12096454SA SHQ12096455SA SHQ12096462SA SHQ12096464SA SHQ12106468SA SHQ12106469SA SHQ12106470SA SHQ12106471SA SHQ12106472SA SHQ12106473SA SHQ12106474SA SHQ12106475SA SHQ12106476SA SHQ12106477SA SHQ12106478SA SHQ12106479SA SHQ12106480SA SHQ12106481SA SHQ12106482SA SHQ12106483SA SHQ12106484SA SHQ12106485SA SHQ12106486SA SHQ12106487SA SHQ12106490SA SHQ12106491SA SHQ12106493SA SHQ12106494SA SHQ12106496SA SHQ12106498SA SHQ12106499SA SHQ12106500SA SHQ12106504SA SHQ12106506SA SHQ12106507SA SHQ12106511SA SHQ12106512SA SHQ12106514SA SHQ12106528SA SHQ12106529SA SHQ12106530SA SHQ12106531SA SHQ12106532SA SHQ12106533SA SHQ12106534SA SHQ12106535SA SHQ12106537SA SHQ12106540SA SHQ12106541SA SHQ12106542SA SHQ12106545SA SHQ12106548SA SHQ12106552SA SHQ12112923GA SHQ12112925GA SHQ12112927GA SHQ12112928GA SHQ12112929GA SHQ12112930GA SHQ12112931GA SHQ12112935GA SHQ12112959GA SHQ12112960GA SHQ12112961GA SHQ12112962GA SHQ12116554SA SHQ12116555SA SHQ12116558SA SHQ12116560SA SHQ12116561SA SHQ12116562SA SHQ12116563SA SHQ12116564SA SHQ12116565SA SHQ12116566SA SHQ12116567SA SHQ12116568SA SHQ12116569SA SHQ12116570SA SHQ12116572SA SHQ12116573SA SHQ12116574SA SHQ12116575SA SHQ12116576SA SHQ12116577SA SHQ12116578SA SHQ12116579SA SHQ12116580SA SHQ12116581SA SHQ12116582SA SHQ12116583SA SHQ12116585SA SHQ12116587SA SHQ12116588SA SHQ12116590SA SHQ12116591SA SHQ12116592SA SHQ12116596SA SHQ12116597SA SHQ12116598SA SHQ12116599SA SHQ12116600SA SHQ12116601SA SHQ12116602SA SHQ12116603SA SHQ12116604SA SHQ12116605SA SHQ12116606SA SHQ12116607SA SHQ12116608SA SHQ12116609SA SHQ12116610SA SHQ12116611SA SHQ12116612SA SHQ12116613SA SHQ12116614SA SHQ12116615SA SHQ12116616SA SHQ12116617SA SHQ12116618SA SHQ12116619SA SHQ12116620SA SHQ12116621SA SHQ12116622SA SHQ12116623SA SHQ12116624SA SHQ12116625SA SHQ12116627SA SHQ12116628SA SHQ12116630SA SHQ12116631SA SHQ12116632SA SHQ12116633SA SHQ12116635SA SHQ12116636SA SHQ12116637SA SHQ12116638SA SHQ12116639SA SHQ12116640SA SHQ12122972GA SHQ12122973GA SHQ12122974GA SHQ12122981GA SHQ12122984GA SHQ12122986GA SHQ12126641SA SHQ12126642SA SHQ12126643SA SHQ12126644SA SHQ12126645SA SHQ12126648SA SHQ12126649SA SHQ12126651SA SHQ12126652SA SHQ12126654SA SHQ12126655SA SHQ12126656SA SHQ12126657SA SHQ12126658SA SHQ12126663SA SHQ12126665SA SHQ12126666SA SHQ12126671SA SHQ12126673SA SHQ12126674SA SHQ12126675SA SHQ12126677SA SHQ12126678SA SHQ12126679SA SHQ12126680SA SHQ12126681SA SHQ12126683SA SHQ12126684SA SHQ12126685SA SHQ12126686SA SHQ12126688SA SHQ12126690SA SHQ12126691SA SHQ12126692SA SHQ12126693SA SHQ12126694SA SHQ12126695SA SHQ12126696SA SHQ12126698SA SHQ12126699SA SHQ12126700SA SHQ12126701SA SHQ12126702SA SHQ12126703SA SHQ12126704SA SHQ12126705SA SHQ12126706SA SHQ12126707SA SHQ12126708SA SHQ12126709SA SHQ12126710SA SHQ12126711SA SHQ12126712SA SHQ12126713SA SHQ12126714SA SHQ12126715SA SHQ12126716SA SHQ12126717SA SHQ12126718SA SHQ12126720SA SHQ12126722SA SHQ12126723SA SHQ12136731SA SHQ12136732SA SHQ12136733SA SHQ12136734SA SHQ12136735SA SHQ12136736SA SHQ12136737SA SHQ12136739SA SHQ12136740SA SHQ12136741SA SHQ12136742SA SHQ12136743SA SHQ12136744SA SHQ12136745SA SHQ12136746SA SHQ12136747SA SHQ12136748SA SHQ12136749SA SHQ12136750SA SHQ12136751SA SHQ12136752SA SHQ12136753SA SHQ12136754SA SHQ12136755SA SHQ12136756SA SHQ12136757SA SHQ12136758SA SHQ12136759SA SHQ12136760SA SHQ12136761SA SHQ12136762SA SHQ12136763SA SHQ12136764SA SHQ12136765SA SHQ12136766SA SHQ12136767SA SHQ12136768SA SHQ12136769SA SHQ12136770SA SHQ12136773SA SHQ12136774SA SHQ12136776SA SHQ12136777SA SHQ12136779SA SHQ12136783SA SHQ12136784SA SHQ12136785SA SHQ12136787SA SHQ12136788SA SHQ12143019GA SHQ12146825SA SHQ12146826SA SHQ12146827SA SHQ12146828SA SHQ12146829SA SHQ12146830SA SHQ12146831SA SHQ12146832SA SHQ12146833SA SHQ12146834SA SHQ12146835SA SHQ12146836SA SHQ12146837SA SHQ12146839SA SHQ12146840SA SHQ12146841SA SHQ12146857SA SHQ12146858SA SHQ12146859SA SHQ12146860SA SHQ12146861SA SHQ12146862SA SHQ12146863SA SHQ12146864SA SHQ12146865SA SHQ12146866SA SHQ12146867SA SHQ12146868SA SHQ12146869SA SHQ12146870SA SHQ12146871SA SHQ12146872SA SHQ12146873SA SHQ12146874SA SHQ12146875SA SHQ12146876SA SHQ12153052GA SHQ12156887SA SHQ12156888SA SHQ12156889SA SHQ12156890SA SHQ12156891SA SHQ12156892SA SHQ12156893SA SHQ12156894SA SHQ12156895SA SHQ12156896SA SHQ12156897SA SHQ12156898SA SHQ12156899SA SHQ12156900SA SHQ12156901SA SHQ12156902SA SHQ12156903SA SHQ12156904SA SHQ12156905SA SHQ12156906SA SHQ12156907SA SHQ12156908SA SHQ12156909SA SHQ12156910SA SHQ12156911SA SHQ12156912SA SHQ12156914SA SHQ12156915SA SHQ12156916SA SHQ12156918SA SHQ12156920SA SHQ12156921SA SHQ12156922SA SHQ12156923SA SHQ12156924SA SHQ12156925SA SHQ12156926SA SHQ12156927SA SHQ12156928SA SHQ12156929SA SHQ12156930SA SHQ12156931SA SHQ12156932SA SHQ12156933SA SHQ12156934SA SHQ12156935SA SHQ12156936SA SHQ12156937SA SHQ12156938SA SHQ12156939SA SHQ12156940SA SHQ12156941SA SHQ12156942SA SHQ12156943SA SHQ12156944SA SHQ12156945SA SHQ12156946SA SHQ12156947SA SHQ12156948SA SHQ12156949SA SHQ12156950SA SHQ12156951SA SHQ12156952SA SHQ12156953SA SHQ12156954SA SHQ12156955SA SHQ12156956SA SHQ12156957SA SHQ12156958SA SHQ12156959SA SHQ12156960SA SHQ12156961SA SHQ12156962SA SHQ12156963SA SHQ12156964SA SHQ12156965SA SHQ12156967SA SHQ12156968SA SHQ12156969SA SHQ12156970SA SHQ12156971SA SHQ12156972SA SHQ12156973SA SHQ12156974SA SHQ12156975SA SHQ12156977SA SHQ12156978SA SHQ12156979SA SHQ12156980SA SHQ12156981SA SHQ12156982SA SHQ12156983SA SHQ12156986SA SHQ12156987SA SHQ12156988SA SHQ12156989SA SHQ12156990SA SHQ12156991SA SHQ12156992SA SHQ12156993SA SHQ12156994SA SHQ12156995SA SHQ12156996SA SHQ12156997SA SHQ12156998SA SHQ12156999SA SHQ12157000SA SHQ12157001SA SHQ12157004SA SHQ12157005SA SHQ12157031SA SHQ12157032SA SHQ12157033SA SHQ12157034SA SHQ12157035SA SHQ12157036SA SHQ12157037SA SHQ12157039SA SHQ12157040SA SHQ12157041SA SHQ12157042SA SHQ12157043SA SHQ12157044SA SHQ12157045SA SHQ12157046SA SHQ12157047SA SHQ12157048SA SHQ12157051SA SHQ12157052SA SHQ12157053SA SHQ12157055SA SHQ12167057SA SHQ12167059SA SHQ12167060SA SHQ12167061SA SHQ12167065SA SHQ12167069SA SHQ12167070SA SHQ12167075SA SHQ12167076SA SHQ12167078SA SHQ12167083SA SHQ12167086SA SHQ12167087SA SHQ12167092SA SHQ12167093SA SHQ12167094SA SHQ12167095SA SHQ12167097SA SHQ12167098SA SHQ12167101SA SHQ12167103SA SHQ12167107SA SHQ12167108SA SHQ12167109SA SHQ12167110SA SHQ12167111SA SHQ12167112SA SHQ12167113SA SHQ12167115SA SHQ12167116SA SHQ12167117SA SHQ12167118SA SHQ12167119SA SHQ12167120SA SHQ12167121SA SHQ12167122SA SHQ12167123SA SHQ12167124SA SHQ12167125SA SHQ12167126SA SHQ12167128SA SHQ12167129SA SHQ12167130SA SHQ12167131SA SHQ12167132SA SHQ12167133SA SHQ12167134SA SHQ12167135SA SHQ12167136SA SHQ12167137SA SHQ12167139SA SHQ12167140SA SHQ12167142SA SHQ12167143SA SHQ12167144SA SHQ12167145SA SHQ12167146SA SHQ12167147SA SHQ12167148SA SHQ12167149SA SHQ12167150SA SHQ12167151SA SHQ12167152SA SHQ12167153SA SHQ12167154SA SHQ12167155SA SHQ12167156SA SHQ12167157SA SHQ12167158SA SHQ12167164SA SHQ12167165SA SHQ12167173SA SHQ12167175SA SHQ12167176SA SHQ12167178SA SHQ12167179SA SHQ12167180SA SHQ12167181SA SHQ12167182SA SHQ12167183SA SHQ12167184SA SHQ12167185SA SHQ12167186SA SHQ12167187SA SHQ12167188SA SHQ12167189SA SHQ12167190SA SHQ12167192SA SHQ12167193SA SHQ12167195SA SHQ12167196SA SHQ12167198SA SHQ12167201SA SHQ12167202SA SHQ12167203SA SHQ12167204SA SHQ12167205SA SHQ12167206SA SHQ12167207SA SHQ12167208SA SHQ12167209SA SHQ12167210SA SHQ12167211SA SHQ12167212SA SHQ12167213SA SHQ12167214SA SHQ12167216SA SHQ12167223SA SHQ12167224SA SHQ12167225SA SHQ12167226SA SHQ12167228SA SHQ12167229SA SHQ12167231SA SHQ12167232SA SHQ12167233SA SHQ12167234SA SHQ12167235SA SHQ12167236SA SHQ12173170GA SHQ12173208GA SHQ12173210GA SHQ12173214GA SHQ12173216GA SHQ12173222GA SHQ12173229GA SHQ12173252GA SHQ12173253GA SHQ12177240SA SHQ12177242SA SHQ12177246SA SHQ12177247SA SHQ12177249SA SHQ12177251SA SHQ12177254SA SHQ12177255SA SHQ12177261SA SHQ12177266SA SHQ12177267SA SHQ12177268SA SHQ12177269SA SHQ12177272SA SHQ12177274SA SHQ12177276SA SHQ12177280SA SHQ12177282SA SHQ12177285SA SHQ12177288SA SHQ12177294SA SHQ12177296SA SHQ12177297SA SHQ12177300SA SHQ12177302SA SHQ12177304SA SHQ12177305SA SHQ12177307SA SHQ12177308SA SHQ12177309SA SHQ12177311SA SHQ12177313SA SHQ12177314SA SHQ12177315SA SHQ12177316SA SHQ12177317SA SHQ12177318SA SHQ12177319SA SHQ12177321SA SHQ12177322SA SHQ12177323SA SHQ12177324SA SHQ12177325SA SHQ12177327SA SHQ12177328SA SHQ12177329SA SHQ12177330SA SHQ12177331SA SHQ12177332SA SHQ12177333SA SHQ12177334SA SHQ12177335SA SHQ12177336SA SHQ12177338SA SHQ12177339SA SHQ12177340SA SHQ12177342SA SHQ12177343SA SHQ12177344SA SHQ12177345SA SHQ12177346SA SHQ12177347SA SHQ12177348SA SHQ12177349SA SHQ12177350SA SHQ12177351SA SHQ12177352SA SHQ12177353SA SHQ12177354SA SHQ12177355SA SHQ12177356SA SHQ12177357SA SHQ12177358SA SHQ12177359SA SHQ12177360SA SHQ12177361SA SHQ12177362SA SHQ12177363SA SHQ12177364SA SHQ12177365SA SHQ12177366SA SHQ12177367SA SHQ12177368SA SHQ12177369SA SHQ12177371SA SHQ12177376SA SHQ12177378SA SHQ12177382SA SHQ12177387SA SHQ12177388SA SHQ12177390SA SHQ12177392SA SHQ12177393SA SHQ12177394SA SHQ12177397SA SHQ12177398SA SHQ12177400SA SHQ12177402SA SHQ12177403SA SHQ12177404SA SHQ12177405SA SHQ12177407SA SHQ12177408SA SHQ12177409SA SHQ12183279GA SHQ12183286GA SHQ12183291GA SHQ12187414SA SHQ12187415SA SHQ12187418SA SHQ12187422SA SHQ12187430SA SHQ12187433SA SHQ12187434SA SHQ12187436SA SHQ12187437SA SHQ12187439SA SHQ12187440SA SHQ12187441SA SHQ12187442SA SHQ12187443SA SHQ12187444SA SHQ12187446SA SHQ12187447SA SHQ12187448SA SHQ12187449SA SHQ12187450SA SHQ12187451SA SHQ12187452SA SHQ12187454SA SHQ12187455SA SHQ12187456SA SHQ12187457SA SHQ12187458SA SHQ12187459SA SHQ12187460SA SHQ12187462SA SHQ12187463SA SHQ12187464SA SHQ12187466SA SHQ12187467SA SHQ12187468SA SHQ12187469SA SHQ12187470SA SHQ12187471SA SHQ12187472SA SHQ12187474SA SHQ12187477SA SHQ12187481SA SHQ12187482SA SHQ12187483SA SHQ12187484SA SHQ12187485SA SHQ12187486SA SHQ12187487SA SHQ12187492SA SHQ12187494SA SHQ12187497SA SHQ12187498SA SHQ12187499SA SHQ12187503SA SHQ12187504SA SHQ12187505SA SHQ12187506SA SHQ12187507SA SHQ12187509SA SHQ12187510SA SHQ12187511SA SHQ12187512SA SHQ12187514SA SHQ12187515SA SHQ12187516SA SHQ12187517SA SHQ12187518SA SHQ12187519SA SHQ12187520SA SHQ12187521SA SHQ12187522SA SHQ12187523SA SHQ12187524SA SHQ12187525SA SHQ12187527SA SHQ12187529SA SHQ12187531SA SHQ12187535SA SHQ12187536SA SHQ12187538SA SHQ12187539SA SHQ12187540SA SHQ12187541SA SHQ12187543SA SHQ12187544SA SHQ12187545SA SHQ12187546SA SHQ12187547SA SHQ12187548SA SHQ12187549SA SHQ12187550SA SHQ12187551SA SHQ12187552SA SHQ12187553SA SHQ12193341GA SHQ12193342GA SHQ12193343GA SHQ12193344GA SHQ12193345GA SHQ12193347GA SHQ12197557SA SHQ12197562SA SHQ12197564SA SHQ12197565SA SHQ12197568SA SHQ12197569SA SHQ12197570SA SHQ12197571SA SHQ12197572SA SHQ12197574SA SHQ12197579SA SHQ12197583SA SHQ12197584SA SHQ12197585SA SHQ12197587SA SHQ12197588SA SHQ12197591SA SHQ12197592SA SHQ12197594SA SHQ12197595SA SHQ12197596SA SHQ12197597SA SHQ12197598SA SHQ12197599SA SHQ12197600SA SHQ12197601SA SHQ12197602SA SHQ12197603SA SHQ12197604SA SHQ12197605SA SHQ12197606SA SHQ12197607SA SHQ12197609SA SHQ12197610SA SHQ12197611SA SHQ12197612SA SHQ12197613SA SHQ12197614SA SHQ12197627SA SHQ12197628SA SHQ12197629SA SHQ12197630SA SHQ12197631SA SHQ12197632SA SHQ12197633SA SHQ12197634SA SHQ12197635SA SHQ12197636SA SHQ12197637SA SHQ12197638SA SHQ12197639SA SHQ12197640SA SHQ12197641SA SHQ12197642SA SHQ12197643SA SHQ12197644SA SHQ12197645SA SHQ12197646SA SHQ12197647SA SHQ12197648SA SHQ12197649SA SHQ12197650SA SHQ12197651SA SHQ12197652SA SHQ12197653SA SHQ12197654SA SHQ12197656SA SHQ12197658SA SHQ12197659SA SHQ12197660SA SHQ12197661SA SHQ12197662SA SHQ12197663SA SHQ12197664SA SHQ12197665SA SHQ12197666SA SHQ12197667SA SHQ12197668SA SHQ12197669SA SHQ12197670SA SHQ12197671SA SHQ12197672SA SHQ12197673SA SHQ12197674SA SHQ12197675SA SHQ12197676SA SHQ12197677SA SHQ12197678SA SHQ12197679SA SHQ12197680SA SHQ12197681SA SHQ12197682SA SHQ12197683SA SHQ12197684SA SHQ12197685SA SHQ12197686SA SHQ12197687SA SHQ12197688SA SHQ12197689SA SHQ12197690SA SHQ12197691SA SHQ12197692SA SHQ12197693SA SHQ12197694SA SHQ12207710SA SHQ12207711SA SHQ12207712SA SHQ12207715SA SHQ12207717SA SHQ12207718SA SHQ12207719SA SHQ12207720SA SHQ12207721SA SHQ12207722SA SHQ12207723SA SHQ12207724SA SHQ12207725SA SHQ12207726SA SHQ12207727SA SHQ12207728SA SHQ12207729SA SHQ12207730SA SHQ12207731SA SHQ12207732SA SHQ12207733SA SHQ12207734SA SHQ12207735SA SHQ12207736SA SHQ12207737SA SHQ12207738SA SHQ12207739SA SHQ12207740SA SHQ12207741SA SHQ12207742SA SHQ12207743SA SHQ12207744SA SHQ12207745SA SHQ12207746SA SHQ12207748SA SHQ12207749SA SHQ12207750SA SHQ12207771SA SHQ12207772SA SHQ12207773SA SHQ12207774SA SHQ12207775SA SHQ12207776SA SHQ12207777SA SHQ12207778SA SHQ12207779SA SHQ12207780SA SHQ12207781SA SHQ12207782SA SHQ12207783SA SHQ12207784SA SHQ12207785SA SHQ12207786SA SHQ12207790SA SHQ12207791SA SHQ12207793SA SHQ12207794SA SHQ12207795SA SHQ12207796SA SHQ12207797SA SHQ12207799SA SHQ12207801SA SHQ12207805SA SHQ12207806SA SHQ12207807SA SHQ12207809SA SHQ12207810SA SHQ12207813SA SHQ12207814SA SHQ12207815SA SHQ12207817SA SHQ12207819SA SHQ12207821SA SHQ12207822SA SHQ12207823SA SHQ12207825SA SHQ12207827SA SHQ12207828SA SHQ12207830SA SHQ12207831SA SHQ12207832SA SHQ12207833SA SHQ12207835SA SHQ12207836SA SHQ12207839SA SHQ12217870SA SHQ12217871SA SHQ12217872SA SHQ12217874SA SHQ12217875SA SHQ12217876SA SHQ12217878SA SHQ12217879SA SHQ12217880SA SHQ12217881SA SHQ12217883SA SHQ12217885SA SHQ12217886SA SHQ12217888SA SHQ12217889SA SHQ12217894SA SHQ12217903SA SHQ12217906SA SHQ12217908SA SHQ12217909SA SHQ12217910SA SHQ12217916SA SHQ12217917SA SHQ12217922SA SHQ12217923SA SHQ12217924SA SHQ12217926SA SHQ12217927SA SHQ12217929SA SHQ12217930SA SHQ12217931SA SHQ12217933SA SHQ12217934SA SHQ12217935SA SHQ12217936SA SHQ12217937SA SHQ12217938SA SHQ12217939SA SHQ12217942SA SHQ12217943SA SHQ12217945SA SHQ12217946SA SHQ12217947SA SHQ12217948SA SHQ12217950SA SHQ12217951SA SHQ12217952SA SHQ12217956SA SHQ12217965SA SHQ12217972SA SHQ12217973SA SHQ12217978SA SHQ12217981SA SHQ12217985SA SHQ12217986SA SHQ12217987SA SHQ12217992SA SHQ12217994SA SHQ12217995SA SHQ12217996SA SHQ12217997SA SHQ12217998SA SHQ12218000SA SHQ12218005SA SHQ12218006SA SHQ12218007SA SHQ12218008SA SHQ12218012SA SHQ12218013SA SHQ12218015SA SHQ12218017SA SHQ12218023SA SHQ12218024SA SHQ12218025SA SHQ12218026SA SHQ12218028SA SHQ12218030SA SHQ12218031SA SHQ12218032SA SHQ12218033SA SHQ12218034SA SHQ12218036SA SHQ12218037SA SHQ12218038SA SHQ12218039SA SHQ12218041SA SHQ12218042SA SHQ12218043SA SHQ12218044SA SHQ12218045SA SHQ12218046SA SHQ12218047SA SHQ12218049SA SHQ12218051SA SHQ12218052SA SHQ12228100SA SHQ12228106SA SHQ12228109SA SHQ12228110SA SHQ12228112SA SHQ12228113SA SHQ12228114SA SHQ12228117SA SHQ12228128SA SHQ12228130SA SHQ12228131SA SHQ12228132SA SHQ12228133SA SHQ12228135SA SHQ12228138SA SHQ12228139SA SHQ12228140SA SHQ12228141SA SHQ12228142SA SHQ12228148SA SHQ12228153SA SHQ12228154SA SHQ12228155SA SHQ12228156SA SHQ12228158SA SHQ12228160SA SHQ12228161SA SHQ12228162SA SHQ12228163SA SHQ12228166SA SHQ12228167SA SHQ12228169SA SHQ12228170SA SHQ12228176SA SHQ12228178SA SHQ12228181SA SHQ12228182SA SHQ12228193SA SHQ12228195SA SHQ12228196SA SHQ12228197SA SHQ12228198SA SHQ12228201SA SHQ12228205SA SHQ12228206SA SHQ12228209SA SHQ12228210SA SHQ12228211SA SHQ12228212SA SHQ12228213SA SHQ12228214SA SHQ12228216SA SHQ12228217SA SHQ12228218SA SHQ12228219SA SHQ12228222SA SHQ12228230SA SHQ12228233SA SHQ12228234SA SHQ12228245SA SHQ12228246SA SHQ12228247SA SHQ12228248SA SHQ12228249SA SHQ12228278SA SHQ12238280SA SHQ12238281SA SHQ12238282SA SHQ12238283SA SHQ12238284SA SHQ12238286SA SHQ12238287SA SHQ12238289SA SHQ12238293SA SHQ12238295SA SHQ12238296SA SHQ12238298SA SHQ12238299SA SHQ12238300SA SHQ12238301SA SHQ12238302SA SHQ12238303SA SHQ12238304SA SHQ12238305SA SHQ12238306SA SHQ12238307SA SHQ12238309SA SHQ12238311SA SHQ12238314SA SHQ12238316SA SHQ12238317SA SHQ12238318SA SHQ12238319SA SHQ12238323SA SHQ12238325SA SHQ12238330SA SHQ12238333SA SHQ12238334SA SHQ12238335SA SHQ12238336SA SHQ12238339SA SHQ12238347SA SHQ12238350SA SHQ12238352SA SHQ12238353SA SHQ12238354SA SHQ12238355SA SHQ12238356SA SHQ12238357SA SHQ12238358SA SHQ12238359SA SHQ12238366SA SHQ12238369SA SHQ12238388SA SHQ12238390SA SHQ12238392SA SHQ12238393SA SHQ12238395SA SHQ12238397SA SHQ12238398SA SHQ12238399SA SHQ12238400SA SHQ12238401SA SHQ12238402SA SHQ12238403SA SHQ12238404SA SHQ12238405SA SHQ12238406SA SHQ12238408SA SHQ12238409SA SHQ12238410SA SHQ12238413SA SHQ12238415SA SHQ12238416SA SHQ12238417SA SHQ12238418SA SHQ12238419SA SHQ12238420SA SHQ12238421SA SHQ12238422SA SHQ12238423SA SHQ12238424SA SHQ12238425SA SHQ12238426SA SHQ12238427SA SHQ12238428SA SHQ12238429SA SHQ12238430SA SHQ12238431SA SHQ12238432SA SHQ12238433SA SHQ12238434SA SHQ12238435SA SHQ12238437SA SHQ12238438SA SHQ12238439SA SHQ12238440SA SHQ12238441SA SHQ12238442SA SHQ12238443SA SHQ12238444SA SHQ12238445SA SHQ12238446SA SHQ12238447SA SHQ12238448SA SHQ12238449SA SHQ12238450SA SHQ12238451SA SHQ12238452SA SHQ12238453SA SHQ12238454SA SHQ12238455SA SHQ12238456SA SHQ12238457SA SHQ12238458SA SHQ12238459SA SHQ12238460SA SHQ12238461SA SHQ12238462SA SHQ12238463SA SHQ12238464SA SHQ12238466SA SHQ12238467SA SHQ12238468SA SHQ12238470SA SHQ12238471SA SHQ12238473SA SHQ12238475SA SHQ12238476SA SHQ12238477SA SHQ12238478SA SHQ12238479SA SHQ12238480SA SHQ12238481SA SHQ12238482SA SHQ12238484SA SHQ12238485SA SHQ12238486SA SHQ12238487SA SHQ12238490SA SHQ12238491SA SHQ12238492SA SHQ12238493SA SHQ12238494SA SHQ12238495SA SHQ12238496SA SHQ12238497SA SHQ12238498SA SHQ12238499SA SHQ12238500SA SHQ12238501SA SHQ12238502SA SHQ12238503SA SHQ12238504SA SHQ12238505SA SHQ12238506SA SHQ12238507SA SHQ12238508SA SHQ12238509SA SHQ12238510SA SHQ12238511SA SHQ12238513SA SHQ12238514SA SHQ12238515SA SHQ12238516SA SHQ12238517SA SHQ12238519SA SHQ12238520SA SHQ12238521SA SHQ12238522SA SHQ12238523SA SHQ12238524SA SHQ12238526SA SHQ12238527SA SHQ12238528SA SHQ12238530SA SHQ12238531SA SHQ12238532SA SHQ12238533SA SHQ12248534SA SHQ12248535SA SHQ12248536SA SHQ12248537SA SHQ12248538SA SHQ12248539SA SHQ12248540SA SHQ12248541SA SHQ12248542SA SHQ12248543SA SHQ12248544SA SHQ12248545SA SHQ12248546SA SHQ12248547SA SHQ12248548SA SHQ12248549SA SHQ12248550SA SHQ12248551SA SHQ12248552SA SHQ12248553SA SHQ12248554SA SHQ12248555SA SHQ12248556SA SHQ12248557SA SHQ12248558SA SHQ12248559SA SHQ12248560SA SHQ12248561SA SHQ12248562SA SHQ12248563SA SHQ12248564SA SHQ12248565SA SHQ12248566SA SHQ12248567SA SHQ12248568SA SHQ12248569SA SHQ12248570SA SHQ12248571SA SHQ12248572SA SHQ12248573SA SHQ12248574SA SHQ12248576SA SHQ12248579SA SHQ12248584SA SHQ12248589SA SHQ12248597SA SHQ12248598SA SHQ12248599SA SHQ12248600SA SHQ12248601SA SHQ12248602SA SHQ12248603SA SHQ12248605SA SHQ12248606SA SHQ12248607SA SHQ12248608SA SHQ12248609SA SHQ12248610SA SHQ12248611SA SHQ12248612SA SHQ12248613SA SHQ12248614SA SHQ12248615SA SHQ12248616SA SHQ12248617SA SHQ12248618SA SHQ12248619SA SHQ12248620SA SHQ12248621SA SHQ12248622SA SHQ12248623SA SHQ12248624SA SHQ12248625SA SHQ12248626SA SHQ12248627SA SHQ12248628SA SHQ12248629SA SHQ12248630SA SHQ12248631SA SHQ12248632SA SHQ12248633SA SHQ12248634SA SHQ12248635SA SHQ12248636SA SHQ12248637SA SHQ12248638SA SHQ12248639SA SHQ12248640SA SHQ12248641SA SHQ12248642SA SHQ12248643SA SHQ12248644SA SHQ12248646SA SHQ12248648SA SHQ12248649SA SHQ12248650SA SHQ12248651SA SHQ12248652SA SHQ12248653SA SHQ12248654SA SHQ12248655SA SHQ12248656SA SHQ12248657SA SHQ12248658SA SHQ12248659SA SHQ12248660SA SHQ12248662SA SHQ12248663SA SHQ12248664SA SHQ12248665SA SHQ12248666SA SHQ12248667SA SHQ12248668SA SHQ12248669SA SHQ12248670SA SHQ12248671SA SHQ12248672SA SHQ12248673SA SHQ12248676SA SHQ12248679SA SHQ12248680SA SHQ12248681SA SHQ12248683SA SHQ12248684SA SHQ12248686SA SHQ12248688SA SHQ12248689SA SHQ12248692SA SHQ12248693SA SHQ12248694SA SHQ12248696SA SHQ12248698SA SHQ12248700SA SHQ12248701SA SHQ12248704SA SHQ12248706SA SHQ12248707SA SHQ12248708SA SHQ12258737SA SHQ12258738SA SHQ12258739SA SHQ12258740SA SHQ12258741SA SHQ12258742SA SHQ12258743SA SHQ12258744SA SHQ12258745SA SHQ12258747SA SHQ12258749SA SHQ12258751SA SHQ12258752SA SHQ12258753SA SHQ12258754SA SHQ12258755SA SHQ12258756SA SHQ12258757SA SHQ12258758SA SHQ12258759SA SHQ12258760SA SHQ12258761SA SHQ12258762SA SHQ12258764SA SHQ12258767SA SHQ12258768SA SHQ12258769SA SHQ12258770SA SHQ12258771SA SHQ12258772SA SHQ12258773SA SHQ12258774SA SHQ12258775SA SHQ12258776SA SHQ12258777SA SHQ12258778SA SHQ12258779SA SHQ12258780SA SHQ12258781SA SHQ12258782SA SHQ12258783SA SHQ12258784SA SHQ12258785SA SHQ12258786SA SHQ12258787SA SHQ12258788SA SHQ12258789SA SHQ12258790SA SHQ12258791SA SHQ12258792SA SHQ12258794SA SHQ12258795SA SHQ12258796SA SHQ12258797SA SHQ12258798SA SHQ12258799SA SHQ12258800SA SHQ12258801SA SHQ12258802SA SHQ12258803SA SHQ12258804SA SHQ12258805SA SHQ12258806SA SHQ12258807SA SHQ12258808SA SHQ12258809SA SHQ12258810SA SHQ12258811SA SHQ12258812SA SHQ12258813SA SHQ12258814SA SHQ12258815SA SHQ12258816SA SHQ12258817SA SHQ12258818SA SHQ12258819SA SHQ12258820SA SHQ12258821SA SHQ12258822SA SHQ12258823SA SHQ12258824SA SHQ12268839SA SHQ12268840SA SHQ12268841SA SHQ12268842SA SHQ12268843SA SHQ12268844SA SHQ12268845SA SHQ12268846SA SHQ12268847SA SHQ12268848SA SHQ12268849SA SHQ12268850SA SHQ12268851SA SHQ12268852SA SHQ12268853SA SHQ12268854SA SHQ12268855SA SHQ12268862SA SHQ12268863SA SHQ12268864SA SHQ12268865SA SHQ12268866SA SHQ12268867SA SHQ12268868SA SHQ12268869SA SHQ12268870SA SHQ12268871SA SHQ12268872SA SHQ12268873SA SHQ12268874SA SHQ12268875SA SHQ12268876SA SHQ12268877SA SHQ12268878SA SHQ12268879SA SHQ12268880SA SHQ12268881SA SHQ12268882SA SHQ12268883SA SHQ12268884SA SHQ12268885SA SHQ12268886SA SHQ12268887SA SHQ12268888SA SHQ12268889SA SHQ12268890SA SHQ12268891SA SHQ12268892SA SHQ12268893SA SHQ12268894SA SHQ12268895SA SHQ12268896SA SHQ12268897SA SHQ12268898SA SHQ12268899SA SHQ12268900SA SHQ12268901SA SHQ12268902SA SHQ12268903SA SHQ12268904SA SHQ12268905SA SHQ12268906SA SHQ12268907SA SHQ12268908SA SHQ12268909SA SHQ12268911SA SHQ12268912SA SHQ12268914SA SHQ12268915SA SHQ12268917SA SHQ12268920SA SHQ12268921SA SHQ12268922SA SHQ12268923SA SHQ12268924SA SHQ12268925SA SHQ12268926SA SHQ12268927SA SHQ12268928SA SHQ12268929SA SHQ12268930SA SHQ12268931SA SHQ12268932SA SHQ12268933SA SHQ12278936SA SHQ12278940SA SHQ12278942SA SHQ12278946SA SHQ12278950SA SHQ12278952SA SHQ12278953SA SHQ12278954SA SHQ12278955SA SHQ12278958SA SHQ12278961SA SHQ12278962SA SHQ12278967SA SHQ12278968SA SHQ12279004SA SHQ12279007SA SHQ12279008SA SHQ12279009SA SHQ12289033SA SHQ12289034SA SHQ12289035SA SHQ12289036SA SHQ12289037SA SHQ12289038SA SHQ12289039SA SHQ12289040SA SHQ12289041SA SHQ12289042SA SHQ12289043SA SHQ12289044SA SHQ12289045SA SHQ12289046SA SHQ12289047SA SHQ12289048SA SHQ12289049SA SHQ12289050SA SHQ12289051SA SHQ12289052SA SHQ12289053SA SHQ12289054SA SHQ12289061SA SHQ12289067SA SHQ12289069SA SHQ12289070SA SHQ12289075SA SHQ12289076SA SHQ12289077SA SHQ12289084SA SHQ12289088SA SHQ12289093SA SHQ12289101SA SHQ12289111SA SHQ12289113SA SHQ12289116SA SHQ12289117SA SHQ12289120SA SHQ12289151SA SHQ12289152SA SHQ12289153SA SHQ12289154SA SHQ12289155SA SHQ12289156SA SHQ12289157SA SHQ12289161SA SHQ12289169SA SHQ12289172SA SHQ12289173SA SHQ12289174SA SHQ12289175SA SHQ12289181SA SHQ12289183SA SHQ12289184SA SHQ12289185SA SHQ12289186SA SHQ12289187SA SHQ12289189SA SHQ12289190SA SHQ12289191SA SHQ12289192SA SHQ12289193SA SHQ12289194SA SHQ12289195SA SHQ12289196SA SHQ12289197SA SHQ12289199SA SHQ12289200SA SHQ12289201SA SHQ12289202SA SHQ12289203SA SHQ12289207SA SHQ12289208SA SHQ12289209SA SHQ12289212SA SHQ12289213SA SHQ12289215SA SHQ12289216SA SHQ12289221SA SHQ12289225SA SHQ12289227SA SHQ12289229SA SHQ12289230SA SHQ12289233SA SHQ12289234SA SHQ12289253SA SHQ12299255SA SHQ12299261SA SHQ12299264SA SHQ12299265SA SHQ12299269SA SHQ12299271SA SHQ12299273SA SHQ12299276SA SHQ12299278SA SHQ12299279SA SHQ12299287SA SHQ12299290SA SHQ12299292SA SHQ12299293SA SHQ12299295SA SHQ12299297SA SHQ12299298SA SHQ12299299SA SHQ12299300SA SHQ12299301SA SHQ12299305SA SHQ12299306SA SHQ12299308SA SHQ12299309SA SHQ12299311SA SHQ12299315SA SHQ12299318SA SHQ12299319SA SHQ12299320SA SHQ12299321SA SHQ12299324SA SHQ12299333SA SHQ12299347SA SHQ12299348SA SHQ12299349SA SHQ12299350SA SHQ12299351SA SHQ12299352SA SHQ12299353SA SHQ12299354SA SHQ12299355SA SHQ12299356SA SHQ12299357SA SHQ12299358SA SHQ12299359SA SHQ12299360SA SHQ12299361SA SHQ12299362SA SHQ12299364SA SHQ12299365SA SHQ12299366SA SHQ12299370SA SHQ12299371SA SHQ12299372SA SHQ12299374SA SHQ12299375SA SHQ12299379SA SHQ12299385SA SHQ12299386SA SHQ12299387SA SHQ12299392SA SHQ12299394SA SHQ12299395SA SHQ12299397SA SHQ12299398SA SHQ12299399SA SHQ12299400SA SHQ12299401SA SHQ12299446SA SHQ12299448SA SHQ12299449SA SHQ12299453SA SHQ12299456SA SHQ12299460SA SHQ12309467SA SHQ12309468SA SHQ12309469SA SHQ12309470SA SHQ12309472SA SHQ12309474SA SHQ12309475SA SHQ12309476SA SHQ12309478SA SHQ12309479SA SHQ12309480SA SHQ12309481SA SHQ12309482SA SHQ12309484SA SHQ12309485SA SHQ12309487SA SHQ12309488SA SHQ12309493SA SHQ12309495SA SHQ12309497SA SHQ12309499SA SHQ12309504SA SHQ12309512SA SHQ12309530SA SHQ12309533SA SHQ12309537SA SHQ12309538SA SHQ12309539SA SHQ12309542SA SHQ12309547SA SHQ12309548SA SHQ12309550SA SHQ12309551SA SHQ12309552SA SHQ12309553SA SHQ12309554SA SHQ12309555SA SHQ12309556SA SHQ12309557SA SHQ12309558SA SHQ12309589SA SHQ12309726SA SHQ12309727SA SHQ12309728SA SHQ12309729SA SHQ12309730SA SHQ12309732SA SHQ12309734SA SHQ12309735SA SHQ12299404SA SHQ12289223SA SHQ12320364SA SHQ12320367SA SHQ12320368SA SHQ12320366SA SHQ12319819SA SD011123325SA SD011123355SA SD011123350SA SHQ12319981SA SHQ12361192SA SHQ12361191SA SHQ12330657SA SHQ12309736SA SHQ12309769SA SHQ12371454SA SHQ12391764SA SHQ12391787SA SHQ12310034SA SHQ12310033SA SHQ12258732SA SHQ12319893SA SHQ12330523SA SHQ12310112SA SHQ12310108SA SHQ12310109SA SHQ12289177SA SHQ12320138SA SHQ12310120SA SHQ12320127SA SHQ12310114SA SHQ12310111SA SHQ12320140SA SHQ12320122SA SHQ12320142SA SHQ12320136SA SHQ12320220SA SHQ12310089SA SHQ12320249SA SHQ12320146SA SHQ12310119SA SHQ12330645SA SHQ12381586SA SHQ12330468SA SHQ12330467SA SHQ12330453SA SHQ12320370SA SHQ12320361SA SHQ12320363SA SHQ12320358SA SHQ12320360SA SHQ12320213SA SHQ12320362SA SHQ12320359SA SHQ12320396SA SHQ12320394SA SHQ12320401SA SHQ12320393SA SHQ12320173SA SHQ12320170SA SHQ12320172SA SHQ12320162SA SHQ12289074SA SHQ12289073SA SHQ12320205SA SHQ12320212SA SHQ12320203SA SHQ12320216SA SHQ12320198SA SHQ12320200SA SHQ12320185SA SHQ12320182SA SHQ12320154SA SHQ12320174SA SHQ12320206SA SHQ12320168SA SHQ12320169SA SHQ12320175SA SHQ12320165SA SHQ12320155SA SHQ12320402SA SHQ12320395SA SHQ12320183SA SHQ12320178SA SHQ12320210SA SHQ12320209SA SHQ12320207SA SHQ12320199SA SHQ12320217SA SHQ12320211SA SHQ12320218SA SHQ12320153SA SHQ12310007SA SHQ12320176SA SHQ12320157SA SHQ12320208SA SHQ12320215SA SHQ12299382SA SHQ12299390SA SHQ12319812SA SHQ12319815SA SHQ12319798SA SHQ12319792SA SHQ12319791SA SHQ12319773SA SHQ12340875SA SHQ12340874SA SHQ12299280SA SHQ12299260SA SHQ12299254SA SHQ12309598SA SHQ12309599SA SHQ12309596SA SHQ12309562SA SHQ12309597SA SHQ12309564SA SHQ12309560SA SHQ12309588SA SHQ12309559SA SHQ12309566SA SHQ12309565SA SHQ12309563SA SHQ12309594SA SHQ12289080SA SHQ12371431SA SHQ12330692SA SHQ12309669SA SHQ12309586SA SHQ12309584SA SHQ12289219SA SHQ12289231SA SHQ12289232SA SHQ12309738SA SHQ12299402SA SHQ12319775SA SHQ12309771SA SHQ12319779SA SHQ12319808SA SHQ12319781SA SHQ12319782SA SHQ12309757SA SHQ12309752SA SHQ12309759SA SHQ12309762SA SHQ12309743SA SHQ12309742SA SHQ12319800SA SHQ12309761SA SHQ12319805SA SHQ12319799SA SHQ12309740SA SHQ12299403SA SHQ12319793SA SHQ12299405SA SHQ12319777SA SHQ12299373SA SHQ12299391SA SHQ12299377SA SHQ12299369SA SHQ12299389SA SHQ12299396SA SHQ12309507SA SHQ12309483SA SHQ12309489SA SHQ12309501SA SHQ12309510SA SHQ12309471SA SHQ12309492SA SHQ12309491SA SHQ12309520SA SHQ12309502SA SHQ12309517SA SHQ12309508SA SHQ12309518SA SHQ12309494SA SHQ12309473SA SHQ12309498SA SHQ12309506SA SHQ12309514SA SHQ12309516SA SHQ12309503SA SHQ12309513SA SHQ12309490SA SHQ12309509SA SHQ12309519SA SHQ12299326SA SHQ12299317SA SHQ12319786SA SHQ12319774SA SHQ12319776SA SHQ12319797SA SHQ12319795SA SHQ12319814SA SHQ12319788SA SHQ12319816SA SHQ12319778SA SHQ12319783SA SHQ12319789SA SHQ12319817SA SHQ12319787SA SHQ12319820SA SHQ12319818SA SHQ12319796SA SHQ12319790SA SHQ12319806SA SHQ12299376SA SHQ12299383SA SHQ12309511SA SHQ12299393SA SHQ12310005SA SHQ12320139SA SHQ12310057SA SHQ12310038SA SHQ12310016SA SHQ12310017SA SHQ12310018SA SHQ12310006SA SHQ12278937SA SHQ12319999SA SHQ12309688SA SHQ12258765SA SHQ12167199SA SHQ12310077SA SHQ12187513SA SHQ12309654SA SHQ12258748SA SHQ12309721SA SHQ12309593SA SHQ12309569SA SHQ12319982SA SHQ12309576SA SHQ12309580SA SHQ12309595SA SHQ12310019SA SHQ12310013SA SHQ12320158SA SHQ12320167SA SHQ12320181SA SHQ12320163SA SHQ12320184SA SHQ12320166SA SHQ12320349SA SHQ12320348SA SHQ12310107SA SHQ12310101SA SHQ12310106SA SHQ12320328SA SHQ12320308SA SHQ12299381SA SHQ12299388SA SHQ12319996SA SHQ12320342SA SHQ12310044SA SHQ12320334SA SHQ12320336SA SHQ12310043SA SHQ12310047SA SHQ12299384SA SHQ12310022SA SHQ12320160SA SHQ12320156SA SHQ12320187SA SHQ12279021SA SHQ12279020SA SHQ12320413SA SHQ12320407SA SHQ12320406SA SHQ12320412SA SHQ12320232SA SHQ12320234SA SHQ12320254SA SHQ12320421SA SHQ12320418SA SHQ12320242SA SHQ12320422SA SHQ12330476SA SHQ12330449SA SHQ12330473SA SHQ12330466SA SHQ12330470SA SHQ12330474SA SHQ12330456SA SHQ12330465SA SHQ12320420SA SHQ12320335SA SHQ12320314SA SHQ12320248SA SHQ12319807SA SHQ12320313SA SHQ12320310SA SHQ12320311SA SHQ12319966SA SHQ12319979SA SHQ12319968SA SHQ12319989SA SHQ12319972SA SHQ12330513SA SHQ12330516SA SHQ12330514SA SHQ12319847SA SHQ12319873SA SHQ12330493SA SHQ12330494SA SHQ12310003SA SHQ12320147SA SHQ12320151SA SHQ12330499SA SHQ12320137SA SHQ12310118SA SHQ12320144SA SHQ12330497SA