  Class 2 Device Recall Nanosphere see related information
Date Initiated by Firm October 10, 2013
Date Posted September 23, 2014
Recall Status1 Terminated 3 on June 16, 2015
Recall Number Z-2682-2014
Recall Event ID 69034
510(K)Number K123197  
Product Classification C.Difficile toxin gene amplification assay - Product Code OZN
Product The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic strains Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.
Code Information Product Number: 20-009-022; Lot Number and Expiration Date: Lot: 100313022B, Expiration Date: 04/01/14
Recalling Firm/
Manufacturer
Nanosphere, Inc.
4088 Commercial Ave
Northbrook IL 60062-1829
For Additional Information Contact Customer Support
888-837-4436 Ext. 3
Manufacturer Reason
for Recall
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
FDA Determined
Cause 2
Nonconforming Material/Component
Action Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.
Quantity in Commerce 80 units
Distribution Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = OZN and Original Applicant = NANOSPHERE, INC
-
-