Class 2 Device Recall Zimmer Periarticular Plating System |
|
Date Initiated by Firm |
November 05, 2013 |
Date Posted |
December 27, 2013 |
Recall Status1 |
Terminated 3 on May 05, 2015 |
Recall Number |
Z-0591-2014 |
Recall Event ID |
66499 |
510(K)Number |
K082078
|
Product Classification |
Plate, fixation, bone - Product Code HRS
|
Product |
FEM COND BUTTRESS PLT, LT |
Code Information |
Item #'s: 00-2347-002-12 00-2347-002-14 00-2347-002-16 00-2347-002-18 Lot #s: All lots with an expiration date prior to July 24, 2023 |
Recalling Firm/ Manufacturer |
Zimmer, Inc. 345 E Main St Warsaw IN 46580-2746
|
For Additional Information Contact |
800-613-6131
|
Manufacturer Reason for Recall |
Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure
|
FDA Determined Cause 2 |
Process change control |
Action |
On 11/5/2013, Urgent Medical Device Recall Notifications were sent to affected distributors, hospital Risk Managers, and doctors informing them of the failure.
All distributors were notified via electronic mail. Hospital risk managers, surgeons and distributors with affected inventory within the expiration date of the device will also be notified via courier. The letter identifies the issue, health risks, and their responsibilities. Questions concerning the recall are directed to 1-877-946-2761
Communications outside of the United States will occur approximately two weeks after the United States communications.
Zimmer will conduct effectiveness checks in the following way:
Distributors/Hospitals/Surgeons
100% of the notifications will be tracked to ensure delivery of the notifications.
An inventory certification form will need to be returned from each distributor that certifies that all available affected product from their territory has been returned.
Additional notifications via email and FedEx will be sent to those who have not completed the required certification form or returned affected product.
Accounts will be deemed unresponsive after 3 attempts. |
Quantity in Commerce |
42, 065 units total |
Distribution |
Worldwide Distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HRS and Original Applicant = ZIMMER, INC.
|
|
|
|