SHQ12330496SA SHQ12330495SA SHQ12320134SA SHQ12320356SA SHQ12320437SA SHQ12330526SA SHQ12330509SA SHQ12330527SA SHQ12330510SA SHQ12330507SA SHQ12330520SA SHQ12330519SA SHQ12330508SA SHQ12330511SA SHQ12330518SA SHQ12320141SA SHQ12320124SA SHQ12310117SA SHQ12320135SA SHQ12310121SA SHQ12320128SA SHQ12310110SA SHQ12310116SA SHQ12319930SA SHQ12319857SA SHQ12319859SA SHQ12319853SA SHQ12310030SA SHQ12320385SA SHQ12309653SA SHQ12320441SA SHQ12320130SA SHQ12320125SA SHQ12320145SA SHQ12320129SA SHQ12320131SA SHQ12330492SA SHQ12330591SA SHQ12320414SA SHQ12299459SA SHQ12299458SA SHQ12330585SA SHQ12320416SA SHQ12320415SA SHQ12330587SA SHQ12330586SA SHQ12320159SA SHQ12320189SA SHQ12320188SA SHQ12330604SA SHQ12330590SA SHQ12310099SA SHQ12319975SA SHQ12319988SA SHQ12319962SA SHQ12310096SA SHQ12310100SA SHQ12319980SA SHQ12310093SA SHQ12319986SA SHQ12320196SA SHQ12320190SA SHQ12320192SA SHQ12320193SA SHQ12320195SA SHQ12330575SA SHQ12320186SA SHQ12320191SA SHQ12320400SA SHQ12330501SA SHQ12320164SA SHQ12320177SA SHQ12330581SA SHQ12330573SA SHQ12330578SA SHQ12330571SA SHQ12330570SA SHQ12330569SA SHQ12330583SA SHQ12330577SA SHQ12330580SA SHQ12330579SA SHQ12330572SA SHQ12330522SA SHQ12320221SA SHQ12330739SA SHQ12330525SA SHQ12330512SA SHQ12330749SA SHQ12310091SA SHQ12330517SA SHQ12330529SA SHQ12330531SA SHQ12330521SA SHQ12330588SA SHQ12330568SA SHQ12330781SA SHQ12330498SA SHQ12310071SA SHQ12330532SA SHQ12320347SA SHQ12330574SA SHQ12320403SA SHQ12320411SA SHQ12320398SA SHQ12320399SA SHQ12320410SA SHQ12330582SA SHQ12320392SA SHQ12320405SA SHQ12320409SA SHQ12320408SA SHQ12320404SA SHQ12330596SA SHQ12330486SA SHQ12330487SA SHQ12330488SA SHQ12330485SA SHQ12330484SA SHQ12330490SA SHQ12319902SA SHQ12309768SA SHQ12330592SA SHQ12330524SA SHQ12330769SA SHQ12330763SA SHQ12330744SA SHQ12330772SA SHQ12330754SA SHQ12330747SA SHQ12330756SA SHQ12330777SA SHQ12330766SA SHQ12330771SA SHQ12330731SA SHQ12330741SA SHQ12330757SA SHQ12330770SA SHQ12320447SA SHQ12310074SA SHQ12320446SA SHQ12310076SA SHQ12310072SA SHQ12310087SA SHQ12310079SA SHQ12330458SA SHQ12330735SA SHQ12330762SA SHQ12330775SA SHQ12330576SA SHQ12330790SA SHQ12330505SA SHQ12330500SA SHQ12330504SA SHQ12330785SA SHQ12330502SA SHQ12167200SA SHQ12310078SA SHQ12309574SA SHQ12309577SA SHQ12340985SA SHQ12309571SA SHQ12310086SA SHQ12310088SA SHQ12330654SA SHQ12330680SA SHQ12340986SA SHQ12319987SA SHQ12340915SA SHQ12279027SA SHQ12340980SA SHQ12340950SA SHQ12340988SA SHQ12340956SA SHQ12330715SA SHQ12330606SA SHQ12330647SA SHQ12330641SA SHQ12330643SA SHQ12330608SA SHQ12330603SA SHQ12330702SA SHQ12330648SA SHQ12330664SA SHQ12330665SA SHQ12330663SA SHQ12330703SA SHQ12330649SA SHQ12330607SA SHQ12340864SA SHQ12330656SA SHQ12340888SA SHQ12340876SA SHQ12340859SA SHQ12340886SA SHQ12340885SA SHQ12340887SA SHQ12340882SA SHQ12340884SA SHQ12330601SA SHQ12330720SA SHQ12340878SA SHQ12330691SA SHQ12330544SA SHQ12330812SA SHQ12330808SA SHQ12330813SA SHQ12340890SA SHQ12340889SA SHQ12330809SA SHQ12330810SA SHQ12340894SA SHQ12340893SA SHQ12330814SA SHQ12340892SA SHQ12351049SA SHQ12351056SA SHQ12351042SA SHQ12340946SA SHQ12320236SA SHQ12340957SA SHQ12340063SA SHQ12340954SA SHQ12340955SA SHQ12340922SA SHQ12340960SA SHQ12340961SA SHQ12340945SA SHQ12340827SA SHQ12340953SA SHQ12330689SA SHQ12330535SA SHQ12351063SA SHQ12351054SA SHQ12351053SA SHQ12351041SA SHQ12351066SA SHQ12351062SA SHQ12351051SA SHQ12351088SA SHQ12351082SA SHQ12351080SA SHQ12351077SA SHQ12330786SA SHQ12330784SA SHQ12330789SA SHQ12330788SA SHQ12330783SA SHQ12309549SA SHQ12340901SA SHQ11190011SA SHQ11190006SA SHQ12330542SA SHQ12330688SA SHQ12330684SA SHQ12320375SA SHQ12320379SA SHQ12320386SA SHQ12320382SA SHQ12320383SA SHQ12320384SA SHQ12320377SA SHQ12320387SA SHQ12320381SA SHQ12330543SA SHQ12330681SA SHQ12320380SA SHQ12320369SA SHQ12320378SA SHQ12289081SA SHQ12289062SA SHQ12289162SA SHQ12361238SA SHQ12361251SA SHQ12341002SA SHQ12361258SA SHQ12361254SA SHQ12361240SA SHQ12341001SA SHQ12340989SA SHQ12340995SA SHQ12340908SA SHQ12340920SA SHQ12330768SA SHQ12340916SA SHQ12340914SA SHQ12340921SA SHQ12340917SA SHQ12340912SA SHQ12340909SA SHQ12340924SA SHQ12340918SA SHQ12340913SA SHQ12340948SA SHQ12340964SA SHQ12340966SA SHQ12340947SA SHQ12340828SA SHQ12340963SA SHQ12340967SA SHQ12340959SA SHQ12340870SA SHQ12340962SA SHQ12340829SA SHQ12340869SA SHQ12351064SA SHQ12340938SA SHQ12351024SA SHQ12351028SA SHQ12351030SA SHQ12351031SA SHQ12351022SA SHQ12351033SA SHQ12351027SA SHQ12351025SA SHQ12351021SA SHQ12351038SA SHQ12351029SA SHQ12351026SA SHQ12340943SA SHQ12340934SA SHQ12340942SA SHQ12340940SA SHQ12340944SA SHQ12340939SA SHQ12330761SA SHQ12330758SA SHQ12330767SA SHQ12330746SA SHQ12330776SA SHQ12330751SA SHQ12330753SA SHQ12330750SA SHQ12330745SA SHQ12330752SA SHQ12330759SA SHQ12330773SA SHQ12330733SA SHQ12330755SA SHQ12330732SA SHQ12330740SA SHQ12330736SA SHQ12330760SA SHQ12340958SA SHQ12340949SA SHQ12340975SA SHQ12340952SA SHQ12340997SA SHQ12351037SA SHQ12351055SA SHQ12340999SA SHQ12341000SA SHQ12340910SA SHQ12340919SA SHQ12340983SA SHQ12340971SA SHQ12340981SA SHQ12340968SA SHQ12340911SA SHQ12340907SA SHQ12361250SA SHQ12361262SA SHQ12361182SA SHQ12361133SA SHQ12361137SA SHQ12361136SA SHQ12361143SA SHQ12361139SA SHQ12341005SA SHQ12361253SA SHQ12361263SA SHQ12361249SA SHQ12361256SA SHQ12361243SA SHQ12361156SA SHQ12361176SA SHQ12361178SA SHQ12361160SA SHQ12361180SA SHQ12361177SA SHQ12361179SA SHQ12361183SA SHQ12361170SA SHQ12361167SA SHQ12361169SA SHQ12361168SA SHQ12361166SA SHQ12361158SA SHQ12361162SA SHQ12361159SA SHQ12361161SA SHQ12371320SA SHQ12371324SA SHQ12361260SA SHQ12371323SA SHQ12371316SA SHQ12371321SA SHQ12371318SA SHQ12330850SA SHQ12320444SA SHQ12320438SA SHQ12371304SA SHQ12341004SA SHQ12361141SA SHQ12361135SA SHQ12341006SA SHQ12310051SA SHQ12330709SA SHQ12361206SA SHQ12361205SA SHQ12361202SA SHQ12361199SA SHQ12361203SA SHQ12361204SA SHQ12361196SA SHQ12361193SA SHQ12361201SA SHQ12361157SA SHQ12361189SA SHQ12361187SA SHQ12361185SA SHQ12361181SA SHQ12361171SA SHQ12361188SA SHQ12361164SA SHQ12361155SA SHQ12319998SA SHQ12330678SA SHQ12330695SA SHQ12330714SA SHQ12361186SA SHQ12361190SA SHQ12351110SA SHQ12351114SA SHQ12351115SA SHQ12351081SA SHQ12351091SA SHQ12351019SA SHQ12351008SA SHQ12351013SA SHQ12351083SA SHQ12351079SA SHQ12351085SA SHQ12351086SA SHQ12351089SA SHQ12351087SA SHQ12351078SA SHQ12351076SA SHQ12351015SA SHQ12351012SA SHQ12351016SA SHQ12351011SA SHQ12351014SA SHQ12351017SA SHQ12351007SA SHQ12351010SA SHQ12351009SA SHQ12330638SA SHQ12351106SA SHQ12351095SA SHQ12351093SA SHQ12351098SA SHQ12351105SA SHQ12351103SA SHQ12351090SA SHQ12351097SA SHQ12351096SA SHQ12351092SA SHQ12351099SA SHQ12351102SA SHQ12351104SA SHQ12351108SA SHQ12351113SA SHQ12351112SA SHQ12351100SA SHQ12351111SA SHQ12351109SA SHQ12351084SA SHQ12361195SA SHQ12258733SA SHQ12361194SA SHQ12258734SA SHQ12258730SA SHQ12258731SA SHQ12330651SA SHQ12330632SA SHQ12330642SA SHQ12330660SA SHQ12330634SA SHQ12330635SA SHQ12330646SA SHQ12330637SA SHQ12330661SA SHQ12320427SA SHQ12320424SA SHQ12320435SA SHQ12320428SA SHQ12320429SA SHQ12320425SA SHQ12320434SA SHQ12320432SA SHQ12320431SA SHQ12320426SA SHQ12371380SA SHQ12371377SA SHQ12371379SA SHQ12371306SA SHQ12361154SA SHQ12371456SA SHQ12371378SA SHQ12371382SA SHQ12371384SA SHQ12371469SA SHQ12340883SA SHQ12340880SA SHQ12310053SA SHQ12371478SA SHQ12371476SA SHQ12371479SA SHQ12371481SA SHQ12371470SA SHQ12361153SA SHQ12371430SA SHQ12371433SA SHQ12371432SA SHQ12371451SA SHQ12371427SA SHQ12371424SA SHQ12371453SA SHQ12371483SA SHQ12371450SA SHQ12371425SA SHQ12371423SA SHQ12371434SA SHQ12371426SA SHQ12371455SA SHQ12371360SA SHQ12371449SA SHQ12371452SA SHQ12371466SA SHQ12371474SA SHQ12371467SA SHQ12371482SA SHQ12371490SA SHQ12340871SA SHQ12340872SA SHQ12371473SA SHQ12371477SA SHQ12371468SA SHQ12371472SA SHQ12371465SA SHQ12371492SA SHQ12371495SA SHQ12371498SA SHQ12371499SA SHQ12371504SA SHQ12371475SA SHQ12371493SA SHQ12371503SA SHQ12371500SA SHQ12371501SA SHQ12371497SA SHQ12371488SA SHQ12340867SA SHQ12340865SA SD011123324SA SD011123329SA SHQ12309523SA SHQ12371485SA SHQ12330677SA SHQ12330673SA SHQ12330712SA SHQ12330713SA SHQ12330669SA SHQ12299363SA SHQ12299367SA SD011050924SA SD011123327SA SD011122888SA SD011123301SA SD011122915SA SHQ12381551SA SHQ12381543SA SHQ12381553SA SHQ12381549SA SHQ12381536SA SHQ12381517SA SHQ12381588SA SHQ12381574SA SHQ12381582SA SHQ12381587SA SHQ12391867SA SHQ12320430SA SHQ12391874SA SHQ12391770SA SHQ12391801SA SHQ12371496SA SHQ12371484SA SHQ12082724GA SHQ12309524SA SHQ12381585SA SHQ12381578SA SHQ12381589SA SHQ12381581SA SHQ12381571SA SHQ12381579SA SHQ12381526SA SHQ12381520SA SHQ12381516SA SHQ12381524SA SHQ12381518SA SHQ12381528SA SHQ12381525SA SHQ12381519SA SHQ12381527SA SHQ12381515SA SHQ12381514SA SHQ12381521SA SHQ12381541SA SHQ12381529SA SHQ12381533SA SHQ12371471SA SHQ12381548SA SHQ12381552SA SHQ12381561SA SHQ12381558SA SHQ12381559SA SHQ12381556SA SHQ12381557SA SHQ12381560SA SHQ12381554SA SHQ12381523SA SHQ12381597SA SHQ12381609SA SHQ12381607SA SHQ12381604SA SHQ12381596SA SHQ12381594SA SHQ12381603SA SHQ12381600SA SHQ12381601SA SHQ12381612SA SHQ12381599SA SHQ12381591SA SHQ12381606SA SHQ12381590SA SHQ12381602SA SHQ12381608SA SHQ12371328SA SHQ12371331SA SHQ12371330SA SHQ12371339SA SHQ12371336SA SHQ12371335SA SHQ12371326SA SHQ12066186SA SHQ12066189SA SHQ12381611SA SHQ12381622SA SHQ12391640SA SHQ12391635SA SHQ12391632SA SHQ12391641SA SHQ12391638SA SHQ12391649SA SHQ12391626SA SHQ12391627SA SHQ12391653SA SHQ12391650SA SHQ12391656SA SHQ12391652SA SHQ12391634SA SHQ12391651SA SHQ12381621SA SHQ12391633SA SHQ12391642SA SHQ12391645SA SHQ12391630SA SHQ12391647SA SHQ12391643SA SHQ12391639SA SHQ12391685SA SHQ12391694SA SHQ12391682SA SHQ12391683SA SHQ12381598SA SHQ12381595SA SHQ12381605SA SHQ12381613SA SHQ12381593SA SHQ12391663SA SHQ12391664SA SHQ12391657SA SHQ12391631SA SHQ12391637SA SHQ12391636SA SHQ12391666SA SHQ12391673SA SHQ12391669SA SHQ12391646SA SHQ12391628SA SHQ12391659SA SHQ12391662SA SHQ12391665SA SHQ12391661SA SHQ12310049SA SHQ12371327SA SHQ12391769SA SHQ12391680SA SHQ12391781SA SHQ12391788SA SHQ12320433SA SHQ12381615SA SHQ12381618SA SHQ12371375SA SHQ12371383SA SHQ12381623SA SHQ12391767SA SHQ12391775SA SHQ12381617SA SHQ12381620SA SHQ12381624SA SHQ12391772SA SHQ12391768SA SHQ12391765SA SHQ12391691SA SHQ12391686SA SHQ12391693SA SHQ12391696SA SHQ12391692SA SHQ12391688SA SHQ12381610SA SHQ12391684SA SHQ12391689SA SHQ12391690SA SHQ12391695SA SHQ12391687SA SHQ12391786SA SHQ12391779SA SHQ12391798SA SHQ12391800SA SHQ12391805SA SHQ12391644SA SHQ12391806SA SHQ12391808SA SHQ12371494SA SHQ12391799SA SHQ12391811SA SHQ12391782SA SHQ12391763SA SHQ12391780SA SHQ12391777SA SHQ12391807SA SHQ12391796SA SHQ12391810SA SHQ12391802SA SHQ12391803SA SHQ12391793SA SHQ12391784SA SHQ12391785SA SHQ12391804SA SHQ12391734SA SHQ12391795SA SHQ12391790SA SHQ12391762SA SHQ12391794SA SHQ12391758SA SHQ12391759SA SHQ12391754SA SHQ12391761SA SHQ12391747SA SHQ12391751SA SHQ12391760SA SHQ12391756SA SHQ12391748SA SHQ12391755SA SHQ12391746SA SHQ12391757SA SHQ12391736SA SHQ12391726SA SHQ12391749SA SHQ12391737SA SHQ12391742SA SHQ12391743SA SHQ12391732SA SHQ12391750SA SHQ12391753SA SHQ12391739SA SHQ12391752SA SHQ12391725SA SHQ12391738SA SHQ12391715SA SHQ12391717SA SHQ12391719SA SHQ12391727SA SHQ12391716SA SHQ12391724SA SHQ12391722SA SHQ12391730SA SHQ12391723SA SHQ12391721SA SHQ12391731SA SHQ12391729SA SHQ12391733SA SHQ12391744SA SHQ12391720SA SHQ12391741SA SHQ12391745SA SHQ12391718SA SHQ12391735SA SHQ12391740SA SHQ12391728SA SHQ12340866SA SHQ12340868SA SHQ12309521SA SHQ12391773SA SHQ12299378SA SHQ12340873SA SHQ12340881SA SHQ12381508SA SHQ12371340SA SHQ12371332SA SHQ12371329SA SHQ12371337SA SHQ12391674SA SHQ12391655SA SHQ12391864SA SHQ12391866SA SHQ12391791SA SHQ12391865SA SHQ12391868SA SHQ12391792SA SHQ12391870SA SHQ12391809SA SHQ12391872SA SHQ12391871SA SHQ12371351SA SHQ12371334SA SHQ12391941SA SHQ12330748SA SHQ12330765SA SHQ12330743SA SD008077060GR SD008148506GR SD008419661GR SD008505497GR SD008516288GR SD009047586GR SD009047736GR SD009182612GR SD010470673GA SD010501732GA SD010501853GA SD010501981GA SD010522423GA SD011043164GA SD011043175GA SD011043176GA SD011083575GA SD011083576GA SD011083578GA SD011083580GA SD009078508GR SD009234165GR SD009316921GR SD009388954GR SD009399635GR SD009400086GR SD009409749GR SD009409754GR SD010095486GR SD010106292GR SD010366148GR SD009317065GR SD009399631GR SD008398663GR SD009223766GR SD009316922GR SD009317067GR SD009368553GR SD009409692GR SD009409760GR SD010106266GR SD010074690GR SD010366314GR SD011124191GA SD011124195GA SD011134324GR SD011144462GR SD008077125GR SD008494980GR SD008495011GR SD009244819GR SD009296289GR SD010127165GR SD010460133GR SD009119664GR SD009409943GR SD010095642GR SD010159032GR SD008077091GR SD008209593GR SD008398698GR SD008398701GR SD009037495GR SD009316913GR SD010127168GR SD010159028GR SD010159156GR SD010220641GR SD010459838GR SD010459870GR SD010459897GR SD010501866GRMKE SD007241907GR SD010116744GR SD010159068GR SD010481007GR SD010501860GR SD011134346GR SD011030671SR SD011040835SR SD011040877SR SD011040883SR SD011040905SR SD011050907SR SD011071822SR SD011112274SR SD011112746SR SD011112768SR SD011112770SR SD011143921SR SD011143926SR SD011143930SR SD011143932SR SD011143934SR SD011143946SR SD011143976SR SHQ11260534SR SD009316918GRMKE SD009016742GR SD009327363GR SD010148724GR SD010220658GR SD010430054SR SD010430059SR SD011051067SR SD011143826SR C4EH5806GR! SD010138481GR SD010138486SR SD011051043SR SD011061312SR SD011112642SR SD011112658SR SD011112670SR SD011122833SR SD011122855SR SD011122871SR SD011122998SRBOP SD011123031SR SD011133398SR SD011133399SR SD011133429SR SD011133577SR SD011133629SR SD011133630SR SHQ11240201SR SHQ11240208SR SHQ11240229SR SHQ11250315SR SD009317068GR SD009368560GR SD010138465GR SD010138471GR SD010138473GR SD010138477GR SD010138478GR SD010148551GR SD010148554GR SD010148659GR SD010148663GR SD010148664GR SD010148665GR SD010148666GR SD010148667GR SD010148669GR SHQ11230085SR SHQ11230090SR SHQ11240213SR SHQ11280719SR SD009141370GR SD010138474GR SD010138493GR SD011214883SR SHQ11240214SR SD010127807GR SD007231234GR SD009244555GR SD007387290GR SD007439616GR SD007439617GR SD007439622GR SD007439626GR SD007439630GR SD008356869GR SD010106287GR SBG06315854GR SBG06504719GR SBG07016573GR SBG07131798GR SBG07162973GR SD008388537GR SD009399640GR SD009399642GR SD010127742GR SD010127772GR SD010480120SR SD010500159SR SD010500167SR SD010500197SR SD010510304SR SD010510306SR SD010148696GR SD011051027SR SD011112778SR SHQ11464165SR SHQ11484871SR SHQ11484873SR SD009327365GR SD009327367GR SD010158815GR SD011112747SR SD011143935SR SD011143937SR SD011143964SR SD011143981SR SD011143984SR SHQ12228102SR SHQ12228105SR SD010127657GR SD010127717GR SD010148695GR SD010148702GR SHQ12228098SR SHQ12228099SR SHQ12228101SR SHQ12228103SR SHQ12228108SR SHQ12228116SR SD009233945GR SD010127737GR SD010138490GR SD010138495GR SD010148699GR SD010148701GR SD010148706GR SHQ12238332SR SD010127713GR SD010127735GR SD010148698GR SD010127642GR SD011134318GR SD009255208GR SD010127643GR SD010138284GR SD010138285GR SD010138288GR SD010138289GR SD010148713GR SD010148717GR SD010148720GR SD010148721GR SD010418612GR SD011124208GR SD011134240GR SD011144466GR SD010148710GR SD011020475SR SD011184446SR SD008429969GR SD009400101GR SD009409718GR SD009409719GR SD010127758GR SD010500185SR SD009285824GR SD009400106GR SD009400114GR SD009409772GR SD010127766GR SD010127769GR SD010138428GR SD008377849GR SD008421129GR SD008421144GR SD008429857GR SD011204705SR SD011225019SR SD009172476GR SD010313836GR SD010313839GR SD010313848GR SD010314035GR SD010314164GR SD010491548GR SD010491555GR SD010501864GR SD011215061GR SHQ11433443SR SHQ11433449SR SHQ11433450SR SHQ11474604SR SD011050908SR SD011133387SR SD009409769GR SD010500155SR SD011184421SR SHQ12167185SR SHQ12289063SA SHQ12289065SA SHQ12309731SA SHQ12309733SA SHQ11484735SR SHQ11484752SR SHQ11484757SR SHQ11484760SR SD009244729GR SHQ11342288SR SHQ12309570SA SHQ12309583SA SHQ12309591SA SHQ11240186SR SHQ11423151SR SHQ11423156SR SHQ11423168SR SHQ11433425SR SHQ11433437SR SHQ11433442SR SHQ11474599SR SHQ11474606SR SHQ11362678SR SD009400108GR SHQ11362692SR SHQ11362694SR SHQ11372779SR SHQ11372892SR SHQ12320204SA SHQ12289066SA SHQ12187495SA SHQ12309496SA SHQ12309500SA SHQ12309505SA SHQ12309515SA SHQ12310001SA SHQ12310041SA SHQ12310054SA SHQ12310056SA SHQ12310058SA SHQ12310060SA SHQ12319991SA SHQ12319992SA SHQ12319994SA SHQ12320340SA SD011164129SR SHQ12310009SA SHQ12310024SA SHQ12310027SA SHQ12320161SA SHQ12320180SA SHQ12320343SA SHQ12310042SA SHQ12320344SA SHQ12320374SA SHQ12320376SA SHQ12310105SA SHQ12319965SA SHQ12319969SA SHQ12319976SA SHQ12319978SA SHQ12320223SA SHQ12320246SA SHQ12320337SA SHQ12320339SA SHQ12320341SA SHQ12319909SABOP SHQ12320355SA SHQ12330552SA SHQ11230067SR SHQ12330547SA SHQ12330563SA SHQ12319971SA SHQ12319977SA SHQ12330515SA SHQ12320219SA SHQ11230065SR SHQ11474610SR SHQ11474615SR SHQ12330556SA SHQ12310098SA SHQ12310102SA SHQ12330730SA SHQ12330738SA SHQ12330742SA SHQ12330764SA SHQ12330560SA SHQ12299256SA SHQ12309572SA SHQ12309573SA SHQ12330567SA SD011124216GR SD011134242GR SHQ11484720SR SHQ11484723SR SHQ11484738SR SHQ11240262SR SHQ11484725SR SHQ11484726SR SHQ11484728SR SHQ11484731SR SHQ11484743SR SHQ11484756SR SHQ11250286SR SHQ11362627SR SHQ11484732SR SHQ11484740SR SHQ11484744SR SHQ11230070SR SHQ11230084SR SHQ11230086SR SHQ11230094SR SHQ11240200SR SHQ11423157SR SHQ11433448SR SHQ11474605SR SHQ11474611SR SHQ12330551SA SHQ12309529SA SHQ12309531SA SHQ12309534SA SHQ12309535SA SHQ12309536SA SHQ12309540SA SHQ12309541SA SHQ12309543SA SHQ12309544SA SHQ12309545SA SHQ12309546SA SHQ11190005SA SHQ11190007SA SHQ11190008SA SHQ11190009SA SHQ11190010SA SHQ11240251SA SD011204713SR SD011204717SR SD011224962SR SD011224963SR SD011224964SR SD011224965SR SD011224968SR SD011224969SR SD011224970SR SHQ11250294SR SHQ11250296SR SHQ11250297SR SHQ11362614SR SHQ11484719SR SHQ11484721SR SHQ11484722SR SHQ11484724SR SHQ11484730SR SHQ11484741SR SHQ11484755SR SHQ11433424SR SHQ11433440SR SD011204761SR SHQ11423158SR SHQ12361172SA SHQ12371338SA SHQ12371343SA SHQ12371344SA SHQ12371345SA SHQ12371346SA SHQ12320448SR SHQ11342312SR SHQ12320194SA SHQ11423173SR SHQ11433438SR SHQ11190026SA SD010449574GR SD010449577GR SD010449593GR SD010449657GR SD010449659GR SD010449674GR SD010449705GR SD010459741GR SD010470524GR SD010491536GR SD010491552GR SHQ11280196GR SHQ11240185SR SHQ11423150SR SHQ11423153SR pbiC4EH5806GR SHQ12371348SA SD010127763GR SD007387389GR SD007439670GR SD008045914GR SD008056572GR SD008076852GR SD008077066GR SD008077076GR SD008077106GR SD008077181GR SD008138379GR SD008138388GR SD008138389GR SD008138390GR SD008138398GR SD008388071GR SD009099161GR SD009129899GR SD009182616GR SD009192982GR SD009203155GR SD009234016GR SD009234019GR SD009234023GR SD010210399GR SHQ11461522GR SHQ12381512SA SHQ11190015SA SD011225018SR SD011225020SR SD008429858GR SD009409711GR SD009409778GR SD011112690SR SD011184412SR SHQ11342309SR SHQ11352483SR SHQ11362667SR SHQ11362676SR SHQ11362677SR SHQ11362679SR SHQ11362696SR SHQ11362701SR SHQ11362704SR SHQ11362706SR SHQ11372790SR SHQ11372798SR SHQ11372805SR SHQ11372893SR SHQ11372894SR SHQ11372895SR SHQ11372896SR SHQ12167175SR SHQ12167188SR SHQ11433444SR SHQ12391676SA SHQ12391679SA SHQ12391681SA SHQ12391660SA SHQ12381592SA SD010408067GA SD010460117GA SD010460132GA SD010470347GA SD010502160GA SHQ12309526SA SHQ12289136SA SHQ12309525SA SHQ12309527SA SHQ12351121SA SHQ12351123SA SD010418752GA SD011093737GA SD011093738GA SD011093740GA SD011093749GA SD011144492GA SD011144495GA SD011144496GA SD011144509GA SD011154612GA SD011154613GA SD011154614GA SD011154615GA SD011154616GA SD011154618GA SD011154619GA SD011154620GA SD011154622GA SD011154625GA SD011154626GA SD011154627GA SD011154630GA SD011164690GA SD011164696GA SD011164697GA SD011164701GA SD011164702GA SD011164704GA SD011164705GA SD011164706GA SD011164708GA SD011164710GA SD011215071GA SHQ11421356GA SHQ11441370GA SHQ11441374GA SHQ12217862SA SHQ12320222SA SHQ12320309SA SHQ12319973SA SHQ12320244SA SHQ12320417SA SD010334832GA SD010377120GA SD010377126GA SD010387228GA SD010387367GA SD010387368GA SD010387369GA SD010387453GA SD010387464GA SD010387468GA SD010387470GA SD010387475GA SD010387674GA SD010387677GA SD010387678GA SD010387738GA SD010387751GA SD010387790GA SD010387792GA SD010387797GA SD010387798GA SD010387799GA SD010387800GA SD010387802GA SD010387803GA SD010387804GA SD010387806GA SD010387810GA SD010387813GA SD010387814GA SD010387815GA SD010397988GA SD010408131GA SD010408143GA SD010408249GA SD010408273GA SD010408305GA SD010408313GA SD010418458GA SD010418846GA SD010418864GA SD010418871GA SD010418919GA SD010428927GA SD010428928GA SD010428931GA SD010428932GA SD010428955GA SD010428990GA SD010429138GA SD010429164GA SD010429170GA SD010429175GA SD010429224GA SD010429226GA SD010429264GA SD010429267GA SD010430011SA SD010449329GA SD010459927GA SD010459932GA SD010460126GA SD010470499GA SD010470502GA SD010470518GA SD010470542GA SD010470719GA SD010470790GA SD010470800GA SD010480114SA SD010481082GA SD010491533GA SD010491602GA SD010491603GA SD010491609GA SD010500214SA SD010500236SA SD010500241SA SD010500245SA SD010502307GA SD010532886GA SD010532894GA SD010532907GA SD011051203SA SD011051207SA SD011061343SA SD011061354SA SD011061388SA SD011061390SA SD011061397SA SD011061399SA SD011061400SA SD011071678SA SD011072037SA SD011072039SA SD011072042SA SD011082169SA SD011112597SA SD011112598SA SD011112600SA SD011112687SA SD011112695SA SD011112731SA SD011112741SA SD011122924SA SD011122943SA SD011122946SA SD011122951SA SD011122952SA SD011122962SA SD011122963SA SD011122964SA SD011122965SA SD011122991SA SD011123027SA SD011123237SA SD011123251SA SD011123307SA SD011123339SA SD011123356SA SD011133437SA SD011133739SA SD011143866SA SD011143868SA SD011144490GA SD011154003SA SD011164125SA SD011174179SA SD011174229SA SD011184460SA SD011194494SA SD011194509SA SD011194525SA SD011204677SA SD011204755SA SD011224932SA SD011225064SA SHQ11250293SA SHQ11260528SA SHQ11280626SA SHQ11280742SA SHQ11280756SA SHQ11280762SA SHQ11280770SA SHQ11280777SA SHQ11280804SA SHQ11280807SA SHQ11280810SA SHQ11280815SA SHQ11290835SA SHQ11290845SA SHQ11290849SA SHQ11301221SA SHQ11301252SA SHQ11301253SA SHQ11301280SA SHQ11301281SA SHQ11301282SA SHQ11301284SA SHQ11301287SA SHQ11301290SA SHQ11301294SA SHQ11301302SA SHQ11301308SA SHQ11301312SA SHQ11311384SA SHQ11311389SA SHQ11311421SA SHQ11321711SA SHQ11321753SA SHQ11321759SA SHQ11321766SA SHQ11321771SA SHQ11331984SA SHQ11342196SA SHQ11342285SA SHQ11342286SA SHQ11352387SA SHQ11352402SA SHQ11352407SA SHQ11352408SA SHQ11352516SA SHQ11362643SA SHQ11362659SA SHQ11382993SA SHQ11443658SA SHQ11443760SA SHQ11443768SA SHQ11443770SA SHQ11453921SA SHQ11454064SA SHQ11474579SA SHQ11484751SA SHQ11484753SA SHQ11484762SA SHQ11484763SA SHQ11484766SA SHQ11484796SA SHQ11484807SA SHQ11484877SA SHQ11495098SA SHQ11495099SA SHQ12015412SA SHQ12015413SA SHQ12015418SA SHQ12015420SA SHQ12015424SA SHQ12015425SA SHQ12015435SA SHQ12015439SA SHQ12015441SA SHQ12015453SA SHQ12015460SA SHQ12015461SA SHQ12015465SA SHQ12015473SA SHQ12025538SA SHQ12035604SA SHQ12035655SA SHQ12045764SA SHQ12045798SA SHQ12056002SA SHQ12056004SA SHQ12056005SA SHQ12056006SA SHQ12056007SA SHQ12056008SA SHQ12056009SA SHQ12056011SA SHQ12056014SA SHQ12066091SA SHQ12076260SA SHQ12076261SA SHQ12076262SA SHQ12076265SA SHQ12076267SA SHQ12076310SA SHQ12076314SA SHQ12096379SA SHQ12096384SA SHQ12096386SA SHQ12096391SA SHQ12096405SA SHQ12096407SA SHQ12096457SA SHQ12106488SA SHQ12126687SA SHQ12146877SA SHQ12146879SA SHQ12146883SA SHQ12146886SA SHQ12157020SA SHQ12157022SA SHQ12157028SA SHQ12157029SA SHQ12167064SA SHQ12167071SA SHQ12167080SA SHQ12167082SA SHQ12167084SA SHQ12167088SA SHQ12167091SA SHQ12167102SA SHQ12167105SA SHQ12167106SA SHQ12167168SA SHQ12167170SA SHQ12167171SA SHQ12167172SA SHQ12167215SA SHQ12177306SA SHQ12177337SA SHQ12187453SA SHQ12187465SA SHQ12187473SA SHQ12187532SA SHQ12197566SA SHQ12197567SA SHQ12197573SA SHQ12197580SA SHQ12197581SA SHQ12197582SA SHQ12197586SA SHQ12197590SA SHQ12197608SA SHQ12197701SA SHQ12197707SA SHQ12207803SA SHQ12207808SA SHQ12207811SA SHQ12218035SA SHQ12238285SA SHQ12238290SA SHQ12238292SA SHQ12238294SA SHQ12238320SA SHQ12238321SA SHQ12238324SA SHQ12238327SA SHQ12238329SA SHQ12238331SA SHQ12238389SA SHQ12248575SA SHQ12248582SA SHQ12258763SA SHQ12279024SA SHQ12309532SA SHQ12319964SA SHQ12319970SA SHQ12319983SA SHQ12319984SA SHQ12330851SA SHQ12310080SA SHQ12330530SA SD010387536GA SD010408281GA SD010408283GA SD010408287GA SD010408288GA SD010408290GA SD010408291GA SD010418534GA SD010480802GA SD010481039GA SD010481043GA SD010491409GA SD010491582GA SD010501775GA SD010501793GA SD010502112GA SD010502149GA SD011071558SA SHQ11230114SA SHQ11240250SA SHQ11250303SA SHQ11250472SA SHQ11280635SA SHQ11280664SA SHQ11301151SA SHQ11301153SA SHQ11301161SA SHQ11301200SA SHQ11310428GA SHQ11310430GA SHQ11321756SA SHQ11321764SA SHQ11321774SA SHQ11332024SA SHQ11342222SA SHQ11383004SA SHQ11383044SA SHQ11383046SA SHQ11433457SA SHQ11464083SA SHQ12015387SA SHQ12015396SA SD010470500GA SHQ11300390GA SHQ11300392GA SHQ11300399GA SHQ11300400GA SHQ11300415GA SHQ11300419GA SHQ11300421GA SHQ11332011SA SHQ11332020SA SHQ11372811SA SHQ11423171SA SHQ11423268SA SHQ11423281SA SHQ11423282SA SHQ11423284SA SD010460214GA SD010500230SA SHQ12167056SA SHQ12167058SA SHQ12167062SA SD010481269GA SD010491529GA SD010491535GA SHQ11301146SA SHQ11301147SA SHQ11301148SA SHQ11301149SA SHQ11301192SA SHQ11301205SA SD010481253GA SD010481256GA SD010481272GA SD010481275GA SD010491310GA SD010491313GA SD010491315GA SD010491532GA SD010522624GA SD010522643GA SD010522647GA SD010532767GA SD010532775GA SD011033006GA SD011033007GA SD011033024GA SD011033034GA SD011033083GA SD011033084GA SD011033086GA SD011033090GA SD011033093GA SD011033095GA SD011033142GA SD011033144GA SD011033145GA SD011033147GA SD011033148GA SD011033149GA SD011033150GA SD011033151GA SD011033152GA SD011033154GA SD011033155GA SD011033158GA SD011033161GA SD011033162GA SHQ11484668SA SHQ11484703SA SHQ11484706SA SHQ11484707SA SHQ11484713SA SHQ11484714SA SHQ11484714SABOP SHQ11484769SA SHQ11484771SA SD010324538GA SHQ12132998GA SHQ12132999GA SHQ12133001GA SHQ12133003GA SHQ12143018GA SHQ12143020GA SHQ12143021GA SHQ12143024GA SHQ12143025GA SHQ12143027GA SHQ12153063GA SD010376568GA SD010376585GA SD010376834GA SD010387769GA SD010418668GA SD010491316GA SD010502134GA SD010522612GA SD010522613GA SD010522616GA SD010522632GA SD011050918SA SD011050919SA SD011051049SA SD011051060SA SD011071560SA SD011071836SA SD011071840SA SD011071895SA SD011071914SA SD011071987SA SD011072145SA SD011082383SA SD011112759SA SD011112772SA SD011112789SA SD011143749SA SHQ11280637SA SHQ11280638SA SHQ11280643SA SHQ11280665SA SHQ11311383SA SHQ12096428SA SHQ12096442SA SHQ12096459SA SHQ12106543SA SHQ12106546SA SHQ12116571SA SHQ12116584SA SHQ12116634SA SHQ12126719SA SHQ12126721SA SHQ12126724SA SHQ12126725SA SHQ12126726SA SHQ12126727SA SHQ12126728SA SHQ12126729SA SHQ12126730SA SHQ12167162SA SHQ12187475SA SHQ12187476SA SHQ12187478SA SHQ12187479SA SHQ12187480SA SHQ12238288SA SHQ12238291SA SD010500219SA SD010500227SA SD010500232SA SD010500237SA SD010500251SA SD010500253SA SD010500255SA SD010500258SA SD010500259SA SD010500261SA SD010500262SA SD010500266SA SD011051206SA SD011061441SA SD011061463SA SD011061466SA SD011061493SA SD011071574SA SD011071575SA SD011071586SA SD011071591SA SD011071601SA SD011071606SA SD011071610SA SD011071617SA SD011071628SA SD011071631SA SD011072045SA SD011072075SA SD011072081SA SHQ12032502GA SHQ12062611GA SHQ12062612GA SHQ12062613GA SHQ12062614GA SHQ12062615GA SHQ12062616GA SHQ12062618GA SHQ12062619GA SHQ12062621GA SD010345409GA SD010366442GA SD010377124GA SD010387266GA SD010387269GA SD010387275GA SD010387278GA SD010387282GA SD010397889GA SD010397891GA SD010397897GA SD010397903GA SD010397911GA SD010397912GA SD010397915GA SD010408192GA SD010408325GA SD010408336GA SD010408419GA SD010418480GA SD010418481GA SD010418487GA SD010418499GA SD010418500GA SD010418501GA SD010418502GA SD010418503GA SD010418504GA SD010418505GA SD010418507GA SD010418508GA SD010418509GA SD010418548GA SD010418593GA SD010418598GA SD010418599GA SD010418601GA SD010418602GA SD010418603GA SD010418605GA SD010418660GA SD010418661GA SD010418662GA SD010418663GA SD010418665GA SD010418670GA SD010418671GA SD010418672GA SD010418828GA SD010418829GA SD010418835GA SD010418836GA SD010418856GA SD010418863GA SD010418898GA SD010428921GA SD010428924GA SD010428925GA SD010428926GA SD010428930GA SD010429107GA SD010429150GA SD010429220GA SD010430032SA SD010430034SA SD010430040SA SD010449407GA SD010449408GA SD010449415GA SD010449502GA SD010449506GA SD010449513GA SD010470416GA SD010470710GA SD010470717GA SD010470793GA SD010470799GA SD010480876GA SD010480881GA SD010480884GA SD010480887GA SD010480889GA SD010480890GA SD010480899GA SD010480910GA SD010481071GA SD010481075GA SD010481080GA SD010481081GA SD010481163GA SD010481176GA SD010481184GA SD010481195GA SD010491351GA SD010491354GA SD010491360GA SD010491365GA SD010491366GA SD010491367GA SD010491368GA SD010491370GA SD010491371GA SD010491372GA SD010491373GA SD010491374GA SD010491375GA SD010491376GA SD010491377GA SD010491378GA SD010491444GA SD010491540GA SD010491541GA SD010491543GA SD010491557GA SD010491564GA SD010491567GA SD010491673GA SD010501751GA SD010501766GA SD010501877GA SD010501887GA SD010501889GA SD010501891GA SD010501997GA SD010502000GA SD010502011GA SD010510336SA SD010522678GA SD010522683GA SD010522685GA SD010532686GA SD010532840GA SD010532846GA SD011043179GA SD011053181GA SD011053183GA SD011053187GA SD011053188GA SD011053190GA SD011053192GA SD011053193GA SD011053198GA SD011063338GA SD011063349GA SD011071627SA SD011073385GA SD011073399GA SD011073417GA SD011073432GA SD011073463GA SD011073465GA SD011073467GA SD011073518GA SD011073525GA SD011073527GA SD011073529GA SD011073530GA SD011073535GA SD011073536GA SD011073537GA SD011073540GA SD011073542GA SD011073543GA