• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Codman Cetras

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Codman Cetras see related information
Date Initiated by Firm July 03, 2014
Date Posted August 12, 2014
Recall Status1 Terminated 3 on January 30, 2017
Recall Number Z-2189-2014
Recall Event ID 68837
510(K)Number K112156  K113526  
Product Classification Shunt, central nervous system and components - Product Code JXG
Product Codman Certas - In Line Valve with SIPHONGUARD¿ Device, Catheter and Accessories; Product Code: 82-8805

The Codman Certas Programmable valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus.
Code Information Lot Numbers: CMBC2Y, CMDCN9, CMJC48, CMLC6D, CMNBFF, CMNCDY, CNCB6M, CNCCLB, CNCCZ4, CNDBRG, CNDCNB, CNHC8M, CNJC4P, CNKCL4, CNLCL1, CNMB8Y, CNMDMO, CNNDHY, CNPCLZ, CPBC99, CPBCRZ, CPCC27, CPDBYJ, CPDCKL 
Recalling Firm/
Manufacturer
Codman & Shurtleff, Inc.
325 Paramount Drive
Raynham MA 02767
For Additional Information Contact
866-685-7325
Manufacturer Reason
for Recall
To clarify the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU).
FDA Determined
Cause 2
Device Design
Action Codman Neuro sent an Urgent Medical Device Safety Notification dated July 17, 2014, to clinicians who use the device. The notification clarifies the CODMAN CERTAS Valve virtual off (setting 8) pressure specification outlined in the product's Instructions for Use (IFU). Consignees are also asked to complete the enclosed Acknowledgement Form and fax the completed form to 1-888-239-1305. For questions or concerns regarding this notification, consignees are instructed to contact their local Codman Neuro Representative or Scientific and Medical Affairs at SciMedAffairs@its.jnj.com or (866) 685-7325. For questions regarding this recall call 508-828-2726.
Quantity in Commerce 9,498 devices total
Distribution Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY; and Internationally to: Canada, Australia, Austria, Belgium, Brazil, Czech Republic, Denmark, Egypt, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Kuwait, Lebanon, Luxembourg, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JXG and Original Applicant = MEDOS INTERNATIONAL SARL
-
-