• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Verigene GramPositive Blood Culture Nucleic Acid Test (BCGP)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Verigene GramPositive Blood Culture Nucleic Acid Test (BCGP)see related information
Date Initiated by FirmJune 17, 2013
Date PostedSeptember 30, 2014
Recall Status1 Terminated 3 on March 14, 2016
Recall NumberZ-2735-2014
Recall Event ID 68014
510(K)NumberK122514 
Product Classification Gram-positive bacteria and their resistance markers - Product Code PAM
ProductVerigene BC-GP Kit (Catalog number 20-005-018); 1) 20 Verigene BC-GP Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous detection and identification of selected gram-positive (BC-GP) and gram-negative (BC-GN) bacteria and resistance markers.
Code Information Part Number: 20-009-018; Lot Numbers: 051713018B, 052413018C
Recalling Firm/
Manufacturer
Nanosphere, Inc.
4088 Commercial Ave
Northbrook IL 60062-1829
For Additional Information Contact
847-400-9115
Manufacturer Reason
for Recall
There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.
FDA Determined
Cause 2
Process change control
ActionUndated recall notification letters were sent to all direct accounts (customers) June 17 - 19, 2013. The letters included instructions for customers to: 1) check their inventory for the recalled products; 2) set any recalled products aside for return; and, 3) contact Nanosphere to obtain replacement tip holders. Customers with questions about this recall can contact Nanosphere via e-mail at technicalsupport@nanosphere.us or by calling 888-837-4436.
Quantity in Commerce1,660 extraction trays
DistributionWorldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = PAM
-
-