SD011073544GA SD011073548GA SD011073549GA SD011073551GA SD011073552GA SD011073553GA SD011073554GA SD011073555GA SD011073556GA SD011083586GA SD011083587GA SD011083593GA SD011083597GA SD011083603GA SD011083605GA SD011093685GA SD011093697GA SD011093698GA SD011093739GA SD011093741GA SD011093745GA SD011093747GA SD011093750GA SD011093751GA SD011093752GA SD011093754GA SD011103758GA SD011103769GA SD011113822GA SD011113829GA SD011113831GA SD011113833GA SD011113835GA SD011113838GA SD011113845GA SD011113847GA SD011113852GA SD011113854GA SD011113856GA SD011113860GA SD011113863GA SD011113864GA SD011113866GA SD011113867GA SD011113869GA SD011113871GA SD011124232GA SD011124234GA SD011134297GA SD011134314GA SD011134315GA SD011134341GA SD011164639GA SD011164648GA SD011164650GA SD011164651GA SD011164652GA SD011164653GA SD011164654GA SD011164655GA SD011164658GA SD011164659GA SD011164660GA SD011164661GA SD011164712GA SD011164713GA SD011164714GA SD011164715GA SD011164716GA SD011164717GA SD011164718GA SD011164719GA SD011164720GA SD011164721GA SD011164722GA SD011164723GA SD011164724GA SD011164725GA SD011164726GA SD011174728GA SD011174731GA SD011174732GA SD011174733GA SD011174734GA SD011174735GA SD011174736GA SD011174737GA SD011174738GA SD011174739GA SD011174740GA SD011174741GA SD011174743GA SD011174744GA SD011174745GA SD011174746GA SD011174747GA SD011174810GA SD011174811GA SD011174812GA SD011174813GA SD011174814GA SD011174815GA SD011174816GA SD011174817GA SD011174818GA SD011174819GA SD011174821GA SD011174822GA SD011174823GA SD011174824GA SD011174825GA SD011174826GA SD011174827GA SD011174829GA SD011184832GA SD011184833GA SD011184836GA SD011184838GA SD011184848GA SD011184849GA SD011184897GA SD011184906GA SD011184911GA SD011184913GA SD011184919GA SD011204966GA SD011204968GA SD011204971GA SD011204979GA SD011204988GA SD011204989GA SD011204990GA SD011204991GA SD011204992GA SD011204993GA SD011204994GA SD011204995GA SD011204996GA SD011204999GA SD011205000GA SD011205001GA SD011205002GA SD011205003GA SD011205004GA SD011205005GACOMPLETE SD011205006GA SD011205007GA SD011205008GA SD011205011GA SD011205012GA SD011205015GA SD011205016GA SD011205017GA SD011205019GA SD011205020GA SD011205021GA SD011205022GA SD011205023GA SD011205024GA SD011205025GA SD011205027GA SD011205028GA SD011205029GA SD011205030GA SD011205031GA SD011205032GA SD011205033GA SD011205034GA SD011205035GA SD011205036GA SD011205037GA SD011205038GA SD011205039GA SD011205040GA SD011205041GA SD011205042GA SD011205043GA SD011205044GA SD011205045GA SD011205046GA SD011215048GA SD011215049GA SD011215052GA SD011215053GA SD011215054GA SD011215056GA SD011215059GA SHQ11190005GA SHQ11190006GA SHQ11190007GA SHQ11190008GA SHQ11190009GA SHQ11190010GA SHQ11190011GA SHQ11190012GA SHQ11190014GA SHQ11190015GA SHQ11190016GA SHQ11190017GA SHQ11190018GA SHQ11190019GA SHQ11190020GA SHQ11190021GA SHQ11190026GA SHQ11230082GACOMPLETE SHQ11230106GA SHQ11230124GA SHQ11230125GA SHQ11230126GA SHQ11230127GA SHQ11230128GA SHQ11230129GA SHQ11230130GA SHQ11230131GA SHQ11230132GA SHQ11250159GA SHQ11280172GA SHQ11280173GA SHQ11280174GA SHQ11280175GA SHQ11280176GA SHQ11280201GA SHQ11280231GA SHQ11280240GA SHQ11290282GA SHQ11290285GA SHQ11290286GA SHQ11290287GA SHQ11290289GA SHQ11290349GA SHQ11290353GA SHQ11290354GA SHQ11300358GA SHQ11300360GA SHQ11300365GA SHQ11300375GA SHQ11300388GA SHQ11320542GA SHQ11320546GA SHQ11320547GA SHQ11320549GA SHQ11320551GA SHQ11320552GA SHQ11320554GA SHQ11320555GA SHQ11320556GA SHQ11320600GA SHQ11320603GA SHQ11320604GA SHQ11330646GA SHQ11330647GA SHQ11330652GA SHQ11330653GA SHQ11330664GA SHQ11330667GA SHQ11330712GA SHQ11330715GA SHQ11330716GA SHQ11330717GA SHQ11330721GA SHQ11330722GA SHQ11330723GA SHQ11330724GA SHQ11330726GA SHQ11340728GA SHQ11340729GA SHQ11340732GA SHQ11340739GA SHQ11340743GA SHQ11340746GA SHQ11340755GA SHQ11340756GA SHQ11340760GA SHQ11340764GA SHQ11340765GA SHQ11340766GA SHQ11340767GA SHQ11340768GA SHQ11340769GA SHQ11340770GA SHQ11340771GA SHQ11340772GA SHQ11340773GA SHQ11340774GA SHQ11340775GA SHQ11340776GA SHQ11340777GA SHQ11340778GA SHQ11340779GA SHQ11340780GA SHQ11340781GA SHQ11340783GA SHQ11340784GA SHQ11340785GA SHQ11340786GA SHQ11340787GA SHQ11340788GA SHQ11340790GA SHQ11340791GA SHQ11360824GA SHQ11360825GA SHQ11360826GA SHQ11360827GA SHQ11360828GA SHQ11360829GA SHQ11360830GA SHQ11360831GA SHQ11360832GA SHQ11360833GA SHQ11360834GA SHQ11360835GA SHQ11360836GA SHQ11360837GA SHQ11360838GA SHQ11360839GA SHQ11360840GA SHQ11360841GA SHQ11360842GA SHQ11360844GA SHQ11360845GA SHQ11360847GA SHQ11360848GA SHQ11360849GA SHQ11360850GA SHQ11360851GA SHQ11360853GA SHQ11360855GA SHQ11360856GA SHQ11360857GA SHQ11360861GA SHQ11360863GA SHQ11360864GA SHQ11360865GA SHQ11370868GA SHQ11370869GA SHQ11370871GA SHQ11370878GA SHQ11370880GA SHQ11370886GA SHQ11370908GA SHQ11370910GA SHQ11370914GA SHQ11370915GA SHQ11370916GA SHQ11370917GA SHQ11370918GA SHQ11370919GA SHQ11370920GA SHQ11370921GA SHQ11370922GA SHQ11370923GA SHQ11370924GA SHQ11370925GA SHQ11370926GA SHQ11370927GA SHQ11370939GA SHQ11370954GA SHQ11370955GA SHQ11370956GA SHQ11370957GA SHQ11370959GA SHQ11370960GA SHQ11370966GA SHQ11370967GA SHQ11370968GA SHQ11370969GA SHQ11370970GA SHQ11370971GA SHQ11370972GA SHQ11370973GA SHQ11370974GA SHQ11370975GA SHQ11370977GA SHQ11370978GA SHQ11370979GA SHQ11370980GA SHQ11370981GA SHQ11370982GA SHQ11370983GA SHQ11370984GA SHQ11370987GA SHQ11370989GA SHQ11370990GA SHQ11381039GA SHQ11381041GA SHQ11381042GA SHQ11381043GA SHQ11381044GA SHQ11381045GA SHQ11381046GA SHQ11381047GA SHQ11381048GA SHQ11381049GA SHQ11381050GA SHQ11381051GA SHQ11381052GA SHQ11381053GA SHQ11381054GA SHQ11381055GA SHQ11381056GA SHQ11381057GA SHQ11381058GA SHQ11381059GA SHQ11381060GA SHQ11381061GA SHQ11381063GA SHQ11381064GA SHQ11381067GA SHQ11381068GA SHQ11381069GA SHQ11381070GA SHQ11381071GA SHQ11381072GA SHQ11381073GA SHQ11381074GA SHQ11381075GA SHQ11381077GA SHQ11381078GA SHQ11391157GA SHQ11391158GA SHQ11391160GA SHQ11391162GA SHQ11391163GA SHQ11391164GA SHQ11391165GA SHQ11391166GA SHQ11391168GA SHQ11391170GA SHQ11391172GA SHQ11391173GA SHQ11391175GA SHQ11391176GA SHQ11401178GA SHQ11401180GA SHQ11401182GA SHQ11401184GA SHQ11401188GA SHQ11401189GA SHQ11401190GA SHQ11401191GA SHQ11401193GA SHQ11401196GA SHQ11401197GA SHQ11401208GA SHQ11401219GA SHQ11401220GA SHQ11401223GA SHQ11401251GA SHQ11401253GA SHQ11401261GA SHQ11441379GA SHQ11481736GA SHQ11481832GA SHQ11491955GA SHQ11491958GA SHQ11491963GA SHQ11491964GA SHQ11491967GA SHQ11491968GA SHQ11491969GA SHQ11491970GA SHQ11491971GA SHQ11491972GA SHQ11491975GA SHQ11491978GA SHQ11491981GA SHQ11491984GA SHQ11491985GA SHQ11491995GA SHQ11491996GA SHQ11502011GA SHQ11502014GA SHQ11502015GA SHQ11502017GA SHQ11502020GA SHQ11502022GA SHQ11502023GA SHQ11502024GA SHQ11502028GA SHQ11502176GA SHQ11502177GA SHQ11502178GA SHQ11502179GA SHQ11502180GA SHQ11502181GA SHQ11502182GA SHQ11502183GA SHQ11502184GA SHQ11502185GA SHQ11502186GA SHQ11502187GA SHQ11502188GA SHQ11502189GA SHQ11502190GA SHQ11502191GA SHQ11502192GA SHQ11502193GA SHQ11502194GA SHQ11502195GA SHQ11512242GA SHQ11512243GA SHQ11512244GA SHQ11512245GA SHQ11512246GA SHQ11512262GA SHQ11512270GA SHQ11512273GA SHQ11512290GA SHQ11512292GA SHQ11512294GA SHQ11512295GA SHQ11512296GA SHQ11512298GA SHQ11512303GA SHQ11512304GA SHQ11512305GA SHQ11512308GA SHQ12012341GA SHQ12012342GA SHQ12012343GA SHQ12012349GA SHQ12012365GA SHQ12012367GA SHQ12012368GA SHQ12012369GA SHQ12012370GA SHQ12012371GA SHQ12012372GA SHQ12012373GA SHQ12012374GA SHQ12012375GA SHQ12012378GA SHQ12012382GA SHQ12012383GA SHQ12012384GA SHQ12012385GA SHQ12012386GA SHQ12012387GA SHQ12012388GA SHQ12022399GA SHQ12022400GA SHQ12022401GA SHQ12022402GA SHQ12022406GA SHQ12022409GA SHQ12022410GA SHQ12022412GA SHQ12022421GA SHQ12022426GA SHQ12022439GA SHQ12022440GA SHQ12022441GA SHQ12022442GA SHQ12022443GA SHQ12022444GA SHQ12022445GA SHQ12022447GA SHQ12022449GA SHQ12022450GA SHQ12022451GA SHQ12032471GA SHQ12032472GA SHQ12032473GA SHQ12032474GA SHQ12032475GA SHQ12032476GA SHQ12032477GA SHQ12032478GA SHQ12032479GA SHQ12032480GA SHQ12032481GA SHQ12032482GA SHQ12032483GA SHQ12032484GA SHQ12032485GA SHQ12042525GA SHQ12042526GA SHQ12042527GA SHQ12042528GA SHQ12042529GA SHQ12042530GA SHQ12042531GA SHQ12042532GA SHQ12042533GA SHQ12042560GA SHQ12052598GA SHQ12052599GA SHQ12052600GA SHQ12052601GA SHQ12052602GA SHQ12052603GA SHQ12052604GA SHQ12062647GA SHQ12062648GA SHQ12062650GA SHQ12062652GA SHQ12062654GA SHQ12062655GA SHQ12062656GA SHQ12062662GA SHQ12062663GA SHQ12062664GA SHQ12062665GA SHQ12062667GA SHQ12066180SA SHQ12072689GA SHQ12072690GA SHQ12072691GA SHQ12072701GA SHQ12072702GA SHQ12072704GA SHQ12072705GA SHQ12072706GA SHQ12076234SA SHQ12082726GA SHQ12082729GA SHQ12082730GA SHQ12082731GA SHQ12082732GA SHQ12082733GA SHQ12082735GA SHQ12082736GA SHQ12082737GA SHQ12082738GA SHQ12082756GA SHQ12082757GA SHQ12082759GA SHQ12082760GA SHQ12082764GA SHQ12082773GA SHQ12082777GA SHQ12092784GA SHQ12092790GA SHQ12092806GA SHQ12092812GA SHQ12092815GA SHQ12092832GA SHQ12092833GA SHQ12092834GA SHQ12092835GA SHQ12092842GA SHQ12112924GA SHQ12112926GA SHQ12112937GA SHQ12112938GA SHQ12112940GA SHQ12112942GA SHQ12112943GA SHQ12112944GA SHQ12112945GA SHQ12112946GA SHQ12112949GA SHQ12143028GA SHQ12143029GA SHQ12143030GA SHQ12143031GA SHQ12143032GA SHQ12143033GA SHQ12143034GA SHQ12143035GA SHQ12146878SA SHQ12146880SA SHQ12146881SA SHQ12146885SA SHQ12157006SA SHQ12157007SA SHQ12157008SA SHQ12157009SA SHQ12157010SA SHQ12157011SA SHQ12157012SA SHQ12157013SA SHQ12157014SA SHQ12157015SA SHQ12157016SA SHQ12157017SA SHQ12157018SA SHQ12157019SA SHQ12157021SA SHQ12157023SA SHQ12157025SA SHQ12157026SA SHQ12157027SA SHQ12157030SA SHQ12167169SA SHQ12197621SA SHQ12197623SA SHQ12197700SA SHQ12207751SA SHQ12207752SA SHQ12207753SA SHQ12207754SA SHQ12207755SA SHQ12207757SA SHQ12207759SA SHQ12207760SA SHQ12217841SA SHQ12217843SA SHQ12217844SA SHQ12217845SA SHQ12217846SA SHQ12217847SA SHQ12217848SA SHQ12217849SA SHQ12217850SA SHQ12217851SA SHQ12217852SA SHQ12217853SA SHQ12217854SA SHQ12217855SA SHQ12217856SA SHQ12217857SA SHQ12217858SA SHQ12217859SA SHQ12217860SA SHQ12217861SA SHQ12217863SA SHQ12217864SA SHQ12217867SA SHQ12217868SA SHQ12217869SA SHQ12218018SA SHQ12218019SA SHQ12218020SA SHQ12218021SA SHQ12218058SA SHQ12218059SA SHQ12218060SA SHQ12218062SA SHQ12218063SA SHQ12218064SA SHQ12218065SA SHQ12218066SA SHQ12218067SA SHQ12218071SA SHQ12218072SA SHQ12218073SA SHQ12218075SA SHQ12218077SA SHQ12218080SA SHQ12218082SA SHQ12218084SA SHQ12218086SA SHQ12218087SA SHQ12218089SA SHQ12218090SA SHQ12218094SA SHQ12218095SA SHQ12218096SA SHQ12218097SA SHQ12228118SA SHQ12228119SA SHQ12228120SA SHQ12228121SA SHQ12228122SA SHQ12228123SA SHQ12228124SA SHQ12228125SA SHQ12228126SA SHQ12228127SA SHQ12228253SA SHQ12228255SA SHQ12228259SA SHQ12228264SA SHQ12228265SA SHQ12238370SA SHQ12238371SA SHQ12238372SA SHQ12238373SA SHQ12238374SA SHQ12238375SA SHQ12238376SA SHQ12238377SA SHQ12238378SA SHQ12278970SA SHQ12278971SA SHQ12278972SA SHQ12278973SA SHQ12278974SA SHQ12278975SA SHQ12278976SA SHQ12278977SA SHQ12278978SA SHQ12278979SA SHQ12278981SA SHQ12278984SA SHQ12289235SA SHQ12289236SA SHQ12289237SA SHQ12289238SA SHQ12289239SA SHQ12289241SA SHQ12289242SA SHQ12289246SA SHQ12289249SA SHQ12299406SA SHQ12299407SA SHQ12299408SA SHQ12299410SA SHQ12299411SA SHQ12299412SA SHQ12299413SA SHQ12299414SA SHQ12299415SA SHQ12299416SA SHQ12299417SA SHQ12299418SA SHQ12299419SA SHQ12299421SA SHQ12299423SA SHQ12299426SA SHQ12299429SA SHQ12299431SA SHQ12279000SA SHQ12320292SA SHQ12309623SA SHQ12309621SA SHQ12309710SA SHQ12309703SA SHQ12217913SA SHQ12228173SA SHQ12228174SA SHQ12228175SA SHQ12309696SA SHQ12258750SA SHQ12217884SA SHQ12309630SA SHQ12309678SA SHQ12309707SA SHQ12309704SA SHQ12309674SA SHQ12309711SA SHQ12309659SA SHQ12309650SA SHQ12309663SA SHQ12309709SA SHQ12309719SA SHQ12309660SA SHQ12309661SA SHQ12309640SA SHQ12309644SA SHQ12309695SA SHQ12309679SA SHQ12309668SA SHQ12309724SA SHQ12309677SA SHQ12309694SA SHQ12309637SA SHQ12309705SA SHQ12309657SA SHQ12309645SA SHQ12309649SA SHQ12309635SA SHQ12309652SA SHQ12309665SA SHQ12309656SA SHQ12309642SA SHQ12309651SA SHQ12309636SA SHQ12309658SA SHQ12309634SA SHQ12309648SA SHQ12278988SA SHQ12278997SA SHQ12278995SA SHQ12278996SA SHQ12278990SA SHQ12278999SA SHQ12278992SA SHQ12278998SA SHQ12278994SA SHQ12309715SA SHQ12309646SA SHQ12309713SA SHQ12309670SA SHQ12309667SA SHQ12309722SA SHQ12309700SA SHQ12309682SA SHQ12309720SA SHQ12309708SA SHQ12309632SA SHQ12309717SA SHQ12309633SA SHQ12309643SA SHQ12309675SA SHQ12309684SA SHQ12309641SA SHQ12309680SA SHQ12309693SA SHQ12309718SA SHQ12309685SA SHQ12309666SA SHQ12309697SA SHQ12309690SA SHQ12309687SA SHQ12309639SA SHQ12309662SA SHQ12309692SA SHQ12309699SA SHQ12309664SA SHQ12309631SA SHQ12309714SA SHQ12309701SA SHQ12309638SA SHQ12309706SA SHQ12309647SA SHQ12309712SA SHQ12309673SA SHQ12309702SA SHQ12309723SA SD011214892SA SHQ12289148SA SHQ12289138SA SHQ12289149SA SHQ12289141SA SHQ12289150SA SHQ12289143SA SHQ12289142SA SHQ12289145SA SHQ12309612SA SHQ12309625SA SHQ12309615SA SHQ12309608SA SHQ12309610SA SHQ12309627SA SHQ12309619SA SHQ12309616SA SHQ12309611SA SHQ12309607SA SHQ12309629SA SHQ12309624SA SHQ12309628SA SHQ12309609SA SHQ12309626SA SHQ12309603SA SHQ12309605SA SHQ12309601SA SHQ12309600SA SHQ12309602SA SHQ12309617SA SHQ12309614SA SHQ12320390SA SHQ12320391SA SHQ12289106SA SHQ12289100SA SHQ12289086SA SHQ12289098SA SHQ12289091SA SHQ12289097SA SHQ12289087SA SHQ12289104SA SHQ12289109SA SHQ12289105SA SHQ12289108SA SHQ12289090SA SHQ12320272SA SHQ12320275SA SHQ12309606SA SHQ12320268SA SHQ12320266SA SHQ12320278SA SHQ12320261SA SHQ12320259SA SHQ12320260SA SHQ12320267SA SHQ12320282SA SHQ12320270SA SHQ12320293SA SHQ12289055SA SHQ12289092SA SHQ12320303SA SHQ12309613SA SHQ12330564SA SHQ12330557SA SHQ12320301SA SHQ12330566SA SHQ12320306SA SHQ12320389SA SHQ12320388SA SHQ12320279SA SHQ12289057SA SHQ12330807SA SHQ12330795SA SHQ12330798SA SHQ12289058SA SHQ12289059SA SHQ12320258SA SHQ12320265SA SHQ12320257SA SHQ12320264SA SHQ12320262SA SHQ12320283SA SHQ12330796SA SHQ12330799SA SHQ12330793SA SHQ12330806SA SHQ12330805SA SHQ12330558SA SHQ12330797SA SHQ12330794SA SHQ12320294SA SHQ12330722SA SHQ11495009SA SD011082220SA SHQ11464204SA SHQ11464210SABOP SHQ11464211SA SHQ11464213SA SHQ11464214SA SHQ11464215SA SHQ11464216SA SHQ11464219SA SHQ11464222SA SHQ11464224SA SHQ12217982SA SHQ12258735SA SHQ12066095SA SD011112565SA SD011112614SA SD011112620SA SD011112630SA SHQ11332009SA SD010522570GA SHQ12330534SA SD011123300SA SD011123305SA SD011123333SA SD011123335SA SHQ12187501SA SHQ12187502SA SHQ12228144SA SD011112570SA SD011112703SA SD011112736SA SD011122949SA SD011122960SA SD011123072SA SHQ11454062SA SHQ11484825SA SHQ11484841SA SD011122992SA SD011123023SA SD011133548SA SD011133607SA SD011133609SA SHQ12187534SA SHQ12207713SA SHQ12207714SA SHQ12207716SA SHQ12258736SA SHQ12320440SA SHQ11505333SA SHQ11505338SA SHQ11505340SA SHQ11505347SA SHQ11505348SA SHQ12066111SA SHQ12066114SA SHQ12177260SA SHQ12177262SA SHQ12177277SA SHQ12268916SA SHQ12268918SA SHQ11464301SA SHQ11464303SA SHQ12330852SA SHQ12330840SA SHQ12320423SA SHQ12066124SA SHQ12096463SA SHQ12187490SA SHQ12187493SA SHQ12076226SA SHQ12157003SA SHQ11453956SA SHQ11453968SA SHQ12096451SA SHQ12096456SA SHQ12096458SA SHQ12096461SA SHQ12167068SA SHQ12340923SA SHQ12177320SA SHQ11464316SA SHQ11464317SA SHQ11464322SA SHQ12177259SA SHQ12177270SA SHQ12177278SA SD011122925SA SHQ11443763SA SHQ11443769SA SHQ11443772SA SHQ11454049SA SHQ11505356SA SHQ11505360SA SHQ11505363SA SHQ11505365SA SHQ11505372SA SHQ12228129SA SHQ12228134SA SHQ12228136SA SHQ12258766SA SHQ12330616SA SHQ12330613SA SHQ12330619SA SHQ12330617SA SHQ12330614SA SHQ12330610SA SHQ12330609SA SHQ12330611SA SHQ12330612SA SHQ12330618SA SHQ12330826SA SD011102514SA SD011102526SA SD011112604SA SHQ11464364SA SHQ11495193SA SHQ12289068SA SHQ12289071SA SHQ12289082SA SD011030678SA SD011030679SA SD011082165SA SD011082171SA SHQ12167127SA SD011061219SA SHQ11453787SA SHQ11453789SA SHQ11453795SA SHQ11453798SA SHQ12035695SA SD011102512SA SD011112613SA SHQ12106503SA SHQ12106508SA SHQ12106509SA SHQ12106513SA SHQ12106516SA SHQ12106517SA SHQ11484801SA SD011071885SA SD011082360SA SHQ12106497SA SHQ12106515SA SD011033146GA SD011050930SA SD011050938SA SD011050940SA SD011050948SA SD011050949SA SD011050953SA SD011050963SA SHQ12278986SA SHQ12278987SA SD010460267GA SHQ12066108SA SHQ12066118SA SHQ12330478SA SHQ12330475SA SHQ12330469SA SHQ12330477SA SHQ12330479SA SHQ12330471SA SHQ12330483SA SHQ12330472SA SHQ12330482SA SHQ12330481SA SHQ12330480SA SHQ12320201SA SHQ12330451SA SHQ12330457SA SHQ12320214SA SHQ12330460SA SHQ12330454SA SHQ12320202SA SHQ12330462SA SHQ12330461SA SHQ12320371SA SHQ12320365SA SHQ12330450SA SHQ12320373SA SHQ12330463SA SHQ12320372SA SHQ12330455SA SHQ12330452SA SHQ12330459SA SHQ12330464SA SD010501773GA SD011030628SA SD011030629SA SD011030634SA SD011030636SA SD011030638SA SD011030639SA SD011030642SA SD011030643SA SHQ11352384SA SHQ11352385SA SHQ11484919SA SHQ12096413SA SD010450006GA SD010450009GA SD010502056GA SD010512380GA SD011030577SA SHQ11484815SA SHQ11484836SA SHQ11484838SA SHQ11484846SA SHQ11484849SA SHQ12066150SA SHQ12066162SA SHQ12310059SA SHQ12299380SA SHQ12310050SA SHQ12340879SA SD011040732SA SD011082259SA SD011082274SA SHQ12066089SA SHQ11453923SA SHQ12035700SA SD010470602GA SD011051082SA SD011061217SA SD011061221SA SD011061226SA SHQ11352510SA SHQ11352512SA SHQ11433523SA SHQ11443756SA SHQ12106505SA SD010481298GA SHQ11352515SA SHQ12092821GA SHQ11505302SA SHQ12258826SA SD011051202SA SD011051213SA SD011072070SA SD011072083SA SD011072085SA SHQ11433454SA SHQ11240125SA SHQ11240150SA SHQ11240158SA SD011153997SA SD011154004SA SD011154005SA SD011123265SA SD011123274SA SHQ11332049SA SHQ11372891SA SHQ11290884SA SHQ11301119SA SHQ11301121SA SD010460268GA SD010460284GA SD010470339GA SD010470579GA SD010470594GA SD010470599GA SHQ11291010SA SHQ11291011SA SHQ11291020SA SHQ11291021SA SHQ11291024SA SD010376663GA SD010376560GA SD010376790GA SD010387610GA SD010387612GA SD010408083GA SD010408088GA SD011123060SA SHQ11505361SA SHQ11443774SA SD011174233SA SD011174239SA SHQ12371350SA SHQ12371341SA SHQ12371333SA SHQ12371342SA SHQ12330615SA SHQ11230098SA SHQ11230104SA SHQ11291039SA SD011040728SA SD011061359SA SD011082284SA SHQ11352378SA SHQ11372796SA SHQ11372797SA SHQ11301131SA SD009129851GA SD010220679GA SD010303545GA SD010345211GA SD010376534GA SD010387386GA SD010387404GA SD010387479GA SD010387642GA SD010387656GA SD011143878SA SD011143879SA SD011143881SA SD011143893SA SD011143902SA SD011143917SA SD011143919SA SD011143924SA SD011143925SA SD011143927SA SD011143933SA SD011143941SA SD011143943SA SD011143944SA SD011143949SA SHQ11301207SA SHQ11311563SA SHQ11321751SA SHQ11280712SA SHQ11280736SA SHQ11280813SA SD011204747SA SD011204748SA SD011204750SA SD011030644SA SD011030645SA SD010450005GA SD010470328GA SD010470333GA SD010470335GA SD010470344GA SD010470532GA SD010470572GA SD010470600GA SD010470755GA SD010470758GA SD010491581GA SD011123034SA SD011123051SA SD011123109SA SD011164101SA SD011164108SA SD010470574GA SD010470604GA SD010470605GA SHQ11280782SA SD010430042SA SD010430050SA SD010450007GA SD010376785GA SD010387573GA SD010397995GA SD010459887GA SD010459890GA SD011133734SA SD010376642GA SD010376802GA SD010450042GA SD010450045GA SD010460322GA SD010470387GA SD010470584GAX SD010470634GA SD010480980GA SD010480982GA SD010491467GA SD010491468GA SD010501927GA SD010501928GA SD010501931GA SD010501937GA SD010501938GA SD010501941GA SD010501942GA SD010502037GA SD010502076GA SD010502191GA SD010502194GA SD010502198GA SD010502199GA SD010502201GA SD010502202GA SD010502209GA SD010502210GA SD010502212GA SD010502271GA SD010502275GA SD010502277GA SD010512331GA SD010512339GA SD010512340GA SD010512346GA SD010512348GA SD010512352GA SD010512359GA SD010512367GA SD010512371GA SD010512383GA SD010512384GA SHQ11240142SA SHQ11301223SA SHQ11321817SA SHQ11230087SA SD010470603 SD011071809SA SD011214886SA SHQ12330537SA SHQ12330539SA SD010481052GA SD010397958GA SD011033040GA SD011123005SA SD011133382SA SHQ11280719SA SHQ11311530SA SHQ11311560SA SHQ11383054SA SHQ11383067SA SD010408322GA SD010408398GA SD010408409GA SD010408412GA SD010408413GA SD010418546GA SD010418549GA SD010418572GA SD010418576GA SD010418580GA SD010418584GA SD010418591GA SD010428947GA SD010428949GA SD010428951GA SD010428956GA SD010429084GA SD010429181GA SD010429248GA SD010449480GA SD010449483GA SD010449485GA SD010449494GA SD010460123GA SD010470613GA SD010470636GA SD010470648GA SD010470748GA SD010491355GA SD010491357GA SD010491363GA SD010491546GA SD010491553GA SD010491665GA SD010491669GA SD010501926GA SD010501943GA SD010502002GA SD010502004GA SD010502014GA SD010502169GA SD010502178GA SD010502180GA SD010502182GA SD010502183GA SD010502184GA SD011051198SA SD011051201SA SD011061437SA SD011071585SA SD011073522GA SD011083590GA SD011083599GA SD011083602GA SD011083606GA SD011093696GA SD011134306GA SD011154539GA SD011154540GA SD011154548GA SD011174748GA SD011174764GA SD011174768GA SD011174771GA SD011174773GA SD011174782GA SD011174785GA SD011174786GA SD011174787GA SD011174789GA SD011174791GA SD011174808GA SHQ11230057SA SHQ11240127SA SHQ11250438SA SHQ11250448SA SHQ11250449SA SHQ11360843GA SHQ11360846GA SHQ11360852GA SHQ11360854GA SHQ11360858GA SHQ11360859GA SHQ11360860GA SHQ11362671SA SHQ11370872GA SHQ11370873GA SHQ11372786SA SHQ11372787SA SHQ11403093SA SHQ11441372GA SHQ11441385GA SHQ11441392GA SHQ11441393GA SHQ11443739SA SHQ11443745SA SHQ11443745SABOP SHQ11443775SA SHQ12197615SA SD010116948GA SD010127524GA SD010138498GA SD010138499GA SD010251591GA SD010292822GA SD010293057GA SD010293085GA SD010313846GA SD010313851GA SD010314198GA SD010324431GA SD010324437GA SD010334901GA SD010335011GA SD010345376GA SD010355619GA SD010355681GA SD010365869GA SD010365914GA SD010365956GA SD010365988GA SD010365989GA SD010365990GA SD010365991GA SD010365992GA SD010365995GA SD010365997GA SD010366086GA SD010366088GA SD010366093GA SD010366095GA SD010366100GA SD010366141GA SD010366223GA SD010366228GA SD010366230GA SD010366233GA SD010366234GA SD010366235GA SD010366238GA SD010366241GA SD010366243GA SD010366246GA SD010366247GA SD010366251GA SD010366252GA SD010366253GA SD010366254GA SD010366258GA SD010366259GA SD010366262GA SD010366263GA SD010366264GA SD010366265GA SD010366333GA SD010366343GA SD010366345GA SD010366347GA SD010366350GA SD010366353GA SD010366356GA SD010366357GA SD010366360GA SD010366362GA SD010366363GA SD010366364GA SD010366366GA SD010366367GA SD010366368GA SD010366369GA SD010366370GA SD010366371GA SD010366372GA SD010366373GA SD010366375GA SD010366376GA SD010366377GA SD010366378GA SD010366379GA SD010366432GA SD010366433GA SD010366435GA SD010366436GA SD010366437GA SD010366439GA SD010366440GA SD010366441GA SD010366443GA SD010366456GA SD010376667GA SD010376668GA SD010376670GA SD010376671GA SD010376672GA SD010376673GA SD010376674GA SD010376675GA SD010376676GA SD010376678GA SD010376680GA SD010376682GA SD010376683GA SD010376684GA SD010376685GA SD010376686GA SD010376687GA SD010376688GA SD010376689GA SD010376690GA SD010376691GA SD010376692GA SD010376693GA SD010376694GA SD010376695GA SD010376696GA SD010376931GA SD010376933GA SD010376937GA SD010377089GA SD010377109GA SD010377111GA SD010377116GA SD010377117GA SD010377118GA SD010377122GA SD010377128GA SD010387262GA SD010387263GA SD010387267GA SD010387270GA SD010387271GA SD010387280GA SD010387283GA SD010387284GA SD010387285GA SD010387373GA SD010387375GA SD010387376GA SD010387454GA SD010387455GA SD010387457GA SD010387459GA SD010387577GA SD010387578GA SD010387579GA SD010387580GA SD010387581GA SD010387582GA SD010387583GA SD010387584GA SD010387585GA SD010387586GA SD010387588GA SD010387589GA SD010387629GA SD010387640GA SD010387672GA SD010387673GA SD010387682GA SD010387706GA SD010387807GA SD010397896GA SD010397900GA SD010397905GA SD010397906GA SD010397914GA SD010397917GA SD010408138GA SD010408142GA SD010408147GA SD010408151GA SD010408152GA SD010408171GA SD010408173GA SD010408174GA SD010408175GA SD010408177GA SD010408178GA SD010408182GA SD010408184GA SD010408187GA SD010408188GA SD010408190GA SD010408194GA SD010408196GA SD010408254GA SD010408263GA SD010408265GA SD010408267GA SD010408269GA SD010408302GA SD010408307GA SD010408324GA SD010408327GA SD010408332GA SD010408333GA SD010408335GA SD010408340GA SD010408341GA SD010408399GA SD010408400GA SD010408401GA SD010408402GA SD010408404GA SD010408406GA SD010408411GA SD010408416GA SD010408417GA SD010408418GA SD010408420GA SD010408425GA SD010408433GA SD010418445GA SD010418446GA SD010418448GA SD010418449GA SD010418450GA SD010418452GA SD010418455GA SD010418457GA SD010418459GA SD010418460GA SD010418462GA SD010418463GA SD010418464GA SD010418465GA SD010418466GA SD010418467GA SD010418468GA SD010418469GA SD010418470GA SD010418471GA SD010418472GA SD010418473GA SD010418474GA SD010418476GA SD010418477GA SD010418478GA SD010418482GA SD010418483GA SD010418484GA SD010418485GA SD010418489GA SD010418490GA SD010418491GA SD010418492GA SD010418493GA SD010418494GA SD010418495GA SD010418496GA SD010418497GA SD010418498GA SD010418544GA SD010418545GA SD010418547GA SD010418550GA SD010418551GA SD010418552GA SD010418553GA SD010418554GA SD010418557GA SD010418560GA SD010418562GA SD010418563GA SD010418564GA SD010418565GA SD010418566GA SD010418567GA SD010418568GA SD010418569GA SD010418570GA SD010418573GA SD010418574GA SD010418575GA SD010418577GA SD010418578GA SD010418579GA SD010418581GA SD010418582GA SD010418583GA SD010418586GA SD010418587GA SD010418588GA SD010418589GA SD010418590GA SD010418673GA SD010418674GA SD010418675GA SD010418676GA SD010418677GA SD010418678GA SD010418679GA SD010418680GA SD010418681GA SD010418682GA SD010418683GA SD010418684GA SD010418685GA SD010418686GA SD010418687GA SD010418802GA SD010418803GA SD010418804GA SD010418805GA SD010418806GA SD010418807GA SD010418808GA SD010418809GA SD010418810GA SD010418811GA SD010418812GA SD010418813GA SD010418814GA SD010418815GA SD010418816GA SD010418817GA SD010418818GA SD010418819GA SD010418820GA SD010418821GA SD010418822GA SD010418823GA SD010418824GA SD010418825GA SD010418826GA SD010418827GA SD010418830GA SD010418831GA SD010418832GA SD010418833GA SD010418837GA SD010418838GA SD010418840GA SD010418841GA SD010418842GA SD010418843GA SD010418844GA SD010418845GA SD010418847GA SD010418869GA SD010418870GA SD010418902GA SD010428922GA SD010428923GA SD010428950GA SD010428952GA SD010428958GA SD010428959GA SD010428961GA SD010429072GA SD010429077GA SD010429078GA SD010429085GA SD010429086GA SD010429088GA SD010429089GA SD010429090GA SD010429165GA SD010429168GA SD010429173GA SD010429174GA SD010429176GA SD010429177GA SD010429178GA SD010429179GA SD010429180GA SD010429182GA SD010429183GA SD010429184GA SD010429185GA SD010429186GA SD010429187GA SD010429188GA SD010429189GA SD010429190GA SD010429191GA SD010429192GA SD010429193GA SD010429194GA SD010429195GA SD010429196GA SD010429197GA SD010429198GA SD010429199GA SD010429200GA SD010429201GA SD010429202GA SD010429203GA SD010429204GA SD010429205GA SD010429206GA SD010429207GA SD010429208GA SD010429209GA SD010429210GA SD010429212GA SD010429213GA SD010429214GA SD010429215GA SD010429216GA SD010429217GA SD010429242GA SD010429243GA SD010429244GA SD010429245GA SD010429246GA SD010429247GA SD010429249GA SD010429250GA SD010429251GA SD010429252GA SD010429253GA SD010429254GA SD010429255GA SD010429256GA SD010429258GA SD010429259GA SD010429260GA SD010429261GA SD010429263GA SD010429265GA SD010429266GA SD010429269GA SD010429270GA SD010429271GA SD010429272GA SD010429273GA SD010429275GA SD010429276GA SD010429277GA SD010429278GA SD010429279GA SD010429280GA SD010429281GA SD010429282GA SD010429283GA SD010429284GA SD010429285GA SD010429286GA SD010449328GA SD010449331GA SD010449332GA SD010449333GA SD010449334GA SD010449335GA SD010449336GA SD010449337GA SD010449338GA SD010449339GA SD010449340GA SD010449341GA SD010449365GA SD010449366GA SD010449367GA SD010449368GA SD010449369GA SD010449370GA SD010449371GA SD010449373GA SD010449374GA SD010449375GA SD010449376GA SD010449377GA SD010449378GA SD010449379GA SD010449380GA SD010449381GA SD010449382GA SD010449383GA SD010449384GA SD010449385GA SD010449386GA SD010449387GA SD010449388GA SD010449389GA SD010449391GA SD010449398GA SD010449399GA SD010449400GA SD010449402GA SD010449403GA SD010449405GA SD010449406GA SD010449409GA SD010449410GA SD010449411GA SD010449413GA SD010449414GA SD010449417GA SD010449418GA SD010449419GA SD010449420GA SD010449423GA SD010449424GA SD010449425GA SD010449426GA SD010449431GA SD010449432GA SD010449433GA SD010449434GA SD010449448GA SD010449449GA SD010449450GA SD010449451GA SD010449452GA SD010449453GA SD010449456GA SD010449457GA SD010449458GA SD010449459GA SD010449460GA SD010449462GA SD010449464GA SD010449466GA SD010449467GA SD010449468GA SD010449470GA SD010449471GA SD010449472GA SD010449473GA SD010449474GA SD010449481GA SD010449484GA SD010449487GA SD010449488GA SD010449489GA SD010449492GA SD010449496GA SD010449497GA SD010449499GA SD010449500GA SD010449501GA SD010449517GA SD010449518GA SD010449519GA SD010449520GA SD010449521GA SD010449522GA SD010449523GA SD010449524GA SD010449525GA SD010449526GA SD010449527GA SD010449528GA SD010449529GA SD010449530GA SD010449533GA SD010449534GA SD010449535GA SD010449536GA SD010449538GA SD010449539GA SD010449540GA SD010460313GA SD010470534GA SD010470679GA SD010470680GA SD010470681GA SD010470682GA SD010470683GA SD010470685GA SD010470687GA SD010470688GA SD010470690GA SD010470691GA SD010470697GA SD010470700GA SD010470701GA SD010470702GA SD010470703GA SD010470705GA SD010470706GA SD010470708GA SD010470709GA SD010470712GA SD010470713GA SD010470714GA SD010470716GA SD010470718GA SD010470721GA SD010470722GA SD010470725GA SD010470726GA SD010470788GA SD010480886GA SD010480894GA SD010480906GA SD010481079GA SD010481089GA SD010481140GA SD010481141GA SD010481142GA SD010481143GA SD010481144GA SD010481145GA SD010481146GA SD010481148GA SD010481151GA SD010481152GA SD010481155GA SD010481156GA SD010481162GA SD010481170GA SD010481174GA SD010481186GA SD010481190GA SD010481194GA SD010481196GA SD010491339GA SD010491350GA SD010491369GA SD010491442GA SD010491443GA SD010491544GA SD010491559GA SD010491656GA SD010491661GA SD010491672GA SD010491675GA SD010491676GA SD010491677GA SD010491678GA SD010491679GA SD010491680GA SD010491681GA SD010491682GA SD010491683GA SD010491684GA SD010501740GA SD010501741GA SD010501743GA SD010501744GA SD010501746GA SD010501747GA SD010501749GA SD010501750GA SD010501753GA SD010501754GA SD010501755GA SD010501756GA SD010501757GA SD010501758GA SD010501759GA SD010501761GA SD010501765GA SD010501767GA SD010501768GA SD010501769GA SD010501872GA SD010501873GA SD010501874GA SD010501875GA SD010501876GA SD010501878GA SD010501883GA SD010501884GA SD010501892GA SD010501893GA SD010501894GA SD010501895GA SD010502006GA SD010502008GA SD010502013GA SD010502036GA SD010502106GA SD010502109GA SD010502244GA SD010502304GA SD010502305GA SD010502306GA SD010502310GA SD010502312GA SD010502314GA SD010502316GA SD010502317GA SD010502318GA SD010502319GA SD010502320GA SD010512403GA SD010522682GA SD010532874GA SD010532877GA SD010532880GA SD010532882GA SD010532884GA SD010532887GA SD010532888GA SD010532892GA SD010532898GA SD010532903GA SD011033103GA SD011033106GA SD011033107GA SD011033108GA SD011033112GA SD011033118GA SD011033120GA SD011033123GA SD011033126GA SD011033128GA SD011033129GA SD011033130GA SD011033131GA SD011033133GA SD011033134GA SD011033135GA SD011033136GA SD011033138GA SD011033140GA SD011053182GA SD011053185GA SD011053186GA SD011053189GA SD011053191GA SD011053194GA SD011053195GA SD011053196GA SD011053197GA SD011053199GA SD011053200GA SD011053202GA SD011053203GA SD011053204GA SD011053205GA SD011053206GA SD011053207GA SD011053208GA SD011053209GA SD011053211GA SD011053212GA SD011053213GA SD011053214GA SD011053215GA SD011053216GA SD011053217GA SD011053218GAMKE SD011053219GA SD011063220GA SD011063221GA SD011063222GA SD011063223GA SD011063224GA SD011063225GA SD011063226GA SD011063227GA SD011063228GA SD011063229GA SD011063230GA SD011063231GA SD011063232GA SD011063233GA SD011063234GA SD011063235GA SD011063236GA SD011063237GA SD011063238GA SD011063239GA SD011063240GA SD011063329GA SD011063330GA SD011063331GA SD011063332GA SD011063333GA SD011063334GA SD011063335GA SD011063336GA SD011063337GA SD011063339GA SD011063340GA SD011063341GA SD011063342GA SD011063343GA SD011063344GA SD011063345GA SD011063346GA SD011063347GA SD011063348GA SD011063350GA SD011063351GA SD011063352GA SD011063353GA SD011063354GA SD011063355GA SD011063356GA SD011073357GA SD011073358GA SD011073359GA SD011073360GA SD011073361GA SD011073362GA SD011073363GA SD011073364GA SD011073365GA SD011073366GA SD011073367GA SD011073368GA SD011073369GA SD011073371GA SD011073372GA SD011073373GA SD011073374GA SD011073375GA SD011073376GA SD011073377GA SD011073378GA SD011073379GA SD011073380GA SD011073381GA SD011073382GA SD011073383GA SD011073384GA SD011073386GA SD011073387GA SD011073388GA SD011073389GA SD011073390GA SD011073391GA SD011073392GA SD011073393GA SD011073394GA SD011073396GA SD011073397GA SD011073398GA SD011073400GA SD011073401GA SD011073402GA SD011073403GA SD011073404GA SD011073405GA SD011073406GA SD011073407GA SD011073408GA SD011073409GA SD011073410GA SD011073411GA SD011073412GA SD011073413GA SD011073414GA SD011073415GA SD011073416GA SD011073418GA SD011073419GA SD011073420GA SD011073421GA SD011073422GA SD011073423GA SD011073424GA SD011073425GA SD011073426GA SD011073427GA SD011073428GA SD011073430GA SD011073431GA SD011073433GA SD011073435GA SD011073438GA SD011073439GA SD011073440GA SD011073441GA SD011073442GA SD011073444GA SD011073445GA SD011073446GA SD011073447GA SD011073448GA SD011073449GA SD011073450GA SD011073451GA SD011073452GA SD011073453GA SD011073454GA SD011073455GA SD011073456GA SD011073457GA SD011073458GA SD011073459GA SD011073460GA SD011073461GA SD011073462GA SD011073464GA SD011073466GA SD011073468GA SD011073469GA SD011073470GA SD011073471GA SD011073472GA SD011073473GA SD011073474GA SD011073475GA SD011073476GA SD011073477GA SD011073478GA SD011073479GA SD011073481GA SD011073483GA SD011073484GA SD011073486GA SD011073487GA SD011073488GA SD011073489GA SD011073491GA SD011073494GA SD011073495GA SD011073499GA SD011073500GA SD011073503GA SD011073505GA SD011073507GA SD011073510GA SD011073511GA SD011073512GA SD011073515GA SD011073516GA SD011073517GA SD011073519GA SD011073520GA SD011073521GA SD011073523GA SD011073524GA SD011073526GA SD011073528GA SD011073532GA SD011073534GA SD011073538GA SD011073539GA SD011073541GA SD011073545GA SD011073547GA SD011083557GA SD011083558GA SD011083559GA SD011083560GA SD011083561GA SD011083562GA SD011083564GA SD011083565GA SD011083568GA SD011083570GA SD011083573GA SD011083574GA SD011093724GA SD011093726GA SD011093727GA SD011093728GA SD011093730GA SD011093731GA SD011093733GA SD011093734GA SD011093735GA SD011093736GA SD011093744GA SD011113951GA SD011113966GA SD011113971GA SD011113986GA SD011124005GA SD011124116GA SD011124119GA SD011124120GA SD011124121GA SD011124129GA SD011124165GA SD011124166GA SD011124173GA SD011134331GA SD011134365GA SD011134374GA SD011144489GA SD011144494GA SD011144497GA SD011154512GA SD011154514GA SD011154516GA SD011154517GA SD011154518GA SD011154519GA SD011154520GA SD011154521GA SD011154523GA SD011154524GA SD011154525GA SD011154526GA SD011154527GA SD011154528GA SD011154531GA SD011154572GA SD011154574GA SD011154575GA SD011154578GA SD011154581GA SD011154590GA SD011154592GA SD011154593GA SD011154594GA SD011154595GA SD011154596GA SD011154597GA SD011154598GA SD011154599GA SD011154600GA SD011154601GA SD011154602GA SD011154603GA SD011154604GA SD011154605GA SD011154606GA SD011154607GA SD011154608GA SD011154609GA SD011154610GA SD011154611GA SD011154621GA SD011154624GA SD011154631GA SD011164632GA SD011164633GA SD011164634GA SD011164635GA SD011164636GA SD011164637GA SD011164638GA SD011164640GA SD011164641GA SD011164642GA SD011164643GA SD011164644GA SD011164645GA SD011164646GA SD011164647GA SD011164649GA SD011164656GA SD011164657GA SD011164663GA SD011164680GA SD011164681GA SD011164682GA SD011164683GA SD011164684GA SD011164685GA SD011164687GA SD011164688GA SD011164689GA SD011164691GA SD011164692GA SD011164693GA SD011164694GA SD011164695GA SD011164698GA SD011164699GA SD011164700GA SD011164703GA SD011164707GA SD011164709GA SD011164711GA SD011164727GA SD011174729GA SD011174730GA SD011174742GA SD011174772GA SD011174802GA SD011174820GA SD011184830GA SD011184831GA SD011184834GA SD011184835GA SD011184837GA SD011184839GA SD011184840GA SD011184841GA SD011184842GA SD011184843GA SD011184844GA SD011184845GA SD011184846GA SD011184847GA SD011184850GA SD011184851GA SD011184852GA SD011184853GA SD011184854GA SD011184856GA SD011184857GA SD011184858GA SD011184860GA SD011184861GA SD011184862GA SD011184863GA SD011184864GA SD011184865GA SD011184866GA SD011184867GA SD011184868GA SD011184869GA SD011184870GA SD011184871GA SD011184876GA SD011184879GA SD011184881GA SD011184900GA SD011184910GA SD011184912GA SD011184915GA SD011184916GA SD011184917GA SD011184918GA SD011194920GA SD011194921GA SD011194922GA SD011194923GA SD011194924GA SD011194925GA SD011194926GA SD011194927GA SD011194928GA SD011194929GA SD011194930GA SD011194931GA SD011194934GA SD011194935GA SD011194943GA SD011194946GA SD011194947GA SD011194948GA SD011194950GA SD011194951GA SD011194952GA SD011194953GA SD011194954GA SD011194955GA SD011194956GA SD011194957GA SD011194958GA SD011194959GA SD011204960GA SD011204961GA SD011204962GA SD011204963GA SD011204965GA SD011204967GA SD011204973GA SD011204975GA SD011204976GA SD011204977GA SD011204978GA SD011204981GA SD011204982GA SD011204983GA SD011204984GA SD011204985GA SD011204987GA SD011204997GA SD011204998GA SD011205009GA SD011205010GA SD011205047GA SD011215051GA SD011215055GA SD011215057GA SD011215060GA SD011215061GA SD011215062GA SD011215063GA SD011215064GA SD011215065GA SD011215066GA SD011215067GA SD011215068GA SD011215069GA SD011215070GA SD011215072GA SD011215073GA SD011215074GA SD011215075GA SD011215076GA SD011215077GA SD011215078GA SD011215079GA SD011215080GA SD011215081GA SD011215082GA SD011215083GA SD011215084GA SD011215085GA SD011215086GA SD011215087GA SD011215088GA SD011215089GA SD011215090GA SD011215101GA SD011215102GA SD011215103GA SD011215104GA SD011215105GA SD011215106GA SD011215107GA SD011215108GA SD011215109GA SD011215110GA SHQ11220040GA SHQ11220041GA SHQ11220042GA SHQ11220043GA SHQ11220044GA SHQ11220045GA SHQ11220046GA SHQ11220047GA SHQ11220048GA SHQ11220049GA SHQ11250160GA SHQ11250164GA SHQ11250165GA SHQ11250166GA SHQ11250168GA SHQ11250169GA SHQ11280177GA SHQ11280179GA SHQ11280183GA SHQ11280184GA SHQ11280185GA SHQ11280186GA SHQ11280187GA SHQ11280188GA SHQ11280189GA SHQ11280190GA SHQ11280191GA SHQ11280192GA SHQ11280193GA SHQ11280194GA SHQ11280195GA SHQ11280196GA SHQ11280197GA SHQ11280211GA SHQ11280213GA SHQ11280215GA SHQ11280220GA SHQ11280222GA SHQ11280223GA SHQ11280224GA SHQ11280225GA SHQ11280226GA SHQ11280227GA SHQ11280229GA SHQ11280230GA SHQ11280232GA SHQ11280233GA SHQ11280234GA SHQ11280237GA SHQ11280238GA SHQ11280241GA SHQ11290243GA SHQ11290244GA SHQ11290245GA SHQ11290246GA SHQ11290247GA SHQ11290248GA SHQ11290249GA SHQ11290250GA SHQ11290251GA SHQ11290252GA SHQ11290253GA SHQ11290254GA SHQ11290255GA SHQ11290256GA SHQ11290257GA SHQ11290258GA SHQ11290259GA SHQ11290260GA SHQ11290261GA SHQ11290262GA SHQ11290263GA SHQ11290264GA SHQ11290265GA SHQ11290266GA SHQ11290267GA SHQ11290268GA SHQ11290269GA SHQ11290270GA SHQ11290271GA SHQ11290272GA SHQ11290273GA SHQ11290274GA SHQ11290277GA SHQ11290291GA SHQ11290292GA SHQ11290293GA SHQ11290296GA SHQ11290297GA SHQ11290301GA SHQ11290305GA SHQ11290306GA SHQ11290308GA SHQ11290309GA SHQ11290313GA SHQ11290318GA SHQ11290319GA SHQ11290320GA SHQ11290322GA SHQ11290347GA SHQ11290351GA SHQ11290352GA SHQ11290355GA SHQ11290357GA SHQ11300363GA SHQ11300364GA SHQ11300369GA SHQ11300371GA SHQ11300374GA SHQ11300376GA SHQ11300377GA SHQ11300378GA SHQ11300381GA SHQ11300382GA SHQ11300384GA SHQ11310422GA SHQ11310423GA SHQ11310424GA SHQ11310425GA SHQ11310427GA SHQ11310429GA SHQ11310431GA SHQ11310432GA SHQ11310433GA SHQ11310434GA SHQ11310435GA SHQ11310436GA SHQ11310437GA SHQ11310438GA SHQ11310439GA SHQ11310440GA SHQ11310441GA SHQ11310442GA SHQ11310443GA SHQ11310444GA SHQ11310445GA SHQ11310446GA SHQ11310447GA SHQ11310448GA SHQ11310449GA SHQ11310494GA SHQ11310495GA SHQ11310496GA SHQ11310497GA SHQ11310498GA SHQ11310499GA SHQ11310500GA SHQ11310501GA SHQ11310502GA SHQ11310503GA SHQ11310504GA SHQ11310505GA SHQ11310506GA SHQ11310507GA SHQ11310508GA SHQ11310509GA SHQ11310510GA SHQ11310511GA SHQ11310512GA SHQ11310513GA SHQ11310514GA SHQ11310515GA SHQ11310516GA SHQ11310517GA SHQ11310518GA SHQ11310519GA SHQ11310520GA SHQ11310521GA SHQ11310522GA SHQ11310523GA SHQ11310524GA SHQ11310525GA SHQ11310526GA SHQ11310527GA SHQ11310531GA SHQ11310532GA SHQ11310533GA SHQ11310538GA SHQ11310539GA SHQ11310540GA SHQ11310541GA SHQ11320543GA SHQ11320544GA SHQ11320545GA SHQ11320548GA SHQ11320550GA SHQ11320553GA SHQ11320557GA SHQ11320585GA SHQ11320592GA SHQ11320594GA SHQ11320595GA SHQ11320597GA SHQ11320599GA SHQ11320601GA SHQ11320602GA SHQ11320605GA SHQ11320606GA SHQ11320607GA SHQ11320608GA SHQ11320609GA SHQ11320610GA SHQ11320611GA SHQ11320612GA SHQ11320613GA SHQ11320614GA SHQ11320615GA SHQ11320616GA SHQ11320617GA SHQ11320618GA SHQ11320619GA SHQ11320620GA SHQ11320621GA SHQ11320622GA SHQ11320623GA SHQ11320624GA SHQ11320625GA SHQ11320626GA SHQ11320627GA SHQ11320628GA SHQ11320629GA SHQ11320630GA SHQ11320631GA SHQ11320632GA SHQ11320633GA SHQ11320634GA SHQ11320635GA SHQ11320636GA SHQ11320637GA SHQ11330638GA SHQ11330639GA SHQ11330640GA SHQ11330641GA SHQ11330642GA SHQ11330643GA SHQ11330644GA SHQ11330645GA SHQ11330648GA SHQ11330649GA SHQ11330650GA SHQ11330651GA SHQ11330654GA SHQ11330655GA SHQ11330656GA SHQ11330657GA SHQ11330658GA SHQ11330659GA SHQ11330660GA SHQ11330661GA SHQ11330662GA SHQ11330663GA SHQ11330665GA SHQ11330666GA SHQ11330668GA SHQ11330669GA SHQ11330690GA SHQ11330691GA SHQ11330692GA SHQ11330693GA SHQ11330694GA SHQ11330695GA SHQ11330696GA SHQ11330697GA SHQ11330698GA SHQ11330699GA SHQ11330700GA SHQ11330701GA SHQ11330702GA SHQ11330703GA SHQ11330704GA SHQ11330705GA SHQ11330706GA SHQ11330707GA SHQ11330708GA SHQ11330709GA SHQ11330710GA SHQ11330711GA SHQ11330713GA SHQ11330714GA SHQ11330718GA SHQ11330719GA SHQ11330720GA SHQ11330725GA SHQ11330727GA SHQ11340730GA SHQ11340731GA SHQ11340733GA SHQ11340734GA SHQ11340735GA SHQ11340736GA SHQ11340737GA SHQ11340738GA SHQ11340740GA SHQ11340741GA SHQ11340742GA SHQ11340744GA SHQ11340745GA SHQ11340747GA SHQ11340748GA SHQ11340749GA SHQ11340750GA SHQ11340751GA SHQ11340752GA SHQ11340753GA SHQ11340754GA SHQ11340757GA SHQ11340758GA SHQ11340759GA SHQ11340761GA SHQ11340762GA SHQ11340763GA SHQ11340782GA SHQ11340789GA SHQ11360795GA SHQ11360808GA SHQ11360812GA SHQ11360814GA SHQ11360815GA SHQ11360816GA SHQ11360817GA SHQ11360818GA SHQ11360819GA SHQ11360821GA SHQ11360822GA SHQ11360862GA SHQ11360867GA SHQ11370877GA SHQ11370882GA SHQ11370909GA SHQ11370911GA SHQ11370912GA SHQ11370913GA SHQ11370928GA SHQ11370929GA SHQ11370930GA SHQ11370931GA SHQ11370932GA SHQ11370933GA SHQ11370934GA SHQ11370935GA SHQ11370936GA SHQ11370937GA SHQ11370938GA SHQ11370940GA SHQ11370941GA SHQ11370942GA SHQ11370943GA SHQ11370944GA SHQ11370945GA SHQ11370946GA SHQ11370947GA SHQ11370948GA SHQ11370949GA SHQ11370950GA SHQ11370951GA SHQ11370952GA SHQ11370953GA SHQ11370958GA SHQ11370961GA SHQ11370962GA SHQ11370963GA SHQ11370964GA SHQ11370965GA SHQ11370976GA SHQ11370985GA SHQ11370986GA SHQ11370988GA SHQ11370991GA SHQ11370992GA SHQ11370993GA SHQ11370994GA SHQ11370995GA SHQ11370996GA SHQ11370997GA SHQ11370998GA SHQ11370999GA SHQ11371000GA SHQ11371001GA SHQ11371002GA SHQ11371003GA SHQ11381004GA SHQ11381005GA SHQ11381006GA SHQ11381007GA SHQ11381008GA SHQ11381009GA SHQ11381010GA SHQ11381011GA SHQ11381012GA SHQ11381013GA SHQ11381014GA SHQ11381015GA SHQ11381016GA SHQ11381017GA SHQ11381018GA SHQ11381019GA SHQ11381020GA SHQ11381021GA SHQ11381022GA SHQ11381023GA SHQ11381024GA SHQ11381025GA SHQ11381026GA SHQ11381027GA SHQ11381028GA SHQ11381029GA SHQ11381031GA SHQ11381040GA SHQ11381062GA SHQ11381066GA SHQ11381076GA SHQ11391159GA SHQ11391161GA SHQ11391167GA SHQ11391169GA SHQ11391171GA SHQ11391174GA SHQ11401186GA SHQ11401194GA SHQ11401195GA SHQ11401198GA SHQ11401200GA SHQ11401202GA SHQ11401203GA SHQ11401204GA SHQ11401205GA SHQ11401209GA SHQ11401210GA SHQ11401211GA SHQ11401213GA SHQ11401214GA SHQ11401215GA SHQ11401216GA SHQ11401217GA SHQ11401222GA SHQ11401224GA SHQ11401225GA SHQ11401228GA SHQ11401229GA SHQ11401234GA SHQ11401241GA SHQ11401255GA SHQ11401258GA SHQ11401263GA SHQ11401264GA SHQ11401265GA SHQ11403117SA SHQ11411266GA SHQ11411268GA SHQ11411272GA SHQ11411274GA SHQ11411276GA SHQ11411286GA SHQ11411287GA SHQ11411288GA SHQ11411289GA SHQ11411290GA SHQ11411291GA SHQ11411293GA SHQ11411294GA SHQ11411295GA SHQ11411296GA SHQ11411297GA SHQ11411298GA SHQ11411301GA SHQ11411306GA SHQ11411310GA SHQ11411312GA SHQ11411316GA SHQ11411318GA SHQ11411319GA SHQ11411323GA SHQ11421349GA SHQ11421350GA SHQ11421351GA SHQ11421352GA SHQ11421353GA SHQ11421354GA SHQ11421355GA SHQ11421357GA SHQ11421358GA SHQ11421359GA SHQ11421360GA SHQ11441371GA SHQ11441373GA SHQ11441375GA SHQ11441376GA SHQ11441377GA SHQ11441378GA SHQ11441380GA SHQ11441381GA SHQ11441382GA SHQ11441383GA SHQ11441384GA SHQ11441386GA SHQ11441387GA SHQ11441388GA SHQ11441389GA SHQ11441390GA SHQ11441391GA SHQ11451430GA SHQ11451435GA SHQ11451442GA SHQ11451444GA SHQ11451450GA SHQ11451453GA SHQ11451458GA SHQ11451460GA SHQ11451461GA SHQ11451462GA SHQ11451463GA SHQ11451464GA SHQ11451465GA SHQ11451466GA SHQ11451467GA SHQ11451468GA SHQ11451470GA SHQ11451471GA SHQ11451472GA SHQ11451473GA SHQ11451474GA SHQ11471563GA SHQ11471583GA SHQ11471584GA SHQ11471585GA SHQ11471586GA SHQ11471587GA SHQ11471588GA SHQ11471589GA SHQ11471590GA SHQ11471591GA SHQ11471592GA SHQ11481605GA SHQ11481606GA SHQ11481607GA SHQ11481608GA SHQ11481609GA SHQ11481610GA SHQ11481611GA SHQ11481612GA SHQ11481613GA SHQ11481615GA SHQ11481616GA SHQ11481619GA SHQ11481626GA SHQ11481627GA SHQ11481628GA SHQ11481629GA SHQ11481630GA SHQ11481631GA SHQ11481632GA SHQ11481633GA SHQ11481634GA SHQ11481635GA SHQ11481636GA SHQ11481637GA SHQ11481638GA SHQ11481639GA SHQ11481640GA SHQ11481641GA SHQ11481642GA SHQ11481643GA SHQ11481644GA SHQ11481645GA SHQ11481646GA SHQ11481647GA SHQ11481648GA SHQ11481649GA SHQ11481650GA SHQ11481651GA SHQ11481652GA SHQ11481653GA SHQ11481654GA SHQ11481655GA SHQ11481656GA SHQ11481657GA SHQ11481658GA SHQ11481659GA SHQ11481660GA SHQ11481661GA SHQ11481662GA SHQ11481663GA SHQ11481664GA SHQ11481665GA SHQ11481666GA SHQ11481667GA SHQ11481668GA SHQ11481669GA SHQ11481670GA SHQ11481671GA SHQ11481672GA SHQ11481673GA SHQ11481674GA SHQ11481675GA SHQ11481676GA SHQ11481677GA SHQ11481678GA SHQ11481679GA SHQ11481680GA SHQ11481681GA SHQ11481682GA SHQ11481683GA SHQ11481684GA SHQ11481685GA SHQ11481686GA SHQ11481687GA SHQ11481688GA SHQ11481689GA SHQ11481690GA SHQ11481691GA SHQ11481692GA SHQ11481693GA SHQ11481694GA SHQ11481695GA SHQ11481696GA SHQ11481697GA SHQ11481698GA SHQ11481699GA SHQ11481700GA SHQ11481701GA SHQ11481702GA SHQ11481703GA SHQ11481704GA SHQ11481705GA SHQ11481706GA SHQ11481707GA SHQ11481708GA SHQ11481709GA SHQ11481710GA SHQ11481711GA SHQ11481712GA SHQ11481713GA SHQ11481714GA SHQ11481715GA SHQ11481716GA SHQ11481717GA SHQ11481718GA SHQ11481719GA SHQ11481720GA SHQ11481721GA SHQ11481722GA SHQ11481723GA SHQ11481724GA SHQ11481725GA SHQ11481726GA SHQ11481727GA SHQ11481728GA SHQ11481729GA SHQ11481730GA SHQ11481731GA SHQ11481732GA SHQ11481733GA SHQ11481734GA SHQ11481735GA SHQ11481737GA SHQ11481738GA SHQ11481739GA SHQ11481740GA SHQ11481741GA SHQ11481742GA SHQ11481743GA SHQ11481744GA SHQ11481745GA SHQ11481746GA SHQ11481747GA SHQ11481748GA SHQ11481749GA SHQ11481750GA SHQ11481751GA SHQ11481752GA SHQ11481753GA SHQ11481754GA SHQ11481755GA SHQ11481756GA SHQ11481757GA SHQ11481758GA SHQ11481759GA SHQ11481760GA SHQ11481761GA SHQ11481762GA SHQ11481763GA SHQ11481764GA SHQ11481765GA SHQ11481766GA SHQ11481767GA SHQ11481768GA SHQ11481769GA SHQ11481770GA SHQ11481771GA SHQ11481772GA SHQ11481773GA SHQ11481774GA SHQ11481775GA SHQ11481776GA SHQ11481777GA SHQ11481778GA SHQ11481779GA SHQ11481780GA SHQ11481781GA SHQ11481782GA SHQ11481783GA SHQ11481785GA SHQ11481786GA SHQ11481787GA SHQ11481788GA SHQ11481789GA SHQ11481790GA SHQ11481791GA SHQ11481792GA SHQ11481793GA SHQ11481794GA SHQ11481795GA SHQ11481796GA SHQ11481797GA SHQ11481798GA SHQ11481799GA SHQ11481800GA SHQ11481801GA SHQ11481802GA SHQ11481803GA SHQ11481804GA SHQ11481805GA SHQ11481806GA SHQ11481807GA SHQ11481808GA SHQ11481809GA SHQ11481810GA SHQ11481811GA SHQ11481812GA SHQ11481813GA SHQ11481814GA SHQ11481815GA SHQ11481816GA SHQ11481817GA SHQ11481818GA SHQ11481819GA SHQ11481820GA SHQ11481821GA SHQ11481822GA SHQ11481823GA SHQ11481824GA SHQ11481825GA SHQ11481826GA SHQ11481827GA SHQ11481828GA SHQ11481829GA SHQ11481830GA SHQ11481831GA SHQ11481833GA SHQ11481834GA SHQ11481835GA SHQ11481836GA SHQ11481837GA SHQ11481838GA SHQ11481839GA SHQ11481840GA SHQ11481841GA SHQ11481842GA SHQ11491843GA SHQ11491844GA SHQ11491845GA SHQ11491846GA SHQ11491847GA SHQ11491848GA SHQ11491849GA SHQ11491850GA SHQ11491851GA SHQ11491852GA SHQ11491853GA SHQ11491854GA SHQ11491855GA SHQ11491856GA SHQ11491857GA SHQ11491858GA SHQ11491859GA SHQ11491860GA SHQ11491861GA SHQ11491862GA SHQ11491863GA SHQ11491864GA SHQ11491865GA SHQ11491866GA SHQ11491867GA SHQ11491868GA SHQ11491869GA SHQ11491870GA SHQ11491871GA SHQ11491872GA SHQ11491873GA SHQ11491874GA SHQ11491875GA SHQ11491876GA SHQ11491877GA SHQ11491878GA SHQ11491879GA SHQ11491880GA SHQ11491881GA SHQ11491882GA SHQ11491883GA SHQ11491884GA SHQ11491885GA SHQ11491886GA SHQ11491887GA SHQ11491888GA SHQ11491889GA SHQ11491890GA SHQ11491891GA SHQ11491892GA SHQ11491893GA SHQ11491894GA SHQ11491895GA SHQ11491896GA SHQ11491897GA SHQ11491898GA SHQ11491899GA SHQ11491900GA SHQ11491901GA SHQ11491902GA SHQ11491904GA SHQ11491905GA SHQ11491906GA SHQ11491907GA SHQ11491908GA SHQ11491909GA SHQ11491910GA SHQ11491911GA SHQ11491912GA SHQ11491913GA SHQ11491914GA SHQ11491915GA SHQ11491916GA SHQ11491917GA SHQ11491933GA SHQ11491939GA SHQ11491940GA SHQ11491941GA SHQ11491942GA SHQ11491943GA SHQ11491944GA SHQ11491945GA SHQ11491946GA SHQ11491947GA SHQ11491948GA SHQ11491949GA SHQ11491950GA SHQ11491951GA SHQ11491952GA SHQ11491953GA SHQ11491954GA SHQ11491956GA SHQ11491957GA SHQ11491959GA SHQ11491960GA SHQ11491961GA SHQ11491962GA SHQ11491965GA SHQ11491966GA SHQ11491973GA SHQ11491974GA SHQ11491976GA SHQ11491977GA SHQ11491979GA SHQ11491980GA SHQ11491982GA SHQ11491983GA SHQ11491986GA SHQ11491987GA SHQ11491988GA SHQ11491989GA SHQ11491990GA SHQ11491991GA SHQ11491992GA SHQ11491993GA SHQ11491994GA SHQ11491997GA SHQ11491998GA SHQ11491999GA SHQ11492000GA SHQ11492001GA SHQ11492002GA SHQ11492003GA SHQ11492004GA SHQ11492005GA SHQ11492006GA SHQ11492007GA SHQ11492008GA SHQ11492009GA SHQ11502010GA SHQ11502012GA SHQ11502013GA SHQ11502016GA SHQ11502018GA SHQ11502019GA SHQ11502021GA SHQ11502025GA SHQ11502026GA SHQ11502027GA SHQ11502029GA SHQ11502166GA SHQ11512256GA SHQ11512257GA SHQ11512258GA SHQ11512259GA SHQ11512260GA SHQ11512261GA SHQ11512263GA SHQ11512264GA SHQ11512265GA SHQ11512266GA SHQ11512267GA SHQ11512268GA SHQ11512269GA SHQ11512271GA SHQ11512272GA SHQ11512274GA SHQ11512275GA SHQ11512276GA SHQ11512277GA SHQ11512278GA SHQ11512279GA SHQ11512280GA SHQ11512281GA SHQ11512282GA SHQ11512283GA SHQ11512284GA SHQ11512285GA SHQ11512286GA SHQ11512287GA SHQ11512288GA SHQ11512289GA SHQ11512291GA SHQ11512293GA SHQ11512297GA SHQ11512299GA SHQ11512300GA SHQ11512301GA SHQ11512302GA SHQ11512309GA SHQ11512310GA SHQ11512311GA SHQ11512312GA SHQ11512313GA SHQ11512314GA SHQ11512315GA SHQ11512316GA SHQ12012353GA SHQ12012376GA SHQ12012377GA SHQ12012379GA SHQ12012380GA SHQ12012381GA SHQ12022403GA SHQ12022404GA SHQ12022405GA SHQ12022407GA SHQ12022408GA SHQ12022413GA SHQ12022414GA SHQ12022422GA SHQ12022423GA SHQ12022424GA SHQ12022428GA SHQ12022429GA SHQ12022430GA SHQ12022431GA SHQ12022432GA SHQ12022433GA SHQ12022434GA SHQ12022435GA SHQ12022436GA SHQ12022437GA SHQ12022438GA SHQ12032499GA SHQ12032500GA SHQ12032501GAMKE SHQ12032504GA SHQ12032505GA SHQ12032506GA SHQ12032507GA SHQ12042509GA SHQ12042511GA SHQ12042512GA SHQ12042515GA SHQ12042516GA SHQ12042517GA SHQ12042534GA SHQ12042535GA SHQ12042536GA SHQ12042537GA SHQ12042538GA SHQ12042539GA SHQ12042540GA SHQ12042541GA SHQ12042542GA SHQ12042543GA SHQ12042544GA SHQ12042545GA SHQ12042546GA SHQ12042547GA SHQ12042548GA SHQ12042549GA SHQ12042550GA SHQ12042551GA SHQ12042552GA SHQ12042553GA SHQ12042554GA SHQ12042555GA SHQ12042556GA SHQ12042558GA SHQ12042559GA SHQ12042561GA SHQ12052573GA SHQ12052574GA SHQ12052575GA SHQ12052576GA SHQ12052578GA SHQ12052579GA SHQ12052584GA SHQ12052591GA SHQ12052593GA SHQ12052597GA SHQ12062617GA SHQ12062620GA SHQ12062631GA SHQ12062633GA SHQ12072703GA SHQ12082734GA SHQ12082758GA SHQ12082767GA SHQ12082768GA SHQ12082769GA SHQ12082771GA SHQ12082774GA SHQ12082775GA SHQ12082778GA SHQ12092780GA SHQ12092782GA SHQ12092785GA SHQ12092791GA SHQ12092793GA SHQ12092794GA SHQ12092795GA SHQ12092796GA SHQ12092797GA SHQ12092798GA SHQ12092799GA SHQ12092800GA SHQ12092803GA SHQ12092804GA SHQ12092805GA SHQ12092807GA SHQ12092809GA SHQ12092813GA SHQ12092814GA SHQ12092816GA SHQ12092817GA SHQ12092818GA SHQ12092836GA SHQ12092837GA SHQ12092838GA SHQ12092839GA SHQ12092840GA SHQ12092841GA SHQ12092843GA SHQ12092844GA SHQ12102845GA SHQ12102846GA SHQ12102847GA SHQ12102848GA SHQ12102849GA SHQ12102850GA SHQ12102851GA SHQ12102852GA SHQ12102853GA SHQ12102854GA SHQ12102855GA SHQ12102856GA SHQ12102857GA SHQ12102858GA SHQ12102859GA SHQ12102860GA SHQ12102861GA SHQ12102862GA SHQ12102863GA SHQ12102864GA SHQ12102865GA SHQ12102866GA SHQ12102867GA SHQ12102868GA SHQ12102869GA SHQ12102870GA SHQ12102871GA SHQ12102872GA SHQ12102873GA SHQ12102874GA SHQ12102875GA SHQ12102876GA SHQ12102877GA SHQ12102878GA SHQ12102879GA SHQ12102880GA SHQ12102881GA SHQ12102882GA SHQ12102883GA SHQ12102887GA SHQ12102888GA SHQ12102889GA SHQ12102890GA SHQ12102892GA SHQ12102894GA SHQ12102896GA SHQ12102897GA SHQ12102898GA SHQ12102899GA SHQ12102900GA SHQ12102901GA SHQ12102902GA SHQ12102903GA SHQ12102904GA SHQ12102905GA SHQ12102906GA SHQ12102907GA SHQ12102908GA SHQ12102909GA SHQ12112910GA SHQ12112911GA SHQ12112912GA SHQ12112913GA SHQ12112914GA SHQ12112915GA SHQ12112916GA SHQ12112917GA SHQ12112918GA SHQ12112919GA SHQ12112920GA SHQ12112921GA SHQ12112922GA SHQ12122963GA SHQ12122964GA SHQ12122965GA SHQ12122966GA SHQ12122967GA SHQ12122968GA SHQ12143036GA SHQ12143037GA SHQ12143038GA SHQ12143041GA SHQ12143043GA SHQ12143044GA SHQ12143045GA SHQ12143046GA SHQ12143047GA SHQ12153068GA SHQ12153069GA SHQ12153070GA SHQ12153071GA SHQ12153072GA SHQ12153073GA SHQ12153074GA SHQ12153075GA SHQ12153076GA SHQ12153077GA SHQ12153078GA SHQ12153079GA SHQ12153080GA SHQ12153081GA SHQ12153082GA SHQ12153083GA SHQ12153084GA SHQ12153085GA SHQ12153086GA SHQ12153087GA SHQ12153088GA SHQ12153089GA SHQ12153090GA SHQ12153091GA SHQ12153092GA SHQ12153093GA SHQ12153094GA SHQ12153095GA SHQ12153096GA SHQ12153097GA SHQ12163098GA SHQ12163099GA SHQ12163100GA SHQ12163101GA SHQ12163102GA SHQ12163103GA SHQ12163104GA SHQ12163105GA SHQ12163106GA SHQ12163107GA SHQ12163108GA SHQ12163109GA SHQ12163110GA SHQ12163111GA SHQ12163112GA SHQ12163113GA SHQ12163114GA SHQ12163115GA SHQ12163116GA SHQ12163117GA SHQ12163118GA SHQ12163119GA SHQ12163120GA SHQ12163121GA SHQ12163122GA SHQ12163123GA SHQ12163124GA SHQ12163125GA SHQ12163126GA SHQ12163127GA SHQ12163128GA SHQ12163129GA SHQ12163130GA SHQ12163131GA SHQ12163132GA SHQ12163133GA SHQ12163134GA SHQ12163135GA SHQ12163137GA SHQ12163138GA SHQ12163139GA SHQ12163141GA SHQ12163142GA SHQ12173151GA SHQ12173152GA SHQ12173157GA SHQ12173175GA SHQ12173176GA SHQ12173177GA SHQ12173178GA SHQ12173179GA SHQ12173180GA SHQ12173181GA SHQ12173182GA SHQ12173183GA SHQ12173184GA SHQ12173185GA SHQ12173186GA SHQ12173187GA SHQ12173188GA SHQ12173189GA SHQ12173190GA SHQ12173191GA SHQ12173192GA SHQ12173193GA SHQ12173194GA SHQ12173195GA SHQ12173196GA SHQ12173197GA SHQ12183288GA SHQ12183289GA SHQ12183290GA SHQ12183292GA SHQ12183293GA SHQ12183294GA SHQ12183295GA SHQ12183296GA SHQ12183297GA SHQ12183299GA SHQ12183301GA SHQ12183302GA SHQ12183312GA SHQ12183314GA SHQ12183315GA SHQ12183316GA SHQ12183317GA SHQ12193328GA SHQ12193329GA SHQ12193330GA SHQ12193331GA SHQ12193332GA SHQ12193333GA SHQ12193334GA SHQ12193335GA SHQ12193336GA SHQ12193337GA SHQ12193338GA SHQ12193339GA SHQ12193340GA SHQ12193348GA SHQ12193349GA SHQ12193350GA SHQ12193351GA SHQ12193352GA SHQ12193353GA SHQ12193354GA SHQ12193355GA SHQ12193356GA SHQ12193357GA SHQ12193358GA SHQ12193359GA SHQ12193360GA SHQ12193362GA SHQ12193363GA SHQ12193364GA SHQ12193365GA SHQ12193366GA SHQ12193367GA SHQ12193368GA SHQ12193369GA SHQ12193370GA SHQ12193371GA SHQ12203372GA SHQ12203373GA SHQ12203374GA SHQ12203375GA SHQ12203376GA SHQ12203377GA SHQ12203378GA SHQ12203379GA SHQ12203380GA SHQ12203381GA SHQ12203382GA SHQ12203383GA SHQ12203384GA SHQ12203385GA SHQ12203386GA SHQ12203387GA SHQ12203388GA SHQ12203389GA SHQ12203390GA SHQ12203391GA SHQ12203393GA SHQ12203394GA SHQ12203395GA SHQ12203396GA SHQ12203397GA SHQ12203398GA SHQ12203399GA SHQ12203400GA SHQ12203401GA SHQ12203402GA SHQ12203403GA SHQ12203404GA SHQ12203405GA SHQ12203406GA SHQ12203407GA SHQ12203408GA SHQ12203409GA SHQ12203410GA SHQ12203411GA SHQ12203412GA SHQ12203413GA SHQ12203414GA SHQ12203415GA SHQ12203416GA SHQ12203417GA SHQ12203418GA SHQ12203419GA SHQ12203420GA SHQ12203421GA SHQ12213422GA SHQ12213423GA SHQ12213424GA SHQ12213425GA SHQ12213426GA SHQ12213427GA SHQ12213428GA SHQ12213429GA SHQ12213430GA SHQ12213431GA SHQ12213432GA SHQ12213433GA SHQ12213434GA SHQ12213451GA SHQ12213452GA SHQ12213453GA SHQ12213454GA SHQ12213455GA SHQ12213456GA SHQ12213457GA SHQ12213458GA SHQ12213459GA SHQ12213460GA SHQ12213461GA SHQ12213462GA SHQ12213463GA SHQ12213464GA SHQ12213465GA SHQ12213466GA SHQ12213467GA SHQ12213468GA SHQ12213469GA SHQ12213470GA SHQ12213471GA SHQ12213472GA SHQ12213473GA SHQ12213474GA SHQ12213475GA SHQ12213476GA SHQ12213477GA SHQ12213478GA SHQ12213479GA SHQ12213480GA SHQ12213481GA SHQ12213482GA SHQ12213483GA SHQ12213484GA SHQ12213485GA SHQ12213486GA SHQ12213487GA SHQ12213488GA SHQ12213489GA SHQ12213490GA SHQ12213491GA SHQ12213492GA SHQ12213493GA SHQ12213494GA SHQ12213495GA SHQ12213496GA SHQ12213497GA SHQ12213498GA SHQ12213499GA SHQ12213500GA SHQ12213501GA SHQ12213502GA SHQ12213503GA SHQ12213504GA SHQ12213505GA SHQ12213506GA SHQ12213507GA SHQ12213508GA SHQ12213509GA SHQ12213510GA SHQ12213511GA SHQ12213512GA SHQ12213513GA SHQ12213514GA SHQ12213515GA SHQ12213516GA SHQ12213517GA SHQ12213518GA SHQ12213519GA SHQ12213521GA SHQ12213522GA SHQ12213523GA SHQ12213524GA SHQ12213525GA SHQ12213526GA SHQ12213527GAMKE SHQ12213528GA SHQ12213529GA SHQ12213531GA SHQ12213532GA SHQ12213533GA SHQ12213534GA SHQ12213535GA SHQ12213536GA SHQ12213537GA SHQ12213538GA SHQ12213539GA SHQ12213540GA SHQ12213541GA SHQ12213542GA SHQ12213543GA SHQ12213544GA SHQ12233545GA SHQ12233546GA SHQ12233547GA SHQ12233548GA SHQ12233549GA SHQ12233550GA SHQ12233551GA SHQ12233552GA SHQ12233553GA SHQ12233554GA SHQ12233555GA SHQ12233556GA SHQ12233557GA SHQ12233558GA SHQ12233559GA SHQ12243570GA SHQ12243571GA SHQ12243572GA SHQ12243573GA SHQ12243574GA SHQ12243575GA SHQ12243576GA SHQ12243577GA SHQ12243578GA SHQ12243579GA SHQ12243580GA SHQ12243581GA SHQ12243582GA SHQ12243583GA SHQ12243584GA SHQ12243585GA SHQ12243586GA SHQ12243587GA SHQ12243588GA SHQ12243589GA SHQ12243600GA SHQ12243601GA SHQ12243602GA SHQ12243603GA SHQ12243604GA SHQ12243605GA SHQ12243606GA SHQ12243607GA SHQ12243608GA SHQ12243609GA SHQ12243610GA SHQ12243611GA SHQ12243612GA SHQ12243613GA SHQ12243614GA SHQ12243615GA SHQ12243616GA SHQ12243617GA SHQ12243618GA SHQ12243620GA SHQ12243621GA SHQ12243622GA SHQ12243623GA SHQ12243624GA SHQ12243625GA SHQ12243626GA SHQ12243627GA SHQ12243628GA SHQ12243629GA SHQ12243630GA SHQ12243631GA SHQ12243632GA SHQ12243634GA SHQ12243635GA SHQ12243636GA SHQ12243638GA SHQ12243639GA SHQ12243640GA SHQ12243641GA SHQ12243642GA SHQ12243643GA SHQ12243647GA SHQ12243650GA SHQ12243652GA SHQ12243653GA SHQ12243656GA SHQ12243658GA SHQ12243659GA SHQ12243660GA SHQ12243661GA SHQ12293781GA SHQ12303817GA SHQ12303818GA SHQ12303819GA SHQ12303820GA SHQ12303821GA SHQ12303822GA SHQ12303823GA SHQ12303824GA SHQ12303825GA SHQ12303826GA SHQ12303827GA SHQ12303828GA SHQ12303829GA SHQ12303830GA SHQ12303831GA SHQ12303832GA SHQ12303833GA SHQ12303834GA SHQ12303835GA SHQ12303836GA SHQ12303837GA SHQ12303838GA SHQ12303839GA SHQ12303840GA SHQ12303841GA SHQ12303843GA SHQ12303844GA SHQ12303845GA SHQ12303846GA SHQ12323948GA SHQ12344008GA SHQ12374133GA SHQ12344017GA SHQ12303809GA SHQ12263684GA SHQ12303816GA SHQ12323909GA SHQ12313903GA SHQ12344022GA SHQ12344027GA SHQ12344034GA SHQ12344036GA SHQ12344006GA SHQ12333984GA SHQ12323940GA SHQ12323956GA SHQ12283729GA SHQ12323943GA SHQ12323941GA SHQ12323937GA SHQ12323931GA SHQ12323935GA SHQ12323929GA SHQ12323930GA SHQ12323936GA SHQ12283730GA SHQ12263711GA SHQ12263715GA SHQ12283727GA SHQ12283728GA SHQ12323927GA SHQ12263708GA SHQ12313870GA SHQ12243648GA SHQ12243649GA SHQ12323938GA SHQ12323926GA SHQ12313875GA SHQ12323933GA SHQ12313872GA SHQ12323928GA SHQ12333981GA SHQ12243646GA SHQ12333979GA SHQ12323923GA SHQ12333968GA SHQ12323955GA SHQ12333986GA SHQ12323950GA SHQ12333972GA SHQ12323949GA SHQ12243644GA SHQ12333977GA SHQ12263710GA SHQ12323947GA SHQ12323939GA SHQ12323918GA SHQ12323954GA SHQ12333976GA SHQ12303855GA SHQ12263707GA SHQ12333983GA SHQ12263716GA SHQ12333978GA SHQ12323922GA SHQ12333985GA SHQ12333967GA SHQ12333970GA SHQ12323932GA SHQ12333973GA SHQ12323946GA SHQ12323919GA SHQ12333980GA SHQ12283734GA SHQ12333982GA SHQ12323917GA SHQ12313896GA SHQ12323924GA SHQ12303848GA SHQ12323953GA SHQ12293751GA SHQ12303853GA SHQ12323945GA SHQ12303854GA SHQ12293748GA SHQ12303852GA SHQ12293747GA SHQ12313881GA SHQ12283733GA SHQ12323942GA SHQ12303851GA SHQ12303856GA SHQ12283732GA SHQ12303849GA SHQ12323920GA SHQ12243645GA SHQ12323951GA SHQ12323944GA SHQ12303847GA SHQ12263713GA SHQ12323921GA SHQ12263712GA SHQ12344005GA SHQ12344012GA SHQ12243663GA SHQ12303813GA SHQ12303815GA SHQ12303812GA SHQ12313905GA SHQ12344010GA SHQ12303811GA SHQ12323911GA SHQ12344018GA SHQ12253668GA SHQ12344032GA SHQ12344016GA SHQ12323913GA SHQ12344015GA SHQ12344023GA SHQ12344029GA SHQ12344025GA SHQ12344028GA SHQ12323915GA SHQ12344020GA SHQ12344021GA SHQ12313906GA SHQ12344024GA SHQ12323916GA SHQ12263686GA SHQ12313899GA SHQ12323914GA SHQ12344026GA SHQ12344033GA SHQ12313904GA SHQ12344030GA SHQ12313898GA SHQ12344031GA SHQ12303814GA SHQ12303807GA SHQ12323910GA SHQ12303810GA SHQ12323908GA SHQ12344019GA SHQ12323907GA SHQ12263685GA SHQ12313900GA SHQ12303808GA SHQ12354078GA SHQ12344042GA SHQ12344009GA SHQ12344038GA SHQ12344035GA SHQ12344037GA SHQ12293752GA SHQ12293753GA SHQ12344041GA SHQ12253667GA SHQ12344013GA SHQ12354051GA SHQ12243662GA SHQ12253671GA SHQ12344014GA SHQ12344011GA SHQ12243655GA SHQ12364106GA SHQ12243657GA SHQ12364108GA SHQ12213520GA SHQ12374125GA SHQ12344040GA SHQ12344039GA SHQ12344043GA SHQ12344044GA SHQ12253673GA SHQ12253670GA SHQ12243664GA SHQ12344007GA SHQ12253672GA SHQ12313871GA SHQ12243619GA SHQ12313876GA SHQ12374137GA SHQ12313893GA SHQ12313869GA SHQ12313891GA SHQ12313884GA SHQ12313882GA SHQ12313895GA SHQ12313892GA SHQ12313894GA SHQ12313883GA SHQ12313868GA SHQ12313873GA SHQ12263714GA SHQ12263709GA SHQ12293756GA SHQ12293754GA SHQ12293749GA SHQ12293755GA SHQ12293750GA SHQ12243633GA SHQ12243654GA SHQ12283736GA SHQ12283735GA SHQ12243637GA SHQ12354080GA SHQ12354054GA SHQ12354061GA SHQ12354083GA SHQ12354058GA SHQ12354079GA SHQ12354059GA SHQ12354057GA SHQ12354076GA SHQ12354062GA SHQ12354060GA SHQ12354055GA SHQ12364111GA SHQ12092802GA SHQ12354056GA SHQ12354064GA SHQ12354052GA SHQ12354050GA SHQ12354063GA SHQ12183298GA SHQ12313862GA SHQ12374128GA SHQ12313861GA SHQ12364110GA SHQ12364103GA SHQ12364112GA SHQ12364107GA SHQ12364104GA SHQ12364105GA SHQ12364109GA SHQ12374123GA SHQ12374132GA SHQ12374129GA SHQ12374127GA SHQ12374126GA SHQ12374124GA SHQ12374131GA SHQ12313874GA SHQ12323934GA SHQ12243651GA SHQ12323925GA SHQ12374136GA SHQ12374135GA SHQ12374138GA SHQ12323952GA SHQ12374134GA SHQ12313858GA SHQ12313857GA SHQ12183300GA SHQ12313865GA SHQ12313866GA SHQ12374128GAMKE SHQ12310004SA SD010397932GA SD010470438GA SD010470444GA SD010470483GA SD010470504GA SD010470538GA SD010470672GA SD010470676GA SD010480997GA SD010481008GA SD010481013GA SD010481224GA SD010481229GA SD010481238GA SD010481246GA SD010481248GA SD010491322GA SD010491323GA SD010491453GA SD010491456GA SD010491497GA SD010491503GA SD010491599GA SD010501862GA SD010501969GA SD010501974GA SD010501976GA SD010501985GA SD010501989GA SD010502017GA SD010502019GA SD010502027GA SD010502028GA SD010502029GA SD010502033GA SD010502034GA SD010522471GA SD010522488GA SD010522492GA SD010522493GA SD011124035GA SD011134380GA SD011144418GA SD011144422GA SD011144428GA SD011144430GA SD011144463GA SD011144465GA SD011144467GA SHQ11451452GA SHQ12299288SA SHQ12309578SA SHQ12310020SA SHQ12310023SA SHQ12310026SA SHQ12310028SA SD010138497GA SD010408251GA SD010408256GA SD010408261GA SD010408317GA SD011113982GA SD011154515GA SD011194932GA SD011194933GA SD011194936GA SD011194937GA SD011194938GA SD011194939GA SD011194940GA SD011194941GA SD011194942GA SD011194949GA SHQ12022392GA SHQ12022396GA SHQ12062624GA SHQ12062625GA SHQ12062628GA SHQ12062632GA SHQ12062634GA SHQ12289204SA SHQ12289206SA SHQ12299257SA SHQ12299258SA SHQ12299259SA SHQ12299262SA SHQ12299263SA SHQ12299266SA SHQ12299267SA SHQ12299268SA SHQ12299270SA SHQ12299277SA SHQ12299281SA SHQ12299282SA SHQ12299284SA SHQ12299286SA SHQ12299291SA SHQ12299296SA SHQ12299302SA SHQ12299303SA SHQ12299304SA SHQ12299307SA SHQ12320277SA SHQ12320287SA SHQ12320286SA SHQ12320295SA SD011061347SA SD010387538GA SD010408376GA SD010428939GA SD010449578GA SD010449635GA SD010449652GA SD010449653GA SD010449666GA SD010449667GA SD010449669GA SD010449688GA SD010449693GA SD010449695GA SD010449700GA SD010450048GA SD010450051GA SD010450053GA SD010450054GA SD010450057GA SD010450058GA SD010450063GA SD010450064GA SD010450065GA SD010450066GA SD010450068GA SD010450069GA SD010450072GA SD010450074GA SD010450075GA SD010450083GACOMPLETE SD010450084GA SD010450085GACOMPLETE SD010459707GA SD010459712GA SD010459715GA SD010459725GA SD010459726GA SD010459727GA SD010459734GA SD010459736GA SD010459737GA SD010459742GA SD010459743GA SD010459744GA SD010459782GA SD010459790GA SD010459792GA SD010459793GA SD010459797GA SD010459801GA SD010459804GA SD010459805GA SD010459813GA SD010459821GA SD010459829GA SD010459961GA SD010459963GA SD010459966GA SD010459968GA SD010459971GA SD010459975GA SD010459976GA SD010459979GA SD010459982GA SD010459984GA SD010460091GA SD010460092GA SD010460096GA SD010460098GA SD010460101GA SD010460102GA SD010460107GA SD010460110GA SD010460111GA SD010491647GA SD011030563SA SD011030567SA SD011030575SA SD011030583SA SD011051205SA SD011061464SA SD011133563SA SD011184461SA SD011194635SA SD011204669SA SD011204711SA SD011204721SA SD011204730SA SD011225002SA SD011225009SA SD011225031SA SD011225032SA SD011225040SA SHQ11443536SA SHQ11443540SA SHQ12045813SA SHQ12055957SA SHQ12055966SA SHQ12055968SA SHQ12055972SA SHQ12055974SA SHQ12055975SA SHQ12055980SA SHQ12055985SA SHQ12066085SA SHQ12278938SA SHQ12289217SA SD011123266SA SHQ12340993SA SHQ12340992SA SHQ12340998SA SHQ12310066SA SHQ12310065SA SHQ12310067SA SHQ12320307SA SHQ12320316SA SHQ12320318SA SHQ12320312SA SHQ12310075SA SHQ12310085SA SHQ12310069SA SHQ12310070SA SHQ12320321SA SHQ12310083SA SHQ12310073SA SHQ12310081SA SHQ12309683SA SHQ12310084SA SHQ12310068SA SHQ12320345SA SHQ12217914SA SHQ12289083SA SHQ12309725SA SHQ12289220SA SHQ12289228SA SHQ12309772SA SHQ12320315SA SHQ12320320SA SHQ12320317SA SHQ12320247SA SHQ12320152SA SHQ12310014SA SHQ12320197SA SHQ12320352SA SHQ12289131SA SHQ12320123SA SHQ12320148SA SHQ12320300SA SHQ12320299SA SHQ12320133SA SHQ12320149SA SHQ12320132SA SHQ12320353SA SHQ12330550SA SHQ12253678GA SHQ12330779SA SHQ12289079SA SHQ12330782SA SHQ12330780SA SHQ12330540SA SHQ12330541SA SHQ12330536SA SHQ12330821SA SHQ12330823SA SHQ12310061SA SHQ12310062SA SHQ12330687SA SHQ12330690SA SHQ12330683SA SHQ12330682SA SHQ12330811SA SHQ12330548SA SHQ12340897SA SHQ12340902SA SHQ12340896SA SHQ12340906SA SHQ12340895SA SHQ12340905SA SHQ12340898SA SHQ12340899SA SHQ12340900SA SHQ12340903SA SHQ12340994SA SHQ12320397SA SHQ12340904SA SHQ12330584SA SHQ12330506SA SHQ12330787SA SHQ12330503SA SHQ12330791SA SHQ11484903SA SHQ11484872SA SHQ11484886SA SHQ12361197SA SHQ12133002GA SHQ12132996GA SHQ12143016GA SHQ12173154GA SHQ12173153GA SHQ12371300SA SHQ12319967SA SHQ12371302SA SHQ12371428SA SHQ12371429SA SD011033038GA SD011073531GA SD011083583GA SHQ12371298SA SHQ12371299SA SHQ12381580SA SHQ12381584SA SD010179616GA SD010314239GA SD010314386GA SD010365946GA SD010366101GA SD010376757GA SD010376761GA SD010377085GA SD010377107GA SD010377123GA SD010377130GA SD010377132GA SD010387264GA SD010387279GA SD010387469GA SD010387676GA SD010387679GA SD010387680GA SD010387684GA SD010387789GA SD010387791GA SD010387793GA SD010387805GA SD010387808GA SD010387809GA SD010387811GA SD010387812GA SD010408315GA SD010428936GA SD010429055GA SD010429079GA SD010429081GA SD010429083GA SD010470789GA SD010470791GA SD010491353GA SD010491550GA SD010491551GA SD010500217SA SD010500225SA SD010500240SA SD010500263SA SD010501998GA SD010502012GA SD010532873GA SD010532879GA SD011061506SA SD011061508SA SD011061523SA SD011071524SA SD011071533SA SD011071534SA SD011071535SA SD011071536SA SD011071538SA SD011071540SA SD011071549SA SD011071552SA SD011071619SA SD011071729SA SD011071730SA SD011071731SA SD011071732SA SD011071733SA SD011071735SA SD011071736SA SD011071737SA SD011071738SA SD011072100SA SD011072101SA SD011072104SA SD011072107SA SD011072109SA SD011072110SA SD011072113SA SD011072114SA SD011072115SA SD011072116SA SD011072117SA SD011082302SA SD011082303SA SD011082304SA SD011082306SA SD011082309SA SD011082310SA SD011082312SA SD011082313SA SD011123262SA SD011174321SA SD011174330SA SD011194533SA SD011214894SA SHQ11403097SA SHQ11403139SA SHQ11423177SA SHQ11423182SA SHQ11453841SA SHQ11453848SA SHQ11454072SA SHQ12012354GA SHQ12012355GA SHQ12012356GA SHQ12012357GA SHQ12012358GA SHQ12012359GA SHQ12012360GA SHQ12012361GA SHQ12012362GA SHQ12012363GA SHQ12012364GA SHQ12012366GA SHQ12022389GA SHQ12022391GA SHQ12022393GA SHQ12022446GA SHQ12022448GA SHQ12062646GA SHQ12062649GA SHQ12062651GA SHQ12062653GA SHQ12062657GA SHQ12062658GA SHQ12062659GA SHQ12062660GA SHQ12062661GA SHQ12062666GA SHQ12062668GA SHQ12062669GA SHQ12062670GA SHQ12062671GA SHQ12062672GA SHQ12062673GA SHQ12062674GA SHQ12062675GA SHQ12062676GA SHQ12062677GA SHQ12062678GA SHQ12082727GA SHQ12082728GA SHQ12082770GA SHQ12082772GA SHQ12092781GA SHQ12092783GA SHQ12092788GA SHQ12092789GA SHQ12092792GA SHQ12112939GA SHQ12112941GA SHQ12112947GA SHQ12112948GA SHQ12112950GA SHQ12112951GA SHQ12112952GA SHQ12112953GA SHQ12112954GA SHQ12112955GA SHQ12112956GA SHQ12112957GA SHQ12112958GA SHQ12143039GA SHQ12143040GA SHQ12143042GA SHQ12163136GA SHQ12163140GA SHQ12173148GA SHQ12173149GA SD011071553SA SHQ11401226GA SHQ11401227GA SHQ11411311GA SD010366391GA SD010418514GA SD010459799GA SD010459951GA SD010491512GA SD010501977GA SHQ12066059SA SHQ12066187SA SHQ12066047SA SD010272434GA SD010303379GA SD010303440GA SD010314420GA SD010324505GA SD010324757GA SD010334915GA SD010345218GA SD010345256GA SD010366327GA SD010376664GA SD010376697GA SD010376698GA SD010376699GA SD010376700GA SD010376701GA SD010376702GA SD010376810GA SD010387299GA SD010387654GA SD010387655GA SD010387661GA SD010387664GA SD010387776GA SD010387788GA SD010397992GA SD010408038GA SD010408041GA SD010408042GA SD010408044GA SD010408068GA SD010408072GA SD010408075GA SD010408078GA SD010408080GA SD010408087GA SD010408089GA SD010408239GA SD010408240GA SD010418666GA SD010418895GA SD010429098GA SD010429131GA SD010429133GA SD010429144GA SD010429151GA SD010429154GA SD010429155GA SD010449505GA SD010460179GA SD010460215GA SD010480110SA SD010480817GA SD010480878GA SD010480898GA SD010480904GA SD010480907GA SD010481074GA SD010481122GA SD010491563GA SD010491566GA SD010491568GA SD010491569GA SD010491574GA SD010491577GA SD010491578GA SD010491583GA SD010502119GA SD010502301GA SD010510359SA SD010512394GA SD010512397GA SD010520430SA SD010522552GA SD010532835GA SD010532843GA SD010532844GA SD010532845GA SD011020492SA SD011030557SA SD011030558SA SD011030561SA SD011030571SA SD011032911GA SD011032913GA SD011032915GA SD011032921GA SD011032971GA SD011032997GA SD011033002GA SD011040795SA SD011040812SA SD011040815SA SD011040817SA SD011040873SA SD011040882SA SD011040895SA SD011040898SA SD011050997SA SD011050998SA SD011051021SA SD011051036SA SD011051040SA SD011051075SA SD011051076SA SD011051085SA SD011051086SA SD011061278SA SD011061280SA SD011061284SA SD011061285SA SD011061293SA SD011061357SA SD011061366SA SD011061369SA SD011061372SA SD011061375SA SD011061376SA SD011061377SA SD011061380SA SD011061386SA SD011061394SA SD011071572SA SD011071573SA SD011071676SA SD011071703SA SD011071816SA SD011071823SA SD011071831SA SD011071833SA SD011071891SA SD011072013SA SD011072122SA SD011072137SA SD011072139SA SD011082164SA SD011082170SA SD011082286SA SD011082358SA SD011092432SA SD011112691SA SD011112744SA SD011112748SA SD011112749SA SD011112753SA SD011112760SA SD011112766SA SD011112775SA SD011112779SA SD011112780SA SD011112782SA SD011112783SA SD011112784SA SD011112786SA SD011112787SA SD011112797SA SD011112803SA SD011112804SA SD011112805SA SD011122806SA SD011122810SA SD011122816SA SD011122818SA SD011123062SA SD011123063SA SD011123065SA SD011123067SA SD011123068SA SD011123069SA SD011123071SA SD011123075SA SD011123076SA SD011123077SA SD011123078SA SD011123080SA SD011123081SA SD011123082SA SD011123083SA SD011133674SA SD011133677SA SD011133681SA SD011133684SA SD011133685SA SD011133686SA SD011133687SA SD011143880SA SD011143883SA SD011143884SA SD011143887SA SD011143888SA SD011143892SA SD011143895SA SD011143897SA SD011143898SA SD011143900SA SD011143901SA SD011143905SA SD011143906SA SD011143958SA SD011143985SA SD011164074SA SD011164084SA SD011164104SA SD011164109SA SD011164112SA SD011164114SA SD011174181SA SD011174182SA SD011174265SA SD011184352SA SD011184361SA SD011184371SA SD011184372SA SD011184374SA SD011184375SA SD011194516SA SD011194528SA SD011194570SA SD011194609SA SD011235107SA SHQ11184457SA SHQ11230100SA SHQ11230101SA SHQ11230116SA SHQ11230118SA SHQ11230123SA SHQ11250283SA SHQ11250295SA SHQ11250298SA SHQ11250313SA SHQ11250314SA SHQ11250387SA SHQ11250392SA SHQ11250396SA SHQ11250504SA SHQ11260529SA SHQ11260531SA SHQ11260538SA SHQ11260540SA SHQ11260541SA SHQ11260542SA SHQ11260543SA SHQ11260546SA SHQ11280601SA SHQ11280645SA SHQ11280654SA SHQ11280763SA SHQ11280766SA SHQ11280769SA SHQ11280788SA SHQ11290829SA SHQ11290832SA SHQ11291009SA SHQ11291015SA SHQ11291016SA SHQ11291025SA SHQ11291029SA SHQ11291037SA SHQ11291038SA SHQ11291049SA SHQ11291055SA SHQ11291063SA SHQ11291070SA SHQ11291075SA SHQ11301132SA SHQ11301136SA SHQ11301156SA SHQ11301185SA SHQ11301350SA SHQ11311372SA SHQ11311409SA SHQ11311410SA SHQ11311426SA SHQ11311440SA SHQ11321738SA SHQ11321749SA SHQ11321807SA SHQ11331978SA SHQ11332101SA SHQ11332110SA SHQ11332113SA SHQ11342187SA SHQ11342195SA SHQ11352354SA SHQ11352366SA SHQ11352368SA SHQ11352389SA SHQ11352398SA SHQ11352412SA SHQ11362609SA SHQ11362630SA SHQ11362631SA SHQ11372907SA SHQ11383032SA SHQ11383040SA SHQ11423237SA SHQ11423342SA SHQ11423349SA SHQ11423357SA SHQ11443557SA SHQ11443682SA SHQ11443684SA SHQ11443694SA SHQ11443695SA SHQ11443698SA SHQ11443704SA SHQ11443765SA SHQ11453777SA SHQ11453875SA SHQ11453876SA SHQ11453878SA SHQ11453879SA SHQ11453891SA SHQ11453899SA SHQ11454058SA SHQ11464081SA SHQ11464084SA SHQ11464089SA SHQ11464101SA SHQ11464102SA SHQ11464103SA SHQ11464105SA SHQ11464106SA SHQ11464114SA SHQ11464127SA SHQ11464128SA SHQ11464130SA SHQ11464184SA SHQ11464185SA SHQ11464186SA SHQ11464188SA SHQ11464189SA SHQ11464190SA SHQ11464191SA SHQ11464192SA SHQ11464248SA SHQ11464355SA SHQ11464360SA SHQ11464363SA SHQ11464365SA SHQ11464366SA SHQ11464368SA SHQ11464369SA SHQ11464370SA SHQ11464371SA SHQ11464372SA SHQ11474418SA SHQ11474445SA SHQ11474453SA SHQ11474513SA SHQ11474517SA SHQ11474519SA SHQ11474523SA SHQ11474529SA SHQ11474535SA SHQ11474542SA SHQ11474543SA SHQ11484648SA SHQ11484887SA SHQ11484890SA SHQ11484891SA SHQ11484896SA SHQ11495052SA SHQ11495056SA SHQ11495069SA SHQ11505278SA SHQ11505286SA SHQ11505376SA SHQ12015455SA SHQ12015466SA SHQ12015471SA SHQ12025524SA SHQ12025528SA SHQ12035541SA SHQ12035548SA SHQ12035549SA SHQ12035550SA SHQ12035551SA SHQ12035552SA SHQ12035553SA SHQ12035556SA SHQ12035557SA SHQ12035558SA SHQ12035559SA SHQ12035560SA SHQ12035562SA SHQ12035563SA SHQ12035574SA SHQ12035580SA SHQ12035581SA SHQ12035583SA SHQ12045775SA SHQ12045787SA SHQ12045788SA SHQ12055915SA SHQ12055918SA SHQ12055922SA SHQ12055924SA SHQ12055927SA SHQ12066035SA SHQ12066036SA SHQ12066039SA SHQ12066182SA SHQ12066183SA SHQ12066192SA SHQ12072695GA SHQ12072698GA SHQ12096352SA SHQ12096353SA SHQ12096356SA SHQ12096357SA SHQ12096359SA SHQ12096360SA SHQ12096361SA SHQ12096362SA SHQ12096363SA SHQ12096364SA SHQ12096365SA SHQ12096366SA SHQ12096367SA SHQ12096368SA SHQ12096369SA SHQ12096371SA SHQ12096373SA SHQ12096374SA SHQ12106501SA SHQ12116586SA SHQ12116589SA SHQ12116593SA SHQ12116594SA SHQ12116595SA SHQ12156984SA SHQ12157049SA SHQ12167073SA SHQ12167077SA SHQ12167099SA SHQ12167100SA SHQ12167104SA SHQ12167114SA SHQ12167166SA SHQ12177256SA SHQ12177257SA SHQ12177263SA SHQ12177271SA SHQ12177275SA SHQ12177406SA SHQ12187496SA SHQ12197554SA SHQ12197556SA SHQ12197560SA SHQ12197561SA SHQ12197563SA SHQ12207820SA SHQ12207834SA SHQ12207837SA SHQ12207838SA SHQ12217953SA SHQ12217954SA SHQ12217955SA SHQ12217959SA SHQ12217960SA SHQ12217979SA SHQ12217980SA SHQ12217983SA SHQ12217984SA SHQ12218004SA SHQ12218016SA SHQ12228177SA SHQ12228179SA SHQ12228200SA SHQ12228202SA SHQ12228203SA SHQ12228204SA SHQ12228207SA SHQ12228208SA SHQ12228215SA SHQ12228277SA SHQ12228279SA SHQ12238308SA SHQ12238310SA SHQ12238312SA SHQ12238313SA SHQ12238328SA SHQ12238337SA SHQ12238338SA SHQ12238340SA SHQ12238341SA SHQ12238342SA SHQ12238343SA SHQ12238344SA SHQ12238345SA SHQ12238346SA SHQ12238348SA SHQ12238349SA SHQ12238351SA SHQ12238360SA SHQ12238361SA SHQ12238362SA SHQ12238363SA SHQ12238364SA SHQ12238365SA SHQ12238367SA SHQ12238368SA SHQ12238512SA SHQ12238518SA SHQ12238525SA SHQ12238529SABOP SHQ12248590SA SHQ12248591SA SHQ12248592SA SHQ12248593SA SHQ12248594SA SHQ12248595SA SHQ12248596SA SHQ12248604SA SHQ12248645SA SHQ12248647SA SHQ12248677SA SHQ12248690SA SHQ12248691SA SHQ12248697SA SHQ12248709SA SHQ12258833SA SHQ12268859SA SHQ12268919SA SHQ12278944SA SHQ12278945SA SHQ12278947SA SHQ12278948SA SHQ12278949SA SHQ12278951SA SHQ12278956SA SHQ12278957SA SHQ12278959SA SHQ12278960SA SHQ12278965SA SHQ12278966SA SHQ12278969SA SHQ12289095SA SHQ12289103SA SHQ12289107SA SHQ12289178SA SHQ12289179SA SHQ12289180SA SHQ12289182SA SHQ12289252SA SHQ12299310SA SHQ12299313SA SHQ12299314SA SHQ12299368SA SHQ12299427SA SHQ12299432SA SHQ12299433SA SHQ12309655SA SHQ12309671SA SHQ12309672SA SHQ12309676SA SHQ12309681SA SHQ12309686SA SHQ12309689SA SHQ12309691SA SHQ12309716SA SHQ12310021SA SHQ12310012SA SHQ12310015SA SHQ12279026SA SHQ12319993SA SHQ12320338SA SHQ12319963SA SHQ12310082SA SHQ12371376SA SHQ12330686SA SHQ12330685SA SHQ12371381SA SHQ12320439SA SHQ12371311SA SHQ12371315SA SHQ12371325SA SHQ12371307SA SHQ12371309SA SHQ12371312SA SHQ12371314SA SHQ12391667SA SHQ12391675SA SD010127467GA SD010138103GA SD010303432GA SD010303530GA SD010303611GA SD010314085GA SD010314113GA SD010324649GA SD010335165GA SD010335189GA SD010345230GA SD010345316GA SD010345332GA SD010345337GA SD010345460GA SD010345529GA SD010345533GA SD010345535GA SD010345539GA SD010345542GA SD010345544GA SD010345545GA SD010345546GA SD010345547GA SD010345548GA SD010365917GA SD010365950GA SD010365994GA SD010365996GA SD010366001GA SD010366085GA SD010366087GA SD010366089GA SD010366096GA SD010366098GA SD010366102GA SD010366103GA SD010366231GA SD010366237GA SD010366239GA SD010366270GA SD010366273GA SD010366339GA SD010366340GA SD010366341GA SD010366346GA SD010366349GA SD010366355GA SD010376714GA SD010376719GA SD010376724GA SD010376731GA SD010376736GA SD010376762GA SD010376934GA SD010376938GA SD010376940GA SD010376941GA SD010376943GA SD010387188GA SD010387189GA SD010387197GA SD010387198GA SD010387200GA SD010387206GA SD010387207GA SD010387210GA SD010387214GA SD010387215GA SD010387222GA SD010387224GA SD010387225GA SD010387234GA SD010387298GA SD010387300GA SD010387301GA SD010387303GA SD010387304GA SD010387305GA SD010387306GA SD010387360GA SD010387363GA SD010387396GA SD010387441GA SD010387445GA SD010387651GA SD010387671GA SD010387683GA SD010387685GA SD010387689GA SD010387774GA SD010387777GA SD010387794GA SD010387795GA SD010387796GA SD010387801GA SD010387817GA SD010387818GA SD010387819GA SD010387820GA SD010387821GA SD010387824GA SD010387826GA SD010387829GA SD010387833GA SD010387836GA SD010387837GA SD010387840GA SD010397961GA SD010408047GA SD010408105GA SD010408107GA SD010408111GA SD010408114GA SD010408116GA SD010408119GA SD010408121GA SD010408126GA SD010408130GA SD010408137GA SD010408141GA SD010408144GA SD010408149GA SD010408150GA SD010408172GA SD010408207GA SD010408210GA SD010408233GA SD010408235GA SD010408250GA SD010408255GA SD010408258GA SD010408301GA SD010408347GA SD010408348GA SD010408349GA SD010408353GA SD010408356GA SD010418461GA SD010418475GA SD010418479GA SD010418594GA SD010418595GA SD010418596GA SD010418669GA SD010418689GA SD010418690GA SD010418691GA SD010418692GA SD010418693GA SD010418698GA SD010418701GA SD010418779GA SD010418782GA SD010418783GA SD010418784GA SD010418785GA SD010418792GA SD010418793GA SD010418794GA SD010418795GA SD010418799GA SD010418801GA SD010418839GA SD010418878GA SD010418880GA SD010418883GA SD010418884GA SD010418905GA SD010418916GA SD010418917GA SD010428929GA SD010428934GA SD010428940GA SD010428986GA SD010428989GA SD010429053GA SD010429061GA SD010429073GA SD010429080GA SD010429082GA SD010429087GA SD010429105GA SD010429115GA SD010429121GA SD010429125GA SD010429129GA SD010430003SA SD010449503GA SD010459708GA SD010459803GA SD010460137GA SD010460172GA SD010460173GA SD010460174GA SD010460270GA SD010460281GA SD010470469GA SD010470473GA SD010470536GA SD010470588GA SD010470598GA SD010470623GA SD010470632GA SD010470698GA SD010470707GA SD010470732GA SD010480821GA SD010480861GA SD010480870GA SD010480879GA SD010480909GA SD010480989GA SD010481021GA SD010481032GA SD010481041GA SD010481127GA SD010481130GA SD010481131GA SD010481136GA SD010481138GA SD010481149GA SD010481172GA SD010481179GA SD010481181GA SD010481183GA SD010481185GA SD010481192GA SD010481219GA SD010481226GA SD010481233GA SD010491385GA SD010491390GA SD010491437GA SD010491439GA SD010491445GA SD010491558GA SD010491579GA SD010491580GA SD010491587GA SD010491588GA SD010491589GA SD010491591GA SD010491594GA SD010491601GA SD010491604GA SD010491605GA SD010491610GA SD010491621GA SD010491623GA SD010491624GA SD010491629GA SD010491631GA SD010491685GA SD010491694GA SD010500213SA SD010500215SA SD010500222SA SD010500242SA SD010500248SA SD010500260SA SD010500264SA SD010501742GA SD010501762GA SD010501763GA SD010501975GA SD010501979GA SD010501983GA SD010501984GA SD010502024GA SD010502030GA SD010502031GA SD010502113GA SD010502118GA SD010502122GA SD010502123GA SD010502124GA SD010502128GA SD010502129GA SD010502146GA SD010502185GA SD010502234GA SD010502238GA SD010502309GA SD010510337SA SD010512405GA SD010512408GA SD010512419GA SD010512420GA SD010520373SA SD010520435SA SD010520436SA SD010522548GA SD010522557GA SD010532818GA SD010532822GA SD010532823GA SD010532825GA SD010532826GA SD010532828GA SD010532829GA SD010532836GA SD010532841GA SD011020494SA SD011020516SA SD011020517SA SD011030608SA SD011030618SA SD011033010GA SD011033080GA SD011040713SA SD011040715SA SD011050969SA SD011050976SA SD011051022SA SD011051193SA SD011061286SA SD011061291SA SD011061297SA SD011061298SA SD011061299SA SD011061300SA SD011061303SA SD011061311SA SD011061314SA SD011061316SA SD011061337SA SD011061351SA SD011061353SA SD011061391SA SD011061393SA SD011061395SA SD011061405SA SD011061406SA SD011061409SA SD011061410SA SD011061433SA SD011061442SA SD011061470SA SD011061475SA SD011061481SA SD011071556SA SD011071561SA SD011071568SA SD011071570SA SD011071571SA SD011071598SA SD011071600SA SD011071620SA SD011071621SA SD011071634SA SD011071652SA SD011071653SA SD011071658SA SD011071670SA SD011071673SA SD011071727SA SD011071735SABOP SD011071817SA SD011071825SA SD011071874SA SD011071888SA SD011071903SA SD011072059SA SD011072063SA SD011072077SA SD011072089SA SD011072124SA SD011072127SA SD011072128SA SD011072131SA SD011072134SA SD011072140SA SD011073546GA SD011082162SA SD011082193SA SD011082198SA SD011082199SA SD011082203SA SD011082208SA SD011082210SA SD011082213SA SD011082221SA SD011082222SA SD011082228SA SD011082229SA SD011082230SA SD011082231SA SD011082232SA SD011082233SA SD011082235SA SD011082239SA SD011082241SA SD011082252SA SD011082334SA SD011083584GA SD011112564SA SD011112587SA SD011112638SA SD011112680SA SD011112705SA SD011112724SA SD011112729SA SD011112730SA SD011113945GA SD011113947GA SD011113953GA SD011113955GA SD011113957GA SD011122809SA SD011122814SA SD011122898SA SD011122930SA SD011122954SA SD011122957SA SD011122970SA SD011122990SA SD011122995SA SD011122996SA SD011123008SA SD011123018SA SD011123022SA SD011123138SA SD011123191SA SD011123295SA SD011123310SA SD011123318SA SD011123331SA SD011123332SA SD011123351SA SD011123352SA SD011123357SA SD011123362SA SD011133386SA SD011133401SA SD011133418SA SD011133441SA SD011133622SA SD011133700SA SD011133702SA SD011133704SA SD011133706SA SD011133736SA SD011133740SA SD011133741SA SD011133743SA SD011143759SA SD011143789SA SD011143791SA SD011143800SA SD011143809SA SD011143814SA SD011143816SA SD011143821SA SD011143908SA SD011144470GA SD011153998SA SD011154000SA SD011154009SA SD011164076SA SD011164127SA SD011174154SA SD011174156SA SD011174270SA SD011174273SA SD011174290SA SD011174300SA SD011174306SA SD011174308SA SD011184340SA SD011184380SA SD011184443SA SD011184446SA SD011184453SA SD011194566SA SD011194573SA SD011194614SA SD011194619SA SD011204701SA SD011204735SA SD011204738SA SD011204739SA SD011204743SA SD011204746SA SD011204751SA SD011204752SA SD011204753SA SD011204754SA SD011214798SA SD011214804SA SD011214854SA SD011214861SA SD011214868SA SD011224972SA SD011224973SA SD011224975SA SD011224981SA SD011224988SA SD011224989SA SD011224990SA SD011224993SA SD011224994SA SD011224995SA SD011225022SA SD011225027SA SD011235085SA SD011235108SA SHQ11230033SA SHQ11230043SA SHQ11230045SA SHQ11230064SA SHQ11230075SA SHQ11230077SA SHQ11230088SA SHQ11240141SA SHQ11240147SA SHQ11240151SA SHQ11240164SA SHQ11240191SA SHQ11240207SA SHQ11240224SA SHQ11250277SA SHQ11250312SA SHQ11250325SA SHQ11250330SA SHQ11250335SA SHQ11250337SA SHQ11250455SA SHQ11250488SA SHQ11250495SA SHQ11260512SA SHQ11280580SA SHQ11280588SA SHQ11280591SA SHQ11280592SA SHQ11280599SA SHQ11280600SA SHQ11280602SA SHQ11280609SA SHQ11280615SA SHQ11280668SA SHQ11280708SA SHQ11280722SA SHQ11280743SA SHQ11280767SA SHQ11280772SA SHQ11280776SA SHQ11280779SA SHQ11280780SA SHQ11280781SA SHQ11280783SA SHQ11280784SA SHQ11280786SA SHQ11280791SA SHQ11280792SA SHQ11280797SA SHQ11280798SA SHQ11280809SA SHQ11290846SA SHQ11290848SA SHQ11290850SA SHQ11290851SA SHQ11290859SA SHQ11290860SA SHQ11290862SA SHQ11290863SA SHQ11290865SA SHQ11290891SA SHQ11290946SA SHQ11290948SA SHQ11290958SA SHQ11290962SA SHQ11290965SA SHQ11290966SA SHQ11290967SA SHQ11290972SA SHQ11290985SA SHQ11290989SA SHQ11290996SA SHQ11290998SA SHQ11291001SA SHQ11291003SA SHQ11291056SA SHQ11291057SA SHQ11291060SA SHQ11291061SA SHQ11291069SA SHQ11291076SA SHQ11291086SA SHQ11301180SA SHQ11301186SA SHQ11301188SA SHQ11301199SA SHQ11311385SA SHQ11311425SA SHQ11311428SA SHQ11321671SA SHQ11321699SA SHQ11321717SA SHQ11321718SA SHQ11321723SA SHQ11331917SA SHQ11332021SA SHQ11332030SA SHQ11332035SA SHQ11332037SA SHQ11342185SA SHQ11362608SA SHQ11362610SA SHQ11362611SA SHQ11362612SA SHQ11362613SA SHQ11362615SA SHQ11362616SA SHQ11362618SA SHQ11362619SA SHQ11362620SA SHQ11372735SA SHQ11372829SA SHQ11382999SA SHQ11423326SA SHQ11433431SA SHQ11433447SA SHQ11433525SA SHQ11443541SA SHQ11443543SA SHQ11443708SA SHQ11453785SA SHQ11453786SA SHQ11453793SA SHQ11453855SA SHQ11453950SA SHQ11453958SA SHQ11453959SA SHQ11453960SA SHQ11453961SA SHQ11453962SA SHQ11453963SA SHQ11453990SA SHQ11453996SA SHQ11454047SA SHQ11454059SA SHQ11454066SA SHQ11454067SA SHQ11454077SA SHQ11464126SA SHQ11464129SA SHQ11464134SA SHQ11464260SA SHQ11464289SA SHQ11464290SA SHQ11464293SA SHQ11464310SA SHQ11464375SA SHQ11474421SA SHQ11474422SA SHQ11474463SA SHQ11474464SA SHQ11474465SA SHQ11474466SA SHQ11474467SA SHQ11474569SA SHQ11474572SA SHQ11474583SA SHQ11474584SA SHQ11474585SA SHQ11474586SA SHQ11474587SA SHQ11474589SA SHQ11474592SA SHQ11484689SA SHQ11484708SA SHQ11484742SA SHQ11484768SA SHQ11484774SA SHQ11484776SA SHQ11484777SA SHQ11484790SA SHQ11484791SA SHQ11484820SA SHQ11484824SA SHQ11484831SA SHQ11484865SA SHQ11484875SA SHQ11484878SA SHQ11484878SABOP SHQ11484967SA SHQ11484971SA SHQ11484972SA SHQ11484974SA SHQ11484979SA SHQ11495072SA SHQ11495075SA SHQ11505295SA SHQ11505319SA SHQ11505320SA SHQ11512333GA SHQ11512334GA SHQ11512335GA SHQ11512336GA SHQ11512337GA SHQ11512338GA SHQ11512339GA SHQ12015442SA SHQ12015451SA SHQ12015472SA SHQ12025516SA SHQ12025518SA SHQ12025534SA SHQ12025535SA SHQ12035577SA SHQ12035579SA SHQ12035587SA SHQ12035609SA SHQ12035610SA SHQ12035626SA SHQ12035642SA SHQ12035649SA SHQ12035669SA SHQ12035673SA SHQ12035674SA SHQ12035678SA SHQ12045738SA SHQ12045740SABOP SHQ12045746SA SHQ12045771SA SHQ12045772SA SHQ12045773SA SHQ12045832SA SHQ12045833SA SHQ12045834SA SHQ12045835SA SHQ12055959SA SHQ12055962SA SHQ12055967SA SHQ12055969SA SHQ12055971SA SHQ12055971SABOP SHQ12055976SA SHQ12056018SABOP SHQ12066037SA SHQ12066110SA SHQ12066112SA SHQ12066116SA SHQ12066121SA SHQ12096377SA SHQ12096380SA SHQ12096382SA SHQ12096385SA SHQ12096388SA SHQ12096392SA SHQ12096393SA SHQ12096397SA SHQ12096398SA SHQ12096411SA SHQ12096440SA SHQ12106502SA SHQ12116626SA SHQ12116629SA SHQ12133000GA SHQ12136771SA SHQ12136772SA SHQ12136775SA SHQ12136778SA SHQ12136780SA SHQ12136781SA SHQ12136782SA SHQ12136786SA SHQ12136789SA SHQ12136790SA SHQ12136791SA SHQ12136792SA SHQ12136793SA SHQ12136809SA SHQ12143022GA SHQ12143023GA SHQ12143026GA SHQ12146842SA SHQ12146843SA SHQ12146844SA SHQ12146845SA SHQ12146846SA SHQ12146847SA SHQ12146848SA SHQ12146849SA SHQ12146850SA SHQ12146851SA SHQ12146852SA SHQ12146853SA SHQ12146854SA SHQ12146855SA SHQ12146856SA SHQ12153058GA SHQ12153059GA SHQ12153060GA SHQ12153061GA SHQ12153062GA SHQ12153065GA SHQ12156985SA SHQ12167074SA SHQ12167161SA SHQ12177238SA SHQ12177244SA SHQ12177248SA SHQ12177253SA SHQ12177258SA SHQ12177264SA SHQ12177273SA SHQ12177279SA SHQ12177341SA SHQ12187489SA SHQ12193361GA SHQ12197559SA SHQ12197625SA SHQ12197626SA SHQ12203392GA SHQ12207761SA SHQ12207763SA SHQ12217918SA SHQ12217921SA SHQ12217928SA SHQ12217941SA SHQ12217944SA SHQ12217957SA SHQ12217958SA SHQ12217962SA SHQ12217963SA SHQ12217964SA SHQ12217966SA SHQ12217967SA SHQ12217967SABOP SHQ12217988SA SHQ12217989SA SHQ12217990SA SHQ12217991SA SHQ12217993SA SHQ12217999SA SHQ12218001SA SHQ12218009SA SHQ12218010SA SHQ12218011SA SHQ12218014SA SHQ12218027SA SHQ12218029SA SHQ12218040SA SHQ12228145SA SHQ12228146SA SHQ12228147SA SHQ12228150SA SHQ12228151SA SHQ12228159SA SHQ12228164SA SHQ12228165SA SHQ12228185SA SHQ12228187SA SHQ12228194SA SHQ12228221SA SHQ12228223SA SHQ12228224SA SHQ12228225SA SHQ12228226SA SHQ12228227SA SHQ12228229SA SHQ12228231SA SHQ12228232SA SHQ12238380SA SHQ12238381SA SHQ12238469SA SHQ12248661SA SHQ12248678SA SHQ12248682SA SHQ12248685SA SHQ12248687SA SHQ12248695SA SHQ12248699SA SHQ12248702SA SHQ12248703SA SHQ12248705SA SHQ12253677GA SHQ12253679GA SHQ12253683GA SHQ12258710SA SHQ12258711SA SHQ12258712SA SHQ12258713SA SHQ12258714SA SHQ12258715SA SHQ12258716SA SHQ12258717SA SHQ12258718SA SHQ12258719SA SHQ12258720SA SHQ12258721SA SHQ12258722SA SHQ12258723SA SHQ12258724SA SHQ12258725SA SHQ12258726SA SHQ12258727SA SHQ12258728SA SHQ12258729SA SHQ12258829SA SHQ12258831SA SHQ12268857SA SHQ12268858SA SHQ12268910SA SHQ12268913SA SHQ12278985SA SHQ12279011SA SHQ12289064SA SHQ12289096SA SHQ12289112SA SHQ12289114SA SHQ12289115SA SHQ12289118SA SHQ12289119SA SHQ12289158SA SHQ12289159SA SHQ12289160SA SHQ12289163SA SHQ12289164SA SHQ12289165SA SHQ12289166SA SHQ12289167SA SHQ12289168SA SHQ12289170SA SHQ12289171SA SHQ12289188SA SHQ12289198SA SHQ12289205SA SHQ12289210SA SHQ12289211SA SHQ12289214SA SHQ12289218SA SHQ12289222SA SHQ12289226SA SHQ12293787GA SHQ12293788GA SHQ12293789GA SHQ12293790GA SHQ12293791GA SHQ12293792GA SHQ12293793GA SHQ12293794GA SHQ12293795GA SHQ12293796GA SHQ12293797GA SHQ12293798GA SHQ12293799GA SHQ12293800GA SHQ12293801GA SHQ12293802GA SHQ12293803GA SHQ12293804GA SHQ12293805GA SHQ12293806GA SHQ12299272SA SHQ12299275SA SHQ12299283SA SHQ12299289SA SHQ12299294SA SHQ12299328SA SHQ12299329SA SHQ12299330SA SHQ12299331SA SHQ12299332SA SHQ12299334SA SHQ12299335SA SHQ12299336SA SHQ12299337SA SHQ12299338SA SHQ12299339SA SHQ12299340SA SHQ12299341SA SHQ12299342SA SHQ12299343SA SHQ12299344SA SHQ12299345SA SHQ12299346SA SHQ12299409SA SHQ12299422SA SHQ12299438SA SHQ12299439SA SHQ12299440SA SHQ12299441SA SHQ12299442SA SHQ12299443SA SHQ12299444SA SHQ12299445SA SHQ12299447SA SHQ12299450SA SHQ12299461SA SHQ12309477SA SHQ12278993SA SHQ12278989SA SHQ12279001SA SHQ12278991SA SHQ12309575SA SHQ12309581SA SHQ12289132SA SHQ12320350SA SHQ12289124SA SHQ12320351SA SHQ12330491SA SHQ12320357SA SHQ12330545SA SHQ12330546SA SHQ12330549SA SHQ12330554SA SHQ12361200SA SHQ12313859GA SHQ12313863GA SHQ12313860GA SHQ12313864GA SHQ12371310SA SHQ12371489SA SHQ12333971GA SHQ12333974GA SD010480111SA SD011051208SA SD011061259SA SD011061416SA SD011061423SA SD011061440SA SD011061488SA SD011071990SA SD011071995SA SD011071999SA SD011072006SA SD011072024SA SD011072036SA SD011072079SA SD011072093SA SD011133715SA SD011133718SA SD011133719SA SD011133724SA SD011133725SA SD011133727SA SD011133730SA SD011133731SA SD011133732SA SD011133733SA SD011174248SA SD011174250SA SD011174252SA SD011174255SA SD011174256SA SD011174258SA SD011174260SA SD011174262SA SD011174264SA SD011174268SA SD011174277SA SD011174282SA SD011174283SA SD011174284SA SD011174292SA SD011174293SA SD011174296SA SD011174297SA SHQ12167096SA SHQ12177284SA SHQ12177301SA SHQ12177310SA SHQ12177372SA SHQ12177374SA SHQ12177379SA SHQ12177380SA SHQ12177383SA SHQ12177384SA SHQ12177385SA SHQ12177386SA SHQ12177396SA SHQ12177399SA SD010470368GA SD010480983GA SD010502200GA SHQ11433460SA SHQ11433463SA SHQ11484966SA SHQ11484970SA SHQ11484977SA SHQ12228104SA SD010345404GA SD010345406GA SD010345410GA SD010345411GA SD010345427GA SD010450018GA SD010450024GA SD010450034GA SD010459845GA SD010459851GA SD010459869GA SD010459873GA SD010460147GA SD010460148GA SD010460157GA SD010460235GA SD010460242GA SD010460247GA SHQ11441413GA SHQ11451433GA SD010500228SA SD010500229SA SD010500233SA SD010500247SA SD011061422SA SD011061438SA SD011061449SA SD011061479SA SD011071592SA SD011123270SA SD011123276SA SD011174317SA SD011174320SA SD011174323SA SD011174324SA SD011174329SA SD011174331SA SHQ11290951SA SHQ11290978SA SD010481038GA SD010387413GA SD010460228GA SD010470336GA SD011102513SA SD011102530SA SD011102532SA SD011112554SA SD011112555SA SD011112556SA SD011112558SA SD011112559SA SD011112560SA SD011112561SA SD011112562SA SD011112568SA SD011112605SA SD011122936SA SD011133438SA SD011133644SA SD011143855SA SD011143861SA SD011143864SA SD011143870SA SD011143872SA SD011143875SA SD011184351SA SD011184357SA SD011184358SA SD011184378SA SD011194596SA SD011194605SA SD011225023SA SD011225033SA SD011225046SA SD011225068SA SHQ11474595SA SHQ11484653SA SHQ11484661SA SHQ11484711SA SHQ11484782SA SHQ11484784SA SHQ11484787SA SHQ12218055SA SD010366448GA SD010366505GA SD010387423GA SD010397983GA SD010397985GA SD010408405GA SD010418556GA SD010418558GA SD010428941GA SD010428942GA SD010428944GA SD010428957GA SD010428962GA SD010460161GA SD010470330GA SD010470551GA SD010470552GA SD010470554GA SD010470586GA SD010470658GA SD010470765GA SD010470771GA SD010470777GA SD010500218SA SD010501782GA SD010501806GA SD010501811GA SD010501814GA SD010501922GA SD010501923GA SD010501930GA SD010501933GA SD010501936GA SD010502179GA SD010502188GA SD010502260GA SD010510335SA SD010510346SA SD010522648GA SD010522662GA SD010522663GA SD010522664GA SD010522665GA SD010522666GA SD010522667GA SD010522668GA SD010522669GA SD010522670GA SD010522671GA SD010522672GA SD011030677SA SD011030681SA SD011032933GA SD011033005GA SD011033019GA SD011033023GA SD011033068GA SD011033096GA SD011033153GA SD011033157GA SD011033160GA SD011112621SA SD011122854SA SD011122923SA SD011122934SA SD011122945SA SD011122950SA SD011122955SA SD011122956SA SD011123004SA SD011123046SA SD011123048SA SD011123050SA SD011123052SA SD011133404SA SD011133474SA SD011133478SA SD011133594SA SD011133667SA SD011143863SA SD011184350SA SD011184353SA SD011184354SA SD011184356SA SD011184359SA SD011184360SA SD011184377SA SD011184385SA SD011194607SA SD011225001SA SD011225008SA SD011225053SA SD011225057SA SHQ11280695SA SHQ11280699SA SHQ11280706SA SHQ11280735SA SHQ11280741SA SHQ11280744SA SHQ11280745SA SHQ11280746SA SHQ11280752SA SHQ11280753SA SHQ11280759SA SHQ11301139SA SHQ11301164SA SHQ11301184SA SHQ11301187SA SHQ11311534SA SHQ11331996SA SHQ11342213SA SHQ11342215SA SHQ11342221SA SHQ11342225SA SHQ11342254SA SHQ11423376SA SHQ11433422SA SHQ11433427SA SHQ11443619SA SHQ11464292SA SHQ11464410SA SHQ11474600SA SHQ11474613SA SHQ11484705SA SHQ11484712SA SHQ11484778SA SHQ11484959SA SHQ11484961SA SHQ11495168SA SHQ12066098SA SHQ12066106SA SHQ12066113SA SHQ12066115SA SHQ12066117SA SHQ12066119SA SHQ12066120SA SHQ12066123SA SHQ12066125SA SHQ12066126SA SHQ12076237SA SHQ12076238SA SHQ12076240SA SHQ12076241SA SHQ12076244SA SHQ12076247SA SHQ12076250SA SHQ12076256SA SHQ12076258SA SHQ12167067SA SHQ12177283SA SHQ12177286SA SHQ12177287SA SHQ12177289SA SHQ12177290SA SHQ12177291SA SHQ12177292SA SHQ12177293SA SHQ12177295SA SHQ12177298SA SHQ12177299SA SHQ12177312SA SHQ12197575SA SHQ12197576SA SHQ12197578SA SHQ12197589SA SHQ12197593SA SHQ12207787SA SHQ12207788SA SHQ12207789SA SHQ12207792SA SHQ12207798SA SHQ12207800SA SHQ12207802SA SHQ12207804SA SHQ12217891SA SHQ12217901SA SHQ12217902SA SHQ12217911SA SHQ12217912SA SHQ12218050SA SHQ12218053SA SHQ12218054SA SHQ12218056SA SHQ12218057SA SHQ12228115SA SHQ12309579SA SHQ12320346SA SHQ12330593SA SHQ12330589SA SHQ12330734SA SHQ12330774SA SHQ12330737SA SHQ12310095SA SHQ12310097SA SHQ12309567SA SHQ12319974SA SHQ12310104SA SHQ12309568SA SHQ12310092SA SHQ12319985SA SHQ12310103SA SHQ12310090SA SHQ12330778SA SHQ12340969SA SHQ12371352SA SHQ12351101SA SHQ12371303SA SHQ12371356SA SHQ12371357SA SHQ12371353SA SHQ12371349SA SHQ12371358SA SHQ12371354SA SHQ12371347SA SHQ12371355SA SHQ12320443SA SHQ12351018SA SHQ12351094SA SHQ12320445SA SHQ12320419SA SHQ12351107SA SHQ12320436SA SHQ12279031SA SHQ11190030SA SD011071632SA SHQ11464388SA SHQ11464389SA SHQ11464390SA SHQ11464391SA SHQ11464392SA SHQ11464393SA SHQ11464394SA SHQ11464395SA SHQ11464396SA SHQ11464397SA SHQ11484655SA SHQ11484657SA SHQ11484658SA SHQ11484662SA SHQ11484664SA SHQ11484665SA SHQ11484666SA SHQ11484667SA SHQ11484669SA SHQ11484670SA SHQ11484671SA SHQ11484672SA SHQ11484673SA SHQ11484674SA SHQ11484675SA SHQ11484676SA SHQ11484677SA SHQ11484681SA SHQ12025537SA SHQ12045730SA SHQ12045734SA SHQ12299466SA SHQ12330841SA SHQ12361237SA SHQ12309592SA SHQ12309582SA SHQ12309585SA SHQ12309590SA SHQ12320256SA SHQ12320239SA SHQ12320243SA SHQ12320238SA SHQ12320237SA SHQ12320240SA SHQ12320235SA SHQ12320225SA SHQ12320230SA SHQ12320226SA SHQ12320251SA SHQ12320241SA SHQ12320327SA SHQ12320331SA SHQ12320326SA SHQ12320322SA SHQ12320325SA SHQ12320324SA SHQ12320330SA SHQ12320332SA SHQ12320319SA SHQ12320323SA SHQ12320329SA SHQ12320333SA SHQ12320253SA SHQ12320250SA SHQ12320252SA SHQ12320233SA SHQ12320227SA SHQ12320224SA SHQ12310094SA SHQ12320231SA SHQ12320228SA SHQ12320245SA SHQ12320255SA SHQ12320229SA SHQ12217904SA SHQ12330844SA SHQ12330830SA SHQ12330836SA SHQ12330837SA SHQ12330838SA SHQ12330839SA SHQ12299463SA SHQ12330831SA SHQ12330842SA SHQ12299465SA SHQ12330843SA SHQ12299464SA SHQ12299462SA SHQ12330846SA SHQ12299436SA SHQ12299437SA SHQ12330853SA SHQ12330856SA SHQ12330854SA SHQ12330855SA SHQ12330857SA SHQ12330858SA SHQ12330848SA SHQ12330849SA SHQ12330528SA SHQ12361270SA SHQ12361266SA SHQ12351044SA SHQ12351047SA SHQ12351040SA SHQ12351036SA SHQ12351034SA SHQ12351032SA SHQ12351046SA SHQ12351039SA SHQ12351043SA SHQ12351050SA SHQ12351035SA SHQ12351048SA SHQ12351052SA SHQ12351069SA SHQ12351067SA SHQ12351061SA SHQ12351045SA SHQ12340064SA SHQ12351060SA SHQ12361246SA SHQ12351057SA SHQ12351058SA SHQ12351068SA SHQ12351070SA SHQ12351071SA SHQ12361229SA SHQ12361264SA SHQ12361275SA SHQ12361272SA SHQ12361273SA SHQ12361269SA SHQ12361259SA SHQ12361265SA SHQ12361268SA SHQ12361274SA SHQ12361271SA SHQ12361267SA SHQ12361255SA SHQ12361261SA SHQ12361247SA SHQ12361248SA SHQ12361257SA SHQ12361231SA SHQ12361235SA SHQ12361224SA SHQ12361222SA SHQ12361226SA SHQ12361225SA SHQ12361207SA SHQ12361230SA SHQ12361227SA SHQ12361221SA SHQ12361212SA SHQ12361208SA SHQ12361147SA SHQ12361152SA SHQ12361140SA SHQ12361142SA SHQ12361144SA SHQ12361148SA SHQ12361150SA SHQ12361149SA SHQ12361134SA SHQ12361146SA SHQ12361145SA SHQ12361138SA SHQ12361217SA SHQ12361210SA SHQ12361228SA SHQ12361215SA SHQ12361220SA SHQ12361209SA SHQ12361218SA SHQ12361219SA SHQ12361214SA SHQ12361216SA SHQ12361211SA SHQ12361213SA SHQ12340972SA SHQ12340978SA SHQ12351132SA SHQ12341003SA SHQ12340996SA SHQ12351059SA SHQ12340991SA SHQ12351065SA SHQ12351074SA SHQ12351072SA SHQ12351073SA SHQ12351023SA SHQ12351075SA SHQ12340984SA SHQ12340965SA SHQ12340977SA SHQ12340970SA SHQ12340974SA SHQ12340973SA SHQ12340982SA SHQ12340990SA SHQ12351131SA SHQ12340987SA SHQ12340951SA SHQ12340976SA SHQ12340979SA SHQ12361233SA SHQ12361252SA SHQ12361244SA SHQ12361223SA SHQ12361242SA SHQ12361239SA SHQ12361232SA SHQ12361245SA SHQ12361234SA SHQ12361241SA SHQ12361236SA SD010408164GA SD010408170GA SD010408176GA SD010408180GA SD010430035SA SD010430038SA SD010449396GA SD010449422GA SD010449435GA SD010449475GA SD010449477GA SD010449479GA SD010449482GA SD010449491GA SD010449493GA SD010449495GA SD010450010GA SD010480104SA SD010480105SA SD010480107SA SD010480108SA SD010480109SA SD010480115SA SD010491545GA SD010500223SA SD010500226SA SD010500231SA SD010510331SA SD010510332SA SD010510334SA SD010510338SA SD010510339SA SD010510341SA SD010510343SA SD010510345SA SD011032908GA SD011032909GA SD011032910GA SD011032912GA SD011032914GA SD011032916GA SD011032917GA SD011032918GA SD011032919GA SD011032922GA SD011032968GA SD011032970GA SD011032972GA SD011032973GA SD011032989GA SD011032990GA SD011032991GA SD011032993GA SD011032995GA SD011032996GA SD011032998GA SD011032999GA SD011033000GA SD011033001GA SD011033003GA SD011033043GA SD011033044GA SD011033046GA SD011033048GA SD011033049GA SD011033050GA SD011033051GA SD011033053GA SD011033055GA SD011033057GA SD011033058GA SD011033061GA SD011061345SA SD011061352SA SD011061361SA SD011071636SA SD011071637SA SD011071646SA SD011071649SA SD011072119SA SD011072142SA SD011082163SA SD011133714SA SD011143761SA SD011143766SA SD011174189SA SD011174193SA SD011174203SA SD011184452SA SD011184458SA SD011184463SA SHQ12167217SA SHQ12167218SA SHQ12167219SA SHQ12167220SA SHQ12167221SA SD010459955GA SD010459957GA SD011072103SA SD011072105SA SD011082314SA SD011082315SA SD011082316SA SD011133545SA SD011133620SA SD010138420GA SD010272389GA SD010272393GA SD010272407GA SD010272444GA SD010303406GA SD010313705GA SD010335101GA SD010345402GA SD010345431GA SD010345434GA SD010345437GA SD010355793GA SD010355825GA SD010355828GA SD010355829GA SD010355830GA SD010355831GA SD010355833GA SD010365863GA SD010365883GA SD010365890GA SD010365892GA SD010365897GA SD010365899GA SD010365900GA SD010365925GA SD010365959GA SD010366039GA SD010366041GA SD010366049GA SD010366054GA SD010366192GA SD010366193GA SD010366194GA SD010366195GA SD010366196GA SD010366197GA SD010366198GA SD010366199GA SD010366200GA SD010366201GA SD010366202GA SD010366203GA SD010366205GA SD010366206GA SD010366266GA SD010376643GA SD010376645GA SD010376646GA SD010376649GA SD010376650GA SD010376652GA SD010376653GA SD010376654GA SD010376655GA SD010376657GA SD010376658GA SD010376662GA SD010376793GA SD010376799GA SD010376805GA SD010376807GA SD010377093GA SD010387319GA SD010387456GA SD010387461GA SD010387614GA SD010387657GA SD010387675GA SD010387681GA SD010387734GA SD010387737GA SD010387739GA SD010387743GA SD010387750GA SD010387754GA SD010387761GA SD010397989GA SD010397993GA SD010397994GA SD010397999GA SD010398000GA SD010398005GA SD010408090GA SD010408092GA SD010408132GA SD010408133GA SD010408134GA SD010408135GA SD010408136GA SD010408140GA SD010408145GA SD010408146GA SD010408148GA SD010408166GA SD010408181GA SD010408189GA SD010408260GA SD010408300GA SD010408311GA SD010408319GA SD010408320GA SD010408321GA SD010408326GA SD010408329GA SD010408330GA SD010408339GA SD010408342GA SD010408368GA SD010408381GA SD010418521GA SD010418540GA SD010418716GA SD010428933GA SD010428945GA SD010428948GA SD010428960GA SD010428963GA SD010428965GA SD010428966GA SD010428967GA SD010428968GA SD010428969GA SD010428970GA SD010429075GA SD010429113GA SD010429240GA SD010449330GA SD010449478GA SD010449486GA SD010449490GA SD010449543GA SD010449545GA SD010449547GA SD010449548GA SD010449549GA SD010449553GA SD010449554GA SD010449555GA SD010449556GA SD010449560GA SD010449561GA SD010449562GA SD010449563GA SD010449564GA SD010449565GA SD010449569GA SD010449570GA SD010449572GA SD010449573GA SD010449575GA SD010449576GA SD010449582GA SD010449584GA SD010449585GA SD010449586GA SD010449588GA SD010449589GA SD010449590GA SD010449591GA SD010449592GA SD010449594GA SD010449595GA SD010449596GA SD010449597GA SD010449598GA SD010449599GA SD010449600GA SD010449601GA SD010449602GA SD010449603GA SD010449604GA SD010449605GA SD010449606GA SD010449607GA SD010449608GA SD010449609GA SD010449610GA SD010449611GA SD010449612GA SD010449613GA SD010449614GA SD010449615GA SD010449632GA SD010449634GA SD010449636GA SD010449637GA SD010449638GA SD010449639GA SD010449640GA SD010449641GA SD010449642GA SD010449643GA SD010449645GA SD010449648GA SD010449649GA SD010449650GA SD010449651GA SD010449654GA SD010449655GA SD010449656GA SD010449660GA SD010449662GA SD010449664GA SD010449670GA SD010449673GA SD010449676GA SD010449678GA SD010449679GA SD010450086GA SD010460152GA SD010460167GA SD010480101SA SD010480102SA SD010480891GA SD010481036GA SD010481040GA SD010481164GA SD010481169GA SD010481180GA SD010481197GA SD010491396GA SD010491404GA SD010491434GA SD010491436GA SD010491440GA SD010491441GA SD010491560GA SD010491570GA SD010500216SA SD010500243SA SD010501764GA SD010502147GA SD010502161GA SD010502162GA SD010502163GA SD010502166GA SD010502167GA SD010502168GA SD010502172GA SD010502173GA SD010502250GA SD010502253GA SD010502315GA SD010522499GA SD010522513GA SD010522522GA SD010522523GA SD010522524GA SD010522525GA SD010522528GA SD010522529GA SD010522536GA SD010522608GA SD011040881SA SD011040902SA SD011051030SA SD011051165SA SD011051209SA SD011061277SA SD011061283SA SD011061338SA SD011061382SA SD011071583SA SD011071615SA SD011071623SA SD011071624SA SD011071880SA SD011073550GA SD011083581GA SD011083588GA SD011083589GA SD011083591GA SD011083592GA SD011083594GA SD011083595GA SD011083596GA SD011083598GA SD011083600GA SD011083604GA SD011093743GA SD011093757GA SD011103789GA SD011103790GA SD011103804GA SD011123267SA SD011123292SA SD011133710SA SD011133711SA SD011133712SA SD011143918SA SD011143923SA SD011143947SA SD011143948SA SD011143955SA SD011143960SA SD011143961SA SD011143962SA SD011143963SA SD011143965SA SD011143966SA SD011143968SA SD011143969SA SD011143971SA SD011143973SA SD011143982SA SD011143982SABOP SHQ11280639SA SHQ11280698SA SHQ11290841SA SHQ11290842SA SHQ11301137SA SHQ11301172SA SHQ11311415SA SHQ11311427SA SHQ11311429SA SHQ11311462SA SHQ11321661SA SHQ11331915SA SHQ11331916SA SHQ11331918SA SHQ11331920SA SHQ11331921SA SHQ11331922SA SHQ11331923SA SHQ11331924SA SHQ11331926SA SHQ11331929SA SHQ11331935SA SHQ11331936SA SHQ11331939SA SHQ11332106SA SHQ11342152SA SHQ11342170SA SHQ11383082SA SHQ11423328SA SHQ11453892SA SHQ11453947SA SHQ11453951SA SHQ11453952SA SHQ11453953SA SHQ11453999SA SHQ11454000SA SHQ11454001SA SHQ11454002SA SHQ11454003SA SHQ11454004SA SHQ11454005SA SHQ11454006SA SHQ11454007SA SHQ11454008SA SHQ11454009SA SHQ11454012SA SHQ11464187SA SHQ11474598SA SHQ11474634SA SHQ11474635SA SHQ11474638SA SHQ11474640SA SHQ11474644SA SHQ11474645SA SHQ11484718SA SHQ11484770SA SHQ11484781SA SHQ11484783SA SHQ11484792SA SHQ11484805SA SHQ11484811SA SHQ11484814SA SHQ11484858SA SHQ11484859SA SHQ11484895SA SHQ11505305SA SHQ11505306SA SHQ11505310SA SHQ11505312SA SHQ11505314SA SHQ12035691SA SHQ12045724SA SHQ12045726SA SHQ12045751SA SHQ12045752SA SHQ12045753SA SHQ12045756SA SHQ12045760SA SHQ12045763SA SHQ12045766SA SHQ12045768SA SHQ12055956SA SHQ12066029SA SHQ12082739GA SHQ12082740GA SHQ12082741GA SHQ12082742GA SHQ12082743GA SHQ12082744GA SHQ12082745GA SHQ12082746GA SHQ12106525SA SHQ12218061SA SHQ12228258SA SHQ12228268SA SHQ12228272SA SHQ12289240SA SHQ12289243SA SHQ12289244SA SHQ12289245SA SHQ12289247SA SHQ12289248SA SHQ12289250SA SHQ12289251SA SHQ12330820SA SHQ12330817SA SHQ12330816SA SHQ12330818SA SHQ12330815SA SHQ12330819SA SHQ12371317SA SHQ12371319SA SHQ12371322SA SHQ12371308SA SHQ12371305SA SHQ12371444SA SHQ12258828SA SD010303625GA SD010334806GA SD010377175GA SD010377177GA SD010377179GA SD010387331GA SD010387348GA SD010387358GA SD010387384GA SD010387387GA SD010387405GA SD010387478GA SD010387481GA SD010387482GA SD010387483GA SD010387484GA SD010387485GA SD010387486GA SD010387487GA SD010387488GA SD010387489GA SD010387490GA SD010387494GA SD010387770GA SD010387780GA SD010387781GA SD010387784GA SD010397952GA SD010397955GA SD010397964GA SD010397978GA SD010397986GA SD010408226GA SD010418513GA SD010418516GA SD010418517GA SD010418518GA SD010418613GA SD010418769GA SD010418780GA SD010429001GA SD010429006GA SD010429008GA SD010429009GA SD010429013GA SD010459748GA SD010459749GA SD010459750GA SD010459761GA SD010459933GA SD010459946GA SD010459947GA SD010460171GA SD010460176GA SD010460178GA SD010460204GA SD010470349GA SD010470377GA SD010470624GA SD010470638GA SD010470751GA SD010470753GA SD010480803GA SD010480822GA SD010481022GA SD010481024GA SD010481030GA SD010491324GA SD010491402GA SD010491576GA SD010501826GA SD010501991GA SD010502016GA SD010502096GA SD010502114GA SD010502117GA SD010502125GA SD010502137GA SD010502139GA SD010502142GA SD010502148GA SD010502150GA SD010502151GA SD010502295GA SD010512414GA SD010522474GA SD010522485GA SD010522489GA SD010522496GA SD010522506GA SD010522519GA SD010522521GA SD010522535GA SD010522538GA SD010522539GA SD010522654GA SD010532796GA SD010532797GA SD010532798GA SD010532801GA SD010532803GA SD010532808GA SD010532809GA SD010532810GA SD010532817GA SD010532820GA SD010532827GA SD010532832GA SD011032938GA SD011032941GA SD011032969GA SD011032992GA SD011033045GA SD011033047GA SD011033054GA SD011033056GA SD011033059GA SD011033062GA SD011033063GA SD011033064GA SD011033065GA SD011033066GA SD011053259GA SD011053260GA SD011053261GA SD011053262GA SD011063244GA SD011063247GA SD011063249GA SD011063250GA SD011063252GA SD011063253GA SD011063254GA SD011063255GA SD011063256GA SD011063299GA SD011063300GA SD011063301GA SD011063302GA SD011063303GA SD011063304GA SD011063305GA SD011063306GA SD011063308GA SD011063309GA SD011063310GA SD011063311GA SD011063312GA SD011063313GA SD011103794GA SD011113849GA SD011113858GA SD011113865GA SD011113868GA SD011113876GA SD011113877GA SD011113925GA SD011113939GA SD011184348SA SD011224923SA SD011224950SA SHQ11250410SA SHQ11250444SA SHQ11250461SA SHQ11250462SA SHQ11250463SA SHQ11250466SA SHQ11250468SA SHQ11250469SA SHQ11260532SA SHQ11260533SA SHQ11260547SA SHQ11280567SA SHQ11280570SA SHQ11280583SA SHQ11280618SA SHQ11280619SA SHQ11280621SA SHQ11280622SA SHQ11280627SA SHQ11280633SA SHQ11280636SA SHQ11280646SA SHQ11280660SA SHQ11280667SA SHQ11280673SA SHQ11280675SA SHQ11280684SA SHQ11280697SA SHQ11280794SA SHQ11290864SA SHQ11291028SA SHQ11291044SA SHQ11301152SA SHQ11301171SA SHQ11301176SA SHQ11301239SA SHQ11301240SA SHQ11301327SA SHQ11301329SA SHQ11311403SA SHQ11321655SA SHQ11321656SA SHQ11321659SA SHQ11321660SA SHQ11321663SA SHQ11321666SA SHQ11321672SA SHQ11321704SA SHQ11321706SA SHQ11321707SA SHQ11321709SA SHQ11321710SA
SHQ11331919SA SHQ11331925SA SHQ11331943SA SHQ11331981SA SHQ11352377SA SHQ11352388SA SHQ11362621SA SHQ11372748SA SHQ11382982SA SHQ11382991SA SHQ11382997SA SHQ11382998SA SHQ11383000SA SHQ11383001SA SHQ11383003SA SHQ11383006SA SHQ11383010SA SHQ11383011SA SHQ11383015SA SHQ11383017SA SHQ11383025SA SHQ11383029SA SHQ11383036SA SHQ11383038SA SHQ11423234SA SHQ11423356SA SHQ11433402SA SHQ11433518SA SHQ11443547SA SHQ11443553SA SHQ11443699SA SHQ11453784SA SHQ11453788SA SHQ11453797SA SHQ11453809SA SHQ11453858SA SHQ11453860SA SHQ11453894SA SHQ11453895SA SHQ11453897SA SHQ11453898SA SHQ11453900SA SHQ11453901SA SHQ11453903SA SHQ11453904SA SHQ11453969SA SHQ11453972SA SHQ11453977SA SHQ11464108SA SHQ11464193SA SHQ11464227SA SHQ11464232SA SHQ11464357SA SHQ11484733SA SHQ11484734SA SHQ11484736SA SHQ11484739SA SHQ11484746SA SHQ11484750SA SHQ11484754SA SHQ11505359SA SHQ12015383SA SHQ12015384SA SHQ12015389SA SHQ12015395SA SHQ12015399SA SHQ12015401SA SHQ12015405SA SHQ12015408SA SHQ12015434SA SHQ12015443SA SHQ12015447SA SHQ12015448SA SHQ12015449SA SHQ12015458SA SHQ12025485SA SHQ12025491SA SHQ12025513SA SHQ12025525SA SHQ12025529SA SHQ12025531SA SHQ12025532SA SHQ12035545SA SHQ12035554SA SHQ12035682SA SHQ12035683SA SHQ12035684SA SHQ12035686SA SHQ12035688SA SHQ12035689SA SHQ12035690SA SHQ12035694SA SHQ12035696SA SHQ12035697SA SHQ12035698SA SHQ12035699SA SHQ12035718SA SHQ12035720SA SHQ12045774SA SHQ12045778SA SHQ12045782SA SHQ12045783SA SHQ12045793SA SHQ12045794SA SHQ12045795SA SHQ12045844SA SHQ12055917SA SHQ12055920SA SHQ12055925SA SHQ12055997SA SHQ12096372SA SHQ12156919SA SHQ12157054SA SHQ12167072SA SHQ12167227SA SHQ12167230SA SHQ12187445SA SHQ12187461SA SHQ12197655SA SHQ12197657SA SHQ12217961SA SHQ12228107SA SHQ12228143SA SHQ12228152SA SHQ12228180SA SHQ12228183SA SHQ12228189SA SHQ12228190SA SHQ12228191SA SHQ12228192SA SHQ12228199SA SHQ12258832SA SHQ12268856SA SHQ12279025SA SHQ12279030SA SHQ12289060SA SHQ12289072SA SHQ12289078SA SHQ12299327SA SHQ12320143SA SHQ12217899SA SHQ12217873SA SHQ12320150SA SHQ12320126SA SHQ12289032SA SD010418853GA SD010491361GA SD010522615GA SD010522619GA SD010522623GA SD011032928GA SD011032930GA SD011032934GA SD011033008GA SD011033009GA SD011033015GA SD011033016GA SD011033025GA SD011033026GA SD011033028GA SD011033031GA SD011033033GA SD011033041GA SD011033072GA SD011033075GA SD011033076GA SD011123316SA SD011123343SA SD010532839GA SD011093695GA SD011093732GA SD011093746GA SD011093748GA SD011093753GA SD011093756GA SD011113892GA SD011113893GA SD011113894GA SD011113895GA SD011113896GA SD011113897GA SD011113898GA SD011113899GA SD011113900GA SD011113901GA SD011113902GA SD011113903GA SD011113904GA SD011113907GA SD011113910GA SD011113915GA SD011113968GA SD011113969GA SD011113970GA SD011113972GA SD011113973GA SD011113974GA SD011113975GA SD011113976GA SD011113979GA SD011113981GA SD011113983GA SD011113984GA SD011113987GA SD011113988GA SD011113989GA SD011113990GA SD011113991GA SD011124004GA SD011124006GA SD011124007GA SD011124008GA SD011124009GA SD011124010GA SD011124011GA SD011124012GA SD011124013GA SD011124014GA SD011124015GA SD011124016GA SD011124017GA SD011124018GA SD011124019GA SD011124020GA SD011124021GA SD011124022GA SD011124023GA SD011124024GA SD011124025GA SD011124026GA SD011124027GA SD011124028GA SD011124029GA SD011124105GA SD011124106GA SD011124108GA SD011124109GA SD011124110GA SD011124111GA SD011124112GA SD011124113GA SD011124114GA SD011124117GA SD011124123GA SD011124126GA SD011124127GA SD011124130GA SD011124133GA SD011124142GA SD011124143GA SD011124150GA SD011124151GA SD011124154GA SD011124155GA SD011124156GA SD011124158GA SD011124159GA SD011124161GA SD011124162GA SD011124168GA SD011144397GA SD011144402GA SD011144405GA SD011144410GA SD011144413GA SD011144415GA SD011144499GA SD011144501GA SD011144502GA SD011144504GA SD011144505GA SD011144506GA SD011144507GA SD011144508GA SHQ11280236GA SHQ11391177GA SHQ11502081GA SHQ11512237GA SD011072099SA SD011123282SA SD011133713SA SD011133717SA SHQ11240260SA SHQ11260535SA SHQ11403101SA SHQ11423176SA SD010127652GA SD010200054GA SD010366222GA SD010366236GA SD010366277GA SD010366336GA SD010387274GA SD010387286GA SD010387287GA SD010387288GA SD010397948GA SD010408200GA SD010408247GA SD010408253GA SD010408271GA SD010408299GA SD010408304GA SD010408309GA SD010408312GA SD010408314GA SD010408328GA SD010408334GA SD010408338GA SD010408397GA SD010408403GA SD010408408GA SD010408434GA SD010418741GA SD010418745GA SD010418749GA SD010418751GA SD010418756GA SD010418849GA SD010418851GA SD010418852GA SD010418857GA SD010418859GA SD010418867GA SD010418872GA SD010418874GA SD010418876GA SD010418877GA SD010428920GA SD010428935GA SD010366137GA SD010428943GA SD010366229GA SD010428946GA SD010428964GA SD010428971GA SD010366352GA SD010387227GA SD010387230GA SD010387232GA SD010387235GA SD010428972GA SD010428974GA SD010428975GA SD010428976GA SD010428977GA SD010428978GA SD010408183GA SD010428979GA SD010428980GA SD010408252GA SD010428981GA SD010428983GA SD010428984GA SD010428985GA SD010429054GA SD010429057GA SD010429058GA SD010429067GA SD010430037SA SD010449390GA SD010449393GA SD010449397GA SD010449416GA SD010449421GA SD010449428GA SD010418742GA SD010449429GA SD010418746GA SD010418748GA SD010449430GA SD010449498GA SD010470684GA SD010418848GA SD010470686GA SD010470695GA SD010470711GA SD010470720GA SD010470723GA SD010418862GA SD010418865GA SD010470794GA SD010418868GA SD010480974GA SD010481150GA SD010418875GA SD010481153GA SD010481154GA SD010418912GA SD010418915GA SD010481157GA SD010481165GA SD010481166GA SD010481167GA SD010481171GA SD010481173GA SD010481175GA SD010481177GA SD010481178GA SD010481182GA SD010481188GA SD010481189GA SD010481191GA SD010481193GA SD010481198GA SD010491379GA SD010491435GA SD010491537GA SD010428993GA SD010491538GA SD010491542GA SD010491657GA SD010491663GA SD010429112GA SD010491664GA SD010491668GA SD010501745GA SD010501752GA SD010501879GA SD010501880GA SD010501881GA SD010501882GA SD010501886GA SD010501888GA SD010501992GA SD010502001GA SD010502101GA SD010502285GA SD010502303GA SD010510342SA SD010512388GA SD010512391GA SD010512393GA SD010512396GA SD010512402GA SD010522518GA SD010522530GA SD010522531GA SD010522537GA SD010522540GA SD010522541GA SD010522673GA SD010522674GA SD010522675GA SD010522676GA SD010522677GA SD010522679GA SD010522680GA SD010522681GA SD010522684GA SD010532687GA SD010532812GA SD010532872GA SD010532875GA SD010532876GA SD010532885GA SD010532889GA SD010532890GA SD010532891GA SD010532893GA SD010532895GA SD010532896GA SD010532897GA SD010532900GA SD010532901GA SD010532902GA SD010532904GA SD010532905GA SD010532906GA SD011033122GA SD011051199SA SD011051215SA SD011053201GA SD011061417SA SD011061418SA SD011061419SA SD011061420SA SD011061421SA SD010502107GAMKE SD011061426SA SD011061427SA SD011061432SA SD011061445SA SD011061447SA SD011061450SA SD011061452SA SD011061453SA SD011061455SA SD011061456SA SD011061462SA SD011061476SA SD011061483SA SD011071576SA SD011071614SA SD011071629SA SD011072040SA SD011072041SA SD011072080SA SD011072118SA SD011073395GA SD011073429GA SD011073434GA SD011073436GA SD011073437GA SD011073480GA SD011073485GA SD011073492GA SD011073493GA SD011073506GA SD011073533GA SD011083582GA SD011083585GA SD011093742GA SD011103772GA SD011103774GA SD011103775GA SD011103776GA SD011103777GA SD011103778GA SD011103781GA SD011103782GA SD011103784GA SD011103785GA SD011103786GA SD011103787GA SD011103788GA SD011103791GA SD011103801GA SD011112728SA SD011113977GA SD011113980GA SD011113985GA SD011122926SA SD011122942SA SD011123243SA SD011123245SA SD011123247SA SD011123248SA SD011123250SA SD011123253SA SD011123255SA SD011123257SA SD011123258SA SD011123259SA SD011123260SA SD011123263SA SD011123277SA SD011123279SA SD011123280SA SD011123286SA SD011123287SA SD011124118GA SD011124122GA SD011124124GA SD011124125GA SD011124128GA SD011124157GA SD011124163GA SD011124164GA SD011073502GA SD011124167GA SD011133688SA SD011133689SA SD011133690SA SD011133696SA SD011133698SA SD011133708SA SD011133709SA SD011133720SA SD011133726SA SD011133729SA SD011144493GA SD011144498GA SD011144510GA SD011144511GA SD011154002SA SD011154628GA SD011164662GA SD011174153SA SD011174155SA SD011174157SA SD011174158SA SD011174160SA SD011174162SA SD011174163SA SD011174254SA SD011174257SA SD011174261SA SD011174267SA SD011174312SA SD011174315SA SD011174316SA SD011174322SA SD011174326SA SD011214887SA SHQ11250503SA SHQ11260509SA SHQ11260510SA SHQ11280774SA SHQ11290828SA SHQ11290831SA SHQ11290834SA SHQ11290838SA SHQ11290839SA SHQ11403091SA SHQ11403132SA SHQ11423192SA SHQ11423327SA SHQ11423330SA SHQ11423332SA SHQ11423334SA SHQ11423335SA SHQ11423337SA SHQ11423338SA SHQ11443597SA SHQ11454019SA SHQ11454021SA SHQ11464313SA SHQ11464321SA SHQ11495074SA SHQ11505307SA SHQ11505309SA SHQ11505311SA SHQ11505313SA SHQ12015476SA SHQ12015480SA SHQ12015482SA SHQ12045826SA SHQ12056003SA SHQ12066170SA SHQ12076264SA SHQ12076266SA SHQ12076268SA SHQ12076303SA SHQ12096355SA SHQ12096415SA SHQ12106495SA SHQ12136808SA SHQ12146882SA SHQ12146884SA SHQ12157024SA SHQ12197619SA SHQ12197620SA SHQ12197624SA SHQ12217840SA SHQ12218022SA SHQ12228238SA SHQ12228252SA SHQ12228263SA SHQ12228269SA SHQ12228270SA SHQ12228273SA SHQ12072692GA SHQ12217842SA SHQ12217865SA SHQ12217866SA SHQ12228236SA SHQ12228239SA SHQ12278982SA SHQ12278983SA SHQ12279012SA SHQ12279013SA SHQ12279016SA SHQ12309620SA SHQ12289146SA SHQ12289121SA SHQ12309622SA SHQ12289137SA SHQ12228237SA SHQ12309528SA SHQ12289147SA SHQ12289140SA SHQ12289089SA SHQ12289110SA SHQ12289094SA SHQ12289099SA SHQ12289102SA SHQ12309618SA SHQ12289085SA SHQ12309604SA SHQ12299435SA SHQ12299434SA SHQ12330825SA SHQ12330561SA SHQ12330559SA SHQ12330562SA SHQ12330565SA SHQ12361278SA SHQ12351116SA SHQ12330822SA SHQ12361277SA SHQ12361286SA SHQ12361281SA SHQ12361276SA SHQ12371291SA SHQ12371369SA SHQ12371290SA SHQ12371368SA SHQ12371371SA SHQ12371361SA SHQ12371363SA SHQ12371370SA SHQ12371365SA SHQ12371366SA SHQ12371372SA SHQ12371374SA SHQ12371287SA SHQ12371294SA SHQ12371296SA SHQ12371293SA SHQ12371289SA SHQ12371292SA SHQ12371288SA SHQ12371295SA SHQ12320274SA SHQ12351118SA SD010324798GA SD010366215GA SD010366219GA SD010366221GA SD010366282GA SD010376705GA SD010387332GA SD010387333GA SD010428953GA SD010429056GA SD010429060GA SD010429097GA SD010429099GA SD010429100GA SD010429101GA SD010429102GA SD010429103GA SD010512410GA SD011040706SA SD011040709SA SD011040710SA SD011040714SA SD011071554SA SD011071559SA SD011071563SA SD011071567SA SD011071569SA SD011071685SA SD011071686SA SD011071687SA SD011071688SA SD011071690SA SD011071691SA SD011071693SA SD011071695SA SD011071756SA SD011102518SA SD011112616SA SHQ11230038SA SHQ11362578SA SHQ11423344SA SHQ11423346SA SHQ11474538SA SHQ11474541SA SHQ11484650SA SHQ11484652SA SHQ11484879SA SHQ11484880SA SHQ12066083SA SHQ12066084SA SHQ12066092SA SHQ11451440GA SHQ11451446GA SD010376721GA SD010377125GA SD010377163GA SD010418511GA SD010418600GA SD010449395GA SD010449412GA SD010449427GA SD010449509GA SD010480112SA SD010520437SA SD011061302SA SD011061304SA SD011061305SA SD011061306SA SD011061308SA SD011061309SA SD011061315SA SHQ11423187SA SD010408259GA SD010408316GA SD010459931GA SD010459934GA SD010459939GA SD010430041SA SD010449394GA SD010449401GA SD010480103SA SD010500221SA SD010502255GA SD011030603SA SD011030604SA SD011051204SA SD011061389SA SD011061392SA SD011061396SA SD011061398SA SD011061404SA SD011061411SA SD011061412SA SD011061435SA SD011061436SA SD011061439SA SD011061443SA SD011061454SA SD011061458SA SD011061473SA SD011061490SA SD011061491SA SD011071578SA SD011071582SA SD011071584SA SD011071590SA SD011071596SA SD011071603SA SD011071605SA SD011071609SA SD011071630SA SD011071644SA SD011071654SA SD011071655SA SD011071656SA SD011071657SA SD011072050SA SD011072051SA SD011072052SA SD011072054SA SD011072067SA SD011072068SA SD011072069SA SD011154006SA SD011154007SA SD011154008SA SD011154010SA SD011154013SA SD011154015SA SD011154016SA SD011154018SA SD011154019SA SD011154020SA SD011174259SA SD011174272SA SD011174279SA SD011174287SA SD011174299SA SD011174301SA SHQ11260505SA SHQ11290830SA SHQ11290833SA SHQ11290840SA SHQ11290844SA SHQ11433395SA SHQ11443534SA SHQ11443535SA SHQ11443537SA SHQ11443539SA SHQ11474426SA SD010293018GA SD010303408GA SD010313740GA SD010345453GA SD010365969GA SD010387431GA SD010387530GA SD010387535GA SD010387548GA SD010387627GA SD010387638GA SD010387696GA SD010387698GA SD010387708GA SD010387710GA SD010387711GA SD010387713GA SD010387714GA SD010387715GA SD010387716GA SD010387717GA SD010387718GA SD010387724GA SD010387725GA SD010387732GA SD010387767GA SD010387849GA SD010387855GA SD010387858GA SD010387862GA SD010408360GA SD010418542GA SD010500252SA SD010460216GA SD010460218GA SD010460237GA SD010460245GA SD010460246GA SD010460248GA SD010460249GA SD010460251GA SD010470601GA SD010470621GA SD010501919GA SHQ11505327SA SHQ11505352SA SHQ11505379SA SD010365848GA SD010387372GA SD010387476GA SD010397888GA SD010397890GA SD010397892GA SD010397893GA SD010397894GA SD010397895GA SD010397898GA SD010397899GA SD010397901GA SD010397902GA SD010397904GA SD010397907GA SD010397908GA SD010397909GA SD010397910GA SD010397913GA SD010397916GA SD010470692GA SD010470694GA SD010470724GA SD010470787GA SD010470792GA SD010470795GA SD010470796GA SD010470797GA SD010470798GA SD010480877GA SD010480880GA SD010480893GA SD010481084GA SD010481085GA SD010481147GA SD010481158GA SD010481159GA SD010481160GA SD010481161GA SD010481168GA SD010481187GA SD010491359GA SD010491362GA SD010491438GA SD010491554GA SD010491556GA SD010491561GA SD010491658GA SD010491659GA SD010491660GA SD010491662GA SD010501760GA SD010501993GA SD010501995GA SD010502003GA SD010502010GA SD010502015GA SD010502245GA SD010512387GA SD010512398GA SD011063246GA SHQ11250161GA SHQ11250163GA SHQ11250167GA SHQ11280178GA SHQ11280180GA SHQ11280181GA SHQ11280182GA SHQ11280198GA SHQ11280199GA SHQ11280200GA SHQ11280202GA SHQ11280203GA SHQ11280204GA SHQ11280205GA SHQ11280206GA SHQ11280207GA SHQ11280208GA SHQ11280209GA SHQ11280210GA SHQ11280212GA SHQ11280214GA SHQ11280216GA SHQ11280217GA SHQ11280218GA SHQ11280219GA SHQ11280221GA SHQ11280228GA SHQ11280235GA SHQ11280239GA SHQ11290275GA SHQ11290276GA SHQ11290278GA SHQ11290279GA SHQ11290280GA SHQ11290281GA SHQ11290283GA SHQ11290284GA SHQ11290288GA SHQ11290290GA SHQ11290295GA SHQ11290298GA SHQ11290299GA SHQ11290300GA SHQ11290302GA SHQ11290303GA SHQ11290304GA SHQ11290307GA SHQ11290310GA SHQ11290311GA SHQ11290312GA SHQ11290314GA SHQ11290315GA SHQ11290316GA SHQ11290317GA SHQ11290321GA SHQ11310426GA SHQ11360823GA SHQ11370874GA SHQ11370875GA SHQ11370881GA SHQ11370885GA SHQ11370887GA SHQ11401199GA SHQ11401206GA SHQ11401212GA SHQ11411282GA SHQ11411283GA SHQ11411284GA SHQ11411292GA SHQ11411299GA SHQ11411300GA SHQ11411304GA SHQ11411305GA SHQ11411307GA SHQ11411308GA SHQ11411317GA SHQ11411320GA SHQ11411321GA SHQ11411324GA SHQ11411325GA SHQ11411326GA SHQ11411327GA SHQ11421328GA SHQ11421329GA SHQ11421330GA SHQ11421331GA SHQ11421332GA SHQ11421333GA SHQ11421334GA SHQ11421335GA SHQ11421336GA SHQ11421337GA SHQ11421338GA SHQ11421339GA SHQ11421340GA SHQ11421341GA SHQ11421342GA SHQ11421343GA SHQ11421344GA SHQ11421345GA SHQ11421346GA SHQ11421347GA SHQ11421348GA SHQ11451455GA SHQ11451456GA SHQ11451457GA SHQ11451459GA SHQ11451469GA SHQ11451475GA SHQ11451476GA SHQ11451477GA SHQ11451478GA SHQ11471559GA SHQ11471560GA SHQ11471561GA SHQ11471562GA SHQ11471564GA SHQ11471565GA SHQ11471566GA SHQ11471567GA SHQ11471568GA SHQ11471569GA SHQ11471570GA SHQ11471593GA SHQ11471594GA SHQ11481614GA SHQ11481617GA SHQ11481618GA SHQ11481620GA SHQ11481621GA SHQ11481622GA SHQ11481624GA SHQ11481625GA SHQ11502168GA SHQ11502170GA SHQ11502171GA SHQ11502172GA SHQ11502174GA SHQ11502175GA SHQ11512234GA SHQ11512235GA SHQ11512236GA SHQ11512238GA SHQ11512240GA SHQ11512241GA SHQ12052577GA SHQ12052580GA SHQ12052581GA SHQ12052582GA SHQ12052583GA SHQ12062606GA SHQ12062608GA SHQ12062609GA SHQ12092786GA SHQ12092787GA SHQ12092801GA SHQ12092808GA SHQ12092810GA SHQ12092811GA SHQ12102886GA SHQ12102895GA SHQ12374142GA SHQ12283731GA SHQ12374139GA SD010522545GA SHQ12374141GA SHQ12374143GA SD010481114GA SD010512413GA SHQ12374144GA SD010512323GA SHQ11381034GA SHQ11381038GA SHQ11403096SA SHQ11403098SA SHQ11403099SA SHQ11403102SA SHQ11403103SA SHQ11403104SA SHQ11403105SA SHQ11403106SA SHQ11403107SA SHQ11403109SA SHQ11403110SA SHQ11403113SA SHQ11403114SA SHQ11403115SA SHQ11403116SA SHQ11403120SA SHQ11403121SA SHQ11403125SA SHQ11403126SA SHQ11403127SA SHQ11403128SA SHQ11403129SA SHQ11403130SA SHQ11403131SA SHQ11403134SA SHQ11403135SA SHQ11403137SA SHQ11403140SA SHQ11403141SA SHQ11403142SA SHQ11403144SA SHQ11403146SA SHQ11403147SA SHQ11403148SA SHQ11411313GA SD010366051GA SD010418486GA SD010418488GA SD010418559GA SD010418571GA SD010418834GA SD010428954GA SD010429065GA SD010429171GA SD010510344SA SD010520371SA SD010520372SA SD010520374SA SD010520375SA SD010520376SA SD010520377SA SD010520378SA SD010520379SA SD010520380SA SD010520381SA SD010520383SA SD010520384SA SD010520385SA SD010520386SA SD010366380GA SD010418617GA SD010418777GA SD010428998GA SD010430010SA SD011040900SA SD011051183SA SD011061310SA SD011061348SA SD011061387SA SD011061511SA SD011071957SA SD011072132SA SD011072138SA SD011082175SA SD011092429SA SD011092431SA SD011092436SA SD011092439SA SD011092441SA SD011112740SA SD011112742SA SD011112757SA SD011112758SA SD011112763SA SD011112765SA SD011112790SA SD011112792SA SD011112793SA SD011112798SA SD011112799SA SD011123320SA SD011123328SA SD011133738SA SD011174242SA SD011174246SA SHQ11311584SA SHQ11311586SA SHQ11311595SA SHQ11311596SA SHQ11311601SA SHQ11311605SA SHQ11311606SA SHQ11311607SA SHQ11311608SA SHQ11311609SA SHQ11311611SA SHQ11311612SA SHQ11311613SA SHQ11321810SA SHQ11321815SA SHQ11362577SA SHQ11362580SA SHQ11362583SA SHQ11362586SA SHQ11372945SA SHQ11372949SA SHQ11372950SA SHQ11372951SA SHQ11372954SA SHQ11372957SA SHQ11372959SA SHQ11372965SA SHQ11423333SA SHQ11453938SA SHQ12391668SA SHQ12391670SA SHQ12391658SA SHQ12238474SAFRE SD009389253GA SD010303187GA SD010324498GA SD010376712GA SD010377164GA SD010377165GA SD010377166GA SD010377167GA SD010377170GA SD010377171GA SD010377173GA SD010387439GA SD010387594GA SD010387596GA SD010387600GA SD010387823GA SD010387832GA SD010408211GA SD010408212GA SD010408214GA SD010408215GA SD010408217GA SD010408219GA SD010408221GA SD010408222GA SD010408225GA SD010408227GA SD010408229GA SD010408230GA SD010408231GA SD010418512GA SD010418611GA SD010418615GA SD010418619GA SD010418623GA SD010418628GA SD010418629GA SD010418702GA SD010418757GA SD010418797GA SD010459920GA SD010459921GA SD010459922GA SD010459923GA SD010459925GA SD010459928GA SD010470506GA SD010470523GA SD010481231GA SD010481236GA SD010481244GA SD010481245GA SD010491335GA SD010491341GA SD010491342GA SD010491460GA SD010491584GA SD010491585GA SD010491586GA SD010491593GA SD010491595GA SD010491596GA SD010491597GA SD010491600GA SD010501987GA SD010502226GA SD010502227GA SD010502241GA SD010502274GA SD010502278GA SD010510357SA SD010512355GA SD010522505GA SD010522515GA SD010522520GA SD010532789GA SD010532799GA SD010532800GA SD010532802GA SD010532805GA SD010532807GA SD010532811GA SD010532813GA SD010532814GA SD011032936GA SD011032939GA SD011032948GA SD011032959GA SD011040718SA SD011082180SA SD011082182SA SD011082183SA SD011082185SA SD011082186SA SD011082188SA SD011082329SA SD011082336SA SD011082342SA SD011164123SA SD011194483SA SD011204654SA SD011204656SA SD011204662SA SD011214789SA SD011214816SA SD011214867SA SHQ11230053SA SHQ11230055SA SHQ11230059SA SHQ11235137SA SHQ11250475SA SHQ11250477SA SHQ11250492SA SHQ11250499SA SHQ11291036SA SHQ11362530SA SHQ11433415SA SHQ11453818SA SHQ11453886SA SHQ11453890SA SHQ11474518SA SHQ11484928SA SHQ11484944SA SHQ11484945SA SHQ11484955SA SHQ11484962SA SHQ11484963SA SHQ11495068SA SHQ11495161SA SHQ11495169SA SHQ11495170SA SHQ11495173SA SHQ12066093SA SHQ12066190SA SHQ12066191SA SHQ12066193SA SHQ12096389SA SHQ12126689SA SHQ12126697SA SHQ12218002SA SHQ12218003SA SHQ12361174SA SHQ12361175SA SHQ12361173SA SHQ12361163SA SHQ12361165SA SHQ12371486SA SD010387839GA SD010470410GA SD010470432GA SD010481230GA SD010481247GA SD010491331GA SD010491504GA SD010502018GA SHQ11321802SA SD010430076SA SD010481290GA SD010491507GA SD010491511GA SD010502133GA SD011030648SA SD011040782SA SD011040816SA SD011050944SA SD011053266GA SD011071827SA SD011071890SA SD011112796SA SD011143891SA SD011164099SA SD011164105SA SD011164113SA SD011194569SA SD011204741SA SD011214784SA SD011235118SA SHQ11240178SA SHQ11260536SA SHQ11301203SA SHQ11311422SA SHQ11443607SA SHQ11453973SA SHQ11453976SA SHQ11453978SA SHQ11453981SA SHQ11453982SA SHQ11453983SA SHQ11453984SA SHQ11453989SA SHQ11464282SA SHQ11484729SA SHQ11505208SA SHQ11505210SA SHQ11505211SA SHQ11505212SA SHQ11505215SA SHQ11505217SA SHQ12025510SA SHQ12035571SA SHQ12055999SA SHQ12056000SA SHQ12096448SA SHQ12106489SA SHQ12106492SA SHQ12106510SA SHQ12126672SA SHQ12156917SA SHQ12197555SA SHQ11250162GA SHQ11250171GA SHQ12299420SA SHQ12299424SA SHQ12299425SA SHQ12299428SA SHQ12299430SA SHQ12371362SA SHQ12371373SA SHQ12371367SA SD010366042GA SD011124087GA SD011124148GA SD010313836GA SD010313839GA SD010313848GA SD010314035GA SD010314164GA SD010345403GA SD010345414GA SD010345419GA SD010345422GA SD010345430GA SD010365912GA SD010366006GA SD010366358GA SD010376641GA SD010376644GA SD010376665GA SD010376811GA SD010376836GA SD010376930GA SD010376932GA SD010376935GA SD010387272GA SD010387273GA SD010387277GA SD010387371GA SD010387576GA SD010387587GA SD010408163GA SD010408165GA SD010408167GA SD010408168GA SD010408179GA SD010408185GA SD010408186GA SD010408191GA SD010408303GA SD010408421GA SD010408422GA SD010408423GA SD010408424GA SD010408426GA SD010408427GA SD010408428GA SD010408429GA SD010408430GA SD010408431GA SD010408432GA SD010408435GA SD010408436GA SD010408437GA SD010408438GA SD010408439GA SD010408440GA SD010408441GA SD010408442GA SD010408444GA SD010418739GA SD010418740GA SD010418747GA SD010418750GA SD010418753GA SD010418755GA SD010429127GA SD010449574GA SD010449577GA SD010449593GA SD010449657GA SD010449659GA SD010449674GA SD010449705GA SD010450020GA SD010450021GA SD010459741GA SD010459854GA SD010459871GA SD010459872GA SD010460127GA SD010460155GA SD010460163GA SD010460165GA SD010470571GA SD010501924GA SD010502177GA SD010522441GA SD010522443GA SD010522447GA SD010522448GA SD010522450GA SD010522453GA SD010522459GA SD010522461GA SD010522462GA SD010522468GA SD011033097GA SD011033098GA SD011033102GA SD011043165GA SD011043166GA SD011043167GA SD011043168GA SD011043169GA SD011043170GA SD011043171GA SD011043172GA SD011043173GA SD011043174GA SD011043178GA SD011043180GA SD011073482GA SD011073490GA SD011073496GA SD011073497GA SD011073498GA SD011073501GA SD011073504GA SD011073508GA SD011073509GA SD011073513GA SD011073514GA SD011083563GA SD011083566GA SD011083567GA SD011083569GA SD011083571GA SD011083633GA SD011083634GA SD011083635GA SD011083636GA SD011083637GA SD011083639GA SD011083640GA SD011083641GA SD011083642GA SD011083643GA SD011083644GA SD011083645GA SD011083646GA SD011083647GA SD011083648GA SD011083649GA SD011083650GA SD011083651GA SD011083652GA SD011083653GA SD011083654GA SD011083655GA SD011083656GA SD011083657GA SD011083658GA SD011083659GA SD011083661GA SD011083662GA SD011083663GA SD011083664GA SD011083665GA SD011083666GA SD011083667GA SD011083668GA SD011083669GA SD011083670GA SD011083671GA SD011083672GA SD011083674GA SD011083675GA SD011083676GA SD011083677GA SD011083678GA SD011083679GA SD011083680GA SD011083681GA SD011083682GA SD011083683GA SD011083684GA SD011093686GA SD011093687GA SD011093688GA SD011093690GA SD011093691GA SD011093692GA SD011093693GA SD011093694GA SD011113851GA SD011113942GA SD011124084GA SD011124131GA SD011124139GA SD011124141GA SD011124146GA SD011144477GA SD011144481GA SD011144487GA SD011144491GA SHQ11320576GA SHQ11320582GA SHQ11320583GA SHQ11320589GA SHQ11381030GA SHQ11381032GA SHQ11381033GA SHQ11381035GA SHQ11381036GA SHQ11381037GA SHQ11411314GA SHQ11411315GA SHQ11441414GA SHQ11441415GA SHQ11441416GA SHQ11441418GA SHQ11441421GA SHQ11451437GA SHQ12032455GA SHQ12032459GA SHQ12032462GA SHQ12032465GA SHQ12062605GA SHQ12062607GA SHQ12062610GA SHQ12062623GA SHQ12062627GA SHQ12062629GA SHQ12062630GA SHQ12062635GA SHQ12062636GA SHQ12092828GA SHQ12092830GA SHQ12092831GA SHQ12153064GA SHQ12173158GA SHQ12183323GA SHQ12183324GA SHQ12183325GA SHQ12183326GA SHQ12183327GA SHQ12193346GA SHQ12253674GA SHQ12253680GA SHQ12092825GA SHQ12173164GA SHQ12253675GA SHQ12253676GA SD010335029GA SD010366276GA SD010418604GA SD010418658GA SD010418664GA SD010418899GA SD010418914GA SD010418918GA SD010428938GA SD010428973GA SD010429062GA SD010429074GA SD010429076GA SD010429172GA SD011061339SA SD011061341SA SD011071528SA SD011071616SA SD011071639SA SD011071642SA SD011071650SA SD011072108SA SD011082205SA SD011082207SA SD011082209SA SD011082211SA SD011082214SA SD011082216SA SD011082224SA SD011082225SA SD011082237SA SD011082240SA SD011082305SA SD011082307SA SD011082308SA SD011123269SA SD011123281SA SD011123283SA SD011174170SA SD011174314SA SD011174325SA SD011174327SA SD011174328SA SHQ11240176SA SHQ11240177SA SHQ11240179SA SHQ11352351SA SHQ11403100SA SHQ11433479SA SHQ11433533SA SHQ11443549SA SHQ11443550SA SHQ11443551SA SHQ11443552SA SHQ11443556SA SHQ11443700SA SHQ11453776SA SHQ11453778SA SHQ11453779SA SHQ11453780SA SHQ11453781SA SHQ11453790SA SHQ11453791SA SHQ11453854SA SHQ11453887SA SHQ11464112SA SHQ11464118SA SHQ11464122SA SHQ11464123SA SHQ11464373SA SHQ11464374SA SHQ11464377SA SHQ11474423SA SHQ11474430SA SHQ11474444SA SHQ11474454SA SHQ11474460SA SHQ11474462SA SHQ11474520SA SHQ11474527SA SHQ11484798SA SHQ11484864SA SHQ11484980SA SHQ11484981SA SHQ11495186SA SHQ12045827SA SHQ12167167SA SHQ12197622SA SHQ12207765SA SHQ12258834SA SHQ12258835SA SHQ12258836SA SHQ12258837SA SHQ12258838SA SD011123249SA SD011123271SA SD010459719GA SD010459827GA SD010459831GA SD011033087GA SD011033088GA SD011033094GA SHQ11301128SA SHQ11301138SA SHQ11311437SA SHQ11382981SA SHQ11382983SA SHQ11382984SA SHQ11382985SA SHQ11382986SA SHQ11382987SA SHQ11382988SA SHQ11382989SA SHQ11382990SA SHQ11382992SA SHQ11382994SA SHQ11382996SA SHQ12096449SA SHQ12096453SA SHQ12096460SA SHQ12116553SA SHQ12116557SA SHQ12279028SA SHQ12279029SA SD010470359GA SD011122899SA SD011123315SA SD011123359SA SD011174224SA SD011174225SA SD011174234SA SD010116805GA SD010449571GA SD010449644GA SD011061519SA SD011071537SA SD011124227GA SD011124230GA SD011134301GA SD011134343GA SD011154001SA SD011154011SA SHQ11454020SA SHQ11464378SA SHQ11464379SA SHQ11464380SA SHQ11464381SA SHQ11464382SA SHQ12052590GA SHQ12052592GA SHQ12238472SA SHQ12238474SA SHQ12238483SA SHQ12238488SA SHQ12238489SA SHQ12248674SA SHQ12248675SA SD010138316GA SD010138470GA SD010272368GA SD010272379GA SD010313850GA SD010313872GA SD010313876GA SD010313955GA SD010314260GA SD010314368GA SD010345520GA SD010376634GA SD010376666GA SD010376679GA SD010376800GA SD010377042GA SD010377053GA SD010377058GA SD010377078GA SD010387233GA SD010387374GA SD010387377GA SD010387500GA SD010387537GA SD010387545GA SD010387632GA SD010408310GA SD010418606GA SD010418667GA SD010418896GA SD010418897GA SD010428937GA SD010429052GA SD010429063GA SD010429064GA SD010429066GA SD010429070GA SD010429071GA SD010429118GA SD010429124GA SD010429128GA SD010429130GA SD010429135GA SD010429222GA SD010449476GA SD010449504GA SD010449511GA SD010449512GA SD010449514GA SD010449516GA SD010450001GA SD010450008GA SD010450013GA SD010450029GA SD010450030GA SD010450032GA SD010450033GA SD010450036GA SD010450040GA SD010450043GA SD010459753GA SD010459756GA SD010459759GA SD010459760GA SD010459770GA SD010470471GA SD010481096GA SD010481098GA SD010481101GA SD010481104GA SD010481105GA SD010481110GA SD010481116GA SD010481270GA SD010501994GA SD010501999GA SD010502104GA SD010502120GA SD010502131GA SD010502141GA SD010512322GA SD010512409GA SD010512417GA SD010522542GA SD010522547GA SD010522553GA SD010522558GA SD010522559GA SD010522561GA SD010522563GA SD010522564GA SD010522565GA SD010532787GA SD010532824GA SD010532830GA SD010532831GA SD010532833GA SD011033004GA SD011033036GA SD011033100GA SD011103771GA SD011103773GA SD011103779GA SD011103783GA SD011103798GA SD011103799GA SD011113842GA SD011113843GA SD011113908GA SD011113912GA SD011113923GA SD011113924GA SD011113933GA SD011113934GA SD011113935GA SD011113937GA SD011113938GA SD011113940GA SD011113943GA SD011124044GA SD011124048GA SD011124080GA SD011124082GA SD011124083GA SD011124085GA SD011124088GA SD011124089GA SD011124137GA SD011124138GA SD011124140GA SD011124147GA SD011124149GA SD011124211GA SD011134273GA SD011134275GA SD011134276GA SD011134279GA SD011134280GA SD011134281GA SD011134282GA SD011134283GA SD011134284GA SD011134285GA SD011134377GA SD011134386GA SD011134387GA SD011134388GA SD011134392GA SD011144437GA SD011144438GA SD011144439GA SD011144440GA SD011144441GA SD011144442GA SD011144443GA SD011144445GA SD011144446GA SD011144447GA SD011144450GA SD011144451GA SD011144452GA SD011144453GA SD011144454GA SD011144488GA SD011144500GA SD011144503GA SD011184914GA SHQ11190013GA SHQ11320590GA SHQ11320591GA SHQ11320593GA SHQ11320596GA SHQ11320598GA SHQ11360792GA SHQ11360793GA SHQ11360794GA SHQ11360796GA SHQ11360797GA SHQ11360798GA SHQ11360799GA SHQ11360800GA SHQ11360801GA SHQ11360802GA SHQ11360803GA SHQ11360804GA SHQ11360805GA SHQ11360806GA SHQ11360807GA SHQ11360809GA SHQ11360810GA SHQ11360811GA SHQ11360813GA SHQ11360820GA SHQ11381065GA SHQ11391146GA SHQ11391153GA SHQ11401179GA SHQ11401181GA SHQ11401183GA SHQ11401185GA SHQ11401187GA SHQ11401192GA SHQ11401201GA SHQ11401207GA SHQ11401218GA SHQ11401221GA SHQ11401233GA SHQ11401235GA SHQ11401236GA SHQ11401237GA SHQ11401238GA SHQ11401239GA SHQ11401240GA SHQ11401252GA SHQ11401254GA SHQ11401256GA SHQ11401257GA SHQ11401259GA SHQ11401260GA SHQ11401262GA SHQ11411267GA SHQ11411270GA SHQ11411271GA SHQ11411273GA SHQ11411275GA SHQ11411277GA SHQ11411278GA SHQ11411279GA SHQ11411280GA SHQ11411281GA SHQ11411285GA SHQ11411302GA SHQ11411303GA SHQ11411309GA SHQ11411322GA SHQ11441394GA SHQ11441395GA SHQ11441396GA SHQ11441397GA SHQ11441398GA SHQ11441399GA SHQ11441400GA SHQ11441401GA SHQ11441402GA SHQ11441403GA SHQ11441404GA SHQ11441405GA SHQ11441406GA SHQ11441408GA SHQ11441412GA SHQ11441417GA SHQ11441420GA SHQ11451429GA SHQ11451434GA SHQ11451436GA SHQ11451441GA SHQ11451443GA SHQ11451445GA SHQ11451447GA SHQ11451448GA SHQ11451449GA SHQ11451451GA SHQ11481784GA SHQ11491924GA SHQ11491925GA SHQ11491926GA SHQ11491927GA SHQ11491928GA SHQ11491930GA SHQ11491931GA SHQ11491932GA SHQ11491934GA SHQ11491935GA SHQ11491936GA SHQ11491937GA SHQ11491938GA SHQ11502030GA SHQ11502076GA SHQ11502164GA SHQ11502165GA SHQ12022390GA SHQ12022395GA SHQ12022397GA SHQ12022398GA SHQ12022411GA SHQ12022427GA SHQ12032453GA SHQ12032454GA SHQ12032456GA SHQ12032457GA SHQ12032458GA SHQ12032460GA SHQ12032461GA SHQ12032463GA SHQ12032464GA SHQ12045722SA SD010272460GA SD010366139GA SD010366232GA SD010366278GA SD010376772GA SD010387364GA SD010387370GA SD010408193GA SD010408197GA SD010408199GA SD010470356GA SD010480976GA SD010501824GA SD010501828GA SD010501920GAMKE SD010501956GA SD010502270GA SD010512337GA SD010512353GA SD010512354GA SD010512363GA SD010512364GA SD010512375GA SD010512376GA SD010512378GA SD010512382GA SD010522449GA SD010522503GA SD010522507GA SD010522509GA SD010522511GA SD010522526GA SD010522533GA SD010532788GA SD010532792GA SD010532793GA SD010532794GA SD010532795GA SD010532806GA SD010532815GA SD010532816GA SD011112606SA SD011112710SA SD011122916SA SD011122933SA SD011123264SA SD011123340SA SD011133716SA SD011133721SA SD011133722SA SD011133723SA SD011133728SA SD011174243SA SD011174313SA SD011174318SA SD011194583SA SD011194584SA SD011194586SA SD011194587SA SD011194590SA SD011194601SA SD011194608SA SD011194612SA SD011194618SA SHQ11250470SA SHQ11250473SA SHQ11301196SA SHQ11311442SA SHQ11311447SA SHQ11311448SA SHQ11311451SA SHQ11321847SA SHQ11321849SA SHQ11321854SA SHQ11321857SA SHQ11321859SA SHQ11321861SA SHQ11321862SA SHQ11321874SA SHQ11321877SA SHQ11321879SA SHQ11321880SA SHQ11321883SA SHQ11321884SA SHQ11331947SA SHQ11331948SA SHQ11331954SA SHQ11332050SA SHQ11332051SA SHQ11332053SA SHQ11332054SA SHQ11332057SA SHQ11342133SA SHQ11342134SA SHQ11342136SA SHQ11342139SA SHQ11342141SA SHQ11342149SA SHQ11342262SA SHQ11342263SA SHQ11342264SA SHQ11342265SA SHQ11342266SA SHQ11342267SA SHQ11342268SA SHQ11342269SA SHQ11342271SA SHQ11342272SA SHQ11342273SA SHQ11342274SA SHQ11342275SA SHQ11342317SA SHQ11352474SA SHQ11352479SA SHQ11352487SA SHQ11352489SA SHQ11352492SA SHQ11352495SA SHQ11352496SA SHQ11352500SA SHQ11352501SA SHQ11352502SA SHQ11352503SA SHQ11352505SA SHQ11352508SA SHQ11362579SA SHQ11362700SA SHQ11372775SA SHQ11372802SA SHQ11372803SA SHQ11372890SA SHQ11372953SA SHQ11372955SA SHQ11372958SA SHQ11372962SA SHQ11372963SA SHQ11423199SA SHQ11423201SA SHQ11423202SA SHQ11423203SA SHQ11423204SA SHQ11423205SA SHQ11423206SA SHQ11423207SA SHQ11423208SA SHQ11423209SA SHQ11423210SA SHQ11423211SA SHQ11423212SA SHQ11423214SA SHQ11423215SA SHQ11423217SA SHQ11423219SA SHQ11423221SA SHQ11423223SA SHQ11423224SA SHQ11423225SA SHQ11423227SA SHQ11423231SA SHQ11423233SA SHQ11454033SA SHQ11454041SA SHQ11454043SA SHQ11454052SA SHQ11484795SA SHQ11484803SA SHQ11484926SA SHQ11484935SA SHQ11484940SA SHQ11484941SA SHQ11484946SA SHQ12066153SA SHQ12076306SA SHQ12076312SA SHQ12126646SA SHQ12126650SA SHQ12126653SA SHQ12126659SA SHQ12126660SA SHQ12126667SA SHQ12126669SA SHQ12126670SA SHQ12167063SA SHQ12167197SA SHQ12187526SA SHQ12187528SA SHQ12187530SA SHQ12187533SA SHQ12258746SA SHQ12268861SA SHQ12279002SA SHQ12279003SA SHQ12279005SA SHQ12279006SA SHQ12310008SA SHQ12310010SA SHQ12310011SA SHQ12310025SA SHQ12309698SA SD010127683GA SD010189702GA SD010303265GA SD010313741GA SD010313840GA SD010314167GA SD010324440GA SD010345405GA SD010345416GA SD010345428GA SD010345449GA SD010365944GA SD010366168GA SD010366302GA SD010376571GA SD010376584GA SD010376656GA SD010387415GA SD010387424GA SD010387425GA SD010387426GA SD010387427GA SD010387429GA SD010387437GA SD010387505GA SD010387528GA SD010387529GA SD010387540GA SD010387543GA SD010387546GA SD010387547GA SD010387550GA SD010387553GA SD010387554GA SD010387625GA SD010387628GA SD010387639GA SD010387691GA SD010387699GA SD010387705GA SD010387729GA SD010387746GA SD010387764GA SD010387845GA SD010387848GA SD010387853GA SD010387857GA SD010387859GA SD010387860GA SD010387861GA SD010387863GA SD010397924GA SD010397935GA SD010397938GA SD010397949GA SD010397950GA SD010397951GA SD010397953GA SD010397954GA SD010397956GA SD010397957GA SD010397959GA SD010397960GA SD010397962GA SD010397963GA SD010397965GA SD010397966GA SD010397967GA SD010397979GA SD010397980GA SD010397981GA SD010397982GA SD010397984GA SD010397987GA SD010408077GA SD010418707GA SD010429166GA SD010429169GA SD010449454GA SD010449461GA SD010449463GA SD010449465GA SD010449469GA SD010449531GA SD010449532GA SD010449537GA SD010449681GA SD010449682GA SD010449686GA SD010449689GA SD010449692GA SD010449701GA SD010450011GA SD010459709GA SD010460112GA SD010460120GA SD010470439GA SD010470485GA SD010470491GA SD010470512GA SD010470519GA SD010470522GA SD010470526GA SD010470528GA SD010470535GA SD010470539GA SD010470543GA SD010470544GA SD010470568GA SD010470576GA SD010470654GA SD010470734GA SD010470744GA SD010480914GA SD010480953GA SD010480993GA SD010481006GA SD010481018GA SD010481227GA SD010481276GA SD010491326GA SD010491447GA SD010491449GA SD010501851GA SD010501852GA SD010501858GA SD010501863GA SD010512406GA SD010512411GA SD010512415GA SD010512416GA SD010512418GA SD010522611GA SD010522618GA SD010522649GA SD010522651GA SD010522656GA SD010522659GA SD010522660GA SD010522661GA SD010532756GA SD010532759GA SD010532768GA SD010532769GA SD010532770GA SD010532772GA SD010532773GA SD010532774GA SD010532778GA SD010532784GA SD011032923GA SD011032924GA SD011033011GA SD011033012GA SD011033014GA SD011033021GA SD011033032GA SD011033037GA SD011033039GA SD011033042GA SD011033067GA SD011033069GA SD011033070GA SD011033077GA SD011033079GA SD011033082GA SD011033085GA SD011033089GA SD011033099GA SD011033101GA SD011063258GA SD011103764GA SD011103808GA SD011113850GA SD011113880GA SD011113887GA SD011113941GA SD011113944GA SD011113949GA SD011113956GA SD011113960GA SD011113961GA SD011113962GA SD011113963GA SD011113964GA SD011113965GA SD011113967GA SD011113993GA SD011113995GA SD011113996GA SD011113997GA SD011113998GA SD011113999GA SD011114000GA SD011114001GA SD011114002GA SD011114003GA SD011124054GA SD011124055GA SD011124056GA SD011124057GA SD011124058GA SD011124059GA SD011124060GA SD011124061GA SD011124062GA SD011124063GA SD011124064GA SD011124065GA SD011124066GA SD011124067GA SD011124068GA SD011124069GA SD011124070GA SD011124071GA SD011124072GA SD011124073GA SD011124074GA SD011124075GA SD011124076GA SD011124077GA SD011124079GA SD011124081GA SD011124086GA SD011124090GA SD011124091GA SD011124092GA SD011124093GA SD011124094GA SD011124095GA SD011124096GA SD011124097GA SD011124098GA SD011124099GA SD011124100GA SD011124101GA SD011124102GA SD011124104GA SD011124135GA SD011124144GA SD011124145GA SD011124169GA SD011124170GA SD011124171GA SD011124172GA SD011124175GA SD011124176GA SD011124177GA SD011124179GA SD011124185GA SD011124186GA SD011124187GA SD011124188GA SD011124189GA SD011124190GA SD011124192GA SD011124196GA SD011124197GA SD011124198GA SD011124200GA SD011124201GA SD011124202GA SD011124203GA SD011124204GA SD011124220GA SD011124221GA SD011124222GA SD011124223GA SD011124224GA SD011124225GA SD011124226GA SD011124229GA SD011133368SA SD011134247GA SD011134248GA SD011134249GA SD011134255GA SD011134258GA SD011134288GA SD011134289GA SD011134290GA SD011134292GA SD011134293GA SD011134295GA SD011134296GA SD011134310GA SD011134339GA SD011134342GA SD011134344GA SD011134347GA SD011134350GA SD011134352GA SD011134354GA SD011134356GA SD011134357GA SD011134358GA SD011134359GA SD011134360GA SD011134361GA SD011134363GA SD011134364GA SD011134366GA SD011134367GA SD011134368GA SD011134369GA SD011134370GA SD011134371GA SD011134372GA SD011134373GA SD011134375GA SD011134376GA SD011144449GA SD011144455GA SD011144456GA SD011154532GA SD011154533GA SD011154534GA SD011154536GA SD011154537GA SD011154542GA SD011154543GA SD011154545GA SD011154546GA SD011154550GA SD011154573GA SD011154576GA SD011154577GA SD011154579GA SD011154580GA SD011154582GA SD011154583GA SD011154587GA SD011154588GA SD011154589GA SD011164686GA SD011174751GA SD011174763GA SD011174769GA SD011174770GA SD011174774GA SD011174775GA SD011174776GA SD011174777GA SD011174779GA SD011174780GA SD011174783GA SD011174784GA SD011174788GA SD011174790GA SD011174792GA SD011174793GA SD011174795GA SD011174796GA SD011174798GA SD011174800GA SD011174806GA SD011174807GA SD011174809GA SD011184890GA SD011184893GA SD011184898GA SD011184904GA SD011184905GA SD011184907GA SD011205013GA SD011205014GA SD011205018GA SD011205026GA SD011215050GA SD011215058GA SHQ11290342GA SHQ11290348GA SHQ11290350GA SHQ11290356GA SHQ11300359GA SHQ11300361GA SHQ11300362GA SHQ11300366GA SHQ11300367GA SHQ11300368GA SHQ11300370GA SHQ11300372GA SHQ11300373GA SHQ11300379GA SHQ11300380GA SHQ11300383GA SHQ11300385GA SHQ11300386GA SHQ11300387GA SHQ11300391GA SHQ11300393GA SHQ11300394GA SHQ11300395GA SHQ11300396GA SHQ11300397GA SHQ11300398GA SHQ11300401GA SHQ11300402GA SHQ11300403GA SHQ11300404GA SHQ11300405GA SHQ11300407GA SHQ11300408GA SHQ11300409GA SHQ11300412GA SHQ11300413GA SHQ11300418GA SHQ11300420GA SHQ11310528GA SHQ11310529GA SHQ11310530GA SHQ11310534GA SHQ11310535GA SHQ11310536GA SHQ11310537GA SHQ11320575GA SHQ11321667SA SHQ11321697SA SHQ11321724SA SHQ11342299SA SHQ11342315SA SHQ11342316SA SHQ11342319SA SHQ11352393SA SHQ11352397SA SHQ11352400SA SHQ11362623SA SHQ11362634SA SHQ11362636SA SHQ11362638SA SHQ11362647SA SHQ11362652SA SHQ11362654SA SHQ11362655SA SHQ11362656SA SHQ11362661SA SHQ11362664SA SHQ11362665SA SHQ11362666SA SHQ11370870GA SHQ11370879GA SHQ11372784SA SHQ11372791SA SHQ11372820SA SHQ11372830SA SHQ11372837SA SHQ11423162SA SHQ11423165SA SHQ11423172SA SHQ11423280SA SHQ11423285SA SHQ11423301SA SHQ11423307SA SHQ11423313SA SHQ11423315SA SHQ11423318SA SHQ11423321SA SHQ11423324SA SHQ11433428SA SHQ11433452SA SHQ11443743SA SHQ11443747SA SHQ11443750SA SHQ11443751SA SHQ11443752SA SHQ11443753SA SHQ11443754SA SHQ11443755SA SHQ11443758SA SHQ11443762SA SHQ11443764SA SHQ11443767SA SHQ11454023SA SHQ11454023SABOP SHQ11454026SA SHQ11454026SABOP SHQ11454030SA SHQ11454030SABOP SHQ11454031SA SHQ11454031SABOP SHQ11454032SA SHQ11454032SABOP SHQ11454034SA SHQ11454034SABOP SHQ11454036SA SHQ11454036SABOP SHQ11454037SA SHQ11454037SABOP SHQ11454038SA SHQ11454038SABOP SHQ11454040SA SHQ11454040SABOP SHQ11454044SA SHQ11454044SABOP SHQ11454050SA SHQ11454050SABOP SHQ11464336SA SHQ11464336SABOP SHQ11464338SA SHQ11464340SA SHQ11464340SABOP SHQ11464341SA SHQ11464342SA SHQ11464342SABOP SHQ11464343SA SHQ11464344SA SHQ11464345SA SHQ11464346SA SHQ11464347SA SHQ11464348SA SHQ11464352SA SHQ11474548SA SHQ11474549SA SHQ11474550SA SHQ11474552SA SHQ11474553SA SHQ11474554SA SHQ11474555SA SHQ11474556SA SHQ11474557SA SHQ11474561SA SHQ11474562SA SHQ11474637SA SHQ11495184SA SHQ11495185SA SHQ11495187SA SHQ11495188SA SHQ11495190SA SHQ11495191SA SHQ11495192SA SHQ11495194SA SHQ11495195SA SHQ11495196SA SHQ11495197SA SHQ11495198SA SHQ11512227GA SHQ11512230GA SHQ11512231GA SHQ12012345GA SHQ12012346GA SHQ12012347GA SHQ12012348GA SHQ12012350GA SHQ12012351GA SHQ12012352GA SHQ12022394GA SHQ12022425GA SHQ12025508SA SHQ12025508SABOP SHQ12032452GA SHQ12032466GA SHQ12032467GA SHQ12032486GA SHQ12042557GA SHQ12052589GA SHQ12052594GA SHQ12052595GA SHQ12052596GA SHQ12062626GA SHQ12112932GA SHQ12112934GA SHQ12112936GA SHQ12122969GA SHQ12122970GA SHQ12122971GA SHQ12122977GA SHQ12122978GA SHQ12122979GA SHQ12122982GA SHQ12122983GA SHQ12122985GA SHQ12122987GA SHQ12122988GA SHQ12122989GA SHQ12122990GA SHQ12132991GA SHQ12132992GA SHQ12132993GA SHQ12153048GA SHQ12153049GA SHQ12153050GA SHQ12153051GA SHQ12153053GA SHQ12153054GA SHQ12153055GA SHQ12153056GA SHQ12153057GA SHQ12167089SA SHQ12167090SA SHQ12167174SA SHQ12167177SA SHQ12167191SA SHQ12173165GA SHQ12173166GA SHQ12173167GA SHQ12173168GA SHQ12173169GA SHQ12173205GA SHQ12173206GA SHQ12173207GA SHQ12173209GA SHQ12173211GA SHQ12173212GA SHQ12173213GA SHQ12173215GA SHQ12173217GA SHQ12173218GA SHQ12173219GA SHQ12173220GA SHQ12173221GA SHQ12173223GA SHQ12173224GA SHQ12173225GA SHQ12173226GA SHQ12173227GA SHQ12173228GA SHQ12173230GA SHQ12173231GA SHQ12173232GA SHQ12173233GA SHQ12173234GA SHQ12173235GA SHQ12173236GA SHQ12173237GA SHQ12173238GA SHQ12173239GA SHQ12173240GA SHQ12173241GA SHQ12173242GA SHQ12173243GA SHQ12173244GA SHQ12173245GA SHQ12173246GA SHQ12173247GA SHQ12173248GA SHQ12173249GA SHQ12173250GA SHQ12173251GA SHQ12173254GA SHQ12173255GA SHQ12173256GA SHQ12173257GA SHQ12173258GA SHQ12173259GA SHQ12177303SA SHQ12177370SA SHQ12177373SA SHQ12177375SA SHQ12177377SA SHQ12177381SA SHQ12177389SA SHQ12177391SA SHQ12177395SA SHQ12183273GA SHQ12183274GA SHQ12183275GA SHQ12183276GA SHQ12183277GA SHQ12183278GA SHQ12183280GA SHQ12183281GA SHQ12183282GA SHQ12183283GA SHQ12183284GA SHQ12183285GA SHQ12187410SA SHQ12187411SA SHQ12187412SA SHQ12187413SA SHQ12187416SA SHQ12187417SA SHQ12187419SA SHQ12187420SA SHQ12187421SA SHQ12187423SA SHQ12187424SA SHQ12187425SA SHQ12187426SA SHQ12187427SA SHQ12187428SA SHQ12187429SA SHQ12187431SA SHQ12187432SA SHQ12187438SA SHQ12218079SA SHQ12279010SA SHQ12279014SA SHQ12279015SA SHQ12279017SA SHQ12279018SA SHQ12279019SA SHQ12289224SA SHQ12187537SA SD011033113GA SD011224927SA SD011224928SA SD011224977SA SHQ11505308SA SHQ12045830SA SHQ12066181SA SHQ12076235SA SHQ12278963SA SHQ12278964SA SD010501779GA SD010501783GA SD010501784GA SD010501786GA SD010501789GA SD010501792GA SD011020519SA SD011020527SA SD011072000SA SD011072032SA SD011072146SA SD011072157SA SD011113946GA SD011113948GA SD011113954GA SD011113958GA SD011113959GA SD011194599SA SHQ11250400SA SHQ11260539SA SHQ11280821SA SHQ11280822SA SHQ11383005SA SHQ11383008SA SHQ12045836SA SHQ12207818SA SHQ12207824SA SHQ12207829SA SD010127746GA SD010127764GA SD010262001GA SD010272403GA SD010272412GA SD010366097GA SD010366104GA SD010376808GA SD010387460GA SD010408318GA SD010429059GA SD010429068GA SD010429069GA SD010430039SA SD010430063SA SD010430064SA SD010430066SA SD010430068SA SD010430069SA SD010430070SA SD010430071SA SD010481067GA SD010481087GA SD010481132GA SD010491386GA SD010491387GA SD010491696GA SD010500235SA SD010502100GA SD010502102GA SD010502302GA SD010510333SA SD010520382SA SD010522473GA SD010522486GA SD010522491GA SD010522642GA SD011030624SA SD011030673SA SD011030674SA SD011030675SA SD011030676SA SD011030680SA SD011030682SA SD011040755SA SD011050934SA SD011051163SA SD011051179SA SD011051182SA SD011051187SA SD011051191SA SD011051200SA SD011053184GA SD011061401SA SD011071530SA SD011071532SA SD011071587SA SD011071899SA SD011072094SA SD011072095SA SD011072096SA SD011072097SA SD011072098SA SD011072106SA SD011072111SA SD011082311SA SD011112548SA SD011133735SA SD011133737SA SD011143907SA SD011143909SA SD011174247SA SD011194486SA SD011194491SA SD011225067SA SHQ11260506SA SHQ11280760SA SHQ11280764SA SHQ11291023SA SHQ11301232SA SHQ11423329SA SHQ11474420SA SHQ11474424SA SHQ11474425SA SHQ11474429SA SHQ11474431SA SHQ11474432SA SHQ11505300SA SHQ12025533SA SHQ12025536SA SHQ12025539SA SHQ12035585SA SHQ12035641SA SHQ12045723SA SHQ12045742SA SHQ12066061SA SHQ12066062SA SHQ12153067GA SHQ12238379SA SHQ12238465SA SHQ12309587SA SD010376713GA SD010376720GA SD010377156GA SD010428992GA SD010449455GA SD010491521GA SD010491523GA SD010491526GA SD010491575GA SD010500239SA SD010500254SA SD010502308GA SD010502311GA SD010502313GA SD010510340SA SD010512390GA SD010512392GA SD010512400GA SD010522584GA SD010522601GA SD011051078SA SD011051090SA SD011051190SA SD011061294SA SD011061403SA SD011071651SA SD011071698SA SD011071859SA SD011071864SA SD011082172SA SD011143865SA SD011164072SA SD011164086SA SHQ11280796SA SHQ11290857SA SHQ11372782SA SHQ11443757SA SHQ11453796SA SHQ11453799SA SHQ11505358SA SHQ11505364SA SHQ12066022SA SHQ12066038SA SHQ12126661SA SHQ12156913SA SHQ12157038SA SHQ12177401SA SHQ12187542SA SHQ12217949SA SHQ12218048SA SHQ12228111SA SHQ12228275SA SHQ12228276SA SHQ12278980SA SD010355766GA SD010408306GA SD010449392GA SD010512386GA SD010512389GA SD010512395GA SD010512399GA SD011174174SA SD011174175SA SD011174178SA SD011174180SA SD011214885SA SD011224922SA SHQ11214777SA SHQ11260511SA SHQ11280655SA SHQ11280657SA SHQ11280672SA SHQ11280812SA SHQ11290904SA SHQ11301209SA SHQ11301210SA SHQ11301216SA SHQ11301218SA SHQ11301233SA SHQ11301286SA SHQ11301292SA SHQ11301293SA SHQ11311424SA SHQ11311431SA SHQ11311432SA SHQ11311433SA SHQ11311434SA SHQ11311439SA SHQ11311618SA SHQ11311623SA SHQ11311628SA SHQ11311632SA SHQ11311634SA SHQ11311635SA SHQ11311636SA SHQ11311637SA SHQ11311638SA SHQ11311639SA SHQ11311640SA SHQ11311641SA SHQ11311643SA SHQ11311644SA SHQ11311646SA SHQ11311649SA SHQ11321653SA SHQ11321842SA SHQ11321868SA SHQ11321872SA SHQ11321882SA SHQ11372771SA SHQ11423341SA SHQ11423343SA SHQ11423345SA SHQ11423347SA SHQ11423348SA SHQ11423350SA SHQ11423352SA SHQ11423354SA SHQ11423355SA SHQ11423358SA SHQ11423360SA SHQ11423361SA SHQ11443701SA SHQ11474510SA SHQ11474511SA SHQ11474512SA SHQ11474515SA SHQ11474516SA SHQ11474521SA SHQ11474544SA SHQ11474546SA SHQ11484822SA SHQ11484827SA SHQ11484828SA SHQ11484840SA SHQ11484845SA SHQ12015386SA SHQ12015388SA SHQ12035573SA SHQ12045748SA SHQ12056010SA SHQ12056013SA SHQ12066178SA SHQ12157050SA SHQ12167159SA SHQ12167163SA SHQ12197558SA SHQ12217896SA SHQ12217907SA SHQ12340891SA SHQ12330555SA SD010491674GA SD010272452GA SD010376556GA SD010376659GA SD010377061GA SD010387736GA SD010387742GA SD010408244GA SD010408245GA SD010408257GA SD010408264GA SD010449325GA SD010449326GA SD010449327GA SD010450070GA SD010450076GA SD010450078GA SD010450080GA SD010450082GA SD010450089GA SD010450090GA SD010459915GA SD010470581GA SD010481077GA SD010481139GA SD010491607GA SD010491608GA SD010491612GA SD010491613GA SD010491614GA SD010491616GA SD010491617GA SD010491618GA SD010491634GA SD010491635GA SD010491655GA SD010502103GA SD010502105GA SD010502108GA SD010502110GA SD010502111GA SD010502248GA SD010502251GA SD010502254GA SD010522467GA SD011033104GA SD011033109GA SD011033119GA SD011033121GA SD011073370GA SD011083579GA SD011093699GA SD011093701GA SD011093702GA SD011093703GA SD011093704GA SD011093706GA SD011093707GA SD011093708GA SD011093710GA SD011093729GA SD011103761GA SD011103765GA SD011103797GA SD011103800GA SD011103802GA SD011103803GA SD011113841GA SD011113873GA SD011113905GA SD011113906GA SD011113909GA SD011113911GA SD011113913GA SD011113914GA SD011124036GA SD011134319GA SD011174828GA SHQ11250170GA SHQ11290294GA SHQ11370876GA SHQ11370883GA SHQ11370884GA SHQ11391128GA SHQ11401230GA SHQ11401231GA SHQ11401232GA SHQ11411269GA SHQ11441407GA SHQ11441409GA SHQ11451438GA SHQ11451439GA SHQ11451454GA SHQ11491929GA SHQ11502071GA SHQ11502077GA SHQ11502078GA SHQ11502079GA SHQ11502080GA SHQ11502167GA SHQ11502169GA SHQ11502173GA SHQ11512233GA SHQ11512239GA SHQ12012344GA SHQ12042510GA SHQ12042513GA SHQ12042514GA SHQ12042518GA SHQ12102884GA SHQ12102885GA SHQ12102891GA SHQ12102893GA SHQ12173150GA SHQ12253665GA SHQ12253666GA SHQ12253669GA SHQ12333992GA SHQ12333969GA SHQ12333994GA SHQ12333991GA SHQ12303842GA SHQ12323912GA SHQ12313902GA SHQ12333963GA SHQ12333962GA SHQ12333964GA SHQ12333965GA SHQ12333966GA SHQ12333997GA SHQ12334003GA SHQ12344048GA SHQ12354077GA SHQ12333993GA SHQ12354053GA SHQ12354081GA SHQ12354075GA SHQ12354084GA SHQ12344047GA SHQ12344049GA SHQ12333990GA SHQ12334004GA SHQ12333999GA SHQ12333995GA SHQ12384172GA SHQ12384171GA SHQ12384150GA SHQ12384153GA SHQ12384151GA SHQ12384146GA SHQ12384165GA SHQ12384154GA SHQ12384148GA SHQ12384147GA SHQ12384152GA SHQ12384145GA SHQ12344046GA SHQ12384149GA SHQ12344045GA SHQ12333988GA SHQ12384162GA SHQ12384167GA SHQ12384164GA SHQ12384160GA SHQ12384174GA SHQ12384163GA SHQ12384161GA SHQ12384159GA SHQ12384158GA SHQ12384166GA SHQ12354082GA SHQ12333987GA SHQ12384173GA SHQ12384178GA SHQ12384175GA SHQ12384179GA SHQ12384170GA SHQ12384177GA SHQ12333989GA SHQ12384176GA SHQ12384190GA SHQ12384196GA SHQ12384195GA SD010251518GA SD010335149GA SD010345216GA SD010345325GA SD010345461GA SD010345469GA SD010345534GA SD010345551GA SD010387509GA SD010387621GA SD010387695GA SD010387707GA SD010387709GA SD010387712GA SD010387728GA SD010387828GA SD010398019GA SD010398022GA SD010398023GA SD010398024GA SD010398027GA SD010398028GA SD010398029GA SD010398031GA SD010398032GA SD010408096GA SD010408097GA SD010408103GA SD010408118GA SD010408120GA SD010408125GA SD010408331GA SD010408345GA SD010408351GA SD010408357GA SD010418703GA SD010449436GA SD010449446GA SD010459832GA SD010460104GA SHQ12096387SA SD010377121GA SD010387590GA SD010408246GA SD010429257GA SD010429268GA SD010430036SA SD010449372GA SD010449404GA SD010460146GAMKE SD010470689GA SD010470693GA SD010470696GA SD010470699GA SD010470715GA SD010481237GA SD010481240GA SD010491349GA SD010491501GA SD010491686GA SD010491690GA SD010491691GA SD010491692GA SD010491693GA SD010491697GA SD010491699GA SD010491703GA SD010491704GA SD010491707GA SD010491708GA SD010491709GA SD010500224SA SD010500234SA SD010500250SA SD010500256SA SD010500265SA SD010501725GA SD010501728GA SD010501738GA SD010501855GA SD010512326GA SD010512333GA SD010512334GA SD010512341GA SD010512342GA SD010512344GA SD010512347GA SD010512356GA SD010512358GA SD010512360GA SD010512362GA SD010512368GA SD010512369GA SD010512373GA SD010512379GA SD010512381GA SD010532878GA SD010532881GA SD010532883GA SD010532899GA SD011033124GA SD011033125GA SD011033127GA SD011033132GA SD011033137GA SD011033139GA SD011033141GA SD011051196SA SD011051197SA SD011051210SA SD011051212SA SD011051214SA SD011061413SA SD011061425SA SD011061459SA SD011061460SA SD011061468SA SD011061471SA SD011071579SA SD011071597SA SD011071602SA SD011071604SA SD011071607SA SD011071611SA SD011071618SA SD011083572GA SD011083601GA SD011083638GA SD011083660GA SD011083673GA SD011093689GA SD011093755GA SD011113978GA SD011123244SA SD011123246SA SD011123252SA SD011123256SA SD011123261SA SD011123273SA SD011123275SA SD011123291SA SD011124107GA SD011124115GA SD011124152GA SD011124153GA SD011124160GA SD011133415SA SD011133458SA SD011144398GA SD011144399GA SD011144400GA SD011144401GA SD011144403GA SD011144404GA SD011144406GA SD011144407GA SD011144408GA SD011144409GA SD011144411GA SD011144412GA SD011144414GA SD011144416GA SD011154513GA SD011154522GA SD011154529GA SD011154530GA SD011154617GA SD011154623GA SD011154629GA SD011174319SA SD011184855GA SD011184859GA SD011184872GA SD011184873GA SD011184874GA SD011184875GA SD011184877GA SD011184878GA SD011184880GA SD011184882GA SD011184883GA SD011184884GA SD011184885GA SD011184886GA SD011184887GA SD011184888GA SD011184889GA SD011194549SA SD011194944GA SD011194945GA SD011204964GA SD011204969GA SD011204970GA SD011204972GA SD011204974GA SD011204980GA SD011204986GA SD011214895SA SD011225013SA SD011225037SA SD011225042SA SD011225054SA SD011225055SA SD011225056SA SD011225058SA SHQ11280720SA SHQ11280806SA SHQ11301140SA SHQ11301141SA SHQ11301142SA SHQ11301143SA SHQ11301154SA SHQ11301167SA SHQ11301168SA SHQ11301191SA SHQ11301193SA SHQ11372785SA SHQ11403092SA SHQ11403095SA SHQ11403111SA SHQ11403112SA SHQ11403118SA SHQ11403119SA SHQ11403122SA SHQ11403123SA SHQ11403133SA SHQ11403136SA SHQ11403138SA SHQ11403143SA SHQ11403149SA SHQ11423174SA SHQ11423175SA SHQ11423178SA SHQ11423179SA SHQ11423180SA SHQ11423181SA SHQ11423183SA SHQ11423184SA SHQ11423185SA SHQ11423186SA SHQ11423188SA SHQ11423189SA SHQ11423190SA SHQ11423191SA SHQ11423193SA SHQ11423194SA SHQ11423195SA SHQ11423196SA SHQ11423197SA SHQ11423198SA SHQ11443596SA SHQ11443598SA SHQ11443599SA SHQ11443600SA SHQ11453802SA SHQ11453808SA SHQ11453816SA SHQ11453822SA SHQ11453824SA SHQ11453825SA SHQ11453834SA SHQ11453836SA SHQ11453839SA SHQ11453840SA SHQ11453843SA SHQ11453849SA SHQ11453975SA SHQ11454071SA SHQ11464383SA SHQ11464384SA SHQ11464385SA SHQ11464386SA SHQ11464387SA SHQ11464408SA SHQ11464413SA SHQ11484789SA SHQ11484823SA SHQ11484843SA SHQ11495073SA SHQ11495076SA SHQ11495077SA SHQ11495078SA SHQ11495079SA SHQ11495080SA SHQ11495081SA SHQ11495082SA SHQ11495083SA SHQ11495084SA SHQ11495085SA SHQ11495086SA SHQ11505315SA SHQ11505316SA SHQ11505317SA SHQ11505318SA SHQ11505321SA SHQ11505322SA SHQ11505323SA SHQ11505324SA SHQ12015475SA SHQ12015477SA SHQ12015478SA SHQ12015479SA SHQ12015481SA SHQ12045721SA SHQ12045725SA SHQ12045727SA SHQ12045728SA SHQ12045729SA SHQ12045731SA SHQ12045732SA SHQ12045733SA SHQ12045735SA SHQ12045736SA SHQ12045737SA SHQ12045739SA SHQ12045740SA SHQ12055931SA SHQ12055932SA SHQ12055933SA SHQ12055934SA SHQ12055935SA SHQ12055936SA SHQ12055937SA SHQ12055938SA SHQ12055939SA SHQ12055940SA SHQ12055941SA SHQ12055942SA SHQ12055943SA SHQ12055944SA SHQ12055945SA SHQ12055946SA SHQ12055947SA SHQ12055948SA SHQ12055949SA SHQ12055950SA SHQ12055990SA SHQ12055996SA SHQ12056012SA SHQ12056015SA SHQ12056016SA SHQ12056017SA SHQ12056018SA SHQ12056019SA SHQ12056020SA SHQ12056021SA SHQ12066042SA SHQ12066043SA SHQ12066044SA SHQ12066045SA SHQ12066046SA SHQ12066163SA SHQ12066164SA SHQ12066165SA SHQ12066166SA SHQ12066167SA SHQ12066168SA SHQ12066169SA SHQ12066171SA SHQ12066172SA SHQ12066173SA SHQ12066174SA SHQ12066175SA SHQ12066176SA SHQ12066177SA SHQ12106518SA SHQ12106519SA SHQ12106520SA SHQ12106521SA SHQ12106522SA SHQ12106523SA SHQ12106524SA SHQ12106526SA SHQ12106527SA SHQ12136794SA SHQ12136795SA SHQ12136796SA SHQ12136797SA SHQ12136798SA SHQ12136799SA SHQ12136800SA SHQ12136801SA SHQ12136802SA SHQ12136803SA SHQ12136804SA SHQ12136805SA SHQ12136806SA SHQ12136807SA SHQ12146810SA SHQ12146811SA SHQ12146812SA SHQ12146813SA SHQ12146814SA SHQ12146815SA SHQ12146816SA SHQ12146817SA SHQ12146818SA SHQ12146819SA SHQ12146820SA SHQ12146821SA SHQ12146822SA SHQ12146823SA SHQ12146824SA SHQ12197616SA SHQ12197617SA SHQ12197618SA SHQ12197695SA SHQ12197696SA SHQ12197697SA SHQ12197698SA SHQ12197699SA SHQ12197702SA SHQ12197703SA SHQ12197704SA SHQ12197705SA SHQ12197706SA SHQ12197708SA SHQ12197709SA SHQ12207756SA SHQ12207758SA SHQ12207762SA SHQ12207764SA SHQ12207766SA SHQ12207767SA SHQ12207768SA SHQ12207769SA SHQ12207770SA SHQ12218068SA SHQ12218069SA SHQ12218070SA SHQ12218074SA SHQ12218076SA SHQ12218078SA SHQ12218081SA SHQ12218083SA SHQ12218085SA SHQ12218088SA SHQ12218091SA SHQ12218092SA SHQ12218093SA SHQ12228250SA
SHQ12228251SA SHQ12228254SA SHQ12228256SA SHQ12228257SA SHQ12228260SA SHQ12228261SA SHQ12228262SA SHQ12228266SA SHQ12228267SA SHQ12228271SA SHQ12228274SA SHQ12278934SA SHQ12278935SA SHQ12319995SA SHQ12310048SA SHQ12310046SA SHQ12310039SA SHQ12319997SA SHQ12310052SA SHQ12330726SA SHQ12330804SA SHQ12330803SA SHQ12330724SA SHQ12330729SA SHQ12310002SA SHQ12310040SA SHQ12310045SA SHQ12310055SA SHQ12320273SA SHQ12320296SA SHQ12330801SA SHQ12330800SA SHQ12330727SA SHQ12330723SA SHQ12320289SA SHQ12330792SA SHQ12320291SA SHQ12330725SA SHQ12330721SA SHQ12330728SA SHQ12330802SA SHQ12320269SA SHQ12320285SA SHQ12320290SA SHQ12320271SA SHQ12289056SA SHQ12320288SA SHQ12320305SA SHQ12320298SA SHQ12320276SA SHQ12320284SA SHQ12320263SA SHQ12320304SA SHQ12320302SA SHQ12320297SA SHQ12361280SA SHQ12361284SA SHQ12351124SA SD010366468GA SD010303496GA SD010334916GA SD010334920GA SD010334924GA SD010334925GA SD010334926GA SD010345444GA SD010365930GA SD010365987GA SD010366113GA SD010366167GA SD010366182GA SD010366308GA SD010366323GA SD010366426GA SD010376567GA SD010376570GA SD010376574GA SD010376575GA SD010376576GA SD010376587GA SD010376594GA SD010376784GA SD010376823GA SD010376954GA SD010377127GA SD010387265GA SD010387268GA SD010387276GA SD010387411GA SD010387503GA SD010387504GA SD010387508GA SD010387512GA SD010387513GA SD010387522GA SD010387523GA SD010387526GA SD010387534GA SD010387741GA SD010387843GA SD010397991GA SD010398004GA SD010408066GA SD010408073GA SD010408076GA SD010408084GA SD010408098GA SD010408139GA SD010408169GA SD010408241GA SD010408243GA SD010408359GA SD010408382GA SD010408386GA SD010408389GA SD010408443GA SD010418515GA SD010418520GA SD010418530GA SD010418537GA SD010418646GA SD010418648GA SD010418743GA SD010418854GA SD010418855GA SD010418860GA SD010418861GA SD010418866GA SD010418903GA SD010429109GA SD010429111GA SD010429114GA SD010429167GA SD010429274GA SD010430026SA SD010449683GA SD010449684GA SD010449687GA SD010449702GA SD010450019GA SD010450022GA SD010450035GA SD010459739GA SD010459855GA SD010459886GA SD010459926GA SD010459930GA SD010459937GA SD010459944GA SD010459945GA SD010460093GA SD010460278GA SD010460293GA SD010460294GA SD010460296GA SD010460297GA SD010460298GA SD010460299GA SD010460302GA SD010460303GA SD010460304GA SD010460305GA SD010470398GA SD010470399GA SD010470741GA SD010470754GA SD010470762GA SD010480113SA SD010480883GA SD010480901GA SD010480902GA SD010480908GA SD010480944GA SD010480946GA SD010480947GA SD010481069GA SD010481070GA SD010481076GA SD010481086GA SD010481279GA SD010481289GA SD010491311GA SD010491320GA SD010491329GA SD010491336GA SD010491338GA SD010491340GA SD010491397GA SD010491400GA SD010491401GA SD010491405GA SD010491448GA SD010491451GA SD010491452GA SD010491458GA SD010491459GA SD010491500GA SD010491506GA SD010491508GA SD010491622GA SD010491626GA SD010491648GA SD010501736GA SD010501748GA SD010501771GA SD010501772GA SD010501774GA SD010501777GA SD010501780GA SD010501791GA SD010501971GA SD010502089GA SD010502127GA SD010502136GA SD010502249GA SD010512330GA SD010512345GA SD010512349GA SD010512350GA SD010512351GA SD010512357GA SD010512365GA SD010512377GA SD010522476GA SD010522480GA SD010522487GA SD010522490GA SD010522634GA SD010522644GA SD010522646GA SD010532760GA SD010532762GA SD010532777GA SD010532834GA SD011020493SA SD011020503SA SD011030572SA SD011030573SA SD011030587SA SD011030606SA SD011032927GA SD011033105GA SD011033115GA SD011033116GA SD011033117GA SD011050925SA SD011051015SA SD011051023SA SD011051031SA SD011051051SA SD011051054SA SD011051055SA SD011051059SA SD011051062SA SD011051064SA SD011051072SA SD011051073SA SD011051079SA SD011051094SA SD011051120SA SD011051121SA SD011051122SA SD011051123SA SD011051125SA SD011051127SA SD011053267GA SD011061222SA SD011061235SA SD011061260SA SD011061344SA SD011061370SA SD011061385SA SD011061428SA SD011071525SA SD011071527SA SD011071529SA SD011071542SA SD011071543SA SD011071566SA SD011071819SA SD011071828SA SD011071834SA SD011071909SA SD011071911SA SD011071964SA SD011071988SA SD011071996SA SD011071997SA SD011072001SA SD011072002SA SD011072008SA SD011072011SA SD011072014SA SD011072017SA SD011072019SA SD011072027SA SD011072031SA SD011072129SA SD011072133SA SD011072155SA SD011072159SA SD011082242SA SD011082267SA SD011082296SA SD011092404SA SD011092454SA SD011092460SA SD011092464SA SD011092468SA SD011092470SA SD011092472SA SD011092473SA SD011092475SA SD011092476SA SD011092477SA SD011092480SA SD011092481SA SD011092482SA SD011092484SA SD011092485SA SD011092486SA SD011092488SA SD011112585SA SD011112631SA SD011112641SA SD011112655SA SD011112761SA SD011112774SA SD011112785SA SD011112795SA SD011112800SA SD011112802SA SD011122836SA SD011122881SA SD011122883SA SD011122967SA SD011122982SA SD011122986SA SD011123015SA SD011123095SA SD011123110SA SD011123132SA SD011123140SA SD011123175SA SD011123178SA SD011123268SA SD011123299SA SD011123312SA SD011123317SA SD011123319SA SD011123326SA SD011123330SA SD011123341SA SD011133376SA SD011133383SA SD011133390SA SD011133392SA SD011133395SA SD011133396SA SD011133403SA SD011133434SA SD011133477SA SD011133481SA SD011133533SA SD011133535SA SD011133537SA SD011133648SA SD011133658SA SD011133659SA SD011133701SA SD011133707SA SD011143810SA SD011143813SA SD011143818SA SD011143822SA SD011143877SA SD011143882SA SD011143885SA SD011143886SA SD011143889SA SD011143890SA SD011143896SA SD011143899SA SD011143903SA SD011143904SA SD011143920SA SD011143922SA SD011143928SA SD011143939SA SD011143940SA SD011143952SA SD011143953SA SD011143954SA SD011143956SA SD011153999SA SD011164111SA SD011164117SA SD011174231SA SD011174232SA SD011174236SA SD011174237SA SD011194604SA SD011204671SA SD011204686SA SD011204691SA SD011214907SA SD011214913SA SD011214914SA SD011214916SA SD011214918SA SD011225069SA SD011225070SA SHQ11230056SA SHQ11230058SA SHQ11230060SA SHQ11230121SA SHQ11235139SA SHQ11240228SA SHQ11240235SA SHQ11250305SA SHQ11250307SA SHQ11250388SA SHQ11260507SA SHQ11280559SA SHQ11280624SA SHQ11280629SA SHQ11280630SA SHQ11280682SA SHQ11280707SA SHQ11280727SA SHQ11280731SA SHQ11280747SA SHQ11280757SA SHQ11311566SA SHQ11321760SA SHQ11321762SA SHQ11321864SA SHQ11321866SA SHQ11321867SA SHQ11321875SA SHQ11331990SA SHQ11332115SA SHQ11342155SA SHQ11342162SA SHQ11342166SA SHQ11342167SA SHQ11342174SA SHQ11342182SA SHQ11342194SA SHQ11352405SA SHQ11352409SA SHQ11362517SA SHQ11362518SA SHQ11362519SA SHQ11362522SA SHQ11362523SA SHQ11362524SA SHQ11362525SA SHQ11362526SA SHQ11362527SA SHQ11372838SA SHQ11403124SA SHQ11403145SA SHQ11433401SA SHQ11433472SA SHQ11433475SA SHQ11443538SA SHQ11443542SA SHQ11443583SA SHQ11443620SA SHQ11443628SA SHQ11443633SA SHQ11443639SA SHQ11453792SA SHQ11453805SA SHQ11453810SA SHQ11453811SA SHQ11453965SA SHQ11453967SA SHQ11454073SA SHQ11464087SA SHQ11464306SA SHQ11464309SA SHQ11464311SA SHQ11464312SA SHQ11464326SA SHQ11464356SA SHQ11464405SA SHQ11474508SA SHQ11474509SA SHQ11474528SA SHQ11474530SA SHQ11474532SA SHQ11474534SA SHQ11474617SA SHQ11474626SA SHQ11484651SA SHQ11484659SA SHQ11484927SA SHQ11495091SA SHQ12015426SA SHQ12015429SA SHQ12015431SA SHQ12015436SA SHQ12015454SA SHQ12015456SA SHQ12015457SA SHQ12025519SA SHQ12035582SA SHQ12045743SA SHQ12045744SA SHQ12045745SA SHQ12045747SA SHQ12045749SA SHQ12045750SA SHQ12045829SA SHQ12045874SA SHQ12055921SA SHQ12066024SA SHQ12066025SA SHQ12066030SA SHQ12066031SA SHQ12066032SA SHQ12066033SA SHQ12066127SA SHQ12066179SA SHQ12076209SA SHQ12076230SA SHQ12076231SA SHQ12076232SA SHQ12076233SA SHQ12076246SA SHQ12076248SA SHQ12076249SA SHQ12076263SA SHQ12076269SA SHQ12096416SA SHQ12096418SA SHQ12096420SA SHQ12096422SA SHQ12096424SA SHQ12096425SA SHQ12096429SA SHQ12096431SA SHQ12096432SA SHQ12096433SA SHQ12096434SA SHQ12096435SA SHQ12096436SA SHQ12096437SA SHQ12096437SABOP SHQ12096438SA SHQ12096439SA SHQ12096443SA SHQ12096444SA SHQ12096445SA SHQ12096446SA SHQ12106536SA SHQ12106538SA SHQ12106539SA SHQ12106544SA SHQ12106547SA SHQ12106549SA SHQ12106549SABOP SHQ12106550SA SHQ12106551SA SHQ12116556SA SHQ12116559SA SHQ12126647SA SHQ12126662SA SHQ12126664SA SHQ12126668SA SHQ12146838SA SHQ12157002SA SHQ12167160SA SHQ12177237SA SHQ12177239SA SHQ12177241SA SHQ12177243SA SHQ12177245SA SHQ12177250SA SHQ12177252SA SHQ12177265SA SHQ12177281SA SHQ12187488SA SHQ12187491SA SHQ12187500SA SHQ12207812SA SHQ12207816SA SHQ12207826SA SHQ12217877SA SHQ12217882SA SHQ12217887SA SHQ12217890SA SHQ12217892SA SHQ12217893SA SHQ12217895SA SHQ12217897SA SHQ12217900SA SHQ12217905SA SHQ12217915SA SHQ12228137SA SHQ12228186SA SHQ12228242SA SHQ12228243SA SHQ12238297SA SHQ12238315SA SHQ12238322SA SHQ12238326SA SHQ12238382SA SHQ12238385SA SHQ12238386SA SHQ12238387SA SHQ12238394SA SHQ12238396SA SHQ12238407SA SHQ12238411SA SHQ12238412SA SHQ12238414SA SHQ12248577SA SHQ12248578SA SHQ12248580SA SHQ12248581SA SHQ12248583SA SHQ12248585SA SHQ12248586SA SHQ12248587SA SHQ12248588SA SHQ12258793SA SHQ12258825SA SHQ12258827SA SHQ12258830SA SHQ12299312SA SHQ12299316SA SHQ12299322SA SHQ12299323SA SHQ12299325SA SHQ12340941SA SHQ12340929SA SHQ12340932SA SHQ12340930SA SHQ12340935SA SHQ12340925SA SHQ12340931SA SHQ12340933SA SHQ12340936SA SHQ12340937SA SHQ12340926SA SHQ12340928SA SHQ12340927SA SHQ12361151SA SHQ12371502SA SHQ12371480SA SD009441172GA SD009534362GA SD01048386GA SD01040876GA SD010116605GAFRE SHQ1206030SA SHQ1223831DSA SHQ1228385SA SHQ1228387SA SHQ12299285SA SHQ12309561SA SD010303256GA SD010387499GA SD010429262GA SD010449579GA SD010449665GA SD010459794GA
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
800-558-7044
|
Manufacturer Reason for Recall |
GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module.
ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
GE Healthcare "Urgent Medical Device Correction" letter dated April 3, 2013 to all affected customers. The letter was addressed to Healthcare Administrator / Risk manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Product, Safety Instructions, Affected Product Details, and Product Correction. For questions contact Technical Support at 1-800-558-7044. |
Quantity in Commerce |
28,120 *14,709 US & 13,411 OUS) |
Distribution |
Worldwide Distribution - USA Nationwide including DC and PR except ND and the countries of: ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, CHANA, GREECE, HONDURAS, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, NEPAL, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAK IA, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TUNSIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, BOLIVIA, HONG KONG, JAMAICA, PHILIPPINES, SOUTH KOREA, TRINIDAD TOBAGO, URUGUAY, VENEZUELA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
|
|
|
|